throbber

`
`
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`MEDTRONIC, INC. AND MEDTRONIC VASCULAR, INC.
`
`Petitioner,
`
`v.
`
`TELEFLEX INNOVATIONS S.A.R.L.
`
`Patent Owner.
`
`
`
`
`
`SUPPLEMENTAL DECLARATION OF
`STEPHEN JON DAVID BRECKER, MD, FRCP, FESC, FACC
`SUBMITTED IN SUPPORT OF PETITIONER’S REPLIES
`
`
`
`
`
`i
`
`
`
`
`
`IPR2020-00126/-127/-128/-129/-130/-132/-134/-135/-136/-137/-138
`
`Medtronic Ex. 1806
`Medtronic v. Teleflex
`
`Page 1
`
`

`

`
`
`Table of Contents
`
`Page
`
`Table of Contents ....................................................................................................... 2
`I.
`Overview .......................................................................................................... 4
`II.
`Claim Construction .......................................................................................... 4
`A.
`“interventional cardiology devices” ...................................................... 4
`
`B.
`
`C.
`
`“coaxial” ................................................................................................ 9
`
`Claim 25 of the ʼ380 Patent ................................................................. 14
`
`III.
`
`Claim 25: “Means for Receiving and Guiding” ....................... 14
`1.
`Claim 25: “Interventional Device” ........................................... 19
`2.
`Itou-Based Grounds .......................................................................................20
`A.
`Types of interventional cardiology devices insertable through
`catheter (2) ........................................................................................... 20
`
`B. Distal end protective catheter (5) ........................................................ 24
`
`C.
`
`D.
`
`E.
`
`Itou and Ressemann as prior art .......................................................... 25
`
`Itou and Ressemann collar 2141 ......................................................... 29
`
`Itou and Kataishi ................................................................................. 34
`
`IV. Ressemann-Based Grounds ...........................................................................36
`A.
`Ressemann discloses a “tubular structure defining a lumen
`coaxial and in fluid communication with the lumen of the guide
`catheter” ............................................................................................... 36
`
`B.
`
`C.
`
`D.
`
`E.
`
`Ressemann in view of Takahashi ........................................................ 39
`
`Ressemann collar 2141 ........................................................................ 42
`
`Ressemann anticipates the means-plus-function claims of the
`’380 patent ........................................................................................... 45
`
`Ressemann or Itou + Kataishi renders claim 27 of the ’380
`patent obvious ...................................................................................... 49
`
`2
`
`Page 2
`
`

`

`
`
`V.
`
`Kontos-Based Grounds ..................................................................................50
`A.
`Tube 16 of Kontos’s body 12 is not a narrow tube ............................. 50
`
`B.
`
`Body 12 of Kontos’s support catheter 10 has a “cross-sectional
`inner diameter through which interventional cardiology devices
`are insertable” ...................................................................................... 55
`
`C. Kontos necessarily resists axial and shear forces that would
`otherwise tend to dislodge the guide catheter. .................................... 57
`
`D.
`
`The side opening claims are obvious. ................................................. 61
`
`1.
`
`2.
`
`Replacing Kontos’s proximal funnel with a side opening would
`maximize the usable “real estate” inside the catheter
`assembly. ................................................................................... 61
`After removing Kontos’s proximal funnel, a posita would have
`been motivated to maximize the usable real estate, but even if
`no further modifications were made, Kontos would not have a
`“problem gap.” .......................................................................... 65
`After removing Kontos’s proximal funnel, Kontos’s support
`catheter 10 would remain coaxial with the guide catheter. ....... 69
`Even under patent owner’s interpretation, Kontos still has a
`“flexible cylindrical reinforced portion.” ............................................ 72
`
`3.
`
`E.
`
`F.
`
`A POSITA would have been motivated with a reasonable
`expectation of success to achieve the no-more-than-one-french
`differential between inner diameters of the guide catheter and
`body 12. ............................................................................................... 73
`
`G. My opinion remains unchanged: the claims of the ʼ776 patent
`that recite a two-inclined side opening are obvious. ........................... 76
`
`Kontos + Ressemann and Kontos + Ressemann + Takahashi .. 76
`1.
`Kontos + Ressemann + Kataishi ............................................... 78
`2.
`VI. Teleflex’s Evidence Regarding Secondary Considerations Does Not Alter
`My Opinion That the Challenged Claims Are Obvious. ...............................80
`VII. Conclusion .....................................................................................................83
`
`
`
`3
`
`Page 3
`
`

`

`
`
`I.
`
`Overview
`
`1.
`
`I have been retained by Robins Kaplan LLP on behalf of Medtronic,
`
`Inc. and Medtronic Vascular, Inc. (“Medtronic”) as an independent expert to
`
`provide my opinions concerning U.S. Patent Nos. 8,048,032; RE45,380;
`
`RE45,776; RE45,760; and RE47,379 (the “Teleflex Patents”) in IPR2020-00126,
`
`IPR2020-00127, IPR2020-00128, IPR2020-00129, IPR2020-00130, IPR2020-
`
`00132, IPR2020-00134, IPR2020-00135, IPR2020-00136, IPR2020-00137 and
`
`IPR2020-00138.1
`
`2.
`
`I set forth the information below as a supplement to my original
`
`declarations, as Teleflex raised new issues in their Patent Owner Responses on
`
`which I had not previously been given an opportunity to offer testimony. Having
`
`considered Teleflex’s arguments, and the testimony of Teleflex’s declarants, my
`
`opinions on the invalidity of the Teleflex Patents remain the same.
`
`II. Claim Construction
`
`“interventional cardiology devices”
`
`The opinions in this section generally relate to at least the following
`
`A.
`
`3.
`
`IPRs:
`
`
`1 Citations to exhibits refer to exhibits filed in IPR2020-00126, unless noted
`
`otherwise. I understand that most of Patent Owner’s and Petitioners’ exhibits are
`
`numbered consistently across all seven IPRs.
`
`4
`
`Page 4
`
`

`

`
`
` IPR2020-00126 (’032 patent)
`
` IPR2020-00127 (’032 patent)
`
` IPR2020-00128 (’380 patent)
`
` IPR2020-00130 (’380 patent)
`
` IPR2020-00135 (’776 patent)
`
`4.
`
`I am aware that when the Board instituted the IPR2020-00126 ’032
`
`petition it considered whether the limitation “interventional cardiology devices”
`
`required that “all four enumerated devices (guidewires, balloon catheters, stents
`
`and stent catheters)” be insertable into the lumen of the claimed “device for use
`
`with a standard guide catheter.” I.D., Paper 22, 10-13 (IPR2020-00126). This
`
`limitation appears in independent claims 1 and 11. This limitation also appears in
`
`claims 1 and 12 of the ’380 patent. (See I.D., Paper 22, 9-11) (IPR2020-00128).
`
`5.
`
`At institution, the Board determined that “interventional cardiology
`
`devices” refers to “at least two types of the devices selected from the group that
`
`includes but is not limited to guidewires, balloon catheters, stents, and stent
`
`catheters.” I.D., Paper 22, 12 (IPR2020-00126). Additionally, it determined that
`
`Medtronic demonstrated that the tubular portion of Itou’s catheter (2) has an inner
`
`diameter through which both guidewire (6) and distal end protective catheter (5)
`
`may be inserted. Id., 20.
`
`5
`
`Page 5
`
`

`

`
`
`6.
`
`I have reviewed Teleflex’s claim construction arguments, POR, 9-14
`
`(IPR2020-00126) (citing Ex-2138 (Declaration of Peter Keith) and Ex-2145
`
`(Declaration of Doctor John J. Graham)); POR, 7-10 (IPR2020-00135) (same), and
`
`I disagree.
`
`7.
`
`First, I disagree that claims 1 and 11 (’032 patent) or 1 and 12 (’380
`
`patent) require that “at least the set of four devices . . . guidewires, balloon
`
`catheters, stents, and stent catheters” be insertable into the claimed lumen. POR, 9
`
`(IPR2020-00126). I have reviewed the Teleflex patent and do not see that the
`
`claims require this. The claims do not, for example, recite “all interventional
`
`cardiology devices,” or “at least four interventional cardiology devices.”
`
`8.
`
`Second, I disagree that “interventional cardiology devices” should be
`
`construed to be limited to the four enumerated devices and to “any other device
`
`that is delivered beyond the end of the device for use with a standard guide catheter
`
`to a location in the coronary vasculature requiring treatment, to provide treatment
`
`to that location.” 2 POR, 11 (IPR2020-00126).
`
`9.
`
`I understand Teleflex’s argument to be, in part, that the patent
`
`specification uses the terms “cardiac treatment device” and “interventional
`
`
`2 For example, it is my opinion that the distal protection device described in Adams
`
`(Ex-1035 IPR2020-00126) is an interventional cardiology device.
`
`6
`
`Page 6
`
`

`

`
`
`cardiology treatment device” interchangeably with “interventional cardiology
`
`device[s].” POR, 12 (IPR2020-00126). I disagree that the patent uses these three
`
`terms as synonyms. The specification does not limit its discussion to treatment
`
`devices, or even to treatment devices and catheters that deliver those devices. For
`
`example, it explains that “[i]n order to diagnose and treat obstructive coronary
`
`artery disease it is commonly necessary to pass a guidewire or other instrument
`
`through and beyond the occlusion or stenosis of the coronary artery.” Ex-1001,
`
`1:26-29. And it also discusses that “the interventional cardiology art would benefit
`
`from the availability of a system that would be deliverable through standard guide
`
`catheters for providing backup support . . . .” Id., 2:45-49; and see, id., 1:7-8
`
`(describing the invention relating generally to catheters used in interventional
`
`cardiology procedures).
`
`10.
`
`I understand a “cardiac treatment device” and an “interventional
`
`cardiology treatment device” to be subsets of the broader term “interventional
`
`cardiology device[s].” A POSITA understands an “interventional cardiology
`
`device” as a much more general term to include devices that might be placed in the
`
`coronary artery that are not actually treatment devices themselves, for example
`
`intracoronary imaging catheters. However, a POSITA would not even limit the
`
`term “interventional cardiology device” in general only to the coronary arteries.
`
`There is an entire subspecialty within interventional cardiology for non-coronary
`
`7
`
`Page 7
`
`

`

`
`
`interventions. Examples of interventional cardiology devices that are not placed in
`
`a coronary artery would include, for example, intracardiac echocardiography
`
`catheters used for imaging the heart, percutaneous femoral closure devices,
`
`external vascular hemostasis devices, cardiac septal closure devices and cerebral
`
`protection devices amongst many others.
`
`11. Similarly, not even every intracoronary “interventional cardiology
`
`device” is used to provide treatment. Devices in this category include intravascular
`
`ultrasound catheter used for imaging the vessel, optical coherence tomography
`
`catheter used for imaging the vessel and fractional flow reserve measurement
`
`“pressure” wires. These devices are for diagnosis not treatment.
`
`12. As illustrated below, an “interventional cardiology treatment device”
`
`is but a subset of “interventional cardiology device(s).”
`
`
`
`8
`
`Page 8
`
`

`

`
`
`13. Only a subset of the devices specifically identified in the specification
`
`provide treatment. A guidewire does not provide treatment. It instead “provide[s] a
`
`rail over which a series of therapeutic devices can be advanced.” Ex-1015, 95. Dr.
`
`Graham’s testimony supports this. He acknowledges that a guidewire is used in
`
`conjunction with a stent or balloon in order to treat an occlusion. Ex-1801, 89:2-
`
`23.
`
`IPRs:
`
`B.
`
`“coaxial”
`
`14.
`
`The opinions in this section generally relate to at least the following
`
` IPR2020-00127 (ʼ032 Patent)
`
` IPR2020-00130 (ʼ380 Patent)
`
` IPR2020-00134 (’760 Patent)
`
`15.
`
`I understand that Teleflex disputes that Ressemann discloses “a
`
`“tubular structure defining a lumen coaxial and in fluid communication with the
`
`lumen of the guide catheter.” POR, 9-13 (IPR2020-00134) (citing Ex-2138
`
`(Declaration of Peter Keith) and Ex-2145 (Declaration of Doctor John J. Graham)).
`
`Teleflex’s position is that Ressemann’s evacuation sheath includes an evacuation
`
`head with a “multi-lumen tube,” and, as such, the lumen of the tubes are not
`
`coaxial with the lumen of the guide catheter because neither is perfectly concentric
`
`9
`
`Page 9
`
`

`

`
`
`with the lumen of the guide catheter. The images below are reproduced from
`
`Teleflex’s brief and Mr. Keith’s Declaration.
`
`IPR2020-00134, POR at 12; Ex-2138, 65-66.
`
`16.
`
`I disagree. A POSITA does not understand “coaxial” to refer to
`
`perfect concentricity, so that the catheters share exactly the same center line. I have
`
`reviewed the deposition testimony of Mr. Keith, who, now, agrees with this. Ex-
`
`1764, 91:10-18; 94:11-19; Ex-1800, 23:19-24:1; Ex-1805, 107:19-109:6; 117:14-
`
`121:1.
`
`17.
`
`Besides reciting the coaxial arrangement articulated above, the
`
`claims also recite that the guide catheter is positioned adjacent the coronary
`
`ostium, and the guide extension catheter is inserted into the guide catheter to be
`
`advanced through it and partially into the coronary artery. This configuration
`
`allows for the guide extension to receive a stent or balloon catheter. Ex-1601, claim
`
`25 (IPR2020-00134).
`
`10
`
`Page 10
`
`

`

`
`
`18.
`
`This is similar to the disclosure in the patent, which shows a guide
`
`extension catheter inserted into a guide catheter, where the two are in line with the
`
`coronary artery. Id., Fig. 8; 8:10-16.
`
`19.
`
`I am aware that Mr. Keith offered a definition of “coaxial” during
`
`his deposition, which is that “coaxial means essentially a tube-in-a-tube
`
`arrangement that is relatively close-fitting.” Ex-1764, 98:18-21. He was not,
`
`however, able to specify how close fitting “relatively close-fitting” is. Id., 98:22-
`
`99:12. He also testified that his opinion was that the tubes themselves have “radial
`
`symmetry,” id., 101:20, but I see nothing in the patent claims or specification that
`
`would require radial symmetry of the tubes themselves.
`
`20. While the specification describes a 1 French relationship between
`
`the guide catheter and guide extension catheter, not all of the claims of all of the
`
`patents require this. Compare, e.g., ’760 patent (cl. 25, 13:6061) with ’380 patent
`
`(cl. 1), ’032 patent (cl. 1). Thus, “coaxial” cannot require a 1 French differential
`
`between the lumen of the guide extension catheter and the lumen of the guide
`
`catheter. Dr. Graham, Teleflex’s cardiologist, agrees. Ex-1801, 37:22-28:2. So does
`
`Mr. Keith, who testified that a 4-in-6 or a 5-in-7 relationship would still be coaxial.
`
`Ex-1800, 23:19-24:1; Ex-1805, 119:4-121:1.
`
`11
`
`Page 11
`
`

`

`
`
`Ex-1117.
`
`
`
`21.
`
`The import of these relationships is shown below. The larger the
`
`differential between the OD of the extension catheter and the ID of the guide
`
`catheter, the more room there is within the guide catheter for the extension catheter
`
`to drift off-center, such that its axis is not perfectly concentric with that of the
`
`guide catheter.
`
`22.
`
`The schematics below illustrate a 6 French guide catheter described
`
`in the patent, which teaches that 6 French guide catheters have IDs of > 0.070
`
`inches. Ex-1601, 3:42-44 (IPR2020-00134).
`
`23.
`
`On the left, the guide extension catheter is 5 French, which a
`
`POSITA understands to correspond to 5 x (0.33) = 1.65 mm, which is 0.065 inches.
`
`On the right, the guide extension catheter is 4 French, which a POSITA
`
`understands is 4 x (0.33) = 1.32 mm = 0.052 inches.
`
`12
`
`Page 12
`
`

`

`
`
`
`
`
`
`
`
`24.
`
`Thus, “coaxial” must encompass both what the specification
`
`discloses for a 5-in-6 relationship (an 0.005 inch annular gap), as well as the
`
`admitted 4-in-6 relationship (an 0.18 inch annular gap). This is consistent with the
`
`understanding of a POSITA. E.g. Ex-1840, 1 (describing using a 4 Fr angiographic
`
`catheter within a 6 Fr guide catheter in a coaxial fashion); Ex-1841, 1 (describing a
`
`coaxial double catheter technique using an 8 Fr guide catheter and a 5 Fr
`
`diagnostic catheter).
`
`25.
`
`I am also aware that one of the inventors, Mr. Root, testified
`
`
`
`
`
`68:2 86:1-87:7; Ex-2015, 37:10-17. This is also consistent with the testimony of
`
`Teleflex’s experts, as well as the understanding of a POSITA.
`
` Ex-1114, 66:22-
`
`13
`
`Page 13
`
`

`

`
`
`26. Given the claims and specification of the patent, consistent with the
`
`understanding of a POSITA, “coaxial” means the lumens are aligned. Specifically,
`
`it means that the “axis of the lumen of the guide extension catheter is aligned in the
`
`same direction as the axis of the lumen of the guide catheter.”
`
`C. Claim 25 of the ʼ380 Patent
`
`27.
`
`The opinions in this section generally relate to at least the following
`
`IPRs:
`
` IPR2020-00134 (’760 Patent)
`
`1.
`
`Claim 25: “Means for Receiving and Guiding”
`
`28.
`
`Claim 25 of the ʼ380 patent recites a “means for receiving an
`
`interventional device . . . and guiding the interventional device deeper into the
`
`branch vessel.” After the means limitation, “claim 25 provides an extensive
`
`recitation of structure.” I.D., Paper 22, 18 (IPR2020-00129). I understand that at
`
`Institution, the Board explained that “[w]hether the structure identified in claim 25
`
`are sufficient to perform the two recited functions of the means for receiving and
`
`guiding … is a disputed, material issue of fact …. [that] is best resolved upon a full
`
`trial record.” Id., 18. Having reviewed the testimony of Mr. Keith and Dr. Graham,
`
`I still believe that claim 25 recites sufficient structure to overcome the means-plus-
`
`function presumption.
`
`14
`
`Page 14
`
`

`

`
`
`a.
`
`The means-plus-function presumption is overcome because
`Claim 25 recites sufficient structure.
`
`29.
`
`I have been informed by counsel that if a “means” claim recites
`
`sufficient structure for performing the claimed function, then that limitation is not
`
`a means-plus-function limitation.
`
`30.
`
`I understand that Mr. Keith and Dr. Graham believe that the means-
`
`plus-function presumption for claim 25 is not overcome because the claim
`
`language allegedly “does not recite that the tip portion and the reinforced portion
`
`have a lumen that communicates with the claimed opening into which an
`
`interventional device can be received and be guided ‘deeper into the branch
`
`vessel.’” Ex-2138, ¶ 108 (IPR2020-00129); Ex-2145, ¶ 97 (IPR2020-00129). I
`
`disagree and believe that a POSITA reviewing just the language of claim 25 would
`
`conclude that the reinforced portion and tip portion have a lumen.
`
`31.
`
`Although Mr. Keith agrees that the distal end of the claimed side
`
`opening has a full circumference portion, he believes structure is still lacking from
`
`claim 25 because “[t]here is no claim language requiring a lumen extending
`
`through the reinforced portion and the tip portion.” Ex-2138, ¶109 (IPR2020-
`
`00129). I do not agree with Mr. Keith’s assessment. Mr. Keith’s conclusion does
`
`not make sense, as there would be no reason to have a side opening if its lumen
`
`terminated prior to the reinforced and tip portions. I am unaware of any catheter
`
`assembly (i.e., means for receiving and guiding) having a side opening where the
`
`15
`
`Page 15
`
`

`

`
`
`lumen terminates prior to the distally-located portions (i.e., the sections found
`
`distal to the side opening). In my opinion, there would be no reason to have a side
`
`opening if the lumen terminated prior to the distally-located portion(s). I cannot
`
`think of any function or reason to have a catheter assembly with a side opening that
`
`terminates into a lumenless portion.
`
`32.
`
`Further, by the time of the alleged invention, reinforcing catheters—
`
`in particular, with metallic braiding or coiling—was common practice in the art.
`
`E.g. Ex-1008, 7:4-7. The reason for reinforcement is to allow the catheter to
`
`maintain its shape in regions of tortuosity and calcification. E.g. id., 6:66-7:4. In
`
`other words, by recited the claim language “reinforced portion,” a POSITA would
`
`have understood that a lumen was found therein, otherwise there would be no need
`
`for reinforcement.
`
`33.
`
`A POSITA would also expect the tip portion to have a lumen. This is
`
`especially true because if the side opening and reinforced portion both have
`
`lumens, it would be illogical for the tip portion to not also have a lumen (i.e., dead
`
`end). Further, a POSITA knows that the only way to receive and guide a device
`
`into the vasculature would require the tip portion to have a lumen. As a result,
`
`based on the claim language alone, I believe that claim 25 describes that the
`
`reinforced and tip portions have a lumen, and the means-plus-function presumption
`
`is overcome.
`
`16
`
`Page 16
`
`

`

`
`
`34.
`
`I do not believe it is necessary to do so here, but the specification of
`
`the ʼ380 patent, if consulted, reinforces my conclusion that the means-plus-
`
`function presumption is overcome. For example, each Figure showing the
`
`reinforced portion 18 or the tip portion 16 show a lumen. See Ex-1001, Figs. 1, 3,
`
`4. The specification also explains that interventional devices are “inserted through
`
`the lumen of coaxial guide catheter 12.” Id., 10:16-20. And “[t]he coaxial guide
`
`catheter includes a tip portion, a reinforced portion, and a substantially rigid
`
`portion.” Id., 4:51-53. There are no examples or discussion of a lumenless
`
`reinforced portion or tip portion in the specification of the ʼ380 patent. In other
`
`words, if the specification of the ʼ380 patent is considered, it further supports my
`
`conclusion that the reinforced and tip portion of claim 25 have a lumen.
`
`Accordingly, I am still of the opinion that the means-plus-function-presumption is
`
`overcome.
`
`b.
`
`If the means-plus-function presumption is not overcome, the
`corresponding structure is an extension catheter lacking the
`unnecessary structure added by PO.
`
`35.
`
`If the means-plus-function presumption is not overcome, the
`
`corresponding structure for the means for receiving and guiding is an extension
`
`catheter.
`
`17
`
`Page 17
`
`

`

`
`
`36.
`
`I understand that Mr. Keith and Dr. Graham agree that the
`
`corresponding structure in the specification is a guide catheter, but they believe that
`
`the guide catheter must also have the following structure:
`
`a distal tubular portion with a single lumen3 that is coaxial with the
`
`lumen of the guide catheter attached to a substantially rigid rail
`
`structure that allows interventional devices to be advanced alongside
`
`and into the lumen of the tubular portion, with the total length of the
`
`device being longer than the guide catheter, and at least a distal part
`
`of the tubular portion is flexible.4
`
`Ex-2138, ¶114 (IPR2020-00129) (emphasis added); Ex-2145, ¶¶99-100 (IPR2020-
`
`00129).
`
`37.
`
`I have been informed by counsel that when identifying the
`
`corresponding structure, a POSITA should only incorporate structure that is
`
`necessary to perform the claimed function. Here the claimed function is “receiving
`
`the interventional device from an intermediate or distal portion of the [guide
`
`catheter . . . and guiding the interventional device deeper into the branch vessel.”
`
`
`3 I note that the specification never affirmatively states that the described guide
`
`catheter is limited to a single lumen.
`
`4 For ease of discussion, I refer to the extension catheter with Patent Owner’s
`
`additional recitation of structure as “Extension Catheter + Structure.”
`
`18
`
`Page 18
`
`

`

`
`
`Ex-1001, claim 25 (IPR2020-00129). To perform this means for receiving and
`
`guiding, I do not believe that the additional structure recited by Patent Owner is
`
`necessary. In particular, I do not agree with Patent Owner’s position that the
`
`receiving and guiding function requires an extension catheter with a coaxial tubular
`
`portion and a single lumen.
`
`38.
`
`For example, Ressemann is an example of a non-coaxial (under
`
`Patent Owner’s construction),5 multi-lumen catheter that receives and guides an
`
`interventional device. Ex-1008, Figs. 6A-6F.
`
`39.
`
`As a result, if the means-plus-function is not overcome, the
`
`corresponding structure for the means for receiving and guiding is a guide catheter
`
`that does not include Patent Owner’s additional recitation of structure.
`
`2.
`
`Claim 25: “Interventional Device”
`
`40. Other than in claim 25 (and its dependent claims), the phrase
`
`“interventional device” is not used in the ʼ380 patent. The plain and ordinary
`
`meaning of an “interventional device” is any device used in an interventional
`
`
`5 As set forth above, I disagree with Patent Owner’s construction of coaxial.
`
`§ II(B), supra. It is my opinion that Ressemann’s evacuation assembly is coaxial to
`
`the guide catheter. See, e.g., § IV(A), infra.
`
`19
`
`Page 19
`
`

`

`
`
`procedure. Under this definition, Itou’s protective catheter 5 is an interventional
`
`device.
`
`III.
`
`Itou-Based Grounds
`
`41. For the purpose of my analysis I have been asked to assume that Itou
`
`is prior art to the Teleflex patents.
`
`A. Types of interventional cardiology devices insertable through
`catheter (2)
`
`42.
`
`The opinions in this section generally relate to at least the following
`
`IPRs:
`
` IPR2020-00126 (’032 patent)
`
` IPR2020-00128 (’380 patent)
`
` IPR2020-00129 (ʼ380 patent)
`
` IPR2020-00132 (’760 patent)
`
` IPR2020-00135 (’776 patent)
`
`43. Assuming that the claims are construed to require that two
`
`interventional cardiology devices must be insertable through the claimed lumen,
`
`Itou satisfies this limitation.
`
`44. Even if Teleflex were correct both as to its claim construction position
`
`and as to the “effective size” of the tubular structure of catheter (2) being 0.046
`
`inches (and I disagree for the reasons discussed below), Itou nonetheless inherently
`
`20
`
`Page 20
`
`

`

`
`
`discloses a lumen through which all four, enumerated interventional cardiology
`
`devices (guidewires, balloon catheters, stents and stent catheters) are insertable.
`
`45.
`
`Itou teaches that guidewire (6) is insertable through the lumen of
`
`catheter (2). Ex-1007, Fig. 5; 4:64-65; 5:11-20. Guidewires were known to range in
`
`size from 0.010 to 0.018 inches. Ex-1015, 98. They would therefore be insertable
`
`through catheter (2) even if its “effective size” were 0.046 inches.
`
`46. Similarly, even if the “effective size” of the opening to the lumen of
`
`catheter (2) were 0.046 inches, angioplasty catheters were known to be insertable
`
`through catheters with lumen of 0.045 inches. Ex-1009, 4:49-63; Ex-1833, 1.
`
`47.
`
`There were also a variety of stents with crossing profiles that were
`
`small enough to be advanced through a lumen of 0.046 inches (1.16 mm) that I
`
`discuss below. Necessarily, each of the stents that are discussed were advanced into
`
`the coronary vasculature on a “stent catheter.”
`
`48. Baim explains that “[a]ll current slotted tube designs are ‘bare
`
`mounted’ on a delivery balloon, with deflated profiles smaller than 0.040-in.
`
`(1mm)” Ex-1015, 189.
`
`49.
`
`The 4th edition of the Handbook of Coronary Stents disclosed the
`
`following: at pp. 4, 7 (balloon-expandable Genic® stent with less than 0.9 mm
`
`(0.035 inch) profile), 15-16 (balloon-mounted Lunar stent with 0.0382 inch
`
`profile), 21 (Spiral Force stent with 0.039 to 0.042 inch profile and “Runner” stent
`
`21
`
`Page 21
`
`

`

`
`
`delivery balloon catheter), 25 (balloon-expandable Tsunami stent with 0.95 mm
`
`(0.038 inch) profile). Ex-1802.
`
`50.
`
`Similarly, US Pat. No. 6,726,714 taught a stent crimped against a
`
`balloon with a “relatively small loaded diameter having an OD of 0.44 inches.” Ex-
`
`1803, 4:54-56.
`
`51. And an article by Colombo described several balloon-expandable
`
`stents with crossing profiles well under 0.046 inches, including those with profiles
`
`of 0.99 mm (0.038 inch), 0.93 mm (0.036 inch) and 0.84 mm (0.033 inch)). Ex-
`
`1804, Table 1; Fig. 3.
`
`52.
`
`In sum, all of the devices I describe above are insertable through
`
`catheter (2), even if the “effective size” of its opening was 0.046 inches (1.16 mm)
`
`(3.48 Fr). I disagree, however, that the lumen of catheter (2) has an effective size of
`
`0.046 inches.
`
`53. Teleflex’s argument that Itou’s catheter (2) does not have a lumen that
`
`is 1.5 mm (0.059 inches), as Itou teaches in Table 1, but, instead is effectively only
`
`1.16 mm (0.046 inches) is based on the idea that because Itou’s wire (25) has a
`
`diameter of 0.45 mm (0.004 inches), it partially blocks the opening to lumen (2).
`
`Ex-2138, ¶129 (IPR2020-00126).
`
`54.
`
`This calculation does not make sense to a practicing interventional
`
`cardiologist. It assumes, first, that catheter (2) is entirely rigid, such that wire (25)
`
`22
`
`Page 22
`
`

`

`
`
`cannot be bent away from the angled proximal opening of the tubular portion of
`
`catheter (2).
`
`55.
`
`Itou explicitly teaches this is not the case, as it discloses that catheter
`
`(2)—and in particular its distal tubular portion— must be navigated into the
`
`coronary vasculature. Ex-1007, Abstract (“[A]ssembly is insertable into a blood
`
`vessel . . . .”); 1:66-2:5 (teaching that catheter is advanced to a “deep location in a
`
`coronary artery”); Fig. 6; 5:29-42. This means that catheter (2) is necessarily
`
`flexible, including in the region of the junction between the distal end of wire 25
`
`and the proximal part of catheter (2)’s tubular structure. A POSITA understands
`
`that an interventional cardiology device may be introduced into catheter (2) at an
`
`angle.
`
`56. Additionally, Itou teaches that wire (25) is flattened at its distal most
`
`end, “crushed into the form of a flat plate.” Ex-1007, 4:33-36. This is an additional
`
`reason that a POSITA would not understand that wire 25 reduces the effective size
`
`of the opening of the tubular portion of catheter (2). The flattened area adjacent to
`
`the proximal opening —in conjunction with catheter (2)’s flexibility—means that
`
`the entirety of the opening of catheter (2) will be accessible.
`
`57. My opinion that the claims of the ’032, ’380, ’760, and ’776 patents
`
`are invalid remains unchanged.
`
`23
`
`Page 23
`
`

`

`
`
`IPRs:
`
`B. Distal end protective catheter (5)
`
`58.
`
`The opinions in this section generally relate to at least the following
`
` IPR2020-00126 (’032 patent)
`
` IPR2020-00128 (’380 patent)
`
` IPR2020-00129 (ʼ380 patent)
`
` IPR2020-00135 (’776 patent)
`
`59.
`
`Itou describes inserting catheter (5) into catheter (2), and then
`
`advancing the two catheters, together, into guiding catheter (1) and into the
`
`vasculature. Ex-1007, 5:43-46; 7:13-18. In practice, however, a POSITA would
`
`advance catheters (2) and (5) together, until catheter (2) was in a desired location.
`
`As taught by Itou, catheter (5) would be withdrawn before catheter (2) would be
`
`used, either for suction, or to deliver a different interventional cardiology device.
`
`60. Not uncommonly during a procedure an interventional cardiologist
`
`will find that he needs to advance a catheter more deeply into the vasculature. In
`
`this instance, catheter (2) would be left in place, and catheter (5) would be re-
`
`introduced by insertion into the guide catheter, and advancement into catheter (2),
`
`while catheter (2)’s proximal opening remained in the guide catheter. Only then
`
`would it be safe to maneuver catheter (2) to a deeper location.
`
`24
`
`Page 24
`
`

`

`
`
`61. Sequential delivery would be an obvious alternative to preassembly,
`
`albeit not routine.
`
` Ex-1115, 165:3-25.
`
`
`
`62. My opinion that the claims of the ’032, ’380, and ’776 patents are
`
`invalid remains unchanged.
`
`Itou and Ressemann as prior art
`
`The opinions in this section generally relate to at least the following
`
`C.
`
`63.
`
`IPRs:
`
` IPR2020-00126 (’032 patent)
`
` IPR2020-00128 (’380 patent)
`
` IPR2020-00132 (’760 Patent)
`
` IPR2020-00135 (’776 patent)
`
` IPR2020-00137 (’379 patent)
`
`64.

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket