throbber
Paper No. __________
`
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`___________________________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`__________________________
`
`
`
`
`MEDTRONIC, INC.
`
`Petitioner,
`
`v.
`
`TELEFLEX INNOVATIONS S.À.R.L.,
`
`Patent Owner
`_____________________________
`Case No: IPR2020-00132
`U.S. Patent No. RE45,760E
`______________________________
`
`PETITIONER’S REPLY
`
`

`

`TABLE OF CONTENTS
`
`Page
`TABLE OF CONTENTS ............................................................................................ i
`
`TABLE OF AUTHORITIES .................................................................................... iii
`
`I.
`
`INTRODUCTION ........................................................................................... 1
`
`II. GROUND 1: PO DOES NOT CHALLENGE THAT ITOU
`ANTICIPATES CLAIMS 25-31, 33-38, 41-42, 44 AND 47. ......................... 1
`
`III. GROUND 2: ITOU IN VIEW OF RESSEMANN RENDERS
`CLAIMS 32, 39 AND 40 OBVIOUS. ............................................................. 1
`
`A.
`
`PO does not challenge Medtronic’s evidence on the obviousness of
`claim 40. ................................................................................................ 1
`
`B.
`
`Claim 39 is rendered obvious by Itou in view of Ressemann. .............. 2
`
`1.
`
`2.
`
`Itou discloses a tubular structure configured to receive a “stent
`releasably joined to the distal end of [an] elongate balloon
`catheter.” ..................................................................................... 4
`
`The opening of Itou is actually larger, allowing larger devices
`through as well. ........................................................................... 5
`
`C.
`
`Claim 32 ................................................................................................ 7
`
`1.
`
`2.
`
`3.
`
`Adding “at least two inclined slopes” was not inventive. .......... 7
`
`Collar 2141 has more than two inclined slopes. ......................... 8
`
`Itou in view of Ressemann renders claim 32 obvious. ............. 10
`
`IV. GROUND 3: CLAIM 32 IS RENDERED OBVIOUS BY ITOU IN
`VIEW OF KATAISHI. ................................................................................... 15
`
`V.
`
`PO’S “COPYING” ALLEGATIONS DO NOT OVERCOME
`PETITIONER’S STRONG OBVIOUSNESS SHOWING. .......................... 19
`
`A.
`
`B.
`
`Side openings existed on prior art devices. ......................................... 20
`
`PO’s copying arguments are without merit. ........................................ 21
`
`
`
`i
`
`

`

`VI. AIA PATENT ................................................................................................. 24
`
`VII. CONCLUSION .............................................................................................. 25
`
`
`
`
`
`
`
`ii
`
`

`

`TABLE OF AUTHORITIES
`
` Page(s)
`
`Cases
`
`Allied Erecting & Dismantling Co. v. Genesis Attachments, LLC,
`825 F.3d 1373 (Fed. Cir. 2016) .......................................................................... 10
`
`Amazon.com, Inc. v. Barnesandnoble.com, Inc.,
`239 F.3d 1343 (Fed. Cir. 2001) .................................................................... 22, 23
`
`Google LLC v. Lee,
`759 F. App’x 992 (Fed. Cir. 2019) ..................................................................... 10
`
`In re Applied Materials
`692 F.3d 1289 (Fed. Cir. 2012) .......................................................................... 11
`
`In re Magna Elecs., Inc.,
`611 F. App’x 969 (Fed. Cir. 2015) ....................................................................... 7
`
`Johns Hopkins Univ. v. Datascope Corp.,
`543 F.3d 1342 (Fed. Cir. 2008) .......................................................................... 23
`
`KSR Int’l Co. v. Teleflex, Inc.
`550 U.S. 398 (2007) .......................................................................................... 1, 7
`
`Tyco Healthcare Grp. LP v. Ethicon Endo-Surgery, Inc.,
`774 F.3d 968 (Fed. Cir. 2014) .............................................................................. 3
`
`ZUP, LLC v. Nash Mfg., Inc.,
`896 F.3d 1365 (Fed. Cir. 2018) .......................................................................... 20
`
`
`
`
`
`
`iii
`
`

`

`I.
`
`INTRODUCTION
`
`Patent Owner (“PO”) does not and cannot refute that Itou discloses each
`
`limitation of 17 of the 20 challenged claims (Compare Paper 1 (“Pet.”), 19-21 with
`
`Paper 44 (“POR”), 9-10), which are invalid as anticipated. PO only attempts to
`
`refute evidence that claims 32 and 39 are obvious. POR, 9-17. To do so, however,
`
`it must argue that a POSITA would never look to the teachings of one coronary
`
`catheter designed to remove unwanted material from the coronary vasculature,
`
`Ressemann (Ex-1008), to inform as to the use of a second, Itou (Ex-1007), which
`
`has exactly the same purpose. Itou and Ressemann are clearly analogous art, as
`
`obviousness inquiries properly take into account the “inferences and creative steps
`
`that a person of ordinary skill in the art would employ.” KSR Int’l Co. v. Teleflex,
`
`Inc. 550 U.S. 398, 418 (2007).
`
`II. GROUND 1: PO DOES NOT CHALLENGE THAT ITOU
`ANTICIPATES CLAIMS 25-31, 33-38, 41-42, 44 AND 47.
`
`As set forth in Paper 78, Itou is prior art.
`
`III. GROUND 2: ITOU IN VIEW OF RESSEMANN RENDERS CLAIMS
`32, 39 AND 40 OBVIOUS.
`
`A.
`
`PO does not challenge Medtronic’s evidence on the obviousness of
`claim 40.
`
`Similar to ground 1, PO has not challenged the disclosure of the art asserted
`
`against claim 40. It is obvious to use an “elongate balloon catheter,” as taught in
`
`Ressemann, with Itou’s catheter 2, as set forth in the Petition.
`
`
`
`1
`
`

`

`B. Claim 39 is rendered obvious by Itou in view of Ressemann.
`
`Like claim 40, it is also obvious to use an elongate balloon catheter with “a
`
`stent releasably joined to [its] distal end.” Ex-1001, 14:46-48. This is just another
`
`commonly known device that goes through a catheter, as explicitly taught in
`
`Ressemann. Ex-1008, 13:57-14:16; see id., Figs. 6E-F, 6H-I. There was ample
`
`motivation for a POSITA to use such a device with Itou, as well as a reasonable
`
`expectation of success. Pet., 68-71. Itou in view of Ressemann renders claim 39
`
`obvious while requiring not a single change to catheter (2)’s structure. See Pet.,
`
`21-38, 47-48.
`
`PO dismisses Itou’s disclosure because it teaches preassembly of catheters
`
`(2) and (5), while claim 39 recites the side opening receive a stent or balloon while
`
`it is positioned within the guide catheter, and while the extension catheter’s distal
`
`end is extended beyond the distal end of the guide catheter. Pet., 46. Itou’s
`
`disclosure is relevant for two reasons. First, it demonstrates that catheter (2) is long
`
`enough so that when its distal end extends from the guide catheter, its side opening
`
`remains in the guide catheter. Ex-1007, 5:35-42, Fig. 6, Table 1. Second, once
`
`catheter (2) is in place, catheter (5) may be withdrawn (Id., 7:16-19), which means
`
`that catheter (2)’s side opening is in just the right location to receive a “stent
`
`releasably joined” to a balloon catheter. Ex-1001, 4:46-48.
`
`
`
`2
`
`

`

`To the extent there is still a question about the motivation to look to
`
`Ressemann to modify Itou, Medtronic notes that Itou and Ressemann are clearly
`
`analogous art. See Tyco Healthcare Grp. LP v. Ethicon Endo-Surgery, Inc., 774
`
`F.3d 968, 979 (Fed. Cir. 2014).
`
`Itou discloses a suction catheter, while Ressemann teaches an embolic
`
`protection device that is also used to deliver a stent. Pet., 19-21, 56-61. The patent
`
`claims, Itou, and Ressemann all share the same field of endeavor, as they are all
`
`coronary catheters designed to provide solutions to the problem of coronary artery
`
`occlusion. Ex-1806 ¶¶ 64-74; see also Ex-1805, 20:7-21:1, 70:13-72:1. As Dr.
`
`Brecker explains, an interventional cardiologist would know that Itou’s catheter (2)
`
`could sequentially deploy a distal end protection device, deliver a stent, and then
`
`suction. And, given its structure, he would know that catheter (2) could be used
`
`solely to deliver a stent. Ex-1806 ¶¶ 75-76; see Ex-1811, Table 1.
`
`Inventor Root explained that the alleged invention of GuideLiner itself grew
`
`from prior work on a suction catheter. Ex-1762, 46:9-20. In fact, VSI used the
`
`same technology it used in its Pronto suction catheter to make GuideLiner’s distal
`
`tube. Id., 45:10-17.
`
`
`
`3
`
`

`

`1.
`
`Itou discloses a tubular structure configured to receive a “stent
`releasably joined to the distal end of [an] elongate balloon
`catheter.”
`
`If the “effective size” of Itou’s opening was 0.046 inches, as PO wrongly
`
`suggests, Itou still inherently discloses that catheter (2) is configured to receive a
`
`balloon-expandable stent, thereby satisfying the requirements of claim 39. Before
`
`the alleged invention of the ’760 patent, a POSITA was aware of numerous
`
`balloon-expandable stents that were deliverable through a catheter with an inner
`
`diameter (“ID”) of 0.046 inches. Ex-1806 ¶¶ 47-51; Ex-1015b, 189; Ex-1802, 4, 7,
`
`14-15, 21, 25; Ex-1803, 4:54-58; Ex-1804, Table 2, Fig. 3.
`
`At least three reasons demonstrate that PO cannot credibly argue otherwise.
`
`First, Mr. Keith, PO’s expert, admits that Itou teaches advancing catheter (2) into
`
`the coronary vasculature, so that the distal part of tubular structure (24) is extended
`
`into the vasculature, but the proximal opening remains within the guide catheter.
`
`Ex-1805, 137:7-138:16; Ex-1007, 5:35-42, 7:13-18.
`
`
`
`4
`
`
`
`

`

`
`
`Ex-1007, Fig. 6.
`
`Second, Keith also admits that a guidewire could be navigated into guiding
`
`catheter 1—and then through the guide catheter and right up to the opening of
`
`catheter (2)—but then has no idea whether it would be able to enter the opening.
`
`Ex-1805, 143:3-145:20.
`
`
`
`Ex-2138 ¶ 164 (modified).
`
`Third, Mr. Keith’s testimony is questionable, as the 5 French version of
`
`GuideLiner v.3 also has an ID of 0.046 inches and is marketed for delivering
`
`stents, which are, of course, much larger than a guidewire. Ex-2164, 5.
`
`2. The opening of Itou is actually larger, allowing larger devices
`through as well.
`
`Itou teaches an opening of .059 inches, not .046 as argued by PO. The
`
`teaching of Itou is correct because the fully circumferential portion of the pushwire
`
`
`
`5
`
`

`

`is not flush with the lumen, and the wire is flexible and bends in use. Ex-1806
`
`¶¶53-57; Ex-1807 ¶¶28-46.
`
`Catheter (2)’s angulated proximal opening presents its widest diameter at a
`
`substantial angle to wire 25, which tapers as it is welded to collar 231 (“crushed
`
`into a form of a flat plate”). The flattening of wire 25 means the fully-
`
`circumferential portion of wire (25) does not block catheter (2)’s proximal
`
`opening. Ex-1806 ¶56; Ex-1807 ¶¶35-39; see Ex-1114, 75:20-76:6 (
`
`
`
`).
`
`
`
`Additionally, Itou teaches that the distal portion of catheter (2) is
`
`emphatically not rigid, as it must be navigated into a coronary artery. Ex-1007,
`
`Abstract, Fig. 6, 5:35-42. This necessarily requires that the distal tubular portion of
`
`catheter (2) bend, as—if it did not—it could never be positioned where Itou
`
`instructs, which is “deep” in the coronary vasculature. Id., 1:66-2:5, 5:35-39, Fig.
`
`6; Ex-1806 ¶¶54-55; Ex-1807 ¶36.
`
`Collectively, the structure and flexibility of catheter (2) mean that the
`
`“effective size” of its’ opening is exactly what Itou teaches: 1.5 mm, or 0.059
`
`inches. Ex-1806 ¶¶53-54; Ex-1807 ¶¶30-39.
`
`
`
`6
`
`

`

`C. Claim 32
`
`Claim 32 adds that “the segment defining the side opening includes at least
`
`two inclined slopes. Ex-1001, 10:51-11:21.
`
`1. Adding “at least two inclined slopes” was not inventive.
`
`PO’s inventors and experts admit that the various iterations of the double-
`
`incline configuration are not important, and provide no benefit over a single
`
`incline.
`
`
`
`. Ex-1108, 75:12-76:1. Similarly, Root testified that a
`
`wide variety of side opening configurations all worked for their intended purpose,
`
`and that there was no difference between the various iterations. Ex-1762, 79:17-
`
`80:12, 91:2-19, 115:6-14, 280:22-210:21. And both PO experts struggled to find an
`
`advantage for the double incline in practice—GuideLiner version 2 had only a
`
`single incline (Ex-1800, 49:7-14)—or the patents. Ex-1813, 48:16-49:3, 98:5-15;
`
`Ex-1800, 39:1-41:18, 49:15-51:21, 55:12-58:15; Ex-1817, 60:10-62:5, 63:10-23.
`
`As a result, adding a second incline to the proximal opening constitutes “the
`
`predictable use of prior art elements according to their established functions.” See
`
`KSR, 550 U.S. at 417; see also In re Magna Elecs., Inc., 611 F. App’x 969, 974
`
`(Fed. Cir. 2015) (finding that “short horizontal lines provide the same information
`
`and functionality as long horizontal lines, and cannot be used as a distinguishing
`
`
`
`7
`
`

`

`factor to render the claims nonobvious”). Regardless, as set forth below, the
`
`claimed inclines are present in the art and obvious.
`
`2. Collar 2141 has more than two inclined slopes.
`
`PO states that GuideLiner version 1 has three inclines. Thus, collar 2141
`
`also has at least three inclines. Ex-1806 ¶87.
`
`Keith testified that he identifies where an incline ends and another begins
`
`“just sort of by eye,” explaining that for version 1 of GuideLiner, as identified
`
`below, incline 1 has a “relatively steeper quality” to it, and incline 3 has “a
`
`shallower quality” to it, and “somewhere in between there is what goes from one to
`
`the next” is incline 2. Ex-1800, 45:21-46:2. His opinion is that even shallow
`
`inclines are inclines so long as they can be identified relative to the catheter’s
`
`longitudinal axis. Id., 47:3-8.
`
`Ex-1122 (color added to arrows for visibility).
`
`
`
`
`
`8
`
`

`

`Similarly, Root testifies that the figure below shows two inclines that are
`
`separated by a flat section. Ex-1762, 91:24-93:24.
`
`Ex-1116 (bottom annotation added to show inclined regions (red) and flat region
`
`
`
`(blue)).
`
`In light of PO’s witness testimony, the Ressemann collar must have more
`
`than two inclines. As shown below, support collar 2141 has at least an incline (C)
`
`that is just proximal to its fully circumferential portion, an incline (B)
`
`corresponding to the curve between 2141a and 2141b (similar to the curve in Fig. 4
`
`of the patent), and an incline (A) at the curved portion of tab 2141b. Ex-1806 ¶¶80-
`
`91.
`
`
`
`9
`
`

`

`
`
`Ex-1008, Fig. 16J (annotated); see Ex-2138 ¶¶140, 148, 152 (Keith’s various 3D
`
`renderings of collar 2141 showing at least a third incline).
`
`3.
`
`Itou in view of Ressemann renders claim 32 obvious.
`
`PO argues that a POSITA would not have been motivated to combine
`
`Ressemann’s collar 2141 with Itou’s suction catheter 2 because in its assembled
`
`state, Ressemann does not teach this configuration of collar 2141. POR, 22-31.
`
`This is not sufficient to overcome Petitioner’s showing of obviousness. Allied
`
`Erecting & Dismantling Co. v. Genesis Attachments, LLC, 825 F.3d 1373, 1380-81
`
`(Fed. Cir. 2016) (explaining that claims can be obvious even if a prior art
`
`combination requires “substantial redesign and reconstruction” of a reference). The
`
`law simply does not require that a component from reference A be combined with
`
`reference B in exactly the same way as is taught in reference A. Google LLC v.
`
`Lee, 759 F. App’x 992, 997 (Fed. Cir. 2019) (“Our obviousness framework permits
`
`modifying one reference and combining it with a second reference.”). And that is
`
`
`
`10
`
`

`

`because a “reference must be considered for everything that it teaches, not simply
`
`the described invention or a preferred embodiment.” In re Applied Materials 692
`
`F.3d 1289, 1298 (Fed. Cir. 2012) (emphasis added).
`
`PO actually does not dispute the first motivation to combine collar 2141
`
`with Itou – larger area of entry for devices. PO argues that not all double inclined
`
`openings have a larger area of entry. But it is undisputed that collar 2141 does have
`
`a larger area, and that a larger opening is beneficial for receiving stent and balloon
`
`catheters. Compare Pet., 66 with POR, 24-29. As Petitioner has proposed to modify
`
`Itou, collar 2141 would increase the area of entry to catheter (2) to over three times
`
`its size. Ex-1807 ¶125. Moreover, the combination is easy because tab 2141b could
`
`be welded to wire (25), just as Itou teaches welding collar 23 to wire 25. Ex-1007,
`
`4:33-36. Keith admits this would be a matter of routine engineering. Ex-1800,
`
`48:5-49:5. Double incline claims are obvious on this basis alone.
`
`PO also argues that a POSITA would not have been motivated to add collar
`
`2141 to catheter (2) to reinforce its opening, because Itou’s metal collar is thicker
`
`than the preferred embodiment of Ressemann. POR, 35. But this is actually another
`
`reason to do it. The Ressemann walls are half as thick. Id.; see Ex-2138 ¶144-46
`
`(comparing Itou’s 0.0043 inch wall thickness to Ressemann’s 0.002 inch
`
`thickness); Ex-1805, 180:17-181:4. This is beneficial as the collar would still
`
`
`
`11
`
`

`

`provide reinforcement but be less bulky, Ex-1807 ¶126; Ex-1806 ¶89, ensuring an
`
`increased area of opening.
`
`An independent reason to combine is to provide a flexibility transition not
`
`present in Itou. Collar 2141’s tab 2141b provides the opportunity to improve on
`
`Itou’s weld point, providing a flexibility transition by extending wire 25 under the
`
`tab.
`
`Itou explains that the weld point is crushed into the form of a flat plate
`
`(Ex-1007, 4:33-36), which creates a potential weakness point. Ex-1813, 161:4-10.
`
`A POSITA understands that crushing a solid metal wire into a flat plate decreases
`
`the metal’s ductility at that location. Ex-1807 ¶127; Ex-1818, 117-18, 121-23. And
`
`because decreasing ductility at critical stiffness transition points is known in the art
`
`to create the risk of kinking, a POSITA had the motivation to look to alternative
`
`options that would have less risk of failure. Ex-1807 ¶127.
`
`Itou’s weld point is located at a critical stiffness transition point, where
`
`“substantially rigid” wire (25) transitions into tubular portion (24). Ex-1005 ¶¶151,
`
`185-87, 197-202, 207-13; Ex-1042 ¶¶27, 64-84; Ex-1807 ¶¶52-66, 127-128, 210.
`
`Before the Teleflex patent, it was widely appreciated that rapid exchange catheters
`
`were susceptible to kinking at the “interface between the stiff proximal portion and
`
`the flexible distal portion due to the change in stiffness at this interface,” and
`
`
`
`12
`
`

`

`various ways were known to address the interface. Ex-1816, 2:41-45; Ex-1807
`
`¶¶52-66, 127-128, 210.
`
`Here, PO does not dispute the flexibility transitions of catheter (2). POR, 43-
`
`45. Nor does its expert have an opinion on them. Ex-1805, 124:9-129:3; 131:11-
`
`132:13. As discussed above, a POSITA knew that transitional regions in catheter
`
`flexibility could be improved upon. This is the very point of tab 2141 of
`
`Ressemann’s collar. Ex-1008, 24:64-67; Ex-1807 ¶¶48, 128, 103.
`
`PO also argues that a POSITA would have recognized the problems created
`
`by placing Ressemann’s support collar atop catheter (2)’s wire (25). POR, 41-42.
`
`Specifically, PO argues that the proposed modification would (i) create a catch-
`
`point and (ii) be “complex” because convex up metal tab to a convex down shaft
`
`would have to be bonded. Id. Both arguments fail.
`
`As is standard practice in the art—and as explicitly taught by Ressemann
`
`(Ex-1008, 7:28-39, 24:2-11)—support collar 2141 would be encased in a polymer
`
`jacket. Ex-1807 ¶¶90-96, 129-132. In the 2D rendering of a 3D structure, below,
`
`Itou’s tubular portion (yellow) is shown, and collar 2141 is embedded between the
`
`tubular structure’s resin layers, similar to Itou’s collar 23. Proximal of the tubular
`
`portion, tab 2141 b extends along the top of wire-like portion (25), secured by
`
`polymer (red). Ex-1807 ¶¶130-132.
`
`
`
`13
`
`

`

`Id. ¶130.
`
`
`
`In so doing, as shown above, the proposed modification would have no
`
`catch-points, and the shape of the collar—specifically at least two inclined
`
`slopes—would be retained and be present in the opening of catheter 2. Ex-1807
`
`¶¶130-132. This is the case if tab 2141b is incorporated atop of wire (25), as
`
`illustrated above, in which case inclines A, B, and C are part of the segment
`
`defining the side opening. It is also the case even if, as PO suggests, tab 2141b is
`
`incorporated underneath wire (25) (POR, 40), in which case at least inclines B and
`
`C are part of the segment defining the side opening. In fact, as Dr. Brecker opines,
`
`there is another incline between inclines B and C. Ex-1806 ¶88.
`
`PO also incorrectly argues that the “convex up” and “convex down”
`
`arrangement would cause the proposed modification to “pop” loose. POR, 41.
`
`Indeed, numerous examples, including those cited on the face of the PO patent
`
`(e.g., Ex-1030), teach catheter assemblies with portions having convex up (distal
`
`lumen) and convex down components (pushrod). Ex-1807 ¶¶93-96. Thus, there
`
`
`
`14
`
`

`

`was a reasonable expectation of success in making the proposed modification, and
`
`claim 32 is rendered obvious by Itou in view of Ressemann.
`
`Finally, PO is incorrect in asserting that using Ressemann’s collar in Itou
`
`would eliminate the 1 Fr differential between guide catheter (1) and catheter (2).
`
`POR, 43-45. Itou teaches a metal collar that is between two layers of polymer (i.e.
`
`“within” the walls of suction catheter (2)). Ex-1007, 4:36-38. There would be no
`
`reason that the circumferential portion of collar 2141 would not be similarly
`
`positioned. Ex-1807 ¶129. And, as collar 2141 is thinner than metal collar 23,
`
`using the former instead of the latter increases the ID of catheter (2). Ex-1807
`
`¶¶126, 129. An increase in the ID of catheter 2 by 0.0046 inches is an increase of
`
`0.11 mm. Thus, catheter 2’s modified ID is 1.61 mm, and the differential between
`
`the lumen of modified catheter 2 and guide catheter 1 is 0.19 mm, which is less
`
`than 1 French. Ex-1062, 545. Claim 39 is also obvious.
`
`IV. GROUND 3: CLAIM 32 IS RENDERED OBVIOUS BY ITOU IN
`VIEW OF KATAISHI.
`
`A POSITA was motivated to combine the shape of Kataishi’s distal opening
`
`with Itou’s proximal opening. Pet., 73-78. Specifically, adding Kataishi’s shape to
`
`Itou’s proximal opening would have improved loading of interventional devices
`
`and improved crossability of the catheter. Id. PO is incorrect to argue otherwise.
`
`PO argues that the benefits of Kataishi’s distal tip do not apply equally to
`
`Itou’s proximal opening. POR, 51. This is belied by explicit teachings in
`
`
`
`15
`
`

`

`Ressemann— and in a patent by their own expert, Mr. Keith—that explain that
`
`the shape of a lumen’s opening functions similarly, regardless of whether it is
`
`placed at the distal end or the proximal end.
`
`The proximal and distal ends 140a, 140b of the evacuation lumen 140
`
`are preferably angled to allow for smoother passage of the evacuation
`
`sheath assembly 100 through a guide catheter, and into a blood vessel,
`
`and to facilitate smoother passage of other therapeutic devices
`
`through the evacuation lumen 140 of the evacuation head 132. The
`
`larger area of the angled open ends also allows for larger deformable
`
`particulate matter to pass through the lumen more smoothly.
`
`Ex-1008, 6:52-60 (emphasis added).
`
`
`
`Id., Fig. 1A; see Ex-1800, 140:18-143:7, 146:16-147:8,148:21-149:14.
`
`PO’s own expert, Mr. Keith, is a named inventor on a related patent, which
`
`contains the verbatim teaching that the shape of both the distal and proximal
`
`opening can improve suction of particulate matter. Ex-1123, 7:54-60; Ex-1800,
`
`149:3-18. The larger area of the angled openings is beneficial both for suctioning
`
`“larger deformable particular matter” as well as for introducing a therapy catheter,
`
`
`
`16
`
`

`

`such as a balloon-expandable stent. Ex-1800, 149:3-8; see Ex-1008, Figs. 6E, 6G,
`
`13:57:14-14:26. Kataishi’s teaching that the shape of its distal tip improves that
`
`catheter’s ability to suction thrombus would be understood by a POSITA to also
`
`apply to the proximal opening of a catheter such as Ressemann or Itou. Ex-1807
`
`¶¶133-143; Ex-1806 ¶96. The shape of Kataishi’s opening increases the area of the
`
`proximal opening as compared to Itou’s opening. Ex-1807 ¶146.
`
`Id. (Schematic illustrating relative area of Itou’s proximal opening as shown in
`
`
`
`(Ex-1007, Fig. 3) versus Kataishi’s proximal opening (Ex-1025, Fig. 12)).
`
`And an increase in the area of opening makes it easier to introduce a stent or
`
`balloon catheter. Ex-1005 ¶¶283-90; Ex-1042 ¶¶104-112; Ex-1807 ¶¶136-138;
`
`Ex-1005 ¶¶287-289.
`
`
`
`17
`
`

`

`PO is also incorrect to argue that adding Kataishi’s shape to Itou’s proximal
`
`opening would not improve catheter crossability. POR, 51-52. As a threshold
`
`matter Ressemann teaches that both the proximal and distal ends of evacuation
`
`lumen 140 assist in ensuring “smoother passage of the evacuation sheath assembly
`
`100 through a guide catheter.” Ex-1008, 6:52-57; Ex-1123, 7:54-60. Further, PO is
`
`incorrect to argue that “crossability” is limited to the suction catheter’s ability to
`
`“extend[] past the end of the guide catheter through tortuous vasculature.” POR,
`
`51. Indeed, Sakurada, which tested the Kataishi device, performed a “passing
`
`ability test” that measured “[t]he length between the ostium and the distal tip of the
`
`aspiration catheter,” and found that Kataishi’s shape had improved crossability. Ex-
`
`1055, 300-02. And it is this improved crossability that prevents kinking (i.e.,
`
`catheter kink when encountering an unnavigable portion of the vasculature). Ex-
`
`1807 ¶¶52-66, 139-142, 210. A POSITA appreciates that modifying Itou’s proximal
`
`opening to include the shape of Kataishi’s distal tip was an obvious modification.
`
`Ex-1806 ¶¶92-98.
`
`PO further argues that Kataishi’s benefits—that allegedly flow from both its
`
`shape and flexibility of the material—is inapplicable to Ressemann’s collar. PO is
`
`again belied by its own expert witness, who testified that once a shape is provided,
`
`a POSITA would readily use routine engineering to create side openings of several
`
`
`
`18
`
`

`

`different materials, some more rigid and some more flexible. Ex-1764, 27:9-17,
`
`30:2-8, 31:16-21.
`
`V.
`
`PO’S “COPYING” ALLEGATIONS DO NOT OVERCOME
`PETITIONER’S STRONG OBVIOUSNESS SHOWING.
`
`The earliest rapid exchange (“Rx”) GEC in the record is Kontos, followed
`
`by Ressemann and Itou. These devices are Rx, configured to deliver a wide variety
`
`of IVCDs, and provide increased back up support when extended partially past the
`
`end of a GC as intended. These devices—and the functionality associated with
`
`them—were all published in the art in advance of the 2004 Transcatheter
`
`Cardiovascular Therapeutics conference, where inventor Root maintains he
`
`conceived of the idea claimed in the Teleflex patents.1 Compare Ex-1755 ¶¶27-38
`
`with Ex-2118 ¶¶5-6; see also Ex-1806 ¶¶205-219; Ex-2138 ¶71; Ex-2145 ¶82;
`
`Ex-1762, 39:1-22; Ex-1114, 111:13-17; Ex-1800, 16:22-17:1; Ex-1762, 194:4-16;
`
`Ex-1817, 25:20-26:9, 63:10-64:8; Ex-2215 ¶22; Ex-1820, 23:15-26:22, 36:3-10;
`
`Ex-2123 ¶20.
`
`PO bases its entire secondary considerations case on alleged copying of
`
`claim 32—a dependent claims that recites a side opening adapted to receive IVCDs
`
`
`1
`
`
`
`. Ex-1762, 39:19-21,
`
`67:23-68:1; Ex-2003; Ex-1114, 61:24-62:9, 77:9-15, 121:12-122:1.
`
`
`
`19
`
`

`

`with “at least two inclined slopes.” POR, 60-67. Support collar 2141 of Ressemann
`
`discloses or teaches a device with at least two inclined slopes. Itou discloses a side
`
`opening, and the only potential obviousness issue is whether it would be obvious to
`
`add another inclined slope to Itou’s side opening. PO cannot overcome Petitioner’s
`
`strong showing of obviousness. ZUP, LLC v. Nash Mfg., Inc., 896 F.3d 1365, 1374
`
`(Fed. Cir. 2018) (“[A] strong showing of obviousness may stand even in the face of
`
`considerable evidence of secondary considerations.”).
`
`A.
`
`Side openings existed on prior art devices.
`
`Side openings existed in the prior art, including prior art devices presented in
`
`these IPRs.
`
`
`
`. Ex-1108, 75:13-18. And PO’s experts were unable to identify particular
`
`advantages to one form of side opening over another. Ex-1800, 39:1-15 (single
`
`incline versus double incline achieve same identified benefits); Ex-1813, 49:4-7
`
`(GuideLiner V2 and V3, with differing side openings, “work the same” if used
`
`“appropriately, as intended”), 94:8-14; see also Ex-1806 ¶205-219.
`
`
`
`
`
`
`
` Ex-1822, 28-29; see
`
`also Ex-1823, 31.
`
`
`
`20
`
`

`

`B.
`
`PO’s copying arguments are without merit.
`
`As an initial matter, neither Telescope nor GuideLiner practice Claim 53 of
`
`the ’776 patent. As admitted by Inventor Root,
`
`. Ex-1115, 154:14-155:4. Further, PO’s own expert previously testified that
`
`“[m]athmatically, those of ordinary skill in the art define one French as .0131
`
`inches.” Ex-1825 ¶¶43-48 (testifying that Booting Catheter infringes due to the
`
`0.0131 difference); see also Ex-1844, 27:3-9 (Judge Schiltz noting that PO’s
`
`attorneys “stood before [him] in QX Medical and told [him] one French meant
`
`0.00131 inches”).
`
`But the difference for both GuideLiner and Telescope is 0.014 inches.
`
`
`
`
`
`
`
`Ex-1082 Ex-A, 39. The District Court recognized as much at the preliminary
`
`injunction hearing. Ex-1844, 31:8-11 (“I just have to say that I give you kudos for
`
`
`
`21
`
`

`

`your creativity [that 1 French means more than 0.0131 inches]. You’re going to be
`
`walking uphill for me on that argument.”).
`
`Further, PO’s “copying” arguments erroneously compare competing
`
`products to various versions of GuideLiner, ignoring the fact that the side opening
`
`described in claim 32 is disclosed or taught in the prior art. Amazon.com, Inc. v.
`
`Barnesandnoble.com, Inc., 239 F.3d 1343, 1366 (Fed. Cir. 2001) (“[E]vidence of
`
`copying by [defendant] and others is not sufficient to demonstrate nonobviousness
`
`of the claimed invention, in view of the substantial question of validity raised by
`
`the prior art references.”).
`
`Like Itou and Ressemann, Guidezilla and The Boosting Catheter are Rx
`
`devices configured to deliver a wide variety of IVCDs and provide increased
`
`backup support when extended partially past the end of a GC. Compare Ex-2046
`
`¶¶70-71 and Ex-2056 ¶¶73-74, with Ex-1007, 4:27-32, Figs. 3-5; Ex-1008
`
`Abstract, 6:18-24, 12:9-14:39, Figs. 6A-6F. Guidezilla2 and The Boosting Catheter
`
`
`2 PO provides no context surrounding its license with Boston Scientific, and it has
`
`not made the license of record. See Iron Grip Barbell Co. v. USA Sports, Inc., 392
`
`F.3d 1317, 1324 (Fed. Cir. 2004) (licenses entitled to little weight “if the patentee
`
`does not demonstrate a nexus between the merits of the invention and the licenses
`
`of record.”) PO and Boston Scientific entered into a settlement agreement and
`
`
`
`22
`
`

`

`have a side opening with more than one incline, like Support Collar 2141 of
`
`Ressemann. Id.
`
`Telescope practices the prior art. 3 Compare Ex-2071, 10, 35 with Ex-1007,
`
`Figs. 3-5, 4:27-32; Ex-1008, Abstract, Figs. 6A-6F, 6:18-24, 12:9-14:39.
`
`PO’s copying argument focuses on Petitioner’s alleged copying of GuideLiner’s
`
`“half-pipe”—a design element only present in the third version of GuideLiner.
`
`POR, 64-67. Telescope does not copy GuideLiner’s “half-pipe.” As Dr. Zalesky
`
`explains the entry ports for inserted therapeutic devices are substantially different
`
`for the two devices. Ex-1830 ¶3-27. Even PO acknowledges that “GuideLiner V3
`
`has a longer side opening with a third inclined region between the two incline
`
`
`terminated their litigation in August 2014 (Ex-2044 ¶34), less than three months
`
`after Boston Scientific filed four petitions for IPR against PO’s alleged GuideLiner
`
`patents. See IPR2014-00759, -760, -762, -763.
`
`3 PO’s suggestion that regulatory approval supports copying is wrong. “FDA
`
`equivalence is irrelevant to patent law because it involves fundamentally different
`
`inquiries.” Johns Hopkins Univ. v. Datascope Corp., 543 F.3d 1342, 1349 n.3 (Fed.
`
`Cir. 2008); see also Ex-1077, 22:3-20 (regulatory equivalence does not equate to
`
`copying).
`
`
`
`23
`
`

`

`regions.” POR, 66; see also Ex-1821, 233:10-23, 236:16-21 (
`
`
`
`).
`
`PO’s “copying” arguments ignore other important differences between
`
`GuideLiner and Telescope. Ex-1830 ¶3-27. PO invites the Board to look at the
`
`parties’ products “side by side,” while ignoring the alleged “small differences,”
`
`including “Telescope’s hydrophilic coating and round push wire.” POR, 65-66.

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket