throbber

`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`___________________________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`__________________________
`
`MEDTRONIC, INC., AND MEDTRONIC VASCULAR, INC.
`
`Petitioners,
`
`v.
`
`TELEFLEX INNOVATIONS S.À.R.L.,
`
`Patent Owner.
`_____________________________
`
`Case No.: IPR2020-00136
`U.S. Patent No. RE 45,776
`______________________________
`
`PETITIONERS’ REPLY
`
`
`
`
`PROTECTIVE ORDER MATERIAL
`
`

`

`TABLE OF CONTENTS
`
`Page
`
`
`TABLE OF CONTENTS .......................................................................................... ii
`TABLE OF AUTHORITIES ................................................................................... iv
`I.
`Introduction ...................................................................................................... 1
`II.
`Claim Construction: Claim 36 Does Not Require Two Inclines. .................... 1
`III. Kontos’s Extension Catheter Is Not a Narrow Tube. ...................................... 2
`IV. The Asserted claims are Obvious. ................................................................... 4
`A.
`The Kontos-Ressemann combination teaches the single-incline and
`double-incline side openings of the ʼ776 patent. .................................. 4
`1.
`Replacing Kontos’s funnel with a side opening maximizes the
`usable area in the catheter assembly. .......................................... 5
`Petitioner’s other motivations are not based in hindsight. .......... 9
`Replacing Kontos’s funnel with a side opening would not create
`a problematic gap. .....................................................................11
`A POSITA had a reasonable expectation of success when
`replacing Kontos’s funnel with Ressemann’s collar. This
`combination creates a two-incline side opening. ......................13
`Even if a POSITA placed Ressemann’s collar underneath the
`pushrod, the Kontos-Ressemann combination still teaches the
`single-incline and double-incline openings. .............................15
`Claim 49: Kontos-Ressemann provides backup support to assist in
`resisting axial and shear forces............................................................18
`Claims 30-31, 53-56: Kontos-Ressemann-Takahashi combination
`teaches the not-more-than-one-French limitation. ..............................20
`The Kontos-Ressemann-Kataishi combination teaches a double-
`inclined side opening. ..........................................................................21
`Teleflex Asserts Secondary Considerations Based Upon Something it Did
`Not Invent—a Rapid Exchange Guide Extension Catheter. .........................25
`A.
`The combination of mother-in-child and Rx was well known. ..........26
`B.
`Side openings existed on prior art devices. .........................................27
`C.
`PO’s copying arguments fail. ..............................................................28
`
`V.
`
`ii
`
`2.
`3.
`
`4.
`
`5.
`
`B.
`
`C.
`
`D.
`
`

`

`VI. CONCLUSION ..............................................................................................32
`VI.
`CONCLUSION .............................................................................................. 32
`
`
`
`
`
`
`iii
`iii
`
`

`

`TABLE OF AUTHORITIES
`
` Page(s)
`
`Cases
`
`Amazon.com, Inc. v. Barnesandnoble.com, Inc.,
`239 F.3d 1343 (Fed. Cir. 2001) .......................................................................... 29
`
`In re Fulton,
`391 F.3d 1195 (Fed. Cir. 2004) ............................................................................ 9
`
`In re Kao,
`639 F.3d 1057 (Fed. Cir. 2011) .......................................................................... 25
`
`In re Schreiber,
`128 F.3d 1473 (Fed. Cir. 1997) .......................................................................... 20
`
`Johns Hopkins Univ. v. Datascope Corp.,
`543 F.3d 1342 (Fed. Cir. 2008) .......................................................................... 31
`
`Mytee Prods., Inc. v. Harris Research, Inc.,
`439 F. App’x 882 (Fed Cir. 2011) ...................................................................... 18
`
`Sakraida v. Ag Pro, Inc.,
`425 U.S. 273 (1976) ............................................................................................ 25
`
`ZUP, LLC v. Nash Mfg., Inc.,
`896 F.3d 1365 (Fed. Cir. 2018) .......................................................................... 26
`
`
`
`iv
`
`

`

`I.
`
`INTRODUCTION
`
`Patent Owner (“PO”) does not dispute, because it cannot, that Kontos
`
`describes its “support catheter” as a “mini guide catheter.” Ex-1409, 3:40-49. Nor
`
`does PO dispute that Kontos teaches, just like the coaxial guide catheter 12 of the
`
`Teleflex patent, that support catheter 10 includes a short distal lumen (body 12)
`
`coupled to a pushrod (wire 14). Other than the various side openings, Kontos
`
`teaches each structural limitation of the Challenged Claims. But as explained
`
`herein, the use of the claimed side openings was an obvious modification. The
`
`Challenged Claims are invalid as obvious.
`
`II. CLAIM CONSTRUCTION: CLAIM 36 DOES NOT REQUIRE TWO
`INCLINES.
`
`Claim 36 recites that “the segment defining the angled proximal end of the
`
`partially cylindrical opening includes at least one inclined region that tapers into a
`
`non-inclined region.” Ex-1401, 14:22-25. PO did not propose a construction for
`
`this claim,1 and just assumes that claim 36 requires a two-incline side opening.
`
`Paper 40 (“POR”), 42.
`
`Claim 36 only requires a “segment” defining a partially cylindrical opening
`
`with an angled proximal end that tapers into a non-inclined region. As
`
`
`1 In co-pending IPR2020-00135, PO proposes a construction of claim 36. Even if
`
`PO had proposed the same construction here, it would not be correct.
`
`1
`
`

`

`demonstrated by rigid portion 20 throughout the specification, this segment could
`
`extend proximally and distally to the actual opening it defines. The claim also does
`
`not dictate that the non-inclined region is part of the segment, stating only that the
`
`angled proximal end tapers into a non-inclined region.
`
`Finally, PO’s attempt to read Figure 4 into the claim is nonsensical. In
`
`Figure 4 the distal incline tapers into the non-inclined region within the opening.
`
`Ex-1401, Fig. 4. The proximal angle claimed here tapers into the pushrod. Id.,
`
`7:10.
`
`
`
`IPR2020-00135, Paper 44, 10-11. If Figure 4 is the basis for this claim, the non-
`
`inclined region is the pushrod, which is not part of the opening.
`
`III. KONTOS’S EXTENSION CATHETER IS NOT A NARROW TUBE.
`
`PO argues that Kontos’s body 12 must be a “narrow tube” that requires “a
`
`snug fit around [the] small-diameter PTCA catheter.” POR, 10. PO is wrong. Ex-
`
`1806, ¶¶142-50; Ex-1807, ¶¶159-63. Kontos never states that it is “important” for
`
`2
`
`

`

`body 12 to be “narrow” or have a “snug fit.” Ex-1806, ¶¶144-45; Ex-1801, 52:5-8
`
`(language is silent concerning “snugness”); Ex-1800, 77:6-11, 78:12-20. In fact,
`
`Kontos teaches the opposite, explaining that the size of body 12 should be suitable
`
`“for existing PTCA catheters” and that “[o]f course, other sizes may be used for
`
`other applications.” Ex-1409, 4:46-48, 4:61-5:2.
`
`PO’s argument is also contradicted by how Kontos actually prevents kinking
`
`of the PTCA catheter. Ex-1806, ¶147; Ex-1807, ¶¶14-27, 160-163. Kontos
`
`prevents kinking by reducing the distance between the distal-most portion of the
`
`catheter assembly and the occlusion. Ex-1806, ¶147; Ex-1807, ¶¶20-21, 163.
`
`“[B]ody 12 functions as a guide catheter extension,” by reducing “the gap that
`
`PTCA catheter 40 must negotiate without assistance …, [which] lessens
`
`considerably the tendency of the PTCA catheter 40 to bend, buckle or kink.” Ex-
`
`1409, 5:49-56. This is because, by comparison to advancement within the
`
`vasculature, significantly less force is needed to advance the interventional
`
`cardiology device (“IVCD”) within body 12 due to its lubricious inner coating. Ex-
`
`1806, ¶147; Ex-1807, ¶¶20-21, 163; Ex-1813, 52:14-53:8, 72:24-73:14. It is
`
`Kontos’s ability to shorten the distance that the IVCD traverses in the
`
`vasculature—especially in regions that are not straight or heavily calcified—and
`
`not the spatial relationship with the IVCD that reduces the likelihood of kinking.
`
`Ex-1806, ¶147; Ex-1807, ¶¶20-21, 163.
`
`3
`
`

`

`This conclusion is buttressed by the real-world experience of PO’s expert.
`
`Dr. Graham testified that he has never had an IVCD kink within a catheter
`
`assembly. Ex-1813, 52:14-53:8. In other words, the closeness of support assembly
`
`10 and the PTCA catheter 40 is irrelevant, as kinking (if it does occur) would not
`
`happen until the IVCD is in the vasculature. Ex-1806, ¶148.
`
`Not only is a close-fitting tube unnecessary to prevent kinking of the PTCA
`
`catheter, but, as Dr. Brecker explains, it would be detrimental. Ex-1806, ¶146. If
`
`the relationship is too “snug,” the IVCD can be damaged when loaded into the
`
`extension catheter, especially post-expansion, when the PTCA balloon is retracted
`
`proximally into the extension catheter. Id.
`
`Finally, even if Kontos taught a small body 12 with a “snug fit” to the
`
`IVCD, PO’s argument still fails because it is premised on Kontos’s invention being
`
`limited to fixed-wire catheters. Ex-1800, 76:2-8, 80:12-81:17; Ex-1801, 56:1-10,
`
`60:16-21. Kontos is not so limited, and instead provides that “the device of this
`
`present invention may be used with almost any type of catheter, including over-
`
`the-wire catheters.” Ex-1409, 9:47-50; Ex-1806, ¶¶149-50.
`
`IV. THE ASSERTED CLAIMS ARE OBVIOUS.
`
`A. The Kontos-Ressemann combination teaches the single-incline
`and double-incline side openings of the ʼ776 patent.
`
`A POSITA was motivated, with a reasonable expectation of success, to
`
`replace Kontos’s proximal funnel with Ressemann’s support collar 2141
`
`4
`
`

`

`(“Ressemann’s collar”). Pet., 25-34. At Institution, the Board agreed, explaining
`
`that “Petitioner ha[d] demonstrated a reasonable likelihood of prevailing” on this
`
`combination. Paper 20 (“I.D.”), 21. Replacing Kontos’s proximal funnel with
`
`Ressemann’s support collar 2141 results in a proximal opening with two inclines.
`
`Pet., 55-56.
`
`PO’s entire opposition is based on the mistaken belief that after adding
`
`Ressemann’s collar, a POSITA would make no further modification to Kontos’s
`
`support assembly 10. Even if a POSITA made no further modifications, though,
`
`PO’s argument still fails.
`
`1.
`
`Replacing Kontos’s funnel with a side opening maximizes
`the usable area in the catheter assembly.
`
`Replacing Kontos’s proximal funnel with a side opening maximizes the
`
`usable real estate within the catheter assembly. Ex-1806, ¶169. This is particularly
`
`important given that catheter assemblies used in PCI procedures are “not much
`
`bigger than a toothpick.” Ex-2137, 199:11-200:16. PO’s expert agrees that
`
`“maximizing the usable real estate inside a guide catheter is extremely important.”
`
`Ex-2145, ¶81; Ex-1801, 102:2-8; Ex-1813, 91:18-92:5. Specifically, PO’s expert
`
`testified that it was “important” to “maximize the inner diameter of the extension
`
`catheter without having to increase the outer diameter of the guide catheter.”
`
`Ex-1801, 102:2-8; Ex-2138, ¶161 (IPR2020-00129). And PO’s expert agreed that
`
`the use of a proximal funnel does not maximize the useable “real estate” within the
`
`5
`
`

`

`catheter assembly. Ex-1813, 92:19-24. Further, inventor Root similarly testified
`
`that he “
`
`
`
`” of the device. Ex-1115, 296:23-297:1.
`
`a. Maximizing usable real estate: reducing diameter of GC.
`
`Replacing Kontos’s funnel with a side opening, reduces the outer diameter
`
`of the extension catheter, which permits the use of a smaller-diameter GC.
`
`
`
`Ex-1405, ¶163. PO does not disagree, but instead argues that a POSITA would not
`
`replace the funnel with a side opening because Kontos was already sized to fit
`
`inside a 6 French GC and that a 5 French GC offered no advantage. POR, 25-26.
`
`Kontos has a 0.065-inch outer diameter at base portion 18 (Ex-1409, Fig. 3),
`
`meaning that to fit inside a 6 French GC having a 0.070-inch inner diameter,
`
`Kontos’s funnel would only have a 0.005 inch profile. Ex-1807, ¶¶164-166. This
`
`would provide no—or significantly limit the—funneling function. Id., ¶167.
`
`Further, even at 0.005 inches, the funnel would “rub” in a 6 French GC, which, as
`
`6
`
`

`

`admitted by PO’s expert, would “hinder the ability to facilitate smooth passage.”
`
`Ex-1801, 115:14-117:7. Thus, replacing Kontos’s funnel with a side opening
`
`would improve the trackability within the GC. Ex-1806, ¶170. Further, a POSITA
`
`would still be motivated to use a side opening because it would more easily permit
`
`compatibility with a 5 French GC, which has advantages over a 6 French GC.2 Id.,
`
`¶171.
`
`b. Maximizing usable real estate: increasing diameter of
`extension catheter.
`
`The corollary is also true: a POSITA would be motivated to replace
`
`Kontos’s funnel with a side opening because the inner diameter of body 12 could
`
`then be increased without similarly increasing the diameter of the GC. Ex-1806,
`
`¶172.
`
`
`2 For compatibility with a 5 French GC, tube 16 would be molded onto tapered
`
`wire 14. § IV.C, infra.
`
`
`
`7
`
`

`

`Ex-1405, ¶165. A POSITA was motivated to make this modification, as PO’s
`
`expert acknowledged, because “the larger the available area within a guiding
`
`device, the larger the variety of types, numbers and sizes of therapeutic devices
`
`that can be delivered.” Ex-2145, ¶129. Doing so, “ultimately improves the
`
`available treatment options for the patient.” Id.; Ex-1813, 132:22-133:10.
`
`Despite its own expert agreeing to the advantages of increasing the diameter
`
`of the extension catheter without causing a commensurate increase in the GC, PO
`
`argues that a POSITA would not apply this teaching because Kontos requires a
`
`“snug” fit. POR, 23. But for the reasons discussed above, Kontos does not mandate
`
`such a fit. § III, supra.
`
`PO also argues that it would not be possible to increase the diameter of body
`
`12, without also increasing the diameter of the GC, due to Kontos’s raised marker
`
`bands. POR, 23-24. Kontos teaches, however, that “[o]f course, numerous other
`
`methods for disposing marker band 30 at distal end 24 will be readily apparent to
`
`those skilled in the art.” Ex-1409, 4:21-24. Embedded marker bands were routine
`
`in the art. Ex-1806, ¶173; Ex-1762, 56:9-23. PO’s expert agreed. Ex-1801, 66:19-
`
`67:6, 70:5-17, 75:15-19. It was well within the skill of a POSITA to recess
`
`Kontos’s marker bands. Ex-1807, ¶¶168-82. PO cannot defeat Petitioners showing
`
`that a POSITA would replace Kontos’s funnel with Ressemann’s collar.
`
`8
`
`

`

`2.
`
`Petitioner’s other motivations are not based in hindsight.
`
`A POSITA was also motivated to replace Kontos’s funnel with Ressemann’s
`
`collar, as it would increase the area of entry—by more than five-fold—into the side
`
`opening. Ex-1807, ¶187. PO’s expert agreed:
`
`Even though [Kontos is] funneled, the opening is perpendicular …. And
`
`I've have agreed on record that an angled opening gives you a greater
`
`surface area than a perpendicular one …. With the -- the combination,
`
`using the -- the angled opening at 2141 tab from Ressemann compared
`
`to what Kontos has originally described … would increase the opening
`
`area into … Kontos. It would increase the area.
`
`Ex-1801, 120:16-121:13.
`
`Additionally, replacing Kontos’s funnel with Ressemann’s collar would, as
`
`noted in the Institution Decision, “reinforce the proximal opening … in the
`
`presence of deforming forces.” I.D., 21 (citing Ex-1408, 24:49-53). This would
`
`have been advantageous given that, according to PO, Kontos’s tube is flexible.
`
`IPR2020-00127, Paper 41, 24-25. PO argues, though, that a POSITA would not
`
`use Ressemann’s collar because it was “simple[r]” to “reinforce Kontos’s existing
`
`opening with a stiffening material (such as metal).” POR, 29 n.5. PO’s argument
`
`misses the mark, however, as the Federal Circuit “does not require that a particular
`
`combination must be the preferred, or the most desirable, combination described in
`
`the prior art in order to provide the motivation for the current invention.” In re
`
`Fulton, 391 F.3d 1195, 1200 (Fed. Cir. 2004).
`
`9
`
`

`

`Further, as noted in the Institution Decision, “Ressemann teaches that ‘[t]he
`
`proximal end 2140a of the evacuation lumen 2140 is preferably angled to facilitate
`
`smooth passage of other therapeutic devices through the evacuation lumen 2140 of
`
`the evacuation head 2132.’” I.D., 21 (citing Ex-1408, 23:17–20). Despite this clear
`
`teaching, PO argues that side openings actually cause device “catching or hang
`
`up.” POR, 19-20. This argument, though, is contradicted by a patent (“Keith”)
`
`issued to PO’s expert, which contains Ressemann’s verbatim disclosure of side
`
`openings “facilitate[ing] smoother passage” of IVCDs. Ex-1123, 7:54-60;
`
`Ex-1800, 149:3-10. PO cannot maintain its argument that a side opening causes
`
`device hang-up, given its own expert obtained a patent reciting the benefits of side
`
`openings.
`
`Side openings also allow for smoother passage of the catheter assembly as it
`
`navigates the vasculature. Pet., 32. PO argues, though, that this teaching is
`
`“unsupported by the evidence.” POR, 26. PO is wrong. Not only does Ressemann
`
`explicitly state that a side opening “allow[s] for smoother passage,” but an
`
`identical teaching is also recited in Keith. Ex-1523, 7:54-60. Further, in prior
`
`litigation, PO’s expert, Mr. Keith, argued that “
`
`
`
` Ex-1819, ¶112.
`
`Finally, a side opening permits smooth re-entry when the extension catheter
`
`is extended beyond the distal end of the GC. Pet., 45-46. PO’s expert previously
`
`10
`
`

`

`found himself in that situation (i.e., passed an extension catheter entirely outside of
`
`the GC),3 Ex-1801, 72:10-73:18, and agreed that a proximal opening, as opposed to
`
`Kontos’s funnel, permits smoother re-entry of the extension catheter. Ex-1801,
`
`79:14-21; Ex-1800, 151:17-152:1.
`
`3.
`
`Replacing Kontos’s funnel with a side opening would not
`create a problematic gap.
`
`Despite the clear teachings for why a POSITA would replace Kontos’s
`
`funnel with a side opening, PO argues that doing so would actually cause
`
`problems. POR, 17-22. First, PO’s argument fails because it is based on a false
`
`premise that a POSITA would not maximize the usable real estate in the catheter
`
`assembly. As explained above, a POSITA would have done so. § IV.A.1, supra.
`
`Even assuming a POSITA made no further modifications after replacing the funnel
`
`with a side opening, Kontos would not create a “problem gap” as alleged by PO.
`
`If Kontos was used with a 6 French GC, the gap between the support
`
`assembly’s outer wall and the inner wall of the GC would be 0.005 inches. Given
`
`that a guidewire or the distal-most wire of a fixed-wire balloon typically have a
`
`
`3 PO is incorrect to argue that “Kontos teaches that generally it is not desirable to
`
`extend the proximal end of body 12 beyond the distal end of the guide catheter.”
`
`POR, 27. PO ignores Figure 7, which teaches the intentional advancement into the
`
`vasculature. Ex-1409, 6:19-28, Fig. 7.
`
`11
`
`

`

`0.014 inch diameter, the PTCA catheter would not catch in the gap between tube
`
`16 and the inner wall of the GC. Ex-1806, ¶¶ 174-77.
`
`Ex-1409, Fig. 1 (modified by Petitioner).
`
`Ressemann and Keith buttress this conclusion. Ex-1806, ¶178. Both
`
`references teach side openings that are “preferably angled” to “facilitate smoother
`
`passage of other therapeutic devices,” and both teach a gap of at least 0.009 inches
`
`as shown below. Id., ¶179. This is almost double Kontos’s alleged “problematic”
`
`gap. Id.
`
`
`
`
`
`12
`
`

`

`Ex-1123, Fig. 1A (annotations added). For these additional reasons, a POSITA
`
`would replace Kontos’s funnel with Ressemann’s collar. Ex-1806, ¶179.
`
`4.
`
`A POSITA had a reasonable expectation of success when
`replacing Kontos’s funnel with Ressemann’s collar. This
`combination creates a two-incline side opening.
`
`PO also argues no reasonable expectation of success to integrate
`
`Ressemann’s collar with Kontos. POR, 41-42. Specifically, PO argues that the
`
`proposed modification would (i) create a catch point and (ii) not be
`
`“straightforward.” Id. Both arguments fail.
`
`In support of its argument that the proposed modification would create a
`
`catch-point, PO presents the following schematic.
`
`Id., 42. It was standard practice in the art—including, as explicitly taught by
`
`Ressemann (Ex-1408, 7:28-39, 24:2-11)—to encase Ressemann’s collar in a
`
`polymer jacket to remove any catch points. Ex-1807, ¶¶188-190, 90-92. In so
`
`doing, as shown in the below schematic, the proposed modification would
`
`
`
`13
`
`

`

`eliminate any concern about creating a problematic catch-point.
`
`
`
`Ex-1807, ¶¶188-190.
`
`PO is also incorrect to argue that while traversing the vasculature, forces
`
`would have “tend[ed] to push the tab off of the wire.” POR, 41. Indeed, numerous
`
`examples, including those cited on the face of the Teleflex patent (see, e.g., Ex-
`
`1463) disclose catheter assemblies with portions having convex up (distal lumen)
`
`and convex down components (pushrod). Ex-1807, ¶¶191, 93-96. Further,
`
`Ressemann’s collar could have been attached to Kontos’s wire 14, as taught by
`
`Ressemann, with a thin layer of solvent casting. Ex-1807, ¶¶90, 91, 101, 131, 190;
`
`Ex-1408, 24:2-11. Doing so would have maintained the proximal incline of
`
`Ressemann’s collar. Ex-1807, ¶190. Thus, PO is incorrect to argue no expectation
`
`of success. Accordingly, the Kontos-Ressemann combination teaches both the
`
`single-incline and two-incline proximal openings recited in the ʼ776 patent. Ex-
`
`14
`
`

`

`1806, ¶194
`
`
`
`Ex-1009, Fig. 1 (modified by Petitioner).
`
`5.
`
`Even if a POSITA placed Ressemann’s collar underneath
`the pushrod, the Kontos-Ressemann combination still
`teaches the single-incline and double-incline openings.
`
`PO states that GuideLiner version 1 has three inclines, which means that
`
`Ressemann must also have at least three inclines. Ex-1806, ¶196. Therefore, even
`
`if Ressemann’s collar was placed underneath Kontos’s pushrod—as alleged by PO
`
`(POR, 39)—the Kontos-Ressemann combination still teaches a two-inclined
`
`opening. Ex-1806, ¶195.
`
`PO’s expert testified that he identifies where one incline ends and another
`
`begins “just sort of by eye,” explaining that for version 1 of GuideLiner, as
`
`identified below, incline 1 has a “relatively steeper quality” to it, and incline 2 has
`
`“a shallower quality” to it, and “somewhere in between there” is incline 3.
`
`Ex-1800, 45:21-46:2. He believes that even shallow inclines are inclines so long as
`
`they can be identified relative to the catheter’s longitudinal axis. Id., 46:21-47:8.
`
`15
`
`

`

`Ex-1122 (color added to arrows for visibility). Similarly, Root testified that the
`
`figure below has two inclines separated by a flat section. Ex-1762, 91:24-94:1.
`
`
`
`16
`
`

`

`Ex-1116 (bottom annotation added to show inclined regions (red) and flat region
`
`
`
`(blue)).
`
`In light of PO’s testimony, Ressemann’s collar must have more than two
`
`inclines. As shown below, Ressemann’s collar has an incline at the proximal end
`
`(“A”), an incline corresponding to the curve between 2121a and 2141b (“C”), and
`
`an incline at the curved portion of tab 2141b (“B”). Ex-1806, ¶¶197-98.
`
`
`
`17
`
`

`

`Therefore, even if a POSITA placed the tab of Ressemann’s collar under
`
`wire 14, the resulting Kontos-Ressemann combination would still have two
`
`inclines. Ex-1806, ¶¶199-200 (identifying B and C).
`
`
`
`Ex-1807, ¶192.
`
`B. Claim 49: Kontos-Ressemann provides backup support to assist in
`resisting axial and shear forces.
`
`The Kontos-Ressemann combination provides back-up support in two ways:
`
`(i) shortening the distance that the IVCD must travel within the vasculature and (ii)
`
`increasing the moment of inertia of the catheter-in-catheter assembly. Ex-1806,
`
`¶¶159-67; Ex-1807, ¶¶14-27, 152-158, 193.4
`
`
`4 PO argues that “Petitioner bears the burden of showing that Kontos …
`
`inherently” satisfies claim 49. POR, 43. Petitioner need only show that Kontos-
`
`Ressemann teaches the structure of claim 49. Indeed, “after establishing a prima
`
`facie case of anticipation,” the burden of production shifts to the PO “to show that
`
`the prior art structure did not inherently possess the functionally defined limitations
`
`of the claimed apparatus.” Mytee Prods., Inc. v. Harris Research, Inc., 439 F.
`
`18
`
`

`

`Kontos’s support catheter 10 shortens the distance that the PTCA catheter
`
`must travel while in contact with the tortuous and potentially calcified vasculature.
`
`Ex-1806, ¶161; Ex-1807, ¶¶153, 155, 163. This type of catheter-in-catheter
`
`assembly that reduces the distance to the target location is a form of “deep
`
`seating.”5 Ex-1813, 75:4-17. And as admitted by PO’s expert, deep seating
`
`provides back-up support. Id., 76:4-6. More particularly, a physician must apply
`
`less force to advance an IVCD within the catheter assembly (by comparison to
`
`advancing within the vasculature), and thus, provides back-up support. § III, supra.
`
`Further, Kontos-Ressemann provides back-up support by making what is
`
`essentially a “double-catheter.” In so doing, the nested assembly increases the (i)
`
`flexural rigidity, (ii) torsional rigidity, and (iii) resistance to buckling forces of the
`
`catheter-in-catheter assembly. Ex-1807, ¶¶22-27. This in turn increases the
`
`moment of inertia, which results in increased back-up support. Ex-1806, ¶162; Ex-
`
`
`App’x 882, 886 (Fed Cir. 2011). Regardless, as set forth herein, Kontos-
`
`Ressemann inherently satisfies claim 49.
`
`5 PO’s expert opined that there are two types of deep seating: (i) extending the GC
`
`past the ostium and into the coronary artery or (ii) positioning the GC at the ostium
`
`and using a child catheter therein to extend into the coronary artery. Ex-1813,
`
`74:21-75:17; Ex-2145, ¶¶51, 78.
`
`19
`
`

`

`1807, ¶27.
`
`PO chose to describe its invention by what it does: provide back-up support.
`
`“[C]hoosing to define an element functionally, i.e., by what it does, carries with it a
`
`risk.” In re Schreiber, 128 F.3d 1473, 1478 (Fed. Cir. 1997). This is because “[i]t
`
`is well settled that the recitation of a new intended use for an old product does not
`
`make a claim to that old product patentable.” Id. at 1477. Indeed, as described in
`
`Schreiber, a functional element will “not have patentable weight if the structure is
`
`already known, regardless of whether it has ever been used in any way [for the
`
`recited function].” Id. Nevertheless, for the reasons set forth herein, even if the
`
`functional claim language is afforded patentable weight, Kontos-Ressemann
`
`teaches this limitation. Ex-1806, ¶¶163-67.
`
`C. Claims 30-31, 53-56: Kontos-Ressemann-Takahashi combination
`teaches the not-more-than-one-French limitation.
`
`PO does not dispute that Takahashi provides motivation to achieve a 1
`
`French differential to increase back-up support. Ex-1801, 117:14-118:5. PO argues
`
`that this motivation is inapplicable, though, because Kontos “teaches away from
`
`maximizing the inner diameter.” POR, 44. As discussed above, PO is incorrect to
`
`argue that Kontos requires a snug fitting body 12. § III, supra. Thus, a POSITA
`
`was motivated to apply the teachings of Takahashi to increase back-up support.
`
`Ex-1806, ¶¶189, 191.
`
`PO also argues that a POSITA would not have an expectation of success to
`
`20
`
`

`

`modify Kontos to achieve the 1 French differential. POR, 45-46. PO is wrong. A
`
`POSITA knew how to (i) replace the proximal funnel with a side opening and (ii)
`
`recess Kontos’s distal marker bands. § IV.A.1.b, supra. Further, Kontos explicitly
`
`teaches that “tube 16 may be molded directly only application wire 14.” Ex-1409,
`
`4:31-32. In so doing, to maximize the usable “real estate” in the catheter assembly,
`
`a POSITA would taper the pushrod for attachment onto the Kontos-Ressemann
`
`combination. Ex-1807, ¶¶183-195. These modifications allow the diameter of tube
`
`16 to be increased (Ex-1409, 4:46-48), and Kontos would achieve the not-more-
`
`than-one-French differential. Ex-1806, ¶¶190, 192.
`
`Id., Fig. 1 (modified by Petitioner).
`
`D. The Kontos-Ressemann-Kataishi combination teaches a double-
`inclined side opening.
`
`A POSITA was motivated to combine the shape of Kataishi’s distal opening
`
`with Ressemann’s proximal opening. Pet., 68-72. Specifically, adding Kataishi’s
`
`shape to Ressemann’s proximal opening would improve loading of interventional
`
`21
`
`

`

`devices and improve crossability of the catheter. Id. PO is incorrect to argue
`
`otherwise.
`
`PO argues that a POSITA would not recognize that the benefit of Kataishi’s
`
`distal opening—“superior loading of thrombus”—applies equally to Ressemann’s
`
`proximal opening for loading interventional cardiology devices. POR, 51. PO is
`
`incorrect, as the art itself and the knowledge of a POSITA in view of that art
`
`shows the benefits of Kataishi’s inclined shape applies to loading both devices
`
`and thrombi through the lumen. PO ignores, for instance, that Ressemann itself
`
`explicitly teaches that both “[t]he proximal and distal ends 140a, 140b of the
`
`evacuation lumen 140 are preferably angled,” and that this shape on either end
`
`“allows for larger deformable particulate matter to pass through the lumen more
`
`smoothly.” Ex-1408, 6:52-60.
`
`Id., Fig. 1A. The Keith patent also teaches that the “proximal and distal ends … are
`
`preferably angled … to facilitate smoother passage of other therapeutic devices.”
`
`Ex-1123, 7:54-60 (emphasis added); Ex-1800, 149:3-18; Ex-1408, Figs. 6E, 6G,
`
`
`
`22
`
`

`

`13:57-14:26. Thus, Kataishi’s teaching that the shape of its distal tip improves that
`
`catheter’s ability to suction thrombus would be understood by a POSITA to also
`
`apply to the proximal opening of a catheter such as Ressemann. Ex-1807, ¶¶196-
`
`199; Ex-1806, ¶¶ 201-04. The shape of Kataishi’s opening provides similar
`
`benefits to Ressemann’s opening, and when implemented in the Kontos-
`
`Ressemann combination, provides roughly the same opening area. Ex-1807,
`
`¶¶204-07.
`
`
`
`Id., ¶205 (illustrating the larger area of Kataishi’s shape versus Ressemann’s
`
`
`
`support collar under wire 14) And an increase in the area of opening makes it
`
`easier to introduce a stent or balloon catheter. Ex-1807, ¶¶196-99; Ex-1806,
`
`¶¶201-04.
`
`23
`
`

`

`
`
`PO also incorrectly alleges that Kataishi’s “crossability” benefits would not
`
`apply to Ressemann because it does not “see” the vasculature. POR, 50-51. As a
`
`threshold matter, Ressemann teaches that both the proximal and distal ends of
`
`evacuation lumen 140 assist in ensuring “smoother passage of the evacuation
`
`sheath assembly 100 through a guide catheter.” Ex-1408, 6:52-57 (emphasis
`
`added); Ex-1123, 7:54-60. A POSITA would recognize Kataishi’s distal shape
`
`would also assist devices as they move through the guide catheter. Indeed, as cited
`
`in the Petition, Sakurada performed a “passing ability test” of Kataishi that
`
`measured “[t]he length between the ostium and the distal tip of the aspiration
`
`catheter.” Ex-1455, 300-02. Sakurada found that Kataishi’s shape had improved
`
`crossability as the catheter advanced along various loops in the vasculature. Id. A
`
`POSITA would understand these benefits apply when navigating around loops in a
`
`guide catheter. Ex-1807, ¶198. Because Kataishi’s shape is being used in a
`
`predictive way to achieve a known and similar benefit, a POSITA appreciates that
`
`modifying Ressemann’s proximal opening to include the shape of Kataishi’s distal
`
`tip was an obvious modification determined through routine experimentation and
`
`optimization. Ex-1807, ¶¶200-206; Ex-1405, ¶¶243-50.
`
`PO further argues that Kataishi’s benefits—that allegedly flow from both its
`
`shape and flexibility of the material—is inapplicable to Ressemann’s collar. PO is
`
`again belied by its own expert witness, who testified that as of the filing date of the
`
`24
`
`

`

`’776 Patent, once a shape is provided, a POSITA would readily use routine
`
`engineering to create side openings of several different materials, some more rigid
`
`and some more flexible. Ex-1764, 27:9-17, 30:2-8, 31:16-21.
`
`V. TELEFLEX ASSERTS SECONDARY CONSIDERATIONS BASED
`UPON SOMETHING IT DID NOT INVENT—A RAPID EXCHANGE
`GUIDE EXTENSION CATHETER.
`
`The

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket