throbber
Declaration previously submitted in conjunction with
`Medtronic, Inc. and Medtronic Vascular, Inc.’s Petition
`for Inter Partes Review in IPR2020-00136, et seq.
`
`
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`MEDTRONIC, INC., AND MEDTRONIC VASCULAR, INC.,
`Petitioners,
`v.
`TELEFLEX INNOVATIONS S.À.R.L.,
`Patent Owner.
`
`
`Case No.: IPR2020-00136
`U.S. Patent No: RE45,776
`
`
`
`DECLARATION OF RICHARD A. HILLSTEAD, PH.D., FAHA
`
`
`
`
`
`
`Page 1
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`
`TABLE OF CONTENTS
`
`
`Overview of Common Knowledge Related to Flexibility and
`
`INTRODUCTION ............................................................................................. 1
`I.
`QUALIFICATIONS .......................................................................................... 1
`II.
`SCOPE OF WORK ........................................................................................... 4
`III.
`IV. OVERVIEW OF ART ....................................................................................... 6
`A. Overview of Kontos ................................................................................ 6
`B.
`Overview of Ressemann and Common Teachings in the Art ................. 9
`C.
`Background on catheter sizes ................................................................13
`D. Overview of Takahashi .........................................................................14
`E.
`Reinforcement .......................................................................................16
`F.
`Overview of Kataishi ............................................................................23
`KONTOS AND RESSEMANN COMBINATION .........................................26
`A. A POSITA would have been motivated to modify Kontos to add
`as taught by Ressemann ........................................................................26
`B.
`proximal opening more rigid than the tubular structure. ......................41
`C.
`structure. ................................................................................................44
`D. Kontos as modified by Ressemann discloses reinforcing the
`polymer. .................................................................................................45
`E.
`or coil that is 20 to 30 cm. .....................................................................50
`F.
`support collar that form a rigid portion of the extension catheter.........51
`G. Kontos as modified by Ressemann discloses a partially
`cylindrical opening and a tubular structure that is reinforced. .............53
`
`V.
`
`a partially cylindrical opening proximal of the tubular structure,
`
`Kontos as modified by Ressemann’s support collar discloses a
`
`Kontos as modified by Ressemann’s support collar includes a
`concave track that is continuous with the lumen of the tubular
`
`tubular structure with a braid or a coil that is surrounded by a
`
`Kontos as modified by Ressemann discloses a reinforcing braid
`
`Kontos as modified by Ressemann discloses a wire 14 and
`
`
`
`
`
`i
`
`
`
`Page 2
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`
`cylindrical opening that has a greater flexural modulus than tube
`
`Kontos as modified by Ressemann discloses at least two inclined
`
`Kontos in combination with Ressemann, in further view of
`Takahashi discloses a tubular structure that is uniform in size
`
`VIEW OF KATAISHI AND/OR THE COMMON KNOWLEDGE OF
`
`view of Kataishi to arrive at proximal opening with two inclines
`
`A POSITA would be motivated to modify Kontos in combination
`with Ressemann in view of Takahashi, in further view of
`
`COMBINATION WITH GUIDE CATHETER 38 ASSISTS IN
`RESISTING AXIAL AND SHEAR FORCES EXERTED BY AN
`INTERVENTIONAL CARDIOLOGY DEVICE THAT TEND TO
`
`H. Kontos, as modified by Ressemann, discloses a partially
`16 of tubular structure. ..........................................................................54
`I.
`regions ...................................................................................................54
`VI. KONTOS IN COMBINATION WITH RESSEMANN IN FURTHER
`VIEW OF TAKAHASHI .................................................................................57
`A. A POSITA would be motivated to combine Takahashi with
`Kontos and Ressemann .........................................................................57
`B.
`from a proximal end to a distal end ......................................................60
`VII. KONTOS IN COMBINATION WITH RESSEMANN IN FURTHER
`A POSITA ........................................................................................................61
`A. A POSITA would be motivated to modify Kontos/Ressemann in
`and non-inclined region ........................................................................61
`B.
`Kataishi and/or the common knowledge of a POSITA .........................65
`VIII. KONTOS DISCLOSES THAT THE SUPPORT ASSEMBLY 10 IN
`DISLODGE A GUIDE CATHETER FROM A BRANCH ARTERY ............66
`
`
`
`
`
`
`
`
`
`
`ii
`
`Page 3
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`
`TABLE OF EXHIBITS
`
`Description
`U.S. Patent No. RE45,776 (“the ’776 patent”)
`
`U.S. Patent No. 7,736,355 (“Itou”)
`
`U.S. Patent No. 7,604,612 (“Ressemann”)
`
`U.S. Patent No. 5,439,445 (“Kontos”)
`
`New Method to Increase a Backup Support of a 6 French Guiding
`Coronary Catheter, Catheterization and Cardiovascular Interventions
`63: 452-456 (2004) (“Takahashi”)
`
`Excerpt from Grossman’s Cardiac Catheterization, Angiography, and
`Intervention (6th edition) (2000) (chapters 1, 4, 11, 23-25).
`
`U.S. Patent No. 5,891,056 (“Ramzipoor”)
`
`Boston Scientific, Summary of Safety and Effectiveness Data,
`TAXUS™ Express2™ Drug-Eluting Coronary Stent System (March
`4, 2004)
`
`U.S. Publication Application No. 2005/0015073 (“Kataishi”)
`
`The sliding rail system (monorail): description of a new technique
`for intravascular instrumentation and its application to coronary
`angioplasty, Z. Kardio. 76:Supp. 6, 119-122 (1987) (“Bonzel”)
`
`U.S. Publication Application No. 2004/0236215 (Mihara)
`
`U.S. Publication Application No. 2004/0010280 (“Adams ’280”)
`
`U.S. Patent No. 6,042,578 (“Dinh”)
`
`No.
`1401
`
`1407
`
`1408
`
`1409
`
`1410
`
`1415
`
`1418
`
`1424
`
`1425
`
`1432
`
`1433
`
`1435
`
`1446
`
`1447 WO 97/37713 (“Truckai”)
`
`1448
`
`Terumo Heartrail II product literature
`
`iii
`
`
`
`
`
`
`Page 4
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`
`Description
`No.
`1449 Medtronic Launcher product literature
`
`1450
`
`1451
`
`1455
`
`1461
`
`1462
`
`1475
`
`U.S. Patent No. 5,980,486 (“Enger”)
`
`U.S. Patent No. 5,911,715 (“Berg”)
`
`Sakurada, Improved Performance of a New Thrombus Aspiration
`Catheter: Outcomes From In Vitro Experiments and a
`Case Presentation (“Sakurada”)
`
`U.S. Patent No. 5,690,613 (“Verbeek”)
`
`lserson, J.-F.-B. Charrière: The Man Behind the “French” Gauge,
`The Journal of Emergency Medicine. Vol. 5 pp 545-548 (1987)
`
`Excerpt from Plaintiff’s infringement allegations in Vascular
`Solutions, LLC. v. Medtronic, Inc., D. Minn., No. 19-cv-01760
`(October 11, 2019), D.I. 1-14.
`
`1479
`
`Complaint in Vascular Solutions, LLC. v. Medtronic, Inc., D. Minn.,
`No. 19-cv-01760 (October 11, 2019), D.I. 1-14.
`
`
`
`
`iv
`
`Page 5
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`Introduction
`I.
`I have been retained by Robins Kaplan LLP on behalf of Medtronic,
`1.
`
`Inc. and Medtronic Vascular, Inc. (“Medtronic”) as an independent expert to
`
`provide my opinion on the disclosures of certain patents.
`
`2.
`
`I am informed that Medtronic intends to use my opinion in support of
`
`its petition to the Patent Trial and Appeal Board (“PTAB”) for Inter Partes Review
`
`(“IPR”) of U.S. Patent No.: RE45,776.
`
`3.
`
`I make this declaration based on my personal education, experience,
`
`and knowledge in the field of medical device product development.
`
`II. Qualifications
`4. My curriculum vitae is being submitted as Ex. 1443.
`
`5.
`
`I have been actively involved in the design and development of
`
`medical devices for more than thirty years. I held several progressive, Product
`
`Research and Development positions with Cordis Corporation (J&J) from 1987 to
`
`1993 where I was responsible for the design and development of numerous
`
`vascular intervention devices including stents and angioplasty balloon catheters. I
`
`pioneered device development in the Cordis Coronary Stent program as a Senior
`
`Corporate Research Engineer. During my tenure at Cordis, I also held the position
`
`of Senior Engineer, Custom Products, where I was responsible for designing a
`
`wide variety of customized catheters and devices for individual physicians.
`
`
`
`
`1
`
`Page 6
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`From 1993 until 1999, I directed new technology development for
`6.
`
`Georgia-based Novoste Corporation, primarily focusing on intravascular
`
`brachytherapy and catheter based delivery systems for the treatment of coronary
`
`restenosis following angioplasty and stenting.
`
`7.
`
`In 1999, I became a founding member of The Innovation Factory, a
`
`private medical device incubator in Duluth, GA. At The Innovation Factory, I
`
`served as Chief Science Officer and was primarily responsible for early clinical
`
`investigations, and overall R&D in a wide variety of life science ventures.
`
`8.
`
`I was a principal partner and founding member in Accuitive Medical
`
`Ventures I and II (2004 – 2008). Accuitive Medical Ventures is a $225M venture
`
`capital fund focused entirely on growing early stage medical device companies into
`
`attractive candidates for acquisition. In 2008, I joined another medical device
`
`venture fund, Georgia Venture Partners (GVP), where I remain a partner today.
`
`9.
`
`I have managed numerous, diverse, multi-disciplinary development
`
`teams from product concept through clinical approval to sales release. I am a
`
`frequent speaker on the importance of innovation and intellectual property creation
`
`and capture as it relates to the entrepreneurial process in the medical device
`
`industry at conferences and scientific sessions.
`
`10. Currently, I am CEO of Richard A. Hillstead Inc., a medical device
`
`development and entrepreneurship consulting firm located near Atlanta, GA. I am
`2
`
`
`
`
`Page 7
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`also the current Chairman and former CEO of Biofisica Inc., a Georgia wound
`
`healing device company. I am a past Chairman and co-founder of Cerebral
`
`Vascular Applications, Inc. (CVA), a Georgia company dedicated to reducing the
`
`incidence of stroke through closure of the atrial appendage of the heart. I am a
`
`former member of the Emory University New Technology Advisory Board, where
`
`I participated in reviewing promising new medical device technologies and
`
`recommended development strategies.
`
`11.
`
`I was the 2019 recipient of the Georgia BIO Golden Helix Award for
`
`Industry Growth, the highest award bestowed in the state of Georgia for medical
`
`device design, development, and entrepreneurship.
`
`12.
`
` I am named inventor on approximately eighty issued U.S. patents and
`
`pending applications as well as dozens of international patents. My patents pertain
`
`to medical device design, and a majority of these patents relate specifically to
`
`catheter design. My patents and patent publications have been cited, globally, over
`
`8000 times.
`
`13.
`
`In 2012 and 2013, I served as an Entrepreneur in Residence to the
`
`United States Food and Drug Administration.
`
`14.
`
`I am a Fellow in the American Heart Association (FAHA) on the
`
`Council of Clinical Cardiology, Fellow on the Council on Lifestyle and
`
`
`
`
`3
`
`Page 8
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`Cardiometabolic Health, and am a current member of the Stroke Council and
`
`Council on Cardiopulmonary Resuscitation.
`
`15.
`
`I have no financial interest in Medtronic. I similarly have no financial
`
`interest in the asserted patents, and have had no contact with the named inventors
`
`of the asserted patent.
`
`III. Scope of Work
`I have been asked to review the ’776 patent and opine on the level of
`16.
`
`ordinary skill in the art as of May 3, 2006.
`
`17.
`
`I have additionally been asked to consider and provide my opinions
`
`on disclosures in the following references:
`
`1)
`
`2)
`
`U.S. Pat. No. 5,439,445 to Kontos (“Kontos”) (Ex. 1409);
`
`U.S. Pat. No. 7,604,612 to Ressemann (“Ressemann”) (Ex.
`
`1408)
`
`3)
`
`“New Method to Increase a Backup Support of a 6 French
`
`Guiding Coronary Catheter,” Catheterization and
`
`Cardiovascular Interventions, 63:452-456 to Takahashi
`
`(“Takahashi”) (Ex. 1410); and
`
`4)
`
`U.S. Pat. Pub. No. 2005/0015073 to Kataishi et al., (“Kataishi”)
`
`(Ex. 1425)
`
`
`
`4
`
`
`
`
`
`
`Page 9
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`V.
`The Person of Ordinary Skill in the Art
`
`18.
`
`I am not a lawyer and have been informed by counsel of the legal
`
`standards set forth herein. I have also been informed by counsel of factors that may
`
`be considered in determining the level of ordinary skill in the art include (a) the
`
`educational level of the inventor; (b) the type of problem encountered in the art; (c)
`
`prior art solutions to those problems; (d) the rapidity with which inventions are
`
`made; (e) sophistication of the technology; and (f) the educational level of those
`
`working in the field.
`
`19.
`
`I have reviewed the ’776 patent. For the ’776 patent, a person of
`
`ordinary skill in the art (“POSITA”) at the time of the alleged invention (May 3,
`
`2006) would have had (a) a medical degree; (b) completed a coronary intervention
`
`training program, and (c) experience working as an interventional cardiologist.
`
`Alternatively, a POSITA would have had (a) an undergraduate degree in
`
`engineering, such as mechanical or biomedical engineering; and (b) three years of
`
`experience designing medical devices, including catheters or catheter-deployable
`
`devices. Extensive experience and technical training might substitute for education,
`
`and advanced degrees might substitute for experience. Additionally, a POSITA
`
`with a medical degree may have access to a POSITA with an engineering degree,
`
`and one with an engineering degree might have access to one with a medical
`
`degree.
`
`
`
`
`5
`
`Page 10
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`IV. Overview of Art
`A. Overview of Kontos
`20. U.S. Patent No. 5,439,445 (“Kontos”) is titled “Support Catheter
`
`Assembly.” See Ex. 1409. It was filed on June 27, 1994 and issued August 8, 1995.
`
`21. Kontos discloses “[a] support catheter assembly” for “facilitating
`
`insertion of a PTCA balloon into a lesion.” Ex. 1409, 1:10-13. PTCA stands for
`
`percutaneous transluminal coronary angioplasty. Kontos states that “the support
`
`catheter can be inserted into and passed through a guide catheter, over a PTCA
`
`catheter, and out the distal end of the guide catheter so as to function as an
`
`extension of the guide catheter to bridge the gap (or at least some of it) between the
`
`end of the guide catheter and the stenosis to be opened.” Id. at 2:16-22.
`
`22. Kontos details that when removing a stenosis, “[t]he guide catheter
`
`[…] can generally reach only to the coronary ostia, whereas the lesion to be opened
`
`is most commonly located in one of the coronary arteries leading from the ostia.”
`
`Ex. 1409, 1:39-42. Because a guide catheter cannot extend beyond the ostia of a
`
`coronary artery, a therapeutic device such as a balloon catheter must traverse the
`
`coronary artery without the beneficial support of a guide catheter. See Ex. 1409,
`
`1:42-45. Since “the distal end of a PTCA catheter is made to be extremely soft and
`
`flexible,” it is “readily susceptible to kinking and bending” if unsupported by a
`
`guide catheter when traversing a coronary artery. See Ex. 1409, 1:30-38. Kontos
`
`
`
`
`6
`
`Page 11
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`solves this problem and “facilitate[s] the passage of the balloon catheter from the
`
`end of the guide catheter to the lesion” by supporting the therapeutic device with a
`
`“support catheter assembly 10” that functions as an extension of the guide catheter.
`
`Ex. 1409, 1:46-49; see also id. 2:16-22.
`
`23. Fig. 6B from Kontos, reproduced below, illustrates the support
`
`assembly catheter 10 functioning as an extension of the guide catheter 38. A PTCA
`
`balloon catheter 40 is shown disposed within the support catheter and readied for
`
`deployment into a lesion. See Ex. 1409 at 1:9-13.
`
`Ex. 1409, Fig. 6B
`
`24. Figure 6C (below) illustrates the balloon catheter 40 extended beyond
`
`the distal end of the support catheter assembly 10 and into the lesion (stenosis).
`
`
`
`Ex. 1409, Fig. 6C.
`
`
`
`
`
`
`7
`
`Page 12
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`25. As illustrated in Fig. 6C, above, “when extending beyond the distal
`
`end of guide catheter 38, body 12 functions as a guide catheter extension, and the
`
`gap that PTCA catheter 40 must negotiate without assistance is made much
`
`shorter.” Ex. 1409, 5:49-52.
`
`26. Kontos describes a support assembly 10 that includes wire 14 and
`
`body 12. “Body 12 [] can be viewed as a mini guide catheter…” Ex. 1409, 3:47.
`
`Body 12 includes a tube 16 and soft tip 28. See Ex. 1409, Fig. 1; 3:45-55.
`
`Ex. 1409, Fig. 1, color added.
`
`27.
`
`Insertion wire 14 is permanently affixed to the distal tube 16. Kontos
`
`
`
`states:
`
`Insertion wire 14 is attached to base portion 18 at proximal end 20 of
`tube 16, and preferably is permanently affixed thereto. For example, as
`best shown in FIGS. 1, 3 and 4, wire 14 may be connected to base
`portion 18 by inserting it into a receiving hole 34, and affixing it therein
`by, for example, gluing, pressure fitting, shrink fitting, or the like.
`Alternatively, tube 16 may be molded directly onto application wire 14.
`
`
`
`
`8
`
`Page 13
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`Numerous other methods of connecting wire 14 to body 12 will readily
`occur to those skilled in the art.
`
`Ex. 1409, 4:25-34.
`
`28. Kontos teaches that “[t]ube 16 may be composed of any pliable
`
`material suitable for percutaneous medical procedures, but preferably is composed
`
`of a molded plastic material, such as polyethylene.” Ex. 1409, 4:1-4. “[S]oft tip 28
`
`generally is cylindrical in shape and extends coaxially from distal end 24 of tube
`
`16.” Id. at 4:5-7. “Soft tip 28 may be composed of any highly flexible material
`
`suitable for percutaneous medical procedures, but preferably is composed of a soft
`
`plastic such as a copolymer of polyethylene and ethyl vinylalcohol (EVA).” Id. at
`
`4:7-12. 1
`
`B. Overview of Ressemann and Common Teachings in the Art
`29. U.S. Pat. No. 7,604,612 to Ressemann (“Ressemann”) was filed
`
`August 9, 2002 and issued October 20, 2009.
`
`30. Ressemann teaches “a partial length evacuation sheath” used to
`
`“evacuat[e] emboli, particulate matter, and other debris from a blood vessel.” (Ex.
`
`
`1 POSITA would appreciate that in this case patentee has acted as his own
`
`lexicographer as the common shorthand for ethylene vinyl alcohol is EVOH, while
`
`EVA typically refers to ethylene vinyl acetate.
`
`
`
`
`9
`
`Page 14
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`1408, 3:59-60; 5:65-67.) The evacuation sheath of Ressemann is intended to be
`
`deployed with the aid of a guide catheter that is positioned within the ostium of a
`
`coronary artery, while the evacuation sheath may be advanced through the guide
`
`catheter and beyond a major side branch of a target vessel. (Ex. 1408. 12:26-30.)
`
`31. Ressemann’s evacuation sheath assembly is for aspirating embolic
`
`material, id., Abstract; 12:9-13:34, and for stent or balloon delivery. Id., 6:25-34;
`
`12:3-8.
`
`32. The evacuation sheath includes a distal evacuation head 132 and shaft
`
`portions (including proximal shaft 110, distal shaft 120). Id., 6:19-20, 10:36,
`
`10:60-62, Figs. 1A, 1C, 11A. The head is “preferably made of a relatively flexible
`
`polymer such as low-density polyethylene, polyurethane, or low durometer
`
`Pebax® material.” Id., 6:36-39. (The head 132 is illustrated below in pink).
`
`33. Ressemann teaches that a coil 139 can be embedded into the tube to
`
`provide kink resistance. Id. 7:1-7. “A covering of polyurethane can then be applied
`
`to contain the coil 139,” and secure it in position within evacuation lumen 140.”
`
`Id., 7:8-12; Fig. 1C. Ressemann describes that the distal end of the tube is “angled”
`
`to impart more flexibility at the distal tube’s end. See id., 7:48-51.
`
`
`
`
`10
`
`Page 15
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`
`34. Ressemann also teaches deploying a stent through evacuation head
`
`
`
`132, as shown below in Figure 6E, excerpted below.
`
`
`
`Ex. 1408, Figure 6E, color and annotations added.
`
`
`
`
`
`
`11
`
`Page 16
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`35. Ressemann also teaches to angle the proximal opening of its
`
`evacuation head. Ressemann teaches that the angle at 140a allows for smoother
`
`passage of therapeutic devices through the lumen 140 and the larger area of an
`
`angled opening (as opposed to the area of an opening formed by a vertical end to
`
`lumen 140) allows for larger material to pass through the lumen more smoothly.
`
`The proximal and distal ends 140 a, 140 b of
`the evacuation
`lumen 140 are preferably angled to allow for smoother passage of the
`evacuation sheath assembly 100 through a guide catheter, and into a
`blood vessel, and to facilitate smoother passage of other therapeutic
`devices through the evacuation lumen 140 of the evacuation head 132.
`The larger area of the angled open ends also allows for larger
`deformable particulate matter to pass through the lumen more
`smoothly.
`
`Ex. 1408, 6:52-60 (100 embodiment); see also id., 24:33-38 (2100 embodiment).
`36. The flexibility of Ressemann’s evacuation sheath assembly is
`
`engineered in multiple ways. The bending stiffness of stiffness transition member
`
`135 “decreases gradually from the proximal end to the distal end of the stiffness
`
`transition member 135.” (Ex. 1408, 11:57-59.) Ressemann also discloses
`
`transitioning the stiffness of the tube 138 by incorporating a kink-resistant coil 139
`
`that includes its proximal and distal ends wound at a reduced pitch to allow the
`
`
`
`
`12
`
`Page 17
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`final coil to be positioned at the marker bands 146a and 146b and avoid kinking at
`
`that interface. (Ex. 1408, 7:19-28.)
`
`37. Either the decreasing rigidity of member 135 (alone), or the absence
`
`of kink-resistant coil 139 at the distal end of the evacuation head made of a flexible
`
`polymer (alone), renders the distal end of the evacuation head more flexible than
`
`its more proximal regions.
`
`C. Background on catheter sizes
`38. Clearly, Ressemann’s evacuation head is sized such that a balloon
`
`deliverable stent can be passed through its lumen. See Ex. 1408, Fig. 6E, 6F. To
`
`understand a bit more about the sizing of catheters and devices, I provide a brief
`
`background on the size of catheters and interventional devices. The French gauge
`
`size is a quantitative measure common in catheter design that was developed in the
`
`19th century. (Ex. 1462 at 545.) The “French size” (commonly abbreviated as
`
`“Fr”) is a standard unit of measure of the diameter of a catheter. One French equals
`
`1/3 mm. (Ex. 1462 at 545.)
`
`39. A guide catheter must be small enough to fit in the artery of the
`
`patient yet big enough to accommodate interventional devices such as balloon
`
`catheters, and stents. As of May 3, 2006, guide catheters generally ranged in size
`
`from 5 French to 8 French. (See Ex. 1410 at 453-454 & Ex. 1415 at 548-549.)
`
`
`
`
`13
`
`Page 18
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`D. Overview of Takahashi
`40. The article “New Method to Increase a Backup Support of a 6 French
`
`Guiding Coronary Catheter” by Takahashi et. al. (“Takahashi”), published in
`
`Catheterization and Cardiovascular Interventions in 2004 and describes a 5Fr in
`
`6Fr mother-and-child system that provides greater backup support than a 7Fr guide
`
`catheter.2 (Ex. 1410.)
`
`41. As described by Takahashi, the 5 Fr Heartrail catheter had an inner
`
`diameter of 0.059 inches (1.50 mm) (see Ex. 1448) and could accept “normal
`
`balloons or stent delivery systems less than 4.0 mm in diameter.” (Ex. 1410 at
`
`452.) (The 4.0 mm diameter notation of the balloon or stent refers to its fully
`
`expanded size and not its size prior to inflation or expansion.) (See also Ex. 1424,
`
`which describes the Taxus rapid exchange delivered stent for use with a 5 Fr guide
`
`catheter (guide catheter inner diameter equal to or greater than 0.058 inches).)
`
`42. Takahashi teaches inserting a 5 Fr guiding catheter with an inner
`
`lumen size of 0.059 inches into a 6 Fr guiding catheter with an inner lumen size of
`
`0.071 inches (at “B”). Ex. 1410 at 452, 454. The difference between the inner
`
`
`2 Vol. 63, pages 452-456.
`
`
`
`
`14
`
`Page 19
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`lumen diameters of the two guiding catheters is 0.012 inches (0.30 mm), which is
`
`not more than one French size3 in difference. Id.
`
`43. Once the 6 Fr guiding catheter is placed at the ostia of the coronary
`
`artery (“E”), Takahashi teaches to extend the 5 Fr guiding catheter beyond the
`
`distal end of the 6 Fr guiding catheter and into the coronary artery. Ex. 1410 at 454.
`
`Then, Takahashi places the balloon at the lesion with the advantage of the
`
`additional backup support provided by the extension of the 5 Fr guiding catheter.
`
`Id.
`
`
`3 One French equals 1/3 of a millimeter or 0.33 mm.
`
`
`
`
`15
`
`Page 20
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`
`
`E. Overview of Common Knowledge Related to Flexibility and
`Reinforcement
`It is important to understand the criteria that must be met by any
`
`44.
`
`device that is required to be inserted into the vasculature and traverse the anatomy
`
`from the groin to the heart (or from the wrist to the heart, in the case of radial
`
`artery access). When referring to relative rigidities of devices that pertain to PTCA,
`
`a POSITA would understand rigidity to refer to the extent to which a device or
`
`component can be flexed or bent in a lateral direction.
`
`1.
`
`First, the device must be flexible enough to traverse the curves
`and bends of the patient’s vasculature.
`
`
`
`
`16
`
`Page 21
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`2.
`
`Second, the device must not damage the vasculature.
`
`3.
`
`Third, the device must be rigid enough or “pushable” so that it
`does not significantly compress or stretch during placement and
`is “steerable” enough to navigate tortuous vasculature.
`
`45. By May 3, 2006, it was well known that a balance between flexibility,
`
`rigidity, and atraumatic properties must be maintained when designing a PTCA
`
`catheter. Numerous teachings of this balance of features existed in the art at this
`
`time. (Ex. 1415 at 549; and Ex. 1451, 2:35-46; see also, Ex. 1407; Ex. 1408; and
`
`Ex. 1409.)
`
`46. For example, it was known in the art that the distal tip of a catheter
`
`should be soft. Guide catheters, for example, included “a very soft material in the
`
`most distal 2 mm of the catheter to reduce the chance of vessel trauma.” Ex. 1415
`
`at 549. Similarly, the distal end of PTCA catheters were “made to be extremely soft
`
`and flexible so as to facilitate its passage through tortuosities and restrictions in the
`
`vascular system.” Ex. 1409, 1:30-33. In addition, support catheters designed to
`
`extend beyond the distal tip of a guide catheter were also designed to “have a soft
`
`tip.” Ex. 1409, 2:53, 4:5. For example, as Kontos describes, a soft tip could be
`
`made of a soft plastic, such as “a copolymer of polyethylene and ethyl-vinylalcohol
`
`(EVA).” Id., 4:10-11.
`
`
`
`
`17
`
`Page 22
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`47. More generally, it was known in the art that in order for coronary
`
`catheters to traverse safely the vasculature, the preferred material used to form the
`
`catheter should either be a polymer or a material coated with a polymer. See e.g.,
`
`Ex. 1407, 3:30-37; 3:50-58; Ex. 1408, 6:37-42; 6:66-7:14. Most commonly around
`
`2000, guide catheters were formed of either polyethylene (Cook Inc.,
`
`Bloomington, IN) or polyurethane (Cordis Corporation, Miami, Fl., and USCI,
`
`Billerica, MA) and contain either steel braid, nylon, or other reinforcing materials
`
`within the catheter wall. (Ex. 1415 at 214.) Occasionally reference to “resin” in the
`
`literature in conjunction with catheter materials and their manufacture, means that
`
`the final catheter material is formed of a polymer. Resins are the raw materials that
`
`are used to form the final polymer. 4 Colloquially in the catheter arts, reference to a
`
`“resin” can mean that the catheter is formed of a polymer. See e.g., Ex. 1407, 3:30-
`
`37; 3:50-58.
`
`48. The way in which scientists and engineers quantitatively measure the
`
`relative rigidity and flexibility of catheter is by determining their flexural modulus.
`
`Flexural modulus is a material property that expresses “[a] measure of resistance . .
`
`
`4 See https://www.ptonline.com/knowledgecenter/profile-extrusion/glossary-of-
`
`terms. Last accessed Oct. 11, 2019. While this reference is from present day, the
`
`meaning of resin has not changed since May 3, 2006.
`
`
`
`
`18
`
`Page 23
`
`Medtronic Exhibit 1442
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1905
`Medtronic v. Teleflex
`
`

`

`IPR2020-00136
`Patent RE45,776
`. to bending.” (Ex. 1440, 772). In other words, flexural modulus measures the
`
`tendency of a material to resist flexure (otherwise known as bending). A material
`
`with a higher flexural modulus is less flexible than a material with a lower flexural
`
`modulus.
`
`49.
`
`It was known in the art to desi

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket