throbber
J Mater Sci (2015) 50:493–518
`DOI 10.1007/s10853-014-8685-2
`
`R E V I E W
`
`Performing elemental microanalysis with high accuracy and high
`precision by scanning electron microscopy/silicon drift detector
`energy-dispersive X-ray spectrometry (SEM/SDD-EDS)
`
`Dale E. Newbury • Nicholas W. M. Ritchie
`
`Received: 14 July 2014 / Accepted: 25 October 2014 / Published online: 12 November 2014
`Ó Springer Science+Business Media New York (outside the USA) 2014
`
`Abstract Electron-excited X-ray microanalysis per-
`formed in the scanning electron microscope with energy-
`dispersive X-ray spectrometry (EDS) is a core technique
`for characterization of the microstructure of materials. The
`recent advances in EDS performance with the silicon drift
`detector (SDD) enable accuracy and precision equivalent to
`that of the high spectral resolution wavelength-dispersive
`spectrometer employed on the electron probe microana-
`lyzer platform. SDD-EDS throughput, resolution, and sta-
`bility
`provide
`practical
`operating
`conditions
`for
`measurement of high-count spectra that form the basis for
`peak fitting procedures that recover the characteristic peak
`intensities even for elemental combination where severe
`peak overlaps occur, such PbS, MoS2, BaTiO3, SrWO4,
`and WSi2. Accurate analyses are also demonstrated for
`interferences involving large concentration ratios: a major
`constituent on a minor constituent (Ba at 0.4299 mass
`fraction on Ti at 0.0180) and a major constituent on a trace
`constituent (Ba at 0.2194 on Ce at 0.00407; Si at 0.1145 on
`Ta at 0.0041). Accurate analyses of low atomic number
`elements, C, N, O, and F, are demonstrated. Measurement
`of trace constituents with limits of detection below 0.001
`mass fraction (1000 ppm) is possible within a practical
`measurement time of 500 s.
`
`Contribution of the United States Government; not subject to U.S.
`copyright.
`D. E. Newbury (&) N. W. M. Ritchie
`Materials Measurement Science Division, National Institute of
`Standards and Technology, Gaithersburg, MD 20899, USA
`e-mail: dale.newbury@nist.gov
`
`Introduction
`
`Origins: electron probe microanalysis with wavelength-
`dispersive spectrometry
`
`Electron-excited X-ray spectrometry for the measurement
`Query of elemental composition on the microstructural
`scale has been an important part of the materials charac-
`terization arsenal since the invention of the electron probe
`microanalyzer (EPMA) in 1951 by Castaing [1, 2]. Cas-
`taing not only produced the first working EPMA instrument
`but he also established the framework for the fundamental
`measurement science of
`the technique,
`including the
`physical basis for a practical quantitative analysis method.
`For the first two decades of the EPMA technique, the dif-
`fraction-based wavelength-dispersive X-ray spectrometer
`(WDS) was the only practical way to measure the X-ray
`intensities. Castaing recognized that the complex depen-
`dence of the WDS efficiency on photon energy made it
`impractical
`to develop a quantification procedure that
`compared different elements measured at different photon
`energies. Besides the variable solid angle of the WDS that
`is dependent on photon energy, four or more diffractors
`with different d-spacings and scattering efficiencies are
`needed to satisfy Bragg’s equation over the photon energy
`range of interest from 100 eV to 10 keV. To overcome this
`measurement dilemma, Castaing developed the ‘‘k-ratio’’
`protocol based on measuring the characteristic X-ray
`intensity, I, for the same element in the unknown and in a
`standard of known composition:
`k ¼ Iunknown=Istandard:
`The characteristic X-ray peak intensity is corrected for
`background and measured under identical conditions of
`beam energy, known dose, and detector efficiency for both
`
`ð1Þ
`
`123
`
`Merck Exhibit 2233, Page 1
`Mylan Pharmaceuticals Inc. v. Merck Sharp & Dohme Corp.
`IPR2020-00040
`
`

`

`494
`
`J Mater Sci (2015) 50:493–518
`
`Fig. 1 Distribution of
`[(measured - true)/true 9
`relative errors
`100 %] using the k-ratio protocol with WDS measurements and
`matrix corrections with the NBS ZAF procedure FRAME (1975) [5].
`Note that the histogram bins have a width of 1 % relative
`
`is
`region that
`the principal
`near-surface composition,
`sampled by electron-excited X-rays, rendering the analyt-
`ical
`results unrepresentative of
`the material being
`measured.
`Throughout the development of quantitative electron-
`excited X-ray microanalysis, researchers rigorously tested
`the method by measuring as unknowns carefully selected
`multi-element materials whose microscopic homogeneity
`could be first confirmed by EPMA and whose overall
`composition was measured by independent chemical ana-
`lysis. The distribution of measured relative errors, defined
`as
`
`Relative error ¼ Measured concentrationReferenceð
`eferenceŠ  100 %;
`
`Þ=
`ð3Þ
`
`½ R
`
`as determined by a mature version of the k-ratio/matrix
`correction procedure in 1975 is shown in Fig. 1 for WDS
`measurements of major1 constituents [5]. This distribution
`can be characterized by a standard deviation of 2.5 %
`relative, so that approximately 95 % of the analyses fall
`within the span of ±5 % relative error.
`
`1 Note: in this paper, the following arbitrary convention for broadly
`classifying the concentration range will be followed:
`
`‘‘major,’’ mass concentration C [ 0.1 (more than 10 wt%)
`‘‘minor’’ 0.01 B C B 0.1 (1–10 wt%)
`‘‘trace’’ C \ 0.01 (\1 wt%).
`
`ð2Þ
`
`unknown and standard. By measuring the same peak under
`identical conditions, the same efficiency value effectively
`appears in both the numerator and denominator of Eq. (1)
`as a multiplier of the intensity, and thus the efficiency
`quantitatively cancels in the k-ratio.
`Castaing further described the basis for the physical
`calculations that are necessary to convert the suite of k-
`ratios into mass concentrations, which after substantial
`further contributions by numerous authors (see Ref. [3] for
`Heinrich’s detailed account of these developments) take
`the following form:
`Ci=Cstd ¼ ki ZAFc;
`where Cstd is the mass concentration of the element of
`interest in the standard; and Z, A, F, and c are the ‘‘matrix
`correction factors’’
`that calculate the compositionally
`dependent interelement effects of electron scattering and
`energy loss (Z), X-ray self-absorption within the specimen
`(A), and secondary X-ray emission following self-absorp-
`tion of the electron-excited characteristic (F) and contin-
`uum (c) X-rays. Importantly for the Castaing standards-
`based k-ratio method, the standards required do not have to
`closely match the composition of the unknown, which is an
`enormous advantage when dealing with complicated multi-
`the k-ratio
`element unknowns. Suitable standards for
`measurements include pure elements (e.g., Al, Si, Cr, Fe,
`Ni, etc.), while stoichiometric compounds can be used for
`those elements that are not in solid form in a vacuum (e.g.,
`MgO for O), that are highly reactive (e.g., KCl for K and
`Cl),
`that deteriorate under electron bombardment (e.g.,
`FeS2 for S), or that have a low melting temperature (e.g.,
`GaP for Ga and P).
`The extremely sharp focal properties of the WDS forced
`EPMA analysts to develop procedures to establish and
`maintain the critical condition of identical detection effi-
`ciency when measuring the separate intensities for the
`unknown and standards required for Eq. (1) [3]. To place
`the specimen reproducibly within the narrow spatial range,
`spanning a few micrometers, over which the WDS had
`constant X-ray transmission, a fixed-position optical
`microscope with a shallow depth of focus was incorporated
`into the EPMA at the coincident focal position for all
`spectrometers. The condition of the specimen surface was
`recognized to be another critical requirement [4]. It came to
`be understood early in the development of EPMA that the
`specimen had to metallographically polished to a very high
`degree of surface finish, but not chemically etched. To
`create contrast in optical metallography, chemical etching
`typically produces topography through orientation-depen-
`dent etch rates in different grains and phases, but even fine-
`scale topography can influence measured X-ray intensities,
`especially for low-energy photons. Moreover,
`in some
`cases, chemical etching induces changes in the surface/
`
`123
`
`Merck Exhibit 2233, Page 2
`Mylan Pharmaceuticals Inc. v. Merck Sharp & Dohme Corp.
`IPR2020-00040
`
`

`

`J Mater Sci (2015) 50:493–518
`
`495
`
`Further development: energy-dispersive X-ray
`spectrometry
`
`The development of semiconductor-based X-ray detection
`in the 1960s led to the first successful energy-dispersive
`X-ray spectrometer (EDS) using lithium-compensated sil-
`icon [Si(Li)-EDS] operated on an electron-column instru-
`ment, an EPMA [6]. As compared to the narrow
`instantaneous energy range of the WDS, the Si(Li)-EDS
`provided a view of the entire excited X-ray spectrum from
`a threshold of approximately 100 eV (modern perfor-
`mance) to the Duane-Hunt limit set by the incident beam
`energy, up to 20 keV or higher. This wide energy range
`enabled detection of all elements, with the exception of H
`and He (modern performance), at every location sampled
`by the beam, which provided an enormous advantage when
`dealing with complex microstructures where local segre-
`gation can create unexpected compositional variation and
`where unexpected elements can be localized as inclusions.
`Comprehensive elemental analysis capability, combined
`with the relative simplicity of non-focusing line-of-sight
`detection, the large solid angle of collection which exceeds
`that of WDS by at least a factor of 10, and the long-term
`operating stability, resulted in the enthusiastic acceptance
`of EDS, especially by the rapidly developing scanning
`electron microscope (SEM) community. The combination
`of SEM imaging with EDS X-ray microanalysis has given
`the materials community one of its most powerful micro-
`structural characterization tools [2].
`The capability of Si(Li)-EDS to perform quantitative
`X-ray microanalysis was established soon after its intro-
`duction by several members of the microanalysis commu-
`nity, most of whom had extensive WDS quantitative
`microanalysis experience [7–11]. Thus, the initial EDS
`implementation of quantitative analysis was based upon
`their experience with the WDS k-ratio protocol. The EDS
`could be used in an equivalent manner by measuring the
`intensities for the unknown and appropriate standards
`under the same carefully controlled conditions of surface
`condition (highly polished), beam energy, known dose
`(beam current 9 detector live time), beam incidence angle,
`detector elevation angle (‘‘take-off angle’’), and detector
`efficiency (e.g., constant detector-to-target distance to yield
`constant detector solid angle). The enormous advantages of
`the EDS over WDS for analysis were quickly recognized:
`(1) all elements in the unknown were measured simulta-
`neously minimizing the dose to the specimen; (2) the large
`solid angle of the EDS relative to WDS further improved
`efficiency of detection which lowered the necessary dose
`relative to WDS; and (3) the stability of the EDS meant
`that the spectra of standards could be archived and recalled
`as needed. Since the measured EDS spectrum consists of
`the characteristic X-ray peaks superimposed on the X-ray
`
`continuum, various strategies were developed to determine
`characteristic intensities,
`including digital filtering for
`background removal followed by multiple linear least
`squares (MLLS) fitting and background modeling under the
`peak window constrained by the continuum intensity
`measured in energy windows where no peaks occurred [8,
`10]. The background-corrected characteristic intensities
`for the unknown and the standards were used to calculate
`k-ratios followed by the matrix correction procedure. The
`k-ratio matrix correction procedure with the Si(Li)-EDS
`was demonstrated to be capable of achieving relative errors
`within the WDS distribution for major constituents when
`the characteristic X-ray peaks did not suffer significant
`overlap from the peaks of other elements. An example is
`shown in Table 1 for the Si(Li)-EDS analysis of gold-
`copper alloys (NIST Standard Reference Material 482)
`where the observed relative errors range from -1.6 to
`1.0 %, falling well within the WDS analysis error distri-
`bution of Fig. 1.2 Further development of the EDS quan-
`titative microanalysis method enabled accurate analyses
`when significant peak overlaps occurred providing the
`intensities of the mutually interfering species were similar.
`
`Typical SEM/EDS microanalysis practice
`
`Despite the level of analytical accuracy demonstrated for
`the EDS k-ratio/matrix corrections protocol and the avail-
`ability of this procedure within most commercial imple-
`mentations of EDS analytical software, modern SEM/EDS
`microanalysis practice has developed along a different
`trajectory that minimizes the need for the user’s expertise.
`As an unintended consequence, EDS microanalysis as
`performed in the SEM has acquired an unfortunate repu-
`tation as a ‘‘semi-quantitative’’ technique. This situation
`has developed because of three contributing factors: (1) the
`rise of standardless analysis which now dominates EDS
`quantitative analysis [14]; (2) the severe effects of speci-
`men geometry on the accuracy of X-ray microanalysis
`which occur no matter which analytical protocol is fol-
`lowed, standards-based or standardless [15]; and (3) the
`occasional but significant failures in qualitative analysis,
`i.e., incorrect elemental identification, which immediately
`undermines confidence in the method [16–18].
`
`The rise of ‘‘standardless’’ quantitative analysis
`
`By necessity, WDS measures each element in the sample
`relative to the same element in an appropriate standard to
`
`2 Materials analyzed in this paper include NIST Standard Reference
`Materials, NIST microanalysis research materials (glasses), natural
`minerals, and stoichiometric compounds confirmed to be homoge-
`neous on a micrometer lateral scale.
`
`123
`
`Merck Exhibit 2233, Page 3
`Mylan Pharmaceuticals Inc. v. Merck Sharp & Dohme Corp.
`IPR2020-00040
`
`

`

`496
`
`J Mater Sci (2015) 50:493–518
`
`Table 1 Si(Li)-EDS analysis of NIST Standard Reference Material 482 Copper–Gold Alloysa (all concentration values in mass fraction)
`
`Cu (certified)
`
`Analyzed
`
`Rel. error (%)
`
`Au (certified)
`
`Analyzed
`
`Rel. error (%)
`
`0.198
`
`0.396
`
`0.599
`
`0.798
`
`0.198 ± 0.002
`
`0.399 ± 0.001
`
`0.605 ± 0.001
`
`0.0
`
`0.8
`
`1.0
`
`0.797 ± 0.001
`
`-0.1
`
`0.801
`
`0.603
`
`0.401
`
`0.200
`
`0.790 ± 0.002
`
`0.594 ± 0.002
`
`0.402 ± 0.002
`
`0.199 ± 0.003
`
`-1.4
`
`-1.6
`
`0.1
`
`-1.2
`
`Total
`
`0.988
`
`0.993
`
`1.007
`
`0.996
`
`Analysis performed with Cu Ka and AuLa; beam energy = 20 keV; quantitative calculations with NIST Desktop Spectrum Analyzer [12];
`uncertainty expressed as 1r from the measured counts (from Ref [13])
`a See footnote 2
`
`eliminate the need to accurately know the spectrometer
`efficiency. Because the EDS spectrum provides, in every
`measurement, the complete photon energy range revealing
`all characteristic X-ray peaks and the X-ray continuum
`background, it became attractive to develop an alternative
`approach
`for
`quantitative EDS microanalysis
`that
`employed the whole spectrum. The so-called ‘‘standardless
`analysis’’ method requires only the EDS spectrum of the
`unknown and eliminates the need to measure standards
`locally or to specify the electron dose [2, 14]. ‘‘Standard-
`less analysis’’ seeks to provide the necessary standard
`intensity for the denominator of Eq. (1) for each element in
`the unknown either by theoretical calculation of X-ray
`generation and propagation in a pure element target (‘‘first
`principles’’ standardless) or by the use of a library of actual
`standards measured on a well-characterized EDS detector
`at several beam energies under defined conditions that can
`be related to the efficiency as a function of photon energy
`of the local EDS (‘‘remote standards’’ standardless). The
`resulting suite of k-ratios is then subjected to the same
`matrix correction calculations of Eq. (2). Because a true
`first principles implementation of standardless analysis
`requires an extensive database of X-ray parameters such as
`the ionization cross section, X-ray fluorescence yield,
`X-ray mass absorption coefficient, and others, many of
`which are poorly known, especially for the L-shell and
`M-shell X-ray families, the ‘‘remote standards’’ method,
`which actually anchors the quantitative calculations to a
`suite of archived experimental measurements, is the basis
`for the typical modern software implementation.
`The performance of a recent commercial version of
`‘‘standardless analysis’’ is shown in the error histogram of
`Fig. 2. While this error distribution appears similar to that
`of the classic k-ratio protocol with WDS or EDS as shown
`in Fig. 1, it is in fact much broader, with the errors binned
`in increments of 5 % relative error, compared to the 1 %
`relative error increments of Fig. 1. For this particular
`implementation of standardless analysis, the width of the
`error range that
`is necessary to capture 95 % of the
`analytical results is approximately ±30 % relative. While
`this level of performance may be adequate for some
`
`123
`
`Fig. 2 Distribution of relative errors observed for a commercial
`implementation of standardless analysis (2013). Note that
`the
`histogram bins have a width of 5 % relative
`
`applications, the prospective user of the analytical results
`of such a procedure needs to be aware of the inherent
`limitations imposed by such a wide error distribution.
`Table 2 provides specific examples of the application of
`this standardless analysis procedure to the analysis of metal
`sulfides. While the analytical accuracy achieved for FeS
`(troilite, a meteoritic mineral) is excellent with relative
`errors less than ±2 % for S and Fe, the relative errors for
`the analysis of FeS2 (pyrite), CuS (covellite), ZnS (sphal-
`erite), and PbS (galena) exceed 20 % relative, a level of
`performance so poor that
`it would not be possible to
`properly assign the formula for these compounds from the
`analyzed mass concentrations.
`Another often overlooked consequence of using the
`standardless analysis procedure is the requirement that the
`calculated concentrations must be internally normalized to a
`sum of unity. This requirement occurs because the relation of
`the electron dose and the absolute EDS efficiency of the
`measured spectrum to the remote standards database is lost
`so that internal normalization is needed to place the calcu-
`lated concentration values on a meaningful scale. That the
`
`Merck Exhibit 2233, Page 4
`Mylan Pharmaceuticals Inc. v. Merck Sharp & Dohme Corp.
`IPR2020-00040
`
`

`

`J Mater Sci (2015) 50:493–518
`
`Table 2 Standardless analysis
`of sulfides (2013 Commercial
`Software)
`
`497
`
`Compound
`
`Metal
`
`Analysis
`
`Relative error (%)
`
`Sulfur
`
`Analysis
`
`Relative error (%)
`
`FeS
`
`FeS
`
`CuS
`
`ZnS
`
`SrS
`
`CdS
`
`Sb2S3
`PbS
`
`0.635
`
`0.466
`
`0.665
`
`0.671
`
`0.732
`
`0.778
`
`0.717
`
`0.866
`
`0.629
`
`0.642
`
`0.764
`
`0.762
`
`0.758
`
`0.808
`
`0.739
`
`0.914
`
`-1.0
`
`38
`
`15
`
`14
`
`3.6
`
`3.8
`
`3.1
`
`5.5
`
`0.365
`
`0.534
`
`0.335
`
`0.329
`
`0.268
`
`0.222
`
`0.283
`
`0.134
`
`0.371
`
`0.358
`
`0.236
`
`0.239
`
`0.242
`
`0.192
`
`0.261
`
`0.086
`
`1.8
`
`-33
`
`-30
`
`-28
`
`-10
`
`-13
`
`-7.8
`
`-36
`
`analyzed mass concentration total of all constituents in a
`standardless analysis equals exactly 1.000 (100 wt%) may
`seem comforting, but the internal normalization that must
`occur does in fact represent a loss of critical information. In
`the standards-based k-ratio/matrix corrections protocol per-
`formed with WDS or EDS, the sum of the individual con-
`stituents rarely coincides exactly with unity, but tends to vary
`from 0.98 to 1.02, as shown in the example presented in
`Table 1, a consequence of the inevitable errors that arise in
`measuring the characteristic intensities and in calculating the
`matrix correction factors. Analytical totals outside of this
`range can occur because of uncontrolled deviations in the
`experimental conditions between measuring the unknown
`and standards (e.g., differences in coating thickness or in the
`thickness of native surface oxides), but a low analytical total
`may also reveal the presence in the analyzed volume of a
`previously unrecognized constituent. For example, a region
`of the specimen that is oxidized rather than metallic will
`contain oxygen at a concentration from 0.2 to 0.3 mass
`fraction. The analytical total if oxygen is not considered
`(either by directly measuring its X-ray intensity and making
`the appropriate matrix correction calculation or by indirectly
`calculating oxygen by the method of assumed stoichiometry
`of the cations) will be 0.7–0.8, significantly below unity,
`which should trigger the curiosity of a careful analyst to
`further examine the measured spectrum and discover the
`oxygen peak. While this may seem a trivial example that
`even a novice analyst should not miss, in fact as we enter an
`era in which much of our data are collected under automa-
`tion, the lack of manual inspection combined with the loss of
`a meaningful analytical total by the standardless method will
`result in questionable data appearing in the final results that
`may be difficult to review after collection and processing. As
`discussed below, a more frequently encountered source of
`deviation in the analytical total is the impact of uncontrolled
`‘‘specimen geometry,’’ i.e., the effects of size, shape, and
`local surface inclination on beam electron—specimen
`interaction and the generation and propagation of X-rays, on
`the measured X-ray intensities. A ‘‘zeroth’’ level assumption
`in standards-based and standardless analysis procedures is
`
`that the specimen composition is the only factor affecting the
`X-ray intensities. When the specimen geometry deviates
`from the ideal flat surface placed at known angles to the
`incident electron beam and the X-ray spectrometer, very
`large effects on the X-ray intensities can occur, especially
`when low-energy and high-energy photons are measured in
`the same analysis.
`Despite these limitations, the simplicity of operation
`required for standardless analysis, which only requires the
`analyst to measure the EDS spectrum of the unknown and
`to specify the beam energy and the X-ray emergence angle,
`has resulted in its widespread acceptance by the SEM/EDS
`community. Based on our informal surveys of the field,
`probably more than 98 % of reported quantitative EDS
`microanalysis results are obtained with some implementa-
`tion of standardless analysis. However, the modest ana-
`lytical performance revealed in Fig. 2 and Table 2 is surely
`a major contributor to the reputation of SEM/EDS as only
`achieving ‘‘semiquantitative’’ results, while the internal
`normalization of all results to unity conveys a false sense of
`accuracy and confidence.
`
`Specimen geometry effects: we can be our own worst
`enemies when it comes to performing accurate
`quantification
`
`When analytical results are automatically normalized, an
`even more egregious source of large, uncontrolled, and
`likely unrecognized errors in SEM/EDS microanalysis
`arises from specimen geometry effects [2]. The line-of-sight
`acceptance of the EDS spectrometer enables the analyst to
`record an X-ray spectrum from almost any location where
`the beam strikes the specimen, which can be a useful feature
`in qualitatively surveying the complex microstructure of a
`specimen with complex topography. However, specimen
`geometry effects such as shape and local surface inclination
`to the beam can have a profound impact on electron scat-
`tering and even more importantly, on the path length along
`which X-rays must travel to the detector and along which
`they suffer absorption. These ‘‘geometric effects’’ modify
`
`123
`
`Merck Exhibit 2233, Page 5
`Mylan Pharmaceuticals Inc. v. Merck Sharp & Dohme Corp.
`IPR2020-00040
`
`

`

`498
`
`J Mater Sci (2015) 50:493–518
`
`Analysis with a compromised sample shape: all forms
`Analysis of K411: Bulk polished, 600 Grit, In-hole, Chips, Shards
`80
`
`BULK
`
`8.8
`8.6
`Mg (normalized weight percent)
`
`11.5
`
`11.4
`
`11.3
`
`Fe (normalized weight percent)
`
`Bulk K411
`
`Shards_overscan
`
`Shards_fixed-beam
`
`Macroscopic chips
`
`Surface voids
`
`Bulk-600grit
`
`0
`
`1
`
`2
`
`9
`8
`7
`6
`5
`4
`3
`Mg (normalized weight percent)
`
`10
`
`11
`
`12
`
`70
`
`60
`
`50
`
`40
`
`30
`
`20
`
`10
`
`0
`
`Fe (normalized weight percent)
`
`Fig. 4 Analysis of NIST SRM 470 (K411 glass) in various geometric
`forms (flat, polished bulk; scratched surface after 600-grit grinding;
`shallow surface holes, chips, and shards) using the k-ratio protocol
`with SDD-EDS measurements and NIST DTSA-II. Plot of Fe
`(normalized weight percent) vs. Mg (normalized weight percent) [19]
`
`the standardless analysis method, the inevitable internal
`normalization to unity eliminates the important information
`provided by the analytical total that would reveal the impact
`of specimen geometry effects.
`An example of how serious the impact of specimen
`geometry can be upon analytical accuracy is illustrated in
`Figs. 3 and 4 for the analysis of a microscopically homo-
`geneous glass, NIST SRM 470 (K411),
`the complete
`composition of which is listed in Table 3. Table 3 also
`contains the results of a standards-based k-ratio protocol
`analysis of NIST SRM 470 (K411 glass) prepared in the
`ideal specimen geometry of a flat, highly polished surface
`(0.1 lm alumina final polish) with a thin (\10 nm) carbon
`conductive coating for charge dissipation. The relative
`errors as compared to the SRM certificate values for the
`average of 20 analyses range from -1.1 to 1.8 %. Fig-
`ure 3a shows a plot of the distribution from 20 analyzed
`locations for the concentrations of magnesium and iron,
`elements which were chosen because the large difference
`in their photon energies, 1.254 keV for MgKa, b and
`6.400 keV for FeKa, make them differentially sensitive to
`geometric effects since the low-energy photons of Mg
`suffer much higher absorption compared to the high-energy
`photons of FeKa [19]. The cluster of the analyses is seen to
`be very narrow, with one exception. A reasonable question
`the analyst might ask is whether this outlier represents an
`actual deviation in the local composition or arises from
`some other factor. Upon review of the analyzed locations,
`the outlier in Fig. 3 was in fact determined to be the
`consequence of an analysis that was performed in a scratch
`that remained after final polishing,
`thus representing a
`
`Fig. 3 a Analysis of NIST SRM 470 (K411 glass) as a flat, highly
`polished bulk sample (final polish with 100 nm alumina) using the
`k-ratio protocol with SDD-EDS measurements and NIST DTSA-II.
`Plot of Fe (normalized weight percent) vs. Mg (normalized weight
`percent) for 20 randomly selected analyses. Note the outlier (circled).
`b Analysis of NIST SRM 470 (K411 glass) as a flat, but slightly
`scratched bulk sample (scratches remaining after 1 lm diamond
`polish) using the k-ratio protocol with SDD-EDS measurements and
`NIST DTSA-II. Plot of Fe (normalized weight percent) vs. Mg
`(normalized weight percent) for 20 randomly selected analyses
`
`the measured relative elemental intensities in ways that
`have nothing to do with the composition of the specimen.
`The physical basis assumed for the matrix correction factors
`of Eq. (2), which are needed for quantification with both the
`k-ratio standards protocol and the standardless method, is
`that the compositional difference between the unknown and
`the standard(s) is the only factor that affects the measured
`X-ray intensities. When a quantitative analysis of a topo-
`graphically irregular specimen is performed following the
`k-ratio/standards protocol, the analytical total will deviate
`significantly from unity in response to deviations from the
`ideal flat, highly polished specimen geometry. However, in
`
`123
`
`Merck Exhibit 2233, Page 6
`Mylan Pharmaceuticals Inc. v. Merck Sharp & Dohme Corp.
`IPR2020-00040
`
`

`

`J Mater Sci (2015) 50:493–518
`
`499
`
`(a)
`
`12
`
`Analysis of K411: Bulk polished and All Geometries
`
`SRMvalue
`
`024681
`
`0
`
`Bulk K411
`
`All Geometries
`
`1
`
`0
`
`20
`
`40
`60
`80
`100
`Raw Analytical Total (weight percent)
`
`120
`
`140
`
`Analysis of K411: Bulk polished and All Geometries
`
`80
`
`70
`
`60
`
`50
`
`40
`
`30
`
`20
`
`Bulk K411 All Geometries
`
`1
`
`SRMvalue
`
`01
`
`0
`
`0
`
`20
`
`40
`
`60
`
`80
`
`100
`
`120
`
`140
`
`Raw Analytical Total (weight percent)
`
`Mg (normalized weight percent)
`
`(b)
`
`Fe (normalized weight percent)
`
`Fig. 5 Analysis of NIST SRM 470 (K411 glass) in various geometric
`forms (flat, polished bulk; scratched surface after 600-grit grinding;
`shallow surface holes, chips, and shards) using the k-ratio protocol
`with SDD-EDS measurements and NIST DTSA-II: a Mg (normalized
`weight percent) vs. the raw analytical total (weight percent), including
`oxygen calculated by assumed stoichiometry [19], b Fe (normalized
`weight percent) vs. the raw analytical total (weight percent), including
`oxygen calculated by assumed stoichiometry [19]
`
`standardless analysis
`information,
`losing this critical
`enables risky analytical behavior with SEM/EDS which
`contributes enormously to the dismissal of SEM/EDS as
`being only ‘‘semi-quantitative.’’ Unless the specimen
`geometry is carefully controlled, SEM/EDS analysis is
`subject to errors so broad as to render the compositional
`results of questionable value for many applications.
`
`Qualitative analysis failures
`
`A separate but extremely significant issue is the reliability
`of elemental identification in the EDS spectrum. The crit-
`ical first step of qualitative analysis obviously must be
`correct if the subsequent quantitative analysis is to have
`any value at all. Automatic identification of the charac-
`teristic peaks in the EDS spectrum is a valuable software
`
`123
`
`Table 3 SEM/SDD-EDS analysis of NIST SRM K411 glass
`
`Element
`
`SRM
`certificate
`
`EDS analysis
`
`SDa
`
`Relative
`error (%)
`
`O
`
`Mg
`
`Si
`
`Ca
`
`Fe
`
`0.424
`
`0.0885
`
`0.254
`
`0.111
`
`0.112
`
`0.428 (stoich)
`
`0.0876
`
`0.258
`
`0.111
`
`0.114
`
`0.022
`
`0.045
`
`0.053
`
`0.026
`
`0.031
`
`0.9
`
`-1.1
`
`1.6
`
`0
`
`1.8
`
`Conditions: polished specimen; E0 = 20 keV; analysis following the
`k-ratio protocol with standards using the NIST DTSA-II software;
`standards included the pure elements Mg, Si, and Fe; Ca from SRM
`470 (glass K412), with oxygen calculated on the basis of assumed
`stoichiometry of the cations
`a 20 analyses
`
`geometric effect rather than a true compositional variation.
`This surface roughness effect
`is illustrated for a more
`severe situation in Fig. 3b, which shows the results from 20
`analyses at randomly selected locations on the scratched,
`irregular surface that remained after polishing with 1 lm
`diamond particles. The results show both a systematic shift
`in the apparent concentrations and a broadening in the
`distribution.
`the analyses reported in
`The surface analyzed for
`Fig. 3b did not appear to be especially rough when viewed
`in an SEM image. When the SRM 470 (K411 glass) is
`prepared in more extreme geometric forms, including the
`scratched surface that remains after grinding with 600 grit
`silicon carbide, fracture surfaces, and fragmented particles,
`and then analyzed following the same k-ratio/standards
`protocol used for the results in Table 3, the normalized
`concentrations for magnesium and iron are found to span a
`much broader range, as shown in Fig. 4 [19]. The range of
`apparent concentrations obtained from the irregularly
`shaped specimens is so extreme, spanning nearly an order
`of magnitude for each constituent, that the results would be
`of dubious value for most applications that depend on an
`accurate composition for proper interpretation.
`Analysis with the standards-based k-ratio/matrix cor-
`rections protocol reveals the impact of geometric effects on
`X-ray microanalysis through the behavior of the raw ana-
`lytical total [19]. Figure 5a and b individually plot the
`normalized magnesium and iron concentrations against the
`total from the k-ratio/standards analysis
`raw analytical
`procedure, showing how well the raw analytical total is
`correlated with the magnitude of the relative error. This
`example illustrates well the pitfalls of blindly attempting to
`quantitatively analyze the EDS spectrum obtained from
`randomly shaped objects. Because of the inevitable nor-
`malization that must occur in the standardless analysis
`procedure, the important clue that the raw analytical total
`provides that will identify such dubious analyses is lost. By
`
`Merck Exhibit 2233, Page 7
`Mylan Pharmaceuticals Inc. v. Merck Sharp & Dohme Corp.
`IPR2020-00040
`
`

`

`J Mater Sci (2015) 50:493–518
`
`Fig. 7 a Gold, beam energy 20 keV, showing correct identification
`of the Au L-family and M-family [16–18]. b Gold, beam energy
`10 keV, showing misidentification of the Au M-family as the Zr
`L-family and Nb L-family [16–18]
`
`robust and reliable results, the prudent analyst will inspect
`each tentative elemental assignment that is suggested by
`the automatic peak identification software and confirm or
`correct the identification by carefully considering possible
`alternatives using software tools to explore the database of
`characteristic X-ray energies.
`In some circumstances,
`achieving identification with a high deg

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket