throbber
This guidance was written prior to the February 27, 1997 implementation of FDA's
`Good Guidance Practices, GGP's. It does not create or confer rights for or on any person
`and does not operate to bind FDA or the public. An alternative approach may be used if
`such approach satisfies the requirements of the applicable statute, regulations, or both.
`This guidance will be up dated in the next revision to include the standard elemnt s of GGP 's .
`
`ALL 2078
`PROLLENIUM V. ALLERGAN
`IPR2019-01505 et al.
`
`

`

`SHELF LIFE OF MEDICAL DEVICES
`
`April 1991
`
`Geoffrey S. Clark
`Microbiologist
`Division of Small Manufacturers Assistance
`Office of Training and Assistance
`Center for Devices and Radiological Health
`Food and Drug Administration
`
`

`

`TABLE OF CONTENTS
`
`Page
`
`1
`
`1
`
`4
`
`.
`
`6
`
`
`
`10
`
`13
`
`23
`
`2 4
`
`. . . . . . . . . . . . . . . . . . . .
`. . . . . . . . . .
`. . . . . . . . . . . . . . . . . . .
`. . . . . . . . . . . . . . . . . . . . .
`. . . . . . . . . . . . . . . . . . . .
`. . . . . . . . . . . . . . . . . . . .
`. . . . . . . . . . . . . . . . . . . . . .
`. . . . . . . . . . . . . . . . . . . . .
`
`INTRODUCTION
`
`STABILITY CRITERIA AND VARIABLES
`
`RECOMMENDATIONS
`
`REGULATIONS
`
`FDA POLICIES
`
`CDRHGUIDANCE
`
`GLOSSARY
`
`REFERENCES
`
`NOTE :
`
`The excerpts of documents included throughout this
`
`paper have been edited .
`
`

`

`INTRODUCTION
`
`The potential benefit from the use of a medical device ranges
`from relieving minor irritations to correcting life threatening
`conditions. If the device design and manufacturing processes are
`done adequately, there is a high probability that the device will
`perform as desired at the time it is manufactured. However,
`there are many naturally occurring factors that can affect how
`long after manufacturing the device will maintain the ability to
`fully perform the intended function.
`
`Shelf life is the term or period during which a commodity remains
`suitable for the intended use. An expiration date is the
`termination of shelf life, after which a percentage of the
`commodity, e.g., medical devices, may no longer function as
`intended. To determine if a particular device requires a shelf
`life and assign an expiration date, there are a number of
`different parameters that must be considered. The device must be
`analyzed to determine if it is susceptible to degradation that
`would lead to functional failure and the level of risk that the
`failure would present. For some devices, e.g., tongue
`depressors, it is not reasonable to assign a shelf life because
`of the small likelihood of time-dependant product degradation and
`the lack of serious consequences if it did fail to perform as
`designed. For certain devices susceptible to degradation that
`are intended to treat life-threatening conditions, e.g.,
`pacemakers, the failure rate should approach zero within the
`labeled shelf life.
`
`The purpose of this document is to:
`
`inform readers of the Food and Drug Administration (FDA)
`regulations and policies relating to shelf life of medical
`devices.
`
`discuss the various parameters that determine the length of
`time a particular device will remain within acceptable
`specifications;
`
`outline the different activities that can be undertaken to
`establish the shelf life of a device; and
`
`STABILITY CRITERIA AND VARIABLES
`
`The United States Pharmacopoeia (USP) defines stability as "the
`extent to which a product retains, within specified limits, and
`throughout its period of storage and use, i.e., its shelf life,
`the same properties and characteristics that it possessed at the
`time of manufacture." There is no one exhaustive set of criteria
`that would apply equally to all medical devices. The USP has a
`section <1191> entitled "Stability Considerations in Dispensing
`Practicen that supplies general information on this topic. It
`includes a list of five sets of criteria for acceptable levels of
`stability for drug products as follows on the next page:
`
`

`

`1. chemical,
`physical,
`2.
`microbiological,
`3.
`therapeutic, and
`4 .
`5. toxicological.
`
`Although this set of criteria applies specifically to the
`evaluation of drug product stability, it is useful as a starting
`point in developing a set of criteria to evaluate the stability
`of medical devices.
`
`The following outline may be useful in identifying parameters
`that could significantly affect the shelf life of a device, even
`though all of the criteria will not apply to every device. This
`outline is based on the criteria listed above with the addition
`of biocompatibility.
`
`Chemical
`
`Degradation: Do any active ingredients or components of the
`device degrade over time in a manner which adversely affects
`device safety or performance?
`
`Interactions: Do ingredients or components interact to
`alter the device? Does the device have interactions among
`the various components that cause degradation of the ability
`to perform the intended function?
`
`Device and Packaging Interaction: Is there interaction
`between the device and package that has undesirable affects?
`
`Radioactive Decay: Does the device contain radioactive
`material with a relatively short half-life? Do the
`radioactive decay by-products alter the safety or
`effectiveness of the device either by themselves or through
`further interaction?
`
`Manufacturing: Do any of the manufacturing processes alter
`the chemistry of the raw materials, components, or finished
`device in a manner which adversely affects device Safety or
`performance?
`
`Physical
`
`Physical Characteristics: Does the device have physical
`characteristics that vary with time; e.g., appearance,
`viscosity, elasticity, tensile strength, burst strength, or
`electrical resistance? In some cases, a significant change
`in appearance may cause concern to the user even though the
`performance of the device is not affected.
`
`

`

`2.2 Manufacturing Process: Do the different steps in the
`manufacture of the device affect the physical
`characteristics of the individual components or the finished
`device in a manner which adversely affects device safety or
`performance?
`
`2.3 Storage Conditions: Do the storage conditions, e.g.,
`temperature, humidity, light, etc., have an affect on the
`device in a manner which adversely affects device safety or
`performance?
`
`3.
`
`Microbiological
`
`3.1 Sterility: Do sterile devices remain sterile? Maintenance
`of sterility is primarily determined by the maintenance of
`package and seal integrity.
`
`3.2 Environmental Control: Is an environmental control program
`needed during manufacturing or storage to monitor and adjust
`the microbial load in or on the device and packaging below
`an established tolerance level to prevent adverse
`degradation of the product?
`
`3.3 Antimicrobial Effectiveness: Does the device lose the
`ability to perform the intended antimicrobial function?
`
`3.4 Integrity: Do the device's barrier characteristics change?
`
`3.5 Preservative Effectiveness: If the device uses a
`preservative system, how long does the preservative system
`retain effectiveness within tolerance levels?
`
`4 .
`
`Therapeutic: Does the ability of the device to perform the
`intended therapeutic or diagnostic function change under
`storage or use conditions?
`
`5. Toxicological: Do device degradation by-products form
`during storage or use that produce an adverse toxic effect?
`
`6. Biocompatibility: Does the biocompatibility of the product
`change adversely during storage or use?
`
`There are numerous variables that affect the shelf life of a
`medical device. Some of these were discussed in the stability
`criteria outline above while additional product dependent
`variables are listed below. Although manufacturers may not be
`able to control all of the variables, their affect on device
`performance can be minimized if properly considered. Each of the
`categories listed below should be addressed during the
`preproduction estimation of product shelf life and the actual
`determination of the shelf life for normal production units. A
`written procedure for determining the necessity of a shelf life
`and setting the shelf life of a finished medical device should be
`used to avoid missing an important aspect and to provide needed
`
`

`

`documentation. While developing these written procedures for
`determining a shelf life of a device, the stability criteria and
`the product specific variables that follow should be assessed.
`
`1. Storage conditions, e.g.! temperature variations, relative
`humidity, ventilation, a n pressure, air-borne
`contamination, visible light and other radiation, etc.
`
`2.
`
`3.
`
`The nature of the device and intended use, e.g., medical
`gloves are made of latex because of the intended use even
`though latex deteriorates with age.
`
`The components used to manufacture the device, e.g., some
`devices contain a battery or other components that would
`suffer degradation of function with the passage of time.
`
`4. Method of manufacture, e.g., an in vitro diagnostic device
`that is aseptically packaged may have a shorter shelf life
`than one that is terminally sterilized.
`
`5. Packaging, e.g., products that are packaged in different
`size containers may each have a different stability shelf
`life due to the different ratio of product to package
`surface area contact.
`
`6. Transportation conditions; e.g., vibration, shock,
`temperature, humidity, etc.
`
`RECOMMENDATIONS
`
`The concept of shelf life or expiration dating should be
`incorporated into product reliability during the process of
`developing or improving a device for commercial distribution.
`The result is to make sure that, if the device is used in
`accordance with the labeling, the device will perform in the
`intended manner. Developing an appropriate set of specifications
`for the device's characteristics and assigning tolerance values
`for these characteristics is essential to the process of shelf
`life determination. The events that can cause a device to no
`longer perform in the intended manner may originate:
`
`a
`
`a
`
`internally, e.g., device component interactions or
`degradation can cause the device operating characteristics
`to fall outside of the prescribed tolerances; or
`
`externally, e.g., the shipping or storage conditions can
`cause a breakage in the device, a failure of the barrier
`properties of a sterile package or degeneration of the
`device itself.
`
`A device's "shelf lifew should not be confused with a device's
`"useful life.'' The useful life of a device is the duration of
`actual use or the number and duration of repeat uses before some
`change results in the device's inability to achieve its intended
`function.
`
`

`

`Establishing a Shelf Life
`
`The best time to begin considering device shelf life is during
`preproduction planning and review while the device is being
`formulated before any concrete decisions about the device have
`been made and changes can be readily adapted. When design,
`material and process factors that influence shelf life are
`identified early in the process, redesigning the device
`parameters involve the least cost. The materials, components and
`packaging that are to be used in the manufacture of the device
`may need to be examined for their individual shelf life
`characteristics in addition to their effect on the shelf life of
`the finished device. Some materials and components may need
`special handling to maintain their characteristics within the
`desired specifications. Also, the intended use of the device is
`an important consideration because this will greatly influence
`the tolerance level of shelf life or other failure of the device.
`
`Begin by establishing a target shelf life for the finished device
`that allows adequate time for shipping, storage and use.
`Evaluate the proposed materials and components used to produce
`and package the device. Also, a review of the literature should
`be conducted and, where feasible, data for similar devices
`collected. The method of manufacturing may need to be tailored
`to the materials or different materials may need to be selected
`that don't interfere with, and are not affected by, the
`manufacturing process. For devices that are found to have good
`stability characteristics, larger production runs may be
`practical to permit savings in production costs. With all
`devices there must be sufficient inventory control to produce,
`ship and sell the device within the shelf life period to avoid
`the monetary losses associated with recovering an out-of-date
`product, reworking a product, or otherwise disposing material to
`prevent unnecessary health risks for the consumer. If the shelf
`life considerations are implemented properly, they will add to
`product quality and actually decrease total production cost.
`
`For devices that are intended to be sterilized, the affect of the
`sterilization procedure on both the device and the package must
`be considered. It is much easier to rule out a package with
`water soluble sealants and select packaging that is compatible
`with steam sterilization, if steam sterilization is selected as
`the most desirable procedure before a contract with a package
`supplier is signed. There are numerous reference works available
`that cover packaging technology and various sterilization
`technologies.
`
`Procedure for Testing Shelf Life
`
`A written procedure for establishing and monitoring shelf life of
`medical devices should include the following:
`
`1. Organizational units responsible for the various phases of
`the shelf life testing program should be included in the
`written procedure.
`
`

`

`A Finished Device Sampling Plan, including the purpose for
`collecting the samples, the number of finished devices to be
`collected, frequency of sampling, sample selection criteria,
`and lots to be sampled.
`
`Raw Materials, Components and Packaging Evaluation Plan, to
`determine if any of these have their own individual shelf
`life considerations and how they affect the shelf life of
`the finished device. This plan should include obtaining
`data from suppliers, obtain new data as appropriate, and
`recording data for current and future designs.
`
`A Plan for Storage of Shelf Life Samples, including storage
`conditions and the environmental conditions to be
`controlled, monitored and recorded. .
`
`Accelerated Aging Parameters, including information that
`validates the accelerated system. The results need to be
`supported by real time testing of shelf life samples to
`confirm the tentative shelf life data collected from the
`accelerated tests.
`
`Simulation of Shipping and Handling Stresses Plan, including
`vibration tests, temperature extremes challenge, actual
`shipping and intentionally mishandling the device to
`determine the affect of unusual circumstances.
`
`Follow-up procedures should be included in the written plan
`that outline the steps to be taken based upon the results of
`the shelf life testing. These procedures must also cover
`conspicuously placing the expiration date on appropriate raw
`materials in storage and on the finished device to aid in
`stock rotation. The results of the shelf life testing may
`indicate a need to set limits on the time the device is held
`in storage, redesign the packaging of the device, or control
`the environmental conditions to minimize the degradation.
`
`REGULATIONS
`
`The Code of Federal Regulations (CFR) is a codification of the
`general and permanent rules published in the Federal Register by
`the Executive Departments and Agencies of the Federal Government.
`FDA regulations are grouped in Title 21 of the CFR, and the
`regulations that govern medical devices are contained in Parts
`800 to 1299. The regulations that apply specifically to shelf
`life and expiration dating of medical devices are listed below:
`In Vitro Diagnostic Regulation - 21 CFR Part 809
`1. The paragraph below outlines the labeling requirements for
`the primary container or wrapper of the retail package.
`
`21 CFR 809.10(a)(5): For a reagent, appropriate storage
`instructions adequate to protect the stability of the
`product. When applicable, these instructions shall include
`
`

`

`such information as conditions of temperature, light,
`humidity, and other pertinent factors. For products
`requiring manipulation, such as reconstitution and/or mixing
`before use, appropriate storage instructions shall be
`provided for the reconstituted or mixed product which is to
`be stored in the original container. The basis for such
`instructions shall be determined by reliable, meaningful,
`and specific test methods such as those described in
`211.166 of this chapter.
`
`2.
`
`The paragraph below outlines the information that is to be
`included in the package insert or other labeling that
`accompanies the product.
`
`21 CFR 809.10(b)(5)(iv): Appropriate storage instructions
`adequate to protect the stability of the product. When
`applicable, these instructions shall include such
`information as conditions of temperature, light( humidity,
`and other pertinent factors. For products requiring
`manipulation, such as reconstitution and/or mixing before
`use, appropriate storage instructions shall be provided for
`the reconstituted or mixed product. The basis for such
`instructions shall be determined by reliable, meaningful,
`and specific test methods such as those described in §
`211 .I66 of this chapter.
`
`3.
`
`The paragraph below describes a requirement for the labeling
`of Itgeneral purpose laboratory reagentsM.
`
`21 CFR 809.10(d)(l)(v): Appropriate storage instructions
`adequate to protect the stability of the product. When
`applicable, these instructions shall include such
`information as conditions of temperature, light, humidity,
`and other pertinent factors. The basis for such
`instructions shall be determined by reliable, meaningful,
`and specific test methods such as those described in §
`211.166 of this chapter.
`. The paragraph below from the drug regulation is referred to
`
`in the three regulations above and specifically requires a
`written stability testing program that may include
`accelerated shelf life testing, but also indicates that real
`time testing must be performed to verify the validity of the
`accelerated testing results.
`
`21 CFR 211.166: (a) There shall be a written testing
`program designed to assess the stability characteristics of
`drug products. The results of such stability testing shall
`be used in determining appropriate storage conditions and
`expiration dates. The written program shall include: (1)
`Sample size and test intervals based on statistical criteria
`for each attribute examined to assure valid estimates of
`stability; (2) Storage conditions for samples retained for
`testing; (3) Reliable, meaningful, and specific test
`methods; ( 4 ) Testing of the drug product in the same
`
`

`

`c o n t a i n e r - c l o s u r e s y s t e m a s t h a t i n which t h e drug product
`i s marketed; ( 5 ) T e s t i n g o f d r u g p r o d u c t s f o r r e c o n s t i t u t i o n
`a t the t i m e o f d i s p e n s i n g ( a s d i r e c t e d i n the l a b e l i n g ) a s
`w e l l a s a f t e r t h e y a r e r e c o n s t i t u t e d .
`(b) An adequate
`number o f b a t c h e s o f e a c h drug product s h a l l be tested t o
`d e t e r m i n e an a p p r o p r i a t e e x p i r a t i o n d a t e and a record o f
`s u c h d a t a s h a l l be m a i n t a i n e d . A c c e l e r a t e d s t u d i e s ,
`combined w i t h b a s i c s t a b i l i t y i n f o r m a t i o n on the components,
`d r u g p r o d u c t s , and c o n t a i n e r - c l o s u r e s y s t e m , may be used t o
`s u p p o r t t e n t a t i v e e x p i r a t i o n d a t e s provided f u l l shelf l i f e
`s t u d i e s a r e n o t a v a i l a b l e and a r e b e i n g c o n d u c t e d . Where
`d a t a from a c c e l e r a t e d s t u d i e s a r e used t o p r o j e c t a
`t e n t a t i v e e x p i r a t i o n d a t e t h a t i s beyond a d a t e s u p p o r t e d b y
`a c t u a l shelf l i f e s t u d i e s , there m u s t be s t a b i l i t y s t u d i e s
`c o n d u c t e d , i n c l u d i n g d r u g product t e s t i n g a t a p p r o p r i a t e
`i n t e r v a l s , u n t i l the t e n t a t i v e e x p i r a t i o n d a t e i s v e r i f i e d
`or the a p p r o p r i a t e e x p i r a t i o n d a t e d e t e r m i n e d .
`
`5. The paragraph below requires an expiration date based upon
`the recommended storage conditions for the reagent unless
`there is a visible alteration of the product or a simple
`test method by which the user can determine if the reagent
`meets appropriate specifications.
`
`An e x p i r a t i o n d a t e based upon t h e
`21 CFR 809.10(a)(6)(i):
`s t a t e d s t o r a g e c o n d i t i o n s .
`Premarket Approval (PMA) - 21 CFR Part 814
`The PMA requirements concerning shelf life do not stipulate which
`devices require a shelf life nor do they stipulate the parameters
`that must be considered in establishing a shelf life. It is the
`responsibility of the sponsor of the PMA to make the initial
`determination of whether their device requires a shelf life and
`the device characteristics that will be considered.
`
`1. The paragraph below outlines the information that is to be
`contained in a PMA and requires a section of the PMA to
`include the results of nonclinical laboratory studies
`including studies of shelf life.
`
`A section c o n t a i n i n g r e s u l t s o f t h e
`21 CFR 814.20(b)(6)(i):
`n o n c l i n i c a l l a b o r a t o r y s t u d i e s w i t h the device i n c l u d i n g
`m i c r o b i o l o g i c a l , t o x i c o l o g i c a l , immunological,
`b i o c o m p a t i b i l i t y , stress, wear, s h e l f l i f e , and o t h e r
`l a b o r a t o r y or animal t e s t s a s a p p r o p r i a t e .
`
`2. Paragraph (a) below covers the type of changes to a device
`that require a supplemental PMA application, and expiration
`dating is specifically covered in section (8). It is
`advantageous to have a protocol for extending shelf life
`approved by FDA in the original PMA, since the change in
`product expiration date can be made as soon as the
`evaluation is completed instead of having to submit an
`additional PMA supplement and awaiting FDA approval.
`
`

`

`21 CFR 814.39(a) (8): Extension o f the e x p i r a t i o n d a t e o f
`the d e v i c e based on d a t a o b t a i n e d under a new or r e v i s e d
`s t a b i l i t y or s t e r i l i t y t e s t i n g protocol t h a t h a s not been
`approved by FDA.
`I f the p r o t o c o l h a s been approved, the
`change s h a l l be r e p o r t e d t o FDA under paragraph ( b ) o f t h i s
`section.
`
`3. The Center for Devices and Radiological Health (CDRH) has
`determined that if a protocol for testing the shelf life of
`a device has been approved by FDA in the original PMA or a
`PMA supplement, revising the shelf life of the device based
`on data collected according to the protocol is a change that
`does not affect the device's safety or effectiveness.
`Therefore, the change can be made and FDA notified in the
`next periodic report.
`
`21 CFR 814.39(b): An a p p l i c a n t may make a change i n a
`d e v i c e a f t e r F D A ' s approval o f a PMA for a device w i t h o u t
`s u b m i t t i n g a PMA supplement i f the change d o e s n o t a f f e c t
`the device's s a f e t y or e f f e c t i v e n e s s and the change i s
`r e p o r t e d t o FDA i n postapproval p e r i o d i c r e p o r t s r e q u i r e d a s
`a c o n d i t i o n t o the approval o f the d e v i c e , e .g., a n
`e d i t o r i a l change i n the l a b e l i n g which d o e s not a f f e c t t h e
`s a f e t y or e f f e c t i v e n e s s o f the device.
`Good Manufacturing Practices (GMP) - 21 CFR Part 820
`The GMP requirements outline elements of a quality assurance
`system that point to the desired goals, not as an intricate set
`of instructions. This approach is beneficial to manufacturers
`that may have already implemented quality control procedures
`which include shelf life; or who are contemplating a new device
`that has unique aspects which would not fit into a l'cookbookv
`approach. Some GMP elements that impact on shelf life are
`reprinted below.
`
`1. The paragraph below requires the establishment of a quality
`assurance program that would include an expiration date if
`appropriate.
`
`21 CFR 820.5: Every f i n i s h e d device m a n u f a c t u r e r s h a l l
`prepare and implement a q u a l i t y assurance program t h a t i s
`a p p r o p r i a t e t o the s p e c i f i c device manufactured and meets
`the r e q u i r e m e n t s o f t h i s p a r t .
`
`2.
`
`This section requires an environmental control program when
`it is necessary to prevent contamination of the device and
`to provide proper conditions for the operations performed to
`manufacture, store and distribute the device.
`
`21 CFR 820.46: Where environmental c o n d i t i o n s a t the
`m a n u f a c t u r i n g s i t e c o u l d have an a d v e r s e e f f e c t on a
`device s f i t n e s s f o r u s e , these environmental c o n d i t i o n s
`s h a l l be controlled t o p r e v e n t c o n t a m i n a t i o n o f t h e device
`and t o p r o v i d e proper c o n d i t i o n s for each o f the o p e r a t i o n s
`
`

`

`8 2 0 . 4 0 . Conditions t o be considered
`performed pursuant t o
`f o r control are l i g h t i n g , v e n t i l a t i o n , temperature,
`humidity, a i r pressure, f i l t r a t i o n , airborne contamination,
`and other contamination. Any environmental control system
`shall be periodically inspected t o v e r i f y t h a t t h e system i s
`properly functioning. Such inspections shall be documented.
`
`3.
`
`Both the manufacturing specifications and the manufacturing
`processes must be written and include a formal method to
`control changes to these procedures. This GMP requirement
`includes an evaluation of any change to the shelf life
`characteristics of the device resulting from a specification
`or process change.
`
`21 CFR 820.100: Written manufacturing s p e c i f i c a t i o n s and
`processing procedures shall be established, implemented, and
`controlled t o assure t h a t t h e device conforms t o i t s
`original design o r any approved changes i n t h a t design.
`(also, see subparts F and J)
`4. The paragraph below requires written procedures for
`warehouse control and distribution. This could include
`controlled storage; e.g., refrigeration, humidity control,
`etc., and first-in, first-out shipping, as needed to protect
`the device.
`
`21 CFR 820.150: There shall be written procedures for
`warehouse control and d i s t r i b u t i o n o f finished devices t o
`assure t h a t only those devices approved f o r release are
`d i s t r i b u t e d . Where a d e v i c e ' s f i t n e s s for use or q u a l i t y
`deteriorates over time, there shall be a system t o assure
`t h a t the oldest approved devices are distributed f i r s t .
`
`FDA POLICIES
`
`CDRH has issued various policy and internal guidance documents to
`assist CDRH employees and the regulated industry in understanding
`what information will be required to satisfy the regulatory
`requirements for commercial distribution of a medical device.
`The policies that are in effect for the applications submitted to
`satisfy these premarketing clearance and distribution
`requirements are discussed below.
`
`Premarket Notification (510(k))
`
`There currently is no written policy concerning the determination
`of shelf life for a device that is the subject of a 510(k)
`submission. However, CDRH has issued an internal guidance
`document, 510(k) Sterility Review Guidance #K90-1, to set forth
`the procedure to be used by the Office of Device Evaluation (ODE)
`and the Office of Compliance and Surveillance (OCS) in reviewing
`510(k)s for sterile devices. The 510(k) Sterility Review
`Guidance document lists the information that is to be included in
`a 510(k) for a sterile device regardless of how the device is
`labeled for shelf life. The guidance also acknowledges that
`certain sterilization processes have a deleterious effect upon a
`
`

`

`nonsterility aspect of the device, e.g., the shelf life of the
`sterilized device could vary due to the use of different
`sterilization procedures. The 510(k) Sterility Review Guidance
`document can be used by medical device manufacturers when
`preparing a 510(k) application for a sterile medical device to
`make sure the application contains all the information required
`by CDRH. An excerpt of #K90-1 follows:
`
`The following information concerning the s p e c i f i c a t i o n s related
`t o s t e r i l i t y should be collected and reviewed by ODE during t h e
`review o f the 510 ( k ) for a s t e r i l e device:
`-
`t h e s t e r i l i z a t i o n method t h a t w i l l be used;
`-
`a description o f t h e method t h a t w i l l be used t o validate
`t h e s t e r i l i z a t i o n cycle, but not t h e validation data i t s e l f ;
`-
`t h e s t e r i l i t y assurance l e v e l (SAL) f o r the device which t h e
`f i r m intends t o meet;
`-
`a description o f t h e packaging t o maintain t h e device's
`s t e r i l i t y ( t h i s i s not t o include packaging i n t e g r i t y
`t e s t i n g data) ;
`i f s t e r i l i z a t i o n involved ETO, the maximum l e v e l s o f
`residues o f ethylene oxide, ethylene chlorohydrin, and
`ethylene glycol which remain on t h e device;
`whether the product i s "pyrogen free1' and a description o f
`t h e method used t o make t h e determination;
`t h e radiation dose, i f radiation s t e r i l i z a t i o n w i l l be used.
`
`-
`-
`Only t h i s information w i l l be collected regardless o f how t h e
`device i s labeled, i . e . , whether it i s labeled s t e r i l e , s t e r i l e
`u n t i l opened or damaged, or s t e r i l e u n t i l a stated expiration
`date.
`
`-
`
`We recognize t h a t certain s t e r i l i z a t i o n processes have
`deleterious e f f e c t upon a n o n s t e r i l i t y aspect o f t h e device.
`These e f f e c t s have t o be factored i n t o t h e equivalency decision.
`
`The pilot program that is outlined in #K90-1 with the Division of
`Cardiovascular Devices (DCD), 510(k) Staff and OCS participating
`is still in effect. As of March 1991, there have been no
`decisions made regarding expanding the program to the other
`divisions, modifying or removing this portion of the guidance or
`evaluating the effectiveness of the program.
`
`Investigational Device Exemption (IDE)
`
`There currently is no written policy concerning the determination
`of shelf life for a device that is the subject of an IDE. The
`current CDRH policy for shelf life determination for devices
`undergoing clinical investigation allows each of the operating
`divisions within ODE to determine:
`
`which devices require an established shelf life prior to the
`clinical study; and
`
`the methods to be used for validating a shelf life of a
`particular device.
`
`

`

`Premarket Approval (PMA)
`
`The current policy concerning the determination of shelf life for
`a device that is the subject of a PMA is entitled I8Office of
`Device valuation Shelf-Life Policy for PMAs8' issued
`March 7 , 1984. This policy for shelf life determination for
`devices marketed through the PMA process allows each of the
`operating divisions within ODE to determine:
`
`which devices require an established shelf life; and
`
`the adequacy of the proposed methods for validating a shelf
`life for a particular device.
`
`The policy also requires that the labeling for the device include
`the shelf life and discusses the process for extending the shelf
`life by submitting a PMA supplement or following an approved
`protocol as previously discussed. The text of the policy
`follows:
`
`The ODE Divisions have the r e s p o n s i b i l i t y t o determine
`whether t h e sponsor must submit data i n support o f s h e l f
`l i f e dating o f a PMA device. The Divisions have the
`r e s p o n s i b i l i t y for maintaining a l i s t o f generic classes o f
`devices which require s h e l f l i f e dating and supplying t h e
`l i s t t o t h e PMA s t a f f .
`
`The Divisions have t h e r e s p o n s i b i l i t y f o r evaluating t h e
`data t o determine whether they are appropriate and adequate
`and t o reach a conclusion regarding t h e s h e l f l i f e o f t h e
`device.
`
`The approval order (approval l e t t e r ) shall include the s h e l f
`l i f e o f t h e device. This i s an

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket