throbber
u-pun
`
`
`1. leont
`Conn-«Ii
`09
`11.2
`1
`
`11.21
`Q11
`11.51
`1A
`13.11
`11.6
`20
`11.71
`11.72
`
`2. Validation
`1
`12.10
`Q10
`Q1 1
`
`2.16
`12.11
`13.13
`
`12.1
`13.13
`3
`7.14
`
`4
`11.44
`3. Chan Control
`1
`4. Re ection and Re-nse 111'Materials
`1
`2
`
`
`
`
`
`4.14
`
`7.44
`
`10.11
`
`11.4
`
`Dated: 10 June 2015
`
`Q7 Q&As
`
`
`
`20
`
`1 8.5l
`
`18.14
`18.2
`
`QSA
`QSD
`Q63
`11
`Q53
`
`19.3
`
`20
`
`
`
`13.17
`
`17.60
`
`
`
`
`
`14.30
`
`P- 21
`
`18
`UT Ex. 2034
`SteadyMed v. United Therapeutics
`|PR2016-00006
`
`|PR2020-00770
`
`United Therapeutics EX2007
`Page 4751 of 7335
`
`
`
`
`
`3
`
`4
`
`1 .3
`2
`9. APII for Use in Clinical Trials
`1
`0. Glenn
`1
`
`2.50
`
`
`
`8. S ecifil: Guidance for APIs Manufactured 11 Cell Culture/Fermentation
`1
`
`14.1
`14.2
`14.3
`14.40
`11.11
`14.41
`11.11
`
`14.43
`5. Com aim; Mid Recalls
`15.10
`1
`In
`
`15.12
`15.13
`2
`1.50
`
`15.14
`
`6. Calm-act Mum Lament (including Llhmtnries)
`1
`2.2
`16.12
`
`2.22
`
`2
`16
`Q10
`3
`16.12
`16.14
`
`7. Agents, Bunkers, Traders, Distributors, Rep“
`, and Relnbeller:
`
`1
`17.1
`
`2
`17.1
`16.12
`3
`17.2
`9.42
`20
`9.43
`17.40
`17.61
`17.20
`17.6
`
` 2
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4751 of 7335
`
`

`

` 7?
`
`W ' Center for Drug Evaluation
`
`and Research (CDER)
`
`Reviewer Guidance'
`
`Validation of
`
`Chromatographic Methods
`
`November 1994
`CMC 3
`
`P- 1
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`IPR2016-00006
`
`|PR2020-00770
`
`United Therapeutics EX2007
`Page47520f7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4752 of 7335
`
`

`

`I.
`
`ll.
`
`III.
`
`IV.
`
`TABLE OF CONTENTS
`
`INTRODUCTION ...............................................
`
`TYPES OF CHROMATOGRAPHY .................................
`
`A.
`
`High Performance Liquid Chromatography (HPLC) ...............
`
`1.
`
`2.
`3.
`
`4.
`
`5.
`
`6.
`
`Chiral Chromatography ...............................
`
`Ion-exchange Chromatography .........................
`lon-pairlAffinity Chromatography ........................
`
`Normal Phase Chromatography .........................
`
`3 A
`
`Reversed Phase Chromatography .......................
`
`3
`
`Size Exclusion Chromatography ........................ .4
`
`1
`
`2
`
`2
`
`2
`
`3
`3
`
`B.
`
`C.
`
`Gas Chromatography (GC) ..................................
`
`Thin-Layer Chromatography (TLC) ............................
`
`REFERENCE STANDARDS ......_ ................................
`
`PARAMETERS FOR VALIDATION OF HPL CHROMATOGRAPHIC
`METHODS FOR DRUG SUBSTANCE AND DRUG PRODUCT ..........
`
`Accuracy .............................................. _.
`
`.
`
`Detection Limit and Quantitation Limit .........................
`
`4
`
`5
`
`5
`
`7
`
`8
`
`8
`
`A.
`
`B.
`
`C.
`
`D.
`
`Linearity ......................................_ .......... 11
`
`Precision ................................... V............. 13
`
`1.
`
`Repeatability ........................................ 13 '
`
`a.
`
`b.
`
`Injection Repeatability .
`
`.
`
`.
`
`.' ..................... 13
`
`Analysis Repeatability ........................... 15
`
`i
`
`P- 2
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`lPR2016-00006
`
`|PR2020-00770
`
`United Therapeutics EX2007
`Page 4753 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4753 of 7335
`
`

`

`E.
`
`F.
`
`G.
`
`H.
`
`i.
`
`J.
`
`2,
`
`3.’
`
`Intermediate Precision ................................ 15
`
`Reproducibility ...................................... 16
`
`Range .................................................. 16
`
`Recovery ................................................ 16
`
`Robustness .............................................. 16
`
`Sample Solution Stability ................................... 17
`
`Specificity/selectivin....................................... 17
`
`System Suitability Specifications and Tests ..................... 21
`
`1.
`
`2.
`
`3.
`
`4.
`
`5.
`
`6.
`
`Capacity factor ..... ' ................................. 22
`
`Precision/Injection repeatability ......................... 22
`
`Relative retention .................................... 22
`
`Resolution .......................... ................ 22
`
`Talling factor ........................................ 23
`
`Theoretical plate number .............................. 26
`
`K.
`
`General Points to Consider ....... ‘ ........................... 28
`
`V.
`
`VI.
`
`COMMENTS AND CONCLUSIONS y................................ 29
`
`ACKNOWLEDGEMENTS ........................................ 29
`
`VII.
`
`REFERENCES ................................................ 29 ,
`
`Ii
`
`P- 3
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`|PR2016-00006
`
`|PR2020-00770
`
`United Therapeutics EX2007
`Page 4754 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4754 of 7335
`
`

`

`REVIEWER GUIDANCE1
`
`VALIDATION OF CHROMATOGRAPHIC METHODS
`
`I.
`
`INTRODUCTION
`
`The purpose of this technical review guide is to present the issues to consider when
`evaluating chromatographic test methods from a regulatory perspective. The
`document discusses the points to note and weaknesses of chromatography so that
`CDER reviewers can ensure that the methods performance claims are properly
`evaluated, and that sufficient information is available for the field chemist to assess the
`method. Analytical terms, as defined by the International Conference of
`Harmonization (ICH), 1993, have been incorporated in this guide.
`
`Chromatographic methods are commonly used for the quantitative and qualitative
`analysis of raw materials, drug substances, drug products and compounds in biological
`fluids. The components monitored include chiral or achiral drug, process impurities,
`residual solvents, excipients such as preservatives, degradation products, extractables
`and leachables from container and closure or manufacturing process, pesticide in drug
`product from plant origin, and metabolites.
`
`The objective of a test method is to generate reliable and accurate data regardless of
`whether it is for acceptance, release, stability or pharmacokinetics study. Data are
`generated for the qualitative and quantitative testing during development and post-
`approval of the drug products. The testing includes the acceptance of raw materials,
`release of the drug substances and products, in-process testing for quality assurance,
`and establishment of the expiration dating period.
`
`Validation of a method is the process by which a method is tested by the developer or
`user for reliability, accuracy and preciseness of its intended purpose. Data thus
`
`1This guidance has been prepared by the Analytical Methods Technical
`Committee of the Chemistry Manufacturing Controls Coordinating Committee (CMC
`CC) of the Center for Drug Evaluation and Research at the Food and Drug
`Administration. Although this guidance does not create or confer any rights for or on
`any person and does not operate to bind FDA or the industry, it does represent the
`agency’s current thinking on the validation of chromatographic methods. For additional
`copies of this guidance, contact the Division of Communications Management, HFD—
`210, CDER, FDA, 5600 Fishers Lane, Rockville, MD 20857 (Phone: 301-594-1012).
`Send one self-addressed adhesive label to assist the offices in processing your
`request. An electronic version of this guidance is also available via Internet the World
`Wide Web (WWW) ( connect to the FDA Home Page at WWW.FDA.GOV/CDER and
`go to the “Regulatory Guidance" section).
`
`P- 4
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`lPR2016—00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4755 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4755 of 7335
`
`

`

`generated become part of the methods validation package submitted to CDER.
`
`Methods validation should not be a one-time situation to fulfil Agency filing
`requirements, but the methods should be validated and also designed by the developer
`or user to ensure ruggedness or robustness. Methods should be reproducible when
`used by other analysts, on other equivalent equipment, on other days or locations, and
`throughout the life of the drug product. Data that are generated for acceptance,
`release, stability, or pharmacokinetics will only be trustworthy if the methods used to
`generate the data are reliable. The process of validation and method design also
`should be early in the development cycle before important data are generated.
`Validation should be on-going in the form of re-validation with method changes.
`
`|l.
`
`TYPES OF CHROMATOGRAPHY
`
`Chromatography is a technique by which the components in a sample, carried by the
`liquid or gaseous phase, are resolved by sorption-desorption steps on the stationary
`phase.
`
`A.
`
`High Performance Liquid Chromatography (HPLC)
`
`HPL chromatographic separation is based on interaction and differential
`partition of the sample between the mobile liquid phase and the stationary
`phase. The commonly used chromatographic methods can be roughly
`divided into the following groups, not necessarily in order of importance:
`
`@SnPFpNT‘
`
`Chiral
`
`|on—-exchange
`|on--pair/affinity
`Normal phase
`Reversed phase
`Size exclusion
`
`1.
`
`Chiral Chromatography
`
`Separation of the enantiomers can be achieved on chiral stationary
`phases by formation of diastere‘omers via derivatizing agents or
`mobile phase additives on achiral stationary phases. When used
`as an impurity test method, the sensitivity is enhanced if the
`enantiomeric impurity elutes before the enantiomeric drug.
`
`P- 5
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`lPR2016—00006
`
`|PR2020-00770
`
`United Therapeutics EX2007
`Page 4756 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4756 of 7335
`
`

`

`Ion—exchange Chromatography
`
`Separation is based on the charge-bearing functional groups,
`anion exchange for sample negative ion (X'), or cation exchange
`for sample positive ion (X+). Gradient elution by pH is common.
`
`Ion—pair/Affinity Chromatography
`
`Separation is based on a chemical interaction specific to the target
`species. The more popular reversed phase mode uses a buffer
`and an added counter-ion of opposite charge to the sample with
`separation being influenced by pH, ionic strength, temperature,
`concentration of and type of organic co-solvent(s). Affinity
`chromatography, common for macromolecules, employs a ligand
`'(biologically active molecule bonded covalently to the solid matrix)
`which interacts with its homologous antigen (analyte) as a
`reversible complex that can be eluted by changing buffer
`conditions.
`'
`
`Normal Phase Chromatography
`
`Normal phase chromatography is a chromatographic technique
`that uses organic solvents for the mobile phase and a polar
`stationary phase. Here, the less polar components elute faster
`than the more polar components.
`
`Reversed Phase Chromatography
`
`The test method most commonly submitted to CDER is the
`reversed phase HPLC method. UV detection is the most common
`detection technique.
`
`Reversed phase chromatography, a bonded phase
`chromatographic technique, uses water as the base solvent.
`Separation based on solvent strength and selectivity also may be
`affected by column temperature and pH.
`In general, the more
`polar components elute faster than the less polar components.
`
`UV detection can be used with all chromatographic techniques.
`The concern for this type of detector is the loss of sensitivity with
`lamp aging, and varying sensitivity at the low level depending on
`design and/or manufacturer. A point to note is that observations
`on the HPL chromatograms, by UV detection in combination with
`reversed-phase HPLC, may not be a true indication of the facts for
`
`3
`
`P- 6
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`IPR2016—00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4757 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4757 of 7335
`
`

`

`the following reasons:
`
`0
`
`0
`
`0
`
`Compounds much more polar than the compound of interest
`may be masked (elute together) in the solvent front/void
`volume.
`
`Compounds very less polar than the analyte may elute
`either late during the chromatographic run or are retained in
`the column.
`
`Compounds with lower UV extinction coefficients or different
`wavelength maxima may not be detectable at the low level
`relative to the visibility of the analyte since only one
`wavelength is normally monitored.
`
`6.
`
`Size Exclusion Chromatography
`
`Also known as gel permeation orfiltration, separation is based on
`the molecular size or hydrodynamic volume of the components.
`Molecules that are too large for the pores of the porous packing
`material on the column elute first, small molecules that enter the
`pores elute last, and the elution rates of the rest depend on their
`relative sizes.
`
`Gas Chromatography (GC)
`
`Gas chromatography is based on the volatilized sample transported by
`the carrier gas as the moving phase through the stationary phase of the
`column where separation takes place by the sorption/desorption process.
`
`Samples for gas chromatographic analysis are normally low molecular
`weight compounds that are volatile and stable at high temperature.
`In
`this respect, residual solvents in drug substances and drug products are
`suitable for gas chromatographic analysis. Chemical derivatives can also
`be formed to achieve volatility and thermal stability;
`
`Common detectors are flame ionization (FID) for carbon—containing
`compounds, electron capture (ECD) for halogenated compounds, flame
`photometric (FPD) for compounds containing sulphur or phosphorous
`and nitrogen-phosphorous (NPD) for compounds containing nitrogen or
`phosphorous. Chiral separation also can be achieved by gas
`chromatography. Separation by the packed column is rapidly being
`replaced by the capillary column that provides improved resolution and
`analysis speed. The location of the analyte on the gas chromatogram is
`
`1
`
`P- 7
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`lPR2016—00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4758 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4758 of 7335
`
`

`

`described by retention time (R) which is similar to HPLC.
`
`C.
`
`Thin-Layer Chromatography (TLC)
`
`Thin-layer chromatography is the simplest of the more common
`chromatographic techniques. Separation is based on migration of the
`sample spotted on a coated (stationary phase) plate with one edge
`dipped in a mixture of solvents (mobile phase). The whole system is
`contained in an enclosed tank.
`
`Detection techniques include fluorescence, UV and sprays (universal and
`specific) for compounds that are not naturally colored. The location of the
`analyte on the TLC plate is described by the Rf value which is the ratio of
`the migration distance of the compound Of interest to the mobile phase
`front.
`
`Of the three techniques, gas, liquid and thin—layer, TLC is the most universal test
`method as all components are present on the plate and with appropriate detection
`techniques, all components can be observed. However, it normally is not as accurate
`or sensitive as HPLC. TLC has a higher analytical variation than HPLC, although one
`sees the "whole picture“ when appropriate detection schemes are selected.
`
`III.
`
`REFERENCE STANDARDS
`
`A reference standard is a highly purified compound that is well characterized.
`Chromatographic methods rely heavily on a reference standard to provide accurate
`data. Therefore the quality and purity of the reference standard is very important. Two
`types of reference standards, chemical and nuclidic, exist. With the latter, the radio-
`label purity should also be considered as well as the chemical purity.
`
`As described in the Guideline for-Submitting Samples and Analytical Data for Methods
`Validation, the two categories _of chemical reference standards are as follows:
`
`0
`
`o
`
`USP/NF reference standard that does not need characterization, and
`
`non-compendial standard that should be of the highest purity that can be
`obtained by reasonable effort and should be thoroughly characterized to
`assure its identity, strength, quality and purity.
`
`The points to note are:
`
`0
`
`Most USP/NF reference standards do not state the purity of the
`compound.
`
`P- 8
`
`'
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`IPR2016—00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4759 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4759 of 7335
`
`

`

`o
`
`0
`
`0
`
`-
`
`The purity correction factor for non-USP, reference standards is
`recommended to be included in the calculation of the test method.
`
`In addition to structurally-related impurities from the synthesis process,
`other process impurities like heavy metals, residual solvents, moisture
`(bound and unbound), pesticides for products of plant origin, and
`degradation products can also contribute to the lack of purity in the
`reference standard.
`
`The drying of the reference standard before use, if stated in the method,
`will eliminate residual solvent(s), unbound moisture and sometimes
`bound moisture (depending on the drying conditions). The drying step is
`always included for hygroscopic compounds. On the other hand, drying
`can result in the loss of a hydrate or cause degradation in heat-sensitive
`compounds.
`
`Chromatographic test methods use either external or internal standards for
`quantitation.
`
`A.
`
`An external standard method is used when the standard is analyzed on a
`separate chromatogram from the sample. Quantitation is based on a
`comparison of the peak area/height (HPLC or GO) or spot intensity (TLC)
`of the sample to that of a reference standard of the analyte of interest.
`
`The external standard method is more appropriate for samples as follows:
`
`1.
`
`2.
`
`3.
`
`Sample with a single target concentration and narrow
`concentration range, e.g., acceptance and release tests.
`
`Simple sample preparation procedure.
`
`Increased baseline time for detection of potential extraneous
`peaks, e.g., impurities test.
`
`B.
`
`With an internal standard method, compound of known purity that does
`not cause interference in the analysis is added to the sample mixture.
`Quantitation is based on the response ratio of compound of interest to the
`internal standard vs the response ratio of a similar preparation of the
`reference standard (HPLC or GC). This technique is rarely used for TLC
`methods.
`
`The internal standard method is more appropriate for samples as follows:
`
`P- 9
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`lPR2016-00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4760 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4760 of 7335
`
`

`

`1.
`
`2.
`
`3.
`
`Complex sample preparation procedures, e.g., multiple
`extractions.
`
`Low concentration sample (sensitivity being an issue), e.g.,
`pharmacokinetics studies.
`
`Wide range of concentrations expected in the sample for analysis,
`e.g., pharmacokinetics studies,
`
`Although CDER does not specify Whether the method must use an internal or
`external standard for quantitation, it is commonly observed that HPLC methods
`for release and stability and TLC methods use external standards; and methods
`for biological fluids and GC methods use internal standards.
`
`The working concentration is the target concentration of the compound of interest as
`described in the method. Keeping the concentrations of the sample and the standard
`close to each other for the external standard method improves the accuracy of the
`method.
`
`Recommendations:
`
`1.
`
`2.
`
`Include the purity correction factor, if known, of the reference standard in
`the calculation.
`
`State the working concentrations of the standard and sample in the
`method.
`
`IV.
`
`PARAMETERS FOR VALIDATION OF HPL CHROMATOGRAPHIC
`METHODS FOR DRUG SUBSTANCE AND DRUG PRODUCT
`
`Though many types of HPL chromatographic techniques are available, the most
`commonly submitted method, the reversed-phase HPLC with UV detection, is selected
`to illustrate the parameters for validation. The criteria for the validation of this
`technique can be extrapolated to other detection methods and chromatographic
`techniques. For acceptance, release or stability testing, accuracy should be optimized
`since the need to show deviation from the actual or true value is of the greatest
`concern.
`-
`
`A.
`
`Accuracy
`
`Accuracy is the measure of how close the experimental value is to the
`true value.
`
`P- 10
`
`UT Ex. 2035
`’
`SteadyMed v. United Therapeutics
`lPR2016-00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4761 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4761 of 7335
`
`

`

`Accuracy studies for drug substance and drug product are recommended
`to be performed at the 80, .100 and 120% levels of label claim as stated in
`the Guideline for Submitting Samples and Analytical Data for Methods
`Validation.
`
`For the drug product, this is performed frequently by the addition of
`known amounts of drug by weight or volume (dissolved in diluent) to the
`placebo formulation working in the linear range of detection of the
`analyte. This would be a true recovery for liquid formulations. For
`formulations such as tablet, suppository, transdermal patch, this could
`mean evaluating potential interaction ofthe active drug with the
`excipients in the diluent. From a practical standpoint, it is difficult to
`manufacture a single unit with known amount of active drug to evaluate
`recovery. This test evaluates the specificity of the method in the
`presence of the excipients under the chromatographic conditions used for
`the analysis of the drug product.
`It will pick up recovery problems that
`could be encountered during the sample preparation and the
`chromatographic procedures. However, it does not count the effect of the
`manufacturing process.
`
`At each recommended level studied, replicate samples are evaluated.
`The RSD of the replicates will provide the analysis variation or how
`precise the test method is. The mean of the replicates, expressed as %
`label claim, indicates how accurate the test method is.
`
`Recommendations:
`
`Recovery data at least in triplicate at each level (80,100 and 120% of
`label claim)Is recommended. The mean is an estimate of accuracy and
`the RSDIs an estimate of sample analysis precision.
`
`Detection Limit and Quantitation Limit
`
`These limits are normally applied to related substances in the drug
`substance or drug product. Specifications on these limits are submitted
`with the regulatory impurities method relating to release and stability of
`both drug substance and drug product.
`Detection limit is the lowest concentration of analyte in a sample that can
`be detected, but not necessarily quantitated, under the stated
`
`P- 11
`
`UT Ex. 2035
`'
`SteadyMed v. United Therapeutics
`lPR2016-00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4762 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4762 of 7335
`
`

`

`experimental conditions. Quantitation limit is the lowest concentration of
`analyte in a sample that can be determined with acceptable precision and
`accuracy under the stated experimental conditions.
`
`WithUV detectors, it is difficult to assure the detection precision of low
`level compounds due to potential gradual loss of sensitivity of detector
`lamps with age, or noise level variation by detector manufacturer. At low
`levels, assurance is needed that the detection and quantitation limits are
`achievable with the test method each time. With no reference standard
`
`for a given impurity or means to assure detectability, extraneous peak(s)
`could "disappear/appear." A crude method to evaluate the feasibility of
`the extraneous peak detection is to use the percentage claimed for
`detection limit from the area counts of the analyte. For example,
`detection limit claim of 0.01% for the analyte integrated area count of
`50,000 will give an area count of 5 that is not detectable.
`
`Though USP expresses detection limit and quantitation limit in terms of 2 ,
`or 3, and 10 times noise level respectively, this concept is not very
`practical. Noise level on a detector during the method development
`phase may be different when samples are assayed on different detectors.
`etc. The use of standard(s) in the test method at the quantitation limit
`level (proposed by the applicant) is assurance that the impurity can be
`observed and quantitated.
`
`Detector sensitivity can vary with the model number and/or manufacturer
`as illustrated in Table ‘_l for the analysis of a compound by two
`commercial detectors. The data should not be taken as the expected
`ratio of sensitivity of the two detectors.
`It is not known if other parameters
`which can also play a part, e.g., age of lamp, column, were considered
`when setting these limits.
`
`P- 12
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`lPR2016-00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4763 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4763 of 7335
`
`

`

`Table 1.
`
`Comparison of Detector Sensitivity Limits in Two
`Commercial Detectors.
`
`_ Detector 1
`
`Detector 2
`
`
`
`
`
`
`
`Quantitatio
`:1 Limit
`
`Detection
`Limit
`
`0 . 21%
`
`0.07%
`
`0.16%
`
`,
`
`0.05%
`
`One also should be cautious that baseline noise is not interpreted as
`extraneous peaks. Undulations may be observed at the void volume if
`the diluent for the sample is different from the solvents (proportion and
`type) used in the mobile phase.
`
`If a reference standard for the compound of interest is available, a
`standard close to the quantitation limit or the specification could be used.
`For monitoring peak(s) with no reference standard for the impurity, a
`diluted reference standard of the drug substance is recommended. The
`method should then check that the high and low concentrations are
`operating in the linear range of detection of the drug substance.
`Otherwise the information that is expressed as % area or height of the
`It
`drug substance peak from the same HPL chromatogram will be biased.
`should also be noted that the extraneous peak using area count does not
`consider the detection response which depends on the UV extinction
`coefficient or absorptivity of the compound.
`
`Recommendations:
`
`1.
`
`2.
`
`Analysis repeatability and injection repeatability data at the
`quantitation limit.
`
`Use of an additional reference standard at the quantitation limit
`level in the test method.
`
`P- 13
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`lPR2016—00006
`
`lPR2020—00770
`
`United Therapeutics EX2007
`Page 4764 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4764 of 7335
`
`

`

`C.
`
`. Linearity
`
`The linear range of detectability that obeys Beer's Law is dependent
`«smonrthe compound analyzed and detector used. The working sample
`concentration and samples tested for accuracy should be in the linear
`range.
`'
`~
`
`Figures-’1 and 2 illustrate the behavior of UV response vs.
`.A point to
`.-concentration of a (a) linear and _(b) non-linear relationship.
`;note is that when monitoring impurity peaks expressed as percent
`'area' of the parent drug substance, the impurity observed may not be
`a true reflection of the theoretical amount if the non-linear section of
`:the concentration curve is employed.
`In addition, the actual amount
`will be obtained only if the extinction coefficient or absorptivity values
`are the same for both impurity and parent compound.
`Impurity
`reference standards are often needed.
`
`'
`
`Concentrations vs. Peak Areas of Standards to Illustrate
`Figure 1.
`Linearity.
`
`tame
`
`um
`
`wand“
`
`.
`'
`regression coefficient = 0.999998
`intercept = 0.103
`slope = 0.000011
`
`
`
`
`
`mama.
`.
`Marking
`
`
`.... Concentration
`8
`ruww
`L
`_
`_
`< _
`.
`x
`”WV”
`fl
`"
`II
`_
`_
`a
`u wwwm
`>
`,
`<
`'
`,
`4000909
`
`
`
`
`
`”Wm
`
`
`-I—l
`l-—-——l-——-‘l~-'--—l-——i--——i——I--——-I'-—--4 eu—l'
`“9
`[IV
`20
`so
`a
`ID
`IO
`7D
`In
`N
`IN
`In
`
`IV
`
`.
`
`,
`
`0
`
`uq-par ul
`
`1 1
`
`P- 14
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`|PR2016—00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4765 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4765 of 7335
`
`

`

`Figure 2.
`
`Concentrations vs. Peak Areas of Standards Outside the Linear Range.
`
`KIO‘D
`
`'31,)
`
`'
`
`.
`
`n no u m a m u n u m 7“
`o-l——I—i—+—i——I—i—i—+—+—i——+—--Hi———i-—+—--i ---+-—-l
`
`Concentration ue/nL
`
`
`Recommendations:
`
`- The linearity range for examination depends on the purpose of the test
`method. For example, the recommended range for an assay method
`-for content would be NLT i 20% and the range for an
`assay/impurities combination method based onrarea % (for impurities)
`would be +20% of target concentration down to the limit of
`quantitation of the drug substance or impurity. Under most
`circumstances, regression coefficient (r) is 2 0.999.
`Intercept and
`slope should be indicated.
`
`12
`
`P- 15
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`lPR2016-00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4766 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4766 of 7335
`
`

`

`Precision
`
`Precision is the measure of how close the data values are to each other
`
`for a number of measurements under the same analytical conditions.
`ICH has defined precision to contain three components: repeatability,
`intermediate precision and reproducibility. Ruggedness as defined in
`USP XXII <1225>, 1990 incorporates the concepts described under the
`terms "intermediate precision”, "reproducibility" and "robustness" of this
`guide.
`
`1.
`
`Repeatability
`
`a.
`
`Injection Repeatability
`
`Sensitivity is the ability to detect small changes in the
`concentration of the analyte in the sample. Sensitivity can
`be partially controlled by monitoring the specification for
`injection reproducibility (system suitability testing).
`
`The sensitivity or precision as measured by multiple
`injections of a homogeneous sample (prepared solution)
`indicates the performance of the HPLC instrument under the
`chromatographic conditions and day tested. The
`information is provided as part of the validation data and as
`a system suitability test. The specification, as the
`coefficient of variation in % or relative standard deviation
`
`(RSD), set here will determine the variation limit of the
`analysis. The tighter the value, the more precise or
`sensitive to variation one can expect the results. This
`assumes that the chromatograph does not malfunction after
`the system suitability testing has been performed. Keep in
`mind,‘ however, that it does not consider variations due to
`the drug product manufacturing and laboratory sample
`preparation procedures. As an illustration for injection and
`_Rt variation, Table 2 provides representative data collected
`when a leak developed in the chromatographic system
`during sampling. The set of four duplicate samples were
`injected sequentially. Variations in peak area and drift of
`retention times are noted. Sets of typical data from a well-
`behaved system for comparison are shown in Table 3.
`
`13
`
`P- 15
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`lPR2016-00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4767 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4767 of 7335
`
`

`

`Table 2.
`
`Representative Injection Repeatability Data for an HPL
`Chromatographic System that Developed a Leak During
`Sampling.
`'
`
`
`_“
`
`2155699
`004
`
`
`2120466
`
`
`A1
`A2
`
`B1
`B2
`
`C1
`C2
`
`562
`566
`
`5.52
`613
`
`6.48
`
`A Peak Area
`
`35233
`
`52555
`
`38213
`
`20128
`
`2255555
`2288355
`
`2265279
`
`2602016
`
`Illlllllliailllll
`lllllllliigilllll
`
`Table 3.
`
`Representative Injection Repeatability Data for Select
`Formulations from a Normally Functional HPL
`Chromatographic System.
`
`Inhalation
`Solution
`
`Solution for
`Inhalation
`
`10
`
`10
`
`
`
`1993162 i
`5029
`
`0.25%
`
`1722253 i
`6288
`
`1744320 i
`3133
`
`
`
`
`
`H
`
`P- 17
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`IPR2016-00006
`
`|PR2020-00770
`
`United Therapeutics EX2007
`Page 4768 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4768 of 7335
`
`

`

`Recommendations:
`
`As part of methods validation, a minimum of 10 injections
`with an RSD of 51% is recommended. With the methods
`
`for release and stability studies, an RSD of s 1% RSD for
`precision of the system suitability tests for at least five
`injections (n 2 5) for the active drug either in drug substance
`or drug product is desirable. For low level impurities, higher
`variations may be acceptable.
`
`b.
`
`Analysis Repeatability
`
`Determination, expressed as the RSD, consists of multiple
`measurements of a sample by the same analyst under the
`same analytical conditions. For practical purpose, it is often
`combined with accuracy and carried out as a single study.
`See section IV.A under Accuracy.
`
`intermediate Precision
`
`Intermediate precision was previously known as part of
`ruggedness. The attribute evaluates the reliability of the method in
`a different environment other than that used during development of
`the method. The objective is to ensure that the method will provide
`the same results when similar samples are analyzed once the
`method development phase is over.
`
`Depending on time and resources, the method canvbe tested on
`multiple days, analysts, instruments, etc.
`
`Intermediate precision in the test method can be. partly assured by
`good system suitability specifications. Thus, it is important to set
`tight, but realistic, systemsuitability specifications.
`
`Recommendations:
`
`As a minimum, data generated as described under section |V.A
`Accuracy, for two separate occasions, is recommended to indicate
`the intermediate precision of the test method.
`
`P- 18
`
`UT Ex. 2035
`SteadyMed v. United Therapeutics
`lPR2016-00006
`
`lPR2020-00770
`
`United Therapeutics EX2007
`Page 4769 of 7335
`
`IPR2020-00770
`United Therapeutics EX2007
`Page 4769 of 7335
`
`

`

`3.
`
`Reproducibility
`
`As defined by lCH, reproduci

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket