throbber
UNITED STATES PATENT AND TRADEMARK OFFICE
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`MEDTRONIC, INC. AND MEDTRONIC VASCULAR, INC.
`Petitioner,
`v.
`TELEFLEX LIFE SCIENCES LIMITED,
`Patent Owner.
`
`Case IPR2020-01341
`U.S. Patent No. 8,142,413
`
`Case IPR2020-01342
`U.S. Patent No. 8,142,413
`
`Case IPR2020-01343
`U.S. Patent No. RE 46,116
`
`Case IPR2020-01344
`U.S. Patent No. RE 46,116
`
`
`
`SUPPLEMENTAL DECLARATION OF STEPHEN JON DAVID BRECKER,
`MD, FRCP, FESC, FACC
`
`
`
`
`
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 1 of 63
`
`

`

`
`
`I.
`II.
`
`III.
`
`D.
`
`E.
`
`Page
`Overview .......................................................................................................... 4
`Claim Construction .......................................................................................... 5
`A.
`“interventional cardiology device”........................................................ 5
`B.
`The claimed step of inserting an interventional cardiology
`device need not be performed only after the coaxial guide
`catheter is inserted into the guide catheter. ......................................... 10
`Itou-based grounds.........................................................................................13
`A.
`Types of interventional cardiology devices insertable through
`catheter (2) ........................................................................................... 13
`B. Distal end protective catheter (5) ........................................................ 16
`C. Advancing the interventional cardiology device . . . into contact
`with or past a lesion in the second artery ............................................ 17
`Itou discloses that protective catheter (5) is inserted through a
`guide catheter “alongside of the substantially rigid portion” of
`the coaxial guide catheter. ................................................................... 21
`Itou discloses “applying a force to . . . the coaxial guide catheter
`such that the distal portion of the coaxial guide catheter remains
`seated . . . in response to an opposing backward force exerted
`by the interventional cardiology device.” ........................................... 22
`Itou discloses that protective catheter (5) is extended through a
`“proximal side opening . . . while the proximal portion remains
`within the lumen of the guide catheter.” ............................................. 23
`It would have been obvious to insert an interventional
`cardiology device such as a stent or balloon through Itou’s
`suction catheter (2). ............................................................................. 24
`Itou and Ressemann collar 2141 ......................................................... 28
`Itou and Kataishi ................................................................................. 32
`
`F.
`
`G.
`
`H.
`I.
`
`Table of Contents
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 2 of 63
`
`

`

`
`
`V.
`
`2.
`
`3.
`
`C.
`
`IV. Kontos–based grounds ...................................................................................34
`A.
`Tube 16 of Kontos’s body 12 is not a narrow tube ............................. 34
`B. Kontos necessarily resists axial and shear forces that would
`otherwise tend to dislodge the guide catheter. .................................... 38
`The side opening claims are obvious. ................................................. 43
`1.
`Replacing Kontos’s proximal funnel with a side opening would
`maximize the usable “real estate” inside the catheter assembly.
` ...................................................................................................43
`After removing Kontos’s proximal funnel, a POSITA would
`have been motivated to maximize the usable real estate, but
`even if no further modifications were made, Kontos would not
`have a “problem gap.” ...............................................................47
`After removing Kontos’s proximal funnel, Kontos’s support
`catheter 10 would remain coaxially aligned with the guide
`catheter. .....................................................................................51
`D. A POSITA would have been motivated with a reasonable
`expectation of success to achieve the no-more-than-one-french
`differential between inner diameters of the guide catheter and
`body 12. ............................................................................................... 53
`E. My opinion remains unchanged: the claims that recite a two-
`inclined side opening are obvious. ...................................................... 55
`1.
`Kontos + Ressemann and Kontos + Ressemann + Takahashi ..55
`2.
`Kontos + Ressemann + Takahashi + Kataishi ..........................57
`Teleflex’s evidence regarding secondary considerations does not alter my
`opinion that the challenged claims are obvious. ............................................59
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 3 of 63
`
`

`

`
`
`
`I.
`
`Overview
`I have been retained by Robins Kaplan LLP on behalf of Medtronic,
`1.
`
`Inc. and Medtronic Vascular, Inc. (“Medtronic”) as an independent expert to
`
`provide my opinions concerning U.S. Patent Nos. 8,142,413 and RE46,116 (the
`
`“Teleflex Patents”) in IPR2020-01341, IPR2020-01342, IPR2020-01343, and
`
`IPR2020-01344.1
`
`2.
`
`I set forth the information below as a supplement to my original
`
`declarations, as Teleflex raised new issues in their Patent Owner Responses on
`
`which I had not previously been given an opportunity in these IPRs to offer
`
`testimony. Having considered Teleflex’s arguments, and the testimony of Teleflex’s
`
`declarants, my opinions on the invalidity of the Teleflex Patents remain the same.
`
`3.
`
`I have reviewed additional material in conjunction with my analysis
`
`set forth herein. This includes the Final Written Decisions in IPRs on related
`
`Teleflex patents: IPR2020-00126, IPR2020-00127, IPR2020-00128, IPR2020-
`
`00129, IPR2020-00130, IPR2020-00132, IPR2020-00134, IPR2020-00135,
`
`IPR2020-00136, IPR2020-00137 and IPR2020-00138 (“Related IPRs”). A list of
`
`
`1 Citations to exhibits refer to exhibits filed in IPR2020-01341, unless noted
`
`otherwise. I understand that most of Teleflex’s and Medtronic’s exhibits are
`
`numbered consistently across all four IPRs.
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 4 of 63
`
`

`

`
`
`these materials includes everything cited in this declaration, and the materials
`
`disclosed in my original declarations.
`
`II. Claim Construction
`“interventional cardiology device”
`A.
`4.
`The opinion in this section generally relates to at least IPR2020-01341
`
`and IPR2020-01342.
`
`5.
`
`I am aware that when the Board instituted the IPR2020-01341 petition
`
`it determined that the limitation “interventional cardiology device” refers to “at
`
`least two types of the devices selected from the group that includes, but is not
`
`limited to, guidewires, balloon catheters, stents, and stent catheters.” IPR2020-
`
`001341, Paper 11 (“I.D.”) at 16. This limitation appears in independent claim 1 of
`
`the ’413 patent.
`
`6.
`
`At institution, the Board additionally determined that the claims do
`
`not require that more than one of guidewires, stents, stent catheters, and balloon
`
`catheters be simultaneously insertable into and through the lumen of the claimed
`
`coaxial guide catheter. Id. Additionally, it determined that Medtronic demonstrated
`
`where Itou discloses every limitation of claim 1. Id. at 24.
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 5 of 63
`
`

`

`
`
`7.
`
`I have reviewed Teleflex’s claim construction arguments, IPR2020-
`
`001341, Paper 23 (“POR”) at 14-17 (citing Ex-2138 (Declaration of Peter Keith)
`
`and Ex-2145 (Declaration of Doctor John J. Graham)),2 and I disagree.
`
`8.
`
`First, to the extent that Teleflex appears to argue this, I disagree that
`
`the term requires that all “four standard ICDs enumerated” in the specification be
`
`insertable. Id. at 14. I have reviewed the Teleflex patent and do not see that the
`
`claims require this. The claims do not, for example, recite “all interventional
`
`cardiology devices,” or “at least four interventional cardiology devices.” Ex-1001,
`
`claim 1.
`
`9.
`
`Second, I disagree that “interventional cardiology device” should be
`
`construed to be limited to the four enumerated devices and to “any other device
`
`delivered beyond the end of the device for use with a standard guide catheter to a
`
`location in the vasculature requiring treatment, to provide treatment to that
`
`location.”3 POR at 14.
`
`
`2 Where I cite to a POR in this Declaration, I have also reviewed the material cited
`
`therein, including the opinions of Teleflex’s experts.
`
`3 For example, it is my opinion that the distal protection device described in Adams
`
`(IPR2020-01342, Ex-1035) is an interventional cardiology device.
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 6 of 63
`
`

`

`
`
`10.
`
`I understand Teleflex’s argument to be, in part, that the patent
`
`specification uses the terms “cardiac treatment device” and “interventional
`
`cardiology treatment device” interchangeably with “interventional cardiology
`
`device[s].” POR at 14-15. I disagree that the patent uses these three terms as
`
`synonyms. The specification does not limit its discussion to treatment devices, or
`
`even to treatment devices and catheters that deliver those devices. For example, it
`
`explains that “[i]n order to diagnose and treat obstructive coronary artery disease it
`
`is commonly necessary to pass a guidewire or other instrument through and
`
`beyond the occlusion or stenosis of the coronary artery.” Ex-1001, 1:31-34. And it
`
`also discusses that “the interventional cardiology art would benefit from the
`
`availability of a system that would be deliverable through standard guide catheters
`
`for providing backup support . . . .” Ex-1001, 2:51-55; see also, id., 1:13-14
`
`(describing the invention relating generally to catheters used in interventional
`
`cardiology procedures).
`
`11.
`
`I understand an “interventional cardiology treatment device” to be
`
`subsets of the broader term “interventional cardiology device[s].” A POSITA
`
`understands an “interventional cardiology device” as a much more general term to
`
`include devices that might be placed in the coronary artery that are not actually
`
`treatment devices themselves, for example intracoronary imaging catheters.
`
`However, a POSITA would not even limit the term “interventional cardiology
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 7 of 63
`
`

`

`
`
`device” in general only to the coronary arteries. There is an entire subspecialty
`
`within interventional cardiology for non-coronary interventions. Examples of
`
`interventional cardiology devices that are not placed in a coronary artery would
`
`include, for example, intracardiac echocardiography catheters used for imaging the
`
`heart, percutaneous femoral closure devices, external vascular hemostasis devices,
`
`cardiac septal closure devices, and cerebral protection devices, amongst many
`
`others.
`
`12. Similarly, not even every intracoronary “interventional cardiology
`
`device” is used to provide treatment. Devices in this category include intravascular
`
`ultrasound catheter used for imaging the vessel, optical coherence tomography
`
`catheter used for imaging the vessel, and fractional flow reserve measurement
`
`“pressure” wires. These devices are for diagnosis not treatment.
`
`13. As illustrated below, an “interventional cardiology treatment device”
`
`is but a subset of “interventional cardiology device(s).”
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 8 of 63
`
`

`

`
`
`
`
`14. Only a subset of the devices specifically identified in the specification
`
`provide treatment. A guidewire does not provide treatment. It instead “provide[s] a
`
`rail over which a series of therapeutic devices can be advanced.” Ex-1015, 95. Dr.
`
`Graham’s testimony supports this. He acknowledges that a guidewire is used in
`
`conjunction with a stent or balloon in order to treat an occlusion. Ex-1801,
`
`89:2-23.
`
`15. As I have testified in Related IPRs, a guidewire does not treat a lesion,
`
`it is merely the mechanism by which a treatment device is delivered. IPR2020-
`
`00126, Ex-2238, 19:6-16, 20:7-8. Following insertion of a guidewire in a blocked
`
`artery, there may be instances in which some blood flow is restored. In some
`
`patients, there may also be improvement in ST segments on an ECG, but other
`
`patients may deteriorate with partial reperfusion. Id., 21:17-22:20. This is
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 9 of 63
`
`

`

`
`
`illustrative of why guidewires cannot be considered treatment devices, insofar as a
`
`POSITA is aware that while a guidewire may be necessary for delivering a therapy,
`
`it cannot and indeed is not, considered a treatment device.
`
`16. Moreover, when the inventors of the Teleflex Patents desired to claim
`
`a “treatment” device, they did so explicitly. See Ex-1845,4 16:1-5 (reciting
`
`advancement of a “treatment catheter”); see also IPR2020-01341, Paper 26
`
`(proposing ’413 patent substitute claims 15-17 that recite a “treatment catheter”
`
`instead of “an interventional cardiology device”). They did not do so in the
`
`originally issued ’413 patent.
`
`B.
`
`The claimed step of inserting an interventional cardiology device
`need not be performed only after the coaxial guide catheter is
`inserted into the guide catheter.
`17. Limitation 1.f of claim 1 of the ’413 patent recites “inserting the
`
`interventional cardiology device into and through the continuous lumen of the
`
`standard guide catheter alongside of the substantially rigid portion and advancing
`
`the interventional cardiology device through and beyond a lumen of the flexible tip
`
`portion into contact with or past a lesion in the second artery.” Ex-1001, 11:1-6.
`
`
`4 Reference to Ex-1845 here is with respect to IPR2020-01341 and IPR2020-01342
`
`only. Ex-1845 was previously filed as Ex-1001 in IPR2020-01343 and Ex-1401 in
`
`IPR2020-01344.
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 10 of 63
`
`

`

`
`
`18.
`
`I have reviewed Teleflex’s claim construction arguments, IPR2020-
`
`001341, POR at 9-14, and I disagree.
`
`19. As I have testified, claim 1 does not mandate insertion of an
`
`interventional cardiology device after insertion of a coaxial guide catheter.
`
`Ex-2245, 82:7-83:3. Premounted or preformed devices may be advanced within a
`
`catheter so long as the whole delivery system is de-aired, which guards against the
`
`danger of introducing an air embolism. Id., 94:19-23; see also Ex-1846, 39:7-14,
`
`44:5-14, 46:8-23, 48:21-49:14, 49:23-50:25; Ex-1797, 23:24-26:3.
`
`20.
`
`I am aware of other Teleflex claims in which the inventors were
`
`specific that an interventional cardiology device, such as a treatment catheter, is
`
`inserted into a guide extension catheter only after the guide extension catheter had
`
`already been advanced into the guide catheter. See e.g., Ex-1845, 16:1-5 (reciting
`
`advancement of a treatment catheter through a guide catheter and a guide extension
`
`catheter, and into a coronary artery, while the distal end of the guide extension
`
`catheter is positioned beyond the distal end of the guide catheter). Similar language
`
`is not found in the claims of the ’413 patent.
`
`21. Additional support for the fact that claim 1 does not require sequential
`
`insertion of a guide extension catheter (coaxial guide catheter) and interventional
`
`cardiology device comes from dependent claims 10 and 11:
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 11 of 63
`
`

`

`
`
`10. The method of claim 9, further comprising extending the
`interventional cardiology device through the proximal side opening;
`advancing the interventional cardiology device through structure
`defining a full circumference portion; and advancing the interventional
`cardiology device through structure defining a partially cylindrical
`portion.
`11. The method of claim 9, further comprising extending the
`interventional cardiology device through a flexible cylindrical distal tip
`portion and a flexible cylindrical reinforced portion of the tubular
`structure proximal to the flexible distal tip portion.
`
`22.
`
`If the claims of the ’413 patent required that each step in each claim
`
`be performed sequentially, then claim 10 makes no sense. An interventional
`
`cardiology device cannot be advanced through a full circumference portion of a
`
`coaxial guide catheter’s side opening before it is advanced through a partially
`
`cylindrical portion. And that is because an interventional cardiology device is
`
`advanced through the side opening in a proximal to distal direction, and the
`
`opening’s full circumference portion is more distal than its partially cylindrical
`
`portion. Ex-1001, Figs. 4, 12-16. In other words, an interventional cardiology
`
`device will first pass through the partially cylindrical portion before it reaches the
`
`full circumference portion. This sequence is the opposite of what is recited in claim
`
`10. This strongly suggests that the claim steps need not be performed in the
`
`specific order in which they appear in the claims.
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 12 of 63
`
`

`

`
`
`23. Similarly, for claim 11, an interventional cardiology device will first
`
`advance through a cylindrical reinforced portion before it reaches the flexible
`
`cylindrical distal tip portion. Ex-1001, Fig. 1 (depicting reinforced portion 18 and
`
`tip portion 16).
`
`III.
`
`Itou-based grounds
`I have been asked to assume that Itou is prior art to the claims of the
`24.
`
`Teleflex Patents.
`
`25. As I discuss below, my opinion that the claims of the ’413 and ’116
`
`patents are invalid remains unchanged.
`
`A. Types of interventional cardiology devices insertable through
`catheter (2)
`26. Assuming that the claims of the ’413 patent are construed to require
`
`that two interventional cardiology devices must be insertable through the claimed
`
`lumen, Itou satisfies this limitation. Ex-1005, ¶ 221.
`
`27. Even if Teleflex were correct both as to its claim construction position
`
`and as to the “effective size” of the tubular structure of catheter (2) being 0.046
`
`inches, POR at 45, (and I disagree for the reasons discussed herein), Itou
`
`nonetheless inherently discloses a lumen through which all four, enumerated
`
`interventional cardiology devices (guidewires, balloon catheters, stents and stent
`
`catheters) are insertable.
`
`28.
`
`Itou teaches that guidewire (6) is insertable through the lumen of
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 13 of 63
`
`

`

`
`
`catheter (2). Ex-1007, Fig. 5, 4:64-65, 5:11-20. Guidewires were known to range in
`
`size from 0.010 to 0.018 inches. Ex-1015, 98. They would therefore be insertable
`
`through catheter (2) even if its “effective size” were 0.046 inches.
`
`29. Similarly, even if the “effective size” of the opening to the lumen of
`
`catheter (2) were 0.046 inches, angioplasty catheters were known to be insertable
`
`through catheters with lumen of 0.045 inches. Ex-1009, 4:48-64; Ex-1833, 1.
`
`30. There were also a variety of stents with crossing profiles that were
`
`small enough to be advanced through a lumen of 0.046 inches (1.16 mm) that I
`
`discuss below. Necessarily, each of the stents that are discussed were advanced into
`
`the coronary vasculature on a “stent catheter.”
`
`31. Baim explains that “[a]ll current slotted tube designs are ‘bare
`
`mounted’ on a delivery balloon, with deflated profiles smaller than 0.040-in.
`
`(1mm).” Ex-1015, 189.
`
`32. The 4th edition of the Handbook of Coronary Stents (Ex-1802)
`
`disclosed the following: balloon-expandable Genic® stent with less than 0.9 mm
`
`(0.035 inch) profile, id. at 4, 7; balloon-mounted Lunar stent with 0.0382 inch
`
`profile, id. at 15-16; Spiral Force stent with 0.039 to 0.042 inch profile and
`
`“Runner” stent delivery balloon catheter, id. at 21; and balloon-expandable
`
`Tsunami stent with 0.95 mm (0.038 inch) profile, id. at 25.
`
`33. Similarly, US Pat. No. 6,726,714 taught a stent crimped against a
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 14 of 63
`
`

`

`
`
`balloon with a “relatively small loaded diameter having an OD of 0.44 inches.” Ex-
`
`1803, 4:54-56. And an article by Colombo described several balloon-expandable
`
`stents with crossing profiles well under 0.046 inches, including those with profiles
`
`of 0.99 mm (0.038 inch), 0.93 mm (0.036 inch) and 0.84 mm (0.033 inch). Ex-
`
`1804, Table 1, Fig. 3.
`
`34.
`
`In sum, all of the devices I describe above are insertable through
`
`catheter (2), even if the “effective size” of its opening was 0.046 inches (1.16 mm)
`
`(3.48 Fr). I disagree, however, that the lumen of catheter (2) has an effective size
`
`of 0.046 inches.
`
`35. Teleflex’s argument that Itou’s catheter (2) does not have a lumen that
`
`is 1.5 mm (0.059 inches), as Itou teaches in Table 1, but, instead is effectively only
`
`1.16 mm (0.046 inches) is based on the idea that because Itou’s wire (25) has a
`
`diameter of 0.45 mm (0.004 inches), it partially blocks the opening to lumen (2).
`
`Ex-2138, ¶¶ 173-174.
`
`36. This calculation does not make sense to a practicing interventional
`
`cardiologist. It assumes, first, that catheter (2) is entirely rigid, such that wire (25)
`
`cannot be bent away from the angled proximal opening of the tubular portion of
`
`catheter (2).
`
`37.
`
`Itou explicitly teaches this is not the case, as it discloses that catheter
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 15 of 63
`
`

`

`
`
`(2)—and in particular its distal tubular portion—must be navigated into the
`
`coronary vasculature. Ex-1007, Abstract (“[A]ssembly is insertable into a blood
`
`vessel . . . .”), 1:66-2:5 (teaching that catheter is advanced to a “deep location in a
`
`coronary artery”), Fig. 6, 5:29-42. This means that catheter (2) is necessarily
`
`flexible, including in the region of the junction between the distal end of wire 25
`
`and the proximal part of catheter (2)’s tubular structure. A POSITA understands
`
`that an interventional cardiology device may be introduced into catheter (2) at an
`
`angle.
`
`38. Additionally, Itou teaches that wire (25) is flattened at its distal most
`
`end, “crushed into a form of a flat plate.” Ex-1007, 4:33-36. This is an additional
`
`reason that a POSITA would not understand that wire 25 reduces the effective size
`
`of the opening of the tubular portion of catheter (2). The flattened area adjacent to
`
`the proximal opening—in conjunction with catheter (2)’s flexibility—means that
`
`the entirety of the opening of catheter (2) will be accessible.
`
`B. Distal end protective catheter (5)
`Itou discloses distal end protective catheter (5), which is an
`39.
`
`interventional cardiology device. Ex-1005, ¶¶ 221-225.
`
`40.
`
`Itou describes inserting catheter (5) into catheter (2), and then
`
`advancing the two catheters, together, into guiding catheter (1) and into the
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 16 of 63
`
`

`

`vasculature. Ex-1007, 5:43-46, 7:13-18. Once the assembly was positioned in the
`
`desired location, and as taught by Itou, catheter (5) would be withdrawn before
`
`catheter (2) would be used for suction. Catheter (2) could also be used to deliver a
`
`different interventional cardiology device.
`
`41. Not uncommonly during a procedure, an interventional cardiologist
`
`will find that he needs to advance a catheter more deeply into the vasculature. In
`
`this instance, catheter (2) would be left in place, and catheter (5) would be
`
`reintroduced by insertion into the guide catheter, and advancement into catheter
`
`(2), while catheter (2)’s proximal opening remained in the guide catheter. Only
`
`then would it be safe to maneuver catheter (2) to a deeper location.
`
`42. Sequential delivery would be an obvious alternative to preassembly,
`
`albeit not routine.
`
` Ex-1851, 165:3-25.
`
`
`
`C.
`
`43.
`
`Advancing the interventional cardiology device . . . into contact
`with or past a lesion in the second artery
`I understand that Teleflex argues that Itou does not disclose an
`
`interventional cardiology device that comes “into contact with” a “lesion.” POR at
`
`38-40. I disagree. Ex-1005, ¶¶ 221-225.
`
`44. Dr. Graham states that the target location for Itou’s suction catheter is
`
`“just on the proximal side of the thrombus.” Ex-2145, ¶ 119. He relies on teachings
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 17 of 63
`
`

`

`
`
`in Itou, however, which do not say this. Id., (citing Ex-1007, 7:1-27); see also Ex-
`
`2138, ¶ 156 (citing to the same).
`
`45. As an initial matter, Itou describes that the guiding catheter is
`
`positioned “proximal” to the target location, such as within the coronary ostium.
`
`Ex-1007, 1:66-2:5. This is illustrated in Itou’s Fig. 6, below. Guiding catheter (1) is
`
`positioned in the coronary ostium, proximal of target location (80).
`
`
`
`Ex-1007, Fig. 6 (color added).
`
`46.
`
`Itou explains that the “target location” is located deep in the coronary
`
`artery. Id., Abstract, 1:6-9, 5:35-38. Such a location is depicted in Figure 6 at (80).
`
`47.
`
`Itou additionally teaches that suction catheter (2) and distal end
`
`protective catheter (5) are inserted together to the target location. Ex-1007, 7:13-
`
`18; see also id., 5:43-46. The distal tip of catheter (2) is in contact with the target
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 18 of 63
`
`

`

`
`
`location (80). Furthermore, Itou explains that the distal tip of catheter (5) is
`
`designed to extend distally of the distal opening of catheter (2). Id., Fig. 5, 4:48-55,
`
`Table 1 (describing catheter (2) as having a length of 1250 mm and catheter (5) as
`
`having a length of 1320 mm). Mr. Keith admits this. Ex-2138, ¶ 172. As Itou
`
`explains, the tubular portion of catheter (5) should range from 20-50 mm in length,
`
`so as to adequately provide a protective function. Ex-1007, 6:36-41. Necessarily,
`
`the tip of protective catheter (5) comes into contact with the target location.
`
`48. That the distal ends of both suction catheter (2) and catheter (5) come
`
`into contact with the foreign matter at the target location is not surprising. Dr.
`
`Graham explains that suction catheters are typically designed to extend beyond a
`
`thrombus so as to be able to aspirate it in its entirety. Ex-2145, ¶ 120; see also
`
`Ex-1025 (Kataishi). This is consistent with Itou’s teaching that the distal tips of
`
`catheter (2) and catheter (5) are soft and flexible tip, so as to reduce damage to the
`
`blood vessel. Ex-1007, 3:22-28, 4:48-55. Itou contemplates that the distal ends of
`
`catheter (2) and catheter (5) come into contact with the coronary artery walls at the
`
`target location (80).
`
`49.
`
`Itou is explicit that the target location is one at which there is foreign
`
`matter, “such as a thrombus or an embolus” that cause constriction of the blood
`
`vessel. Id., Abstract, 1:6-9, 5:35-38. This is also depicted in Figure 6 at (80).
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 19 of 63
`
`

`

`
`
`50. Finally, Itou explains that, after catheter (2) is withdrawn, then
`
`catheter (2) is manipulated (“rotated, pushed or pulled”) until its distal tip is “in
`
`front of a thrombus,” as seen in Figure 6. Ex-1007, 7:20-23. This disclosure does
`
`not alter my opinion. By the time catheter (2) is manipulated in this fashion, its
`
`distal end—and the distal tip of catheter (5) —have necessarily already come into
`
`contact with the foreign matter at the target location.
`
`51. Mr. Keith states that “lesion” and “thrombus” are not synonymous.
`
`Ex-2138, ¶ 155. He is partially correct but misunderstands the relationship of one
`
`to the other. They are not synonymous, but the overwhelming majority of inter-
`
`coronary thrombus form on pre-existing atherosclerotic lesion.
`
`52. As taught in Baim, coronary lesions may be characterized according
`
`to their complexity.
`
`Ex-1015, 127.
`
`
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 20 of 63
`
`

`

`
`
`53. While minimally complex lesions do not have thrombus, more
`
`complex lesions do. Stated differently, from the disclosure of Itou, a POSITA
`
`would understand there was a lesion at target location 80 in the coronary artery.
`
`Ex-1007, Fig. 6.
`
`D.
`
`Itou discloses that protective catheter (5) is inserted through a
`guide catheter “alongside of the substantially rigid portion” of the
`coaxial guide catheter.
`54. As I discuss above, claim 1 does not require sequential insertion of a
`
`coaxial guide catheter into a guide catheter, followed by insertion of an
`
`interventional cardiology device.
`
`55. When pre-assembled, wire-like portion (55) of catheter (5) is
`
`“alongside of the substantially rigid portion” of suction catheter (2). See Ex-1005,
`
`¶¶ 221-225.
`
`
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 21 of 63
`
`

`

`
`
`
`
`
`
`
`
`56. Under Mapping-1, the “substantially rigid portion” of catheter (2)
`
`includes both wire-like portion (25) and proximal tip 23. Under Mapping-2, the
`
`“substantially rigid portion” of catheter (2) is wire-like portion (25).
`
`57.
`
`Itou additionally teaches that—following preassembly—suction
`
`catheter (2) and protective catheter (5) are inserted into and through the continuous
`
`lumen of the standard guide catheter. Ex-1007, 5:43-46.
`
`E.
`
`58.
`
`Itou discloses “applying a force to . . . the coaxial guide catheter
`such that the distal portion of the coaxial guide catheter remains
`seated . . . in response to an opposing backward force exerted by
`the interventional cardiology device.”
`I understand that Teleflex argues that Itou does not disclose claim 2.
`
`POR at 40-41. I disagree. Ex-1005, ¶ 226 (citing id., ¶¶ 160-173).
`
`59.
`
`Itou explicitly discloses applying force to the proximal end of suction
`
`catheter (2). Ex-1007, 5:43-46.
`
`60.
`
`Itou also teaches that the distal end of catheter (2) is advanced to a
`
`target location in the coronary artery and is maintained at that location, even as the
`
`distal end of protective catheter (5) extends further into the coronary artery. Ex-
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 22 of 63
`
`

`

`
`
`1007, Figs. 5, 6, 4:48-55, 5:35-51, 6:36-41, Table 1 (describing catheter (2) as
`
`having a length of 1250 mm and catheter (5) as having a length of 1320 mm). The
`
`extension of distal end protective catheter (5) beyond the distal tip of catheter (2)
`
`necessarily generates an opposing backwards force. Ex-1807, ¶¶ 47-51.
`
`61.
`
`I have also opined that it would be obvious to use catheter (2) to
`
`deliver a balloon catheter or stent. In the event that a balloon catheter or stent are
`
`delivered through suction catheter (2), this limitation is also necessarily met.
`
`F.
`
`62.
`
`Itou discloses that protective catheter (5) is extended through a
`“proximal side opening . . . while the proximal portion remains
`within the lumen of the guide catheter.”
`Itou discloses this limitation. Ex-1005, ¶¶ 231-235.
`
`63. Contrary to Patent Owner’s arguments, POR at 41-42—and similar to
`
`claim 1—claim 9 does not require that the proximal portion of the coaxial guide
`
`catheter already be within the guide catheter before an interventional cardiology
`
`device is extended through the coaxial guide catheter’s side opening.
`
`64.
`
`Itou teaches that an assembly of catheter (5) and catheter (2) be
`
`advanced into the coronary artery, and that the proximal end of catheter (2)’s
`
`tubular portion remain within the guide catheter. Ex-1007, 5:35-46, 7:13-18.
`
`65. Prior to its withdrawal, catheter (5) is extended through the proximal
`
`side opening of catheter (2)’s tubular structure, and extends for a distance along the
`
`
`
`IPR2020-01343
`
`Medtronic Ex-1806
`Medtronic v. Teleflex
`Page 23 of 63
`
`

`

`
`
`longitudinal axis of the proximal portion of the tubular structure. Id., Fig. 5, 4:48-
`
`52.
`
`G.
`
`66.
`
`It would have been obvious to insert an interventional cardiology
`device such as a stent or balloon through Itou’s suction catheter
`(2).
`I understand that Teleflex disputes that a POSITA would first insert
`
`suction catheter (2) into a guide catheter, and then separately and sequentially
`
`insert an interventional cardiology device through the suction catheter. POR at 43-
`
`51. I also understand that Teleflex disputes that a POSITA’s use of Itou’s catheter
`
`(2) would be informed by the teachings of Ress

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket