throbber
Trials@uspto.gov
`Tel: 571-272-7822
`
`
`Paper 26
`Entered: June 14, 2021
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`APPLE INC.,
`Petitioner,
`v.
`OMNI MEDSCI, INC.,
`Patent Owner.
`
`IPR2020-00175
`Patent 10,188,299 B2
`
`
`
`
`
`
`
`
`
`Before GRACE KARAFFA OBERMANN, JOHN F. HORVATH, and
`SHARON FENICK, Administrative Patent Judges.
`Opinion for the Board filed by Administrative Patent Judge FENICK.
`
`Opinion Concurring filed by Administrative Patent Judge HORVATH.
`
`
`
`JUDGMENT
`Final Written Decision
`Determining All Challenged Claims Unpatentable
`35 U.S.C. § 318(a)
`
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`
`I.
`
`INTRODUCTION
`A. Background
`Apple Inc. (“Petitioner”) filed a Petition requesting inter partes
`review of claims 7 and 10–14 (“the challenged claims”) of U.S. Patent No.
`10,188,299 B2 (Ex. 1001, “the ’299 patent”). Paper 1 (“Pet.”), 3. Omni
`MedSci Inc. (“Patent Owner”) filed a Preliminary Response. Paper 6
`(“Prelim. Resp.”). After considering the Petition, Preliminary Response, and
`additional briefing, we instituted inter partes review of all challenged claims
`on all grounds raised. Paper 11 (“Dec. Inst.”).
`Patent Owner filed a Response to the Petition (Paper 13, “PO Resp.”),
`Petitioner filed a Reply (Paper 15, “Pet. Reply”), and Patent Owner filed a
`Sur-reply (Paper 18, “PO Sur-reply”). An oral hearing was held on
`March 25, 2021, and the hearing transcript is included in the record. See
`Paper 25 (“Tr.”).
`We have jurisdiction under 35 U.S.C. § 6(b)(4). This is a Final
`Written Decision under 35 U.S.C. § 318(a) and 37 C.F.R. § 42.73. For the
`reasons set forth below, we find Petitioner has shown by a preponderance of
`evidence that claims 7 and 10–14 of the ’299 patent are unpatentable.
`B. Real Parties-in-Interest
`Petitioner and Patent Owner each identifies only itself as a real party-
`in-interest. Pet. x; Paper 4, 1.
`C. Related Matters
`Petitioner and Patent Owner identify the following as matters that
`could affect or be affected by a decision in this proceeding:
`Issued patents: U.S. Patent No. 9,651,533 and U.S. Patent No.
`9,164,032.
`
`2
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`
`Litigation: Omni MedSci, Inc. v. Apple Inc., Action No. 2-19-cv-
`05673-YGR (N.D. Cal.); Omni MedSci, Inc. v. Apple Inc., Action No.
`2-19-cv-05924 (N.D. Cal.); Omni MedSci, Inc. v. Apple Inc., Action
`No. 2-18-cv-00429-RWS (E.D. Tex.) (terminated); and Omni MedSci,
`Inc. v. Apple Inc., Action No. 2-18-cv-00134-RWS (E.D. Tex.)
`(terminated).
`Inter partes review proceedings: IPR2019-00913 (terminated) and
`IPR2019-00916 (instituted).
`See Pet. xii–xiii; Paper 4, 1–2.
`D. Overview of the ’299 Patent
`The ’299 patent was filed on May 12, 2017, and claims priority to a
`utility application filed on December 17, 2013 and a provisional application
`filed on December 31, 2012. Ex. 1001, codes (22), (60), (63), 1:7–13. The
`’299 patent is directed to a system for measuring physiological parameters.
`Id. at code (54).
`The system, in one embodiment, includes a wearable measurement
`device for measuring physiological parameters. Id. at 6:48–50. This
`measurement device includes a light source including multiple
`semiconductor sources configured to generate an output optical beam in
`which a portion of the wavelengths of the output optical beam are of a near-
`infrared wavelength between 700 and 2500 nanometers. Id. at 6:50–55. A
`portion of this output optical beam is delivered to a sample and some portion
`of the beam reflected from or transmitted through the sample is received and
`processed. Id. at 6:55–63.
`A system including a wearable device is depicted in Figure 24 of the
`’299 patent, which is reproduced below.
`
`3
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`
`
`Figure 24 is a schematic illustration of a medical measurement device
`that is part of a personal or body area network. Id. at 8:25–30. Wearable
`measurement device 2401 is a physiological measurement or blood
`constituent measurement device that includes processor 2402 and transmitter
`2403. Id. at 30:16–19. Communication link 2404 allows communication
`between measurement device 2401 and personal device 2405. Id. at 30:16–
`22. Personal device 2405, which may be a smart phone, optionally has a
`receiver, a transmitter, a display, a voice control, a speaker, one or more
`buttons or knobs, and/or a touch screen. Id. at 30:39–43. Personal device
`2405 stores, processes, displays, and transmits at least a portion of the output
`signal generated by wearable measurement device 2401. Id. at 30:37–39.
`Personal device 2405 also transmits some of the data or the processed output
`
`4
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`signal over a wireless transmission link to internet or “cloud” 2407. Id.
`at 30:46–49. Internet or cloud 2407 may provide services including storage,
`processing, and retransmission of data to the originator or to another
`designated recipient such as a health care provider or doctor. Id. at 30:55–
`67.
`
`The light source of the measurement device can increase a signal-to-
`noise ratio by increasing either the LED intensity or pulse rate. Id. at code
`(57), 3:11–16. Additionally, “change detection schemes may be used, where
`the detection system captures the signal with the light source on and with the
`light source off. . . . Then, the signal with and without the light source is
`differenced. This may enable the sun light changes to be subtracted out.”
`Id. at 29:13–18.
`
`E. Illustrative Claims
`Claim 7 of the ’299 patent is the sole independent claim of the
`challenged claims and is reproduced below, with one limitation (the “pulse
`rate limitation”) italicized for emphasis.
`7. A system for measuring one or more physiological parameters
`comprising:
`a light source comprising a plurality of semiconductor sources
`that are light emitting diodes, each of the light emitting
`diodes configured to generate an output optical beam having
`one or more optical wavelengths, wherein at least a portion
`of the one or more optical wavelengths is a near-infrared
`wavelength between 700 nanometers and 2500 nanometers;
`a lens configured to receive a portion of at least one of the
`output optical beams and to deliver a lens output beam to
`tissue;
`a detection system configured to receive at least a portion of the
`lens output beam reflected from the tissue and to generate an
`output signal having a signal-to-noise ratio, wherein the
`
`5
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`
`detection system is configured to be synchronized to the
`light source;
`a personal device comprising a wireless receiver, a wireless
`transmitter, a display, a microphone, a speaker, one or more
`buttons or knobs, a microprocessor and a touch screen, the
`personal device configured to receive and process at least a
`portion of the output signal, wherein the personal device is
`configured to store and display the processed output signal,
`and wherein at least a portion of the processed output signal
`is configured to be transmitted over a wireless transmission
`link;
`a remote device configured to receive over the wireless
`transmission link an output status comprising the at least a
`portion of the processed output signal, to process the
`received output status to generate processed data, and to
`store the processed data;
`wherein the output signal is indicative of one or more of the
`physiological parameters, and the remote device is
`configured to store a history of at least a portion of the one
`or more physiological parameters over a specified period of
`time;
`the system configured to increase the signal-to-noise ratio by
`increasing light intensity of at least one of the plurality of
`semiconductor sources from an initial light intensity and by
`increasing a pulse rate of at least one of the plurality of
`semiconductor sources from an initial pulse rate; and
`the detection system further configured to:
`generate a first signal responsive to light received while the
`light emitting diodes are off,
`generate a second signal responsive to light received while at
`least one of the light emitting diodes is on, and
`increase the signal-to-noise ratio by differencing the first signal
`and the second signal.
`Ex. 1001, 33:29–34:11, Certificate of Correction.
`
`6
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`
`Claims 10 and 11 depend from claim 7, claim 12 depends from claim
`11, claim 13 depends from claim 12, and claim 14 depends from claim 13.
`Id. at 34:21–37.
`
`F. Evidence
`
`Reference
`Mannheimer
`
`US 5,746,206
`
`Lisogurski et al. US 9,241,676 B2
`
`Park et al.
`
`US 9,596,990 B2
`
`Date
`May 5, 1998
`
`May 31, 2012
`
`Nov. 6, 2013
`
`Carlson et al.
`
`US 2005/0049468 A1 Mar. 3, 2005
`
`Exhibit
`1008
`
`1011
`
`1010
`
`1009
`
`Petitioner also relies upon the Declaration of Brian Anthony, Ph.D.
`(Ex. 1003). Patent Owner relies upon two declarations of Duncan L.
`MacFarlane, Ph.D., P.E. (Ex. 2122; Ex. 2131). Petitioner cross-examined
`Dr. MacFarlane by deposition. Ex. 1065.
`G. Asserted Grounds of Unpatentability
`Petitioner asserts that the challenged claims would have been
`unpatentable on the following grounds:
`Claims
`Challenged
`7, 11–13
`12, 13
`10, 14
`
`References
`Lisogurski, Carlson
`Lisogurski, Carlson, Mannheimer
`Lisogurski, Carlson, Park
`
`35 U.S.C. §1
`103
`103
`103
`
`
`1 The Leahy-Smith America Invents Act, Pub. L. No. 112-29, 125 Stat. 284
`(2011), amended 35 U.S.C. § 103 effective March 16, 2013. Because the
`’299 patent claims priority to a provisional filed prior to the effective date of
`the applicable AIA amendment, we refer to the pre-AIA version of § 103.
`
`7
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`
`Claims
`Challenged
`142
`
`35 U.S.C. §1
`103
`
`References
`Lisogurski, Carlson, Park, Mannheimer
`
`
`
`II. ANALYSIS
`A. Legal Standards
`It is a petitioner’s burden to demonstrate unpatentability. See
`Dynamic Drinkware, LLC v. Nat’l Graphics, Inc., 800 F.3d 1375, 1378
`(Fed. Cir. 2015) (citing Tech. Licensing Corp. v. Videotek, Inc., 545 F.3d
`1316, 1326–27 (Fed. Cir. 2008)).
`A claim is unpatentable as obvious if “the differences between the
`claimed invention and the prior art are such that the subject matter as a
`whole would have been obvious at the time the invention was made to a
`person having ordinary skill in the art to which said subject matter pertains.”
`35 U.S.C. § 103(a) (2012). The question of obviousness is resolved on the
`basis of underlying factual determinations, including: (1) the scope and
`content of the prior art; (2) any differences between the claimed subject
`matter and the prior art; (3) the level of skill in the art; and (4) objective
`evidence of nonobviousness.3 Graham v. John Deere Co. of Kansas City,
`383 U.S. 1, 17–18 (1966).
`
`
`2 Although Petitioner styles this ground as relating to both claims 10 and 14,
`claim 10 depends from claim 7, and the Petition does not provide any
`indication of what portion of claims 7 or 10 Petitioner asserts is taught or
`suggested by Mannheimer. Pet. 3, 60–66, 70; Inst. Dec. 3. Accordingly, we
`treat this ground as directed solely to claim 14.
`3 No argument or evidence concerning secondary considerations has been
`adduced in the current record.
`
`8
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`
`Even if prior art references disclose all claim limitations when
`combined, there must be evidence to explain why a person of ordinary skill
`in the art would have combined the references to arrive at the claimed
`invention. Kinetic Concepts, Inc. v. Smith & Nephew, Inc., 688 F.3d 1342,
`1366–67 (Fed. Cir. 2012) (citing Innogenetics, N.V. v. Abbott Labs., 512
`F.3d 1363, 1374 (Fed. Cir. 2008) (holding that “some kind of motivation
`must be shown from some source, so that the [trier of fact] can understand
`why a person of ordinary skill would have thought of either combining two
`or more references or modifying one to achieve the patented [invention]”)).
`An invention “composed of several elements is not proved obvious merely
`by demonstrating that each of its elements was, independently, known in the
`prior art.” KSR Int’l Co. v. Teleflex Inc., 550 U.S. 398, 418 (2007). Rather,
`“it can be important to identify a reason that would have prompted a person
`of ordinary skill in the relevant field to combine the elements in the way the
`claimed new invention does.” Id.
`An obviousness determination “cannot be sustained by mere
`conclusory statements; instead, there must be some articulated reasoning
`with some rational underpinning to support the legal conclusion of
`obviousness.” Id. (quoting In re Kahn, 441 F.3d 977, 988 (Fed. Cir. 2006));
`see In re Magnum Oil Tools Int’l, Ltd., 829 F.3d 1364, 1380 (Fed. Cir.
`2016).
`
`B. Level of Ordinary Skill in the Art
`Petitioner, relying on the testimony of Dr. Anthony, identifies a
`person of ordinary skill in the art (“POSITA”) as someone who “would have
`[had] a good working knowledge of optical sensing techniques and their
`applications, and familiarity with optical system design and signal
`
`9
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`processing techniques.” Pet. 15; Ex. 1003 ¶ 37. Such a person would have
`obtained such knowledge through “an undergraduate education in
`engineering (electrical, mechanical, biomedical, or optical) or a related field
`of study, along with relevant experience studying or developing
`physiological monitoring devices . . . in industry or academia.” Pet. 15–16.
`Patent Owner does not comment on this definition or provide an alternative
`definition of a person of ordinary skill.
`We find Petitioner’s undisputed definition of the person of ordinary
`skill in the art to be consistent with the problems and solutions disclosed in
`the patent and prior art of record, and adopt it as our own. See, e.g., In re
`GPAC Inc., 57 F.3d 1573, 1579 (Fed. Cir. 1995).
`C. Claim Construction
`In inter partes reviews, we interpret a claim “using the same claim
`construction standard that would be used to construe the claim in a civil
`action under 35 U.S.C. 282(b).” 37 C.F.R. § 42.100(b) (2019). Under this
`standard, we construe the claim “in accordance with the ordinary and
`customary meaning of such claim as understood by one of ordinary skill in
`the art and the prosecution history pertaining to the patent.” Id. Only claim
`terms that are in controversy need to be construed and only to the extent
`necessary to resolve the controversy. See Nidec Motor Corp. v. Zhongshan
`Broad Ocean Motor Co., 868 F.3d 1013, 1017 (Fed. Cir. 2017).
`Initially, Petitioner requested construction of the terms “beam” and
`“lens,” and Patent Owner, in the Preliminary Response, requested
`construction of the pulse rate limitation. Pet. 18–19; Prelim. Resp. 10. In
`the Decision on Institution we determined that, for the purposes of
`
`10
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`institution, no constructions were necessary. Dec. Inst. 18–19 (citing Nidec,
`868 F.3d at 1017).
`In its Response, Patent Owner contends that no claim construction is
`necessary for any term. PO Resp. 10. Petitioner does not further address the
`construction of “beam” or “lens” in its Reply, but does obliquely address the
`construction of the pulse rate limitation, casting Patent Owner’s arguments
`regarding patentability as “impos[ing] special requirements for how the
`claimed devices increase signal-to-noise ratio [SNR].” Reply 1. However,
`Petitioner does not describe or request a specific construction for the pulse
`rate limitation. Id. We address the parties’ arguments relating to the pulse
`rate limitation below, at Section II.D.2.i. With respect to claim construction,
`however, we determine that no constructions of “beam,” “lens,” or the pulse
`rate limitation are necessary to resolve the controversy before us.
`D. Patentability of Claims 7, 11, 12, and 13 over Lisogurski and Carlson
`Petitioner argues claims 7 and 11 are unpatentable over the
`combination of Lisogurski and Carlson. Pet. 20–60. Patent Owner argues
`that Lisogurski does not teach or suggest the pulse rate limitation. PO Resp.
`11–18; PO Sur-reply 1–11.
`1. Overview of the Prior Art
`a) Lisogurski
`Lisogurski discloses a “physiological monitoring system [that]
`monitor[s] one or more physiological parameters of a patient . . . using one
`or more physiological sensors.” Ex. 1011, 3:44–46. The physiological
`sensors may include a “pulse oximeter [that] non-invasively measures the
`oxygen saturation of a patient’s blood.” Id. at 3:62–64. The pulse oximeter
`includes “a light sensor that is placed at a site on a patient, typically a
`
`11
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`fingertip, toe, forehead, or earlobe.” Id. at 4:6–7. The light sensor “pass[es]
`light through blood perfused tissue and photoelectrically sense[s] the
`absorption of the light in the tissue.” Id. at 4:8–11. The light sensor emits
`“one or more wavelengths [of light] that are attenuated by the blood in an
`amount representative of the blood constituent concentration,” and may
`include red and infrared (IR) wavelengths of light. Id. at 4:42–48.
`Figure 3 of Lisogurski is reproduced below.
`
`
`
`
`Figure 3 of Lisogurski is “a perspective view of a physiological
`monitoring system.” Id. at 2:23–25. The system includes sensor 312,
`monitor 314, and multi-parameter physiological monitor 326. Id. at 17:35–
`36, 18:44–45. Sensor 312 includes “one or more light sources 316 for
`emitting light at one or more wavelengths,” and detector 318 for “detecting
`the light that is reflected by or has traveled through the subject’s tissue.” Id.
`at 17:37–42. Sensor 312 may have “[a]ny suitable configuration of light
`
`12
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`source 316 and detector 318,” and “may include multiple light sources and
`detectors [that] may be spaced apart.” Id. at 17:42–45. Light source 316
`may include “LEDs of multiple wavelengths, for example a red LED and an
`IR [LED].” Id. at 19:25–27. Sensor 312 may be “wirelessly connected to
`monitor 314.” Id. at 17:57–59.
`Monitor 314 “calculate[s] physiological parameters based at least in
`part on data relating to light emission . . . received from one or more sensor
`units such as sensor unit 312.” Id. at 17:59–62. Monitor 314 includes
`“display 320 . . . to display the physiological parameters,” and “speaker 322
`to provide an audible . . . alarm in the event a subject’s physiological
`parameters are not within a predefined normal range.” Id. at 18:3–10.
`Monitor 314 is “communicatively coupled to multi-parameter physiological
`monitor 326” (“MPPM 326”) and “may communicate wirelessly” with
`MPPM 326. Id. at 18:58–61. Monitor 314 may also be “coupled to a
`network to enable the sharing of information with servers or other
`workstations.” Id. at 18:62–65.
`Multi-parameter physiological monitor 326 may also “calculate
`physiological parameters and . . . provide a display 328 for information from
`monitor 314.” Id. at 18:49–52. MPPM 326 may also be “coupled to a
`network to enable the sharing of information with servers or other
`workstations.” Id. at 18:62–65. The remote network servers may also “be
`used to determine physiological parameters,” and may display the
`parameters on a remote display, display 320 of monitor 314, or display 328
`of MPPM 326. Id. at 20:53–58. The remote servers may also “publish the
`data to a server or website,” or otherwise “make the parameters available to
`a user.” Id. at 20:58–60.
`
`13
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`
`Lisogurski discloses that the monitoring system shown in Figure 3,
`described above, “may include one or more components of physiological
`monitoring system 100 of FIG. 1.” Id. at 17:32–35. Lisogurski further
`discloses that although “the components of physiological monitoring system
`100 . . . are shown and described as separate components. . . . the
`functionality of some of the components may be combined in a single
`component,” and “the functionality of some of the components . . . may be
`divided over multiple components.” Id. at 15:66–16:8. Figure 1 of
`Lisogurski is reproduced below.
`
`Figure 1 of Lisogurski is a “block diagram of an illustrative physiological
`monitoring system.” Id. at 2:11–13. The system includes “sensor 102 and
`monitor 104 for generating and processing physiological signals of a
`subject.” Id. at 10:44–46. Sensor 102 includes “light source 130 and
`
`
`
`14
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`detector 140.” Id. at 10:48–49. Light source 130 includes “a Red light
`emitting source and an IR light emitting source,” such as Red and IR
`emitting LEDs, with the IR LED emitting light with a “wavelength [which]
`may be between about 800 nm and 1000 nm.” Id. at 10:52–58. Detector
`140 “detect[s] the intensity of light at the Red and IR wavelengths,” converts
`them to an electrical signal, and “send[s] the detection signal to monitor 104,
`where the detection signal may be processed and physiological parameters
`may be determined.” Id. at 11:9–10, 11:20–23.
`Monitor 104 includes user interface 180, communication interface
`190, and control circuitry 110 for controlling (a) light drive circuitry 120, (b)
`front end processing circuitry 150, and (c) back end processing circuitry 170
`via “timing control signals.” Id. at 11:33–38, Fig. 1. Light drive circuitry
`120 “generate[s] a light drive signal . . . used to turn on and off the light
`source 130, based on the timing control signals.” Id. at 11:38–41. The light
`drive signal “control[s] the intensity of light source 130 and the timing of
`when the light source 130 is turned on and off.” Id. at 11:50–54. Front end
`processing circuitry 150 “receive[s] a detection signal from detector 140 and
`provides one or more processed signals to back end processing circuitry
`170.” Id. at 12:42–45. Front end processing circuitry 150 also
`“synchronize[s] the operation of an analog-to-digital converter and a
`demultiplexer with the light drive signal based on the timing control
`signals.” Id. at 11:43–46.
`Back end processing circuitry 170 “use[s] the timing control signals to
`coordinate its operation with front end processing circuitry 150.” Id. at
`11:46–49. Backend processing circuitry 170 includes processor 172 and
`memory 174, and “receive[s] and process[es] physiological signals received
`
`15
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`from front end processing circuitry 150” in order to “determine one or more
`physiological parameters.” Id. at 14:56–57, 14:60–64. Backend processing
`circuitry 170 is “communicatively coupled [to] use[r] interface 180 and
`communication interface 190.” Id. at 15:16–18. User interface 180 includes
`“user input 182, display 184, and speaker 186,” and may include “a
`keyboard, a mouse, a touch screen, buttons, switches, [and] a microphone.”
`Id. at 15:19–22. Communication interface 190 allows “monitor 104 to
`exchange information with external devices,” and includes transmitters and
`receivers to allow wireless communications. Id. at 15:43–44, 15:48–57.
`Lisogurski teaches the physiological monitoring system may modulate
`the light drive signal to have a “period the same as or closely related to the
`period of [a] cardiac cycle.” Id. at 25:49–51. Thus, “[t]he system may vary
`parameters related to the light drive signal including drive current or light
`brightness, duty cycle, firing rate, . . . [and] other suitable parameters.” Id.
`at 25:52–55. Lisogurski further teaches “the system may alter the cardiac
`cycle modulation technique based on the level of noise, ambient light, [and]
`other suitable reasons.” Id. at 9:46–48. Thus, “[t]he system may increase
`the brightness of the light sources in response to [any] noise to improve the
`signal-to-noise ratio.” Id. at 9:50–52. The system may also “change from a
`modulated light output to a constant light output in response to noise, patient
`motion, or ambient light.” Id. at 9:57–60.
`b) Carlson
`Carlson discloses an “optical pulsoximetry [device] used for non-
`invasive measurement of pulsation and oxygen saturation in arterial human
`or animal blood.” Ex. 1009 ¶ 2. The device measures the light “absorption
`of reduced (Hb)—and oxidized (HbO2) h[e]moglobin at two optical
`
`16
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`wavelengths, where the relative absorption coefficients differ significantly.”
`Id. ¶ 3. Figure 2 of Carlson is reproduced below.
`
`
`Figure 2 of Carlson is a schematic illustration of an ear clip sensor 1
`of a pulsoximeter device. Id. ¶¶ 33, 49. Sensor 1 includes light source 15,
`which transmits light beam 8 through a patient’s earlobe 2, and light detector
`11 to detect the transmitted light. Id. ¶ 49. Light source 15 emits light at
`two wavelengths—660 nm and 890 nm—and can consist of two LEDs. Id.
`¶ 50.
`Carlson’s pulsoximeter can be used to “survey the heath condition of
`a person or an animal [that] is mobile,” and is “not restricted for use in, e.g.,
`a hospital.” Id. ¶ 72. Carlson teaches that “standard pulsoximeter sensors
`suffer from signal instability and insufficient robustness versus
`environmental disturbances.” Id. ¶ 4. For example, when a sensor is worn
`by a person driving along a tree-lined avenue, the sensor will receive
`sunlight “at a certain frequency” such that “every time when passing a tree,
`sunlight is attenuated and between the trees sunlight is influencing the
`measurement of the pulsoximeter sensor.” Id. ¶ 68. To address such
`
`17
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`problems, Carlson includes “optical and/or electronic means for increasing
`Signal-to-Noise ratio (S/N) . . . of a pulsoximeter sensor for robust
`application of pulsoximetry in telemedicine- and near patient testing
`applications in rough (optical) environmental conditions.” Id. ¶ 10. In
`particular, the LEDs in Carlson’s sensor emit light “not as a current or
`continuous light but as pulsed light.” Id. ¶ 69. Carlson’s sensor also uses
`“AC-Coupling or Lock-In Amplification (synchronous detection) . . . to
`temporarily modulate the amplitude of the optical radiation of . . . the LED
`at a carrier frequency f0 in order to shift the power spectrum of the
`pulsoximeter signals into a higher frequency range.” Id. ¶ 20. Modulation
`frequency f0 is selected to be “outside the frequency spectrum of sunlight
`and of ambient light.” Id. ¶ 69. This allows the pulsoximeter signal to be
`easily discriminated from environmental signals, such as sunlight and
`ambient light, and “increas[es] significantly the Signal-to-Noise and Signal-
`to-Background ratio.” Id.
`Carlson further discloses that sensor 1 can be wirelessly connected to
`“a special unit worn by [a] person or patient,” where “a signal is generated if
`[a] measured value is not within a predetermined range.” Id. ¶¶ 77–78. The
`generated signal can be “transmitted to a respective person, to a medical
`doctor, to a hospital, etc.” Id. ¶ 78. The pulsoximeter can also include a
`“GPS device which at any time gives the location of the person using the
`pulsoximetric sensor monitoring configuration.” Id.
`c) Petitioner’s Proposed Combination
`Petitioner proposes combining Lisogurski’s physiological monitoring
`system shown in Figures 1 and 3, in which sensor 102/312 wirelessly
`communicates with monitor 104/314, with Carlson’s teachings regarding
`
`18
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`selecting an LED pulse frequency to increase signal-to-noise in a wireless
`pulse oximeter sensor. See Pet. 22–24, 29–31, 34–37, 50–52. Petitioner’s
`proposed combination also relocates some of the components of
`Lisogurski’s monitor 104/314 to sensor 102/312, as illustrated in a series of
`Petitioner-modified versions of Figure 1 of Lisogurski, which we combine
`below into a single modified version of Figure 1. See id. at 35, 38, 45.
`
`
`
`Modified Figure 1 of Lisogurski demonstrates Petitioner’s proposed
`combination, which relocates some components of monitor 104 (i.e., control
`circuitry 110, light drive circuitry 120, and front end processing circuitry
`150) to sensor 102 as illustrated in the Petitioner-modified versions of Figure
`1 provided in the Petition. Id.
`Petitioner articulates sufficient reasoning with rational underpinning
`to demonstrate why a person of ordinary skill in the art would have modified
`Lisogurski’s sensor 102/312 and monitor 104/314 in the manner proposed.
`See id. at 7–11, 34–37, 44–47 (citing/quoting Ex. 1003 ¶¶ 50–58, 109, 113–
`
`19
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`117, 139–143; Ex. 1005 ¶ 3; Ex. 1009 ¶ 4; Ex. 1011, 11:20–27, 11:38–41,
`11:50–54, 16:2–9, 17:32–35, 17:55–59, 18:16–31, 25:52–55, Fig. 1;
`Ex. 1020, 3, 6–7, 12; Ex. 1021, 2–4; Ex. 1022, 1; Ex. 1023, 1, 2, 5, 6;
`Ex. 1024, 459, 460, 462; Ex. 1027, 9, 10, 15–31, 33, 35, 40–49; Ex. 1029,
`221).
`
`First, Lisogurski expressly suggests the modification by teaching
`embodiments in which “the functionality of some of the components may be
`combined in a single component” and embodiments in which “the
`functionality of some of the components of monitor 104 . . . may be divided
`over multiple components.” Ex. 1011, 16:2–4, 16:7–9. Second, numerous
`industry trends would have motivated the modification. These include
`improving the capabilities of wearable sensors for use in sports and personal
`fitness applications and wirelessly connecting wearable sensors to networks
`to remotely monitor patient health. See Ex. 1005 ¶ 3; Ex. 1009 ¶ 4;
`Ex. 1020, 3; Ex. 1021, 2–4; Ex. 1022, 1; Ex. 1024, 462, Ex. 1027, 9, 10, 15,
`33, 35, 40–49; Ex. 1029, 221; see also KSR, 550 at 418 (“When a work is
`available in one field of endeavor, design incentives and other market forces
`can prompt variations of it . . . . If a person of ordinary skill can implement a
`predictable variation, § 103 likely bars its patentability.”). Patent Owner
`does not dispute these contentions. See PO Resp. 10–37.
`Petitioner also articulates sufficient reasoning with rational
`underpinning to demonstrate why a person of ordinary skill in the art would
`have modified Lisogurski’s light source to incorporate the teachings of
`Carlson. See Pet. 22–24, 30–31, 50–52 (citing/quoting Ex. 1003 ¶¶ 50–58,
`79–83, 103–104, 157–162; Ex. 1009 ¶¶ 2, 4, 10, 14, 48, 52, 67–69;
`
`20
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`Ex. 1011, 1:4–6, 1:16–18, 1:67–2:3, 3:50–53, 4:15–20, 4:63–67, 5:55–61,
`6:3–6, 9:46–60, 13:60–14:10, 14:40–55, 17:51–58, 37:6–20; Ex. 1019, 765).
`Patent Owner argues that “Carlson adds nothing relevant to
`Lisogurski because the two references contain identically the same teaching
`– right down to the same exemplary 1000 Hz modulation rate” and “[a]n
`ordinary artisan reading Carlson would learn nothing new beyond what
`Lisogurski already discloses.” PO Resp. 3–4 (citing Ex. 1011 6:30;
`Ex. 1009 ¶ 69). Patent Owner further argues that, “[i]n fact, Carlson adds
`less than nothing because it merely modulates, temporarily, an unmodulated
`light source . . . and without changing the predetermined modulation
`frequency.” Id. at 3.
`Lisogurski and Carlson teach complementary and combinable
`methods for increasing signal-to-noise in a wearable pulsoximeter in the
`presence of ambient light or sunlight. For example, Lisogurski “may alter
`the cardiac cycle modulation technique based on the level of noise, ambient
`light, [or] other suitable reasons.” Ex. 1011, 9:46–48. In particular,
`Lisogurski “may increase the brightness of the light sources in response to
`the noise to improve the signal-to-noise ratio. In some embodiments, the
`system may increase brightness throughout the cardiac cycle . . . .” Id.
`at 9:50–54. Carlson includes “electronic means for increasing the Signal-to-
`Noise . . . in rough (optical) environmental conditions, e.g., at changing light
`influences, such as sunlight, shadow, artificial light, etc.” Ex. 1009 ¶ 10. In
`particular, Carlson pulses LEDs at a “frequency [that] is chosen in such a
`way that it is outside the frequency spectrum of sunlight and of ambient
`light. . . . [thereby] increasing significantly the Signal-to-Noise.” Id. ¶ 69.
`
`21
`
`

`

`IPR2020-00175
`Patent 10,188,299 B2
`
`
`We are not persuaded by Patent Owner’s argument that one of
`ordinary skill would not have combined Carlson and Lisogurski because the
`teachings are redundant. While Patent Owner points out similarities
`between aspects of the two references, these similarities would have
`motivated the combination rather than militated a

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket