throbber
O R I G I N A L A R T I C L E
`
`Ability to handle, and patient preference for,
`insulin delivery devices in visually impaired
`patients with type 2 diabetes
`
`Charles Fox*, Carol McKinnon, Anthony Wall, Simon A Lawton
`
`ABSTRACT
`In this comparative study, 86 patients with type 2 diabetes and visual impairment were evaluated on their preference for, and ability
`to operate, three different insulin delivery systems – InnoLet® (NovoNordisk), Humulin® NPH Pen (Eli Lilly) and vial and syringe
`(Becton Dickinson). Patients found the clocklike dose scale on InnoLet® significantly easier to read than that of the other systems
`(92% were able to read four doses, versus 45% and 61% with Humulin® Pen and syringe respectively, both p < 0.001), and showed
`greater ability to set and dispense a 20 unit dose without instruction. After reading the packaging information leaflet and brief ver-
`bal instruction, 99% of patients were able to correctly set and dispense three consecutive insulin doses with InnoLet® compared with
`85% with Humulin® Pen and 64% with syringe (p = 0.001 and p < 0.001, respectively). Ability to set and deliver insulin was not
`however clearly related to visual acuity. On questionnaire, 87% of patients expressed an overall preference for InnoLet®, and 13%
`for Humulin® Pen (p < 0.001); no patients preferred the syringe. In conclusion, insulin delivery systems designed to simplify accu-
`rate, reliable insulin delivery for people with visual impairment can improve the ability of such patients to repeatedly set and deliver
`the correct insulin dose. Copyright © 2002, John Wiley & Sons, Ltd
`
`Practical Diabetes Int, 2002; 19(4): 104–107
`
`KEY WORDS
`type 2 diabetes; insulin delivery system; visual impairment
`
`Introduction
`Type 2 diabetes is a progressive disorder,
`associated with a steady decline in beta-cell
`function, from a mean 51% at diagnosis,
`to 28% after 6 years1. This reduction in
`glucose-lowering capacity necessitates
`introduction of insulin in the majority of
`patients, within 15 years of diagnosis2. By
`lowering HbA1c through the use of inten-
`sive therapy, morbidity associated with
`
`Charles Fox, BM FRCP Consultant
`Physician, Northampton General Hospital,
`Northampton, UK
`
`Carol McKinnon, MBChB, MRCGP,
`DRCOG, General Practitioner, Castlemilk
`Group Practice, Glasgow, UK
`
`Anthony Wall, LRCP MRCS MB.BS DCH,
`General Practitioner and Clinical Trials
`Coordinator, St John’s Health Centre,
`Woking, UK
`
`Simon A Lawton, BHB, MBChB,
`International Medical Advisor, Novo
`Nordisk A/S, 2880 Bagsvaerd, Denmark
`
`*Correspondence to Dr Charles Fox,
`Northampton General Hospital, Cliftonville,
`Northampton NN1 5BD, UK
`
`Received: 28 June 2001
`Accepted: 7 January 2002
`
`microvascular, and probably macrovascu-
`lar, complications can be significantly
`reduced3,4. Any reduction in HbA1c, how-
`ever small, contributes to improving prog-
`nosis3. Nonetheless, many eligible patients
`are denied the benefits of timely insulin
`therapy due to misconceptions about
`appropriate use of insulin in type 2 dia-
`betes, and anxiety concerning the practical
`and organisational skills required to follow
`complex dosing regimens. From the
`patients’ perspective the anticipated pain
`and social stigma of injections compound
`the delay in receiving optimal glucose-low-
`ering therapy5.
`Current insulin delivery systems vary in
`format and ease of use. Insulin is available
`in vials, to be drawn up using a conven-
`tional syringe, but this method provides
`poor dose accuracy6. Patients who experi-
`ence particular difficulty in accurate self-
`dosing include the elderly, as well as
`younger people with poor vision or
`impaired manual dexterity. Visual impair-
`ment is common in diabetes, with acuity
`6/12 or worse in 16% of people over the
`age of 65, rising to nearly 27% by age 757.
`Thus, insulin delivery systems that are easy
`to use by people with poor vision are essen-
`tial for effective diabetes management, and
`to allow patients to maintain their inde-
`pendence. Prefilled insulin cartridges,
`
`inserted into disposable or durable insulin
`delivery devices with short needles, can
`improve dose accuracy and increase com-
`pliance rates compared with more tradi-
`tional methods, and quality of life is often
`enhanced through their use8–11. Since
`patients with type 2 diabetes perform sig-
`nificantly worse than age-matched controls
`on tests of learning and reasoning12, this
`patient group – of whom over 50% will
`require daily insulin therapy at some
`stage13 – may derive particular benefit from
`delivery systems that are easy to handle and
`simple to operate. It is anticipated that
`rational insulin delivery systems that make
`maximum use of existing sensory capacity
`will be preferred by patients, and will help
`those with sensory impairment capitalise
`on their remaining abilities.
`
`Materials and methods
`In this multicentre, open, randomised,
`comparative
`study, 86
`insulin-naive
`patients with type 2 diabetes were drawn
`from three centres in the UK. During a
`handling test, patients tested three insulin
`delivery systems – InnoLet® (Novo
`Nordisk, Denmark), Humulin® NPH Pen
`(Eli Lilly, USA), and vial and 0.5 ml
`(Becton Dickinson, USA).
`syringe
`NovoFine® 30G 8 mm needles (Novo
`Nordisk) were used with the prefilled pen
`
`104 Pract Diab Int May 2002 Vol. 19 No. 4
`
`Copyright © 2002 John Wiley & Sons, Ltd.
`
` PFIZER, INC. v. NOVO NORDISK A/S - IPR2020-01250, Ex. 1041, p. 1 of 4
`
`

`

`O R I G I N A L A R T I C L E
`
`Insulin delivery devices in the visually impared
`
`devices; the syringe had a pre-attached
`Micro-Fine + 8 mm needle. The study was
`conducted in accordance with good clini-
`cal practice.
`
`Patients
`Patients were included if they were ≥55
`years with diet- and/or oral-hypogly-
`caemic-agent-treated type 2 diabetes. Since
`one of the study objectives was to assess
`the ease with which patients could learn
`how to operate the respective delivery sys-
`tems, people with a history of insulin self-
`administration were excluded to avoid
`potential bias from previous exposure to
`the systems tested. Visual acuity (corrected
`near vision) in the best eye was between
`0.5 (20/40) and 0.1 (20/200) as assessed
`with a Rosenbaum card.
`
`Assessment methods
`Accuracy when viewing dose
`scale
`Ability to administer insulin safely and
`independently relies on repeated accurate
`dose reading. Patients were therefore asked
`to read four randomly selected whole-
`numeral dose settings on each insulin
`delivery system under standardised direct
`illumination.
`
`Handling test
`The handling test comprised three parts.
`
`Part I: intuitive test
`Patients were asked to set and discharge a
`20 unit insulin dose with a minimum of
`standardised instruction and no specific
`training, using each of the three delivery
`systems. The time allowed to complete the
`task was 2 min. Patients were assessed as
`successful in completing this section if the
`dose was set within 5% of that requested
`and the full dose dispensed. In this, and all
`other parts of the study, doses were deliv-
`ered into a disposable container.
`
`Part II: written instruction
`Patients were provided with those sections
`of the manufacturers’ package information
`leaflets giving details on how to operate the
`device. After a maximum of 3 min reading
`time, participants set and dispensed three
`doses of insulin. Doses were randomly
`selected (dose range 4–50 units, with one
`dose > 30 units) but consistent between
`devices and patients (dose 1 = 23 units,
`dose 2 = 42 units, dose 3 = 17 units), and
`dose setting was deemed accurate if it fell
`within 1 unit for doses < 20 units and
`within 5% for doses > 20 units. Successful
`
`completion of this section required all three
`doses to be correctly set and delivered.
`
`Part III: verbal instruction and
`demonstration
`Patients only entered this section of the
`trial if they were unable to complete the
`previous
`section
`successfully. Verbal
`instruction and a demonstration on how
`to operate the system correctly were pro-
`vided, and a 5 min limit was set. The
`patients were then requested to set and dis-
`pense three further randomly selected
`doses (dose 1 = 40 units, dose 2 = 29 units,
`dose 3 = 19 units). Successful completion
`of this section was assessed on the same cri-
`teria as in part II.
`
`Questionnaire
`On finishing the handling test, partici-
`pants completed a questionnaire on their
`preference for certain features of each
`delivery system. Only the question on
`overall preference required both qualitative
`and quantitative answers.
`
`Data analysis
`Setting sample size to a minimum of 80
`patients ensured that group differences
`would be detected with a power of 85% at
`a 0.05 level of significance (a = 5%). The
`global difference between delivery systems
`was calculated with Fisher’s exact test using
`SAS© proc freq. If global difference was
`significant (p ≤ 0.05), all three pairwise
`comparisons between devices were made
`using Fisher’s exact test. In the case of pair-
`wise comparisons a Bonferroni correction
`to
`the value was applied. Because
`InnoLet® was tested twice, the threshold
`value was lowered from 5% to 2.5%.
`
`Results
`Study population
`characteristics
`The study population comprised 86
`patients (51 male, 35 female), mean age 69
`years and mean disease duration 6 years.
`Visual acuity ranged between 0.29 and 0.5
`in 46 patients (54.1%), 0.2 and 0.29 in 17
`patients (20%) and 0.1 and 0.2 in 22
`(25.9%) patients. Only five (5.8%)
`reported a subjective hand disability.
`
`Withdrawal and non-
`completion of handling test
`No patients withdrew prematurely from
`the trial, and only two subjects (2.3%)
`were unable to complete the handling test.
`In both cases this was due to inability to
`handle the vial and syringe.
`
`Figure 1. More patients were able to read
`four consecutive doses settings with
`InnoLet® than with Humulin® Pen or
`syringe
`
`Visual accuracy when reading
`dose scale
`The majority of patients could read all
`four doses correctly with InnoLet®
`(79 patients, 92%), compared with 39
`(45%) with Humulin® Pen, and 52 (61%)
`with the syringe (Figure 1). The differences
`between InnoLet® and both the other
`devices were highly statistically significant
`(p < 0.001 for both). The number of
`patients in each visual acuity group who
`were able to read all four doses correctly
`decreased with decreasing acuity for all
`devices, but the decline was less apparent
`with InnoLet® than with Humulin® Pen
`or syringe.
`
`Handling test
`Intuitive test
`Eighty-three patients (97%) set the 20 unit
`dose correctly (i.e. not evaluating ability to
`dispense the dose) with InnoLet®, com-
`pared with 41 (48%) and 33 (39%) with
`Humulin® Pen and syringe, respectively.
`Following this, 72 patients (84%) also suc-
`ceeded in dispensing the complete 20 unit
`dose with InnoLet®, compared with 35
`(41%) with Humulin® Pen and 27 (31%)
`with the syringe (p < 0.001 between
`InnoLet® and both Humulin® Pen and
`syringe for setting and dispensing; Table 1).
`Mean time taken to set and dispense the
`correct dose was significantly shorter with
`InnoLet® (26 s) than with Humulin® Pen
`(65 s, p < 0.001 versus InnoLet®) or
`syringe (53 s, p < 0.001 versus InnoLet®;
`Figure 2). There was no clear relationship
`between visual acuity and ability to set and
`dispense the correct dose.
`
`Written instruction
`After receiving written guidance, 83 patients
`(97%) correctly set all three insulin doses
`using InnoLet®, while 69 (80%) set and dis-
`pensed all doses accurately (Table 1). This
`compared with 56 (65%) and 24 (28%)
`
`Pract Diab Int May 2002 Vol. 19 No. 4
`
`Copyright © 2002 John Wiley & Sons, Ltd.
`
`105
`
` PFIZER, INC. v. NOVO NORDISK A/S - IPR2020-01250, Ex. 1041, p. 2 of 4
`
`

`

`O R I G I N A L A R T I C L E
`
`Insulin delivery devices in the visually impared
`
`Table 1. InnoLet® is the most reliable device for accurate dose-setting and insulin delivery
`
`InnoLet® Humulin® Pen
`
`Syringe
`
`84a,b
`
`80a,c
`
`99a,b
`
`41
`
`61
`
`85
`
`31
`
`27
`
`64
`
`% of patients able to
`intuitively set and
`dispense 20 U insulin dose
`
`% of patients able to set
`and dispense three insulin
`doses after written
`instruction
`
`% of patients able to set
`and dispense three insulin
`doses after both written
`instruction and brief verbal
`instruction/demonstration
`a p < 0.001 versus syringe.
`b p ≤ 0.001 versus Humulin® Pen.
`c p < 0.01 versus Humulin® Pen.
`
`between InnoLet® and syringe). Again,
`there was no clear relationship between
`visual acuity and ability to use the devices
`correctly.
`
`Questionnaire
`InnoLet® was the preferred delivery system
`of 75 (87%) patients, compared with 11
`patients (13%) who chose Humulin® Pen
`(p < 0.001); no patients preferred the
`syringe. All the specific device features
`assessed were rated as significantly superior
`for InnoLet® (p < 0.001 where applicable;
`Figure 3). All five patients with subjective
`hand impairment found that the button
`on InnoLet® was the easiest to depress
`during injection. Four patients found
`InnoLet® the easiest to hold while setting
`a dose, while one preferred Humulin® Pen
`in this respect. Two found InnoLet® the
`easiest to hold while injecting, whereas two
`preferred Humulin® Pen. Formal statisti-
`cal testing of these differences was omitted
`because of the small sample size.
`
`Discussion
`The need to self-inject can be a barrier to
`starting insulin therapy at any age. Many
`people with type 2 diabetes are introduced
`to insulin self-injection during their more
`advanced years, when confidence in learning
`new skills is declining and both physical and
`psychological limitations add to the burden
`of performing complex tasks. If patients
`believe that assistance from caregivers will be
`necessary, reluctance to be dependent may
`be an additional barrier to starting therapy5.
`Visual impairment – most commonly
`due to cataract – affects 16% of people
`with type 2 diabetes over the age of 65, and
`
`Figure 2. Patients were intuitively able to
`set and dispense insulin significantly more
`quickly using InnoLet® than Humulin®
`Pen or syringe
`
`who set the dose correctly using Humulin®
`Pen and syringe respectively, and 52 (61%)
`and 23 (27%) who dispensed insulin accu-
`rately. Significantly more patients were suc-
`cessful operating InnoLet® compared with
`the other two delivery systems (dose-setting,
`p < 0.001 versus both other systems; dose-
`setting plus delivery, p = 0.007 versus
`Humulin® Pen, p < 0.001 versus syringe).
`Only when using the syringe did decreasing
`visual acuity impair patients’ ability to set
`and fully dispense the correct dose.
`
`Verbal instruction and
`demonstration
`Patients who entered part III of the trial
`were those unable to successfully complete
`part II. InnoLet® accounted for 17
`patients, Humulin® Pen for 34 and syringe
`for 62. After verbal instruction and
`demonstration, 17 (100%) correctly set all
`doses with InnoLet®, compared with 23
`(68%) and 31 (50%) with Humulin® Pen
`and syringe, respectively (p = 0.009 for the
`
`106 Pract Diab Int May 2002 Vol. 19 No. 4
`
`difference with Humulin® Pen; p < 0.001
`for syringe). Following this, 94% (n = 16)
`of patients correctly dispensed all three
`insulin doses with InnoLet®, compared
`with 62% (n = 21) and 50% (n = 31) with
`Humulin® Pen and syringe, respectively (p
`= 0.019 between InnoLet® and Humulin®
`Pen, and p < 0.001 between InnoLet® and
`syringe; Table 1). The proportion of sub-
`jects who set and fully dispensed all three
`doses correctly showed no clear relation-
`ship with visual acuity.
`Following written or verbal instruction
`and demonstration, insulin was accurately
`set and dispensed by 85 (99%) patients
`using InnoLet®, compared with 73 (85%)
`and 54 (64%) with Humulin® Pen and
`syringe, respectively (p = 0.001 between
`InnoLet® and Humulin® Pen, p < 0.001
`
`Figure 3. Patients preferred InnoLet® to Humulin® Pen or syringe
`% of patients
`10 20 30 40 50 60 70 80 90 100
`
`0
`
`Easiest number scale to read
`
`Easiest to set insulin
`
`Simplest to operate
`
`Easiest to learn how to use
`
`Easiest button for injection
`
`Easiest to hold when setting dose
`
`Easiest to hold while injecting
`
`Feel most confident about
`
`Overall preference
`
`InnoLet®
`Humulin®
`Pen
`Syringe
`
`Copyright © 2002 John Wiley & Sons, Ltd.
`
` PFIZER, INC. v. NOVO NORDISK A/S - IPR2020-01250, Ex. 1041, p. 3 of 4
`
`

`

`O R I G I N A L A R T I C L E
`
`Insulin delivery devices in the visually impared
`
`Key points
`● Visual impairment is common in people with type 2 diabetes.
`● Accurate, reliable insulin delivery is facilitated by devices that are tailored to
`physical and psychological ability.
`● With only minimal instruction – of a duration reasonable within a busy clinic
`– the majority of patients, including those with severe visual impairment,
`were able to demonstrate the ability to repeatedly set and dispense insulin
`accurately using InnoLet®.
`● In comparison with Humulin® Pen and syringe, patients could accurately set
`and dispense insulin more reliably with InnoLet®.
`● Patients became competent in using InnoLet® more rapidly, preferred the
`design features of InnoLet® and felt more confident about using it.
`
`this figure rises to 27% at age 757.
`Repeated setting and delivery of the correct
`insulin
`dose
`poses
`a
`challenge.
`Furthermore, over 50% of people with type
`2 diabetes have limited joint mobility in
`the hands14, and 25% have symptomatic
`peripheral neuropathy15. Arthritis and
`tremor increase with advancing age16,
`adding to the difficulty in manually operat-
`ing handheld devices. In order for patients
`to be willing to self-inject insulin regularly,
`both regimen and device should be tailored
`to physical and psychological ability.
`In this study of 86 patients with type 2
`diabetes and visual impairment, patients
`were able to read the InnoLet® dosing
`scale more reliably than that of the other
`systems. Although accuracy fell with
`reduced visual acuity, this was least pro-
`nounced with InnoLet®, and over 80% of
`patients with severe visual impairment
`(<0.2 to ≥0.1) were still able to read the
`dose scale, compared with less than 20%
`with Humulin® Pen and 41% with
`syringe. At every level of instruction, a
`greater number of patients were able to
`accurately set and dispense insulin using
`InnoLet®. In the absence of specific train-
`ing, 84% set and administered the dose
`reliably, compared with 41% and 32%
`with Humulin® Pen and syringe, respec-
`tively. With only minimal training, in the
`form of written instruction or a 5 min
`explanation, 99% became proficient.
`Overall, 87% of patients preferred
`InnoLet® to Humulin® Pen or the syringe.
`
`Ratings for specific design features were
`significantly higher for InnoLet® across all
`eight categories assessed.
`Since people with type 2 diabetes are sig-
`nificantly more likely to have difficulty per-
`forming tasks that require recall of learned
`skills12,17,18, one could speculate that the
`patient preference for InnoLet® demon-
`strated in this study rests on design features
`that address the specific patient needs asso-
`ciated with type 2 diabetes. The large clock-
`like dial, which is set like a familiar kitchen-
`timer, and large-scale numbers assist accu-
`rate dose selection; audible clicks accom-
`pany delivery of each insulin unit dialled,
`reassuring patients about dose selection.
`This may account for the greater number of
`patients in the handling study being able to
`repeatedly set and dispense insulin with
`InnoLet® (99%) than were able to read the
`dose scale without error (92%). Although
`the preference of the five patients with sub-
`jective hand impairment in this study
`tended towards InnoLet®, there were too
`few patients to provide conclusive results.
`However, it seems logical that the comfort-
`able handling of the device and large dose-
`delivery buttons would simplify insulin
`delivery for patients with weak hands.
`In conclusion, the rational design of
`InnoLet® renders it a device suitable for
`accurate, reliable delivery of insulin, even
`in patients with visual impairment. Given
`a choice, the majority of patients favour
`InnoLet® over currently available alterna-
`tives.
`
`Conflict of interest: This research project was funded by NovoNordisk and Charles Fox
`received an honorarium for taking part in the launch of the InnoLet® injection device.
`However, he has no ongoing commercial interest in this product.
`
`References
`1. UK Prospective Diabetes Study Group. UK
`Prospective Diabetes Study 16. Diabetes 1995; 44:
`1249–1258.
`2. Clauson P, Linnarsson R, Gottsater A et al.
`Relationships between diabetes duration, metabolic
`control and beta-cell function in a representative pop-
`ulation of type 2 diabetic patients in Sweden. Diabet
`Med 1994; 11: 794–801.
`3. UK Prospective Diabetes Study (UKPDS) Group.
`Intensive blood glucose control with sulphonylureas
`or insulin compared with conventional treatment and
`the risk of complications in patients with type 2 dia-
`betes (UKPDS33). Lancet 1998; 352: 837–853.
`4. Ohkubo Y et al. Intensive insulin therapy prevents the
`progression of diabetic microvascular complications
`in Japanese patients with NIDDM. Diabetes Res Clin
`Pract 1995; 28: 103–117.
`5. Bashoff E, Beaser R. Insulin therapy and the reluctant
`patient. Postgrad Med 1995; 97: 86–90.
`6. Puxty J, Hunter D, Burr W. Accuracy of insulin injec-
`tion in elderly patients. BMJ 1983; 287: 1726.
`7. Klein R, Klein B, Moss S. Visual impairment in dia-
`betes. Ophthalmology 1984; 91; 1–9.
`8. Hörnquist J, Wikby A, Andersson P, Dufva A.
`Insulin-pen treatment, quality of life and metabolic
`control:
`retrospective
`intra-group evaluations.
`Diabetes Res Clin Pract 1990; 10: 221–230.
`9. Paterson K, Sandler D. Pen injectors – the way for-
`ward? Diabet Med 1992; 9: 684.
`10. Coscelli C, Lostia S, Lunetta M et al. Safety, efficacy,
`acceptability of a pre-filled insulin pen in diabetic
`patients over 60 years old. Diabetes Res Clin Pract
`1995; 28: 173–177.
`11. Corsi A et al. Pre-filled insulin pen in newly insulin-
`treated diabetic patients over 60 years old. Diab Nutr
`Metab 1997; 10: 78–81.
`12. Reaven G, Thompson L, Nahum D, Haskins E.
`Relationship between hyperglycaemia and cognitive
`function in older NIDDM patients. Diabetes Care
`1990; 13: 16–21.
`13. Barnett A. Tablet and insulin therapy in type 2 dia-
`betes in the elderly. J Royal Soc Med 1994; 87:
`612–614.
`14. Starkman H, Gleason R, Rand L et al. Limited joint
`mobility (LJM) of the hand in patients with diabetes
`mellitus: relation to chronic complications. Ann
`Rheum Dis 1986; 45: 130–135.
`15. Ziegler D, Gries F, Spuler M et al. The epidemiology
`of diabetic neuropathy. Diabet Med 1993; 10:
`82S–86S.
`16. Frey M, Barrett-Connor E, Sledge P et al. The effect
`of non insulin dependent diabetes mellitus on the
`prevalence of clinical osteoarthritis. A population-
`based study. J Rheumatol 1996; 23: 716–722.
`17. Strachan M, Diary I, Ewing F, Frier B. Is type II dia-
`betes associated with an increased risk of cognitive
`dysfunction? A critical review of published studies.
`Diabetes Care 1997; 20: 438–445.
`18. Cosway R, Strachan MWJ, Dougal A et al. Cognitive
`function and information processing in type 2 dia-
`betes. Diabet Med 2001; 18: 803–810.
`
`Pract Diab Int May 2002 Vol. 19 No. 4
`
`Copyright © 2002 John Wiley & Sons, Ltd.
`
`107
`
` PFIZER, INC. v. NOVO NORDISK A/S - IPR2020-01250, Ex. 1041, p. 4 of 4
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket