throbber
IPR2020-01341, -01342
`Patent 8,142,413
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`
`
`
`
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`
`
`
`
`
`MEDTRONIC, INC., AND MEDTRONIC VASCULAR, INC.,
`
`Petitioner,
`
`v.
`
`TELEFLEX LIFE SCIENCES LIMITED,
`
`Patent Owner.
`
`
`
`
`Case No. IPR2020-01341
`Case No. IPR2020-01342
`U.S. Patent No. 8,142,413
`
`
`
`
`
`
`PETITIONER’S OPPOSITION TO
`PATENT OWNER’S CONTINGENT MOTION TO AMEND
`
`
`
`
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`
`TABLE OF CONTENTS
`
`I.
`II.
`
`B.
`
`INTRODUCTION ........................................................................................... 1
`PROPOSED CLAIMS 15-17 ARE UNPATENTABLE. ............................... 1
`A.
`Substitute claims 15-17 are unpatentable over Itou in view of
`Ressemann or Kataishi. ......................................................................... 1
`1.
`Substitute Claim 15 ..................................................................... 2
`2.
`Substitute Claim 16 ..................................................................... 6
`3.
`Substitute Claim 17 ..................................................................... 7
`Substitute claims 15-17 are unpatentable over Kontos in view of
`Ressemann and Takahashi and/or Kataishi. ........................................12
`1.
`Substitute claims 15-17 are unpatentable over Kontos in
`view of Ressemann and Takahashi. ..........................................13
`Substitute claim 17 is unpatentable over Kontos in view of
`Ressemann, Takahashi, and Kataishi. .......................................18
`SECONDARY CONSIDERATIONS DO NOT OVERCOME THE
`STRONG SHOWING OF OBVIOUSNESS. ...............................................19
`A.
`Teleflex is not entitled to a presumption of nexus. .............................19
`B.
`Teleflex’s secondary consideration evidence all relates to prior
`art features and functionality. ..............................................................22
`1.
`Teleflex’s alleged evidence of copying is actually copying
`of the prior art. ..........................................................................23
`2. Medtronic did not copy GuideLiner. ........................................24
`IV. CONCLUSION ..............................................................................................25
`
`
`2.
`
`III.
`
`
`
`i
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`
`TABLE OF AUTHORITIES
`
` Page(s)
`
`Cases
`Fox Factory, Inc. v. SRAM, LLC,
`944 F.3d 1366 (Fed. Cir. 2019) .......................................................................... 19
`In re Kao,
`639 F.3d 1057 (Fed. Cir. 2011) .............................................................. 23, 24, 25
`Interval Licensing LLC v. AOL, Inc.,
`766 F.3d 1364 (Fed. Cir. 2014) .......................................................................... 21
`Iron Barbell Co. v. USA Sports, Inc.,
`392 F.3d 1317 (Fed. Cir. 2004) .................................................................... 24, 25
`KSR Int’l Co. v. Teleflex, Inc.,
`550 U.S. 398 (2007) ............................................................................ 9, 12, 13, 18
`Tyco Healthcare Grp. LP v. Ethicon Endo-Surgery, Inc.,
`774 F.3d 968 (Fed. Cir. 2014) .............................................................................. 3
`WBIP, LLC v. Kohler Co.,
`829 F.3d 1317 (Fed. Cir. 2016) .......................................................................... 19
`ZUP, LLC v. Nash Mfg., Inc.,
`896 F.3d 1365 (Fed. Cir. 2018) .......................................................................... 23
`
`
`ii
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`INTRODUCTION
`I.
`Medtronic, Inc., and Medtronic Vascular, Inc., (collectively “Petitioner” or
`
`“Medtronic”) submits this opposition to Patent Owner’s Contingent Motion to
`
`Amend (MTA).1 Patent Owner (“Teleflex”) seeks to amend claims 1, 7, and 8 and
`
`proposes substitute claims 15-17. MTA, 1, Appendix A (“App.”). But the substitute
`
`claims are unpatentable over the prior art. Teleflex’s motion should be denied.
`
`II.
`
`PROPOSED CLAIMS 15-17 ARE UNPATENTABLE.
`Substitute claims 15-17 are unpatentable over Itou in view of
`A.
`Ressemann or Kataishi.
`Unlike original claim 1, substitute claim 15 requires each step of the claimed
`
`method to be performed in the order recited. MTA, App. It also recites a “treatment
`
`catheter that includes a stent,” instead of an “interventional cardiology device.”
`
`Additional limitations added to the substitute claims relate to a size differential
`
`between the lumen of the guide catheter and the lumen of the coaxial guide catheter
`
`(claim 15), or recite specific sizes for each (claim 16). Limitations added to substitute
`
`claim 17 add limitations to the shape of the coaxial guide catheter’s side opening.
`
`But these additions cannot overcome the prior art of record. The analysis below
`
`focuses on the newly added limitations while briefly addressing the original
`
`
`1 Papers or Exhibits cited herein are in Case No. IPR2020-01341 unless indicated
`
`otherwise. Counterpart exhibits beginning Ex-14XX are filed in IPR2020-01342.
`
`1
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`limitations, which are thoroughly addressed by the original Petition and supporting
`
`testimonial evidence.
`
`Substitute Claim 15
`1.
`First, even if the preamble of substitute claim 15 is construed to be limiting,
`
`see Petition (Pet.), 21 n.6, Itou renders the claim obvious. Itou discloses providing
`
`backup support. A POSITA knew that to advance an interventional cardiology
`
`device through a guide catheter (“GC”) and into the coronary vasculature, the GC
`
`had to provide sufficient backup support. Ex-1015, 548; Ex-1847 ¶17. The ’413
`
`patent similarly teaches that because the disclosed coaxial guide catheter is
`
`“extended through the lumen of the guide catheter and beyond the distal end of the
`
`guide catheter and inserted into the branch artery,” it “assists in resisting axial and
`
`shear forces exerted by an interventional cardiology device passed through the
`
`second lumen and beyond the flexible distal tip portion.” Ex-1001, Abstract, 4:64-
`
`5:9. As the ’413 patent makes clear, it is the combination of a GC and an extension
`
`catheter inserted into the coronary ostium that provides the claimed backup support.
`
`Pet., 22. The exact same configuration of catheters is disclosed in Itou.2 Id., 23-24;
`
`Ex-1847 ¶¶13-24.
`
`
`2 As explained in Petitioner’s Reply in IPR2020-01341 (filed concurrently
`
`herewith), Itou is prior art.
`
`2
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`Second, adding the limitation “a treatment catheter that includes a stent” to
`
`claim 15 does not render it patentable. Teleflex dismisses Itou’s catheter 2 because
`
`it is “meant for an entirely different purpose” than delivering a stent. MTA, 16. But
`
`while Itou does not disclose a “treatment catheter that includes a stent,” Itou in view
`
`of Ressemann renders this limitation obvious. Itou and Ressemann are analogous
`
`art. Pet., 17-21. See Tyco Healthcare Grp. LP v. Ethicon Endo-Surgery, Inc., 774
`
`F.3d 968, 979 (Fed. Cir. 2014) (finding that references to two different surgical
`
`instruments were analogous even though they did not share the same field of
`
`endeavor because they were reasonably pertinent to a particular problem with which
`
`the inventor was involved). Here, Itou, Ressemann, and the ’413 claims all share the
`
`same field of endeavor, which is removing occlusions from the coronary vasculature.
`
`Ex-1847 ¶¶26-60, 118; see also Ex-1805, 20:7-21:1, 70:13-72:1.
`
`Itou discloses that catheter 2 is placed in
`
`precisely the correct location to deliver a
`
`treatment catheter that includes a stent across a
`
`coronary artery lesion, which is shown below
`
`at 80. Ex-1847 ¶28. See Ex-1007, Fig. 6 (color
`
`added), 5:35-42, 7:13-19.
`
`Itou also teaches that catheter 2 is
`
`suitable for delivering a treatment catheter that
`
`3
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`includes a stent. Pet., 72-74; Ex-1847 ¶28. Teleflex wrongly suggests that the
`
`“effective size” of the opening of catheter 2 is 0.046 inches. MTA, 17. That is
`
`inaccurate for reasons set forth in Petitioner’s Reply (Reply) (IPR2020-01341),
`
`§V.B., filed herewith.
`
`But—even if that were correct—Itou still discloses that catheter 2’s tubular
`
`structure is a suitable one through which to advance a treatment catheter that includes
`
`a stent. Before the alleged invention of the ’413 patent, a POSITA was aware of
`
`several stents that were deliverable through a catheter with an I.D. of 0.046 inches.
`
`Ex-1847 ¶52; see, e.g., Ex-1015, 641 (“All current slotted tube designs are ‘bare
`
`mounted’ on a delivery balloon, with deflated profiles smaller than 0.040-in.
`
`(1mm)....”); Ex-1802, 101, 104 (balloon-expandable Genic® stent with less than 0.9
`
`mm (0.035 inch) profile), 142-143 (balloon-mounted Lunar stent with 0.0382 inch
`
`profile); Ex-1804, Table 1, Fig. 3 (disclosing several commercially available stents
`
`with crossing profiles well under 0.046 inches, including those with profiles of 0.99
`
`mm (0.038 inch), 0.93 mm (0.036 inch) and 0.84 mm (0.033 inch)).
`
`Teleflex cannot credibly argue otherwise. Its expert, Keith, admits (as he
`
`must) that Itou’s catheter 2 is suitably placed for delivering a stent. Ex-1805, 137:7-
`
`138:16. He also admits that a guidewire could be navigated into guiding catheter 1—
`
`and then through the guide catheter and right up to the opening of catheter 2—but
`
`then he has no idea whether it would be able to enter the opening. Id., 143:3-145:20.
`
`4
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`Keith’s opinion is questionable given that the 5 French version of GuideLiner v.3
`
`also has an I.D. of 0.046 inches and is marketed for delivering stents, which are, of
`
`course, much larger than a guidewire. Ex-2164, 5.
`
`A POSITA would be motivated to use catheter 2 to deliver a treatment catheter
`
`that includes a stent because this is exactly what is taught in Ressemann. Pet., 77-
`
`83; Ex-1847 ¶¶30-31. Teleflex argues that “using a suction catheter to deliver a stent
`
`could be dangerous.” MTA, 16. In doing so, it ignores explicit teachings in the prior
`
`art to do just that. Ex-1019, 3:4-6, 3:34-36; Ex-1847 ¶¶ 39, 49. Moreover, Teleflex’s
`
`position is belied by real-world evidence that the same catheter may be used to
`
`suction a thrombus—and then deliver a stent—without removing the catheter from
`
`the coronary vasculature. Teleflex’s expert’s patent explicitly teaches using a suction
`
`catheter to suction thrombus, and then—while the aspiration sheath is still in
`
`position—use it to place a stent. Ex-1930, [0084]-[0085], Figs. 28-29. Similarly,
`
`even the GuideLiner has been used to first aspirate thrombus from “thrombus-laden
`
`lesion[s]”—and to then deliver stents—with no requirement that it be removed from
`
`the coronary vasculature between the two procedures. Ex-1127, 250.
`
`Finally, the remaining limitations added to claim 15 do nothing more than
`
`recite additional disclosures in Itou: (A) guide catheter 1 has a distal end adapted to
`
`be placed in an ostium of a branch artery that branches off from a first artery
`
`(Ex-1007, Abstract, 5:29-34); (B) catheter 2 has a cylindrical flexible tip portion and
`
`5
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`a single lumen that is coaxial with the lumen of guide catheter 1 (id., Fig. 3, 3:47-
`
`58, Table 1); (C) catheter 2 has an inner diameter not more than about one French
`
`smaller than the inner diameter of guide catheter 1 (id., Table 1); and (D) the distal
`
`portion of catheter 2’s substantially rigid portion includes a side opening configured
`
`to receive a treatment catheter that includes a stent. Supra, 3-5; Ex-1847 ¶¶28, 74.
`
`Claim 15 is obvious.
`
`Substitute Claim 16
`2.
`One of the limitations added to substitute claim 16 recites that when the distal
`
`portion of the coaxial guide catheter’s tubular structure is extended beyond the distal
`
`end of the guide catheter, its proximal portion remains in the guide catheter lumen.
`
`This is disclosed in Itou. Ex-1007, Figs. 5-6, 3:1-3, 5:26-46, Table 1; Ex-1847 ¶85.
`
`Claim 16 also requires a 6 French guide catheter, and a coaxial guide catheter
`
`that has a tubular structure with an inner diameter that is “at least 0.056 inches.” This
`
`is also taught in Itou. Table 1 explains that guide catheter 1 has an inner diameter of
`
`1.8 mm. Ex-1007. This is equivalent to 0.070 inches, and a 6 French guide catheter
`
`has an internal diameter greater than or equal to 0.070 inches. Ex-1847 ¶88.
`
`Similarly, Table 1 teaches that the inner diameter of the tubular structure of catheter
`
`2 is 1.5 mm, Ex-1007, Table 1, which is 0.059 inches. Ex-1847 ¶89. Thus, Itou
`
`teaches a coaxial guide catheter tubular structure with an inner diameter that is “at
`
`least 0.056 inches.” Claim 16 is obvious.
`
`6
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`Substitute Claim 17
`3.
`It would have been obvious to modify the proximal opening of Itou’s catheter
`
`2, in view of Ressemann or Kataishi, so that it had two inclined sidewalls spaced
`
`apart by a non-inclined concave track. Ex-1847 ¶¶92-123. Like Itou, both
`
`Ressemann and Kataishi disclose catheters for treating occluded vessels. Ex-1008,
`
`Abstract; Ex-1025, Abstract, [0001], and a POSITA had the motivation to modify
`
`Itou based on the teachings of Ressemann and Kataishi for this reason and for those
`
`discussed below. Ex-1847 ¶¶92, 117-23.
`
`Ressemann
`a)
`In one embodiment, Ressemann’s evacuation sheath includes a support collar
`
`2141 on the proximal end of the tubular portion of the evacuation sheath. Ex-1847
`
`¶¶93-95. As shown in Figure 16J, below, support collar 2141 includes a concave
`
`track that runs from the proximal to the distal end of the collar. Circumferential
`
`portion 2141a fits into the proximal opening of the tubular portion’s lumen. Ex-1008,
`
`24:54-56.
`
`
`
`7
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`Id., Fig. 16J (bottom) and schematic of Fig. 16J (top).
`
`Collar 2141 has the claimed side-opening configuration—two inclined
`
`sidewalls (1, 2) spaced apart by a non-inclined concave track. Ex-1800, 166:8-12,
`
`168:9-19 (identifying two inclines); Ex-1847 ¶106.
`
`A POSITA would have been motivated to modify Itou’s proximal tip 23 to
`
`incorporate the structure of Ressemann’s support collar 2141 because Itou’s suction
`
`catheter 2 could be used to deliver a treatment catheter that includes a stent, as taught
`
`in Ressemann. Supra, 3-5. And the reason a POSITA would do so is because
`
`modifying catheter 2’s proximal opening with Ressemann’s collar 2141 increased
`
`the area for receiving a stent. Ex-1847 ¶97; Ex-1807 ¶¶58-59.
`
`Two additional motivations for modifying Itou with Ressemann’s collar are
`
`explicitly taught in Ressemann. The collar serves to reinforce the proximal opening
`
`of the catheter lumen, as well as provide a flexibility transition between the distal
`
`tubular structure and the proximal shaft. Ex-1008, 24:49-67; Ex-1847 ¶¶94-95. A
`
`POSITA would have looked to Ressemann when modifying Itou because both
`
`references disclose devices that address the same problem—removing coronary
`
`vessel occlusions—in the same way—by using a rapid-exchange child catheter, the
`
`distal end of which is extended past a guiding catheter’s distal end, into a coronary
`
`artery. Ex-1007, Abstract, 1:13-16, 2:2-5, 2:29-38, 3:59-63, 5:32-34, Figs. 1A, 1B,
`
`5, 6; Ex-1008, Abstract, 6:18-24, 12:9-12, 19-30, Figs. 6A, 6B. In both Itou and
`
`8
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`Ressemann, occlusive material is suctioned out of the coronary vasculature. Id. A
`
`POSITA would have had a reasonable expectation of success because modifying the
`
`side opening of Itou’s suction catheter to have the structure of Ressemann’s support
`
`collar is nothing more than combining prior art elements according to known
`
`methods to yield predictable results. KSR Int’l Co. v. Teleflex, Inc., 550 U.S. 398,
`
`417 (2007).
`
`b) Kataishi
`Kataishi similarly discloses a side opening on the distal end of a catheter that
`
`has two inclined sidewalls spaced apart by a non-inclined concave track (below, top-
`
`left). Ex-1025, Figs. 2, 12, [0010] (annotation added).
`
`
`
`
`
`
`
`This shape is nearly identical to that taught in the ’413 patent in Figure 4 (the figure
`
`Patent Owner frequently identifies as providing support for this “complex side
`
`opening” limitation) (above, right). Ex-1001, Fig. 4 (rotated, annotation added).
`
`9
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`Kataishi teaches specific benefits to the shape of this side opening. First, it
`
`improves a catheter’s “crossing ability”—the ability to reach a desired target site by
`
`advancing past bends in the vasculature. Ex-1025, Abstract, [0001]; Ex-1847 ¶109;
`
`and see Ex-1055, 6, 8 (testing the Kataishi device via a “passing ability test” that
`
`measured “[t]he length between the ostium and the distal tip of the aspiration
`
`catheter” and found that Kataishi’s shape had improved crossability). Another
`
`benefit is the improved ability of the two-incline-opening is that it improves the
`
`ability to receive material through the opening (e.g., an improved ability to suction
`
`thrombi). Ex-1025, Abstract, [0026]-[0027]; Ex-1847 ¶¶110-11.
`
`
`
`Ex-1025, Fig. 10.
`
`In light of these benefits, taught by Kataishi, a POSITA would have been
`
`motivated to incorporate the shape of Kataishi’s distal opening in Itou’s proximal tip
`
`23. Ex-1847 ¶122. Regarding the first benefit, a POSITA would have recognized
`
`that improved crossability is a desirable feature in a proximal opening of a short tube
`
`for the same reasons as it is desirable in the distal opening of such a tube. Ex-1008,
`
`6:52-60 (“The proximal and distal ends 140a, 140b of the evacuation lumen 140 are
`
`10
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`preferably angled to allow for smoother passage….”). In particular, when retracting
`
`the device, the angled features of the Kataishi opening would provide improved
`
`crossability as the proximal end would then be the leading side.
`
`For the second benefit, a POSITA would have recognized that the improved
`
`ability of an opening to accept material would have been applicable to a proximal
`
`opening accepting a treatment catheter that includes a stent in the same way as it
`
`improved Kataishi’s distal end to suction a thrombus. Similar to modifying the
`
`proximal opening of Itou’s catheter 2 with Ressemann’s collar 2141, a POSITA had
`
`the motivation to modify catheter 2’s proximal opening with the shape of Kataishi’s
`
`distal end because this increased the area for receiving a treatment catheter that
`
`includes a stent. Ex-1847 ¶¶117-22; Ex-1807 ¶¶89-93.
`
`Allowing an improved entry of materials into the lumen of Itou’s suction
`
`catheter without increasing the diameter of the device would have been desirable to
`
`a POSITA, who would have understood that this modification allows the catheter to
`
`receive a therapy catheter and still be advanced to distal locations into the coronary
`
`vasculature (compared to catheters with larger diameters). Ex-1025, Abstract,
`
`[0026]-[0027], Fig. 10; Ex-1055, 300, 304 (disclosing a better ability to load because
`
`of shape of distal opening); Ex-1847 ¶123; Ex-1807 ¶91.
`
`Similar to modifying Itou’s catheter 2 to include the Ressemann collar at its
`
`proximal end, a POSITA would have a reasonable expectation of success in
`
`11
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`modifying Itou’s suction catheter with the claimed side-opening features, as taught
`
`by Kataishi. Ex-1847 ¶124. Creating two different inclined slopes would have been
`
`a routine task when manufacturing an extension catheter. Id.
`
`B.
`
`Substitute claims 15-17 are unpatentable over Kontos in view of
`Ressemann and Takahashi and/or Kataishi.
`Teleflex’s MTA alleges two principal reasons why the substitute claims are
`
`valid over the Kontos grounds: (1) none of the art discloses or suggests the claimed
`
`“backup support” provided by a 6 French guide catheter with a guide extension
`
`catheter less than “about” one French size smaller used to deliver a stent and a stent
`
`catheter, and (2) none of the art discloses or suggests the claimed double incline side
`
`opening. MTA, at 8-10. But Kontos inherently teaches backup support, and
`
`Takahashi taught that 5-in-6 French catheter sizing improved backup support for
`
`delivery of interventional cardiology devices like stents. Ressemann also teaches
`
`delivery of stents through a guide extension catheter with a side opening—indeed,
`
`Teleflex itself admitted that Ressemann anticipated similar claims in related patents
`
`in other IPRs. For the double incline, the exact structure claimed is already disclosed
`
`for use in two suction catheters, Ressemann and Kataishi. It would be no more than
`
`common sense and routine engineering to take known structures (with known
`
`benefits for better trackability and better entry/exit of matter through the opening)
`
`and apply those teachings to Kontos to achieve predictable results.
`
`12
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`Kontos discloses a support catheter assembly 10 (“guide extension catheter”)
`
`for use with a guide catheter 38. Ex-1009, 5:35-38, 5:43-46, 7:1-23, 7:35-43, Figs.
`
`5-6, 8. 6 French was a common size for the guide catheter that would have an inner
`
`diameter of 1.8mm (0.071 inches). Ex-1847 ¶¶186, 189. Support catheter assembly
`
`10 has an insertion/manipulation wire 14 (“substantially rigid segment”) and body
`
`12 (i.e., “tubular structure”), which includes soft tip 28 and tube 16. Ex-1009, 3:45-
`
`55, 4:5-7, 4:25-38, 5:25-30, Figs. 6A-C. This support catheter assembly is used to
`
`“pass further through the coronary ostia than can guide catheter 38” and to function
`
`as a “guide extension catheter” when “extending beyond the distal end of guide
`
`catheter 38” for a PTCA balloon catheter. Ex-1009, 5:40-56. This procedure “lessens
`
`considerably the tendency of the PTCA catheter 40 to bend, buckle or kink.” Id.
`
`1.
`
`Substitute claims 15-17 are unpatentable over Kontos in
`view of Ressemann and Takahashi.
`Sequence Delivery of Stents
`a)
`Substitute claim 15 specifies that the method limitations are performed “in the
`
`order recited.” MTA, App. The parties dispute whether the last element in substitute
`
`claim 15 (to insert a treatment catheter that includes a stent) is suggested by the art
`
`in the order claimed. It is. Kontos itself explains that “[b]ody 12 could be inserted
`
`first” into the guide catheter 38, and then “followed by the PTCA catheter 40.”
`
`Ex-1009, 7:45-49; see also id., 5:16-28; Ex-1847 ¶¶181-82. Teleflex cannot dispute
`
`this explicit disclosure. And while Kontos does not disclose delivery of a stent and
`
`13
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`stent catheter, Kontos is sufficiently sized and configured to permit the passage of
`
`stents and stent catheters through guide catheter 38. Ex-1847 ¶125. Kontos’s tube 16
`
`of the guide catheter, which is positioned inside guide catheter 38, has an inner
`
`diameter of 0.045 inches. Ex-1009, 4:46-50. It was known that stents and stent
`
`catheters could be advanced through the guide catheter of Kontos. Id., see also Ex-
`
`1015, 641 (“All current slotted tube designs are ‘bare mounted’ on a delivery
`
`balloon, with deflated profiles smaller than 0.040-in. (1mm)”); Ex-1097, 104
`
`(Genic® stent with less than 0.9 mm (0.035 inch) profile), 143 (Lunar stent with
`
`0.0382 inch profile), 269 (Spiral Force stent with 0.039 to 0.042 inch profile), 274
`
`(Tsunami stent with 0.95 mm (0.038 inch) profile). Further, stent delivery was
`
`disclosed in both Ressemann and Takahashi—Takahashi says its inner lumen of the
`
`5 French child catheter has a 0.059 inch inner diameter and can accept both balloons
`
`and stent delivery systems less than 4.0 mm in diameter. Ex-1010, 5.
`
`b)
`
`6 French Guide Catheter and One French Size
`Smaller Lumen
`Kontos discloses a cross-sectional outer diameter and inner diameter of body
`
`12 that is 0.055 inches and 0.045 inches, respectively. Ex-1009, 3:56-59, 4:48-52.
`
`At base portion 18, Kontos discloses a 0.065-inch outer diameter. Id., 4:50-53.
`
`Kontos does not disclose the cross-sectional inner diameter of the guide catheter.
`
`Takahashi, however, discloses a “five-in-six” system wherein the inner diameter of
`
`14
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`the 5 French catheter is 0.059 inches. Ex-1847 ¶¶186, 189; Ex-1442 ¶¶97-103; see
`
`also Ex-1010, 452.
`
`It would have been obvious to look at modifying Kontos in light of Ressemann
`
`and Takahashi to use an extension catheter with a 0.056 or greater inner diameter.
`
`Ex-1847 ¶¶186, 189. Indeed, Kontos, Ressemann, and Takahashi are directed to the
`
`same type of device, are in the same field of endeavor, and are reasonably pertinent
`
`to the problem faced by the inventors of the ʼ413 patent. Ex-1847 ¶186.
`
`A POSITA would have been motivated to combine Takahashi with Kontos
`
`and Ressemann, given that the former teaches that using a catheter-in-catheter
`
`assembly—in particular, an assembly that achieves 1 French differential between
`
`the inner and outer catheters—can improve back-up support. Ex-1847 ¶¶186, 189;
`
`Ex-1442 ¶¶97-103. Further, a POSITA would have been able to accomplish the
`
`claimed combination with a reasonable expectation of success given the teachings
`
`of Kontos, Ressemann, Takahashi, and/or a POSITA’s knowledge. A POSITA knew
`
`how to (i) replace the proximal funnel with a side opening, as discussed above, and
`
`(ii) recess Kontos’s distal marker bands. Ex-1847 ¶189. Further, Kontos explicitly
`
`teaches that “tube 16 may be molded directly only application wire 14.” Ex-1009,
`
`4:31-32.
`
`In so doing, to maximize the usable “real estate” in the catheter assembly, a
`
`POSITA would taper the pushrod for attachment onto the Kontos-Ressemann
`
`15
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`combination. Ex-1847 ¶152; Ex-1807 ¶¶131-34, 143-44. Making
`
`these
`
`modifications would permit a POSITA to use an extension catheter with at least a
`
`0.056-inch inner diameter in a 6 French GC. Ex-1807 ¶¶134-36. This is modification
`
`would increase the outer diameter of Kontos’s body 12, but this modification was
`
`well within the skill of a POSITA, as appropriately sized catheters were ubiquitous
`
`in the art. See, e.g., Ex-1847 ¶¶163-65; Ex-1442 ¶¶141-42; Ex-1009, 4:64-65;
`
`Ex-1010, 452. Combining the teachings of Kontos with Ressemann and Takahashi
`
`to achieve the not-more-than-one French differential would have been nothing more
`
`than combining prior art elements according to known methods to yield predictable
`
`results. Ex-1847 ¶192; Ex-1442 ¶103.
`
`Complex Side Opening
`c)
`Kontos’s support catheter assembly 10 does not have a side opening.
`
`However, Ressemann discloses a complex side opening. A POSITA would have
`
`been motivated to modify Kontos’s tube 16 in view of Ressemann to have a complex
`
`side opening. In particular, a POSITA would have been motivated to modify Kontos
`
`to add Ressemann’s support collar 2141 to Kontos for multiple reasons.
`
`
`
`16
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`First, a POSITA would have known that use of a side opening would optimize
`
`the inner diameter of tube 16 without causing a reduction in the area of the point of
`
`entry to the extension catheter 10. Ex-1847 ¶¶162-63. For example, as an alternative
`
`to the flared proximal opening 26, a POSITA would have been motivated to use a
`
`side opening like in Ressemann as then the diameter of the GC could be reduced
`
`without causing a commensurate reduction in the area of the proximal opening of
`
`extension catheter. Id. Alternatively, a POSITA would have been motivated to
`
`remove Kontos’s proximal funnel, as it would permit the inner diameter of the
`
`extension catheter to be increased without causing a commensurate increase in the
`
`outer diameter of the guide catheter. Ex-1847 ¶¶193-94.
`
`Second, the side opening formed by support collar 2141 facilitates “smooth[]”
`
`reception of the interventional cardiology device as it enters the lumen of the
`
`extension catheter. Ex-1008, 6:52-57 (100 embodiment), 24:38-41 (2100
`
`embodiment); see also Ex-1847 ¶194. That is, the side opening reduces the
`
`likelihood of device hang-up, meaning it promotes better advancement of the therapy
`
`catheter as it travels to the occlusion. Ex-1847 ¶¶194-95.
`
`Third, the side opening, as taught by Ressemann, promotes “smoother
`
`passage” of the catheter assembly as it navigates the tortuous vasculature. Ex-1008,
`
`6:52-57; see also Ex-1847 ¶167; Ex-1442 ¶87. Adding a side opening in the form of
`
`support collar 2141 to the lumen of the extension catheter reduces the amount of
`
`17
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`force that a physician must exert to advance the catheter through winding
`
`vasculature. Ex-1847 ¶162.
`
`Teleflex alleges this motivation is impermissible hindsight, but “[t]he
`
`combination of familiar elements according to known methods is likely to be
`
`obvious when it does no more than yield predictable results.” KSR Int’l Co. v.
`
`Teleflex Inc., 550 U.S. 398, 416 (2007). Teleflex narrowly focuses on specific
`
`embodiments and argues because they are not combinable without making some
`
`modifications, the invention is not obvious. Teleflex ignores that these modifications
`
`were well-understood routine engineering concepts within the capability of a
`
`POSITA, who had motivation to make them. Taking well-defined structures with
`
`known benefits, like the shape of various catheter openings, and applying those
`
`structures to other catheters to achieve predictable results is not inventive. Thus, the
`
`proposed amended claims are invalid.
`
`2.
`
`Substitute claim 17 is unpatentable over Kontos in view of
`Ressemann, Takahashi, and Kataishi.
`Ressemann (and Keith’s patent, Teleflex’s own expert) gives an explicit
`
`teaching why Kataishi’s double incline shape would be combined into Kontos:
`
`The proximal and distal ends 140a, 140b of the evacuation lumen 140
`are preferably angled to allow for smoother passage of the evacuation
`sheath assembly 100 through a guide catheter, and into a blood vessel,
`and to facilitate smoother passage of other therapeutic devices
`through the evacuation lumen 140 of the evacuation head 132. The
`
`18
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`larger area of the angled open ends also allows for larger deformable
`particulate matter to pass through the lumen more smoothly.
`
`Ex-1008, 6:52-60 (emphasis added). The shape gives benefits to trackability of the
`
`catheter itself, as well as smoother passage of catheters and other interventional
`
`devices and particular matter. Kataishi offers those same benefits and a POSITA
`
`would recognize that Kataishi’s shape would have similar and predictable benefits
`
`if incorporated onto the proximal opening. Ex-1847 ¶¶215-16.
`
`III. SECONDARY CONSIDERATIONS DO NOT OVERCOME THE
`STRONG SHOWING OF OBVIOUSNESS.
`A. Teleflex is not entitled to a presumption of nexus.
`Teleflex recites secondary considerations for substitute claims 15 and 17, each
`
`of which requires the inner lumen to have “an inner diameter not more than about
`
`one French smaller than the inner diameter of the standard guide catheter.” MTA,
`
`App. A rebuttable presumption of nexus exists “when the patentee shows that the
`
`asserted objective evidence is tied to a specific product and that product is the
`
`invention disclosed and claimed in the patent.” WBIP, LLC v. Kohler Co., 829 F.3d
`
`1317, 1329 (Fed. Cir. 2016). Conversely, “when the thing that is commercially
`
`successful is not coextensive with the patented invention . . . the patentee is not
`
`entitled to a presumption of nexus.” Fox Factory, Inc. v. SRAM, LLC, 944 F.3d 1366,
`
`1373 (Fed. Cir. 2019).
`
`19
`
`

`

`IPR2020-01341, -01342
`Patent 8,142,413
`GuideLiner does not practice substitute claim 15 of the ’413 patent. See
`
`Ex-2250, App’x B.
`
`
`
`
`
`. Ex-1851, 154:13-155:4 (testifying about “not more than one French” language
`
`of similar claim). Further, Teleflex’s own expert previously testified that
`
`“[m]athmatically, those of ordinary skill in the art define one French as .0131
`
`inches.” Ex-1825 ¶¶43-48 (testifying that Boosting Catheter infringes due to the
`
`0.0131 difference); see also Ex-1844, 27:3-9 (Judge Schiltz noting that Teleflex’s
`
`attorneys “stood before [him] in QXMedical and told [him] one French meant 0.0131
`
`inches”).
`
`But the difference for both GuideLiner and Telescope is 0.014 inches.
`
`Ex-1082 ¶58 (red

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket