throbber
Validation of
`
`Pharmaceutical
`
`Processes
`
`Third Edition
`
`BRIHSH LIBRARY
`DOCUMENW RUPPIY CENTRE
`
`2 1 JUL 2009
`
`_ m09].
`-— 2597a
`
`—-
`"‘
`
`Edited by
`James Agalloco
`Age/loco & Associates
`Bel/e Mead. New Jersey. USA
`
`Frederick J. Carleton
`
`Carleton Technologies Incorporated
`Boynlon Beach. Florida. USA
`
`informa
`healthcare
`Mr.» Inn tendon
`
`Regeneron Exhibit 1052.001
`
`

`

`Infomia llealthcare USA. Inc.
`:2 Vanderbilt Avenue
`New York, N‘i 10017
`
`I." 2008 by Informa Healthcare USA, Inc.
`Int'nrma Healthcare is an Informa business
`
`No claim to original L33. Government works
`I'rinted in the United States of America on acid-tree paper
`lll987b§4321
`
`International Standard Book Number-l0: 0—84934055-3 (Hardcover)
`International Standard Book Number-13: 978-084934055-7 (Hardcover)
`
`This book contains information obtained from authentic and highly regarded sources. Reprinted material is quoted with
`permission, and sources are indicated. A wide variety of references are listed. Reasonable efforts have been made to publish
`reliable data and information, but the author and the publisher cannot assume responsibility for the validity of all materials or
`for the consequence of their use.
`
`No part of this book may be reprinted, reproduced, transmitted, or utilized in any form by any electronic, mechanical, or other
`means, now known or hereafter invented. including photocopying, microfilming, and recording, or in any information storage
`or retrieval system. without written permission from the publishers.
`
`For permission to photocopy or use material electronically from this work, please access www.copyright.com (httpzf/www.
`copyrightcom/l or contact the Copyright Clearance Center, Inc. (CCC )i 222 Rosewood Drive. Danvers. MA 01923, 978—750-8400.
`CCC is a not-for—profit organization that provides licenses and registration for a variety of users. For organizations that have been
`granted a photocopy license by the CCC, a separate system of payment has been arranged.
`
`Trademark Notice: Product or corporate names may be trademarks or registered trademarks. and are used only for
`identification and explanation without intent to infringe.
`
`
`Library of Congress Catalogingdn-Publicatinn Data
`
`Validation of pharmaceutical processes / edited by James P. Agalltx‘o, Frederick I. Carleton.
`- 3rd ed.
`P. J (m.
`Includes bibliographical references and index.
`[SEN-13: 97WS-l93-7055—7 (hardcover : alk. paper)
`ISBN-10: 0-8493—7055-8 (hardcover : alk. paper)
`I. Sterilization.
`
`2. Pharmaceutical technology—Quality control.
`3. Pharmaceutical technology—Standards.
`I. Agalloco, James I’., 1948- ll. Carleton, Frederick J., I923
`[DNLM2 1. Technology, Pharmaceutical. 2. Quality Control. 3 Sterilization—methods. Q's" 778 V172 2007"]
`R5] 99.573V345 2007
`615'. l q—chZ
`
`20070 I 821%
`
`Vi5it the Informa Web site at
`www.infonna.com
`
`and the Informa Healthcare Web site at
`www.informahealthcare.com
`
`Regeneron Exhibit 1052.002
`
`

`

`
`
` This material may be protected by Copyright law (Title 17 U.S. Code)
`
`
`16—
`
`Validation of Ethylene Oxide Sterilization Processes
`
`Iohn R. Gillie
`SGM Biotech. inc. Bozeman. Montana. U.S.A.
`
`Gregg Mosley
`Biotest Laboratories. Inc, Minneapolis. Minnesota. USA
`
`INTRODUCTION
`
`The recent history of EU sterilization has been dominated
`by advances in engineering technology These advances
`include not only the computerized controls for the
`operation of sterilizers, but also the physical environ-
`mental controls that permit safe use of 100% E0 gas.
`Most of these improvements have been driven by econ-
`omics. Faster. cheaper processes are indeed worthy
`objectives. However. sometimes it seems we have lost
`the focus on what we are really attempting to accomplish
`in the sterilization processes. Any sterilization process
`must deliver a lethality that kills the naturally occurring
`bioburden microbes that contaminate the products and
`materials. if the process does not render the products or
`materials free from living microorganisms, then steriliza-
`tion has not been achieved.
`
`The microbiological dimension of the EC) process
`has been overwhelmed by the recent strides in
`engineering and the physical process controls. The
`increasing complexity of medical products would
`be much more difficult
`to sterilize without
`these
`
`improvements.
`corresponding engineering, process
`However. failure to properly address microbial lethality
`renders all of these engineering advancements. mean-
`ingless it the resulting product is not sterile.
`Prix‘ess validation means establishing by obiective
`evidence that a process consistently produce» a result or
`product meeting its predetermined specifications ( l).
`The EU sterilization process is expected to deliver
`sterile products that possess all other specified quality
`attributes. Validation must document all critical process
`controls. The products to be sterilized must be challenged
`with an appropriate microbial system located in the
`"Worst case" or "least
`lethal" product
`location.
`In
`
`‘il'l'fl'i'hlfmlt‘ mm! in ”“5 chapter. AAMl, .-\:~,~'ocialion for the -\dvance—
`munl of Medical Instrumentation; Lil, biological
`indicator BIER.
`biological
`indicator evaluator reshtometi-r, ("(13, carbon dioxide;
`DEC. dynamic environmental conditioning, DLFT, device under
`test; F0. E10. ethylene oxide; FDA, Food and Drug Administration:
`(EC, gas l hromamgraphv, ll‘, inoculated product. IR, mtrared; MW,
`molecular in eight; NlUSi l, National institute of Occupational Safety
`and Health; \llm’. National Institute of Science and Tet hnologv; 1“qu
`permissible mposure limits, RH, relative humidity; R'l‘l“), resistance
`temperature detector. SAC, static atmoiepheric conditioning; SAL,
`sterility assurance level, Si R. spore lo} reduction: TAR. test accuracy
`ratio; Tt'.
`thermocouple. TISR. test uncertaintv ratio. HV‘A. time-
`“ t’tgl‘l ted .n eraize.
`
`addition, this microbial challenge product must be posi-
`tioned in the worst case,
`least lethal
`locationts) in the
`production load.
`If the microbial challenges are not
`located in these least lethal locations, then the resulting
`documented evidence may be biased and result in false
`Conclusions about the adequacy of the sterilization vali—
`dation program.
`The validation of the E0 gas sterilization process is
`one of the more complm programs facing process engin-
`eers and micmbiolngists because some critical process
`parameters are interactive. EO gaseous sterilization has
`been shown to be an extremely effective process that can
`be performed with an infinite number of combinations
`of parameters. Key parameters that affect sterilization
`efficacy are (1') concentration of E0 gas.
`(ii) RH,
`(iii) temperature of the process,
`(it’) accessibility of
`the product and packaging for these parameters, and
`tr) time.
`A validation program must demonstrate that the
`selected combination of these interactive process pare
`ameters result
`in an effective physical and biological
`process. The effectiveness of this process is measured by
`calibrated physical instruments and a calibrated micro—
`bial challenge. These process parameters must then be
`correlated to a calculated SAL for the product. SAL is the
`probability of a single i iable miermirganism occurring on
`a product. The required assurance level may vary
`depending on the product itself or the end use of the
`product. but is typically less than one chance in a million
`of a non-sterile unit or SAL of lit
`Another challenge is the task of assuring that
`the E0 gas used does not create a health hazard
`for
`the employees in the working area or
`leave
`unacceptable residuals in the product delivered to the
`consumer. Adsorbed E0 gas is removed fairly rapidly
`from processed materials, while absorbed E0 gas is
`released much more slowly. This absorption rate is
`highly dependent on the specific process conditions.
`material being, processed, as well as the geometry of the
`product, which affects material surface—to—volume ratios.
`Appropriate measures must also be taken to assure that
`E0 gas used in the sterilizing environment is Controlled
`and contained so that environmental insult in affected
`work areas is within acceptable regulated limits.
`During the E0 gas sterilization proceees, the gas
`interacts with the materials processed by reaction,
`absorption or adsorption. The EU gas is also trapped in
`the air spaces within the product or material being
`
`Regeneron Exhibit 1052.003
`
`

`

`242
`
`m
`
`STERILIZATION SANITIZATION AND STERlLIT‘i’ ASSURANCE
`
`sterilized. Unreacted residual gas is rapidly removed
`through evacuation, heated nitrogen or air exchanges.
`Product
`that
`is removed from a sterilizer must be
`
`to the
`insult
`controlled to prevent environmental
`workers, The best procedure is to place the sterilized
`materials in an environment that aids the desorption of
`the gas and is environmentally controlled to minimize
`workplace contamination.
`
`CHARACTERISUCS 0F El]
`
`Chemical Properties
`l, l-epoxyethane, and
`ED is also referred to as EtO,
`dimethylene oxide (2).
`It has a formula of CgHiO. The
`following structure is illustrated:
`
`rug—CH:
`0
`
`is a colorless gas, with a molecular weight of
`it
`44.05. It has a characteristic ether—like odor at toxic levels.
`
`EU has a boiling point of lt).7“C (51.3°Fi at 760 mmHg
`pressure, a melting point of —ilZ.6”C (, 1707?), a
`specific gravity of 0.87“ apparent at 20"C (60’1"), or a
`specific gravity of 0.897 at PC. E0 has a vapor density of
`1.5, with dry air being, equal to 1.0. and a vapor pressure
`at ZUT of [095 mmHg. It is completely miscible in water,
`alcohol. acetone, benzene, ether, carbon tetrachloride.
`HCFCs, and most organic solvents, and is a powerful
`solvent for fats, oils, greases, waxes. some rubber formu-
`lations, and paints. it is highly exothermic and potentially
`explosive when heated or mixed with (i) alkali metal
`hydroxides.
`(ii) highly active catalytic surfaces such
`as anhydrochlorides of iron,
`tin. or aluminum, and
`(iii)
`the oxides of iron and aluminum. The explosive
`limits are 3"?"
`to 97% by volume in air.
`It has a flash
`point of —ti“C (20”F).
`It
`is relatively noncorrosive for
`materials. E0 is relatively stable in neutral aqueous
`solutions and when diluted with liquid or gaseous
`carbon dioxide or halocarbons such as HCFCs. E0 is
`
`relatively unstable in either acidic or alkaline aqueous
`solutions and may rapidly form ethylene glycol.
`
`Biological Activity
`EO. reacts irreversibly with numerous chemical moieties
`on cellular molecules by an alkylation reaction where the
`[CHgOH-{Hg‘l alkyl group is covalently bonded with
`the available moiety via an addition reaction. Reactions
`with ~NH3, —SH, —COOH. and (Ii—[30H groups are
`common and illustrated in Figure l (3).
`Reaction rates vary and depend on the Specific pic.
`for each moiety and the existent pH. For a more compre-
`hensive review of possible reactions we refer the reader to
`Russell H). First-order lethality kinetics require that only
`one molecule per cell is the critical target (5—7). Reactions
`other than the critical reaction leading to microbial
`inactivation must be considered collateral damage
`reactions. Not all microbial
`inactivation obeys firsts
`order kinetics. However, even where multiple sites or
`molecules may be required for inactivation, the concept
`regarding critical reactions and collateral reactions is the
`same. Where inactivation is the result of cumulative
`
`then some
`damage, which is not first-order kinetics.
`damaging reactions must be considered more important
`to the events leading to microbial inactivation tcritical)
`
`
`E0 Alkylatlon Action
`
`
`
`NH-CchHQOH
`S-CHZCHQOH
`I
`
`.
`
`CH2
`
`COO-CHZCHEOH
`
`Illustration of the alkylation reaction of ethylene oxide
`Figure 1
`with chemically active moieties in the bacterial coil.
`
`than other reactions (collateral). Winaro and Stumbo (ti)
`identified E0 reactions with DNA as the critical reactions
`
`resulting in microbial inactivation. Lawley and Brooke:
`(9) defined specific reactions of ED with the nucleic acid
`tertiary heterocyclic nitrogen sites (=i\i-i in numerous
`experiments resulting in more than 10 publications
`between l957 (l0) and 1963 where they state (9):
`1. Sites in the nucleic acids reactive towards alkylating
`agents are shown to be,
`in order of decreasing
`reactivity: for RNA, N—7 of guanine. N-l of adenine.
`N-l of cytosine and N-3 of adenine for DNA, N-7 of
`guanine. N—3 of adenine and N-l of cytosine.
`Denatured DNA behaves in this respect like RNA.
`The observed differences between DNA and RNA are
`ascribed to the involvement of N-l of adenine and of
`
`[-J
`
`3.
`
`cytosine in hydrogen bond formation in DNA.
`In all cases alkylation results in destabilization of the
`nucleosides or the corresponding moieties in the
`nucleic acids. At neutral pH, with DNA, 7-alkyl-
`guanines and 3—all-tyladenines are slowly liberated
`by hydrolysis, the latter at the greater rate, whereas
`with RNA slow rearrangements occur,
`l-alkyla-
`denim: moieties yielding b-methlaminopurine
`moieties and l-alkylcytosines giving the corre-
`sponding l-alkyluracils.
`More recent studies suggest that disruption of the
`DNA molecule may occur differently depending on
`various repair mechanisms (ll).
`In the case of certain
`repair mechanisms, the reactions with cytosine may be
`the injury which ultimately leads to the inactivation of
`the microbe.
`
`VALIDA'I'IDN UF E0 STEHILIZATIDN PROCESSES
`
`Validation of the E0 process is divided into two phases:
`Engineering Qualification and Process Qualification.
`When these activities are completed successfully and all
`aspects of the process are documented. the process can be
`certified for routine use for manufacturing goods.
`
`Engineering Qualification
`Engineering Qualification deals with the sterilizer and
`associated equipment used in the process. This phase is
`divided into three segments: Installation Qualification,
`Calibration, and Operational Qualification.
`
`Regeneron Exhibit 1052.004
`
`

`

`Installation Qualification
`
`installation Qualification requires an audit of the equip-
`ment as it has been installed in the facility. This audit
`includes checking all utilities and supplies to the equip-
`ment
`to make sure that
`they meet
`the manufacturer's
`recommended specifications. Engineering drawings must
`be evaluated to assure that (i) the equipment is assembled
`according to the manufacturer’s prints, (ii) the equipment
`is installed according to the installation schematics, and
`(iii) all aspects of the equipment are documented with
`appropriate engineering drawings or sketches. These
`drawings are essential for future reference to compare
`the ha rd wa rc validated to any future configurations. This
`segment of the validation program is probably the most
`abused with sterilizers using nonexplosive containers of
`E0. Systems using 1011“» are extremely well documented,
`which is d m en by the safety issue. Once the equipment is
`hooked up and it "runs," little more is ever documented.
`With the pressure to get things working, little attention is
`paid to the documentation for future reference.
`inade-
`quately treated items typically include documentation of
`utilities, spare parts lists and preventive maintenance
`procedures. Many validations have been performed
`with all the necesmiry tests on the hardware relating to
`product loads, but with no record as to the exact configu-
`ration of
`the equipment when the validation was
`executed. Since any mechanical device will routinely
`malfunction, or wear out and require replacement, it is
`absolutely essential
`that a well—prepared Installation
`Qualification document be assembled for each piece of
`equipment to be validated. If this is not done, subsequent
`validation data may prove meaningless.
`
`Calibration
`
`.
`
`The second segment of the Engineering Qualification is
`the calibration of all process sensing, controlling, indicat—
`ing,
`and
`recording
`devices
`on
`the
`sterilizer
`or independent systems associated with it. Recording
`instruments that appear on the control panel are typically
`calibrated, but many of the control instruments are often
`located out of sight and should not be ignored since they
`may have a tremendous impact on the cycle function. For
`example with the DEC phase of an E0 sterilizing, process.
`it is extremely important to calibrate the stall point of the
`vacuum pump before the actual pressure or temperature
`set points are calibrated. This measurement is critical to
`balance the steam input into the chamber in relation to the
`capacity of the vacuum pump to remove the steam from
`the chamber. All critical process control instruments that
`are recorded and displayed by the control system must be
`calibrated. This is even more complicated when micro
`processor Control units are employed, because not only
`are there specific operating set points for those systems,
`there are also high- and low—limit alarm and other default
`systems that must be documented and calibrated. The
`calibration program will also vary depending on the type
`of computerized system,
`The calibration program should be performed with
`instruments referenced as secondary standards. The
`secondary or transfer standard is a standard that can be
`transported to and from the actual sterilization equip-
`ment because most instruments associated with the
`sterilixer must be calibrated at
`the sterilizer's location.
`
`t6
`
`VALIDATION a; ETHVLENE OXIDE sremumuow PROCESSES
`
`243
`
`Secondary standards must be traceable to a recognized
`standard such as those maintained by the hilt-ST.
`A measurement or calibration compares a BET to a
`standard or reference. This standard should outperform
`the DUT by a specific ratio, called the ”TUR“ also known
`as the TAR. As a rule of thumb, the TUR should be greater
`or equal to 4:] (12).
`Primary standards should have an even greater
`sensitivity.
`It
`is recommended that these primary stan—
`dards be submitted to the NIST for calibration and
`
`recertification on a periodic basis. Primary standards
`are usually recertified annually. It is extremely important
`that detailed procedures be established including limits
`and acceptable correction variances allowed and cali—
`bration frequency for all
`the instruments on the
`sterilizer. Adequate records must be maintained. A
`tracking system is essential to assist metrology, assuring
`that required calibrations occur at their designated
`frequencies. A history file should be maintained for
`each instrument. and the records reviewed to assure
`
`established calibration frequencies are appropriate.
`
`Operational Qualification
`The third segment of Engineering Qualification is Oper—
`ational Qualification that deals with the operating
`parameters of the sterilizer: their function, adjustment,
`and control. These tests are performed with an empty
`chamber. The various parameters for the cycle are eval-
`uated to determine if they perform as specified by the
`manufacturer. Temperature controllers are set and eval-
`uated to determine performance. The temperature
`distribution within the sterilizer is documented. The
`
`unit is sequenced through its operating steps to assure
`that the sequencing is appropriate. Every operating
`parameter must be documented to determine its compli-
`ance with the manufacturer's operating specification. The
`Operational Qualification protocol will serve as the basis
`for developing the Standard Operating Procedure for
`routine operation of the sterilizer. The Operational Quali—
`fication testing specifies in detail how the equipment
`operates.
`
`Process (Perturmance) Qualificatinn
`The final phase of validation deals with Process Qualifica-
`tion. Even though the unit functions appropriately with
`an empty chamber, it must now be demonstrated that it
`sterilizes product This phase may require repetition with
`different products and loads.
`
`Load Configuration
`There are several key aspects of Process Qualification.
`First. the specific product and all its packaging must be
`defined. The next step is to define the way master cartons
`are arranged into pallets. Pallet arrangement within the
`sterilizer is also part of the load configuration definition.
`Many manufacturers have numerous products that must
`be mixed together in order to achieve effective sterilizer
`throughput.
`Categoriring product for the sterilizing load is an
`extremely important element.
`It
`is important
`that
`the
`particular product mix is configured with a rationale
`that packaging is similar and products should be of
`consistent mass and materials and actual product
`
`Regeneron Exhibit 1052.005
`
`

`

`244
`
`ill
`
`summation SANITIZATlON aria STERiLirmssutiatict
`
`is possible that a manufacturer may
`it
`configuration,
`hat e in its catalog, hundreds ol different products.
`It
`these all have the same characteristics and packaging,
`it
`is possible that the} could be sterilized within one or
`two dilterent sterili/ation cycles.
`It
`is also possible that
`a manufacturer ma}, produce unit a few products, each
`hemp so different from the other products manutattured
`that each product sterilization prm‘ess hill hate to he
`validated in different cvcies.
`
`Unce product categories have been identified, it is
`also possible to \‘ar\' the load configurations. Loads must
`be e\ti‘eiiielt specific in the i\ at thei are defined. Small
`tolerances are permissible without changing the oiei'all
`impact on the biological effectiveness of the sterilization.
`However. changes in the qualified load must he evalie
`ated and properly documented to determine the
`potential
`impact on the biological ett‘ectii‘eness ot
`the process.
`Conditions which influence the lethality delitered
`by the process are mass. density. packaging. product
`design and materials. External preconditioning of
`product
`loads is common and allows easy process
`measurements. Preconditioning time may vary with
`different
`loads. Determining moisture. EO gas and
`temperature penetration into the palletized load is
`much more difficult in the sterilizer.
`if time to achieve
`
`acceptable levels of the- . parameters is similar to the
`originally defined load.
`then loads can be considered
`equivalent. Measured lethality should be similar with
`similar loads.
`It should he noted that configuration
`changes may influence the location of ”worst case—
`least
`lethal" position. Confirmation will have to he
`performed and appropriate adiustmei‘its may have to
`be made to assure that a proper monitoring location
`is documented.
`
`Once the product and load have been defined, then
`the worst case—least lethal locations in the pmduct, within
`each pallet and within the vessel must be determined,
`These locations will have to be monitored physically and
`biologically to provide data on all critical process
`parameters
`
`Pal/e! Configurations
`Pallet construction mat depend in part on how much
`shipping will take place between the time of construction
`until sterilization, When processing was pertormed
`in-house, it was easy toconstruct pallets with “chimneys”
`configured between columns of master cartons l'hese
`chimneys assured that more surface area of the master
`cartons was directly accessible for thermal transfer and
`gas ewhange. This ti'pe of configuration provides the
`greatest homogeneity of steriliration conditions across
`the product load.
`('ontract EU sterilization is extremely popular
`today and provides users with “state~ot—thc~art‘
`systems at reasonable expense. the problem comes not
`from the sterilizer. but from the logistics involied in
`transporting the product
`tiff~site to the contractor,
`Pallets are constructed at
`the product manufacturing
`site with transportation in mind, not sterili/ation
`Pallets are denseli packed because they survii e the
`rigors of overland shipping much better than pallets
`configured with \‘oid spaces
`(chimneys)
`tor gas
`
`
`
`Flgura 2 An example 01 a banded pallet ol product prowding
`mammurn surface exposure to sterilization vapors. Note: Corner
`protectors on pallet protecting the master cartons.
`
`permeation. Stretch wrap is commonly used to hold
`the palletired hows together. Stretch W rap is eweptional
`tor maintaining pallet
`integriti during shipping,
`but
`it ma). create a tremendous, barrier to sterilizing,
`\apor penetration. Stretch wrap manutactiirers are now
`uttering a "net" type of o rapping material which signi-
`ticantly increases the surface ot
`the master cartons.
`directly exposed to the sterilizing vapors. The best tech-
`nique from a sterilization perspective is to use strapping:
`to band the pallets together. ”its requires the use of
`corner protectors so 1h not to crush the outside Corners
`of the master cartons (Fig. 2], An example of uniformly
`constructed pallets loaded into a sterilizer vessel appears
`in Figure 3. More pallet configurations can be sterilized
`successfully of course, but prix‘i‘ss times may be longer
`
`
`
`Figure 3 An example ot a uniform load configuration Two
`identical pallets side by SlCle All pallets are exactly the same in
`construction and product
`
`Regeneron Exhibit 1052.006
`
`

`

`lethality and £0 residuals
`and variations of microbial
`across the load may be greater.
`
`gas supply lines allowing makeup charges to maintain the
`target gas concentration.
`
`‘16
`
`VALIDATION OF ETHYLEME OXIDE STERIL|ZATl0N PROCESSES
`
`245
`
`CRITICAL STEHILIZATION PROCESS PARAMETERS
`
`There are several major considerations to be aware of in
`order to structure a validation program that will assure
`that the sterilization process does what it
`is intended
`to do.
`
`These considerations include: (1') controlled process
`parameters and their interaction; (ii) an integration of the
`physical process conditions; (iii) the selection of appro-
`priate process conditions; (in) the product design; to) how
`the product is pretreated prior to exposure; (hi) how the
`product is handled following sterilization: (I-iil how the
`process is monitored. including physical, chemical, and
`biological methods; and (viii) the effect of residual E0 and
`its reaction products on the material being sterilized.
`There are [our critical interactive parameters that
`must be controlled for E0 sterilization process: ii) E0 gas
`concentration; (ii) moisture; (iii)
`temperature: and (mi
`time. All these parameters interact to affect the lethality
`delivered by the process.
`
`E0 Gas Concentration
`
`General Use Range of E0 Gas Concentration
`50 gas concentrations below 300 mg/L and above
`1200 mg/ L are not commonly used in the industry F0
`gas concentrations less than 300 mg/ L are not effective in
`practical process times. Concentrations above 901)
`to
`1200 mg/L do not shorten the process times sufficiently
`to warrant the additional cost of gas. Sterilization effec-
`tiveness is dependent on the molecular collision of the F0
`molecule and the biological entity that is being sterilized.
`Therefore. more E0 molecules lead to more rapid micro-
`bial lethality. A sterilizing process using (>01) mg/l. of E0
`delivers approximately twice the lethality as a process
`using Still mg/ L in the same time. However, considering
`the cost of E0. processes are generally designed toward
`the lower concentrations of E0. Concentrations of «100 to
`
`btltlmg/I. appear to be the more popular conditions
`today for operations to balance the cost of ED, equipment
`and throughput time.
`
`ED Gas Concentration Controllers
`
`The I30 gas concentration is controlled in one of two
`ways. The most common method of control is the indirect
`method through the use of a pressure control system. The
`EU gas concentration desired is calculated as to the
`corresponding increase in pressure. The desired pressure
`settings are then maintained by conventional pressure
`controllers. The direct control method uses analytical
`instruments that achaally detect the E0 gas concentration
`in the environment inside the sterilizer.
`The analytical systems are either gas chrom~
`atographic,
`[R
`or microwave
`detectors. These
`instruments are installed directly to the sterilizer. Periodic
`gas samples are withdrawn from the sterilizer or gas
`circulation lines and passed through the detector. Some
`[it or microwave detectors maybe mounted on the exterior
`chamber wall using an access port or in the gas circulation
`system. Electronic signals are sent to control valves in the
`
`Indirect Methods
`
`There are two approaches for the indirect method
`of measuring E0 gas concentration in the sterilizer—
`they are weight and pressure. The indirect methods are
`dependent on using gas cylinders containing certified
`mixtures of EO. When the chamber is pressurized, it is
`assumed that the mixture contains the given percentage of
`E0 relative to the change in pressure. Therefore, this
`change in pressure can be equated to an aSSumed gas
`Concentration. This system is very easy to monitor using
`pressure transducers and recorders.
`The second indirect method measures the weight of
`the gas cylinder contents dispensed into the vessel. This
`method assumes that a uniform mixture of the E0 and
`
`diluent gas was dispersed into the vessel, yielding an
`assumed concentration of gas in the sterilizing chamber.
`This system is easy to monitor using acceptably
`sensitive scales.
`These indirect methods are reasonably good est-is
`mates
`for most gas mixtures. Neither method
`compensates for absorption of E0 by the packaging
`materials or the product. Different materials absorb E0
`at different rates than they do diluent gases (I3). Further-
`more, indirect methods do not consider physical leaks in
`the sterilization system. Thus the indirect method. at best,
`provides an approximation of the E0 gas concentration in
`the vessel.
`
`Direct Gas Measurement
`
`Direct analysis of the E0 in a sterilizing chamber can be
`performed by specific analytical instruments. Two of the
`most common analytical methods are the GC and the
`IR spectrophotometer.
`
`Gas Chromatography. CC has been the most widely
`used method for determining the level of EC.)
`in the
`sterilizing environment Some EO processes operate at
`atmospheric or positive pressure, making withdrawal of
`a gas sample easy. Sterilization processes that use 100%
`EU with a nitrogen blanket may operate at slightly
`subatmospheric pressures and sampling is slightly more
`difficult. CC is not used in It
`":2, E0 processes with no
`nitrogen overlay because they operate under a deep
`vacuum. When dealing with explosive mixtures of E0
`or pure E0, only intrinsically safe instrumentation must
`be used.
`
`Sample removal is extremely important in order to
`assure meaningful data. Sample lines must be heated and
`insulated upon exiting the sterilizer. It cold spots occur in
`the sampling lines, the E0 and water vapor may condense,
`yielding false data. These samples may be collected using
`gas collection bottles or with lines attached directly to the
`CC if an automatic injection system such as a gas sampling
`loop is used.
`Multiple sample sites also present a problem.
`Representative sites are generally selected throughout
`the sterilizing chamber. Small capillary tubes serving as
`sample delivery lines are fitted to the gas sample ports.
`Care must be taken to permit these sample tubes to be
`flushed to assure that
`the sample being extracted is,
`
`Regeneron Exhibit 1052.007
`
`

`

`245
`
`m STEHILIZATION. summation AND STERlLlTY ASSURANCE
`
`from the chamber environment and not a
`indeed.
`in the sample delivery line. For this reason,
`residual
`this method is not acceptable for sampling,r within the
`product or product packages. The flushing of the sample
`lines accelerates the gas penetration into these restricted
`locations and yields data that are not representative of
`actual load conditions,
`
`Gas samples can be extracted from the gas recircula—
`tion system. This provides a good estimate of the gas
`concentration in the chamber.
`
`The GC unit must be calibrated prior to sample
`analysis with a certified standard gas. This certified
`standard may be either a diluted gas mixture or lll"'..
`EO. Most laboratories that are established to perform CC
`analysis are qualified to use 100% E0 as the standard for
`calibration. However,certitied mixtures are available from
`
`gas suppliers. The (SC is calibrated at one point with this
`standard gas and expressed as mole percent. These cali-
`bration results are independent of temperature and
`pressure. The mole percent concentration of the steriliza-
`tion chamber is compared to the standard gas and is then
`converted into mg/ L
`
`100%
`x
`
`mol
`273°C
`( 273°C + ”Cl
`
`g
`=xr
`
`Ill-L7 psia 4: l' psigl
`14.7 psia
`
`where Y psig, pressure of the sterilizing chamber;
`"C, temperature in the sterilizing chamber; XF, scaling
`factor.
`Therefore, the E0 concentration from the GC data
`in mole percent multiplied by the scaling factor (XF)
`yields mg/ L.
`
`[R Analysis. Most gases have a characteristic IR spec-
`trum that can be used to identify them. These spectra are
`usually rather complex; however, each usually contains a
`small number of strong analytical bands that are used in
`this analysis.
`These lR analyzers incorporate a fixed wave-length
`filter that corresponds to one of these strong bands. An
`optical path is also chosen that provides the sensitivity
`range required for the particular analysis.
`The analytical wavelengt

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket