`An expert offers a primer on popular medical device coatings and their applications.
`
`Kevin Guen ther
`
`A catheter with a lubricious coating.
`
`In the world of low friction, iubricious coatings for medical devices are diverse and complex. There
`are many different types of coatings, each with their own advantages, disadvantages, and
`applications. This is a basic overview of the most popular medical device coatings and what can be
`expected from each type.
`
`Lubricious coatings are most commonly distinguished by their degree of lubricity, or the amount of
`reduction in friction they provide. To understand what these measurements mean, we must first
`learn how these measurements are obtained. In the medical device industry, friction is measured by
`
`what is known as a pinch test. In a pinch test, a coated device is secured on an instrument that pulls
`the device between the jaws of a clamp which produces a load on the part. While clamped, the coated
`
`_
`
`
`Page 1
`
`Teleflex Ex. 2055
`Medtronic v. Teleflex
`IPR2020-00131
`
`
`
`part is pulled at a fixed speed for a fixed distance, and the resistance to the pull is measured in grams
`of pull force. The grams of pull force needed to overcome the friction produced by the clamps divided
`by the clamp force yields the coefficient of friction (CoF). CoF = Forcepun/ Forcedamp
`Visit Harland Medical Systems at Booth #1615 during MD&M West in Anaheim,_CA,
`Februag 7-9, 2017 (http:/lmdmwest.mddionline.coml?
`_mc=arti_x_mddir_le_tsnr_thibault_mddir_med_44_x-harlandmedicalsystems)
`
`
`
`
`
`A BOO-gram clamp force (Force Clamp) is a commonly used setting. The higher the pull force needed,
`the higher the CoF and the lower the lubricity of the surface. The lower the pull force, the lower the
`CoF and the higher the lubricity.
`
`When measuring friction between two surfaces, another characteristic
`is that of static friction pull force vs. dynamic friction pull force. Static
`friction is the force needed to break the initial friction force 50 that the
`two surfaces can move relative to each other. Dynamic friction pull
`force is that force needed to maintain the motion once it has been
`initiated. Static friction (sometimes referred to as "stiction") is almost
`always significantly higher than dynamic friction.
`
`Now that we have an understanding of how the COP is obtained and the
`
`significance of the different amounts of grams of pull force required, we
`can take a look at some commonly used coatings and compare them.
`
`Before we look at the different types of coatings, let's begin by looking
`
`at the properties of an uncoated sample to provide a reference point. A
`very common material used in medical devices is PEBAX® (a registered
`trademark of Arkema, lnc.), the brand name of polyether block amide.
`When uncoated, PEBAX® (hardness of 55 ShoreD) generally produces
`
`
`
`pinch test results of about 500 to 600 grams of pull force. Assuming a
`BOO-gram clamp force was used, the COP would be (1.0—1.2), a very low
`level of lubricity. Other uncoated catheter materials of construction
`such as nylon 12, polyethylene, and others exhibit similar CoF. Such a high level of friction would
`make many of today's catheter-delivered minimally invasive procedures nearly infeasible.
`
`A friction test system to
`measure lubricity.
`
`Device manufacturers have turned to a variety of coatings to provide the surface lubricity not
`
`offered by the bare catheter material, The first class of coating is silicone oil (siloxane). This type of
`coating has long been used to reduce surface friction on medical catheters, introducers, etc. It is
`generally applied by the manufacturer, and is relatively inexpensive. it offers a much reduced CoF, at
`about 0.13, and approximately 125—175 grams of pull force. However, devices coated with silicone oil
`can be difficult to apply as they are very mobile and can spread to other manufacturing areas where
`they are undesirable. For example, the presence of silicone oil on a surface can prevent other
`surface-enhancing coatings from adhering to the device surface. Once silicone oil is present, it can be
`very difficult to clean from a device or work surface. As a result, these coatings can be a major
`contaminant risk to the manufacturer.
`
`The second class of coating is PTFE (Polytetrafluoroethylene), known best by the brand name
`Teflon®, another popularly used material to reduce friction. It has been employed as a material of
`construction for catheter-type products or as an added coating for guidewires and other devices
`made from metal. PTFE typically exhibits between 150 and 200 grams of pull force, or a CoF of about
`0.3 , 0.4. On bare metals such as stainless steel and nitinol, PTFE offers performance similar to that
`of silicone oil. The main advantage of PTFE coatings is that they provide some level of lubricity while
`dry and do not require any source of moisture to perform.
`
`
`Page 2
`
`Teleflex Ex. 2055
`Medtronic v. Teleflex
`IPR2020-00131
`
`
`
`Hydrophilic coatings are the third class of coatings employed to reduce
`surface friction and enhance lubricity The term "hydrophilic" means
`that it is water loving. In other words, it readily wets out when exposed ‘
`to moisture. Most hydrophilic coatings consist of a loosely cross-linked
`polymer coating that will readily uptake liquid when exposed to a
`source. This coating that surrounds the device acts like a microscopic
`sponge and the water (or other liquid) provides excellent low friction
`process.
`characteristics as long as the coating remains wet. Compared to PTFE
`coatings, which are hydrophobic and repel water, hydrophilic coatings provide superior lubricity
`(lower CoF) when wet. Devices coated with these coatings demonstrate pull forces as low as 3 grams
`and CoF values as low as 0.01. In clinical use, devices coated with hydrophilic material demonstrate
`
`
`
`The dip coating and curing
`
`superior control, exhibit lower stiction (static friction) behavior, and require far less force to perform
`any given movement. Ranging from 3-25 grams of pull force, hydrophilic coatings offer the best CoF,
`averaging between .01 and .05. This high level of lubricity enables devices to navigate tortuous
`anatomical pathways while reducing tissue irritation and adding to patient comfort.
`
`Some types of hydrophilic coatings may generate particulates from their surface in use. For these
`applications where particle generation from a coating is critical. such as cardiovascular and
`neurovascular, the selected formulation must be one designed specifically for low particle
`
`generation. The only real limitation of this category of coatings is that they must be hydrated to
`perform.
`
`Commercially available hydrophilic coatings average between 8 and 12 grams of pull force that
`reduces friction over an uncoated surface by as much as 98%. This type of low—friction performance
`increases the device‘s ability to navigate through tortuous anatomical pathways, improves device
`control, reduces tissue damage, and adds to patient comfort.
`
`Advances in surface enhancements, especially low—friction hydrophilic coatings, have accelerated
`minimally invasive surgery's growth. These coatings now impact a variety of surgical specialties,
`including cardiovascular, orthopedic, urological, peripheral, and neurological procedures.
`
`Kevin Guenther is the methods director at Harland Medical Systems (http://directoryqmed.com/harland-
`medical-systems-inc-camp237833.html), a medical device coating technologies company based in Eden
`Prairie, MN. He can be reached at inftharlandmedical.com (mailto:info@harlandmedicalcoml.
`
`[I mages courtesy of HARLANQ MEDICAL SYSTEMS ihttp://www.harlandmedical.com[)]
`
`Filed Under:
`
`Surface Treatment (/primary-keywords/surface-treatment-O)
`
`RECOMMENDED FOR YOU
`
`BIOMEDEVEv An MDaMEuni
`
`
`
`
`
`D+D
`Omed
`
`(/supplier—stories—week-iuly-7)
`
`
`Page 3
`
`Teleflex Ex. 2055
`Medtronic v. Teleflex
`IPR2020-00131
`
`
`
`(https://biomedboston.com/?
`_mc=arti_x_mddir_edt_aud_allen_bmner_med_902_x-
`
`Supplier Stories for the Week
`
`of July 7 (lsupplier-stories-
`
`eventspage)
`
`week-july-7)
`
`BlOMEDevice Boston (https://biomedboston.com/?
`
`_mc=arti_x_mddir_edt_aud_a|len_bmner_med_902_x-
`
`eventspage)
`May 6, 2020, to May 7, 2020
`
`l/how—dry-lubricants-
`
`improve-medical-device-
`assembly)
`
`How Dry Lubricants
`
`Improve Medical Device
`Assembly (/how—dry-
`lubricants-improve-
`
`medical-device-assembly)
`Recent medical device design
`incorporates single-use devices
`that make...
`
`Nu—Med Files Patent for a
`
`(/could-protect-
`
`Reactive Tube Coating
`
`implantable-devices-
`
`(/nu-med-flles-patent-
`
`reactive-tu be—coating)
`Nu—Med Plus, a Salt Lake City,
`UT-based medical device
`
`company, recentlyfiled...
`
`dangerous-bacteria)
`Could This Protect
`
`Implantable Devices from
`
`Dangerous Bacteria?
`(Icould-protect-
`
`implantable-devices-
`
`dangerous-bacteria)
`Bacteria may need to find a
`new home inside the body, as
`researchers from...
`
`Using Laser Marking to Meet UDI Requirements (/using-laser-marking-meet-udi-
`
`requirements)
`Laser marking is an effective and safe solution for marking unique device...
`
`Comment *
`
`See All in Surface Treatment » (/taxonomy/term/109)
`
`
`500 characters remaining
`
`Get Medtech News in Your
`
`+—
`
`Inbox Daily
`
`Subscribe Now
`
`(subscribe)
`
`
`
`TRENDING
`
`
`Page 4
`
`Teleflex Ex. 2055
`Medtronic v. Teleflex
`IPR2020-00131
`
`
`
`Medtronic Rumored to Be Looking at a Deal with LivaNova (/medtronic-rumored-be-
`
`looking—deal-livanova)
`
`This Robot Gives Birth and We Can't Stop Watching (/robot-gives-birth-and-we-cant-
`
`stop-watching)
`
`FDA Strikes Hard at Coronavirus Outbreak with New Policy (/fda-strikes-hard-
`
`coronavirus-outbreak-new-policy)
`
`Thermo Fisher Scientific Set to Acquire Qiagen for $11.5B (/thermo-fisher-scientific-
`
`set—acquire-qiagen-115b)
`
`The Anatomy of a Medical Device Startup (lanatomy-medical-device-startup)
`
`SweynTooth Cybersecurity Vulnerabilities Put Dozens of Medical Devices at Risk
`
`(/sweyntooth-cybersecurity-vulnerabilities-put-dozens-medical-devices-risk)
`
`Livongo is Still Blowing Minds with Its Success (/livongo-still-blowing-minds-its-
`success)
`
`Why NuVasive's Pulse Delay ls Actually a Good Thing (/why-nuvasives-pulse-delay-
`actually-good-thing)
`
`Search for products &
`
`manufacturers:
`
`Search IVIDDl/Qmecl
`
`Q
`
`(https://directory.qmed.com/index.html)
`
`OUR NEXT EVENT
`
`- IOMEDeVic-
`An Mm“ Even
`
`
`
`E evice
`
`Boston Convention & Exhibition Center,
`Boston
`
`May 06, 2020, to May 07, 2020
`
`(https://biomedboston.com/?_mc=arti_x_mddir_edt_aud_a|Ien_bmner_med_902_x-eventspage)
`
`
`Page 5
`
`Teleflex Ex. 2055
`Medtronic v. Teleflex
`IPR2020-00131
`
`
`
`News (/understanding-low-friction-coatings-medica|—devices?qt-q med_resources_widget=O#qt-q med_resources_V\
`
`I__I
`L.
`,
`I.__L_
`mwmimmimrlrglibanfifiamTedmagtdMammwmfiemiéwufisbgom’aget 1#qt-qmed_resou
`I.--I. _.I.
`u; A LI an}: 3-13
`IUUI\' uaLn‘aI'HIE" IIiSLCIiyCII' iuuuLiCESSIDLcu' IIIC1U‘I'
`IQ III.IIII
`
`The Tooling Equation: Are You Maximizing Value? (https://directory.q med.com/too|ing-for-
`molded-parts-is-often-an-area-that-file101011.html)
`
`IP Protection: Four Areas to Audit in Selecting A Contract Manufacturer
`
`(https://directory.qmed.com/p-intellectual-property-ip-is-the-core-of-file099136.htm|)
`
`Outsourcing in Mexico: Is This Location the Best Fit for Your Project?
`(https://directory.qmed.com/p-the-Iast-few-years—have-challenged-every-supply-
`file104233.html)
`
`Finding the Optimal Metal Tubing for Medical and Critical Applications
`(https://directory.qmed.com/selecting-the-best-al|oy-for-a-tubing-app|ication-file086470.html)
`
`MORE FROM MD+D| & QMED
`_—
`
`MORE SITES
`
`
`Design News
`Mission (/mission)
`(http5://www.designnews.com/}
`Marketing Planner
`(http://media.ubmamevents.com/LP=1111) PlasticsToday
`Staff (/staff)
`(https://www.p|asticstoday.com/)
`Add Your Company to Qmed
`Packaging Digest
`(http://dircctory.qmed.com/add-your-
`(http://www.packagingdigest.com/)
`company-page000134.html)
`Qmed (http://directory.qmed.com/)
`Subscribe (/subscribe)
`Powder & Bulk Solids
`(http://www.powderbu|ksolids.com/)
`
`SOCIAL
`
`acebook
`
`
`# F
`
`Linkedln
`
`RSS Feed
`
`ma l’l<E‘l'S
`(http://www.informamarkets.coml
`
`
`Copyright © 2020. All rights reserved. Informa Markets, a trading division
`of Informa PLC.
`
`Accessibility (http://www.informamarkeiscom/accessibility)
`| Privacy Policy (http://www.informamarkels.com/privacy-policy)
`
`
`Page 6
`
`Teleflex Ex. 2055
`Medtronic v. Teleflex
`IPR2020-00131
`
`
`
`| Cookie Policy (http://www.informamarkets.com/cookie-policy)
`I Terms of Use (http://www.informamarkets‘com/terms-of-use)
`
`
`Page 7
`
`Teleflex Ex. 2055
`Medtronic v. Teleflex
`IPR2020-00131
`
`