International Journal of Cardiology 222 (2016) 141–147
`
`Contents lists available at ScienceDirect
`
`International Journal of Cardiology
`
`journal homepage: www. else vi er. com/ locate/ijcard
`
`Review
`Guide extension, unmissable tool in the armamentarium of modern
`interventional cardiology. A comprehensive review
`Enrico Fabris MD a,c, Mark W. Kennedy MB, BCh, BAO a, Carlo Di Mario MD, PhD b, Gianfranco Sinagra MD c,
`Vincent Roolvink MD a, Jan Paul Ottervanger MD, PhD a,
`Arnoud W.J. van't Hof MD, PhD a, Elvin Kedhi MD, PhD a,Ì
`a ISALA Hartcentrum, Zwolle, The Netherlands
`b Royal Brompton Hospital & NHLI Imperial College, London, United Kingdom
`c Cardiovascular Department, Ospedali Riuniti and University of Trieste, Trieste, Italy
`
`a r t i c l e
`
`i n f o
`
`a b s t r a c t
`
`Due to the aging population undergoing percutaneous coronary intervention (PCI), interventional cardiologists
`are confronted daily with treatment of lesions with complex anatomy. Despite improvements in stent devices
`and PCI techniques, these lesions remain a challenge in terms of procedural success. Guide-extensions (GE) are
`coaxial “mother and child” catheters employed to facilitate device delivery but they can be used in many different
`complex scenarios. A comprehensive review of the possible applications of GE and of the GuideLiner™ (GL), the
`most widely used GE device, is missing. We therefore aim to provide a comprehensive review of all the potential
`applications of the GL and other GE devices, describe its limitations as well as tips and tricks for successful usage
`of this GE catheter.
`
`© 2016 Elsevier Ireland Ltd. All rights reserved.
`
`Article history:
`Received 15 June 2016
`Accepted 27 July 2016
`Available online 29 July 2016
`
`Keywords:
`Complex PCI
`Guide extension catheter
`GuideLiner
`
`1. Introduction
`
`With the increasing age of the population undergoing percutaneous
`coronary intervention (PCI) and advances in interventional cardiology,
`increasingly complex procedures are being faced on a regular basis. In
`coronary lesions with extreme tortuosity and/or calci cations, stent de-
`livery to the target lesion may be challenging despite adequate lesion
`preparation and improved deliverability of novel stent devices.
`Adequate back-up support remains the cornerstone for successful PCI.
`Extra back-up support guiding catheters and active support techniques,
`such as deep intubation in combination with either buddy wires and/or
`an anchor balloon, are commonly used to improve support. The “mother
`and child” concept of having a smaller catheter for intracoronary
`insertion through the conventional guiding catheter was introduced
`by skilful Japanese colleagues advising Terumo on the design of the
` rst dedicated device, the Heartrail™ system (Terumo, Japan). Unfortu-
`nately this system was cumbersome and required removal of the
`haemostatic valve followed by advancement over the coronary wire
`into and through the mother guide, with subsequent reconnection of
`the haemostatic valve to the proximal end of the Heartrail catheter.
`The GuideLiner™ (GL) catheter (Vascular Solutions, Minneapolis, MN),
`connecting a distal exible 20 cm catheter to steel rod, and allowing
`
`Ì Corresponding author at: Isala Hartcentrum, Docter van Heesweg 2, Zwolle, The
`Netherlands.
`E-mail address: e.kedhi@isala.nl (E. Kedhi).
`
`http://dx.doi.org/10.1016/j.ijcard.2016.07.168
`0167-5273/© 2016 Elsevier Ireland Ltd. All rights reserved.
`
`for rapid exchange, being a monorail system, overcame most of these
`limitations and became the prototype of other similar guide-extension
`IMDS,
`the
`(GE) systems (Guidion™ Flexible Guide Extension,
`Netherlands and Guidezilla™ Guide Extension Catheter, Boston
`Scienti c, Boston). Since the rst-in-man report of the successful use
`of GL for distal stent delivery following failure of conventional
`techniques [1], this device has gained popularity and has been used
`for many other applications. The aim of this manuscript is to offer a
`complete overview of the current possible applications of the GL and
`other GE devices, to describe its limitations and to report results obtain-
`ed in prior studies.
`
`2. The GuideLiner and other guide extension devices
`
`The GL is a coaxial “mother and child” GE system, which has been
`developed for deep vessel engagement and device delivery; providing
`active guide support by its long exible tubular end, which can be deep-
`ly advanced into the target vessel. Unlike deep intubation of a guiding
`catheter, the GL has no primary curve and its soft distal tip promises a
`low dissection risk compared to deep seating of regular guides. The
` rst generation GL received FDA approval and CE marking in 2009 and
`was designed as a single lumen rapid exchange catheter with a exible
`20-cm tip connected by a metal collar to a 115-cm stainless steel shaft.
`The second-generation V2 system has a 5 cm longer exible tubular end
`and an all polymer collar for increased exibility; in the more recent V3
`the 25 cm rapid exchange section is maintained but with an additional
`
`
`Page 1
`
`Teleflex Ex. 2194
`Medtronic v. Teleflex
`
`

`

`142
`
`E. Fabris et al. / International Journal of Cardiology 222 (2016) 141–147
`
`Fig. 1. Guide-extension catheters A) The GuideLiner™-V3 extension catheter. B) Guidion™ Flexible Guide Extension and C) Guidezilla™ Guide-Extension Catheter.
`
`“half-pipe” to assist device alignment when passing through the collar
`transition (Fig. 1) and to limit the incidence of stent collar interactions.
`The GL reduces the inner guide diameter by approximately 1Fr and is
`currently available in 4 sizes: 5.5, 6, 7, and 8 Fr. Once the guide catheter
`and guide wire are placed, the GL catheter can be advanced over the
`guide wire through the haemostatic valve as an extension to the guide
`catheter. Subsequently, the procedure can be continued as usual,
`without need for disconnection and reattachment. The Guidezilla™ is
`available only in a 6 Fr compatible version, with an inner lumen mini-
`mally larger than the corresponding GL (0.057″ instead of 0.056″),
`with a hypotube rather than a steel ribbon to improve pushability, a
`proprietary coating rather than silicone and with a polymer coated
`metal collar to facilitate device insertion at the transition point. The
`Guidion™ has a more exible atraumatic distal end; the most recent
`version promises to have better pushability and is available also in a
`5 Fr compatible version. Besides being usable with smaller 5 Fr guiding
`catheters, the 5 Fr option allows insertion alongside a balloon catheter
`(for instance during anchoring) and its use in combination with a 6 Fr
`guide extension to create a telescopic system able to reach more distal
`locations and offer greater support.
`
`3. Principal indication for device utilization
`
`Current published data shows that GL is primarily used in patients
`who are signi cantly older, more frequently have multi-vessel disease
`[2] and present lesions which are signi cantly more complex [2],
`more calci ed, and have longer lesion length [3] compared to patients
`
`in which PCI can be performed without using the device. The main GL
`case series/registries, which have described target lesions using the
`AHA/ACC lesion characteristics, report a percentage of B2/C ranging
`between 91 and 97% [3–5], and in the largest series [2] the percentage
`of Type-C lesions was 78% (Fig. 2). In the context of such complex
`anatomy, reported procedural success using the GL, in series with at
`least ten patients, ranges from 80% to 100% (Table 1).
`
`3.1. Presence of vessel angulation and tortuosity
`
`Despite improvements in second-generation stent design, including
`improved stent trackability, pushability and overall deliverability,
`severe vessel angulation and tortuosity remain the most frequent
`cause of procedural failure. The GL provides an elegant method to
`overcome this challenge, and represents one of the most common indi-
`cations for its use. Eddin et al. [2] have shown that proximal vessel an-
`gulation and lesion angle were the main predictors of GL use. A 45°
`proximal vessel angle predicted the need for GL use with a sensitivity
`of 73% and speci city of 74%. Chan et al. [6] have also reported that an
`angulated take-off of the target vessel and tortuosity represented the
`31% and 7%, respectively, of the indication for GL usage. Furthermore,
`coronary artery tortuosity, de ned as ≥3 bends of ≥45° or ≥1 bend of
`≥90° change in vessel direction, was identi ed in 43% of cases in a series
`of PCI with GL support [7]. Finally, proximal tortuosity was found in
`21.9% of cases by Dursun et al. [5] and severe tortuosity was reported
`in 35% of case in one of the largest series published [8].
`
`
`Page 2
`
`Teleflex Ex. 2194
`Medtronic v. Teleflex
`
`

`

`E. Fabris et al. / International Journal of Cardiology 222 (2016) 141–147
`
`143
`
`Fig. 2. Percentage of AHA/ACC type lesions characteristics reported in published case series/registries. Left side: percentage of b2/C type lesions; right side: percentage of C type lesion.
`The width of the columns is representative of the sample size. Pts = patients.
`
`3.2. Presence of severe calci cation
`
`In vessels with heavy proximal calci cation, advancement of a stent
`to a distal target lesion can remain challenging, owing to signi cant
`friction between stent struts and the calci ed vessel wall. GL utilization
`in this particular situation has a dual bene t: it provides robust back-up
`and therefore improves stent trackability and when advanced until just
`proximal to the target lesion; it avoids friction between the stent and
`calci ed vessel wall. In the main case series, the presence of heavy
`calci cation in PCI performed using the device ranges from one-third
`to two-third of cases [1,5,7–9]. While use of rotational atherectomy
`may be necessary to provide full deployment of the stent in calci ed
`lesions [10], burr delivery to distal lesion sites can remain dif cult,
`particularly in tortuous anatomy, and in such situations is associated
`with an increased risk of vessel perforation. Anecdotal case reports
`have described the passage of the GL beyond the proximal tortuous
`segment facilitating burr delivery to distal calci ed lesions [11,12].
`
`1.25–1.5 mm burrs can be accommodated through a 7Fr GL [11,12].
`However the insertion of the burr through the collar can be challenging
`[11]. A way to overcome the problem is to manually load the burr inside
`the distal end of the GE outside the body advancing both together up to
`the ostium and then sliding the GE.
`
`3.3. Chronic total occlusion
`
`Chronic total occlusion (CTO) intervention represents one of the
`most challenging subsets of PCI. During CTO PCI, procedural failure is
`often due to an inability to deliver a balloon or micro-catheter across
`the lesion [13]. GL use achieves better support and deeper intubation
`of target vessels and its value has been demonstrated in numerous se-
`ries [6,13,14]. A single centre experience has reported use of GL in
`17.8% (66/372) of CTO procedures [13]. In 28 of these cases, GL assisted
`initial balloon or micro-catheter advancement to the culprit CTO lesion
`and in 6 of these cases was used to support the initial wiring of the
`
`Table 1
`Registries/case series including at least 10 patients.
`
`Studies
`
`Date of
`publication
`
`Number of
`pts
`
`Age
`(years)
`
`By pass
`n(%)
`
`Procedural
`success (%)
`
`Stent deformation or
`dislodgment (%)
`
`Coronary
`dissection (%) Other complications
`
`97%
`1(7.6%)
`67 ± 13
`90%
`–
`71 ± 10
`100%
`3(18.7%)
`68 ± 8
`90%
`2(9.5%)
`65 ± 13
`10(14.2%) 93%
`67 ± 13
`100%
`–
`69 ± 12
`–
`82%
`64 ± 10
`–
`87.5%
`72 ± 8
`12(21.8%) 98%
`69 ± 13
`3(4.6%)
`95%
`70 ± 10
`71.5 ± 11 28(8.5%)
`80%
`70.5 ± 10 2(1%)
`86%
`69 ± 10
`2(1.6%)
`93%
`
`2010
`Mamas et al. [1]
`13
`2011
`Cola et al. [9]
`10
`2012
`Dardas et al. [7]
`16
`2012
`Luna et al. [32]
`21
`2012
`De Man et al. [4]
`70
`2013
`Eddin et al. [3]
`22
`28a
`2013
`Kovavic et al. [13]
`15b
`Garcia-Blas et al. [34] 2014
`55c
`Chan et al. [6]
`2015
`Dursun et al. [5]
`2015
`64
`347d
`Waterbury et al. [2]
`2015
`188e
`Alkhall et al. [8]
`2016
`124f
`Alkhall et al. [8]
`2016
`a 1 patient had successful renal denervation.
`b 16 consecutive procedures in 15 patients.
`c All CTO procedures.
`d 363 procedures in 347 patients.
`e All procedures in which GuideLiner-V2 were used.
`f All procedures in which GuideLiner-V3 were used. n = number, Pts = patients.
`
`6.2%
`20%
`–
`–
`1.4%
`–
`–
`–
`–
`–
`2.3%
`11.1%
`2.4%
`
`–
`30%
`12.5%
`–
`–
`–
`–
`6.25%
`–
`–
`3.3%
`0.5%
`–
`
`–
`–
`–
`1 acute vessel closure; pressure damping in 57%
`1 air embolism
`–
`1 wire perforation
`–
`–
`–
`1 ventricular arrhythmia
`–
`–
`
`
`Page 3
`
`Teleflex Ex. 2194
`Medtronic v. Teleflex
`
`

`

`144
`
`E. Fabris et al. / International Journal of Cardiology 222 (2016) 141–147
`
`lesion. The GL successfully facilitated micro-balloon crossing of the CTO
`lesion in 85.7% of cases (24/28), when a guidewire had already crossed
`the CTO. Similar procedure success (97%) was also reported from Chan
`et al. [6] in 33 CTO GL-facilitated PCI. Furthermore, during reverse
`controlled antegrade and retrograde subintimal tracking (CART) the
`advancement of the GL from the antegrade guide catheter can offer a
`visible and accessible target for the retrograde wire. “GuideLiner reverse
`CART” is an elegant modi cation of reverse CART shortening the dis-
`tance between the site of re-entry of the retrograde guidewire and the
`antegrade guiding catheter [15–17] (Fig. 3).
`
`3.4. Anomalous origin of coronary arteries or grafts
`
`Extreme take-off angulation, particularly in case of anomalous origin
`of coronary arteries represents a challenge to adequate catheter engage-
`ment, and may result in dissection or insuf cient support during PCI. GL
`utilization may improve both guide catheter coaxial orientation and
`support (Fig. 4). Ramanathan et al. [18] described an anomalous origin
`of a right coronary artery (RCA) arising from the left coronary cusp,
`where GL facilitated coaxial engagement and stent delivery after failure
`of conventional guides to achieve selective cannulation. Similarly, in a
`case with critical left circumex artery (LCX) stenosis, which arose
`from an anomalous LM coronary artery originating from the right
`sinus of Valsalva, GL was placed in the distal LM for additional support
`and avoided guide disengagement [19]. In a recent case, GL was success-
`fully used to enhance backup and increase pushing strength allowing
`stent deployment in distal LCX arising from a single left coronary artery
`[20].
`Similarly, one of the main dif culties in graft PCI, may be the near
`impossibility of coaxial engagement of the ostia. GE often provides the
`only solution to properly engage with adequate support both vein grafts
`in the ascending aorta or the mammary artery, especially in the
`presence of an extremely tortuous subclavian. Farooq et al. [21] have
`highlighted the feasibility of GE devices, including GL, to overcome
`many of these problems and facilitate trans-radial graft interventions
`in a series of selected complex cases. Park et al. [22] reported a
`challenging PCI in which GL allowed stent deployment in left anterior
`descending (LAD) artery through a tortuous left internal mammary
`artery (LIMA). Additional backup support for equipment delivery can
`be achieved by deep intubation of GL in the graft or indeed deep inser-
`tion of guide catheter over GL system (“Rail-Roading”). Finally, when
`
`Fig. 3. “GuideLiner reverse CART”. GuideLiner is advanced antegradely (arrows) to
`reduce the distance between the site of re-entry of the retrograde guidewire and the
`antegrade guiding catheter.
`
`faced with a proximal graft lesion, which may prevent deep guide intu-
`bation, further backup may be achieved with a ‘Swan-Neck’ maneuver
`[21] with the guide catheter positioned in the aortic sinus and tip of
`the GL extended to the vein graft ostium.
`
`4. Other possible applications
`
`Although GL was primarily designed to access discrete regions of the
`coronary vasculature, and to facilitate placement of interventional
`devices, other versatile applications of this device have been reported.
`
`4.1. Selective injection of target coronary segments to reduce contrast usage
`
`As the GL catheter is based upon a monorail rapid-exchange
`platform, contrast injections through the guiding catheter enter the
`proximal portion of the GL and exit it distally (selectively in the target
`segment). Since there is no loss of contrast into proximal side branches,
`the volume of contrast necessary for target vessel opaci cation is typi-
`cally lower than if injected through the guiding catheter alone [23].
`Serajian et al. [24] using the above-mentioned property, have shown
`that in situations with competitive LIMA ow, GL may be used for selec-
`tive LAD angiography. Using such a technique provided visualization of
`an LAD lesion, which otherwise was not well appreciated.
`Pershad et al. [25] after failed attempts to visualize a distal vessel
`because of the presence of proximal aneurysm with swirling of con-
`trast in the proximal segment, successfully used a GL catheter placed
`beyond the aneurysm to facilitate sub-selective injection into the
`distal vessel.
`
`4.2. Optical coherence tomography assessment of a vessel through the
`GuideLiner
`
`Optical coherence tomography (OCT) assessment through the GL
`may be another potential application. Mitomo et al. [26] reported a
`case in which a GL was used to facilitate OCT catheter delivery and min-
`imize the contrast dose used in a challenging complex lesion.
`
`4.3. GuideLiner for thrombectomy
`
`Stys et al. [27] and Farooq et al. [28] presented successful use of the
`GL for thrombus aspiration after failure of dedicated manual aspiration
`thrombectomy devices. The GL should be removed with the guide cath-
`eter fully intubated in the coronary ostium to prevent any potential em-
`bolization of thrombus to the systemic vasculature [28]. Further blood
`should be then aspirated from the guide catheter to ensure no thrombus
`remains within the guide catheter.
`An overview of potential indications for GE use is represented in
`Table 2.
`
`5. Potential complications
`
`As described, GL may signi cantly facilitate PCI; nonetheless it also
`carries some device-speci c complications; these include stent damage,
`stent loss, stent catheter deformation, dissection, air embolism and
`pressure dampening (see Tables 1 and 3). Awareness of the system lim-
`itations can simplify and render safer use of the device.
`
`5.1. Stent and balloon disruption
`
`The stainless steel collar present in the rst generation GL has been
`implicated in several instances of stent and balloon disruption [1,13,
`29]; and in fact the main limitation encountered at the beginning expe-
`rience of the device [1] was stent damage which occurred in two of the
`32 stents delivered (6.2%) as the stents were advanced through the
`catheter portion of the device. Similarly, Seto et al. [30] witnessed dam-
`age of stent struts in two cases upon attempted advancement through
`
`
`Page 4
`
`Teleflex Ex. 2194
`Medtronic v. Teleflex
`
`

`

`E. Fabris et al. / International Journal of Cardiology 222 (2016) 141–147
`
`145
`
`Fig. 4. GuideLiner use for anomalous coronary origin. Anomalous origin of the right coronary artery from the left sinus (arrows). With only guiding catheters, it was very dif cult to
`visualize angiographically the vessel (A–B). The use of GuideLiner (*) is useful to properly engage the vessel and for a better diagnostic evaluation (C–D).
`
`Table 2
`Potential indications for guide-extension use.
`
`Potential indications
`
`the proximal GL collar, such resistance to stent advancement through
`the device, may be due to the lifting up (owering) of stent struts
`against the collar of the GL [31].
`
`Common indications
`Devices delivery in presence of vessel angulation and tortuosity [2,5–7]
`Devices delivery in presence of diffuse coronary calci cations [1,3–5,7–9]
`Facilitation of catheter engagement and intervention in anomalous coronary
`arteries [18–20] and grafts [21,22]
`Supporting the initial wiring or to assist advancement of balloon or micro-catheter
`during chronic total occlusion interventions as well as to facilitate re-entry of the
`retrograde guidewire.
`
`Other possible indications
`Selective injection of target coronary segments to reduce contrast usage [23]
`Focused injection to selective opaci cation of distal vessels in presence of proximal
`aneurisms [25] or presence of graft competitive ow [24]
`Thrombectomy in bail-out situations dealing with very large thrombus burden [27,28]
`Optical coherence tomography assessment of a tortuous vessel through the
`GuideLiner [26]
`Delivery of the burr and enabling safe rotational atherectomy in tortuous calci c
`lesion [11,12]
`
`Anecdotal cases
`Retrieval of entrapped rotablator burr using counter-traction with a GuideLiner [35]
`Maintenance of coronary guidewire position during exchange of the guide catheter [36]
`Renal denervation with the aid of a GuideLiner catheter [6]
`
`Table 3
`Lists of possible complications associate with GuideLiner use.
`
`Complications
`
`Notes
`
`Stent deformation on
`advancement or withdraw
`[1,2,4,8,30,31]
`
`Disruption of the stent catheter
`[8,29]
`
`Coronary Dissection [2,6,8,32]
`
`Pressure dampening [32]
`Air embolism [4]
`
`Dislodgement of the distal marker
`[31]
`
`It is more likely to occur when using guide
`catheters with a secondary bend
`Newer GL-V3 should prevent deformation
`and/or damage on advancement
`The proximal collar cut the stent delivery balloon
`shaft while GL was withdrawn backwards to
`‘unsheathe’ a delivered stent. Occurred using a
`sharply angled Kimny guide catheter.
`Also observed with a 6-Fr GL catheter in 7-Fr
`Kimny guiding catheter during forceful dye
`injection with GL catheter ejection forward
`creating the dissection.
`Mainly observed during “6-in-7” Fr GL
`As a result of insuf cient venting of the
`wedged GL
`Occurred after extensive manipulations.
`Dislodgement of the distal marker was
`possibly due to damage of the GL inner lining
`from a deformed stent.
`
`
`Page 5
`
`Teleflex Ex. 2194
`Medtronic v. Teleflex
`
`

`

`146
`
`E. Fabris et al. / International Journal of Cardiology 222 (2016) 141–147
`
`Stent deformation or dislodgement may also occur on attempted
`withdrawal of a stent through the distal band of the GL. Moreover
`when combined with a sharply angled guide catheter, upon retraction
`of the GL, the proximal collar may kink and disrupt the stent shaft cre-
`ating a leak which prevents transmission of deployment pressure to
`the balloon [29].
`In a recent registry of 363 consecutive PCI procedures stent
`disruption and loss occurred in 8/363 (2.2%) of cases [2]. The incidence
`of this complication has decreased over time, with cases involving the
` rst generation GL device (V1) (n = 144) accounting for six out of
`eight cases of stent loss, two cases occurring with the second generation
`device (V2) (n = 119) and none with the third generation (V3) (n =
`100) [2]. However, a recent experience [8] has suggested that the
`V2-GL had a higher incidence of complications (10%, n = 19/188, of
`stent/proximal collar interaction), especially when using the right radial
`approach, compared with V1-GL historical data [1,4,32]. The possible
`mechanistic explanation for the reported increase in complications
`with V2, may reside in its length (25 cm) which frequently placed the
`proximal collar within the brachiocephalic and right subclavian seg-
`ments; an area which may be subject to ovoid distortion with respira-
`tion and therefore a potential space between the guide catheter and
`the GL may be created, allowing interaction with the stent [8]. The
`same authors demonstrated that the latest modi cations with the GL
`version 3 (V3) seem to have ameliorated these complications when
`compared with older generation [8]: with zero cases of proximal
`collar-stent interaction, only one case (0.8%, n = 1/124) of longitudinal
`stent deformation and two incidents (1.6%, n = 2/124) of stent
`interaction with the distal edge of the V3 GL described.
`
`5.2. Pressure dampening
`
`In one case series [32], using primarily 7F GL catheters, pressure
`dampening was frequently observed in 57% (12/21) patients, with one
`patient developing acute vessel closure, likely due to dissection [32].
`Pressure dampening occurring with GL use may directly contribute to
`unsuccessful of GL use.
`
`5.3. Coronary dissections
`
`Despite the GL being equipped with a softer tip than standard cath-
`eters, dissection of the coronary vessels may occur, particularly during
`deep intubations. The incidence of this event observed in a large registry
`[2] was 3.3%. Chang et al. [15] reported a LM dissection which occurred
`with a 6-Fr GL in 7-Fr Kimny guiding catheter during forceful dye
`injection: with forceful injection, the GL was ejected forward, resulting
`in a type D LM dissection. Therefore, removing GL from the loaded
`
`guiding catheter after completion of PCI for a nal control angiogram
`is recommended.
`
`6. Practical tip and tricks
`
`• On initial insertion of the GL the at push-rod should be oriented in a
`lateral position within the guiding catheter and should be advanced
`within the guiding catheter without rotation to avoid wrapping the
`guidewire around it [33].
`• In situations where the GL itself cannot be easily advanced due to the
`primary vessel anatomy, this can be facilitated by inating a balloon in
`the target lesion and use it as an anchor improving in this way the
`support of the guide catheter while advancing the GL through com-
`plex anatomy [2] (Fig. 5).
`• In the case of resistance while inserting a wire or stent through the GL:
`1) ensure the combination of the wire and stent is compatible with
`the internal diameter of the GL; 2) If a secondary wire is in use,
`check for wire wrapping of the secondary wire around the GL; if
`wire wrap is evident, consider pulling back the secondary wire and
`re-advancing it; 3) check the location of stent in relationship to the
`collar of the GL. If resistance occurred in that segment the stent should
`be inspected for signs of damage prior to re-advancement [33].
`
`Advancement of coronary stents across the rapid exchange
`transition zone can be eased by positioning the transition zone of the
`GL in a straight portion of the guiding catheter in order to facilitate
`the passage of devices through it [6]. Ideally when using a radial
`approach, the connection to the exible segment should be situated in
`the ascending aorta [8].
`
`• Interventions requiring nal kissing balloon ination are feasible with
`the use of 6-in-7 (and, where necessary, 7-in-8) GL [6]. If coronary
`guidewires have been placed in the main vessel and side branch, GL
`must to be inserted over both coronary guidewires [6]. Inserting a
`new guidewire can be dif cult and if a guidewire is needed while
`the GL is already in situ, meticulous attention has to be paid to ensure
`the guidewire passes through the lumen of the GL tubing rather than
`outside by direct visualization on uoroscopy.
`• Every effort should be undertaken to minimize pressure dampening,
`however some degree of pressure dampening seems almost inevita-
`ble. Meticulous attention must be paid to ensure that adequate
`antegrade coronary ow is preserved and that no injury has occurred
`before proceeding with the intervention [32]. Stenting of long lesions
`with multiple stents is preferably done from distally to proximally,
`however when the proximal segment of the vessel is also severely ste-
`nosed, treatment of this segment initially can facilitate the engage-
`ment of the GL catheter proximal to the distal lesion [1].
`
`Fig. 5. GuideLiner anchoring balloon technique. Insertion of the guidewire (A), then the GuideLiner is passed up to the distal end of the guiding catheter (arrow) followed by a balloon
`that is inated in the distal lesion (B). The GuideLiner is then advanced into the vessel over the inated balloon reaching the position (arrow) that allows stent delivery (*)(C).
`
`
`Page 6
`
`Teleflex Ex. 2194
`Medtronic v. Teleflex
`
`

`

`E. Fabris et al. / International Journal of Cardiology 222 (2016) 141–147
`
`147
`
`• After removal of GL/GE is important to allow back bleeding to prevent
`air embolism during subsequent contrast injection
`
`7. Conclusion
`
`The GL “mother in child” extension catheter has signi cantly im-
`proved procedural outcomes in complex lesion anatomy and broadened
`the subset of lesions where PCI can be successfully performed. Compli-
`cations with GL application are possible and operators should be aware
`of them, however the latest improvements in catheter design have
`signi cantly reduced the risk of potential device related complications.
`GL can be nowadays considered a vital tool in the armamentarium of
`the modern interventional cardiologist.
`
`Conict of interest
`
`The authors have no conict of interest to declare.
`
`References
`
`[1] M.A. Mamas, F. Fath-Ordoubadi, D.G. Fraser, Distal stent delivery with guideliner
`catheter: rst in man experience, Catheter. Cardiovasc. Interv. 76 (2010) 102–111.
`[2] T.M. Waterbury, P. Sorajja, M.R. Bell, R.J. Lennon, V. Mathew, M. Singh, S. GS, G. R.,
`Experience and complications associated with use of guide extension catheters in
`percutaneous coronary intervention, Catheter. Cardiovasc. Interv. (2015), http://
`dx.doi.org/10.1002/ccd.26329.
`[3] M.J. Eddin, E.J. Armstrong, U. Javed, J.H. Rogers, Transradial interventions with the
`guideliner catheter: role of proximal vessel angulation, Cardiovasc. Revasc. Med.
`14 (2013) 275–279.
`[4] F.H. de Man, K. Tandjung, M. Hartmann, K.G. van Houwelingen, M.G. Stoel, H.W.
`Louwerenburg, M.W. Basalus, H. Sen, M.M. Lowik, C. von Birgelen, Usefulness and
`safety of the guideliner catheter to enhance intubation and support of guide
`catheters: insights from the twente guideliner registry, EuroIntervention 8 (2012)
`336–344.
`[5] H. Dursun, A. Tastan, Z. Tanriverdi, E. Ozel, D. Kaya, Guideliner catheter application
`in complex coronary lesions: experience of two centers, Anatol. J. Cardiol. 16 (5)
`(2016) 333–339.
`[6] P.H. Chan, E. Alegria-Barrero, N. Foin, M. Paulo, A.C. Lindsay, N. Viceconte, C. Di
`Mario, Extended use of the guideliner in complex coronary interventions,
`EuroIntervention 11 (2015) 325–335.
`[7] P.S. Dardas, N. Mezilis, V. Ninios, D. Tsikaderis, E.K. Theo logiannakos, The use of the
`guideliner catheter as a child-in-mother technique: an initial single-center experi-
`ence, Heart Vessel. 27 (2012) 535–540.
`[8] M. Alkhalil, A. Smyth, S.J. Walsh, C. McQuillan, M.S. Spence, C.G. Owens, C.G.
`Hanratty, Did the use of the guideliner v2(tm) guide catheter extension increase
`complications? A review of the incidence of complications related to the use of
`the v2 catheter, the inuence of right brachiocephalic arterial anatomy and the re-
`design of the v3(tm) guideliner and clinical outcomes, Open Heart 3 (2016)
`e000331.
`[9] C. Cola, F. Miranda, B. Vaquerizo, A. Fantuzzi, J. Bruguera, The guideliner catheter for
`stent delivery in dif cult cases: tips and tricks, J. Interv. Cardiol. 24 (2011) 450–461.
`[10] E. Barbato, D. Carrie, P. Dardas, J. Fajadet, G. Gaul, M. Haude, A. Khashaba, K. Koch, M.
`Meyer-Gessner, J. Palazuelos, K. Reczuch, F.L. Ribichini, S. Sharma, J. Sipotz, I.
`Sjogren, G. Suetsch, G. Szabo, M. Valdes-Chavarri, B. Vaquerizo, W. Wijns, S.
`Windecker, A. de Belder, M. Valgimigli, R.A. Byrne, A. Colombo, C. Di Mario, A.
`Latib, C. Hamm, European expert consensus on rotational atherectomy,
`EuroIntervention 11 (2015) 30–36.
`[11] M. Vo, K. Minhas, M. Kass, A. Ravandi, Novel use of the guideliner catheter to deliver
`rotational atherectomy burrs in tortuous vessels, Case Rep. Cardiol. 2014 (2014)
`594396.
`[12] P. Costanzo, K. Aznaouridis, A. Hoye, A. Alahmar, Guideliner-facilitated rotational
`atherectomy in calci ed right coronary artery: The “child” makes the difference,
`JACC Cardiovasc. Interv. 9 (5) (2016) e47–e48.
`
`[13] J.C. Kovacic, A.B. Sharma, S. Roy, J.R. Li, R. Narayan, D.B. Kim, S.K. Sharma, A.S. Kini,
`Guideliner mother-and-child guide catheter extension: a simple adjunctive tool in
`pci for balloon uncrossable chronic total occlusions, J. Interv. Cardiol. 26 (2013)
`343–350.
`[14] E.B. Hanna, T.W. Dasari, T.A. Hennebry, Use of the guideliner catheter for the treat-
`ment of a bifurcational total occlusion of the native left anterior descending artery
`through a tortuous composite venous graft, J. Invasive Cardiol. 23 (2011) E40–E42.
`[15] Y.C. Chang, H.Y. Fang, T.H. Chen, C.J. Wu, Left main coronary artery bidirectional
`dissection caused by ejection of guideliner catheter from the guiding catheter,
`Catheter. Cardiovasc. Interv. 82 (2013) E215–E220.
`[16] A.M. Mozid, J.R. Davies, J.C. Spratt, The utility of a guideliner catheter in retrograde
`percutaneous coronary intervention of a chronic total occlusion with reverse cart-
`the “capture” technique, Catheter. Cardiovasc. Interv. 83 (2014) 929–932.
`[17] M. Vo, E.S. Brilakis, Faster, easier, safer: “guideliner reverse cart” for retrograde
`chronic total occlusion interventions, Catheter. Cardiovasc. Interv. 83 (2014)
`933–935.
`[18] P.K. Raman

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket