throbber
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
`
`I.
`
`GENERAL INFORMATION
`
`Device Generic Name:
`
`Coronary Stent System
`
`Device Trade Name:
`
`DriverTM Over-The-Wire, Rapid Exchange and Multi-
`Exchange Coronary Stent System
`
`6%-
`
`Applicant's Name and Address:
`
`Medtronic Vascular
`3576 Unocal Place
`Santa Rosa, CA 95403
`USA
`
`Date(s) of Panel Recommendation: None
`
`Premarket Approval Application (PMA) Number: PO30009
`
`Date of Notice of Approval to Applicant:
`
`October 1,2003
`
`11.
`
`INDICATIONS FOR USE
`
`The Medtronic DriverTM Over-The-Wire, Rapid Exchange, and Multi-Exchange Coronary
`Stent Systems (hereinafter called the DriverTM Coronary Stent System) are indicated for
`improving coronary luminal diameter in patients with symptomatic ischemic heart disease
`due to discrete de novo or restenotic lesions with reference vessel diameters of 3.0 - 4.0
`mm and 5 30 mm in length using direct stenting or pre-dilatation. Outcome beyond 270
`days for thispermanent implant is unknown at present.
`
`>
`
`111. DEVICE DESCRIPTION
`
`The Medtronic DriverTM Coronary Stent System is comprised of two components: the
`Stent and the Delivery System. The DriverTM stent is identical despite which delivery
`system it is mounted on.
`
`The DriverTM stent consists of a series of 1 .O mm segments that are constructed from a
`continuous toroid ring manufactured from a Co-Ni-Cr-Mo alloy conforming to ASTM F
`562-00. The ring is formed into alternating upper and lower crowns with 10 radiused
`crowns per end connected by axial struts for a total of 20 crowns and 20 axial struts in a
`zigzag pattern.
`
`Page 1
`
`Medtronic Exhibit 1823
`
`

`

`The use of a new alloy facilitates the implementation of thinner struts without affecting
`the radial strength or radiopacity of the stent. Unlike 3 16L stainless steel, the new alloy
`does not contain significant levels of iron (1 .O percent by weight maximum), thereby
`enhancing MRI compatibility. In addition, the radiopacity of the Co-Ni-Cr-Mo alloy is
`greater than that of 3 16L stainless steel, due primarily to increased levels of molybdenum.
`As such, the radiopacity of 3 16L stainless steel stents is maintained with the Co-Ni-Cr-
`Mo alloy stents even with thinner stent struts
`
`The stent is mounted on one of three delivery systems. Each delivery system provides a
`means for delivering the stent through the coronary vasculature and, once in the desired
`location, expands the stent through balloon inflation. The three delivery systems
`available with the DriverTM stent are the Over-The-Wire (OTW), Rapid Exchange (RX)
`and Multi Exchange (MX) Delivery System.
`
`All delivery systems have a minimal amount of working length extending beyond the
`stent on each side. The inner lumen of the catheter is designed to accommodate a
`maximum guidewire diameter of 0.0 14 inches.
`
`Table 1 provides the product labeling specifications for the Medtronic DriverTM Coronary
`Stent and the three Delivery Systems
`
`Table 1: Product Labeling Specifications for the Medtronic DriverTM Coronary
`S tent Svs terns
`
`Nominal
`Pressure
`(atm)
`
`Rated
`Burst
`Pressure
`(atm)
`
`Medtronic DriverTM
`Over-the- Wire
`Coronary StenGystem
`
`Over-the-
`Wire
`
`Medtronic DriverTM
`Rapid Exchange
`Coronary Stent System
`
`Rapid
`Exchange
`
`Medtronic DriverTM
`Multi-Exchange MX
`Coronary Stent System
`
`Multi-
`Exchange
`
`3.0
`3.5
`4.0
`
`3 .O
`3.5
`4.0
`
`3.0
`3.5
`4.0
`
`Inner
`Diameter
`(inches)
`.056
`
`.056
`
`.064
`
`9
`12
`15
`18
`24
`30
`9
`12
`15
`18
`24
`30
`9
`12
`15
`18
`24
`30
`
`9
`
`16
`
`16
`
`16
`
`Page 2
`
`Medtronic Exhibit 1823
`
`

`

`IV.
`
`CONTRAINDICATIONS, WARNINGS AND PRECAUTIONS
`
`Contraindications
`
`The Driver Coronary Stent System is contraindicated for use in:
`Patients in whom antiplatelet and/or anticoagulation therapy is
`contraindicated.
`Patients who are judged to have a lesion that prevents complete inflation of
`an angioplasty balloon.
`
`k
`
`Warnings and Precautions
`
`A list of warnings and precautions can be found in the device labeling.
`
`V.
`
`ALTERNATIVE PRACTICES AND PROCEDURES
`
`Patients with early coronary artery disease receive exercise, diet, and drug therapy as
`appropriate. If the disease progresses, alternative practices specific to the treatment of
`coronary artery disease include: percutaneous transluminal coronary angioplasty (PTCA),
`drug therapy (e.g., thrombolytic agents, antiplatelet agents, and anticoagulant agents),
`Coronary Artery Bypass Graft Surgery (CABG), and stenting with other commercially
`available stents.
`
`VI. MARKETING HISTORY
`
`The Driver Coronary Stent System has been approved for commercial distribution in the
`European Union, Canada, China, Australia and Singapore. The Driver Coronary Stent
`System has not been withdrawn from marketing for any reason relating to the safety and
`effectivenesuf the device.
`
`*
`
`VII. SUMMARY OF PRECLINICAL STUDIES
`
`Biocompatibility
`
`All biocompatibility testing was performed in accordance with;
`
`0
`
`Guidance for the Submission of Research and Marketing Applications for
`Interventional Cardiology Devices; published by the Interventional Cardiology Devices
`Branch, Division of Cardiovascular, Respiratory and Neurology Devices, Office of
`Device Evaluation in May 1995.
`
`0
`
`IS0 10993, Biological Evaluation of Medical Devices
`
`Page 3
`
`Medtronic Exhibit 1823
`
`

`

`The following biocompatibility tests were conducted and passed on the DriverTM
`Coronary Stent System:
`
`Biocompa
`Component
`Stent Raw
`Material
`
`d i t y Testing on (
`Biological Effect
`Cytotoxicity
`
`Sensitization
`
`k
`
`Irritation
`
`Systemic Toxicity
`
`Mutagenicity /
`Genotoxicity
`
`e
`
`Implantation /
`Subchronic toxicity
`
`Hemocompatibility
`
`Skin Sensitization Kligman
`Maximization test-IS0
`
`Intracutaneous Injection
`Test-IS0
`
`System Injection test-IS0
`
`~~
`
`I-Ni-Cr-Mo alloy (DriverTM Stent)
`Test
`Criterion
`L929 MEM Elution Test-
`01-2293-GI
`IS0
`Results must indicate a
`grade of 2 or lower.
`01-2293-GI1
`No dermal inflammatory
`response greater than the
`control.
`01-2293-GS
`No greater adverse
`reactions or responses
`when compared to the
`controls.
`0 1 -2293-G9
`No greater adverse
`reactions or responses
`when compared to the
`controls.
`01-2293437
`The test article must not
`induce a significant
`increase in the number of
`micronucleated cells when
`compared to controls.
`0 1-229346
`The test article must not
`induce a significant
`increase in reversion in
`the genomes of the
`organisms tested, when
`compared to controls.
`01-2293-G5
`Test article extraEts must
`not be considered
`mutagenic when
`compared to controls.
`0052D640 P1056
`No capsule formation or
`other adverse reaction to
`the implanted test article.
`01-2293-G3
`The test article must not
`induce complement
`activation when compared
`to controls.
`
`Rodent Bone Marrow
`Micronucleus Assay-IS0
`
`Chromosomal Aberration
`Assay-IS0
`
`S. typhimurium/ E.coli
`Reverse Mutation Assay-
`IS0
`
`Biocompatibility Implant
`Study
`
`C3a Complement Activation
`Assay
`
`Results
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Page 4
`
`Medtronic Exhibit 1823
`
`

`

`Unactivated Partial
`Thromboplastin Time
`Assay-IS0
`
`~~~~
`
`In Vitro Hemocompatibility
`Assay-IS0
`
`Bc
`
`Pyrogenicity
`
`Rabbit Pyrogen Test
`(Material Mediated)-IS0
`
`Hemocompatibil ity
`
`C3a Complement Activation
`Assay
`
`Implantation /
`Subchronic toxicity
`
`IS0 Muscle Implantation, in
`3 Rabbits (6 weeks),
`whistopathology
`
`0 1-2293-G4
`The test article must not
`significantly effect the
`clotting time of human
`plasma when compared to
`controls.
`
`01-2293-G2
`The test article must not
`adversely effect selected
`hematological parameters
`of human blood when
`compared to controls.
`0 1 -2293-G 10
`No single animal must
`exhibit an increase of
`0.5"C or more above its
`baseline temperature.
`01T 10878 01
`The test article must not
`induce complement
`activation when compared
`to controls.
`OIC 12404 00
`No capsule formation or
`other adverse reaction to
`the implanted test article.
`
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Component
`Stent and
`Delivery
`System
`
`Type
`Cytotoxicity
`
`Sensitization
`
`Irritation
`
`Systemic Toxicity
`
`Test
`L929 MEM Elution Test-
`IS0
`
`IS0 Sensitization Study in
`the Guinea Pig
`
`IS0 Acute Intracutaneous
`Reactivity Study in the
`Rabbit
`
`IS0 Acute Systemic
`Toxicity
`in the Mouse
`
`Criterion
`OIC 11441 00
`Results must indicate a grade of
`2 or lower.
`01C 11441 00
`No dermal inflammatory
`response at the test sites greater
`than that seen in the control.
`01C 11441 00
`No adverse reactions or
`responses when compared to
`the controls.
`01C 11441 00
`No greater adverse reactions or
`responses when compared to
`the controls.
`
`Res u Its
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Page 5
`
`Medtronic Exhibit 1823
`
`

`

`Mutagenicity I
`Genotoxicity
`
`Genotoxicity: Bacterial
`Reverse Mutation (Ames)
`Test, Salmonella and E. coli
`Strains
`
`k
`
`Hemocompatibility
`
`Hemolysis
`
`White Blood Cell
`Morphology Study -In vitro
`
`Plasma Recalcification Time
`Coagulation Study (With
`Protocol Amendment I, 11 &
`III)
`voo20-000
`
`IS0 Rabbit Pyrogen Study
`(Material Mediated)
`
`Hemocompatibility
`
`Pyrogenicity
`
`01C 12359 00
`The test article must not induce
`a significant increase in
`reversion in the genomes of the
`organisms tested, when
`compared to controls.
`Test article extracts must not be
`considered mutagenic when
`compared to controls.
`01C 11441 00
`The mean hemolytic index must
`be 2% or less.
`01T 10878 00
`The number of while cells with
`morphological abnormalities
`must be -3%.
`01T 10878 00
`The test article must not exhibit
`a significant difference in
`recalcification times and effects
`on fibrin clot formations when
`comoared to the control.
`OIC 11441 00
`No single animal must exhibit
`an increase of 0.5"c or more
`above its baseline temperature.
`
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Biocompa
`Component
`Delivery
`System
`
`bility Testing on 1
`Type
`Cytotoxicity
`
`Hemocompatibility
`
`! Multi-Exchange Delivery System
`Test
`Criterion
`L929 MEM Elution Test-
`02T0757800
`Results must indicate a grade of
`I S 0
`2 or lower.
`02T0757800
`The mean hemolytic index must
`be 2% or less.
`
`Hemolysis
`
`C3a Complement Activation
`
`Plasma Recalcification
`
`02T0626800 --
`
`The test article must not induce
`complement activation when
`compared to controls.
`02T0626800
`The test article must not exhibit
`a significant difference in
`recalcification times when
`compared to the control.
`
`Thromboresistance
`
`02T0692000
`Test articles are evaluated for
`the presence of thrombus as
`comoared to the controls.
`
`Resu I ts
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`Pass
`
`13
`
`Page 6
`
`Medtronic Exhibit 1823
`
`

`

`Irritation
`
`IS0 Acute Intracutaneous
`Reactivity
`
`Systemic Toxicity
`
`IS0 Acute System Toxicity
`
`Genotoxicity
`
`Ames
`
`Physicochemical
`
`Physicochemical (aq)
`
`Physicochemical (Non-aq)
`
`Sensitization
`
`IS0 Sensitization
`
`02T0626800
`02T0692000
`No greater adverse reactions or
`responses when compared to
`the controls.
`02T0626800
`02T0692000
`No greater adverse reactions or
`responses when compared to
`the controls.
`02T0626800
`02T0692000
`Test article extracts must not be
`considered mutagenic.
`02T0757800
`Non-volatile residue: < 15mg
`Residue on Ignition: <5mg
`Heavy Metals: 5 lppm
`Buffering Capacity: <lorn1
`02T0757800
`No acceptance criteria have
`been determined. For
`characterization onlv.
`02T0332300
`02T0692000
`96T0594800
`97C0609000
`OOC 10 14200
`No dermal inflammatory
`response at the test sites greater
`than that seen in the control.
`
`Pass
`Pass
`
`Pass
`Pass
`
`Pass
`Pass
`
`Pass
`
`Pass
`
`Pass
`Pass
`Pass
`Pass
`Pass
`
`The Medtronic DriverTM Stent Delivery System successhlly passed all the above
`referenced biocompatibility testing. The DriverTM Over-The-Wire Delivery System was
`not tested since the materials found in it are identical to the materials found in the two
`systems (Rapid Exchange and Multi-Exchange) which passed all tests noted above.
`
`->
`
`Sterilization
`The DriverTM Stent System will be sterilized using Electron Beam sterilization and has
`been validated per AAMIhSO 11 137: Sterilization of health care products - Requirements
`for validation and routine control - Radiation sterilization.
`
`Results obtained from the sterilization studies show that the device satisfies a minimum
`Sterility Assurance Level (S.A.L.) of
`
`In Vitro Testing
`
`In vitro bench testing of the Medtronic DriverTM Stent Delivery Systems was conducted
`in accordance with the FDA ODE “Guidance for the Submission of Research and
`Marketing Applications for Interventional Cardiology Devices: PTCA Catheters,
`
`Page 7
`
`Medtronic Exhibit 1823
`
`

`

`Atherectomy Catheters, Lasers, Intravascular Stents”, May 1995. The relevant tests
`outlined in the guidance were conducted to demonstrate the functional performance
`characteristics of the device.
`
`The following is a brief summary of the bench testing conducted on the device:
`
`Material Analysis (stent)
`This study was conducted to verify that DriverTM stents are produced from material that
`conforms to the chemical composition requirements of ASTM F562-00, “Standard
`Specification for Wrought Cobalt-3 5 Nickel-20 Chromium- 10 Molybdenum Alloy for
`Surgical Implant Applications”. Data for this test was collected from Co-Ni-Cr-Mo alloy
`tubes as processing of this material into a completed stent does not change the overall
`composition of the material. The material used to manufacture the DriverTM stent
`conformed to ASTM F562-00.
`
`Scanning Electron Microscopy (SEM)
`SEM analysis was conducted at 200x magnification to identify trace contaminants that
`may be present on the DriverTM stent surface. No foreign material was observed on any
`of the production lot stents tested. Several small particles inherent to the base stent
`material were observed on the surface, which most probably were exposed during stent
`electropolishing during the manufacturing procedure.
`
`Mechanical Properties
`This testing was conducted to characterize the following mechanical properties of
`annealed Co-Ni-Cr-Mo alloy: 0.2% offset yield strength, ultimate tensile strength, percent
`elongation, and reduction of area. Five (5) tensile bars were machined from barstock
`conforming to ASTM F562-00. The bars were then vacuum annealed to product an
`ASTM grain size of 8.5 in the gage sections. The bars were tensile tested to failure while
`engineering stress and strain were continuously recorded. The offset yield strength,
`ultimate tensile strength, percent elongation, and percent reduction of area for the
`DriverTM stent material are in conformance with ASTM F562-00.
`*
`
`Corrosion
`This test was conducted to evaluate the relative susceptibility to localized corrosion of
`finished Driver stent components. The analysis was based upon ASTM G61 , “Standard
`Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements for
`Localized Corrosion Susceptibility of Iron, Nickel, or Cobalt-Based Alloys”. The results
`indicated that the DriverTM stent possesses a high resistance to the onset of localized
`corrosion.
`
`Stent Free-area
`The Percent Free-Area is a value calculated using Driver stent nominal dimensional
`values and is based in the ratio of stent area to the area of the vessel. Metal to artery
`percentage ratios ranged from 15.4% to 20.5% for 3.0mm to 4.0mm stent, respectively.
`
`Page 8
`
`Medtronic Exhibit 1823
`
`

`

`Length Change vs. Diameter
`This test was performed to determine the change in length as a function of stent internal
`diameter. The testing was performed on complete stent delivery systems that had been
`subjected to all manufacturing and sterilization procedures. Each stent was measured for
`initial length while still on the delivery system, then deployed to nominal pressure of 9
`ATM and measured after removal from the delivery system. Length change ranged from
`-3% to +4% and met the acceptance criteria.
`
`rk
`
`Stent Uniformity
`This test was conducted to demonstrate that the deployed internal diameter of the Driver
`stent is consistent with labeling. The testing was performed on complete stent delivery
`systems that had been subjected to all manufacturing and sterilization procedures. Each
`stent was deployed to nominal pressure of 9ATM. The stent inner diameter was
`measured at each end. All Driver stents met the uniformity expansion specification of +/-
`10% from nominal labeled diameter to 9ATM. The results indicated that the Driver stent
`expands uniformly at all diameters and maintains this uniformity upon withdrawal of the
`balloon.
`
`Radial Strength
`This test is conducted to determine and graphically represent the change in stent internal
`diameter as a function of circumferential pressure and to determine the pressure at which
`deformation is no longer completely reversible for the Driver stent. Fifteen (1 5) 3.0mm
`and fifteen (1 5 ) 4.0mm stents were subjected to all stent-manufacturing procedures. The
`stents were deployed to nominal pressure and removed from the delivery system. The
`stents were placed into a sleeve approximately lmm larger then the stent diameter. A
`vacuum was then applied and outer diameter measurements taken at various pressures.
`All samples maintained a minimum of at least 50 percent of the original stent diameter
`after a 254mm Hg pressure was radially applied. All samples met the acceptance criteria.
`
`Bent Radial Fatigue Testing
`This test was conducted to determine the radial fatigue integrity on a bend of a Driver
`stent under afcelerated cycling designed to test expected in vivo conditions. Eighty-two
`(82) Driver stents were deployed up to a maximum of 5.lmm in distendable tubing and
`attached to fatigue equipment that underwent a 0.005” distention. There were no stent
`failures noted after 420 million cycles which is equivalent to 10 years in vivo.
`
`Stent Recoil
`This test was conducted to quantify the amount of elastic recoil. Fifteen (1 5) stent
`delivery systems of each length and diameter (excluding 30 12 and 40 12) were subjected
`to all manufacturing and sterilization procedures. The stent delivery system was inflated
`to nominal pressure (9ATM) and the stent was removed allowing for recoil to occur. The
`inner diameter at each end of the stent was recorded. Recoil was calculated subtracting
`the recoiled stent inner diameter from the pre-recoil inner diameter. Average recoil
`ranged from 0.002 to 0.004 inches. All stents met the acceptance criteria after balloon
`inflation to nominal pressure.
`
`Page 9
`
`Medtronic Exhibit 1823
`
`

`

`MRI
`Driver Stents were tested to ensure they did not react to a magnetic field in such a manner
`that might pose a clinical issue. The stents should neither deflect nor produce appreciable
`susceptibility image artifacts in the presence of a magnetic field.
`
`Testing for image artifacts created by the Driver stents in a 1.5 Tesla magnetic field
`produced minimal artifact susceptibility. This result was expected in the light of the lack
`of iron in the material composition. The mean deflection angle measured for the Driver
`stent in association with testing performed using a 3.0 Tesla field was 3".
`
`e,
`
`In conclusion, the Driver stents tested did not cause distortion of the magnetic field
`during the MRI procedure. If an implant deflects less than 45", then the magnetically
`induced deflection force is less than the force on the implant due to gravity. Accordingly,
`since the stents tested had a mean deflection of 3", a negligible risk of stent movement in
`a 3.0 Tesla MRI field was indicated.
`
`Stent Expansion
`This test is conducted to determine if the plastic deformation experienced by the stent
`when expanded from the compressed profile to the final maximum deployed diameter can
`produce crack initiation for the Driver stent. Fifteen (1 5) samples from each length and
`diameter (excluding the 30 12 and 40 12) were deployed to their largest possible diameters
`by inflating each delivery system to balloon failure. Each stent was examined at 45X
`magnification for potential cracks. All samples met the acceptance criteria with no
`visible cracks or notches.
`
`Dimensional Verification (stent)
`This test is conducted to verify that dimensional and visual specifications for the Driver
`stent can be achieved. Fifteen (1 5 ) samples from each length within 3.0 and 3.5mm
`diameters were measured with a toolmaker's microscope. All samples met the
`dimensional specifications.
`
`Maximum Pressure (burst test)
`This test is conducted to demonstrate that the delivery system (with mounted stent) will
`not experience balloon, shaft, proximal adaptation or proximal/distal seal loss of integrity
`at or below the pressure required to expand the stent to its labeled diameter. Stent
`delivery systems that had been subjected to all manufacturing and sterilization procedures
`were pressurized to 9Opsi with pressure held for 15 seconds and released for 3 seconds.
`The cycle was then repeated, increasing inflation pressure by 15psi each cycle until
`failure. The shaft of all delivery systems remained intact throughout the test and resulted
`in a labeled rated burst pressure of 16 ATM.
`
`t
`
`Stent Diameter vs. Pressure (stent compliance)
`This test is conducted to quantify the stent inner diameter as a function of balloon
`inflation pressure. Each delivery system was inflated to 6ATM and the outer diameter at
`the distal and proximal point of the stent was measured. This was repeated at 1ATM
`increments. Once the OD measurement was taken at 9ATM, the system was deflated, the
`
`1 '3
`
`Page 10
`
`Medtronic Exhibit 1823
`
`

`

`stent was removed and the inner diameter was measured. The stent was placed back on
`the balloon and continued 1 ATM increments until balloon failure. All stent delivery
`system met the acceptance criteria requirements.
`
`Bond Strength
`This test was performed to provide data to support the bond strength specifications for the
`Driver Stent Delivery Systems. All bonds were trimmed to manageable lengths and
`loaded into the tensile tester and pulled to failure. All samples met the acceptance
`criteria.
`
`a
`
`Diameter and Profile
`This test was conducted to provide dimensional data for the tip, balloon bond, distal shaft,
`intermediate shaft, and proximal shaft of the Driver Coronary Stent System. All catheter
`diameters and profiles met all acceptance criteria.
`
`Balloon Deflatibility
`This test is conducted to provide assurance that the deflated delivery system balloons will
`not interfere with the Driver deployed stent in vivo. Complete stent delivery systems
`were inflated to nominal pressure (9ATM) with a replica vessel placed over the
`premounted stent. The stent delivery system was then deflated and vacuum was drawn,
`allowing the stent to release from the delivery system. All samples passed the acceptance
`requirement of either gravity or vessel release.
`
`Balloon Deflation Time
`This test is conducted to quantify the deflation time of the Driver stent delivery system.
`The complete stent delivery systems were inflated to 9 ATM. Simultaneously, a vacuum
`was pulled and a timer started. The amount of time to deflate the balloon from nominal
`pressures was recorded. The results indicated that all Driver Coronary Stent Systems
`could be deflated within a clinically acceptable time.
`
`Balloon Fatigue
`This report Symmarizes the testing done to satisfy all regulatory requirements for balloon
`fatigue on the Driver Coronary Stent System. The stent delivery systems were repeatedly
`inflated to demonstrate the balloon could sustain multiple repeated inflations. The stent
`delivery systems were capable of at least 20 inflations to rated burst pressure.
`
`Crossing Profile
`This test is conducted to verify the crossing profile as per label claims and to discuss the
`clinical relevance of this test. The stent delivery systems are inserted through a crossing
`profile block until the smallest block that the stent could pass through was determined.
`All samples passed through the 0.056” hole gauge with a maximum observed crossing
`profile of 0.049”.
`
`Tortuous Path
`Stent retention is evaluated using a test which involves tracking the device back and forth
`five times through a track fixture that has four bends of varying tortuosity. After tracking,
`
`Page 11
`
`Medtronic Exhibit 1823
`
`

`

`the force required to remove the stent from the delivery system is measured and the initial
`peak force observed during testing is recorded. Devices are required to meet a pre-
`determined retention force with a 95% confidence and 95% reliability.
`
`In vivo (animal) studies
`Medtronic has conducted six GLP animal studies utilizing the Driver Coronary Stent
`Systems in accordance with 2 1 CFR 0 5 8. These studies evaluated safety, efficacy and
`overall device performance; summaries of these studies are included in the following
`table:
`
`hmrr
`Study
`#
`FS62
`
`FS65
`
`ry of GLP Studies Perform
`Stent Design
`(Delivery System)
`3.04.5mm* dia. x 18mm Ingth.
`(OTW and RX Delivery
`Systems) Control = S7
`3.0-3.5mm dia. x 18mm Ingth.
`(OTW Delivery System)
`Control = S7
`
`FS66
`
`3.0-3.5mm dia. x 18mm Ingth.
`(OTW Delivery System)
`Control = S7
`
`FS67
`
`3.0-3.5mm dia. x 18mm Ingth.
`(OTW Delivery System)
`Control = S7
`
`FS8 1
`
`3.0-3.5mm dia. x 18mm Ingth.
`(OTW Delivery System)
`Control = S7
`
`FS87
`
`Endpoints
`Acute delivery, mechanical
`performance, and chronic
`vascular response.
`Acute delivery and deployment
`performance, angiographic
`estimates of in-stent restenosis at
`sacrifice and histomorphometric
`analysis of histology.
`Acute delivery and deployment
`performance, angiographic
`estimates of in-stent restenosis at
`sacrifice and histomorphometric
`analvsis of histolow.
`Acute delivery and deployment
`performance, angiographic
`estimates of in-stent restenosis at
`sacrifice and histomorphometric
`analysis of histology.
`Acute delivery and deployment
`performance, angiographic
`estimaes of in-stent restenosis at
`sacrifice and histomorphometric
`analysis of histology.
`3.5mm dia. x 18mm Ingth.
`Acute delivery and mechanical
`performance of Over-the-Wire
`(OTW and RX Delivery
`and Rapid Exchange stent
`Systems)
`Control = S7
`systems.
`* The 4.5mm Driver stent was included in this pre-clinical study but is not part of this PMA
`Application
`
`d
`
`# of
`Animals
`
`# of
`Stents
`
`FOIIOW-UP
`Duration
`
`8
`
`6
`
`4
`
`6
`
`6
`
`1
`
`19
`
`28-days
`
`16
`
`28-days
`
`10
`
`90-days
`
`16
`
`180-days
`
`180-days
`
`Acute
`
`Page 12
`
`Medtronic Exhibit 1823
`
`

`

`VIII.
`
`POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
`
`The Medtronic Driver DeNovo and Restenotic Registry enrolled 298 patients in a non-
`randomized, multi-center study. These patients form the basis of the observed events
`reported in the following section.
`
`Driver DeNovo and Restenotic Registry
`
`A total of 298 patients were enrolled in a multi-center registry to evaluate the safety and
`efficacy of the Medtronic Driver Coronary Stent System for treatment of symptomatic
`coronary artery disease.
`
`rk
`
`The primary endpoint of MACE at 180 days was compared to an objective performance
`criterion (OPC) of 15% based on a pooled MACE rate derived by pooling the data from
`the Bard XT Stent EXTRA RCT, Medtronic Micro Stent I1 SMART RCT, and Medtronic
`BeStent I (BEST) & BeStent I1 Registries.
`
`Adverse events reported during the first six and nine months are shown in the table
`below. A total of 25 of 298 patients (8.4%) who received the Medtronic Driver stent
`experienced one or more adverse events during the nine months of follow-up.
`
`A total of 4 of the 298 (1.3%) patients who received the Driver stent died during the
`clinical study. These out-of-hospital deaths were all non-cardiac related: one secondary
`to ovarian cancer at 43 days post-procedure, one secondary to a brain tumor at 136 days
`post-procedure, one due to acute respiratory failure at 242 days post-procedure and one
`non-specified cancer death at 268 days post-procedure. There were no instances of stent
`thrombosis during the first 270 days. The incidence of vascular complications was 3.4%
`(10/298). The rate of bleeding complications was 2.3% (7/298).
`
`There were no (0/298) delivery or device failures reported.
`Prjpcipal Adverse Events Through 180 & 270-Days
`Driver DeNovo and Restenotic Registry
`%, (Number) [95% confidence interval]
`
`Medtronic Driver Stent (N=298)
`180-Day Results
`7.7% (23)[5.0%,11.4%]
`5.7% (17)[3.4%,9.0%]
`2.0% (6) [0 .7%, 4.3%]
`5.7% (1 7)[3.4%,9.0%]
`1.7% (5)[O. 5%, 3.9%)
`4.0% (12)[2.1%,6.9%]
`0.7% (2)[0.1%,2.4%]
`0.0% (0)[0.0%,1.2%]
`
`Medtronic Driver Stent (N=298)
`270-Day Results
`8.4% (25)[5.5%,12.1%]
`5.7% (1 7)[3.4% ,9 .O%]
`2.7% (8){1.2%,5.2%]
`10.1% (30) [6.9%,14.1%]
`I .7% (5)[0.5°h,3.9%)
`8.4% (25) [5.5%,12.1%]
`1.3% (4) [0.4%, 3.4%]**
`0.0% (0)[0.0%,1.2%]
`
`I.
`
`omplication"
`Adverse Event§
`In-Hospital
`Out-of-Hospital
`MACE
`In-Hospital
`Out-of-Hospital
`
`In-Hospital
`
`Page 13
`
`Medtronic Exhibit 1823
`
`

`

`I n-Hospital
`Out-of-Hospital
`Non Q-wave MI
`In-Hospital
`
`In Hospital -PTCA
`In Hospital-CABG
`Out-of-Hospital PTCA
`d I
`Out-of-HosDital CABG
`-Emergent CABG
`In-Hospital
`Out-of-Hospital
`Stent Thrombosis
`I n-Hospital
`Out-of-Hospital
`Bleeding (procedural transfusion)
`I n-Hospital
`Out-of-Hostital
`CVA
`In-Hospital
`Out-of-Hospital
`Vascular Complications
`In-Hospital
`Out-of-Hospital
`Stent Delivery Failures
`
`0.7% (2)[0.1%,2.4%]
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.0%,1.2%]
`0.0% (O)[O.O%, 1.2%]
`1.7% (5)[0.5%,3.9%]
`1.7% (5)[0.5% ,3.9%]
`0.0% (0)[0.0%,1.2%]
`3.4% (10) [1.6%,6.1%]
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.0%,1.2%]
`2.7% (8)[1.2%,5.2%]
`0.7% (2) [0.1%,2.4%]
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.0%,1.2%]
`2.3% (7)[0.9%,4.8%]
`2.0% (6)[0.7%,4.3%]
`0.3% (1)[0.0%,1.9%]
`0.3% (1)[0.0%,1.9%,1
`0.0% (0)[0.0%,1.2%]
`0.3% (1)[0.0%,1.9%]
`3.4% (1 0)[1.6%,6.1 Oh]
`2.7% (8)[1.2%,5.2%]
`0.7% (2)[0.1%,2.4%]
`0.0% (0) [O.O%, 1.2%]
`
`1.3% (4) [0.4%,3.4%]**
`0.0% (0)[0.0%,1.2%]
`0.0% (O)[O.O%, 1.2%]
`0.0% (0)[0.0%,1.2%]
`1.7% (5)[0.5%,3.9%]
`1.7% (5)[0.5%, 3.9%]
`0.0% (0)[0.0%,1.2%]
`7.0% (21) [4.4%, 10.6%)
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.00/0,1.2%]
`6.4% (19) [3.9%, 9.8%]
`0.7% (2) [0.1%,2.4%]
`0.0% (O)[O.O%, 1.2%]
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.0%,1.2%]
`0.0% (0)[0.0%,1.2%]
`2.3% (7)[0.9%,4.8%]
`2.0% (6)[0.7%,4.3%]
`0.3% (1)[0.0%,1.9%]
`0.3% (1)[0.0%,1.9%]
`0.0% (O)[O.O%, 1.2%]
`0.3% (1)[0.0%,1.9%]
`3.4% (1 0)[ 1.6% ,6. 1°/o]
`2.7% (8)[1.2%,5.2%]
`0.7% (2)[0.1%,2.4%]
`0.0% (0) [O.O%, 1.2%]
`
`*Complications are based on patient totals. Seven patients had multiple adverse events: one patient hac
`3 vascular complications, four patients had 2 vascular complications), one patient had 1 non-Q MI and 1
`bleeding. complication), and one patient had1 non-Q MI and 1 vascular complication.
`
`'Adverse Event= Death, Q or Non-Q wave MI, Emergent CABG, Stent Thrombosis, CVA, Bleeding
`Complications, and Vaucular Complications.
`*
`**All four deaths were non-cardiac.
`Definitions for the terms used in Table 2 are found in the footnotes to Table 3.
`
`Potential adverse events that may be associated with the use of a coronary stent in native
`coronary arteries (including those listed in Table 2) are listed below in order of severity:
`
`e
`e
`
`e
`e
`e
`e
`
`e
`
`Death
`Emergency Coronary Artery Bypass Graft Surgery (CABG)
`Stroke/Cerebrovascular Accidents
`Cardiac tamponade
`Stent thrombosis or occlusion
`Total occlusion of coronary artery
`Acute myocardial infarction
`
`Page 14
`
`Medtronic Exhibit 1823
`
`

`

`,
`
`Restenosis of stented segments
`Perforation
`Arrhythmias, including ventricular fibrillation & ventricular tachycardia
`Dissection
`Emboli, distal (air, tissue or thrombotic emboli)
`Stent embolization
`Hemorrhage requiring transfusion
`Pseudoaneurysm, femoral
`Spasm
`Myocardial ischemia
`HypotensiodHypertension
`Allergic reaction to drugskontrast mediudstent material
`Peripheral ischemia
`Peripheral nerve injury
`Infection and pain at the insertion site
`Hematoma
`
`IX.
`
`SUMMARY OF CLINICAL STUDIES
`
`Purpose
`
`The purpose of the Driver Registry was to evaluate the safety and efficacy of the
`Medtronic Driver stent for the treatment of single de novo or restenotic post-PTCA (non-
`stented) lesions in native coronary arteries.
`
`Conclusions
`
`The Driver Registry demonstrated the 1 SO-day and 270-day safety and efficacy of the
`Driver stent for treatment of patients with de novo or restenotic lesions in native coronary
`arteries.
`
`Design e
`
`-?-
`
`A prospective, multi-center non-randomized study was conducted in 23 North American
`clinical sites enrolling 298 patients. Patients were 18 years of age or older with clinical
`evidence of ischemic heart disea

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket