throbber

`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`___________________________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`__________________________
`
`MEDTRONIC, INC., AND MEDTRONIC VASCULAR, INC.
`
`Petitioner,
`
`v.
`
`TELEFLEX LIFE SCIENCES LIMITED,
`
`Patent Owner.
`_____________________________
`
`Case No.: IPR2020-1342
`U.S. Patent No. 8,142,413
`______________________________
`
`PETITIONER’S REPLY
`
`
`
`
`

`

`TABLE OF CONTENTS
`
`B.
`
`C.
`
`Page
`
`TABLE OF CONTENTS .......................................................................................... ii
`TABLE OF AUTHORITIES ................................................................................... iii
`I.
`Introduction ...................................................................................................... 1
`II.
`Claim Construction .......................................................................................... 1
`A.
`“Interventional Cardiology Device(s)” ................................................. 1
`B.
`Claim-Step Order .................................................................................. 2
`III. Kontos’s Extension Catheter is not a Narrow Tube. ....................................... 5
`IV. The Asserted Claims are Obvious. .................................................................. 7
`A.
`Claims 1 & 11: Kontos’s support catheter 10 has a “circular cross-
`sectional inner diameter sized such that interventional cardiology
`devices are insertable.” .......................................................................... 7
`Claims 2 & 12: Kontos provides backup support to assist in resisting
`axial and shear forces exerted by the IVCD. ......................................... 7
`Claims 4, 9-12, and 14: Kontos-Adams combination teaches the
`recited proximal openings. ..................................................................11
`1.
`Replacing Kontos’s funnel with a side opening maximizes the
`usable area in the catheter assembly. ........................................11
`Petitioner’s other motivations are not based in hindsight. ........15
`2.
`Claims 13: Kontos-Adams-Takahashi combination teaches the not-
`more-than-one-French limitation. .......................................................19
`Secondary considerations do not overcome the strong showing of
`obviousness. ...................................................................................................20
`A.
`There is no nexus where Itou and Ressemann disclose every element
`of claims 4, 9, and 14. .........................................................................20
`Teleflex’s secondary considerations evidence does not establish
`copying. ...............................................................................................22
`1.
`Teleflex’s alleged evidence of copying is actually copying of
`the prior art. ...............................................................................22
`2. Medtronic did not copy GuideLiner. ........................................23
`VI. CONCLUSION ..............................................................................................24
`
`V.
`
`D.
`
`B.
`
`ii
`
`

`

`TABLE OF AUTHORITIES
`
`
`
`
`
`Page(s)
`
`Cases
`Altiris, Inc. v. Symantec Corp.,
`318 F.3d 1363 (Fed. Cir. 2003) ............................................................................ 2
`In re Kao,
`639 F.3d 1057 (Fed. Cir. 2011) .............................................................. 20, 22, 23
`Interactive Gift Express, Inc. v. Compuserve, Inc.,
`256 F.3d 1323 (Fed. Cir. 2001) ............................................................................ 2
`Iron Grip Barbell Co. v. USA Sports, Inc.,
`392 F.3d 1317 (Fed. Cir. 2004) .................................................................... 23, 24
`Mantech Environmental Corp. v. Hudson Environmental Servs., Inc.,
`152 F.3d 1368 (Fed. Cir. 1998) ...................................................................passim
`Mformation Tech. v. Research in Motion Ltd.,
`764 F.3d 1392 (Fed. Cir. 2014) ............................................................................ 3
`Ormco Corp. v. Align Tech.,
`463 F.3d 1299 (Fed. Cir. 2006) .......................................................................... 20
`ZUP, LLC v. Nash Mfg., Inc.,
`896 F.3d 1365 (Fed. Cir. 2018) .......................................................................... 22
`
`
`
`iii
`
`

`

`I.
`
`INTRODUCTION
`Patent Owner (“PO”) does not dispute, because it cannot, that Kontos
`
`describes its “support catheter” as a “mini guide catheter.” Ex-1409, 3:40-49. Nor
`
`does PO dispute that Kontos teaches, just like the coaxial guide catheter 12 of the
`
`Teleflex patent, that support catheter 10 includes a short distal lumen (body 12)
`
`coupled to a pushrod (wire 14). Kontos also describes the method of passing the
`
`support catheter “further through the coronary ostia than can guide catheter 38” to
`
`function as a “guide extension catheter” when “extending beyond the distal end of
`
`guide catheter 38” to help prevent a PTCA balloon catheter from bending,
`
`buckling, or kinking en route to treat a lesion. Other than the side opening, Kontos
`
`teaches each structural limitation of the method claims. But as explained herein,
`
`the use of a side opening was an obvious modification from several pieces of prior
`
`art—including Ressemann, which the Board already found anticipated similar
`
`claims in other IPRs. The challenged claims in this IPR are likewise invalid.
`
`II. CLAIM CONSTRUCTION
`“Interventional Cardiology Device(s)”
`A.
`At institution, “interventional cardiology devices” was construed to refer to
`
`“at least two types of the devices selected from the group that includes, but is not
`
`limited to, guidewires, balloon catheters, stents, and stent catheters.” Institution
`
`Decision (I.D.), Paper 9, 16. Patent Owner’s Response does not dispute this
`
`construction. Thus, the Board should adopt its previous construction.
`
`1
`
`

`

`B. Claim-Step Order
`The parties dispute whether steps in claim 1 must be performed in the
`
`sequence recited. As to limitation 1.f.i, the issue is whether the coaxial guide
`
`catheter must already have been inserted into the standard guide catheter (“GC”)
`
`before an interventional cardiology device (“IVCD”) is inserted thereto. The
`
`answer is no. Claim 1 allows for either preassembly of a coaxial guide and IVCD,
`
`or sequential insertion of the two.
`
`“Unless the steps of a method actually recite an order, the steps are not
`
`ordinarily construed to require one.” Interactive Gift Express, Inc. v. Compuserve,
`
`Inc., 256 F.3d 1323, 1342 (Fed. Cir. 2001). PO argues that claim 1 “only makes
`
`sense if the steps take place in order,” POR, 10, but that is only the case for a
`
`subset of steps. The parties agree that a GC must be inserted into a first artery (1.a)
`
`before its distal end can be positioned in a branch artery (1.b). Similarly, a GC
`
`must already be positioned before the coaxial guide extension catheter can be
`
`inserted thereto (1.c-1.e). As discussed in the Petition, however, and below, the
`
`plain language of the claims permits an IVCD to be advanced either
`
`simultaneously with the guide extension catheter or sequentially. Ex-1405, ¶¶200-
`
`205; Ex-1806, ¶¶17-23; Petition (Pet.), Paper 1, 49-52.
`
`The specification’s disclosure of an embodiment in which insertion is
`
`sequential does not limit the claims. Altiris, Inc. v. Symantec Corp., 318 F.3d 1363,
`
`2
`
`

`

`1369-1371 (Fed. Cir. 2003) (finding error where method steps were construed to
`
`require the order used by the sole embodiment). This is not a case in which
`
`sequential insertion is required in order to perform the claimed method. See e.g.
`
`Mformation Tech. v. Research in Motion Ltd., 764 F.3d 1392, 1398-1400 (Fed. Cir.
`
`2014) (agreeing that claims to remotely deploying software require connection
`
`before transmission); Mantech Environmental Corp. v. Hudson Environmental
`
`Servs., Inc., 152 F.3d 1368, 1375-76 (Fed. Cir. 1998) (determining that claims for
`
`remediating groundwater contamination require the provision of wells, before acid
`
`is introduced thereto).
`
`First, PO argues there can be no “backup support” for an IVCD if there is
`
`preassembly. Patent Owner Response (POR), Paper 24, 16. Even assuming the
`
`preamble is limiting, that is nonsensical. Regardless of whether the IVCD is
`
`inserted simultaneously with—or after—the coaxial guide extension catheter, the
`
`result is the same. Before the IVCD is ever advanced into the coronary artery the
`
`guide extension catheter is already in position within the GC. And the patent
`
`teaches that it is the relative orientation of the GC/guide extension catheter—at the
`
`time the IVCD is being advanced into the coronary artery— that provides backup
`
`support. Ex-1401, Fig. 9, 8:4-9.
`
`Second, step 1.f.i requires that two things be positioned “alongside” each
`
`other: (a) the IVCD; and (b) the extension catheter’s substantially rigid portion.
`
`3
`
`

`

`“Alongside” is an adjective that accurately reflects the relative positions of (a) and
`
`(b), even where the two are preassembled. Teleflex’s experts do not disagree. Ex-
`
`1797, 16:6-17:10; 31:9-23; Ex-1846, 49:23-50:25.
`
`Finally, step 1.f.ii requires that the IVCD be advanced “through and beyond”
`
`the lumen of the extension catheter. Testimony from both PO’s experts confirm
`
`that—following preassembly— an IVCD can be inserted into a guide catheter and
`
`then advanced into the coronary artery. Id. Dr. Brecker concurs, Ex-1405, ¶¶200-
`
`05; Ex-1806, ¶¶99-114, and real-world evidence bears this out. Ex-1848, Abstract,
`
`[0204-0207], Figs. 23a-23c (disclosing preassembly and advancement of a
`
`balloon); Ex-1849, Figs. 2C, 4 (disclosing the same).
`
`Accordingly, the default rule that method steps need not be performed in the
`
`order recited applies to claim 1 of the ’413 patent.1
`
`
`1 Claims depending from claim 1 confirm that the ’413 patent does not require
`
`every step to be performed in the order in which it is recited. As written, claims 10
`
`and 11 would require an IVCD to move in a distal to proximal direction
`
`(advancing first through the full circumference portion of a side opening, and then
`
`through its partially cylindrical portion, or advancing through a flexible tip before
`
`advancing through a reinforced portion). Ex-1401, 12:14-25. Neither scenario
`
`makes sense. In use, an IVCD is advanced in the opposite order, in a proximal to
`
`4
`
`

`

`III. KONTOS’S EXTENSION CATHETER IS NOT A NARROW TUBE.
`PO mischaracterizes Kontos and argues that body 12 must be a “narrow
`
`tube” that “make it more akin to a ‘wet noodle’” and “incapable of resisting
`
`backup forces in the manner required by claims 2 and 7.” POR, 25. PO is wrong.
`
`Ex-1806 ¶¶110-11; Ex-1807 ¶109. Kontos never states that it is “important” for
`
`body 12 to be “narrow” or have a “snug fit.” Ex-1806 ¶¶100-04; Ex-1801, 52:5-8
`
`(language is silent concerning “snugness”); Ex-1800, 77:6-11, 78:12-20. In fact,
`
`Kontos teaches the opposite, explaining that the size of body 12 should be suitable
`
`“for existing PTCA catheters” and that “[o]f course, other sizes may be used for
`
`other applications.” Ex-1409, 4:46-48, 4:61-5:2.
`
`PO’s argument is also contradicted by how Kontos actually prevents kinking
`
`of the PTCA catheter. Ex-1806 ¶103-04; Ex-1807 ¶¶109, 117. Kontos prevents
`
`kinking by reducing the distance between the distal-most portion of the catheter
`
`assembly and the occlusion. Ex-1806 ¶103; Ex-1807 ¶¶117. “[B]ody 12 functions
`
`as a guide catheter extension,” by reducing “the gap that PTCA catheter 40 must
`
`negotiate without assistance …, [which] lessens considerably the tendency of the
`
`PTCA catheter 40 to bend, buckle or kink.” Ex-1409, 5:49-56. This is because, by
`
`comparison to advancement within the vasculature, significantly less force is
`
`
`distal direction. Pet., 7-8, 11.
`
`5
`
`

`

`needed to advance the interventional cardiology device (“IVCD”) within body 12
`
`due to its lubricious inner coating. Ex-1806 ¶103; Ex-1807 ¶¶117; Ex-1813, 52:14-
`
`53:8, 72:24-73:14. It is Kontos’s ability to shorten the distance that the IVCD
`
`traverses in the vasculature—especially in regions that are not straight or heavily
`
`calcified—and not the spatial relationship with the IVCD that reduces the
`
`likelihood of kinking. Ex-1806 ¶103; Ex-1807 ¶¶117.
`
`This conclusion is buttressed by the real-world experience of PO’s expert.
`
`Dr. Graham testified that he has never had an IVCD kink within a catheter
`
`assembly. Ex-1813, 52:14-53:8. In other words, the closeness of support assembly
`
`10 and the PTCA catheter 40 is irrelevant, as kinking (if it does occur) would not
`
`happen until the IVCD is in the vasculature. Ex-1806 ¶104.
`
`Not only is a close-fitting tube unnecessary to prevent kinking of the PTCA
`
`catheter, but, as Dr. Brecker explains, it would be detrimental. Ex-1806 ¶102. If the
`
`relationship is too “snug,” the IVCD can be damaged when loaded into the
`
`extension catheter, especially post-expansion, when the PTCA balloon is retracted
`
`proximally into the extension catheter. Id.
`
`Finally, even if Kontos taught a small body 12 with a “snug fit” to the
`
`IVCD, PO’s argument still fails because it is premised on Kontos’s invention being
`
`limited to fixed-wire catheters. Ex-1800, 76:2-8, 80:12-81:20; Ex-1801, 56:1-10,
`
`60:14-21. Kontos is not so limited, and instead provides that “the device of this
`
`6
`
`

`

`present invention may be used with almost any type of catheter, including over-
`
`the-wire catheters.” Ex-1409, 9:47-50; Ex-1806 ¶105. Kontos teaches what it
`
`teaches and is not limited to fixed wire catheters.
`
`IV. THE ASSERTED CLAIMS ARE OBVIOUS.
`A. Claims 1 & 11: Kontos’s support catheter 10 has a “circular
`cross-sectional inner diameter sized such that interventional
`cardiology devices are insertable.”
`PO does not dispute that Kontos’s support catheter 10 is sized to receive at
`
`least two IVCDs. See I.D., 26-27. As discussed above, PO incorrectly alleges that
`
`the claims require a specific order of operations. They do not and are invalid.
`
`B. Claims 2 & 12: Kontos provides backup support to assist in
`resisting axial and shear forces exerted by the IVCD.
`PO does not dispute that Kontos’s support catheter extends beyond the distal
`
`end of the guide catheter and past the ostium to assist delivery of interventional
`
`cardiology devices to lesions and other parts of the vasculature. Ex-1797, 73:10-
`
`74:2; POR, 22. Instead, PO asserts that Petitioner failed to show that Kontos
`
`inherently teaches the recited back-up functionality because Kontos “says nothing
`
`about exerting a force on a proximal portion of body 12.” POR, 22. But the
`
`structural characteristics of Kontos—which PO does not dispute, and which the
`
`Board has already found invalidated similar claims in a separate IPR—provide
`
`back-up support in two ways: (i) shortening the distance that the IVCD must travel
`
`within the vasculature and (ii) by increasing the moment of inertia of the catheter-
`
`7
`
`

`

`in-catheter assembly. Ex-1806 ¶107-11; Ex-1807 ¶¶14-27, 106-12.
`
`Kontos’s support catheter 10 shortens the distance that the PTCA catheter
`
`must travel while in contact with the tortuous and potentially calcified vasculature.
`
`Ex-1806 ¶108; Ex-1807 ¶¶107, 109, 117. This type of catheter-in-catheter
`
`assembly is a form of “deep seating.”2 Ex-1813, 75:4-17. And as admitted by PO’s
`
`expert, deep seating provides back-up support. Id., 76:4-6. More particularly, a
`
`physician must apply less force to advance an IVCD within the catheter assembly
`
`(by comparison to advancing within the vasculature), and thus, provides back-up
`
`support. See also Ex-1797, 73:10-74:2 (“So in Kontos, the physician certainly does
`
`apply a force to the proximal portion of the coaxial guide catheter in order to
`
`advance it through the guide catheter and to advance it out into the vasculature out
`
`to an area to be treated; right? A. Yes, during those parts of the procedure.”).
`
`Further, Kontos provides back-up support by making what is essentially a
`
`“double-catheter.” In so doing, the nested assembly increases the (i) flexural
`
`rigidity, (ii) torsional rigidity, and (iii) resistance to buckling forces of the catheter-
`
`
`2 According to PO’s expert, there are two types of deep seating: (i) extending the
`
`GC past the ostium and into the coronary artery or (ii) positioning the GC at the
`
`ostium and extending the child catheter therein to into the coronary artery.
`
`Ex-1813, 74:21-75:17; Ex-2145 ¶¶60-75.
`
`8
`
`

`

`in-catheter assembly. Ex-1807 ¶¶22-27, 106-08. This in turn increases the moment
`
`of inertia, which results in increased back-up support. Ex-1806 ¶162; Ex-1807
`
`¶¶22-27, 106-08.
`
`PO nonetheless argues that insufficient back-up support is provided because
`
`tube 16 is (i) too flexible and without reinforcement and (ii) positioned with a gap
`
`between its outer wall and the inner wall of the GC. POR, 25. PO is wrong to
`
`characterize Kontos as too flexible, and, as admitted by PO’s expert, a small gap—
`
`such as the 1 French relationship recited in certain claims—is not necessary for
`
`back-up support.
`
`PO argues that tube 16 is “more akin to a ‘wet noodle’” that would “rotate,
`
`bend, or ‘crunch up’” before providing back-up support. POR, 24-25. But as
`
`admitted by PO’s expert, Kontos never says to use “the most pliable molded plastic
`
`from among the range of choices,” and, a POSITA “would expect to pick the
`
`appropriate molded plastic material for the application.” Ex-1800, 69:20-70:13;
`
`Ex-1409, 4:1-4. Further, it makes no sense for tube 16 to be as flexible as PO
`
`suggests, otherwise Kontos could not function for its stated purpose to “permit[] a
`
`physician to deliver, with greatly reduced risk of bending or kinking, a PTCA
`
`catheter to a site of coronary vessel restriction.” Id., 2:22-25; Ex-1806 ¶¶110-12.
`
`Indeed, Kontos contemplates an application that foregoes a GC and uses just
`
`support catheter 10 with PTCA catheter 40. Ex-1409, 8:6-18. Kontos also discloses
`
`9
`
`

`

`use of support assembly 10 as a temporary stent. Id., 6:59-7:5. Foregoing the use of
`
`a GC or using as a temporary stent clearly would not be possible if support
`
`assembly 10 was a wet noodle. Ex-1806 ¶¶110-12.
`
`Kontos’s tube is made of a “molded plastic material, such as polyethylene.”
`
`Ex-1409, 4:1-4. Polyethylene is manufactured in numerous grades and can have
`
`rigid properties (even without metallic reinforcement). Ex-1807 ¶109; see also
`
`Ex-1851, 66:14-16. Finally, PO suggests that the length of tube 16 (~12 inches)
`
`precludes back-up support. POR, 25. PO is wrong, Ex-1807 ¶¶153, 157, and
`
`certain versions of PO’s product had a similar sized 40 cm (<16 inch) tubular
`
`portion. Ex-2138 ¶84.
`
`It was also known that back-up support could be provided without a 1
`
`French relationship between the GC and extension catheter. Ex-1806 ¶¶112-13;
`
`Ex-1807 ¶¶16, 18-19; see also Ex-1814, 5. Kontos’s tube 16 has an outer diameter
`
`of 0.055 inches (Ex-1409, 4:48-50), and, according to PO, could fit in a 6 French
`
`GC, which has an inner diameter of 0.070 inches. Ex-1806 ¶¶112-13. This
`
`difference, referred to as the annular gap, is 0.015 inches. Id. And a 0.015 inch
`
`annual gap, as taught by the Teleflex patent and as admitted by PO’s experts,
`
`provides back-up support. Id.; Ex-2138 ¶61 (showing four-in-six provides back-up
`
`support); Ex-1801, 38:19-20; Ex-1817, 71:23-72:14. Thus, Kontos provides back-
`
`up support even though it does not teach a 1 French relationship. Ex-1806 ¶¶112-
`
`10
`
`

`

`14.
`
`Finally, PO’s argument fails because it misinterprets the back-up claims.
`
`Importantly, PO’s expert admitted that he based his opinion on Kontos failing to
`
`provide “clinically relevant” back-up support. Ex-1801, 63:19-64:19. The back-up
`
`claims, though, only require that the device assists in resisting axial and shear
`
`forces; the device does not have to wholesale prevent axial and shear forces from
`
`exerting backward force. Ex-1806 ¶114. Thus, any contribution by Kontos to resist
`
`such forces will satisfy the back-up claims.
`
`C. Claims 4, 9-12, and 14: Kontos-Adams combination teaches the
`recited proximal openings.
`A POSITA was motivated, with a reasonable expectation of success, to
`
`replace Kontos’s proximal funnel with a side opening. Ex-1405 ¶¶90-101. PO’s
`
`argument assumes a POSITA would make no further modifications to Kontos’s
`
`support assembly 10 after replacing the proximal funnel with a side opening. Even
`
`if a POSITA made no further modifications, though, PO’s argument still fails.
`
`1.
`
`Replacing Kontos’s funnel with a side opening maximizes
`the usable area in the catheter assembly.
`Replacing Kontos’s proximal funnel with a side opening maximizes the
`
`usable real estate within the catheter assembly. Ex-1806 ¶115. This is particularly
`
`important given that catheter assemblies used in PCI procedures are “not much
`
`bigger than a toothpick.” Ex-2137, 199:11-200:16. PO’s expert agrees that
`
`11
`
`

`

`“maximizing the usable real estate inside a guide catheter is extremely important.”
`
`Ex-2145 ¶90; Ex-1801, 102:2-8; Ex-1813, 91:18-92:5. Specifically, PO’s expert
`
`testified that it was “important” to “maximize” the “inner diameter of the extension
`
`catheter without having to increase the outer diameter of the guide catheter.” Ex-
`
`1813, 91:18-92:5; Ex-2138 ¶266. And PO’s expert agreed that the use of a
`
`proximal funnel does not maximize the useable “real estate” within the catheter
`
`assembly. Ex-1813, 92:19-24.
`
`
`
`
`
` Ex-1851, 296:17-297:1.
`
`a. Maximizing usable real estate: reducing diameter of
`GC.
`Replacing Kontos’s funnel with a side opening reduces the outer diameter of
`
`the extension catheter, which permits the use of a smaller-diameter GC.
`
`Ex-1405 ¶211. PO does not disagree, but instead argues that a POSITA would not
`
`replace the funnel with a side opening because Kontos was already sized to fit
`
`
`
`12
`
`

`

`inside a 6 French GC and that a 5 French GC offered no advantage. POR, 37.
`
`Kontos has a 0.065 inch outer diameter at base portion 18 (Fig. 3), meaning that to
`
`fit inside a 6 French GC having a 0.070 inch inner diameter, the funnel would only
`
`have a 0.005 inch profile. Ex-1807 ¶¶120-21. This would provide no—or
`
`significantly limit the—funneling function. Ex-1807 ¶121. Further, even at 0.005
`
`inches, the funnel would “rub” in a 6 French GC, which, as admitted by PO’s
`
`expert, “would hinder the ability to facilitate smooth passage.” Ex-1801, 115:14-
`
`117:7. Thus, replacing Kontos’s funnel with a side opening would improve the
`
`trackability within the GC. Ex-1806 ¶124. Further, a POSITA was motivated to use
`
`a side opening because it would more easily permit compatibility with a 5 French
`
`GC, which has advantages over a 6 French GC.3 Id. ¶125.
`
`b. Maximizing usable real estate: increasing diameter of
`extension catheter.
`The corollary is also true: a POSITA would be motivated to replace the
`
`funnel with a side opening because the inner diameter of body 12 could then be
`
`increased without similarly increasing the diameter of the GC. Ex-1405 ¶¶163-64.
`
`
`3 For compatibility with a 5 French GC, tube 16 would be molded onto a tapered
`
`wire 14. § IV.E, infra.
`
`13
`
`

`

`
`
`Ex-1405 ¶164. A POSITA was motivated to make this modification, as PO’s
`
`expert acknowledged, because “the larger the available area within a guiding
`
`device, the larger the variety of types, numbers and sizes of therapeutic devices
`
`that can be delivered.” Ex-2145 ¶140. Doing so, “ultimately improves the available
`
`treatment options for the patient.” Id.; Ex-1813, 132:22-133:10.
`
`Despite its own expert agreeing to the advantages of increasing the diameter
`
`of the extension catheter without causing a commensurate increase in the GC, PO
`
`argues that a POSITA would not apply this teaching because Kontos requires a
`
`“snug” fit. POR, 28. But as discussed above, Kontos does not mandate such a tight
`
`fit. § III, supra.
`
`PO also argues that it would not be possible to increase the diameter of body
`
`12 without also increasing the diameter of the GC due to Kontos’s raised marker
`
`bands. POR, 27. Kontos teaches, however, that “[o]f course, numerous other
`
`methods for disposing marker band 30 at distal end 24 will be readily apparent to
`
`14
`
`

`

`those skilled in the art.” Ex-1409, 4:21-24. Embedded marker bands were routine
`
`in the art. Ex-1806 ¶119; Ex-1762, 56:3-23. PO’s expert agreed. Ex-1801, 66:19-
`
`67:6, 70:5-17, 75:15-19. It was well within the skill of a POSITA to recess the
`
`marker bands. Ex-1807 ¶¶128-36. PO cannot defeat Petitioners showing that a
`
`POSITA would, based on the teachings of Adams,4 replace Kontos’s funnel with a
`
`side opening.
`
`Petitioner’s other motivations are not based in hindsight.
`2.
`As set forth in the Petition and as taught by Ressemann, side openings
`
`“facilitate smooth[] passage of other therapeutic devices” into the lumen of the
`
`child catheter. Ex-1408, 6:52-57. PO tries to explain away this teaching by noting
`
`Ressemann’s alternative embodiments use a flare or reverse bevel. POR, 28-29.
`
`PO’s expert, though, acknowledged that Ressemann’s primary embodiment does
`
`
`4 Contrary to PO’s argument, POR, 31-34, Adams teaches that the proximal
`
`opening of the guide seal (extension catheter) receives IVCDs. Ex-1435, [0048]
`
`(“The guide seal is open at its distal end to provide for the passage of another
`
`catheter or a distal protection element.”), [0061] (“optionally remov[ing]” guide
`
`seal). Regardless of when the guide seal is removed, Adams teaches that the distal
`
`protection device 15—which is an IVCD (Ex-1801, 85:14-87:2, 121:21-122:3;
`
`Ex-1806 ¶8 n.2)—passes through the proximal opening. Ex-1435, [0064], Fig. 2E.
`
`15
`
`

`

`not use a flare or reverse bevel. Ex-1801, 107:1-108:4.
`
`Moreover, PO’s argument is contradicted by a patent (“Keith”) issued to its
`
`own expert, which contains Ressemann’s verbatim disclosure about side openings
`
`“facilitate[ing] smoother passage” of IVCDs. Ex-1123, 7:54-60; Ex-1800,
`
`149:3-10. Unlike Ressemann, the Keith patent contains no disclosure of a flare or
`
`funnel in any embodiment (whether primary or alternative). Ex-1800, 149:15-18.
`
`Side openings also allow for smoother passage of the catheter assembly as it
`
`navigates the vasculature. Pet., 58. PO argues, though, that this teaching is
`
`“unsupported by the evidence.” POR, 39. PO is wrong. Not only does Ressemann
`
`explicitly state that a side opening “allow[s] for smoother passage of the
`
`evacuation sheath assembly 100 through a guide catheter,” but an identical
`
`teaching is also recited in Keith. Ex-1123, 7:54-60.
`
`
`
`
`
` Ex-1819 ¶112.
`
`Finally, a side opening permits smooth re-entry when the extension catheter
`
`is extended beyond the distal end of the GC. Pet., 58-59. PO’s expert previously
`
`found himself in that situation (i.e., passed an extension catheter entirely outside of
`
`16
`
`

`

`the GC),5 Ex-1801, 72:10-73:18, and agreed that a proximal opening, as opposed
`
`to Kontos’s funnel, permits smoother re-entry of the extension catheter. Ex-1801,
`
`79:14-21; Ex-1800, 151:17-152:1.
`
`Despite the clear teachings for why a POSITA would replace Kontos’s
`
`funnel with a side opening, PO argues that doing so would actually cause
`
`problems. POR, 26-31. First, PO’s argument fails because it is based on a false
`
`premise that a POSITA would not maximize the usable real estate in the catheter
`
`assembly. As explained above, a POSITA would have done so. § IV.C.1, supra.
`
`Even assuming a POSITA made no further modifications after replacing the funnel
`
`with a side opening, Kontos would not create a “problem gap” as alleged by PO.
`
`POR, 26-30. If Kontos was used with a 6 French GC, the gap between the support
`
`assembly 10 and the inner diameter of the GC would be 0.005 inches. Given that a
`
`guidewire or the distal-most wire of a fixed-wire balloon typically have a 0.014
`
`inch diameter, the PTCA catheter would not catch in the gap between tube 16 and
`
`the inner wall of the GC. Ex-1806 ¶¶120-25.
`
`
`5 PO is incorrect to argue that “Kontos teaches that generally it is not desirable to
`
`extend the proximal end of body 12 beyond the distal end of the guide catheter.”
`
`PO ignores Figure 7, which teaches the intentional advancement into the
`
`vasculature. Ex-1409, 6:19-28.
`
`17
`
`

`

`Ex-1409, Fig. 1 (modified by Petitioner).
`
`Ressemann and Keith buttress this conclusion. Ex-1806 ¶124. Both
`
`references teach side openings that are “preferably angled” to “facilitate smoother
`
`passage of other therapeutic devices,” and both teach a gap of at least 0.009 inches
`
`as shown below. Ex-1806 ¶125. This is almost double Kontos’s alleged
`
`“problematic” gap. Id.
`
`Ex-1123, Fig. 1A (annotations added). For these additional reasons, a POSITA
`
`would replace Kontos’s funnel with a side opening. Ex-1806 ¶125.
`
`18
`
`
`
`
`
`

`

`D. Claims 13: Kontos-Adams-Takahashi combination teaches the
`not-more-than-one-French limitation.
`PO does not dispute that Takahashi provides motivation to achieve a 1
`
`French differential to increase back-up support. Ex-1801, 117:14-118:5. PO argues
`
`that this motivation is inapplicable, though, because Kontos “teaches away from
`
`maximizing the inner diameter.” POR, 43. As discussed above, PO is incorrect to
`
`argue that Kontos requires a snug fitting body 12. § III, supra. Thus, a POSITA
`
`was motivated to apply the teachings of Takahashi to increase back-up support.
`
`Ex-1806 ¶¶129-32.
`
`PO also argues that a POSITA would not have an expectation of success to
`
`modify Kontos to achieve the 1 French differential. POR, 42-43. PO is wrong. A
`
`POSITA knew how to (i) replace the proximal funnel with a side opening and (ii)
`
`recess Kontos’s distal marker bands. § IV.C.1.b, supra. Further, Kontos explicitly
`
`teaches that “tube 16 may be molded directly only application wire 14.” Ex-1409,
`
`4:31-32. In so doing, to maximize the usable “real estate” in the catheter assembly,
`
`a POSITA would taper the pushrod for attachment onto the Kontos-Ressemann
`
`combination. Ex-1807 ¶¶131-34. These modifications permit the diameter of tube
`
`16 to be increased (Ex-1409, 4:46-50), and Kontos would achieve the not-more-
`
`than-one-French differential. Ex-1806 ¶¶130-32.
`
`19
`
`

`

`
`
`Ex-1409, Fig. 1 (modified by Petitioner).
`
`V.
`
`SECONDARY CONSIDERATIONS DO NOT OVERCOME THE
`STRONG SHOWING OF OBVIOUSNESS.
`Teleflex bases its secondary considerations arguments on features known in
`
`the prior art. Teleflex argues that the combination of rapid exchange (“RX”), guide
`
`extension, improved backup support, and a side opening drives GuideLiner’s
`
`success. POR, 65-66. This combination is entirely taught by both Itou and
`
`Ressemann.
`
`A. There is no nexus where Itou and Ressemann disclose every
`element of claims 4, 9, and 14.
`Medtronic may rebut Teleflex’s presumption of nexus by demonstrating that
`
`the secondary indicia result from features known in the prior art. In re Kao, 639
`
`F.3d 1057, 1068 (Fed. Cir. 2011). “[I]f the feature that creates the [secondary
`
`consideration] was known in the prior art, the success is not pertinent.” Ormco
`
`Corp. v. Align Tech., 463 F.3d 1299, 1312 (Fed. Cir. 2006).
`
`It is undisputed that RX GECs offering improved backup support and a side
`
`opening existed in the prior art before Root’s conception of the idea claimed in
`
`20
`
`

`

`Teleflex’s patents. Compare Ex-1755 ¶¶27-38 (IPR2020-01341) with Ex-2118 ¶5.
`
`Both Itou and Ressemann disclose the features recited in claims 4, 9, and 14. Pet.,
`
`52-55, 59-60, 67-70, 83, 88-90, 93-95 (IPR2020-01341). Itou and Ressemann are
`
`RX GECs with a side opening that increase backup support when an IVCD is
`
`extended beyond the distal end of the catheter assembly. Ex-1806 ¶¶144-58.
`
`Teleflex does not dispute that Itou recites the necessary structure, despite
`
`being a suction catheter. POR, 41-42 (IPR2020-01341). Itou discloses a
`
`substantially rigid portion that comprises a partially cylindrical portion defining a
`
`side opening. Ex-1407, Figs. 3-4; see, e.g., Ex-1005 ¶227 (IPR2020-01341). Itou
`
`teaches that a protective catheter may be inserted into the lumen through the side
`
`opening and projected from its distal end, and delivered to the target location. Ex-
`
`1407, 4:48-52, 7:1-27, Fig. 5; see, e.g., Ex-1005 ¶¶195, 231-32 (IPR2020-01341).
`
`I

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket