throbber
Paper No. __________
`
`
`
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`___________________________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`__________________________
`
`
`MEDTRONIC, INC., AND MEDTRONIC VASCULAR, INC.
`
`Petitioner,
`
`v.
`
`TELEFLEX INNOVATIONS S.À.R.L.,
`
`Patent Owner
`_____________________________
`Case No: IPR2020-00134
`U.S. Patent No. RE45,760
`______________________________
`
`PETITIONER’S REPLY
`
`

`

`TABLE OF CONTENTS
`
`Page
`TABLE OF CONTENTS ............................................................................................ i
`
`TABLE OF AUTHORITIES .................................................................................... iii
`
`I.
`
`INTRODUCTION ........................................................................................... 1
`
`II. GROUNDS 1-3: PO DOES NOT CHALLENGE ITOU’S
`DISCLOSURE, WHICH ANTICIPATES OR RENDERS OBVIOUS
`CLAIMS 48 AND 51-53. ................................................................................ 2
`
`III. GROUND 4: CLAIMS 48 AND 51-53 ARE RENDERED OBVIOUS
`BY RESSEMANN IN VIEW OF TAKAHASHI. ........................................... 2
`
`A.
`
`Ressemann discloses a “tubular structure defining a lumen
`coaxial and in fluid communication with the lumen of the guide
`catheter.” ................................................................................................ 2
`
`1.
`
`2.
`
`“Coaxial” means “the axis of the lumen of the guide extension
`catheter is aligned in the same direction as the axis of lumen of
`the guide catheter.” ...................................................................... 2
`
`Ressemann’s multi-lumen tube 138 defines evacuation lumen
`140, which is coaxial and in fluid communication with the
`lumen of guiding catheter 160. ................................................... 9
`
`B.
`
`C.
`
`Ressemann in view of Takahashi renders obvious claims 48 and
`51-53. ................................................................................................... 11
`
`PO does not separately challenge that claim 52 is obvious in
`view of Ressemann and the knowledge of a POSITA. ....................... 15
`
`IV.
`
`PO ASSERTS SECONDARY CONSIDERATIONS BASED UPON
`SOMETHING IT DID NOT INVENT—A RAPID-EXCHANGE
`VERSION OF A GUIDE EXTENSION CATHETER. ................................. 16
`
`A. Mother-in-child and Rx devices were well known, and so was
`the combination. .................................................................................. 17
`
`B.
`
`PO’s secondary considerations evidence all relates to prior art
`features and functionality. ................................................................... 18
`
`
`
`i
`
`

`

`V.
`V.
`
`CONCLUSION .............................................................................................. 23
`CONCLUSION .............................................................................................. 23
`
`
`
`
`
`
`
`ii
`ii
`
`

`

`TABLE OF AUTHORITIES
`
` Page(s)
`
`Cases
`
`Amazon.com, Inc. v. Barnesandnoble.com, Inc.,
`239 F.3d 1343 (Fed. Cir. 2001) .......................................................................... 18
`
`In re Kao,
`639 F.3d 1057 (Fed. Cir. 2011) .............................................................. 16, 19, 20
`
`Johns Hopkins Univ. v. Datascope Corp.,
`543 F.3d 1342 (Fed. Cir. 2008) .......................................................................... 20
`
`Ormco Corp. v. Align Tech., Inc.,
`463 F.3d 1299 (2006) .......................................................................................... 18
`
`Raytheon Co. v. Sony Corp.,
`727 F. App’x 662 (Fed. Cir. 2018) ..................................................................... 12
`
`Sakraida v. Ag Pro, Inc.,
`425 U.S. 273 (1976) ............................................................................................ 16
`
`W.M. Wrigley Jr. Co. v. Cadbury Adams USA LLC,
`683 F.3d 1356 (Fed. Cir. 2012) .......................................................................... 19
`
`ZUP, LLC v. Nash Mfg., Inc.,
`896 F.3d 1365 (Fed. Cir. 2018) .......................................................................... 17
`
`
`
`
`
`iii
`
`

`

`I.
`
`INTRODUCTION
`
`The evidence in the Petition establishes that Itou (Grounds 1-3) and
`
`Ressemann (Ground 4) anticipate or render obvious claims 48 and 51-53. PO does
`
`not dispute any of the evidence cited in Grounds 1-3, focusing its substantive
`
`response solely on Ground 4.
`
`PO also cannot dispute that Ressemann discloses a guide extension catheter
`
`(“GEC”) that “is passed through the lumen of a guide catheter, advanced beyond
`
`the distal end of the guide catheter, and inserted into a branch artery of the aorta,”
`
`or that Ressemann’s assembly 100 “facilitate[s] delivery of stents and balloon
`
`catheters.” Compare Paper 41 (“POR”), 4, with Ex-1608, Figs. 6A-6I. Instead, it
`
`focuses on two claimed features, each of which is unremarkable and well known in
`
`the art.
`
`First, PO argues that Ressemann does not disclose a guide extension catheter
`
`with a lumen “coaxial” to the guide catheter lumen, relying on an understanding
`
`that would require a perfect overlap in the axes. PO’s own witness testimony and
`
`documents, however, confirm the accuracy of Medtronic’s position that coaxial
`
`means “the axis of the lumen of the guide extension catheter is aligned in the same
`
`direction as the axis of lumen of the guide catheter.”
`
`Second, PO argues that a POSITA would not be motivated to modify
`
`Ressemann to achieve a “one French size” differential with the guide catheter. For
`
`
`
`1
`
`

`

`the reasons set forth in the Petition and herein, PO is wrong. Claims 48 and 51-53
`
`are obvious.
`
`II. GROUNDS 1-3: PO DOES NOT CHALLENGE ITOU’S
`DISCLOSURE, WHICH ANTICIPATES OR RENDERS OBVIOUS
`CLAIMS 48 AND 51-53.
`
`PO argues that Itou is not prior art. POR, 1. But PO has not established that
`
`its claims were conceived and reduced to practice prior to Itou. Paper 75.
`
`Accordingly, each claim addressed in Grounds 1-3 is anticipated or obvious based
`
`on the evidence Medtronic cites in its Petition.
`
`III. GROUND 4: CLAIMS 48 AND 51-53 ARE RENDERED OBVIOUS BY
`RESSEMANN IN VIEW OF TAKAHASHI.
`
`A. Ressemann discloses a “tubular structure defining a lumen
`coaxial and in fluid communication with the lumen of the guide
`catheter.”
`
`1.
`
`“Coaxial” means “the axis of the lumen of the guide
`extension catheter is aligned in the same direction as the
`axis of lumen of the guide catheter.”
`
`PO has twisted the meaning of coaxial throughout these proceedings. In the
`
`POR, PO took the position that coaxial means the lumens must be perfectly
`
`concentric, such that any deviation off axis by the extension catheter means the
`
`claim element is not met. POR, 11-13. But PO’s lead expert later admitted, twice,
`
`
`
`2
`
`

`

`that it may be off axis the whole time and still be coaxial.1 Ex-1764, 91:10-23,
`
`94:11-19. Clearly the term “coaxial” in this art does not mean “sharing the same
`
`axis” as it might in other arts.
`
`The claims recite a “coaxial” relationship between the lumens of the guide
`
`and guide extension catheters, where the former is configured to be placed in the
`
`coronary artery ostium, and the latter is configured to be extended through the
`
`guide catheter and into the coronary vasculature. Ex-1601, 13:36-14:7. This
`
`coaxial relationship allows for the claimed capability to receive one or more stents
`
`or balloon catheters, which does not require a concentric relationship. Id., 13:67-
`
`14:1.
`
`Similarly, the patent specification explains that a “coaxial arrangement” is
`
`one in which an extension catheter has been inserted into a guide catheter and the
`
`two are in line with the coronary artery.
`
`
`1 The Board also relied on Teleflex’s interpretation of coaxial, now withdrawn, to
`
`deny institution of another petition. IPR2020-00133, Paper 20, 14-15.
`
`
`
`3
`
`

`

`
`
`Figure 8, for example, discloses extension catheter (12) inserted through
`
`guide catheter (56), with each catheter in line with the ostium of the coronary
`
`artery (60). Ex-1601, 8:10-16, Fig. 8. The specification places no limitations on the
`
`size of guide catheter or the amount of radial displacement available to the coaxial
`
`guide catheter.
`
`The specification is consistent with the general use of the word “coaxial” in
`
`this particular art. As Dr. Brecker testified:
`
`From the very first point that I ever put a guide catheter into a coronary
`
`artery, there is a need for the guide catheter to be in line with the coronary
`
`artery, and cardiologists refer to that as being coaxial. They do not need,
`
`and in fact, almost never will share a center line.
`
`Ex-2116, 323:25-324:11; Ex-1806 ¶16; and see ¶¶17-26. The same is true of a
`
`“coaxial guide catheter.
`
`
`
`4
`
`

`

`This understanding is reflected in Patent Owner’s own documents, which
`
`consistently refer to a “coaxial” alignment in reference to a catheter-in-catheter
`
`arrangement that is in line with the coronary ostia.
`
`Ex-1812, 24183 (top); Ex-1815, 131005 (lower). In this art, the term clearly refers
`
`to directional alignment, not mathematical precision or concentricity, which is
`
`impossible.
`
`Long before the current dispute with Medtronic, inventor Root agreed,
`
`testifying that a coaxial guide catheter is “a tube-within-a-tube…for delivering
`
`interventional cardiology devices into the vasculature.” Ex-1715, 37:10-17; see
`
`
`
`5
`
`

`

`Ex-1714, 86:1-87:7
`
`)
`
`
`
`
`
`
`
` Ex-1714, 66:22-67:2. In this
`
`matter, Root initially testified that simply a showing of putting an extension
`
`catheter into a guide catheter is sufficient to establish that the two are coaxial.
`
`Ex-1762, 173:10-25. And while he attempted to backtrack adding the gloss that
`
`coaxial means a “shared” axis (contradicting his testimony from not one, but two
`
`prior depositions), he also admitted that “nothing is perfectly concentric,” and “if
`
`it’s a circle in a circle, that’s coaxial.” Id., 174:13-175:6, 176:11-19.
`
`The fact that “coaxial” cannot be construed to mean a mathematically
`
`perfect overlap is further supported by PO’s expert, Mr. Keith, who testified that
`
`“if you’ve got a tubular structure inside another tubular structure, that is a coaxial
`
`arrangement,” Ex-1764, 90:20-91:2, and agreed that “coaxial” does not require
`
`perfect concentricity at any point. Id., 91:3-9; see Ex-1805, 119:22-120:12
`
`(explaining that catheter engineers and designers know that catheter arrangements
`
`that are tubes inside tubes are…not magically automatically in line (sic) be their
`
`axis, but they are still referred to as coaxial structures and arrangements if they are
`
`relatively close). For example, in describing the catheter arrangement depicted
`
`
`
`6
`
`

`

`below, Keith explained that the catheters are not always in the same arrangement
`
`and are “free to migrate a little bit across the cross section,” but are still referred to
`
`as “coaxial.” Id., 120:13-121:1. Keith testified that testified that a 4-in-6 French,
`
`and a 5-in-7 French (illustrated below) are both coaxial. Ex-1800, 23:19-24:1;
`
`Ex-1805, 119:4-7, 119:12-121:1.
`
`
`
`Ultimately, Keith provided a definition that coaxial means, “essentially a
`
`tube-in-a-tube arrangement that is relatively close fitting.” Ex-1764, 98:18-21. But
`
`he could not say what relatively close fitting means. Id., 98:22-99:4. Based on both
`
`his testimony and that of PO’s cardiologist, Graham, it cannot require a 1 French
`
`differential. Ex-1801, 37:22-38:2. While this definition is an improvement over
`
`
`
`7
`
`

`

`PO’s original position, it still focuses on tubes, whereas the claims recite coaxial
`
`lumens.2
`
`Following Keith’s testimony, it is certain that the lumens may be off-axis,
`
`but unclear by how much. At a minimum any definition must cover the preferred
`
`embodiment of a 5-in-6 French catheter, as well as the admitted coaxial nature of a
`
`4-in-6 French arrangement. These arrangements are shown schematically below,
`
`with 0.005 and 0.018 inches of annular gap space respectively.
`
`
`
`
`
`
`
`
`2 On redirect, Keith construed his construction, stating the tubes must have “radial
`
`symmetry.” Ex-1764, 101:10-24. This new theory has nothing to do with defining
`
`coaxial lumens.
`
`
`
`8
`
`

`

`Ex-1601, 3:42-44 (teaching a 6 French catheter with an inner diameter (“ID”) of
`
`0.070 inches); Ex-1806 ¶23 (explaining that 5 and 4 French catheters have ODs of
`
`0.065 and 0.052 inches respectively).
`
`These examples could be taken to extremes like a 4-in-8 French arrangement
`
`and there is nothing in the specification or elsewhere that suggests that is no longer
`
`coaxial. As used in the Teleflex patents, and the art in general, coaxial simply refers
`
`to the alignment of the lumens. Thus, coaxial means “the axis of the lumen of the
`
`guide extension catheter is aligned in the same direction as the axis of lumen of the
`
`guide catheter.” Any attempt to require further precision would render the claim
`
`indefinite.
`
`2.
`
`Ressemann’s multi-lumen tube 138 defines evacuation
`lumen 140, which is coaxial and in fluid communication
`with the lumen of guiding catheter 160.
`
`As set forth in the Petition, Ressemann meets this claim limitation. Pet.,
`
`61-63. Lumen 140 is defined by tube 138, which is part of assembly 100’s
`
`evacuation head 132. Ex-1608, 6:35-47.
`
`
`
`
`
`
`
`
`
`9
`
`

`

`Id., Figs. 1A-B.
`
`Ressemann teaches that assembly 100 is advanced through guiding catheter
`
`160 until the evacuation head’s distal end is within the coronary vessel, while its
`
`proximal end remains in catheter 160. Id., 12:9-30.
`
`
`
`
`
`Id., Fig. 6B (annotations added).
`
`As shown in Fig. 6B, the axis of lumen 140 within evacuation head 132
`
`(red) is necessarily aligned in the same direction as the axis of the lumen of the
`
`guiding catheter 160 (blue). Ex-1806 ¶101; see Ex-2116, 320:20-321:3 (Dr.
`
`Brecker testifying that he has used coaxial dual lumen catheters).
`
`Moreover, Ressemann teaches an 8 French guide catheter has an ID of 0.090
`
`inches, and that the outer diameter (“OD”) of the tubular structure within which
`
`lumen 140 sits is 0.076 inches, which leaves an annular gap of 0.014 inches. Ex-
`
`
`
`10
`
`

`

`1608, 10:14-29. As illustrated below, this results in a closer fitting within
`
`Ressemann than the admitted co-axial relationship between a 4-in-6 French
`
`catheter.
`
`
`
`
`
`
`
`Ex-1806 ¶¶103-104, and see generally ¶¶100-105.
`
`Lumen 140 is therefore “coaxial” to the lumen of guiding catheter 160, and
`
`is in fluid communication with the same. Pet., 61-63; Ex-1608, 9:30-36.
`
`B. Ressemann in view of Takahashi renders obvious claims 48 and
`51-53.
`
`PO does not dispute the advantages of removing Ressemann’s sealing
`
`balloons in order to achieve the not-more-than-one-French differential. Compare
`
`Pet., 63-67 with POR 18-21. Instead, PO argues that a POSITA would not be
`
`motivated to remove Ressemann’s balloons because that modification would
`
`
`
`11
`
`

`

`“eliminate Ressemann’s primary function of embolic protection.” POR, 13. PO is
`
`wrong.
`
`That the proposed modification impacts Ressemann’s embolic protection
`
`function is not determinative of the obviousness inquiry. A “skilled artisan cannot
`
`be assumed to ignore” Ressemann “merely because it is primarily directed to a
`
`specific application that is different from the application” of PO’s patent, as “a
`
`reference must be considered for everything that it teaches, not simply the
`
`described invention or a preferred embodiment.” Raytheon Co. v. Sony Corp., 727
`
`F. App’x 662, 667 (Fed. Cir. 2018). In Raytheon, for example, a prior art reference
`
`taught “a multi-layer CCD for detecting electromagnetic radiation.” Id. at 666.
`
`Petitioner modified the reference to include silicon, even though the reference
`
`explicitly taught that silicon was not suitable for certain disclosed applications.
`
`Because silicon was suitable for other disclosed applications, the Board’s
`
`obviousness determination was upheld. Id. at 667-68.
`
`Here, PO’s experts do not dispute that Ressemann teaches delivery of a stent
`
`through evacuation lumen 140. Ex-1800, 33:2-16, 34:22-25, 35:16-21; Ex-1801,
`
`33:17-22. Nor is there any dispute that Petitioner’s proposed modifications would
`
`retain Ressemann’s ability to deliver a stent or an angioplasty balloon, as well as
`
`allow assembly 100 to be used with a smaller guide catheter. Pet., 63-67; POR 13-
`
`21; Ex-1807 ¶¶112-117.
`
`
`
`12
`
`

`

`Moreover, PO is mistaken that the modification eliminates embolic
`
`protection. As described by PO’s expert, embolic protection is “something that
`
`prevents debris from flowing downstream.” Ex-1813, 58:15-18. And as also
`
`admitted by PO’s expert, (i) it is possible to “provide embolic protection without
`
`aspiration” and (ii) “there are ways to provide embolic protection without the use
`
`of balloons.” Id., 162:5-11. Before the alleged invention, a POSITA was aware of
`
`non-occlusive embolic protection devices that did not rely upon aspiration of
`
`emboli. Ex-1811, 1443. These devices were often preferable as some patients were
`
`not able to tolerate occlusion. Modified as Petitioner proposes, Ressemann’s
`
`assembly 100 would not only still be suitable for stent or angioplasty balloon
`
`delivery, but also for delivery of filter-based embolic protection systems. Ex-1806
`
`¶¶107-110; Ex-1807 ¶119.
`
`PO’s remaining arguments are that modification of Ressemann to remove
`
`the sealing balloons might somehow alter the ID of evacuation lumen 140, and that
`
`the modifications do not account for “‘stiffness transition member 135’ that sits in
`
`a separate lumen location inside the evacuation lumen 140 and that extends all the
`
`way to the tip of the evacuation sheath.” POR, 19. Neither argument has any
`
`relevance to the articulated reasons to modify Ressemann, and neither is correct.
`
`First, Ressemann teaches that the sealing balloons 134 and 136 are external
`
`to evacuation lumen 140 (yellow, below). Ex-1608, 8:29-40, Figs. 1A-B
`
`
`
`13
`
`

`

`(explaining that the balloons are “secured to an exterior of the multi-lumen tune
`
`138” at their waist portions, including 134b (circled in purple).
`
`
`
`
`
`There is no expectation that removing the balloons from the outside of tube
`
`138 would, in and of itself, impact the size of the interior lumens 140 and 142. Ex-
`
`1807 ¶118; Ex-1806 ¶¶111.
`
`A POSITA, however, had motivation to modify Ressemann so that it could
`
`be used with a smaller guide catheter, for example a 7 French catheter, by
`
`removing the sealing balloons, Pet., 63-67, as well as to increase the size of
`
`evacuation lumen 140. As Dr. Graham admitted, in the 2005-2006 time frame the
`
`concept of “maximizing the usable real estate” was important, and a POSITA
`
`would have looked to have the largest possible ID of an inner catheter without
`
`having to increase the OD of the outer catheter. Ex-1813, 91:18-92:5. While
`
`Ressemann teaches that lumen 140 is “approximately 0.061 inches,” increasing the
`
`lumen size to 0.065 inches had the benefit of increasing the number of stent
`
`catheters that could be delivered through lumen 140. Ex-1806 ¶¶112-114; Ex-1807
`
`
`
`14
`
`

`

`¶116; Ex-1623, 2 (teaching a stent delivery system with a minimum guide catheter
`
`diameter of 0.064 inches); Ex-1802, 131-32 (teaching a stent with a crossing
`
`profile of 0.064 inches). Given that the Teleflex patent teaches that a 7 Fr guide
`
`catheter has a lumen of 0.078 inches (Ex-1601, 3:41-42), the modifications
`
`proposed would result in a differential lumen size of 0.013 inches, which is 1/3 of a
`
`millimeter, or 1 French. Ex-1806 ¶115. This 1 French differential is also explicitly
`
`taught in Takahashi. Pet., 66.
`
`Second, Ressemann is explicit that “stiffness transition member 135” is not
`
`within evacuation lumen 140, but within inflation lumen 142, which is a separate
`
`and distinct structure within multi-lumen tube 138. Ex-1608, 11:29-38, Figs. 1A-B.
`
`It has no bearing on the size of lumen 140.
`
`Claims 48, 51 and 53 are obvious.
`
`C.
`
`PO does not separately challenge that claim 52 is obvious in view
`of Ressemann and the knowledge of a POSITA.
`
`Claim 52 depends from claim 1, adding the limitation that the guide
`
`extension catheter include “a longitudinal length of the reinforced braid or coil”
`
`that is “between 20-30 cm.” Aside from its challenges to the Ressemann disclosure
`
`discussed herein, PO has not challenged that Ressemann in view of the common
`
`knowledge of a POSITA renders claim 52 obvious. POR, 21. Claim 52 is obvious.
`
`
`
`15
`
`

`

`IV. PO ASSERTS SECONDARY CONSIDERATIONS BASED UPON
`SOMETHING IT DID NOT INVENT—A RAPID-EXCHANGE
`VERSION OF A GUIDE EXTENSION CATHETER.
`
`The earliest Rx GEC in the record is Kontos, followed by Ressemann and
`
`Itou. These devices are Rx, configured to deliver a wide variety of IVCDs, and
`
`provide increased back up support when extended partially past the end of a GC as
`
`intended. Ressemann and Itou also have a side opening. These devices and device
`
`features—and the functionality associated with them—were all published in the art
`
`in advance of the 2004 Transcatheter Cardiovascular Therapeutics conference,
`
`where inventor Root maintains he conceived of the idea claimed in the Teleflex
`
`patents. Compare Ex-1755 ¶¶28-38 with Ex-2118 ¶5.
`
`Nevertheless, PO bases its entire secondary considerations case on these
`
`features alone. POR, 43-45. PO makes the fundamental error of confusing
`
`commercialization with invention. Sakraida v. Ag Pro, Inc., 425 U.S. 273, 278
`
`(1976) (secondary considerations, “without invention, will not make
`
`patentability”). It does not matter if PO was first to market when the combination
`
`of features it relies upon were in the prior art. In re Kao, 639 F.3d 1057, 1068−69
`
`(Fed. Cir. 2011) (where “the offered secondary consideration[s] actually results
`
`from something other than what is both claimed and novel in the claim there is no
`
`nexus to the merits of the claimed invention.”).
`
`
`
`16
`
`

`

`In this IPR, PO calls out claims 48, 51, 53 which recite a side opening
`
`adapted to receive IVCDs. Itou has that, and the only potential obviousness issue is
`
`using Ressemann’s teaching to actually do it. But PO does not challenge
`
`Petitioner’s Itou-based grounds. The only obviousness issues raised by PO are
`
`whether (1) Ressemann discloses a GEC with a lumen “coaxial” to the guide
`
`catheter lumen, and (2) a POSITA would not be motivated to modify Ressemann to
`
`achieve a “one French size” differential with the GC. On this record, PO cannot
`
`overcome Petitioner’s strong showing of obviousness. ZUP, LLC v. Nash Mfg.,
`
`Inc., 896 F.3d 1365, 1374 (Fed. Cir. 2018) (“a strong showing of obviousness may
`
`stand even in the face of considerable evidence of secondary considerations”).
`
`A. Mother-in-child and Rx devices were well known, and so was the
`combination.
`
`Rx technology, as opposed to OTW, was developed in the 1980s for balloon
`
`catheters. Ex-2138 ¶71 (“rapid exchange functionality existed for almost two
`
`decades before GuideLiner.”); Ex-2145 ¶82; Ex-1762, 39; Ex-1714, 111:13-17.
`
`And there was a trend in the industry toward full adoption through the 1990s and
`
`early 2000s. Ex-1800, 26:10-27:8; Ex-1817, 25:20-26:9, 37:8-38:6. Full-length
`
`OTW mother-in-child was known to provide guide extension and additional
`
`backup support. Ex-1800, 16:22-17:1; Ex-1762, 194:4-16; Ex-1817, 25:20-26:9;
`
`Ex-1762, 36:3-10; Ex-2123 ¶20. As stated by inventor Root in his earliest notes
`
`and throughout his testimony, the alleged invention was to make an Rx version of a
`
`
`
`17
`
`

`

`mother-in-child catheter. Ex-1762, 39:19-21, 67:23-68:1; Ex-2003; Ex-1714 at
`
`61:24-62:22, 77:9-15, 121:12-122:1. Dr. Thompson concedes that the only
`
`structure you need to enjoy the benefits of modern GEC catheters is Rx, a side
`
`opening, and a lumen relatively close to the size of the GC. Ex-1817, 63:10-64:8;
`
`Ex-2215 ¶22. Dr. Azzalini concedes that all of the drawbacks he lists of mother-in-
`
`child are alleviated by the change to Rx, and all of the relevant benefits he cites for
`
`GECs flow from the change from mother-in-child to Rx. Ex-1762, 23:15-26:22.
`
`But once again, Kontos, Itou and Ressemann are Rx, close in size to the GC,
`
`provide back support, and the latter two have side openings. Supra § III.
`
`B.
`
`PO’s secondary considerations evidence all relates to prior art
`features and functionality.
`
`“If the feature that creates the commercial success was known in the prior
`
`art, the success is not pertinent.” Ormco Corp. v. Align Tech., Inc., 463 F.3d 1299,
`
`1312 (2006) PO’s cited marketing documents prominently emphasize prior art
`
`functionality of “rapid exchange convenience” and “added backup support.” POR,
`
`46-48 (highlighting Ex-2155, Ex-2158, Ex-2161, Ex-2164).
`
`PO’s alleged copying evidence likewise focuses on prior art features and
`
`functionality. Copying is not relevant to prove nonobviousness where, as here, the
`
`feature that is copied is present in the prior art. Amazon.com, Inc. v.
`
`Barnesandnoble.com, Inc., 239 F.3d 1343, 1366 (Fed. Cir. 2001) (“[E]vidence of
`
`copying by [defendant] and others is not sufficient to demonstrate nonobviousness
`
`
`
`18
`
`

`

`of the claimed invention, in view of the substantial question of validity raised by
`
`the prior art references.”); W.M. Wrigley Jr. Co. v. Cadbury Adams USA LLC, 683
`
`F.3d 1356, 1364 (Fed. Cir. 2012); see also In re Kao, 639 F.3d at 1068 (“Where the
`
`offered secondary consideration actually results from something other than what is
`
`both claimed and novel in the claim, there is no nexus to the merits of the claimed
`
`invention.”).
`
`PO compares competing products to various versions of GuideLiner,
`
`ignoring the fact that the features for which is claims nexus (and copying) are
`
`clearly disclosed in the prior art. Like Itou and Ressemann, Guidezilla3 and The
`
`Boosting Catheter are Rx devices configured to deliver a wide variety of IVCDs
`
`and provide increased back up support when extended partially past the end of a
`
`
`3 PO has not submitted the Boston Scientific license or provided any context
`
`surrounding it. See Iron Grip Barbell Co. v. USA Sports, Inc., 392 F.3d 1317, 1324
`
`(Fed. Cir. 2004) (licenses entitled to little weight “if the patentee does not
`
`demonstrate a nexus between the merits of the invention and the licenses of
`
`record.”) PO and Boston Scientific entered into a settlement agreement and
`
`terminated their litigation in August 2014 (Ex-2044 ¶34), less than three months
`
`after Boston Scientific filed four IPR petitions. See IPR2014-00759, -760, -762,
`
`-763.
`
`
`
`19
`
`

`

`GC. Compare Ex-2046 ¶¶70-71 and Ex-2056 ¶¶73-74 with Ex-1607, 4:27-32,
`
`Figs. 3-5; Ex-1808 Abstract, 6:18-24, 12:9-14:39, Figs. 6A-6F. Like Itou and
`
`Ressemann, Guidezilla also had a side opening. Id.
`
`Telescope also practices the prior art. 4 Compare Ex-2071, 10, 35 with
`
`Ex-1607, Figs. 3-5, 4:27-32; Ex-1608, Abstract, Figs. 6A-6F, 6:18-24, 12:9-14:39.
`
`PO’s copying argument focuses on Petitioner’s alleged copying of GuideLiner’s
`
`“half-pipe”—a design element only present in the third version of GuideLiner.
`
`POR, 33-41. None of the claims for which PO claims nexus in this IPR pertain to
`
`its “half-pipe.”5 See IPR2020-00126, Ex-2138 ¶147.
`
`
`4 PO’s suggestion that regulatory approval supports copying is wrong. “FDA
`
`equivalence is irrelevant to patent law because it involves fundamentally different
`
`inquiries.” Johns Hopkins Univ. v. Datascope Corp., 543 F.3d 1342, 1349 n.3 (Fed.
`
`Cir. 2008); see also Ex-1677, 22:3-20.
`
`5 Telescope is not a “copy” of GuideLiner. Ex-1830 ¶3-27. In addition to the
`
`differences acknowledged by PO (POR, 40),
`
`
`
` Ex-
`
`1821, 233:10-23, 236:16-21 (
`
`). Telescope features SmoothPass Technology (Ex-2071, 15-27) for
`
`improved device deliverability,
`
`
`
`20
`
`
`
`
`
`

`

`Indeed, neither Telescope (nor GuideLiner) practice the claims for which PO
`
`claims nexus. Claims 48, 51, and 53 of the ’760 patent require the subject catheter
`
`to have a “tubular structure defining a lumen” with a “uniform cross-sectional
`
`inner diameter that is not more than one French size smaller than the cross-
`
`sectional inner diameter of the lumen of the guide catheter.”
`
`
`
`
`
`
`
` Ex-1115, 154:14-155:4. Further, PO’s own
`
`expert previously testified that “[m]athmatically, those of ordinary skill in the art
`
`define one French as .0131 inches.” Ex-1825 ¶¶43-48 (testifying that Booting
`
`Catheter infringes due to the 0.0131 difference); see also Ex-1844, 27:3-9 (Judge
`
`Schiltz noting that PO’s attorneys “stood before [him] in QX Medical and told
`
`[him] one French meant 0.00131 inches”).
`
`
`
`29; Ex-1823, 11.
`
`
`
`21
`
` Ex-1822, 28-
`
`

`

`But the difference for both GuideLiner and Telescope is 0.014 inches.
`
`GuideLiner" VSGEC‘
`
`0.051
`
`0.063
`
`6F20.066
`
`Telescope" csc
`
`0.056
`
`Gddetiner"vse£c* 0.056
`
`.067
`
`.007
`
`6F20.070
`
`6F20.070
`
`0
`
`Guidezillamll GEC’
`
`0.057
`
`0.067
`
`0 F 2 0.070
`
`- Telescope”G£C
`7
`GuideLinen"V5 GEC‘
`
`0.062
`0.062
`
`0.075
`0.015
`
`7 F 2 0.078
`7 F20.078
`
`7
`
`Guidezflla'" II (5E8
`
`0.06;
`
`0.07}
`
`7 F 2 0.078
`
`5
`
`25
`
`25
`25
`
`25
`
`4
`
`n
`
`N/A, metal c0110!
`
`a
`n
`
`N/A. nwtal collm
`
`250
`
`150
`
`150
`
`150
`150
`
`150
`
`
`
`l
`
`Ex-1682 Ex-A, 39. The District Court recognized as much at the preliminary
`
`injunction hearing. Ex-1844, 31 :8-11 (“I just have to say that I give you kudos for
`
`your creativity [that 1 French means more than 0.0131 inches]. You’re going to be
`
`walking uphill for me on that argument”).
`
`PO’s “copying” arguments also ignore other important differences between
`
`GuideLiner and Telescope. Ex—1830 113-27. PO invites the Board to look at the
`
`parties’ products “side by side,” while ignoring the alleged “small differences,”
`
`including “Telescope’s hydrophilic coating and round push wire.” POR, 41.
`
`“Copying requires the replication of a specific product,” Iron Grip, 392 F.3d at
`
`1325, and these differences are not “small.” Ex-1830 1116-26. When deposed in the
`
`District Court, MS- Welch—
`
`— Ex-1821, 18, 53-54; see also Ex-2071, 7.—
`
`22
`
`

`

` Ex-1822, 13; see also Ex-1823, 18. GEC
`
`coating is clinically relevant. Ex-1817, 65:16-66:10.
`
`
`
`
`
` Ex-1824, 4:22-5:10. These product differences confirm a
`
`
`
`lack of copying.
`
`V. CONCLUSION
`
`For the foregoing reasons, as well as those set forth in the Petition, claims 48
`
`and 51-53 of the ’760 patent are anticipated and/or obvious.
`
`
`
`
`
`Date: December 21, 2020
`
`
`
`
`
`
`
`
`
`
`RESPECTFULLY SUBMITTED,
`
`ROBINS KAPLAN LLP
`
`
`
`/ Cyrus A. Morton /
`Cyrus A. Morton
`800 LaSalle Avenue, Suite 2800
`Minneapolis, MN 55402
`(612) 349-8500
`Attorney for Petitioner
`
`
`
`23
`
`

`

`WORD COUNT CERTIFICATION
`
`I hereby certify that this Reply complies with the word count limit, and
`
`contains 4,061 words. I further certify that, in preparation of this Reply, I used
`
`Microsoft Word, Version 2016, and that this word processing program has been
`
`applied specifically to include all text, including headings, footnotes, and
`
`quotations in the following word count.
`
`
`
`Dated: December 21, 2020
`
`
`
`
`
`
`
`
`
`
`
`
`/ Cyrus A. Morton /
`Cyrus A. Morton
`Registration No. 44,954
`Attorney for Petitioner
`
`
`
`
`
`
`
`
`
`24
`
`

`

`CERTIFICATE OF SERVICE
`
`I hereby certify that on this December 21, 2020, a copy of Petitioner’s Reply
`
`was served in its entirety by electronic mail on Patent Owner’s counsel at the
`
`following addresses indicated in Patent Owner’s Mandatory Notices:
`
`J. Derek Vandenburgh, Reg. No. 32,179
`dvandenburgh@carlsoncaspers.com
`
`Dennis C. Bremer, Reg. No. 40,528
`dbremer@carlsoncaspers.com
`
`Joseph W. Winkels
`jwinkels@carlsoncaspers.com
`
`Peter M. Kohlhepp
`pkohlhepp@carlsoncaspers.com
`
`Tara C. Norgard
`tnorgard@carlsoncaspers.com
`
`Alexander S. Rinn
`pkohlhepp@carlsoncaspers.com
`
`Megan E. Christner, Reg. No. 78,979
`mchristner@carlsoncaspers.com
`
`
`Dated: December 21, 2020
`
`
`
`
`
`Respectfully submitted,
`
`
`
`
`
`
`
`
`
`
`/ Cyrus A. Morton /
`Cyrus A. Morton
`Reg. No. 44,954
`Attorney for Petitioner
`
`
`25
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket