throbber

`
`
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`
`
`
`
`MEDTRONIC, INC., AND MEDTRONIC VASCULAR, INC.,
`
`Petitioner,
`
`v.
`
`TELEFLEX INNOVATIONS S.À.R.L.
`
`Patent Owner
`
`
`
`
`
`
`
`DECLARATION OF MICHAEL JONES
`SUBMITTED IN SUPPORT OF PETITIONER’S REPLIES
`
`
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1807
`Medtronic v. Teleflex
`
`Page 1
`
`

`

`
`
`TABLE OF CONTENTS
`
`
`
`Page
`
`B.
`
`B.
`
`INTRODUCTION ........................................................................................... 1
`I.
`QUALIFICATIONS ........................................................................................ 1
`II.
`III. LEVEL OF ORDINARY SKILL IN THE ART .............................................. 3
`IV. MECHANISM OF BACK-UP FORCE .......................................................... 4
`A. A Catheter-in-Catheter Assembly Decreases the Distance that
`an Interventional Cardiology Device Travels in the Vasculature. ......... 7
`A Catheter-in-Catheter Assembly Increases the Moment-of-
`Inertia, Which is an Increase in the Stiffness of the Assembly. ............ 8
`ITOU ................................................................................................................ 9
`V.
`VI. RESSEMANN ...............................................................................................17
`A.
`Rationale for Modifying Embodiment 100 with Collar 2141 .............17
`1.
`Use of flexibility transition members in the art ........................19
`Feasibility of Modifying Embodiment 100 with Collar 2141 ............25
`1.
`Ressemann and collar 2141 ......................................................38
`C. Modification of Ressemann to Achieve a 1 Fr. Differential
`between the Evacuation Lumen and the Guide Catheter ....................44
`VII. MODIFICATION OF ITOU WITH RESSEMANN COLLAR 2141 ...........49
`VIII. MODIFICATION OF ITOU OR RESSEMANN WITH KATAISHI ...........57
`IX. KONTOS .......................................................................................................64
`A. Kontos Resists Axial and Shear Forces that Would Otherwise
`Tend to Dislodge the Guide Catheter. .................................................66
`Kontos Does Not Require a Narrow Tube 16 to Protect the
`PTCA Catheter Contained Therein. ....................................................69
`After Replacing the Funnel with a Side Opening, Kontos’s
`Support Catheter 10 could be made bigger .........................................72
`D. Modifying Kontos with Ressemann’s support collar ..........................81
`E. Modifying Kontos with Kataishi’s shape ............................................89
`X. GIVEN RESSEMANN’S SUPPORT COLLAR 2141, AND THE
`SHAPE OF KATAISHI’S DESIGN, THE ULTIMATE PROXIMAL
`SIDE OPENING SHAPE IS AN OBVIOUS DESIGN CHOICE
`
`B.
`
`C.
`
`
`
`i
`
`Page 2
`
`

`

`
`
`DETERMINED THROUGH ROUTINE EXPERIMENTATION AND
`OPTIMIZATION ...........................................................................................91
`XI. BY MAY 3, 2006, IT WAS KNOWN IN THE ART TO PROCESS A
`STIFF HYPOTUBE TO IMPROVE STRAIN RELIEF AND
`TRANSITIONING STIFFNESS BETWEEN A HYPOTUBE AND A
`FLEXIBLE CATHETER. ..............................................................................96
`
`
`
`
`
`
`ii
`
`Page 3
`
`

`

`
`
`I.
`
`Introduction
`
`1.
`
`I have been retained by Robins Kaplan LLP on behalf of Medtronic,
`
`Inc. and Medtronic Vascular, Inc. (“Medtronic”) as an independent expert.
`
`2.
`
`I am informed that Medtronic intends to use my opinions in support of
`
`petitions pending at the Patent Trial and Appeal Board (“PTAB”) concerning U.S.
`
`Patent Nos. 8,048,032; RE45,380; RE45,776; RE45,760; and RE47,379 (the
`
`“Teleflex Patents”): IPR2020-00126, IPR2020-00127, IPR2020-00128,
`
`IPR2020-00129, IPR2020-00130, IPR2020-00132, IPR2020-00134,
`
`IPR2020-00135, IPR2020-00136, IPR2020-00137 and IPR2020-00138.
`
`3.
`
`I have been asked to provide my opinion on the disclosures of certain
`
`patents and articles in view of the knowledge of a POSITA as of the date of
`
`invention of the Teleflex Patents.
`
`4.
`
`I am informed that the parties dispute the date of invention of the
`
`patents that are the subject of the IPRs. My opinions herein are the same whether
`
`the relevant date is May 3, 2006 or early 2005.
`
`5.
`
`I make this Declaration based on my personal education, experience,
`
`and knowledge of medical device product development since the mid-1980s. I
`
`have also reviewed the materials cited herein.
`
`II. Qualifications
`
`6. My curriculum vitae is attached as Appendix A.
`
`1
`
`Page 4
`
`

`

`
`
`7.
`
`I received a Bachelor of Science degree in Chemical Engineering
`
`from the University of California, Berkeley in 1984. I later earned a Master of
`
`Science degree in Engineering, Mechanics and Materials, in 1995 from California
`
`State University, Long Beach.
`
`8.
`
`From the mid-1980s through the late-1990s I worked as an engineer
`
`responsible for various aspects of medical device design, development, and
`
`manufacturing for products that included catheters for treating thrombo-embolic
`
`stroke, catheters for treating subarachnoid hemorrhage, balloon catheters for
`
`treating vascular spasm (Micro Therapeutics, Inc.), an electrosurgical balloon
`
`catheter and a ureteral stent product line (Applied Medical Resources), urological
`
`balloon catheters (Baxter Healthcare), embolectomy catheters, endo tracheal and
`
`tracheostomy tubes, and angioplasty catheters (Shiley).
`
`9.
`
`Since 1997, I have worked at Design Development and Fabrication,
`
`Inc. in contract mechanical engineering, focused on product design and
`
`development. My projects included prototype designs for minimally invasive,
`
`surgical, and implantable devices, development and fabrication for surgical
`
`devices, test fixture design and fabrication, and process validation in support of
`
`PMA filings. Clients included: Edwards Lifesciences Inc., SenoRx Inc, Vascular
`
`Control Systems Inc., Micro Therapeutics and Neuroperfusion.
`
`2
`
`Page 5
`
`

`

`
`
`10. Over the course of my work, I have been named as an inventor on
`
`over 100 patents.
`
`11.
`
`I am being compensated at $425 per hour for my time, and my
`
`compensation is not contingent on the results of these proceedings.
`
`III. Level of Ordinary Skill in the Art
`
`12.
`
`I understand that Medtronic proposed the following level of ordinary
`
`skill in the art for these IPRs: a person of ordinary skill in the art (“POSITA”) at
`
`the time of the alleged invention would have had: (a) a medical degree; (b)
`
`completed a coronary intervention training program; and (c) experience working as
`
`an interventional cardiologist. Alternatively, a POSITA would have: (a) an
`
`undergraduate degree in engineering, such as mechanical or biomedical
`
`engineering; and (b) three years of experience designing medical devices,
`
`including catheters or catheter-deployable devices. Extensive experience and
`
`technical training might substitute for education, and advanced degrees might
`
`substitute for experience. Additionally, a POSITA with a medical degree may have
`
`access to a POSITA with an engineering degree, and one with an engineering
`
`degree might have access to one with a medical degree.
`
`13. Based on my education and experience, I meet the definition of a
`
`POSITA. My opinions herein are provided with respect to this definition.
`
`3
`
`Page 6
`
`

`

`
`
`IV. Mechanism of Back-up Force
`
`14.
`
` I have been asked to opine on the mechanism by which a
`
`catheter-in-catheter assembly provides additional back-up support for deployment
`
`of devices, such as guidewires, balloon catheters, stents, and stent catheters, to a
`
`lesion located in a coronary artery.
`
`15. By way of background, tough lesions located deep within the
`
`coronary arteries are difficult to reach. One challenge to treating such distal lesions
`
`is the problem of guide catheters dislodging from the coronary ostium when a
`
`guidewire, balloon catheter, stent or stent catheter meets a particularly tough distal
`
`lesion. This is shown schematically in the figure below. Image “a” illustrates by
`
`arrows the force pushing the guide catheter away from the ostium and coronary
`
`artery due to a guidewire engaging a tough stenosis, for example.
`
`Ex-1814 (Takahiko Suzuki, M.D.)
`http://cct.gr.jp/2003/books/bookspdf/suzuki05cii.pdf
`
`
`
`4
`
`Page 7
`
`

`

`
`
`16. Prior to the 2005-2006 timeframe, a known solution to prevent
`
`back-out of the guide catheter was the use of a “double-catheter,” sometimes
`
`referred to as a “catheter-in-catheter assembly.” By way of the example provided in
`
`image “b” above, Dr. Takahiko Suzuki, M.D. of Toyohashi Heart Center, describes
`
`using a 120 cm 6 Fr Goodman Road-Master™ catheter in a 8 Fr guide catheter “to
`
`make a ‘double-catheter’” to achieve the “back-up force needed to cross [a] wire”
`
`in an ostial RCA or LCA lesion.” Ex-1814.
`
`17. Similarly, in a 2004 article, Takahashi quantitatively illustrates the
`
`increased back-up support of a catheter-in-catheter system. See Ex-10101. In
`
`particular, Takahashi describes achieving the backup support of an 8 Fr guide
`
`
`1 I cite the references I discuss in this declaration according to the way they are
`
`numbered in IPR2020-00126 (e.g. Ex-1010.). However, I understand that the
`
`same set of exhibits was filed in each of the eleven IPRs concerning the five
`
`Root patents, always retaining the same final two numbers. There are
`
`several exceptions to this general rule, including at least the exhibits ending in
`
`“01,” “05,” and “42;” slightly different versions of these documents were filed in
`
`each of the IPRs, although there is much duplication of content. When I cite these,
`
`I indicate the respective IPR number.
`
`
`
`5
`
`Page 8
`
`

`

`
`
`catheter with a 5 Fr-in-6 Fr system. Id. A figure of Takahashi’s catheter-in-catheter
`
`assembly deploying a balloon catheter is shown below.
`
`Id., Fig. 2.
`
`
`
`18. Suzuki and Takahashi disclose a catheter-in-catheter system that
`
`improves back-up support. Takahashi teaches that the distance between the inner
`
`diameters of the nested catheters is less than 1 French. Ex-1010 (disclosing an
`
`inner diameter of the outer 6 Fr catheter of 0.071 inches and inner diameter of the
`
`inner 5 Fr catheter of 0.059 inches—a difference of 0.012 inches (or 0.30 mm,
`
`which is less than 0.33 mm (1/3 mm = 1 Fr)). Suzuki teaches improved back-up
`
`support without necessitating a 1 French relationship between the inner diameters
`
`of the nested catheters (i.e., describing a 6Fr in 8Fr system).
`
`19. As shown in Suzuki, a catheter-in-catheter assembly can provide
`
`back-up support even if the distance between the inner diameters of the nested
`
`catheter is greater than 1 French. This is for two reasons. First, because the inner
`
`6
`
`Page 9
`
`

`

`
`
`catheter extends beyond the ostium and into the coronary artery, the
`
`catheter-in-catheter assembly decreases the distance that an interventional
`
`cardiology device travels in the vasculature. Second, by making a “double
`
`catheter,” the catheter-in-catheter assembly is stiffened, which leads to additional
`
`back-up support.
`
`A. A Catheter-in-Catheter Assembly Decreases the Distance that an
`Interventional Cardiology Device Travels in the Vasculature.
`
`20.
`
` By using a catheter-in-catheter assembly, it is possible to extend the
`
`inner catheter (sometime referred to as the “child catheter”) beyond the ostium of
`
`the coronary artery. In so doing, the interventional cardiology device has to travel a
`
`shorter distance in the vasculature, which in turn reduces the amount of force
`
`necessary to advance the interventional cardiology device to the target location.
`
`This is because the vasculature can be tortuous and/or calcified, thereby requiring
`
`more force to advance the interventional cardiology device.
`
`21. As discussed above, catheters are generally lined with a lubricious
`
`coating, such as Teflon. See Ex-1015, 548. This lining is intended to promote
`
`trackability of an interventional cardiology device within the catheter assembly,
`
`and reduces the amount of force necessary to advance the interventional cardiology
`
`device to the target location. This, in turn, results in reducing the axial and shear
`
`forces exerted by an interventional cardiology device that would otherwise tend to
`
`7
`
`Page 10
`
`

`

`
`
`dislodge from the ostium of a coronary artery the distal portion of a catheter-in-
`
`catheter assembly.
`
`B. A Catheter-in-Catheter Assembly Increases the Moment-of-
`Inertia, Which is an Increase in the Stiffness of the Assembly.
`
`22. The mechanisms involved in the movement of catheters while in the
`
`vasculature explains how a catheter-in-catheter assembly provides back-up
`
`support. In particular, the following mechanical properties of catheters influence
`
`back-up support: flexural rigidity, torsional rigidity, and buckling force. See
`
`Ex-1834. These mechanical properties are proportional to the moment of intertia,
`
`and the construction material. Id.
`
`23.
`
` The moment of inertia, 𝐼(cid:2868), of the cross section of a catheter with
`
`respect to its neutral axis, is expressed by the following, and is proportional to
`
`flexural rigidity and buckling force. Id. at 12. The cross section of a catheter is a
`
`hollow circle with an outer diameter D and an inner diameter d. Id.
`
`
`
`𝐼(cid:2868) =
`
`𝜋(𝐷(cid:2872) − 𝑑(cid:2872))
`
`64
`
`24. The polar moment of inertia of a catheter shaft with outer diameter D
`
`and inner diameter d is expressed as follows, and is proportional to the torsional
`
`rigidity of a catheter. Id., 12-15.
`
`𝐽(cid:2868)𝑐𝑎𝑡ℎ𝑒𝑡𝑒𝑟 =
`
`𝜋(𝐷(cid:2872) − 𝑑(cid:2872))
`
`32
`
`8
`
`Page 11
`
`

`

`
`
`25. From these equations it is clear that when placing a catheter within a
`
`catheter, the outer diameter remains defined and the effective inner diameter
`
`decreases, which results in the following consequences:
`
`a. Flexural rigidity increases;
`
`b. Torsional rigidity increases; and
`
`c. Resistance to buckling force increases.
`
`26. The polar moment of inertia of a catheter shaft with outer diameter D
`
`and inner diameter d is expressed as follows, and is proportional to the torsional
`
`rigidity of a catheter. Id., 12-15.
`
`27. The increase in these properties is why backup support is increased
`
`when a catheter-in-catheter system (whether a full-length catheter in a full-length
`
`catheter or rapid-exchange catheter within a full-length catheter) is used instead of
`
`a single-guide catheter. It also explains why Suzuki—despite not describing a 1
`
`French relationship—provides back-up support.
`
`V.
`
`Itou
`
`28.
`
`I have been asked to opine on whether suction catheter (2) of Itou is
`
`sufficiently sized to receive a balloon or stent catheter. In connection with my
`
`analysis, I have reviewed the position of Medtronic on this limited topic, as well as
`
`Patent Owner’s position as articulated in Patent Owner’s response and the
`
`declaration of Peter T. Keith. See Paper 1 (IPR2020-00135), 30-31, Paper 44
`
`9
`
`Page 12
`
`

`

`
`
`(IPR2020-00135), 11-17; Ex-2138 (IPR2020-00135), ¶¶ 138-41; Paper 44
`
`(IPR2020-00126), 19-21; Paper 44 (IPR2020-00128), 19-21; Paper 44
`
`(IPR2020-00132), 9-10.
`
`29. With reference to Figures 1 and 3 of Itou, Itou teaches an assembly
`
`consisting of a guidewire (6), which fits within a distal end protective catheter (5).
`
`Catheter (5) fits within suction catheter (2), and catheter (2) fits within guide
`
`catheter (1).
`
`Ex-1007, Figs. 1A-F.
`
`
`
`10
`
`Page 13
`
`

`

`
`
`
`
`Id., Fig. 3 (illustrating suction catheter 2, color added).
`
`30. Table 1 of Itou teaches that the inner diameter of catheter 2 is 1.5 mm,
`
`which is 0.059 inches. Id., 7:55-65. In the 2005 time frame, there were
`
`commercially available balloon and stent catheters that could be delivered through
`
`a guide catheter that had an inner diameter (“ID”) of >0.056 inches. One example
`
`is the 5 Fr Heartrail catheter that has an inner diameter of 0.059 inches. Ex-1010,
`
`452. Such an inner diameter can “accept normal balloons or stent delivery systems
`
`less than 4.0 mm in diameter.” Id.
`
`31.
`
`I am aware that Itou teaches that catheter 5 is inserted into catheter 2,
`
`and the catheters are together advanced through guiding catheter 1. Ex-1007, 5:29-
`
`51. I am also aware that Itou teaches that suction catheter 5 should then be
`
`removed from catheter 2. Id., 7:13-19.
`
`32.
`
`I am also aware that Itou teaches that when the distal part of the distal
`
`tubular portion (24) is extended into the coronary artery, the proximal part of
`
`tubular portion (24) remains within the guide catheter. Id., 5:35-42.
`
`11
`
`Page 14
`
`

`

`
`
`33.
`
`I understand Patent Owner has argued that the “effective” size of the
`
`opening of catheter 2 is smaller than Itou discloses. See Paper 44
`
`(IPR2020-00135), 16-17; Ex-2138 (IPR2020-00135), ¶¶ 138-41; Paper 44
`
`(IPR2020-00126), 19-21; Paper 44 (IPR2020-00128), 19-21; Paper 44
`
`(IPR2020-00132), 9-10. Patent Owner argues that wire-like portion 25 to catheter
`
`2 partially blocks the proximal opening of catheter 2. Id. Mr. Keith argues that wire
`
`25 reduces the effective size of the lumen from 0.059 inches to 0.046 inches.
`
`Ex-2138 (IPR2020-00135), ¶ 141. Patent Owner argues that once suction catheter
`
`2 was placed in a guide catheter, it would be impossible to advance a balloon or
`
`stent into catheter 2 for delivery into the coronary vasculature.
`
`34.
`
`I disagree with Mr. Keith and Patent Owner’s position.
`
`35. First, as reported in Itou’s table 1, the wire-like pushrod, 25, does
`
`have a diameter of 0.45 mm (~0.017 inches). However, the portion of wire 25 with
`
`a full, circumferential cross-sectional shape does not abut the proximal opening of
`
`tubular portion 24. Itou is explicit that, adjacent the location of tubular portion 24’s
`
`proximal opening, pushrod 25 is “crushed into a form of a flat plate.” Ex-1007,
`
`4:32-36, Figs. 3, 4. A change in diameter of pushrod 25 is not unsurprising.
`
`Tapering a shaft along its longitudinal axis in the proximal to distal direction was
`
`known. Ex-1063, [0025];
`
`
`
`12
`
`Page 15
`
`

`

`
`
`
`
`
`
`36. Second, catheter (2) bends. It must, in order for its distal tubular
`
`portion to be maneuverable into the coronary vasculature. See, e.g., Ex-1007,
`
`Abstract, Fig. 6. A POSITA would understand Itou’s disclosure to teach that
`
`catheter (2) has varying rigidity along its length, such that its proximal portion is
`
`more rigid, and its distal portion is more flexible. This is not unique. The art taught
`
`that balloon angioplasty catheters should have a decreasing rigidity along their
`
`proximal to distal lengths. Ex-1063, [0025]. The same was taught for embolic
`
`protection catheters, as I discuss herein, in regard to Ressemann, as well as for
`
`aspiration catheters. Ex-1019, 2:60-65 (“The catheters are provided with varying
`
`flexibility along the length of the shaft, such that they are soft and flexible enough
`
`to be navigated through the vasculature of a patient without causing damage, but
`
`are stiff enough to sustain the axial push required to position the catheter properly
`
`and to sustain the aspiration pressures.”)
`
`37. Third, the proximal opening of catheter (2) is cut at an oblique angle,
`
`as illustrated below.
`
`13
`
`Page 16
`
`

`

`
`
`
`
`Ex-1007, Figs. 3 (left), 4.
`
`38. The three features I discuss above — (a) the flattening of wire (25)
`
`adjacent to the proximal opening of catheter (2), (b) catheter 2’s flexibility, and (c)
`
`the angle of the proximal opening means that the proximal opening of catheter (2)
`
`would readily accept a balloon catheter or stent catheter once catheter 2 was
`
`positioned within a guide catheter as shown in Fig. 6.
`
`39.
`
`In order for wire 25 to obstruct Itou in the manner that Patent Owner
`
`and Mr. Keith allege, Itou’s catheter 2 would have to be entirely rigid, which it is
`
`not.
`
`40.
`
`In the alternative, for wire 25 to obstruct Itou in the manner suggested
`
`by Patent Owner and Mr. Keith, catheter 2 would have to be modified such that the
`
`wire 25 is welded to proximal end portion 231 without any flattening of the wire
`
`adjacent the opening. See Ex-1007, 4:32-36. Such a modification would look like
`
`the modified Itou Fig. 3, below.
`
`14
`
`Page 17
`
`

`

`
`
`
`
`Id., Fig. 3 (modified).
`
`
`
`41. This is clearly not what Itou contemplates or teaches. Nor would a
`
`person of ordinary skill in the art understand Itou to disclose such an obstruction at
`
`the proximal opening.
`
`42. Moreover, even if Itou’s proximal opening was obstructed by wire 25
`
`so that the “effective size” of catheter 2’s opening went from 0.059 inches (1.5mm)
`
`to 0.046 inches (1.16 mm) as Patent Owner and Mr. Keith allege (it is not), such an
`
`opening is still large enough to receive a standard coronary stent.
`
`43. By the early 2000s, standard coronary stents, guidewires, balloon
`
`catheters, and stent catheters were available with an outer diameter sufficient to
`
`pass through Itou’s allegedly constricted opening of 0.046 inches. See Ex-1015,
`
`641 (“All current slotted tube designs are ‘bare mounted’ on a delivery balloon,
`
`with deflated profiles smaller than 0.040-in. (1mm)…”); Ex-1802, 104 (Genic®
`
`stent with less than 0.9 mm (0.035 inch) profile), 143 (Lunar stent with 0.0382
`
`15
`
`Page 18
`
`

`

`
`
`inch profile), 269 (Spiral Force stent with 0.039 to 0.042 inch profile), 274
`
`(Tsunami stent with 0.95 mm (0.038 inch) profile); see also, Ex-1803, 4:46-56;
`
`Ex-1804, Table 2.
`
`44. Such sized interventional devices would have sufficient clearance to
`
`be received into a proximal diameter opening of 0.046 inches. Even in the case of a
`
`stent with an outer diameter (“OD”) of 0.044 inches, there would be enough
`
`clearance because the size tolerances on the diameter dimensions can be very
`
`precise (less than plus or minus 0.02 mm, which is less than 0.0008 inches) due to
`
`precision extrusions.
`
`45. Current catheter construction technology using PTFE liners with a tie
`
`layer, stainless steel braids and Pebax outer jackets, and run a tolerance of ±.001
`
`inch on the inner diameter and ±.003 inch on the outer diameter of the completed
`
`catheter body for catheters up to 4.4 mm in outer diameter. These tolerances were
`
`typical in 2005-2006.
`
`46. For the reasons stated above, it is my opinion that the lumen of Itou’s
`
`suction catheter 2 is suitable for receiving a stent catheter once catheter 2 has been
`
`advanced through a guide catheter and has been partially extended from the guide
`
`catheter’s distal end, and when the proximal opening of catheter 2’s tubular portion
`
`is still within the guide catheter.
`
`16
`
`Page 19
`
`

`

`
`
`VI. Ressemann
`
`A. Rationale for Modifying Embodiment 100 with Collar 2141
`
`47.
`
`I have reviewed Petitioner’s arguments regarding reasons to combine
`
`Ressemann collar 2141 with embodiment 100, as well as Teleflex’s arguments to
`
`the contrary. Paper 1 (IPR2020-00129), 45-47; Paper 44 (IPR2020-00129), 28-42;
`
`Ex-2138, ¶¶ 180-88.
`
`48.
`
`I agree that the modification to embodiment 100 of Ressemann is
`
`supported by motivations identified by Petitioner, which include:
`
`1.
`
`2.
`
`to increase the area for receiving a stent and balloon catheter;
`
`the tab portion provides a flexibility transition between the
`
`proximal end of evacuation lumen 140 and shaft 120; and
`
`3.
`
`the support collar reinforces the opening of the lumen.
`
`
`Paper 1 (IPR2020-00129), 46 (citing Ex-1205, ¶ 189; Ex-1242 ¶¶ 86-87.)
`
`
`49. As noted by Medtronic, combining Ressemann’s collar with
`
`Ressemann’s embodiment 100 yields at least two inclines, labeled in the schematic
`
`below.
`
`17
`
`Page 20
`
`

`

`
`
`
`Ex-1008, Fig. 1C (modified with support collar 2141).
`
`
`
`50. Notably, Ressemann’s collar has more than two inclines, as shown in
`
`the schematic below at inclines A, B, and C.
`
`
`Ex-1008, Fig. 16J (annotations showing at least three inclines on the support collar
`
`at the proximal end (A), the transition from the concave track of the tab portion
`
`and incline (B), and the incline near the distal most portion of the opening (C)).
`
`18
`
`Page 21
`
`

`

`
`
`51.
`
`In regard to the motivations identified related to flexibility transition
`
`and proximal opening reinforcement provided by collar 2141, Patent Owner argues
`
`that because Ressemann’s evacuation sheath assembly 100 works without support
`
`collar 2141, POSITA would have no reason to modify it to include the collar. Paper
`
`44, 39. I disagree and set forth my reasons below.
`
`1. Use of flexibility transition members in the art
`
`52. By the relevant time frame it was well known to a POSITA that a
`
`critical region where kinking and buckling can occur in coronary catheters was at
`
`the interface between the stiff proximal portion and the flexible distal portion of
`
`the catheter due to the change in stiffness at this interface.” Ex-1829, 2:38-49.
`
`a. Goodin, U.S. Pat. No. 6,361,529
`
`53. For example, Goodin teaches a rapid-exchange balloon angioplasty
`
`catheter with a proximal section 20 and a distal section 30. Id., 4:26-32. Goodin
`
`also teaches an improvement to the catheter in the form of a stiffness transition
`
`member 50, made, for example, from stainless steel wire, or other material with
`
`appropriate mechanical characteristics. Id., 6:38-41.
`
`19
`
`Page 22
`
`

`

`
`
`Id., Fig. 1 (annotation added).
`
`
`
`54. Goodin teaches positioning member 50 in the interior of a catheter
`
`shaft.
`
`b. U.S. Pat. No. 5,658,251
`
`55. Similarly, a patent that issued to Mr. Keith and others also teaches the
`
`use of a transition member in the interior of a balloon catheter shaft to reinforce the
`
`catheter’s distal end, preventing kinking or closure. Ex-1120 (“the ’251 patent”) at
`
`3:16-19.
`
`56. The inventors explained that the proximal hypotube section 22 of the
`
`catheter was relatively stiff. The catheter’s distal section 24 includes a tubular shaft
`
`23, which was more flexible than section 22. Id., 5:34-38, 6:7-13, 7:6-8. Moreover,
`
`the presence of guidewire 50 within the guidewire lumen 52 meant that there were
`
`relatively stiff sections at both ends of distal section 24. Id., 7:8-10.
`
`“Without the transition member 25, such a configuration will tend to
`
`bend or buckle in the relatively flexible area as an attempt is made to
`
`advance catheter 20 through the vasculature. The responsiveness of the
`
`20
`
`Page 23
`
`

`

`
`
`catheter would thus be substantially diminished. Further, the catheter
`
`20 might tend to kink at the transition section.”
`
`Id., 7:9-14.
`
`57. Similar to Goodin, the ’251 patent teaches adding a transition member
`
`25 to the catheter that increases the stiffness of the distal shaft 23, but is less stiff
`
`(“has less axial strength”) than the hypotube of the main shaft section 22.
`
`Id., Fig. 5 (annotation added).
`
`c. U.S. Pat. No. 5,156,594
`
`
`
`58. Those skilled in the art were also aware that transition members could
`
`be added to a catheter’s exterior, instead of being embedded in the interior of a
`
`catheter shaft. Again, a patent to Mr. Keith illustrates this principle. In a patent
`
`disclosing an over-the-wire balloon dilatation catheter, a guide wire lumen is
`
`disclosed as extending only through the distal portion of the catheter. Ex-1121.
`
`59. Similar to the ’251 patent, the ’594 patent teaches a proximal shaft 22
`
`that is relatively stiff, which includes a more proximal, long shaft section 64
`
`(highlighted below in yellow) and a short shaft section 66 (below, orange).
`
`21
`
`Page 24
`
`

`

`
`
`Id., Fig. 2 (excerpted and annotation added).
`
`
`
`60. The ’594 patent also teaches a distal balloon section 26 (below, pink).
`
`
`
`Id.
`
`
`
`61. Distal section 26 and proximal shaft 22 are joined through an
`
`intermediate sleeve section 24. Id., 7:23-8:14.
`
`22
`
`Page 25
`
`

`

`
`
`Id., Fig. 2 (excerpted and annotation added).
`
`
`
`62. The ’594 patent explains that proximal shaft 22 is relatively stiff
`
`compared to the more flexible polymer intermediate sleeve section 24, and that the
`
`relatively sharp transition in stiffness in the catheter structure creates two potential
`
`concerns. Id., 8:52-55. First, there is the potential for the catheter to kink during
`
`handling. Second, there is the potential for the distal end of the main shaft to “dig
`
`in” to the guide catheter due to the lack of support from the more flexible sleeve.
`
`Id., 8:52-63.
`
`63. The ’594 patent teaches that a kink-resistant structure 110 may be
`
`added to the outside of sleeve 24, to provide intermediate (transitory) stiffness and
`
`kink-resistance between the relatively stiff main shaft and the relatively flexible
`
`intermediate sleeve.
`
`23
`
`Page 26
`
`

`

`
`
`Id., Fig. 1.
`
`
`
`
`64. The ’594 patent teaches that structure 110 may include coil 112,
`
`which may be secured to the exterior of the sleeve by an adhesive like epoxy, and
`
`further secured using a heat shrinkable sheath. Id., 9:15-20. The ’594 illustrates
`
`that there were well known ways to attach additional components to catheters.
`
`65. Collectively, the Goodin, ’251 and ’594 patents illustrate the
`
`well-known principle that coronary catheters may be reinforced to prevent kinking
`
`or buckling, in particular at areas where there is a transition in stiffness, and that
`
`the reinforcement may be placed within a catheter shaft or on the exterior of the
`
`catheter.
`
`66. Eidenschink and Fahey also disclose the need to transition the
`
`stiffness at the critical juncture between a catheter’s relatively stiff proximal
`
`portion and more flexible distal portion, which is described and addressed in detail
`
`in § XI, infra.
`
`24
`
`Page 27
`
`

`

`
`
`B.
`
`Feasibility of Modifying Embodiment 100 with Collar 2141
`
`67.
`
`I have also been asked to opine on how a POSITA would understand
`
`certain embodiments in Ressemann. In particular, for the embodiments shown in
`
`Figures 1A and 1C, I have been asked to detail how POSITA would understand the
`
`relative sizes of the evacuation head 132 as compared to the evacuation lumen 142,
`
`and shafts 110 and 120. A reproduction of Fig. 1A is shown below.
`
`
`
`Ex-1008, Fig. 1A.
`
`
`
`68.
`
`In a preferred embodiment of Ressemann, evacuation head 132 is
`
`intended to be used with an 8 French guide catheter whose inner diameter is about
`
`0.090 inches. Id., 10:13-17. The inner diameter of evacuation lumen 140 is
`
`“approximately 0.061 inches.” Id., 10:18. The targeted outer diameter is 0.076
`
`inches. Id., 10:28. Ressemann’s inflation lumen 142 (shown in cross section in Fig.
`
`1D, below) is “0.005 inches at the widest portion of the crescent (vertical direction
`
`in Fig. 1B).” Id., 10:21-23. I modified Fig. 1D, below, to indicate the vertical
`
`dimension recited by Ressemann.
`
`25
`
`Page 28
`
`

`

`
`
`
`
`
`
`Id., Fig. 1D.
`
`69. Ressemann does not provide a cross sectional view of shafts 110 or
`
`120. Ressemann does state that shaft 110 “forms a hollow tube” preferably “made
`
`of stainless steel.” Id., 10:36-37. Shaft 110 “provides fluid communication between
`
`an inflation apparatus (not shown) and the intermediate and distal shaft portions
`
`120, 130. Id., 10:42-44. Ressemann teaches to coat shaft 110 with “a polymer
`
`sleeve or spray coating for lubricity.” Id., 10:44-46.
`
`70. Shaft 120 also “forms a hollow tube” preferably “formed of
`
`polyethylene or Pebax.” Ex-1008, 10:62-11:1. Shaft 120 also “provides fluid
`
`communication between the proximal shaft portion 110 and distal shaft portion
`
`130.” Id., 11:1-3. Shaft 120 is “secured to the proximal and distal shaft portions
`
`110, 130, preferably by an overlapping weld or bond joint.”
`
`26
`
`Page 29
`
`

`

`
`
`71. While Ressemann does not explicitly show in its figures, POSITA
`
`would understand shaft 110’s and 120’s “hollow tubes” to be circular in cross-
`
`section as was common in the art when Ressemann was filed (2002). The crescent
`
`shape, or D-shape of inflation lumen 142 was also a common inflation lumen
`
`shape known in the art. See, e.g., Ex-1050, Fig. 5, showing a D-shape inflation
`
`lumen for a rapid exchange balloon catheter.
`
`72.
`
`It was common in the art to join a crescent shape or D-shape inflation
`
`lumen with a more proximal tube having a circular cross section. See, e.g,
`
`Ex-1050. Enger illustrates a D-shaped inflation lumen 40 that “communicates”
`
`with lumen 36 of stainless steel hypotube 28 having an outer diameter of 0.022
`
`inches (and thus a circular cross section) and a wall thickness of about 0.003
`
`inches. Id., 5:3-5; see also Ex-1050, Fig. 5. Taking the wall thickness into account,
`
`the inner diameter of hypotube 28 is 0.016 inches. A POSITA would understand
`
`Ressemann’s shaft 120 to be circular in cross section and in communication, and
`
`therefore j

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket