throbber

`
`EXHIBIT A-3
`
`Invalidity Claim Chart of Lam, alone or in combination with any of Sigg, Boulange, Reuter, Scypinski, Fries, Chacornac,
`Nema, D’Souza, Furfine, Badkar, Macugen, Eylea, Lucentis, Stewart, USP789, Liu, Hioki, DC365, Khandke, and/or Dixon
`against U.S. Patent No. 9,220,631.
`
`
`
`Charted Reference:
`
`PCT Patent Publication No. WO 2008/077155 to Lam et al. (“Lam”), in view of Sigg, Boulange, Reuter, Scypinski, Fries, Chacornac,
`Nema, D’Souza, Furfine, Badkar, Macugen, Eylea, Lucentis, Stewart, USP789, Liu, Hioki, DC365, Khandke, and/or Dixon., render
`obvious claims 1-26 of U.S. Patent No. 9,220,631.
`
`This claim chart is based on Regeneron’s current understanding of the asserted claims, and Regeneron’s investigation to date.
`Regeneron is not admitting to the accuracy of any particular construction. Regeneron reserves all rights to amend this invalidity claim
`chart in light of any claim construction developments or any amendments to Novartis’s infringement contentions or domestic industry
`contentions, should such developments occur or amendments be allowed. Further, as discovery is ongoing and Regeneron continues to
`seek discovery from third parties regarding the references identified in Regeneron’s invalidity contentions as well as other potential
`prior art, Regeneron reserves the right to revise its invalidity contentions as appropriate in view of any ongoing discovery.
`
`The claim chart below identifies where each limitation of each asserted claim of the 631 Patent can be found in Lam. The citations
`provided below are exemplary, rather than exhaustive, and Regeneron reserves the right to rely upon any other portion of the cited
`references. Where Regeneron identifies a portion of a reference’s text, the identification should be understood as referencing any
`corresponding figure or diagram, and vice versa.
`
`
`
`
`
`
`Novartis Exhibit 2045.001
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`[1.a-pre] A pre-filled,
`terminally sterilized syringe
`for intravitreal injection
`
`Corresponding Disclosure
`Lam discloses a pre-filled, terminally sterilized syringe for intravitreal injection.
`
`For example, see the following passages and/or figures, as well as all related disclosures:
`
`
`Objects used in medical applications are generally sterilized before use. Sterilization can
`be accomplished by a variety of methods including, e.g., steam sterilization, radiation
`sterilization, gas sterilization (e.g. with ethylene oxide), and chemical sterilization.
`However, these treatments cannot be used for objects containing pharmaceutical
`compositions because their active ingredients are typically sensitive to them. For example,
`steam and gas sterilization are generally performed at high temperatures (approx. 45 0 C
`to 55 0 C or higher) that damage certain active ingredients in pharmaceutical
`compositions. Similarly, the agents used for radiation or chemical sterilization generally
`cause chemical damage to the active ingredients. Consequently, pharmaceutical
`compositions are generally sterilized by an alternative method, e.g. by filtration, and then
`packaged into separately sterilized objects. Because of the complexity of this process, it is
`difficult to also ensure the sterility of the surfaces of the objects.
`
`In many circumstances it would be advantageous to sterilize the surfaces of these objects
`in order to reduce the risk of contamination during subsequent handling. For example,
`there is an increased risk of endophthalmitis after intraocular injection if the surface of the
`syringe used for injection is not sterilized. Thus, there remains a need for efficient and
`cost-effective methods of surface-sterilizing objects containing ethylene-oxide-sensitive,
`temperature-sensitive compounds, such as biological molecules, without a significant
`adverse effect on their activity or integrity.
`
`
`Lam at 1:14-32.
`
`
`The invention relates to methods for surface-sterilizing objects containing ethylene-
`oxide-sensitive, temperature-sensitive compounds, such as biological molecules. The
`invention is based, in part, on the surprising discovery of ethylene-oxide -based
`sterilization conditions that will effectively sterilize the surface of an object but which do
`
`
`
`2
`
`Novartis Exhibit 2045.002
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`Corresponding Disclosure
`not significantly damage ethylene-oxide-sensitive, temperature-sensitive compounds
`contained inside.
`
`
`
`
`
`
`
`In one aspect, the invention provides a method for surface-sterilizing an object having an
`ethylene-oxide(EtO)-impermeable interior space containing a compound with a
`temperature-sensitive and EtO-sensitive activity by exposing the object to EtO under
`conditions such that the object is surface-sterilized and the compound retains at least 50%
`of said activity. In some embodiments, the conditions comprise: a) temperature between
`25 0 C and 35 0 C; b) EtO concentration of between 300 mg/L and 800 mg/L; and c)
`relative humidity between 45% and 60%; for between 1 and 6 hours. In some
`embodiments, the conditions comprise: a) temperature between 27 0 C and 33 0 C; b) EtO
`concentration of between 300 mg/L and 600 mg/L; and c) relative humidity between 48%
`and 52%; for between 1 and 6 hours. In some embodiments, the conditions comprise: a)
`temperature of 3O 0 C; b) EtO concentration of 600 mg/L; and c) relative humidity of
`50%; for 1, 1.5 or 2 hours.
`
`In some embodiments, the compound retains at least 90% of said activity. In some
`embodiments, the compound is a polypeptide, e.g. an antibody, which includes
`monoclonal antibodies, chimeric antibodies, humanized antibodies or human antibodies.
`In some embodiments where the compound is a polypeptide, the percent alkylation of the
`polypeptide is not statistically different from a control polypeptide not exposed to EtO. In
`some embodiments, the antibody is an antigen-binding fragment, e.g. a Fab fragment. In
`some embodiments, the Fab fragment binds VEGF, e.g. ranibizumab (LUCENTIS®). In
`some embodiments, the compound is present in an aqueous pharmaceutical composition,
`e.g. a composition comprising at least one of: an amino acid, a disaccharide and a non-
`ionic surfactant. In some embodiments the pharmaceutical composition comprises
`histidine, trehalose and polysorbate 20.
`
`In some embodiments, the object is a syringe. In some embodiments the syringe
`comprises glass and comprises a stopper comprising D777-7 laminated with FluroTec®;
`and a tip cap comprising D777-7 laminated with FluroTec® or D21-7H laminated with
`
`
`
`3
`
`Novartis Exhibit 2045.003
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`Corresponding Disclosure
`FluroTec®. In some embodiments, the object is contained within a package comprising an
`EtO-permeable material, e.g. TYVEK®.
`
`
`Lam at 2:3-33.
`
`
`In some embodiments, the pharmaceutical composition is designed for intraocular
`injection.
`
`
`Lam at 11:30-31.
`
`
`The methods of the invention are typically used to sterilize objects containing
`pharmaceutical formulations. For example, the methods of the invention may be used with
`syringes, vials or cartridges (such as are used in devices designed for multiple injections).
`In addition, the method of the invention may be used with a syringe with or without a
`needle. In the latter case, some sort of cap or needle shield is generally positioned where
`the needle will subsequently be attached. The following example is intended merely to
`illustrate the practice of the present invention and is not provided by way of limitation.
`The disclosures of all patent and scientific literatures cited herein are expressly
`incorporated in their entirety by reference.
`
`
`Lam at 12:31-13:6.
`
`
`We performed experiments to identify whether there were parameters for EtO sterilization
`that would effectively sterilize the surface of an object but which do not damage an
`ethylene-oxide-sensitive, temperature-sensitive compound contained inside. We
`performed EtO sterilization runs on syringes containing a ranibizumab solution (at a
`protein concentration indicated in Table 2 in a solution with 10 mM histidine HCl, 10% α,
`α - trehalose dehydrate, 0.01% polysorbate 20, pH 5.5) where each run had the following
`standard EtO sterilization steps: (1) set temperature; (2) evacuate chamber to about 5.0"
`HgA; (3) leak test; (4) wash twice with nitrogen; (5) humidify chamber and incubate
`about 30 min; (6) inject EtO gas and incubate for dwell time; (7) evacuate chamber to
`about 5.0" HgA; and (8) wash four times with nitrogen (each wash cycle is about 15-20
`
`
`
`4
`
`Novartis Exhibit 2045.004
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`Corresponding Disclosure
`min). In addition to the syringe, each run also included a paper strip with approximately
`1.9 x 10 6 Bacillus subtilis spores, which was used to monitor the sterilization as follows:
`the strip was soaked in media, vortexed vigorously and then serial dilutions were plated
`and grown for one week. We then varied the following sterilization-critical factors as
`indicated in Table 1 : temperature, relative humidity, time of exposure (gas dwell), and
`EtO concentration
`
`
`Lam at 13:12-26.
`
`
`We also tested several different syringe components: where the stopper on the plunger
`comprised D777-7 laminated with a 125 μm coating of FluroTec® barrier film and where
`the tip cap comprised either D777-7 or D21-7H laminated on both the surface in contact
`with the tip of the syringe and the exterior surface with a 125 μm coating of FluroTec®
`barrier film (all components from West Pharmaceutical Services / Daikyo Seiko). We
`measured the residual EtO in the syringe and the stability of ranibizumab by IEC the same
`day as the treatment and at various monthly time points thereafter. For IEC, we measured
`the percentage of protein in the main peak and in the acidic and basic peaks, with the
`protein in the basic peak representative of alkylation which may have been caused by the
`EtO treatment. As shown in Table 3, under all conditions tested the percentage of protein
`in the basic peaks was at most approximately 1% over control. Further, when the
`FluroTec® barrier film was used on the syringe components, the percentage of protein in
`the basic peak was not statistically different from control.
`
`
`Lam at 15:12-24.
`
`
`
`
`5
`
`Novartis Exhibit 2045.005
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`Corresponding Disclosure
`
`
`Lam at Table 3.
`
`
`The above specification, example and data provide a complete description of the
`manufacture and use of the composition of the invention. Since many embodiments of the
`invention can be made without departing from the spirit and scope of the invention, the
`invention includes all such embodiments.
`
`
`
`
`
`6
`
`Novartis Exhibit 2045.006
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`[1.b] the syringe comprising a
`glass body forming a barrel, a
`stopper and a plunger
`
`Corresponding Disclosure
`This specification contains numerous citations to literature and patents. Each is hereby
`incorporated by reference for all purposes, as if fully set forth.
`
`Lam at 16:3-8.
`
`To the extent Novartis alleges this limitation is not met by any of the disclosures above, it would be
`obvious in view of Sigg, Boulange, Reuter, Scypinski, Metzner, Shah, Fries, Schoenknecht,
`Chacornac, Nema, D’Souza, Furfine, Badkar, Macugen, Eylea, and/or Lucentis. See A-1–A-2, A-
`4–A-13, B-1–B-3 and all references cited therein.
`Lam discloses the syringe comprising a glass body forming a barrel, a stopper and a plunger.
`
`For example, see the following passages and/or figures, as well as all related disclosures:
`
`
`In some embodiments, the object is a syringe. In some embodiments the syringe
`comprises glass and comprises a stopper comprising D777-7 laminated with FluroTec®;
`and a tip cap comprising D777-7 laminated with FluroTec® or D21-7H laminated with
`FluroTec®. In some embodiments, the object is contained within a package comprising an
`EtO-permeable material, e.g. TYVEK®.
`
`
`Lam at 2:29-33.
`
`
`As used herein, an "ethylene-oxide-impermeable" or "EtO-impermeable" object is one in
`which no more than 0.5 ppm EtO is present inside the object after EtO sterilization. As
`EtO-impermable object may comprise, e.g., glass and/or certain plastics.
`
`
`Lam at 3:17-19.
`
`
`We next tested additional conditions with different dwell times during the EtO cycle,
`different numbers of washes and different container components. We performed EtO
`sterilization runs on syringes containing a ranibizumab solution (at 10.0 mg/ml in a
`solution with 10 mM histidine HCl, 10% α, α -trehalose dehydrate, 0.01% polysorbate 20,
`pH 5.6) where each run had the following standard EtO sterilization steps: (1) set
`
`
`
`7
`
`Novartis Exhibit 2045.007
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`Corresponding Disclosure
`temperature to 3O 0 C; (2) evacuate chamber to about 5.0" HgA; (3) leak test; (4) wash
`twice with nitrogen; (5) humidify chamber and incubate about 30 min; (6) inject EtO gas
`and incubate for dwell time indicated in Table 3; (7) evacuate chamber to about 5.0" HgA;
`and (8) wash with with nitrogen the number of time indicated in Table 3. In addition to the
`syringe, each run also included a paper strip with approximately 1.9 x 10 6 Bacillus
`subtilis spores, which was used to monitor the sterilization as follows: the strip was
`soaked in media, vortexed vigorously and then serial dilutions were plated and grown for
`one week. We also tested several different syringe components: where the stopper on the
`plunger comprised D777-7 laminated with a 125 μm coating of FluroTec® barrier film
`and where the tip cap comprised either D777-7 or D21-7H laminated on both the surface
`in contact with the tip of the syringe and the exterior surface with a 125 μm coating of
`FluroTec® barrier film (all components from West Pharmaceutical Services / Daikyo
`Seiko). We measured the residual EtO in the syringe and the stability of ranibizumab by
`IEC the same day as the treatment and at various monthly time points thereafter. For IEC,
`we measured the percentage of protein in the main peak and in the acidic and basic peaks,
`with the protein in the basic peak representative of alkylation which may have been
`caused by the EtO treatment. As shown in Table 3, under all conditions tested the
`percentage of protein in the basic peaks was at most approximately 1% over control.
`Further, when the FluroTec® barrier film was used on the syringe components, the
`percentage of protein in the basic peak was not statistically different from control.
`
`
`Lam at 15:1-24.
`
`
`
`Lam at Claim 21.
`
`
`21. The method of claim 20, wherein said syringe comprises glass and comprises:
`(a) a stopper comprising D777-7 laminated with FluroTec®; and
`(b) a tip cap comprising (i) D777-7 laminated with FluroTec® or (ii) D21-7H laminated
`with FluroTec®.
`
`The above specification, example and data provide a complete description of the
`manufacture and use of the composition of the invention. Since many embodiments of the
`
`
`
`8
`
`Novartis Exhibit 2045.008
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`[1.c] and [the syringe]
`containing an ophthalmic
`solution which comprises a
`VEGF-antagonist, wherein
`
`Corresponding Disclosure
`invention can be made without departing from the spirit and scope of the invention, the
`invention includes all such embodiments.
`
`This specification contains numerous citations to literature and patents. Each is hereby
`incorporated by reference for all purposes, as if fully set forth.
`
`Lam at 16:3-8.
`
`To the extent Novartis alleges this limitation is not met by any of the disclosures above, it would
`have been obvious in view of Sigg, Boulange, Reuter, Scypinski, Metzner, Shah, Fries,
`Schoenknecht, Chacornac, Nema, D’Souza, Furfine, Badkar, Macugen, Eylea, and/or Lucentis. See
`A-1–A-2, A-4–A-13, B-1–B-3 and all references cited therein.
`Lam discloses the syringe contains an ophthalmic solution which comprises a VEGF-antagonist.
`
`For example, see the following passages and/or figures, as well as all related disclosures:
`
`
`In some embodiments, the compound retains at least 90% of said activity. In some
`embodiments, the compound is a polypeptide, e.g. an antibody, which includes
`monoclonal antibodies, chimeric antibodies, humanized antibodies or human antibodies.
`In some embodiments where the compound is a polypeptide, the percent alkylation of the
`polypeptide is not statistically different from a control polypeptide not exposed to EtO. In
`some embodiments, the antibody is an antigen-binding fragment, e.g. a Fab fragment. In
`some embodiments, the Fab fragment binds VEGF, e.g. ranibizumab (LUCENTIS®). In
`some embodiments, the compound is present in an aqueous pharmaceutical composition,
`e.g. a composition comprising at least one of: an amino acid, a disaccharide and a non-
`ionic surfactant. In some embodiments the pharmaceutical composition comprises
`histidine, trehalose and polysorbate 20.
`
`
`Lam at 2:18-24.
`
`
`VEGF or VEGF receptor antibodies including humanized and/or affinity matured VEGF
`antibodies such as the humanized VEGF antibody huA4.6.1 bevacizumab (AVASTIN®)
`
`
`
`9
`
`Novartis Exhibit 2045.009
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`Corresponding Disclosure
`and ranibizumab (LUCENTIS®) (Kim et al., Growth Factors, 7:53-64 (1992),
`International Publication Nos. WO 96/30046 and WO 98/45331)
`
`
`Lam at 7:24-27.
`
`
`In some embodiments, non-polypeptide compounds are used in the methods of the
`invention. These include, e.g., pegaptanib (MACUGEN®; Eyetech Pharmaceuticals),
`steroids, and compounds used for RNA-interference mediated therapy.
`
`
`Lam at 11:9-11.
`
`
`We performed experiments to identify whether there were parameters for EtO sterilization
`that would effectively sterilize the surface of an object but which do not damage an
`ethylene-oxide-sensitive, temperature-sensitive compound contained inside. We
`performed EtO sterilization runs on syringes containing a ranibizumab solution (at a
`protein concentration indicated in Table 2 in a solution with 10 mM histidine HCl, 10% α,
`α - trehalose dehydrate, 0.01% polysorbate 20, pH 5.5) where each run had the following
`standard EtO sterilization steps: (1) set temperature; (2) evacuate chamber to about 5.0"
`HgA; (3) leak test; (4) wash twice with nitrogen; (5) humidify chamber and incubate
`about 30 min; (6) inject EtO gas and incubate for dwell time; (7) evacuate chamber to
`about 5.0" HgA; and (8) wash four times with nitrogen (each wash cycle is about 15-20
`min). In addition to the syringe, each run also included a paper strip with approximately
`1.9 x 10 6 Bacillus subtilis spores, which was used to monitor the sterilization as follows:
`the strip was soaked in media, vortexed vigorously and then serial dilutions were plated
`and grown for one week. We then varied the following sterilization-critical factors as
`indicated in Table 1 : temperature, relative humidity, time of exposure (gas dwell), and
`EtO concentration.
`
`
`Lam at 13:12-26.
`
`
`The above specification, example and data provide a complete description of the
`manufacture and use of the composition of the invention. Since many embodiments of the
`
`
`
`10
`
`Novartis Exhibit 2045.010
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`[1.d] (a) the syringe has a
`nominal maximum fill
`volume of between about 0.5
`ml and about 1 ml,
`
`Corresponding Disclosure
`invention can be made without departing from the spirit and scope of the invention, the
`invention includes all such embodiments.
`
`This specification contains numerous citations to literature and patents. Each is hereby
`incorporated by reference for all purposes, as if fully set forth.
`
`Lam at 16:3-8.
`
`To the extent Novartis alleges this limitation is not met by any of the disclosures above, it would
`have been obvious in view of Sigg, Boulange, Reuter, Scypinski, Metzner, Shah, Fries,
`Schoenknecht, Chacornac, Nema, D’Souza, Furfine, Badkar, Macugen, Eylea, and/or Lucentis. See
`A-1–A-2, A-4–A-13, B-1–B-3 and all references cited therein.
`Lam discloses pre-filled syringes for intravitreal administration, and therefore, it would have been
`obvious to a POSITA that the syringes are small volume syringes, for example, between 0.5-1 ml
`in volume.
`
`For example, see the following passages and/or figures, as well as all related disclosures:
`
`
`VEGF or VEGF receptor antibodies including humanized and/or affinity matured VEGF
`antibodies such as the humanized VEGF antibody huA4.6.1 bevacizumab (AVASTIN®)
`and ranibizumab (LUCENTIS®) (Kim et al., Growth Factors, 7:53-64 (1992),
`International Publication Nos. WO 96/30046 and WO 98/45331)
`
`
`Lam at 7:24-27.
`
`
`In some embodiments, non-polypeptide compounds are used in the methods of the
`invention. These include, e.g., pegaptanib (MACUGEN®; Eyetech Pharmaceuticals),
`steroids, and compounds used for RNA-interference mediated therapy.
`
`
`Lam at 11:9-11.
`
`
`
`
`11
`
`Novartis Exhibit 2045.011
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`Corresponding Disclosure
`In some embodiments, the pharmaceutical composition is designed for intraocular
`injection.
`
`
`Lam at 11:30-31.
`
`
`We next tested additional conditions with different dwell times during the EtO cycle,
`different numbers of washes and different container components. We performed EtO
`sterilization runs on syringes containing a ranibizumab solution (at 10.0 mg/ml in a
`solution with 10 mM histidine HCl, 10% α, α -trehalose dehydrate, 0.01% polysorbate 20,
`pH 5.6) where each run had the following standard EtO sterilization steps: (1) set
`temperature to 3O 0 C; (2) evacuate chamber to about 5.0" HgA; (3) leak test; (4) wash
`twice with nitrogen; (5) humidify chamber and incubate about 30 min; (6) inject EtO gas
`and incubate for dwell time indicated in Table 3; (7) evacuate chamber to about 5.0" HgA;
`and (8) wash with with nitrogen the number of time indicated in Table 3. In addition to the
`syringe, each run also included a paper strip with approximately 1.9 x 10 6 Bacillus
`subtilis spores, which was used to monitor the sterilization as follows: the strip was
`soaked in media, vortexed vigorously and then serial dilutions were plated and grown for
`one week. We also tested several different syringe components: where the stopper on the
`plunger comprised D777-7 laminated with a 125 μm coating of FluroTec® barrier film
`and where the tip cap comprised either D777-7 or D21-7H laminated on both the surface
`in contact with the tip of the syringe and the exterior surface with a 125 μm coating of
`FluroTec® barrier film (all components from West Pharmaceutical Services / Daikyo
`Seiko). We measured the residual EtO in the syringe and the stability of ranibizumab by
`IEC the same day as the treatment and at various monthly time points thereafter. For IEC,
`we measured the percentage of protein in the main peak and in the acidic and basic peaks,
`with the protein in the basic peak representative of alkylation which may have been
`caused by the EtO treatment. As shown in Table 3, under all conditions tested the
`percentage of protein in the basic peaks was at most approximately 1% over control.
`Further, when the FluroTec® barrier film was used on the syringe components, the
`percentage of protein in the basic peak was not statistically different from control.
`
`
`Lam at 15:1-24.
`
`
`
`12
`
`Novartis Exhibit 2045.012
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`Corresponding Disclosure
`
`
`The above specification, example and data provide a complete description of the
`manufacture and use of the composition of the invention. Since many embodiments of the
`invention can be made without departing from the spirit and scope of the invention, the
`invention includes all such embodiments.
`
`This specification contains numerous citations to literature and patents. Each is hereby
`incorporated by reference for all purposes, as if fully set forth.
`
`Lam at 16:3-8.
`
` A
`
` POSITA would have understood that pre-filled syringes for intravitreal administration disclosed
`would be 0.5 ml or 1 ml, or alternatively it would have been obvious for a POSITA to select the
`claimed syringe sizes for use for intravitreal injection. A POSITA would have had a reasonable
`expectation of success combining Lam and Nema, Chacornac, Furfine, Sigg, and/or Macugen in a
`way that satisfies this limitation.
`
`To the extent this limitation is not expressly and/or inherently disclosed, such limitation would
`have been, even without resorting to the disclosures of any other reference, because it was within
`the common knowledge of persons of ordinary skill in the art, and was used in the claims according
`to known methods, to achieve predictable results.
`
`For example, it was common knowledge of persons of ordinary skill in art that pre-filled syringes
`containing an ophthalmic solution for intravitreal administration are generally administered in
`volumes < 0.1 mL. Indeed, the Background Art section of the 631 Patent acknowledges that “[f]or
`small volume syringes, for example those for injections into the eye in which it is intended that
`about 0.1 ml or less of liquid is to be injected. . .” 631 Patent at 1:22-24. A POSITA would have
`understood that the nominal maximum fill volume has to be more than the volume that is desired
`for injection, in order to account for priming and loss of product. Moreover, 0.5 mL and 1 mL are
`standard syringe sizes that are commonly used for intravitreal applications. See, e.g., Macugen
`(VEGF antagonist delivered in a 1 ml prefilled syringe).
`
`
`
`
`13
`
`Novartis Exhibit 2045.013
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`Corresponding Disclosure
`Moreover, if Novartis contends that Lam does not disclose the claimed limitation, then Lam
`renders this limitation obvious in view of numerous prior art references.
`
` A
`
` person of ordinary skill in the art would have been motivated to combine the teachings of Lam
`with Nema, Chacornac, Furfine, Sigg, and/or Macugen, and would have had a reasonable
`expectation of success in doing so, as these references are in the same technical field. It further
`would have been an obvious design choice. Such a person likewise would have understood that
`such combination was merely the application of a known to a known device would have led to
`predictable results.
`
`For example, see the following passages and/or figures, as well as all related disclosures:
`
`
`
`Nema Vol. 1 at Table 1.
`
`
`
`
`
`
`14
`
`Novartis Exhibit 2045.014
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`Corresponding Disclosure
`
`
`
`Nema Vol. 1 at Figure 9.
`
`
`
`
`Additionally, the siliconizing operation comprising a polymerization step (i) is more
`precise and more homogeneous that a simple standard siliconizing operation; and (ii)
`makes it possible to reduce the amount of silicone that is used (that is, loaded on the inner
`surface of the container) by about a factor of 10 without any loss of lubricating effect. For
`example, according to a standard siliconizing process, from 400 to 1000 μg of silicone are
`deposited in a syringe intended to contain doses of 0.5-1 ml (the total inner surface of the
`0.5-1 ml syringe reservoir is about 8 cm2; in this example this surface corresponds to an
`amount of silicone of from about 50 to 125 μg/cm2), whereas from 40 to 100 μg of
`
`
`
`15
`
`Novartis Exhibit 2045.015
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Corresponding Disclosure
`silicone are sufficient for the same syringe (about 5 to 12 μg/cm2) if silicone is deposited
`on the inner surfaces of the container and then polymerized, for example by heating. The
`fact that the inner surface of the syringe is coated with a low amount of polymerized
`silicone in a more homogenous manner than with a low amount of free silicone allows
`non-siliconized plungers to slide smoothly, whereas such plungers are inoperative with
`syringes coated with low amount of free silicone.
`
`
`Chacornac at [0026].
`
`
`Example 4. Stability of 40 mg/ml VEGF Trap Liquid Formulation Stored at 5° C in Pre-
`filled Glass Syringe.
`
` A
`
` liquid formulation containing 40 mg/ml VEGF trap (SEQ ID NO:4), 10 mM phosphate,
`40 mM NaCI, 0.03% polysorbate 20, 5% sucrose, and pH 6.3, was stored at 5 0 C in 1 ml
`prefilled luer glass syringe with 4023/50 FluroTec coated plunger and samples tested at
`0.5, 1 , 2, 3, and 4 months. Stability results are shown in Table 4. Turbidity, percent
`recovered protein and purity was determined as described above.
`
`
`Furfine at [0059]
`
`
`Example 6. Stability of 40 mg/ml VEGF Trap Liquid Formulation Stored at 5 0 C in 1 ml
`Pre- filled Glass Syringe.
`
` A
`
` liquid formulation containing 40 mg/ml VEGF trap (SEQ ID NO:4), 10 mM phosphate,
`135 mM NaCI, 0.03% polysorbate 20, and pH 6.3, was stored at 5 0 C in 1 ml prefilled
`glass luer syringe with 4023/50 FluroTec coated plunger and samples tested at 0.5, 1, 2, 3,
`4, and 5 months. Stability results are shown in Table 6. Turbidity, percent recovered
`protein and purity was determined as described above.
`
`
`Furfine at [0061]
`
`
`Filling of 0.5 mL syringes was performed in a sterile lab for hydrogen peroxide treatment.
`
`Claim Language
`
`
`
`16
`
`Novartis Exhibit 2045.016
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`[1.e] (b) the syringe barrel
`comprises from about 1 μg to
`100 ug silicone oil,
`
`Corresponding Disclosure
`
`Sigg at 20:20-21
`
`
`Additionally, the oxidative stress exerted on a 0.5% Polysorbate 20 solution in prefilled
`glass syringes (1 mL long, ISO) was investigated by measurement of peroxides according
`to standard protocols.
`
`
`Sigg at 22:8-10
`
`Macugen (pegaptanib sodium injection) is supplied in a single use 1 mL glass syringe with a gray
`rubber plunger containing 0.3 mg in a 90 uL deliverable volume. Each syringe is fitted with a fixed
`gauge needle covered with a gray rubber needle shield and a rigid plastic outside sheath. All are
`contained in a foil pouch. The accompanying polystyrene plunger rod and white flange are in a
`separate foil pouch. The two foil pouches are packaged in a carton.
`
`To the extent Novartis alleges this limitation is not met by any of the disclosures above and/or
`obvious based on the knowledge of a POSITA, it would have been obvious in view of Sigg,
`Boulange, Reuter, Scypinski, Metzner, Shah, Fries, Schoenknecht, Chacornac, Nema, D’Souza,
`Furfine, Badkar, Macugen, Eylea, and/or Lucentis. See A-1–A-2, A-4–A-13, B-1–B-3 and all
`references cited therein.
`It would have been obvious to a POSITA to use a known baked silicone method to siliconize the
`syringes disclosed in Lam, which would result in a syringe that meets this limitation. Lam alone or
`in view of Boulange, Fries, Hioki, Chacornac, D’Souza, Reuter, Nema, Khandke, and/or DC365,
`disclose the syringe barrel comprises from about 1 μg to 100 ug silicone oil.
`
`For example, see the following passages and/or figures, as well as all related disclosures:
`
`
`In many circumstances it would be advantageous to sterilize the surfaces of these objects
`in order to reduce the risk of contamination during subsequent handling. For example,
`there is an increased risk of endophthalmitis after intraocular injection if the surface of the
`syringe used for injection is not sterilized. Thus, there remains a need for efficient and
`cost-effective methods of surface-sterilizing objects containing ethylene-oxide-sensitive,
`
`
`
`17
`
`Novartis Exhibit 2045.017
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Claim Language
`
`Corresponding Disclosure
`temperature-sensitive compounds, such as biological molecules, without a significant
`adverse effect on their activity or integrity.
`
`
`Lam at 1:26-32.
`
`
`
`
`
`
`The invention relates to methods for surface-sterilizing objects containing ethylene-
`oxide-sensitive, temperature-sensitive compounds, such as biological molecules. The
`invention is based, in part, on the surprising discovery of ethylene-oxide -based
`sterilization conditions that will effectively sterilize the surface of an object but which do
`not significantly damage ethylene-oxide-sensitive, temperature-sensitive compounds
`contained inside.
`
`In one aspect, the invention provides a method for surface-sterilizing an object having an
`ethylene-oxide(EtO)-impermeable interior space containing a compound with a
`temperature-sensitive and EtO-sensitive activity by exposing the object to EtO under
`conditions such that the object is surfac

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket