throbber
RESEARCH
`
`Distribution of Silicone Oil in Prefilled Glass Syringes
`Probed with Optical and Spectroscopic Methods
`
`ZAI-QING WEN1*, AYLIN VANCE1, FABIAN VEGA2, XIAOLIN CAO1, BRUCE EU2,
`and ROBERT SCHULTHESIS2
`
`1Department of Formulation and Analytical Resources; and 2Department of Drug Delivery Engineering, Amgen Inc.,
`Thousand Oaks, CA 91320 ©PDA, Inc. 2009
`
`ABSTRACT: Prefilled glass syringes (PFSs) have become the most commonly used device for the delivery of
`recombinant protein therapeutics in parenteral formulations. In particular, auto-injectors preloaded with PFSs greatly
`facilitate the convenient and efficient self-administration of protein therapeutics by patients. Silicone oil is used as
`a lubricant in PFSs to facilitate the smooth motion of the plunger during injection. However, there have been few
`sophisticated analytical techniques that can qualitatively and quantitatively characterize in-situ the morphology,
`thickness, and distribution of silicone oil in PFSs. In this paper, we demonstrate the application of three optical
`techniques including confocal Raman microscopy, Schlieren optics, and thin film interference reflectometry to
`visualize and characterize silicone oil distribution in PFS. The results showed that a container coating process could
`produce unevenly distributed silicone oil on the glass barrel of PFSs. An insufficiency of the amount of silicone oil
`on the glass barrel of a PFS can cause stalling when the device is preloaded into an auto-injector. These analytical
`techniques can be applied to monitor the silicone oil distribution in PFSs.
`
`KEYWORDS: Silicone oil, Prefilled glass syringe, Raman microscope, Interference reflectometry, Zebra Schlieren,
`Protein formulation, Drug delivery device.
`
`Introduction
`
`Prefilled glass syringes (PFSs) have become the pri-
`mary container for protein therapeutics in parenteral
`formulation in recent years (1, 2). They offer several
`advantages over conventional primary containers such
`as glass vials. For example, PFSs eliminate several
`steps that are needed to inject a drug from the vial into
`a patient. They contain the exact dose to be injected,
`unlike in vials, which often are overfilled by up to 25%
`to ensure that the desired dose will be available for
`withdrawal and delivery, which may in turn lead to
`dosing errors. The precisely filled dose in PFSs also
`reduces the waste of precious protein therapeutics.
`The majority of protein therapeutics on the market is
`now contained in PFSs. Moreover, PFSs have been
`incorporated into auto-injectors for self administration
`of the drug by patients and for use in clinical centers
`for convenience and needle safety (3).
`
`* Author for correspondence: email: zwen@amgen.com
`
`In contrast to the many benefits PFSs present to pa-
`tients and physicians, the change of primary container
`from glass vials to PFSs poses great challenges for
`formulation scientists and device and packaging engi-
`neers in the biopharmaceutical industry. The degree of
`complexity of a PFS and the auto-injector pre-loaded
`with a PFS is significantly increased compared to that
`of simple glass vials. Protein therapeutics in particular
`pose potential interactions between the protein and the
`additional materials that comprise the PFS. We have
`observed visible particulates that have formed due to
`the interaction of protein products with foreign matters
`left in PFSs during the PFS manufacturing process (4).
`For instance, metals used to make the syringe needle
`may generate iron rust, which could be hazardous to
`protein products. Residual tungsten left in PFS inter-
`acts with the protein to form visible particles and
`generate defects. Silicone oil is an essential material
`used as a lubricant for proper movement of the plunger
`in PFS. In contrast, vials are essentially free of sili-
`cone oil except for an extremely small quantity around
`the vial stopper for assembly purposes. Silicone oil
`has been observed to induce turbidity and particulate
`formation (5, 6) in insulin and other proteins (7).
`
`Vol. 63, No. 2, March–April 2009
`
`149
`
`Novartis Exhibit 2024.001
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`the performance of an auto-injector de-
`Moreover,
`pends on the reduction of friction in the PFS using
`silicone oil.
`
`It is, therefore, important to characterize the distribu-
`tion of silicone oil in PFSs and to understand the
`complex interaction between protein therapeutics and
`silicone oil used in PFSs. In the present study, we
`investigate silicone oil distribution and its correlation
`with the function of PFSs and auto-injectors by using
`a number of optical techniques including optical mi-
`croscopy, confocal Raman microscopy (8), Schlieren
`optics (9), and interference reflectometry (10). The
`results demonstrate that silicone oil could be unevenly
`distributed on the glass barrel of the PFS, or in ex-
`treme cases it could be completely absent in certain
`areas. The deficiency of and uneven distribution of
`silicone oil can result in the stalling of a PFS in an
`auto-injector. On the other hand, external force placed
`on excessive amounts of silicone oil localized in spe-
`cific areas on the PFS can generate silicone oil micro-
`droplets and contribute to particulate formation asso-
`ciated with proteins. These findings should be helpful
`to PFS manufacturers for improving the silicone oil
`coating process in PFSs and improving the overall
`quality of PFSs used in auto-injectors.
`
`Experimental Design
`
`Materials
`
`PFSs with 1.0-mL barrels were obtained from a vendor
`without further treatment. (BD, Franklin Lakes, NJ)
`Syringes from three lots were selected as samples.
`Two lots were coated with silicone oil. However, one
`lot coated with silicone oil showed poor performance
`when the syringes were loaded onto the auto-injector
`for auto-injection testing. The other lot exhibited good
`performance. The third lot of syringes was not coated
`with silicone oil and was used to provide control
`samples. None of the syringes in these three lots were
`filled with product unless it was otherwise noted that
`a product-filled syringe had stalled during auto-injec-
`tion testing. The product in the syringe was not emp-
`tied when it was examined using the Schlieren optical
`method.
`
`Methods
`
`We employed three optical techniques to visualize the
`silicone oil and measure in-situ the thickness of the
`silicone oil droplets. A confocal Raman microscope
`
`and Schlieren optics were used to observe the silicone
`oil droplets on the glass barrel. The Raman spectrum
`of silicone oil was used to confirm its identity. A
`visible reflectometry based on the light interference of
`transparent thin film (10) was applied to determine the
`thickness of silicone oil and its distribution along the
`longitudinal direction of the syringe barrel. The sili-
`cone oil coated on prefilled syringes could be treated
`as transparent thin film.
`
`Optical Microscope: A Carl-Zeiss stereomicroscope
`(Stemi 2000C, Carl-Zeiss, Thornwood, NY) was em-
`ployed to take micrographs of the empty glass sy-
`ringes.
`
`Confocal Raman Microscope: The optical micro-
`scope on the confocal Raman microscope (Senterra威,
`Bruker Optics, Billerica, MA) was employed to exam-
`ine the silicone oil distribution, and the Raman module
`was used to confirm the identity of the silicone oil.
`Three objective lenses attached to the nosepiece with
`20⫻, 50⫻ and 100⫻ magnification, corresponding to
`a spatial size of 5, 2, and 1 ␮m, respectively. Lower
`magnification objective lenses have a larger field of
`view and working distance, which enables ease of
`observation and sample handling. The total magnifi-
`cation is determined by the product of the objective
`lens and binocular eyepieces (10⫻). The syringe was
`placed on a XY-stage, which is precisely controlled by
`micrometers on the microscope. The light beam pro-
`jected onto the sample syringe was vertically focused
`with a 4⫻ objective lens.
`
`The Raman spectrometer module accommodates two
`lasers for exciting Raman scattering: a red 785-nm
`diode laser and a green 532-nm green solid laser to
`facilitate the measurement of samples with different
`scattering natures. Variable laser power can be se-
`lected depending upon the sample signal. The maxi-
`mum laser power is 100 mW for the 785-nm laser and
`20 mW for the 532 nm laser. The spectral coverage is
`from 70 cm⫺1 to 3700 cm⫺1 with a spectral resolution
`of 4 cm⫺1. Two special functions were implemented
`on the Raman microscope: the FlexFocus™ for con-
`focal Raman detection and SureCal威 for continuous
`automated frequency calibration. Both play an impor-
`tant role in the microscope’s performance. The confo-
`cal function provides a significant
`improvement
`in
`both the contrast and the spatial resolution in the
`vertical (z) axis. The properly adjusted confocal opti-
`cal configuration offers the capability of “optical sec-
`tioning” and depth profiling of heterogeneous and thin
`
`150
`
`PDA Journal of Pharmaceutical Science and Technology
`
`Novartis Exhibit 2024.002
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Zebra strip
`
`Syringe Lamps
`
`Optical Microscope Camera
`
`Figure 1
`
`Schematic of a Zebra Schlieren optic set up for visualization of silicone oil in PFS. The light emitted from lamps
`is projected toward the Zebra strip passing through the syringe standing on the holder. The reflected light from
`the syringe is collected by the microscopic lens and a picture is then taken by a digital camera.
`
`specimens, which is crucial for the successful obser-
`vation of silicone oil micro-droplets on the glass barrel
`of PFSs.
`
`Zebra Schlieren Visualization: A Zebra Schlieren
`optics technique was developed to visualize silicone
`oil on the glass syringe barrel. This technique was
`used to observe the optical inhomogeneity in transpar-
`ent media that
`the human eye cannot directly see
`because of an extremely small difference of refractive
`index (9). It exploits the Schlieren phenomena in solid,
`liquids, and gases that have refractive-index gradients
`in one, two, and three dimensions due to inhomoge-
`neity or resulting from temperature change, high-
`speed flow, or the mixing of dissimilar media. This is
`a qualitative visualization method to display the opti-
`cal
`inhomogeneity of a material system. Schlieren
`optics has been widely employed in many sciences and
`engineering fields,
`including the detection of gas
`leaks, jet stream, and liquid flow (9). However, to the
`best of our knowledge,
`this is the first
`time that
`Schlieren optics has been applied to view silicone oil
`droplets on PFSs. The PFSs coated with silicone oil
`
`have a refraction-index gradient along the glass sy-
`ringe barrel due to the presence of silicone oil drop-
`lets. Both silicone oil and the glass barrel are trans-
`parent
`to visible light; however,
`the density and
`refractive index of silicone oil and glass differ from
`each other in that the silicone oil used in PFSs as a
`lubricant has a refractive index of ⬃1.42 at 20 °C,
`while the refractive index of the glass barrel is close at
`⬃1.5. The PFS coated with silicone oil constitutes a
`Schliere. The small difference between the refractive
`index of the silicone oil coated on the boronsilica glass
`barrel makes it difficult to see the silicone oil with the
`naked eye or under a low-resolution optical micro-
`scope.
`
`The optical set-up to observe a Schliere is fairly sim-
`ple, yet it can visualize the silicon oil distribution over
`the PFS in a practical and dramatic manner. Figure 1
`shows a schematic of the optical configuration for the
`Schlieren visualization of silicone oil on the glass
`barrel. A zebra strip pattern was placed in front of a
`screen. This is the key component allowing the
`Schlieren optics to dramatically reveal micron-size
`
`Vol. 63, No. 2, March–April 2009
`
`151
`
`Novartis Exhibit 2024.003
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`droplets of silicone oil on a glass barrel. A glass
`syringe was placed vertically on a syringe holder
`between the zebra strip and a high resolution digital
`optical microscope
`(model VHX-600, Keyence,
`Woodcliff Lake, NJ). Two projector fluorescence
`lamps were placed on the left and right sides of the
`digital microscope to illuminate the syringe. All other
`background light was minimized to reduce disruption.
`The images of the syringes on whose inner surfaces
`silicone oil had been distributed were taken with a
`digital camera attached to the optical microscope. All
`pictures were recorded on a computer without further
`processing.
`
`Interference Reflection Spectroscopy: A rap.ID
`Layer Explorer (rap.ID Particle Systems GmbH, Ber-
`lin, Germany), based on the principle of light interfer-
`ence of thin film and reflection spectrometry, was
`employed for the measurement of silicone oil thick-
`ness. The Layer Explorer employs a beam of white
`light that is projected onto the glass syringe, and the
`reflected light from the thin oil film on the syringe
`surface is collected with an optical fiber probe (10).
`The interference pattern generated by the interaction
`of white light with the thin silicone oil film was
`calculated to determine thickness. The instrument has
`a precisely controlled and automated XY-stage. The
`syringe was laid horizontally on the syringe holder,
`which is fixed on the XY-stage. The software auto-
`matically divides the length of the syringe along the
`longitudinal direction to the number of measurement
`points and moves automatically along the syringe and
`collects the data on each point. It takes only a few
`seconds to measure one point. The size of the spot
`measured was ⬃100 ␮m. The detection limit of the
`thin film thickness was 50 nm. Below 50 nm the
`thickness was displayed as zero because the optical
`interference will not occur when the film thickness is
`much smaller than the wavelengths of visible light.
`
`Results
`
`Microscopic Image of Silicone Oil in PFS
`
`Figure 2 shows the layout of a glass syringe on the
`confocal Raman microscope’s XY-stage, this syringe
`having been selected from a “syringe lot” that showed
`low performance in the syringe injection test. The
`syringe was not filled with product solution. A total of
`30 images were taken of one syringe and are numbered
`accordingly from the plunger side to the needle end
`along the longitudinal direction. We selected 15 out of
`
`1, 2, 3
`Figure 2
`
`29, 30
`
`The layout of the syringe on the XY-stage of the
`Raman Microscope. Each number represents a
`1.5-mm length of a segment of the syringe barrel of
`which a micrograph was taken. A total of 30 mi-
`crographs were taken sequentially with 15 of them
`exhibited in Figure 3.
`
`the 30 images to show the silicone oil distribution on
`the barrel. Figure 3 shows the 15 individual images
`consecutively from the plunger side to the needle side
`of the glass barrel sections. Each picture corresponds
`to a 1.5-mm section of the syringe barrel. The silicone
`oil was seen as micro-droplets surrounded with a thin
`film layer, which under light illumination exhibited a
`typical thin film interference pattern. They resemble
`the Newton rings generated by a thin oil film on water
`under sunlight. These droplets and thin film layers
`were confirmed by Raman spectroscopy to be silicone
`oil. Figure 4 shows the Raman spectra of silicone oils
`on the syringe barrel and of a droplet of reference
`silicone oil on a glass slide. The Raman spectrum of
`the silicone oil droplets on the glass barrel was noisy,
`but the fingerprint pattern unambiguously confirmed
`that it was silicone oil.
`
`The most striking observation was that the distribution
`of silicone oil on the barrel syringe was unevenly
`spread from one end (plunger side) to the other (needle
`side) throughout the syringe barrel. For the syringe
`examined, the first few segments of the syringe barrel
`on the plunger side contained much more silicone oil
`than did segments in other areas. It can be seen that
`there are many small silicone oil droplets that are
`surrounded with the thin layer. However, towards the
`needle end, there were fewer oil droplets with no thin
`oil film surrounding them. The last few segments at
`
`152
`
`PDA Journal of Pharmaceutical Science and Technology
`
`Novartis Exhibit 2024.004
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`2
`
`4
`
`6
`
` 10
`
`12
`
`14
`
`16
`
` 18
`
`20
`
`22
`
`24
`
`Figure 3
`
`26
`
`28
`
` 30
`
`Micrographs of a glass syringe barrel obtained using the confocal Raman microscope. Each image from left to
`right represents a 1.5-mm section of the syringe barrel. An uneven distribution of silicone oil from the plunger
`side to the needle end can be seen.
`
`Vol. 63, No. 2, March–April 2009
`
`153
`
`Novartis Exhibit 2024.005
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Figure 4
`
`Raman spectra of silicone oil on the glass barrel of the PFS (top trace) and of a reference spectrum of silicone
`oil on a glass slide (bottom trace). The positive and negative spikes on the top trace are due to cosmic rays.
`
`the needle side contained the minimum amount of
`silicone oil, and some of the segments showed virtu-
`ally no silicone oil at all. This observation was con-
`sistent with the results of the toluene zone extraction
`of silicone oil obtained with size exclusion chroma-
`tography analysis (11) of syringes with poorly distrib-
`uted silicone oil, showing that the silicone oil is rich
`on the plunger section but poor at the needle-side
`section. They are also consistent with the results ob-
`tained using the gold nano-particle staining method,
`which demonstrated that, for unevenly coated PFSs,
`the amount of silicone oil was deficient at the needle
`end (12). A significant increase of extrusion force at
`the end of the injection was observed when unevenly
`coated PFSs were tested with an Instron Mechanical
`Testing Systems Model 5564 (Instron, Norwood,
`MA). This was attributed to the protein adsorption on
`bare glass surfaces that were deficient of silicone oil
`droplets (12).
`
`Zebra Schlieren Visualization of Silicone Oils
`
`Figure 5 presents images of three representative glass
`syringes that show different silicone oil distributions
`on the glass barrel as revealed by Schlieren optics. The
`N3 syringe was from the syringe lot
`that had no
`silicone oil coating, showing no silicone oil droplets at
`all. The P3 syringe represents the uniformly distrib-
`uted silicone oil on the barrel surface from the syringe
`
`lot that showed good performance in an auto-injection
`test. The A3 syringe shows an uneven distribution of
`silicone oil from the silicone oil-coated syringe lot that
`showed poor injection performance with an auto-in-
`jector. The silicone oil droplets can be clearly seen in
`the two syringes (A3 and P3) that have been coated
`with silicone oil. However, unusually large silicone oil
`droplets are seen (see arrow on A3 in Figure 5) on the
`syringe barrel that has uneven oil distribution. Some
`large silicone oil droplets are estimated to be a few
`millimeters in size, which is apparently too big for
`uniform coating. For the PFS with more uniform dis-
`tributions of silicone oil (P3), the droplets are smaller
`and more uniform. However, they are still seen as
`micro-droplets instead of as a uniform thin film. These
`results are consistent with the pictures obtained with
`the confocal Raman microscope. Schlieren optics,
`thus, provides a quick and dramatic visualization
`method for silicone oil distribution on glass barrels
`and could be developed into a screening technique for
`silicone oil distribution on PFSs for quality control
`and selecting PFSs for auto-injector assemblies. How-
`ever, Schlieren optics is not a quantitative analytical
`method that can determine the thickness of silicone
`oils. For that purpose, we resorted to interference
`reflectometry that can measure the thickness of sili-
`cone oil on a glass barrel, as described in the following
`section.
`
`154
`
`PDA Journal of Pharmaceutical Science and Technology
`
`Novartis Exhibit 2024.006
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`Figure 5
`
`Visualization of the silicone oil used in glass syringes. N3 is not coated with silicone oil. P3 is coated evenly with
`silicone oil. A3 is coated unevenly with silicone oil.
`
`Thickness of Silicone Oil
`
`Figure 6 shows the thickness distribution of silicone
`oil on two syringe barrels in the longitudinal direction
`as obtained using the rap.ID Layer Explorer. The two
`syringes were selected from the syringe lot
`that
`showed poor results in auto-injection testing. The
`thickness of the silicone oil on these two specific
`syringes varies from ⬃100 nm to ⬃700 nm, with a
`median thickness of about ⬃200 nm. However, some
`spots were as thick as 700 nm while other spots show
`no detectable silicone oil (the detection limit is 50
`nm). These quantitative data, in general, are consistent
`with the results obtained from the Schlieren visualiza-
`tion and confocal Raman microscopic data demon-
`strating that the silicone oil was unevenly distributed
`as droplets.
`
`The theoretical thickness of silicone oils on the barrel
`surface can be calculated with the assumption that a
`uniform layer was formed as follows:
`
`
`
`t ⫽ Ri ⫺ 共Ri2 ⫺ m/␲␳h兲1/ 2
`
`(1)
`
`where
`
`t is the thickness of silicone oils on the syringe barrel
`
`Ri is the inner diameter of the syringe barrel
`
`␳ is the density of silicone oil
`
`h is the length of the syringe barrel
`
`For a glass syringe having a barrel length of 50 mm
`and an inner diameter of 6 mm that is coated uniformly
`
`Vol. 63, No. 2, March–April 2009
`
`155
`
`Novartis Exhibit 2024.007
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`1
`
`3
`
`5
`
`7
`
`9 11 13 15 17 19 21 23 25 27 29 31 33 35
`
`Position on Syringe
`
`800
`
`700
`
`600
`
`500
`
`400
`
`300
`
`200
`
`100
`
`0
`
`800
`
`700
`
`600
`
`500
`
`400
`
`300
`
`200
`
`100
`
`0
`
`Thickness (nm)
`
`Thickness (nm)
`
`1
`
`3
`
`5
`
`7
`
`9 11 13 15 17 19 21 23 25 27 29 31 33 35
`
`Position on Syringe
`
`Figure 6
`
`Silicone oil thickness profile along the longitudinal direction of two syringe barrels from plunger end (left) to
`needle end (right).
`
`with 0.4 mg of silicone oil (the thickness of silicone
`oil coating recommended by major PFS manufactur-
`ers), the mean thickness of the silicone oil would be
`⬃430 nm. This means that the silicone oil in some
`areas was nearly twice as thick as the mean value. On
`the other hand, few places showed a complete absence
`of silicone oil. The average measured thickness of 200
`nm was far less than the theoretical mean thickness of
`⬃430 nm. This again suggests that the silicone oil was
`not uniformly distributed. We also measured the thick-
`ness of silicone oil on the syringe barrel at four radial
`positions by rotating the syringe at 45, 90, 135, and
`180 degrees. The distribution pattern of silicone oil
`
`was found to be similar along the longitudinal direc-
`tion at these four angles (data not shown). However, if
`the syringe was placed horizontally for a period of
`time, the silicone oil migrated to the bottom of the
`syringe due to gravity. The thickness of the silicone oil
`at the bottom would be significantly greater than that
`at the glass barrel’s top and sides because the silicone
`oil was not covalently bound to the glass surface.
`
`It has been observed that some prefilled syringes pre-
`loaded onto the auto-injector can be stalled during
`auto-injection testing (12). The auto-injector testing
`for the spring force without loading the PFSs showed
`
`156
`
`PDA Journal of Pharmaceutical Science and Technology
`
`Novartis Exhibit 2024.008
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`consistent performance, which ruled out the malfunc-
`tion of
`the auto-injector. Furthermore, PFSs with
`fairly even distributions of silicone oil (as seen in P3
`in Figure 5) showed no stalling when they were as-
`sembled on an auto-injector and were tested for injec-
`tion. We surmised that the PFS stalling in auto-injec-
`tion was caused by the uneven distribution and
`deficiency of silicone oil on the syringe barrel. The
`stalling of PFSs preloaded onto the auto-injector was
`due to a lack of appropriate lubrication. Several lines
`of evidence support this hypothesis. First, the defi-
`ciency of silicone oils in certain sections on an un-
`evenly coated glass syringe, in particular towards the
`needle end, resulted in high friction force when they
`were tested with the Instron mechanical tester (12).
`Second, post-examination of a Protein X PFS where
`the auto-injector had stalled revealed that there are
`proteins adsorbed on the 1/5 section toward the needle
`end of the barrel. This constitutes indirect evidence of
`the deficiency of silicone oil on this specific section, as
`proteins would preferentially adsorb to glass surfaces
`that were not coated with silicone oil. Third, when two
`stalled PFSs preloaded on auto-injectors were disas-
`sembled and examined using the optical microscope
`and Schlieren visualization, they were found to have
`no silicone oil at all in the PFSs. Figure 7 shows such
`a picture of a stalled PFS that was filled with product
`and then removed from the auto-injector and disas-
`sembled. The picture was taken without removing the
`liquid content after disassembly. Examination using
`Zebra Schlieren optics clearly displays a minimum
`amount of silicone oil on the barrel. We also examined
`the syringe needle hole using a fine tungsten probe
`after the liquid content was removed and found that
`the needle hole was not clogged, this ruling out the
`possibility that a clogged needle had stalled the sy-
`ringe. Finally,
`the above evidence leads us to the
`conclusion that PFS without a silicone oil coating
`would not be able to function smoothly for injection.
`
`Concluding Remarks
`
`We have demonstrated three optical techniques, includ-
`ing confocal Raman microscopy, Schlieren optics, and
`interference reflectometery, as powerful tools to display
`silicone oil distribution, to confirm the identity of the
`coating, and to determine the thickness of silicone oil
`film on the glass barrel of PFS. The first two methods are
`qualitative yet very practical in visualizing silicone oil
`distribution on PFSs. In particular, Schlieren optics is
`very effective in directly observing the silicone oil drop-
`lets on PFSs. These methods may be utilized in a man-
`
`Figure 7
`
`A stalled syringe filled with protein-X solution dis-
`assembled from an auto-injector. Zebra Schlieren
`optics reveals an absence of silicone oil droplets on
`the glass barrel. The picture was taken with the
`liquid content in the syringe.
`
`ufacturing environment to pre-screen the quality of PFSs
`in regard to silicone oil uniformity and so may help to
`ensure the quality of PFSs intended for use in auto-
`injectors. Interference reflection spectrometry can pre-
`cisely measure silicone oil thickness and distribution on
`the glass barrel. All three techniques are non-destructive
`and easy to carry out, and together they offer a set of very
`effective optical tools to characterize silicone oil coat-
`ings on glass surfaces. All three techniques were able to
`show that the silicone oil coating on a syringe’s glass
`barrel was unevenly distributed, accounting for the poor
`performance of some glass syringes. A deficiency of
`silicone oil on the glass barrel can cause the plunger in a
`PFS to stall during injection using the auto-injectors. The
`availability of sophisticated analytical tools to character-
`ize the silicone oil on PFSs in situ can aid in creating a
`uniformly distributed silicone oil coating, which in turn
`
`Vol. 63, No. 2, March–April 2009
`
`157
`
`Novartis Exhibit 2024.009
`Regeneron v. Novartis, IPR2020-01317
`
`

`

`is expected to improve the performance of PFSs for
`protein pharmaceuticals.
`
`Acknowledgements
`
`We are grateful to rap.ID Particle Systems GmbH for
`the use of the Layer Explorer to measure silicone oil
`thickness; and to Drs. Robin Hwang and Ron Forster
`for constructive discussion during this work and Drs.
`Linda O. Narhi and David Brems for critically reading
`the manuscript.
`
`References
`
`1. Karras L.; Wright, L.; Cox, L.; Kouns, T.; Akers,
`M. J. Current issues in manufacturing and control
`of sterile prefilled syringes. Pharm. Technol.
`2000, 24, 188 193.
`
`2. Polin, J. B. The ins and outs of prefilled syringes.
`Pharmaceutical & Medical Packaging News
`2003, 11, 40 – 44.
`
`3. Dahlgren, J.; Veimo, D.; Johansson, L.; Bech, I.
`Patient acceptance of a novel electronic auto-
`injector device to administer recombinant human
`growth hormone: results from an open-label, user
`survey of everyday use. Current Medical Re-
`search and Opinion 2007, 23, 1649 –1655.
`
`5. Baldwin, R. N. Contamination of insulin by sili-
`cone oil: a potential hazard of plastic insulin sy-
`ringe. Diabet. Med. 1988, 5 (8), 789 –790.
`
`6. Chantelau E. Silicone oil contamination of insu-
`lin. Diabet. Med. 1989, 6, 278.
`
`7. Jones, L. S.; Kaufmann, A.; Middaugh, C. R.
`Silicone oil
`induced aggregation of proteins.
`J. Pharm. Sci. 2005, 94 (4), 918 –927.
`
`8. Wen, Z. Q. Raman spectroscopy of protein phar-
`maceuticals. J. Pharm. Sci. 2007, 96, 2861–2878.
`
`9. Settles, G. S. Schlieren & Shadowgraph Tech-
`niques, 2nd ed.; Springer-Verlag: Berlin, 2006.
`
`10. Flaherty, T.; O’Connor, G. The application of
`spectral reflectivity to the measurement of thin
`film thickness. Proc. SPIE (Society of Photo-
`graphic-Instrumentation Engineer) or The Inter-
`national Society for Optical Engineering) 2003,
`4876, 976 –983.
`
`11. Lee, H.; Fujimori, K.; Liu, D.; Perez, L.; Nashed,
`Y. 2008. Extraction and determination of silicone
`oil in glass syringes by HPLC-evaporative light
`scattering detection. To be published.
`
`4. Wen, Z. Q.; Torraca, G.; Yee, C. Y.; Li, G. Inves-
`tigation of contaminants in protein pharmaceuticals
`in pre-filled syringes by multiple micro-spectrosco-
`pies. Am. Pharm. Rev. 2007, 10, 101–107.
`
`12. Eu, B.; Cairns, A.; Hwang, R.; Cao, X.; Wen,
`Z. Q. Direct Visualization of Protein Adsorption
`in Pre-filled Syringes by Gold Nano-particles.
`AAPS-Bio-Conference, Toronto, 2008.
`
`158
`
`PDA Journal of Pharmaceutical Science and Technology
`
`Novartis Exhibit 2024.010
`Regeneron v. Novartis, IPR2020-01317
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket