throbber
The possibilities are endless.
`
`Product Catalog
`
`w w w . g u i d a n t . c o m
`
`
`Page 1
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`GUIDANT CORONARY STENT SYSTEMS
`
`CORONARY GUIDE WIRES
`
`CORONARY GUIDING CATHETERS
`
`CORONARY DILATATION CATHETERS
`
`ACCESSORIES
`
`DIRECTIONAL CORONARY ATHERECTOMY
`
`C U S T O M E R S E R V I C E 8 0 0 2 2 7 - 9 9 0 2
`
`
`Page 2
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`O R D E R I N G I N F O R M A T I O N
`
`R e t u r n P o l i c y
`1. Guidant Corporation’s Vascular Intervention division will accept return of any undamaged and
`unused standard catalog item within 90 days of invoice date. Defective product can be returned
`at any time, with an Authorized Return Goods Number (RGA).
`2. Returned product is subject to examination for acceptable condition or alleged defects.
`3. Acceptable returned product will be credited to customer account subject to a restocking fee.
`4. A Returned Goods Authorization Number is required for all returns and replacements.
`This number can be obtained by calling your account representative, or Customer Service at
`800 227-9902.
`5. The RGA Number must be clearly marked on the outside of all boxes returned. A copy of the
`packing slip and/or Purchase Order Number should be included with the merchandise
`6. Send non-complaint returns to:
`42301 Zevo Drive
`Suite E
`Temecula, CA 92590
`
`A biohazard box is required for all complaint returns. This box can be obtained by calling your
`account representative, or Customer Service.
`
`Send complaint returns to:
`26531 Ynez Road
`Temecula, CA 92591
`
`S h i p p i n g P o l i c y
`Guidant Corporation’s Vascular Intervention division offers a variety of shipping options, including
`overnight delivery. All shipment charges are paid by Guidant. Please specify your shipping preference
`when placing your order with Guidant Customer Service Representatives.
`
`WARNING: Manufactured with CFC-113, a substance which harms public health and environment by
`destroying ozone in the upper atmosphere.
`
`Guidant Corporation’s Vascular Intervention division is diligently working to eliminate the use of
`CFC-113 in the manufacturing process.
`
`
`Page 3
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`CORONARY STENT SYSTEMS
`
`
`Page 4
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`Small Vessel
`
`Small vessel stents
`
`combine low profiles
`
`with a flexible stent
`
`design specifically for
`
`treatment of vessels
`
`that are 2.0 mm
`
`to 2.5 mm in size.
`
`AT T R I B U T E S
`• Thinner struts for increased
`flexibility and larger lumens
`
`• Low system profiles for excellent
`deliverability and crossability
`
`• Noncompliant balloon for
`controlled growth
`
`*ML PIXELTM is indicated for patients presenting
`with abrupt or threatened abrupt closure.
`
`Workhorse
`
`The flexibility,
`
`conformability and
`
`deliverability of
`
`workhorse stents
`
`make them ideally
`
`suited for the majority
`
`of cases.
`
`AT T R I B U T E S
`• Low profiles for easy lesion access,
`enhanced track and stent recross
`
`• Clinically proven stent pattern
`designed to maintain vessel shape
`in tortuous anatomies while providing
`optimal scaffolding
`
`• Minimal balloon overhang to reduce
`balloon interaction outside of the
`stent upon expansion
`
`Large Vessel
`
`AT T R I B U T E S
`
`• Low unsupported surface area and
`high metal-to-artery ratio for excellent
`scaffolding in large native vessels and
`saphenous vein grafts
`
`• High radial strength to minimize recoil
`
`The scaffolding and
`radial strength provided
`by large vessel stents
`make them a superb
`fit for large native
`coronary arteries and
`saphenous vein grafts.
`
`
`Page 5
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`C O R O N A R Y S T E N T
`
`S Y S T E M S
`
`GUIDANT CORONARY STENT SYSTEMS - Cobalt Chromium
`SMALL VESSEL
`MULTI-LINK MINI VISION™ OTW Coronary Stent System
`
`MULTI-LINK MINI VISION™ RX Coronary Stent System
`WORKHORSE
`MULTI-LINK VISION® OTW Coronary Stent System
`
`MULTI-LINK VISION® RX Coronary Stent System
`
`GUIDANT CORONARY STENT SYSTEMS - Stainless Steel
`SMALL VESSEL
`MULTI-LINK OTW PIXEL® Coronary Stent System*
`
`MULTI-LINK RX PIXEL® Coronary Stent System*
`WORKHORSE
`MULTI-LINK OTW ZETA® Coronary Stent System
`
`MULTI-LINK RX ZETA® Coronary Stent System
`LARGE VESSEL
`MULTI-LINK OTW ULTRA™ Coronary Stent System
`
`MULTI-LINK RX ULTRA™ Coronary Stent System
`
`*ML PIXEL® is indicated for patients presenting with abrupt or threatened abrupt closure.
`
`
`
`C U S T O M E R S E R V I C E 8 0 0 2 2 7 - 9 9 0 2C U S T O M E R S E R V I C E 8 0 0 2 2 7 - 9 9 0 2
`
`
`Page 6
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`C O R O N A R Y S T E N T
`
`S Y S T E M S
`
`.014" maximum
`guide wire
`diameter
`
`List
`Price
`($)
`
`$2,150
`$2,150
`$2,150
`
`$2,200
`$2,200
`$2,200
`
`$2,225
`$2,225
`$2,225
`
`$2,260
`$2,260
`$2,260
`
`$2,410
`$2,410
`$2,410
`
`$2,510
`$2,510
`$2,510
`
`Cobalt chromium technology allows for thin
`struts for excellent outcomes without
`compromising strength or radiopacity.
`
`Flexible low profile design
`
`GUIDANT MULTI-LINK MINI VISION™ OTW
`C o r o n a r y S t e n t S y s t e m
`
`Stock
`Number
`
`UPN
`Number
`
`1010152-08
`1010153-08
`1010154-08
`
`00802526331008
`00802526331060
`00802526331121
`
`1010152-12
`1010153-12
`1010154-12
`
`00802526331015
`00802526331077
`00802526331138
`
`1010152-15
`1010153-15
`1010154-15
`
`1010152-18
`1010153-18
`1010154-18
`
`1010152-23
`1010153-23
`1010154-23
`
`00802526331022
`00802526331084
`00802526331145
`
`00802526331039
`00802526331091
`00802526331152
`
`00802526331046
`00802526331107
`00802526331169
`
`Stent Sizes
`Available
`(mm)
`
`Stent Length
`(mm)
`
`Stent
`Nominal
`Pressure
`(atm)
`
`Rated Burst
`Pressure
`(atm)
`
`2.00
`2.25
`2.50
`
`2.00
`2.25
`2.50
`
`2.00
`2.25
`2.50
`
`2.00
`2.25
`2.50
`
`2.00
`2.25
`2.50
`
`8
`8
`8
`
`12
`12
`12
`
`15
`15
`15
`
`18
`18
`18
`
`23
`23
`23
`
`8
`9
`9
`
`8
`9
`9
`
`8
`9
`9
`
`8
`9
`9
`
`8
`9
`9
`
`16
`16
`16
`
`16
`16
`16
`
`16
`16
`16
`
`16
`16
`16
`
`16
`16
`16
`
`16
`8
`28
`2.00
`00802526331053
`1010152-28
`16
`9
`28
`2.25
`00802526331114
`1010153-28
`16
`9
`28
`2.50
`00802526331176
`1010154-28
`Minimum I.D. of Guiding Catheter = .056" Maximum Guide Wire = .014"
`
`
`
`C U S T O M E R S E R V I C E 8 0 0 2 2 7 - 9 9 0 2C U S T O M E R S E R V I C E 8 0 0 2 2 7 - 9 9 0 2
`
`
`Page 7
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`MULTI-LINK MINI VISION™ Coronary Stent System
`
`INDICATIONS
`The Guidant MULTI-LINK MINI VISION™ RX and Guidant MULTI-LINK MINI VISION™ OTW Coronary Stent Systems are
`indicated for improving coronary luminal diameter in patients with abrupt or threatened abrupt closure with failed
`interventional therapy of de novoand restenotic native coronary artery lesions (length ≤ 25 mm) with reference vessel
`diameters from 2.0 to 2.5 mm.
`
`CONTRAINDICATIONS
`The Guidant MULTI-LINK MINI VISION™ Coronary Stent Systems are contraindicated for use in:
`• Patients in whom anti-platelet and / or anti-coagulant therapy is contraindicated.
`• Patients judged to have a lesion that prevents complete inflation of an angioplasty balloon.
`
`WARNINGS AND PRECAUTIONS
`Also see Individualization of Treatment in the IFU.
`
`WARNINGS
`Long term outcome for this permanent implant is unknown at present.
`• Judicious selection of patients is necessary since the use of this device carries the associated risk of subacute thrombosis,
`vascular complications and / or bleeding events.
`• Persons allergic to L-605 CoCr alloy may suffer an allergic reaction to this implant.
`• Implantation of the stent should be performed only by physicians who have received appropriate training.
`• Stent placement should only be performed at hospitals where emergency coronary artery bypass graft surgery can be
`readily performed.
`• Subsequent restenosis may require repeat dilatation of the arterial segment containing the stent. The long-term outcome
`following repeat dilatation of endothelialized stents is unknown at present.
`• When multiple stents are required, stent materials should be of similar composition. A combination of 316L stainless steel
`and L-605 CoCr alloy is acceptable should this be required.
`
`Stent Handling - Precautions
`• For single use only. Do not resterilize or reuse. Note the product "Use By" date.
`• Do not remove the stent from its Delivery System as removal may damage the stent and / or lead to stent embolization.
`Stent system is intended to perform together as a system.
`• Delivery System should not be used in conjunction with other stents.
`• Special care must be taken not to handle or in any way disrupt the stent on the balloon. This is most important during
`catheter removal from packaging, placement over guide wire and advancement through rotating hemostatic valve adapter
`and guiding catheter hub.
`• Do not manipulate (e.g., "roll") the stent with your fingers, as this action may loosen the stent from the delivery balloon.
`• Use only the recommended balloon inflation medium. Never use air or any gaseous medium to inflate the balloon, as this
`may cause uneven expansion and difficulty in deployment of the stent.
`
`Stent Placement - Precautions
`• Do not prepare or pre-inflate Delivery System prior to stent deployment other than as directed. Use balloon purging
`technique described in Delivery System Preparation.
`• Implanting a stent may lead to dissection of the vessel distal and / or proximal to the stent and may cause acute closure of
`the vessel requiring additional intervention (CABG, further dilatation, placement of additional stents, or other).
`• When treating multiple lesions, stent the distal lesion prior to stenting the proximal lesion. Stenting in this order obviates
`the need to cross the proximal stent in placement of the distal stent, and reduces the chance of dislodging the proximal
`stent.
`• Do not expand the stent if it is not properly positioned in the vessel. (See Stent / System Removal – Precautions.)
`• Placement of a stent has the potential to compromise side branch patency.
`• Do not exceed the Rated Burst Pressure as indicated on the product label. Monitor balloon pressures during inflation.
`Use of pressures higher than specified on product label may result in a ruptured balloon with possible intimal damage and
`dissection.
`• An unexpanded stent may be retracted into the guiding catheter one time only. Subsequent movement in and out through
`the distal end of the guiding catheter should not be performed as the stent may be damaged when retracting the
`undeployed stent back into the guiding catheter. Should any resistance be felt at any time during withdrawal of the
`Coronary Stent System, the entire system should be removed as a single unit.
`• Stent retrieval methods (use of additional wires, snares and / or forceps) may result in additional trauma to the coronary
`vasculature and / or the vascular access site. Complications may include bleeding, hematoma or pseudoaneurysm.
`
`Stent / System Removal - Precautions
`Should any resistance be felt at any time during either lesion access or removal of the Delivery System post-stent
`implantation, the entire system should be removed as a single unit.
`
`When removing the Delivery System as a single unit:
`• DO NOT retract the Delivery System into the guiding catheter.
`• Position the proximal balloon marker just distal to the tip of the guiding catheter.
`• Advance the guide wire into the coronary anatomy as far distally as safely possible.
`• Tighten the rotating hemostatic valve to secure the Delivery System to the guiding catheter; then remove the guiding
`catheter and Delivery System as a single unit.
`Failure to follow these steps and / or applying excessive force to the Delivery System can potentially result in loss or damage
`to the stent and / or Delivery System components.
`
`If it is necessary to retain guide wire position for subsequent artery / lesion access, leave the guide wire in place and remove
`all other system components.
`
`Post Implant - Precautions
`When crossing a newly deployed stent with a guide wire, balloon or Delivery System, exercise care to avoid disrupting the
`stent geometry.
`
`MRI (Magnetic Resonance Imaging) Statement
`The Guidant MULTI-LINK MINI VISION™ stent has been shown in non-clinical testing to be MRI safe immediately following
`implantation. MRI test conditions used to evaluate this stent were: for magnetic field interactions, a static magnetic field
`strength of 3 tesla with a maximum spatial gradient magnetic field of 3.3 tesla/meter; for MRI-related heating, a maximum
`whole body averaged specific absorption rate (SAR) of 2.0 W/kg for 15 minutes of MR imaging. While a single stent
`produced a temperature rise of less than 0.6°C and should not migrate under these conditions, the response of overlapping
`stents or stents with fractured struts is unknown. Non-clinical testing has not been performed to rule out the possibility of
`stent migration at field strengths higher than 3 tesla. MR image quality may be compromised if the area of interest is in the
`exact same area or relatively close to the position of the stent.
`
`ADVERSE EVENTS
`Potential Adverse Events
`Adverse events may be associated with the use of a coronary stent in native coronary arteries:
`• Acute myocardial infarction
`• Allergic reaction to contrast
`• Arterial perforation
`• Arterial rupture
`• Arteriovenous fistula
`• Cardiac arrhythmias
`• Bleeding complications (including transfusions)
`• Coronary spasm
`• Death
`• Dissection of the coronary artery
`• Drug reaction to anti-platelet agent
`• Emergency or non-emergent coronary artery bypass graft surgery
`• Entry site complications
`• Hypotension / hypertension
`• Infection
`• Injury to the coronary artery
`• Ischemia
`• Pseudoaneurysm
`• Restenosis of the stented segment
`• Stent embolization
`• Stent thrombosis / emboli
`• Stroke / cerebrovascular accident
`• Total occlusion of the coronary artery
`• Unstable angina pectoris
`• Vascular complications
`
`C U S T O M E R S E R V I C E 8 0 0 2 2 7 - 9 9 0 2
`
`
`Page 8
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`C O R O N A R Y S T E N T
`
`S Y S T E M S
`
`.014" maximum
`guide wire
`diameter
`
`List
`Price
`($)
`
`$2,150
`$2,150
`$2,150
`
`$2,200
`$2,200
`$2,200
`
`$2,225
`$2,225
`$2,225
`
`$2,260
`$2,260
`$2,260
`
`$2,410
`$2,410
`$2,410
`
`$2,510
`$2,510
`$2,510
`
`Cobalt chromium technology allows for thin
`struts for excellent outcomes without
`compromising strength or radiopacity.
`
`Flexible low profile design
`
`GUIDANT MULTI-LINK MINI VISION™ RX
`C o r o n a r y S t e n t S y s t e m
`
`Stock
`Number
`
`1007821-08
`1007822-08
`1007823-08
`
`1007821-12
`1007822-12
`1007823-12
`
`1007821-15
`1007822-15
`1007823-15
`
`1007821-18
`1007822-18
`1007823-18
`
`1007821-23
`1007822-23
`1007823-23
`
`UPN
`Number
`
`00802526365553
`00802526365614
`00802526365676
`
`00802526365560
`00802526365621
`00802526365683
`
`00802526365577
`00802526365638
`00802526365690
`
`00802526365584
`00802526365645
`00802526365706
`
`00802526365591
`00802526365652
`00802526365713
`
`Stent Sizes
`Available
`(mm)
`
`Stent Length
`(mm)
`
`Stent
`Nominal
`Pressure
`(atm)
`
`Rated Burst
`Pressure
`(atm)
`
`2.00
`2.25
`2.50
`
`2.00
`2.25
`2.50
`
`2.00
`2.25
`2.50
`
`2.00
`2.25
`2.50
`
`2.00
`2.25
`2.50
`
`8
`8
`8
`
`12
`12
`12
`
`15
`15
`15
`
`18
`18
`18
`
`23
`23
`23
`
`8
`9
`9
`
`8
`9
`9
`
`8
`9
`9
`
`8
`9
`9
`
`8
`9
`9
`
`16
`16
`16
`
`16
`16
`16
`
`16
`16
`16
`
`16
`16
`16
`
`16
`16
`16
`
`16
`8
`28
`2.00
`00802526365607
`1007821-28
`16
`9
`28
`2.25
`00802526365669
`1007822-28
`16
`9
`28
`2.50
`00802526365720
`1007823-28
`Minimum I.D. of Guiding Catheter = .056" Maximum Guide Wire = .014"
`
`C U S T O M E R S E R V I C E 8 0 0 2 2 7 - 9 9 0 2
`
`
`Page 9
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`MULTI-LINK MINI VISION™ Coronary Stent System
`
`INDICATIONS
`The Guidant MULTI-LINK MINI VISION™ RX and Guidant MULTI-LINK MINI VISION™ OTW Coronary Stent Systems are
`indicated for improving coronary luminal diameter in patients with abrupt or threatened abrupt closure with failed
`interventional therapy of de novoand restenotic native coronary artery lesions (length ≤ 25 mm) with reference vessel
`diameters from 2.0 to 2.5 mm.
`
`CONTRAINDICATIONS
`The Guidant MULTI-LINK MINI VISION™ Coronary Stent Systems are contraindicated for use in:
`• Patients in whom anti-platelet and / or anti-coagulant therapy is contraindicated.
`• Patients judged to have a lesion that prevents complete inflation of an angioplasty balloon.
`
`WARNINGS AND PRECAUTIONS
`Also see Individualization of Treatment in the IFU.
`
`WARNINGS
`Long term outcome for this permanent implant is unknown at present.
`• Judicious selection of patients is necessary since the use of this device carries the associated risk of subacute thrombosis,
`vascular complications and / or bleeding events.
`• Persons allergic to L-605 CoCr alloy may suffer an allergic reaction to this implant.
`• Implantation of the stent should be performed only by physicians who have received appropriate training.
`• Stent placement should only be performed at hospitals where emergency coronary artery bypass graft surgery can be
`readily performed.
`• Subsequent restenosis may require repeat dilatation of the arterial segment containing the stent. The long-term outcome
`following repeat dilatation of endothelialized stents is unknown at present.
`• When multiple stents are required, stent materials should be of similar composition. A combination of 316L stainless steel
`and L-605 CoCr alloy is acceptable should this be required.
`
`Stent Handling - Precautions
`• For single use only. Do not resterilize or reuse. Note the product "Use By" date.
`• Do not remove the stent from its Delivery System as removal may damage the stent and / or lead to stent embolization.
`Stent system is intended to perform together as a system.
`• Delivery System should not be used in conjunction with other stents.
`• Special care must be taken not to handle or in any way disrupt the stent on the balloon. This is most important during
`catheter removal from packaging, placement over guide wire and advancement through rotating hemostatic valve adapter
`and guiding catheter hub.
`• Do not manipulate (e.g., "roll") the stent with your fingers, as this action may loosen the stent from the delivery balloon.
`• Use only the recommended balloon inflation medium. Never use air or any gaseous medium to inflate the balloon, as this
`may cause uneven expansion and difficulty in deployment of the stent.
`
`Stent Placement - Precautions
`• Do not prepare or pre-inflate Delivery System prior to stent deployment other than as directed. Use balloon purging
`technique described in Delivery System Preparation.
`• Implanting a stent may lead to dissection of the vessel distal and / or proximal to the stent and may cause acute closure of
`the vessel requiring additional intervention (CABG, further dilatation, placement of additional stents, or other).
`• When treating multiple lesions, stent the distal lesion prior to stenting the proximal lesion. Stenting in this order obviates
`the need to cross the proximal stent in placement of the distal stent, and reduces the chance of dislodging the proximal
`stent.
`• Do not expand the stent if it is not properly positioned in the vessel. (See Stent / System Removal – Precautions.)
`• Placement of a stent has the potential to compromise side branch patency.
`• Do not exceed the Rated Burst Pressure as indicated on the product label. Monitor balloon pressures during inflation.
`Use of pressures higher than specified on product label may result in a ruptured balloon with possible intimal damage and
`dissection.
`• An unexpanded stent may be retracted into the guiding catheter one time only. Subsequent movement in and out through
`the distal end of the guiding catheter should not be performed as the stent may be damaged when retracting the
`undeployed stent back into the guiding catheter. Should any resistance be felt at any time during withdrawal of the
`Coronary Stent System, the entire system should be removed as a single unit.
`• Stent retrieval methods (use of additional wires, snares and / or forceps) may result in additional trauma to the coronary
`vasculature and / or the vascular access site. Complications may include bleeding, hematoma or pseudoaneurysm.
`
`Stent / System Removal - Precautions
`Should any resistance be felt at any time during either lesion access or removal of the Delivery System post-stent
`implantation, the entire system should be removed as a single unit.
`
`When removing the Delivery System as a single unit:
`• DO NOT retract the Delivery System into the guiding catheter.
`• Position the proximal balloon marker just distal to the tip of the guiding catheter.
`• Advance the guide wire into the coronary anatomy as far distally as safely possible.
`• Tighten the rotating hemostatic valve to secure the Delivery System to the guiding catheter; then remove the guiding
`catheter and Delivery System as a single unit.
`Failure to follow these steps and / or applying excessive force to the Delivery System can potentially result in loss or damage
`to the stent and / or Delivery System components.
`
`If it is necessary to retain guide wire position for subsequent artery / lesion access, leave the guide wire in place and remove
`all other system components.
`
`Post Implant - Precautions
`When crossing a newly deployed stent with a guide wire, balloon or Delivery System, exercise care to avoid disrupting the
`stent geometry.
`
`MRI (Magnetic Resonance Imaging) Statement
`The Guidant MULTI-LINK MINI VISION™ stent has been shown in non-clinical testing to be MRI safe immediately following
`implantation. MRI test conditions used to evaluate this stent were: for magnetic field interactions, a static magnetic field
`strength of 3 tesla with a maximum spatial gradient magnetic field of 3.3 tesla/meter; for MRI-related heating, a maximum
`whole body averaged specific absorption rate (SAR) of 2.0 W/kg for 15 minutes of MR imaging. While a single stent
`produced a temperature rise of less than 0.6°C and should not migrate under these conditions, the response of overlapping
`stents or stents with fractured struts is unknown. Non-clinical testing has not been performed to rule out the possibility of
`stent migration at field strengths higher than 3 tesla. MR image quality may be compromised if the area of interest is in the
`exact same area or relatively close to the position of the stent.
`
`ADVERSE EVENTS
`Potential Adverse Events
`Adverse events may be associated with the use of a coronary stent in native coronary arteries:
`• Acute myocardial infarction
`• Allergic reaction to contrast
`• Arterial perforation
`• Arterial rupture
`• Arteriovenous fistula
`• Cardiac arrhythmias
`• Bleeding complications (including transfusions)
`• Coronary spasm
`• Death
`• Dissection of the coronary artery
`• Drug reaction to anti-platelet agent
`• Emergency or non-emergent coronary artery bypass graft surgery
`• Entry site complications
`• Hypotension / hypertension
`• Infection
`• Injury to the coronary artery
`• Ischemia
`• Pseudoaneurysm
`• Restenosis of the stented segment
`• Stent embolization
`• Stent thrombosis / emboli
`• Stroke / cerebrovascular accident
`• Total occlusion of the coronary artery
`• Unstable angina pectoris
`• Vascular complications
`
`C U S T O M E R S E R V I C E 8 0 0 2 2 7 - 9 9 0 2
`
`
`Page 10
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`Cobalt chromium technology allows for thin
`struts for excellent clinical outcomes without
`compromising strength or radiopacity.
`
`Flexible low profile design
`
`.014" maximum
`guide wire
`diameter
`
`C O R O N A R Y S T E N T
`
`S Y S T E M S
`
`GUIDANT MULTI-LINK VISION® OTW
`C o r o n a r y S t e n t S y s t e m
`
`Stock
`Number
`
`1010142-08
`1010143-08
`1010144-08
`1010145-08
`
`1010142-12
`1010143-12
`1010144-12
`1010145-12
`
`1010142-15
`1010143-15
`1010144-15
`1010145-15
`
`1010142-18
`1010143-18
`1010144-18
`1010145-18
`
`1010142-23
`1010143-23
`1010144-23
`1010145-23
`
`UPN
`Number
`
`00802526328985
`00802526329043
`00802526329128
`00802526329203
`
`00802526328992
`00802526329050
`00802526329135
`00802526329210
`
`00802526329005
`00802526329067
`00802526329142
`00802526329227
`
`00802526329012
`00802526329074
`00802526329159
`00802526329234
`
`00802526329029
`00802526329081
`00802526329166
`00802526329241
`
`Stent Sizes
`Available
`(mm)
`
`Stent Length
`(mm)
`
`Stent
`Nominal
`Pressure
`(atm)
`
`Rated Burst
`Pressure
`(atm)
`
`2.75
`3.00
`3.50
`4.00
`
`2.75
`3.00
`3.50
`4.00
`
`2.75
`3.00
`3.50
`4.00
`
`2.75
`3.00
`3.50
`4.00
`
`2.75
`3.00
`3.50
`4.00
`
`8
`8
`8
`8
`
`12
`12
`12
`12
`
`15
`15
`15
`15
`
`18
`18
`18
`18
`
`23
`23
`23
`23
`
`9
`9
`9
`9
`
`9
`9
`9
`9
`
`9
`9
`9
`9
`
`9
`9
`9
`9
`
`9
`9
`9
`9
`
`16
`16
`16
`16
`
`16
`16
`16
`16
`
`16
`16
`16
`16
`
`16
`16
`16
`16
`
`16
`16
`16
`16
`
`16
`9
`28
`2.75
`00802526329036
`1010142-28
`16
`9
`28
`3.00
`00802526329098
`1010143-28
`16
`9
`28
`3.50
`00802526329173
`1010144-28
`16
`9
`28
`4.00
`00802526329258
`1010145-28
`Minimum I.D. of Guiding Catheter = .056" Maximum Guide Wire = .014"
`
`List
`Price
`($)
`
`$2,150
`$2,150
`$2,150
`$2,150
`
`$2,200
`$2,200
`$2,200
`$2,200
`
`$2,225
`$2,225
`$2,225
`$2,225
`
`$2,260
`$2,260
`$2,260
`$2,260
`
`$2,410
`$2,410
`$2,410
`$2,410
`
`$2,510
`$2,510
`$2,510
`$2,510
`
`C U S T O M E R S E R V I C E 8 0 0 2 2 7 - 9 9 0 2
`
`
`Page 11
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`MULTI-LINK VISION® Coronary Stent System
`
`INDICATIONS:
`The Guidant MULTI-LINK VISION® RX and Guidant MULTI-LINK VISION® OTW Coronary Stent Systems are indicated for
`improving coronary luminal diameter in the following (see Individualization of Treatment):
`• Patients with symptomatic ischemic heart disease due to discrete de novo or restenotic native coronary artery lesions
`(length ≤ 25 mm) with reference vessel diameters ranging from 3.0 mm to 4.0 mm.
`• Patients with symptomatic ischemic heart disease due to lesions in saphenous vein bypass grafts (length ≤ 25 mm) with
`reference vessel diameters ranging from 3.0 mm to 4.0 mm.
`• Restoring coronary flow in patients experiencing acute myocardial infarction, as confirmed by ST segment elevation or
`angiographic findings, who present within 12 hours of symptom onset with native coronary artery lesions of length
`≤ 25 mm with a reference vessel diameter of 3.0 mm to 4.0 mm.
`Outcome (beyond 9 months) for this permanent implant is unknown at present.
`
`CONTRAINDICATIONS:
`The Guidant MULTI-LINK VISION® RX and Guidant MULTI-LINK VISION® OTW Coronary Stent Systems are contraindicated
`for use in:
`• Patients in whom anti-platelet and / or anti-coagulant therapy is contraindicated.
`• Patients judged to have a lesion that prevents complete inflation of an angioplasty balloon.
`
`WARNINGS AND PRECAUTIONS (see Individualization of Treatment):
`
`WARNINGS
`• Judicious selection of patients is necessary since the use of this device carries the associated risk of subacute
`thrombosis, vascular complications and / or bleeding events.
`• Persons allergic to L-605 cobalt chromium alloy may suffer an allergic reaction to this implant.
`•
`Implantation of the stent should be performed only by physicians who have received appropriate training.
`• Stent placement should only be performed at hospitals where emergency coronary artery bypass graft surgery can be
`readily performed.
`• Subsequent restenosis may require repeat dilatation of the arterial segment containing the stent. The long-term outcome
`following repeat dilatation of endothelialized stents is unknown at present.
`• When multiple stents are required, stent materials should be of similar composition. Placing multiple stents of different
`metals in contact with each other may increase the potential for corrosion. The risk of in vivocorrosion does not appear
`to increase based on in vitro corrosion tests using an L-605 CoCr alloy stent (Guidant MULTI-LINK VISION® Coronary
`Stent) in combination with a 316L stainless steel alloy stent (Guidant MULTI-LINK TETRA™ Coronary Stent).
`
`Stent Handling - Precautions
`• For single use only. Do not resterilize or reuse. Note the product "Use By" date.
`• Do not remove stent from its Delivery System as removal may damage the stent and / or lead to stent embolization.
`Stent system is intended to perform as a system.
`• Delivery System should not be used in conjunction with other stents.
`• Special care must be taken not to handle or in any way disrupt the stent on the balloon. This is most important during
`catheter removal from packaging, placement over guide wire and advancement through rotating hemostatic valve
`adapter and guiding catheter hub.
`• Do not manipulate (e.g.,"roll") the stent with your fingers, as this action may loosen the stent from the delivery balloon.
`• Use only the appropriate balloon inflation media. Do not use air or any gaseous medium to inflate the balloon as this
`may cause uneven expansion and difficulty in deployment of the stent.
`
`•
`
`Stent Placement - Precautions
`• Do not prepare or pre-inflate Delivery System prior to stent deployment other than as directed. Use balloon purging
`technique described in Delivery System Preparation (9.3.2).
`Implanting a stent may lead to dissection of the vessel distal and / or proximal to the stent and may cause acute closure
`of the vessel requiring additional intervention (CABG, further dilatation, placement of additional stents, or other).
`• When treating multiple lesions, the distal lesion should be initially stented, followed by stenting of the proximal lesion.
`Stenting in this order obviates the need to cross the proximal stent in placement of the distal stent and reduces the
`chances for dislodging the proximal stent.
`
`• Do not expand the stent if it is not properly positioned in the vessel. (See Stent / System Removal - Precautions (5.3)
`• Placement of a stent has the potential to compromise side branch patency.
`• Do not exceed Rated Burst Pressure (RBP) as indicated on product label. Balloon pressures should be monitored
`during inflation. Use of pressures higher than specified on product label may result in a ruptured balloon with possible
`intimal damage and dissection.
`• An unexpanded stent may be retracted into the guiding catheter one time only. Subsequent movement in and out
`through the distal end of the guiding catheter should not be performed as the stent may be damaged when retracting the
`undeployed stent back into the guiding catheter. Should any resistance be felt at any time during withdrawal of the
`Coronary Stent System, the entire system should be removed as a single unit.
`• Stent retrieval methods (use of additional wires, snares and / or forceps) may result in additional trauma to the coronary
`vasculature and / or the vascular access site. Complications may include bleeding, hematoma or pseudoaneurysm.
`
`Stent / System Removal - Precautions
`Should any resistance be felt at any time during either lesion access or removal of the Delivery System post-stent
`implantation, the entire system should be removed as a single unit.
`When removing the Delivery System as a single unit:
`• DO NOT retract the Delivery System into the guiding catheter.
`• Position the proximal balloon marker just distal to the tip of the guiding catheter.
`• Advance the guide wire into the coronary anatomy as far distally as safely possible.
`• Tighten the rotating hemostatic valve to secure the Delivery System to the guiding catheter; then remove the guiding
`catheter and Delivery System as a single unit.
`Failure to follow these steps and / or applying excessive force to the Delivery System can potentially result in loss or
`damage to the stent and / or Delivery System components.
`If it is necessary to retain guide wire position for subsequent artery / lesion access, leave the guide wire in place and
`remove all other system components.
`
`Post Implant - Precautions
`• Care must be exercised when crossing a newly deployed stent with a coronary guide wire, balloon or Delivery System
`to avoid disrupting the stent geometry.
`
`POTENTIAL ADVERSE EVENTS:
`Adverse events may be associated with the use of a coronary stent in native coronary arteries:
`• Acute myocardial infarction
`• Arrhythmias, including VF and VT
`• Death
`• Dissection
`• Drug reactions to anti-platelet agents / contrast medium
`• Emboli, distal (air, tissue or thrombotic emboli)
`• Emergent Coronary Artery Bypass Surgery
`• Hemorrhage, requiring transfusion
`• Hypotension / Hypertension
`•
`Infection and / or pain at insertion site
`•
`Ischemia, Myocardial
`• Perforation
`• Pseudoaneurysm, Femoral
`• Restenosis of stented segment
`• Spasm
`• Stent embolization
`• Stent thrombosis / occlusion
`• Stroke / cerebrovascular accident
`• Total occlusion of coronary artery
`
`C U S T O M E R S E R V I C E 8 0 0 2 2 7 - 9 9 0 2
`
`
`Page 12
`
`Teleflex Ex. 2211
`Medtronic v. Teleflex
`
`

`

`C O R O N A R Y S T E N T
`
`S Y S T E M S
`
`.014" maximum
`guide wire
`diameter
`
`List
`Price
`($)
`
`$2,150
`$2,150
`$2,150
`$2,150
`
`$2,200
`$2,200
`$2,200
`$2,200
`
`$2,225
`$2,225
`$2,225
`$2,225
`
`$2,260
`$2,260
`$2,260
`$2,260
`
`$2,410
`$2,410
`$2,410
`$2,410
`
`$2,510
`$2,510
`$2,510
`$2,510
`
`Cobalt chromium technology allows for thin
`struts for excellent c

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket