throbber
hon p.1 [100%]
`
`YAKUGAKU ZASSHI 121(6) 459―463 (2001)
`
`459459
`
`―Notes―
`
`Occurrence of Coring in Insulin Vials and Possibility of Rubber Piece
`Contamination by Self-Injection
`
`Toshinari ASAKURA,,a Hiroaki SEINO,b Seishiro NOZAKI,a and Ryuzo ABEc
`Department of Pharmacya and Department of Internal Medicine Diabetes Center,b Ohta Nishinouchi
`Hospital, 2520 Nishinouchi, Koriyama, Fukushima 9638558, Japan, and Department
`of Internal Medicine,c Ohta Memorial Hospital, 524 Nakamachi, Koriyama,
`Fukushima 9638004, Japan
`
`(Received February 5, 2001; Accepted March 23, 2001)
`
`Coring is reported to occur because rubber pieces are shaved oŠ from a rubber stopper when a needle is inserted into
`the rubber stopper of transfusion liquid formulation. We veriˆed whether coring really occurs in insulin vials of self-in-
`jecting patients. We collected insulin cartridges from 30 hospitalized patients and used the primary injection (trial injec-
`tion), the secondary injection and the cartridge remaining preparation as samples. We observed the rubber pieces using a
`microscope and measured the shape, number of pieces. The occurrence rate of coring was 73% for the primary injection,
`47% for the secondary injection and 97% for the cartridge remaining preparation. The rubber pieces in the primary in-
`jection and the secondary injection which went through the needle are mostly in aggregate shape and the rubber pieces in
`the cartridge remaining preparation which did not go through the needle are mostly in needle-like shape. A number of
`small rubber pieces are found in both the primary injection and the secondary injection, indicating a high possibility that
`rubber pieces may be injected under subcutaneous tissue. The coring is considered to occur because needles are repeated-
`ly inserted and rotated at the same spot. It is required to improve the structure to mount a needle to the pen-type injector
`in future. Coring is a very serious problem from the medical and pharmaceutical points of view. Further study should be
`made on the implication to latex allergy and lipodystrophy.
`
`Key words―insulin; coring; needle; vial; diabetes
`
`INTRODUCTION
`
`The pen-type insulin injector (hereunder referred
`to as ``pen-type injector'') has been developed so that
`patients can mount and self-inject insulin from car-
`tridges (hereunder referred to as ``cartridge'') several
`ten times. Therefore, in Japan, many patients are us-
`ing them for 10 years because they are easy to use and
`convenient to carry with, compared with the conven-
`tional disposable syringes.
`Vial preparation is sealed with a rubber stopper
`so that the medical liquid can be used partially for
`several occasions. However, coring is reported to oc-
`cur because rubber pieces are shaved oŠ form rubber
`stoppers if an injector needle of 18 gauge (hereunder
`referred to as ``G'') is used.1,2) Therefore, Japanese
`pharmaceutical ˆrms have provided information to
`prevent coring in vials containing transfusion liquid
`and to give warning to the users. We made the coring
`occurrence test using cartridges because we consi-
`dered it necessary to conˆrm if coring occurs in the
`cartridges, and if the rubber piece generated could
`pass through a needle. As a result, we found that cor-
`ing occurred with extremely high frequency and that
`
`rubber pieces could pass through the needle.3)
`In this test, we checked if the same result could
`be obtained in the cartridges actually used by
`patients.
`
`MATERIALS AND METHODS
`
`We explained the objectives of this test and ob-
`tained consent from 30 patients hospitalized at Ohta-
`Nishinouchi Hospital (hereunder referred to as ``this
`hospital'') and collected the actually used Penˆll300
`(manufactured by Novo Nordisk Pharma, Co., Ltd.)
`as samples. Then, we set the cartridges to NovoPen
`300 (Novo Nordisk Pharma) and, using Micro-Fine
`Plus31 gauge 8 mm needle (Nippon Becton Dickin-
`son, Co., Ltd.), we collected sample A: the primary
`injection, 2 units), sample B: the
`injection (trial
`secondary injection (2 units) and sample C: the car-
`tridge remaining preparation (cartridge remaining
`preparation) (Fig. 1). After the samples are ˆltered
`using 0.45 mm ˆlter (Millipore ˆlter Type: HA, Lot
`No.: H7SH24028), we observed the samples ˆltrated
`using a microscope (OLYMPUS SZ-RT), measured
`the shape, number of pieces and the size and analyzed
`by one-way ANOVA and ScheŠ's test.
`
` PFIZER, INC. v. NOVO NORDISK A/S - IPR2020-01252, Ex. 1073, p. 1 of 5
`
`

`

`hon p.2 [100%]
`
`
`
`460460
`
`Vol. 121 (2001)
`
`Fig. 1. Sampling Methods
`Sample A: the primary injection (trial injection, 2 units), Sample B: the secondary injection (2 units) and Sample C: the cartridge remaining preparation (car-
`tridge remaining preparation).
`
`We selected the rubber pieces of 0.01 mm or larg-
`er in size and measured the longer diameter of ag-
`gregate rubber pieces and the length of needle-shaped
`or spiral-shaped rubber pieces. The test was conduct-
`ed in November 1999. The cartridges were mounted
`on pen-type injectors and stored at room temperature
`(22―25°C), shielded from light, when they were used
`by the patients and when the observation was made.
`Each time the patients made self-injection,
`they
`replaced the needle with new one and wiped the rub-
`ber stopper with alcohol-sterilized cotton before the
`tip of the needle was inserted into the rubber stopper.
`The needle was inserted into the pen-type injector car-
`tridge formulation in accordance with the ``Novo
`Pen300 Utilization Guide'' prepared by Novo Nor-
`disk Pharma, Co., Ltd.
`The average age of the 30 patients from whom
`the samples were collected was 59±18, HbAlc was 8.0
`±1.6%, the number of injections made per day was 3
`±1, the unit of injection per day was 29±16 units,
`the number of insertion made to the collected car-
`tridges was 26±15 and the amount of insulin collect-
`ed was 0.9±0.4 ml.
`
`RESULTS
`
`The
`1. Coring Occurrence Rate (Table 1)
`coring occurrence rate of the 30 samples was 22 sam-
`ples (73%) for Group sample A, 14 (47%) for
`Group sample B and 29 (97%) for Group sample C.
`Most of the rubber pieces in Group sample A and
`sample B were in aggregate shape while the rubber
`pieces in Group sample C are mostly in needle-like
`shape (Fig. 2). Spiral-shaped rubber pieces were
`found only in Group sample C.
`Many of the rub-
`2. Size of Rubber Pieces
`ber pieces of Groups sample A and sample B were
`smaller than 0.01 mm in diameter, while the rubber
`pieces of Group sample C were relatively larger in di-
`ameter, 0.01 mm or more. When we compared the
`rubber pieces of 0.01 mm or more in diameter, we ob-
`tained the following results: the average size of Group
`sample A was 0.07±0.06 mm in diameter. The
`average size of Group sample B was 0.04±0.02 mm
`and the average size of Group sample C was 0.18±
`0.15 mm in diameter (Fig. 3). The size of rubber
`pieces of Group sample C was signiˆcantly larger in
`diameter than those of Groups sample A and sample
`
` PFIZER, INC. v. NOVO NORDISK A/S - IPR2020-01252, Ex. 1073, p. 2 of 5
`
`

`

`hon p.3 [100%]
`
`No. 6
`
`Table 1. Coring Occurrence Rate
`
`Samples
`Coring occurrence
`rate
`Rubber pieces
`
`aggregate shape
`Needle-like shape
`Spiral shape
`
`Sample A
`
`Sample B
`
`Sample C
`
`30
`22(73%)
`
`30
`14(47%)
`
`30
`29(97%)
`
`29(pieces)
`
`22(pieces)
`7
`0
`
`26
`
`22
`4
`0
`
`58
`
`25
`32
`1
`
`Sample A: the primary injection (trial injection, 2 units), sample B: the
`secondary injection (2 units) and sample C: the cartridge remaining prepa-
`ration (cartridge remaining preparation). The ``samples`` and ``the coring
`occurrence rate`` gives insulin vial number. Except for these, gives rubber
`pieces number.
`
`Fig. 2. Needle-Like Shape Rubber
`(magniˆcation×140)
`
`B (p<0.01 by ScheŠ's test).
`3. Observation of the Inside of Rubber Stopper
`By observation of the inside of rubber stoppers, we
`found that the rubber pieces were pushed out by a s-
`eries of needle insertion and depressions were made
`by the dropped rubber pieces (Fig. 4).
`
`
`
`461461
`
`DISCUSSION
`
`There are two problems with injection: One is
`that glass pieces from an ampoule are entrapped into
`injection liquid when it is cut and second is that cor-
`ing of rubber pieces occurs when a needle is inserted
`into a vial, as discussed above. The mechanism of
`coring is that the rubber stopper is shaved when a nee-
`dle is stuck in a rubber stopper of vial. For this rea-
`son, the insertion angle of the needle and the ap-
`proach angle to the heel of the vial are important. At
`present, Japanese transfusion and medicine manufac-
`turers are collaborating to issue brochures to prevent
`coring and to instruct users how to insert an injector
`needle. The brochures tell you to (i) insert a needle
`slowly and vertically, (ii) not to twist the needle while
`it is inserted and (iii) not to insert it at the same spot.
`In Japan, insulin self-injection charge is not co-
`vered by the health insurance since 1981. Since then,
`with the advent of the pen-type injector and the exten-
`sive education to diabetes patients, it has become
`widely introduced for both type 1 and type 2 diabetes
`patients.
`We investigated the occurrence of coring from
`insulin vial formulation for syringes (29G) and insu-
`lin cartridge formulation for pen-type injector (30G,
`31G) at storage temperature (6―8°C/22―25°C), in-
`sertion angle (90°/60°) and the number of insertion
`(10/30 times). As a result, we have found that coring
`may occur when thin 29-31G needles for subcutane-
`ous injection are used. Coring also occurs when vial
`formulation for syringes is stored at a cool place.
`Coring occurred for the cartridge formulation for
`pen-type injector in both cases.3) Therefore, we con-
`ducted this test to check whether coring occurs in the
`cartridges which was used for actual self-injection by
`the patients. It was found that coring occurred at a
`very high ratio of 97% in the cartridges actually used
`by the patients.
`From the pictures showing the inside of the rub-
`ber stopper, it is judged that the rubber pieces come
`out because the needle shaves oŠ the rubber pieces
`when it is mounted on the pen-type injector. The
`material of rubber stoppers used for insulin formula-
`tion has dual layer structure consisting of butyl rub-
`ber and natural rubber.4) The purpose of using the
`material is to prevent coring and to avoid any eŠect
`on the insulin. To minimize the shaving oŠ of rubber
`pieces, the cutting face of the needle (inner needle)
`
` PFIZER, INC. v. NOVO NORDISK A/S - IPR2020-01252, Ex. 1073, p. 3 of 5
`
`

`

`hon p.4 [100%]
`
`
`
`462462
`
`Vol. 121 (2001)
`
`Fig. 3. Size of Rubber Pieces of 0.01 mm or more in Diameter
`Sample A: the primary injection (trial injection, 2 units), Sample B: the secondary injection (2 units) and Sample C: the cartridge remaining preparation (car-
`tridge remaining preparation). Data are presented at means±SD. p<0.01 by ScheŠ's test.
`
`suggested that these small rubber pieces may be in-
`jected into the subcutaneous tissue.
`A ``trial injection'' to release 1 or 2 units of insu-
`lin preparation is performed to take out air after the
`needle is ˆxed before injection and to check if no fault
`with the injector has occurred. Then, the unit of insu-
`lin for injection (2 units in this case) is set and inject-
`ed. At ˆrst, we expected that there would be very little
`chance for the rubber pieces to go through the needle.
`Even if some rubber pieces may have entered in the
`needle when the needle is inserted, they are discharged
`when a trial injection is made and there would be no
`rubber pieces in the actual injection. However, rubber
`pieces were also found in Group sample B in this test.
`Therefore, it is very a serious problem because many
`rubber pieces were found in Group sample B as well
`as in Group sample A, indicating a very high possibil-
`ity that rubber pieces may be injected into the sub-
`cutaneous tissue.
`A number of coring seems to have occurred be-
`cause the needle mounted on Novo Pen300 which is
`used in this test is inserted by rotating it at the same
`spot. However, the pen-type injector currently availa-
`ble in the market employs the same inserting method.
`Therefore, we have to ˆnd out a way to prevent cor-
`ing by enhancing the needle ˆxing method to the car-
`tridge formulation in future.
`In addition, as we found that rubber pieces were
`passed through an injector needle in this test, we have
`to pay attention to the possibility of ``foreign sub-
`stance'' mixture. As the insulin self-injection is sub-
`cutaneous injection into the upper arm, abdominal
`
`Inside of Rubber Stopper
`Fig. 4.
`(magniˆcation×44)
`
`which is inserted into the rubber stopper of the car-
`tridge is so designed that has no blade part. However,
`we observed no improvement in this test and coring
`occurred with high probability. To ``insert needles
`while rotating them on the same spot'' seems to be
`one of the greatest factors to cause coring, rather than
`the rubber material and the cutting face because a
`number of needle-shaped rubber pieces were found in
`the remaining insulin preparation.
`Many small rubber pieces were found in Group
`sample A and Group sample B. The 31G needle has
`an outer diameter of approximately 0.26 mm and the
`inner diameter of approximately 0.15 mm.5) As the
`diameter of a number of rubber pieces found in the
`sample A and sample B of this test is less than 0.15
`mm and could pass through the needles, it is strongly
`
` PFIZER, INC. v. NOVO NORDISK A/S - IPR2020-01252, Ex. 1073, p. 4 of 5
`
`

`

`hon p.5 [100%]
`
`No. 6
`
`wall, femur, etc., and not an intravenous injection,
`there is little possibility of rubber pieces entering into
`blood vessels. However, as reported in many cases of
`latex allergy,6) it is necessary to study the in‰uence of
`coring on insulin allergy7―10) or relationship with
`lipodystrophy.11) Therefore, we should concentrate
`on coring prevention and further tests should be made
`to ˆnd the existence of clinically harmful substance.
`
`ACKNOWLEDGEMENT
`
`We are grateful to Mr. Yoshikatsu Chonan and
`Mr. Masahiko Sato of Koriyama Plant of Novo Nor-
`disk Pharma for cooperation for this test and meas-
`urement.
`
`REFERENCES
`
`1) Sakai T., Kho H., Mastuki A., Jpn. J.
`Anesthesia, 45, 15331535 (1996).
`2) Kubota H., Katahira M., Sumioka T.,
`Wakiichi M., Taniyama E., Miyahara K., ICU
`& CCU, 21, 743745 (1996).
`3) Asakura T., Seino H., Mizuno M., Nozaki S.,
`Abe R., Jpn. J. Hosp. Pharm., 25, 407413
`(1999).
`Internal document of Novo Nordisk Pharma,
`Co., Ltd.
`Internal document of Nippon Becton Dickin-
`son, Co., Ltd.
`6) Akira A., IRYO, 51, 201204 (1997).
`7) Leslie C. G. et al., J. Allergy Clin. Immunol.,
`71, 250254 (1983).
`8) Guntram S., Diabetes Care, 16, 155165
`
`4)
`
`5)
`
`463463
`
`(1993).
`9) Suzuki T., Hikita S., Kizima Y., Okuzumi K.,
`Hurukawa H., Miyaji N., Miyagawa K., Jpn.
`J. Hosp. Pharm., 24, 768773 (1998).
`10) Kastuda E., Shimizu H., Ubukata T., Mori
`M., J. Japan Diab. Soc., 42, 10131015
`(1999).
`11) Sakamoto Y., The Journal of Therapy, 64,
`441447 (1982).
`
`要旨
`注射針を輸液剤等のゴム栓に穿刺するときに,ゴ
`ム栓からゴム片が削り取られるというコアリングが
`報告されている.実際に自己注射を行っている患者
`のインスリンバイアルでもコアリングが発生してい
`るかを確認した.入院患者30名よりインスリンカー
`トリッジを回収し,空打ち液,注入液,カートリッ
`ジ残液を試料とした.発生したゴム片を顕微鏡下で
`観察し形状と個数,大きさを測定した.コアリング
`の発生率は,空打ち液の発生率は73%,注入液は47
`%,カートリッジ残液は97% であった.形状は,
`針を通過している空打ち液と注入液では塊状が多
`く,針を通過していないカートリッジ残液では針状
`が多かった.空打ち液と注入液では小さなゴム片が
`多数確認されたことにより,ゴム片が皮下内に注入
`される可能性が強く示唆された.コアリングの原因
`としては,針を同一個所に回転させて刺すためと考
`えられる.そのため,今後はペン型注射器への針の
`装着について構造上の改良が必要である.コアリン
`グは注射液の異物混入という点で,医学的にも薬学
`的にも非常に重大な問題である.今後はラテックス
`アレルギーやリポディストロフィーなどとの関連も
`検討する必要がある.
`
` PFIZER, INC. v. NOVO NORDISK A/S - IPR2020-01252, Ex. 1073, p. 5 of 5
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket