throbber

`
`
`
`
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`_____________________________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`_____________________________
`
`
`ALPHATEC HOLDINGS, INC. and ALPHATEC SPINE, INC.
`Petitioners,
`
`v.
`
`NUVASIVE, INC.,
`Patent Owner.
`
`_____________________________
`
`Case No. IPR2019-00362
`Patent No. 8,361,156
`_____________________________
`
`
`DECLARATION OF CARL R. MCMILLIN, PH.D.
`
`
`
`
`
`
`
`
`
`NUVASIVE - EXHIBIT 2057
`Alphatec Holdings Inc. et al. v. NuVasive, Inc.
`IPR2019-00362
`
`

`

`
`
`TABLE OF CONTENTS
`
`
`QUALIFICATIONS ........................................................................................ 1 
`I. 
`SCOPE OF WORK.......................................................................................... 4 
`II. 
`III.  LEVEL OF ORDINARY SKILL .................................................................... 5 
`IV.  SUMMARY OF OPINIONS ........................................................................... 5 
`V.  A PERSON OF ORDINARY SKILL IN THE ART WOULD NOT
`HAVE ADOPTED BERRY’S VERTEBRAL BODY DIMENSIONS
`FOR IMPLANT DIMENSIONS ..................................................................... 6 
`A. 
`Brantigan’s Implants Are Design to Fit Within the Annulus
`Fibrosis .................................................................................................. 8 
`B.  Michelson Confirms that a Lumbar Implant Width Should Be
`Less Than that of the Vertebral Body ................................................. 11 
`Frey’s Implants Are Designed to Fit Within the Annulus
`Fibrosis ................................................................................................ 12 
`VI.  A PERSON OF ORDINARY SKILL IN THE ART WOULD NOT
`HAVE SEQUENTIALLY INSERTED LATERAL IMPLANTS ................ 13 
`A.  Michelson Does Not Disclose Sequential Insertion of a Lateral
`Implant or Modular Members ............................................................. 13 
`Brantigan Does Not Disclose Sequential Insertion of a Lateral
`Implant or Modular Members ............................................................. 15 
`VII.  THE PROPOSAL TO INSERT FREY’S IMPLANTS THROUGH A
`HOLLOW TUBE IS NONSENSICAL ......................................................... 20 
`VIII.  BRANTIGAN’S INTERBODY FUSION IMPLANTS DO NOT
`HAVE A LONGITUDINAL APERTURE LENGTH PARALLEL TO
`THE LONGITUDINAL LENGTH OF THE IMPLANT ............................. 21 
`IX.  CONCLUDING STATEMENTS .................................................................. 24 
`
`
`C. 
`
`B. 
`
`-i-
`
`

`

`
`
`I.
`
`I, Carl R. McMillin, declare as follows:
`
`QUALIFICATIONS
`1.
`I received a B.S. in Mechanical Engineering with a material science
`
`major from the General Motors Institute of Technology and a M.S. and a Ph.D. in
`
`Macromolecular Science from the Case Western Reserve University.
`
`2.
`
`I have extensive experience in the field of biomedical engineering.
`
`From 1974 to 1975, I served as a visiting scientist at the Artificial Internal Organs
`
`Department of the Cleveland Clinic Foundation. From 1983 to 1989, I held an
`
`Associate Professor position in the Department of Biomedical Engineering at The
`
`University of Akron where I taught MS/PhD level Biomaterials and Artificial
`
`Organs courses. I also served as the Director of the Cardiovascular Laboratory and
`
`researched accelerated rubber fatigue, blood-materials interactions, implant
`
`associated infections, hemodynamics of artificial hearts, and consulted for
`
`AcroMed Corporation on the initial development of artificial spinal discs and
`
`carbon fiber composite spine fusion plates.
`
`3.
`
`Between 1989 and 1997 I worked at AcroMed Corporation where I
`
`served in various positions including Senior Scientist and Director of R&D. I
`
`conducted and directed research, development, evaluation, and commercialization
`
`of many metal, polymer, polymer/carbon fiber composite and elastomer-based
`
`1
`
`

`

`
`
`orthopedics products. My work and research at AcroMed Corporation was
`
`primarily focused on designing orthopedic products for the spine.
`
`4.
`
`Through my work, I helped design and optimize the world’s first
`
`carbon fiber composite posterior lumbar interbody fusion (PLIF) cages and the
`
`world’s first carbon fiber composite anterior lumbar interbody fusion (ALIF)
`
`cages. I also helped to design the first functional artificial spinal discs implanted in
`
`patients in the U.S.A., as well as stackable implant systems for variable heights of
`
`artificial vertebral bodies. My efforts on these devices involved numerous
`
`considerations, including material selection and validation, cadaver spine
`
`measurements, strength vs. size optimization, etc.
`
`5.
`
`I have served on numerous committees and/or advisory boards in the
`
`biomedical industry. For example, between 1994 and 1998, I served on the
`
`Editorial Advisory Board of the Medical Plastics and Biomaterials for Canon
`
`Communications. Between 1990 and 1996, I served as an Executive Editor for the
`
`international journal, Bio-Medical Materials and Engineering. I also served on the
`
`External Advisory Board (1994-1998) at the Center for Cardiovascular
`
`Biomaterials and have been on technical review panels for the NIH. I have also
`
`served on the Commercialization Cabinet 1997-2002; the Scientific Advisory
`
`Council (1992-1994), and the Executive Advisory Committee (1992-1996), at the
`
`Edison Biotechnology Center and Omeris.
`
`-2-
`
`

`

`
`
`6.
`
`I routinely attend key seminars and conferences in the industry
`
`including the all of the annual meetings for the Society for Biomaterials and every
`
`fourth year World Congress for Biomaterials for the last 30+ years to the present,
`
`and other meetings such as the North American Spine Society meeting in 2011. I
`
`have also presented at over 75 national and international meetings as an author or
`
`co-author.
`
`7.
`
`Throughout my career, I have published over 100 journal articles,
`
`book chapters, theses, etc. I am the inventor on U.S. Patent No. 5,294,391
`
`(“Method of Making a Fiber Reinforced Composite Structure Including
`
`Randomizing the Reinforcing Fibers”) and U.S. Patent No. 5,429,863 (Fiber
`
`Reinforced Composite Structure”) which are the method and device patents on the
`
`PEEK/PEKEKK composite structure that was used in the initial Brantigan PLIF
`
`and ALIF cages I am also an inventor on other spine and orthopedic patents
`
`including U.S. Patent No. 5,824,094 (“Spinal Disc”); U.S. Patent No. 6,669,732
`
`(“Spinal Disc”); and U.S. Patent No. 6,162,252 (“Artificial Spinal Disc”).
`
`8. My professional curriculum vitae is attached as EX2058. It details my
`
`education, experience, and publications, as briefly summarized above, as well as an
`
`overview of some of my experience that is relevant to the matters set forth in this
`
`declaration.
`
`-3-
`
`

`

`
`
`II.
`
`SCOPE OF WORK
`
`9.
`
` I have been retained by Counsel on behalf of NuVasive, Inc. to
`
`provide expert opinions in this matter related to spinal fusion implant sizing and
`
`modular spinal fusion implants. My opinions are based on my skills, knowledge,
`
`training, education, and experience in spinal fusion implant design, and my
`
`examination of the materials cited in this declaration.
`
`10.
`
`I understand that petitions were filed with the United States Patent and
`
`Trademark Office for inter partes review of U.S. Patent Nos. 8,187,334 (“the ’334
`
`patent”) and 8,361,156 (“the ’156 patent”).
`
`11.
`
`I further understand that the Patent Trial and Appeal Board (“PTAB”
`
`or the “Board”) has decided to institute inter partes review of claims 6-9, 16, and
`
`18 of the ’334 patent and claims 1-3, 5, 9–10, 12-21, 23–24, and 27 of the ’156
`
`patent.
`
`12.
`
`I have been specifically asked to provide my expert opinions on the
`
`content of the US2002/0165550 (“Frey,” EX1040), U.S. Pat. No. 5,860,973
`
`(“Michelson,” EX1032), U.S. Pat. No. 5,192,327 (“Brantigan,” EX1007), and
`
`Berry, et al., “A Morphometric Study of Human Lumbar and Thoracic Vertebrae,”
`
`(“Berry,” EX1022). In connection with this analysis I have also reviewed the ’334
`
`and ’156 patents.
`
`-4-
`
`

`

`
`
`13.
`
`I am being compensated at a rate of $400 per hour for my work in this
`
`matter. I am also being reimbursed for reasonable and customary expenses
`
`associated with my work in this investigation. My compensation is not contingent
`
`on the outcome of this matter or the specifics of my testimony.
`
`III. LEVEL OF ORDINARY SKILL
`
`14.
`
`I understand that Dr. Branch asserts that a person of ordinary skill in
`
`the art could be a medical doctor with 2-3 years of experience performing
`
`interbody spinal fusion surgeries or a mechanical or biomechanical engineer with 2
`
`years of experience developing implants and access to orthopedic or
`
`neurosurgeons. This is the standard for a person of ordinary skill in the art I have
`
`applied throughout this declaration. I understand that the relevant time for
`
`consideration is on or before March 29, 2004.
`
`IV. SUMMARY OF OPINIONS
`15. A person of ordinary skill in the art would not have designed a spinal
`
`fusion implant to have a width as wide or wider than the average depth of the
`
`intervertebral space. A person of ordinary skill in the art would not have designed
`
`such a spinal fusion implant to be wider than 32 mm and certainly would not have
`
`designed a spinal fusion implant to have a width as wide as 37.9 mm. Making an
`
`implant wider than 32 mm contradicts how a person of ordinary skill in the art
`
`-5-
`
`

`

`
`
`would understand Berry and ignores the teachings of Brantigan, Michelson, and
`
`Frey.
`
`16. A person of ordinary skill in the art would not have modified the Frey
`
`or Brantigan implants for serial insertion of modular members into the disc space.
`
`Attempting to move a first implant to make room for a second implant would be
`
`quite difficult and would increase the risk of breaking the implant and injuring the
`
`patient. In sum, the proposed modification of the Frey and Brantigan implants and
`
`insertion method would have been unsafe and more invasive.
`
`V. A PERSON OF ORDINARY SKILL IN THE ART WOULD NOT
`HAVE ADOPTED BERRY’S VERTEBRAL BODY DIMENSIONS
`FOR IMPLANT DIMENSIONS
`17. Berry discloses dimensional data obtained from skeletons for lumbar
`
`and three thoracic vertebrae. EX1022 at 1, 3. That is, Berry measures bones that
`
`completely lack the soft tissue found on the vertebral bodies of patients.
`
`18. Figure 1 of Berry (reproduced below) describes the measurements
`
`taken of the skeletal vertebrae. For example, the line DEF (D = superior surface, E
`
`= middle, and F = inferior surface) defines the measurements made along the mid-
`
`sagittal plane of the vertebral body. EX1022, Figure 1. The D and F measurements
`
`are sometimes referred to as the anterior-posterior (“A-P”) dimensions or the depth
`
`of the vertebral body. E.g., EX1032, 2:5-7 (“[T]he depth of a vertebrae being the
`
`dimension of the vertebrae measured from the anterior end to the posterior end of
`
`-6-
`
`

`

`
`
`the vertebrae.”); EX2029 at 8. The line ABC (A = superior surface, B = middle,
`
`and C = inferior surface) defines the measurements made along the mid-coronal
`
`plane of the vertebral body. EX1022, Figure 1. The A and C measurements are
`
`sometimes referred to as the medial-lateral (“M-L”) dimensions or the transverse
`
`width of the vertebral body. EX1032, 3:7-9 (“The transverse width of a vertebra is
`
`measured from one lateral aspect of the spine to the opposite lateral aspect.”);
`
`EX2029 at 7 (“Morphometric studies of lumbar vertebrae [citing Berry] indicate
`
`that at L5, the ratio of A-P width to M-L width of vertebrae averages 0.66.”).
`
`
`
`19.
`
`I understand that Dr. Branch equates the size of vertebral bodies with
`
`the size of lumbar implants. E.g., EX1002, ¶¶152, 153. A person of ordinary skill
`
`in the art would not have designed a spinal fusion implant to be wider than the
`
`-7-
`
`

`

`
`
`depth of vertebral bodies. To the contrary, a person of ordinary skill in the art at
`
`the relevant time would have been familiar with vertebral body dimensions, such
`
`as those disclosed in Berry, and would have designed the width of the implant to
`
`be significantly narrower than the A-P dimension as measured at the sagittal
`
`midplane of the smaller of the two vertebral body endplates that define the
`
`intervertebral space. Certainly, following the actual teachings of Berry, one of
`
`ordinary skill would not have arrived at the absurdly sized spinal fusion implant
`
`(up to 37.9 mm wide) that I understand has been proposed by Petitioner. This is
`
`true regardless of whether one is considering sizing for Brantigan, Michelson, or
`
`Frey implants.
`
`20. A-P implant dimensions should be significantly narrower than the
`
`vertebral body dimensions reported in Berry because the available disc space for a
`
`spinal fusion implant is limited by soft tissue, such as the annulus fibrosis. The
`
`annulus fibrosis is about 5 mm thick, extends around the intervertebral space, and
`
`attaches to the apophyseal ring of adjacent vertebral bodies. EX2040 at Fig. 2. Dr.
`
`Branch’s use of Berry for sizing implants overlooks that a portion of the disc space
`
`is occupied by the annulus fibrosis.
`
`A. Brantigan’s Implants Are Design to Fit Within the Annulus Fibrosis
`
`-8-
`
`

`

`
`
`21. Brantigan’s implants are designed to fit within the annulus fibrosis
`
`(55a) and, as such, are smaller than vertebral bodies. EX1007, Fig. 11 (reproduced
`
`below).
`
`
`
`22. Brantigan expressly instructs that implants should fit within the
`
`annulus fibrosis. E.g., EX1007, 2:26-28 (“[T]he device stretches the disc tissue
`
`creating a tension which will cause the vertebrae to tightly grip the ring on which it
`
`is bottomed.”), 2:63-64 (“…stretching the annular fibrosis so that the vertebrae can
`
`tightly grip the plug…”), 6:25-32 (“A Z-shaped cut 58 through the tubular fibrous
`
`portion of the disc 55 provides access to the interior pulpus portion of the disc
`
`permitting its removal to receive a single plug 11 forming a rigid strut inside of the
`
`column of disc fibers 55a which remain attached to the bottom face 52b of the
`
`upper vertebrae 52 and the top face 57a of the lower vertebrae 57.”), 6:61 (“This
`
`stretches the remaining disc tissue …”).
`
`-9-
`
`

`

`
`
`23.
`
`I worked with Dr. Brantigan during my employment at AcroMed and
`
`I had a role in defining the sizing and other physical properties of the implants
`
`disclosed in the Brantigan patent. I know from firsthand experience that making
`
`implants as large or larger than the vertebral body dimensions reported in Berry is
`
`something that would not make sense to a person of ordinary skill in the art at the
`
`relevant time.
`
`24. Dr. Brantigan and I were well aware of the Berry reference when we
`
`were determining sizing for the lumbar interbody fusion implants. EX2029 at 7
`
`(citing Berry and relying on its disclosure of “the ratio of A-P width to M-L width
`
`of vertebrae”). However, we were also well aware that the space occupied by soft
`
`tissue must be accounted for when sizing implants.
`
`25. For example, lumbar spinal fusion surgeries had been performed prior
`
`to Brantigan’s patent and these earlier surgeries typically used tricortical bone from
`
`the patient’s iliac crest, processed allograft bone, or bovine bone. As a result,
`
`persons of ordinary skill in the art knew that a portion of the disc space is occupied
`
`by soft tissue such as the annulus fibrosis. See also EX2040, Figure 2A (describing
`
`the average width of the annulus as 5 mm).
`
`26. Given the known sizes of vertebral bodies, such as disclosed by Berry,
`
`and the known size of the annulus—the largest Brantigan implant was sized for the
`
`-10-
`
`

`

`
`
`lumbar region to be 28 mm in width (A-P dimension) by 42 mm in length (M-L
`
`dimension). EX2029 at 7; EX2030 at 3.
`
`
`
`27.
`
`I understand that the Petition asserts, based on Berry, that a skilled
`
`artisan would have wanted to increase the width of Brantigan’s implant to be as
`
`wide as 37.9 mm for insertion at L4-L5. This width is about 10 mm wider than the
`
`widest Brantigan implant and as such, it would not fit within the annulus. That is,
`
`the implant Dr. Branch proposes is contrary to Brantigan’s instructions to size an
`
`implant to fit within the annulus. An implant width of approximately 38 mm as
`
`proposed by the Petition is absurdly wide and would not have been made.
`
`B. Michelson Confirms that a Lumbar Implant Width Should Be Less
`Than that of the Vertebral Body
`28. The Michelson reference discloses that the maximum width of even
`
`an oversized lateral lumbar spinal fusion implant is 32 mm. EX1032, Abstract,
`
`-11-
`
`

`

`
`
`10:33-59. Michelson teaches that this dimension “approximates the depth of the
`
`vertebrae.” Id. Michelson is evidence that persons of ordinary skill in the art would
`
`have designed implants with dimensions significantly narrower than the depth of
`
`the vertebrae.
`
`29. Michelson repeatedly discloses that the implant width should not be as
`
`wide as the depth of the vertebrae. For example, Michelson discloses that the
`
`widths of the implants “need” to be “less than the depth of the vertebrae.” Id.,
`
`Abstract, 3:27-32. Michelson also explains that using an implant that exceeds the
`
`depth of the vertebrae is not “possible” and results in “great harm” to the patient:
`
`[T]he maximum possible length for an implant that is inserted from
`either the front or the back of the patient is limited to the depth of the
`vertebrae, the depth of a vertebrae being the dimension of the
`vertebrae measured from the anterior end to the posterior end of the
`vertebrae. It was not previously possible to insert an implant that had
`a length that was greater than the depth of the vertebrae from front to
`back as such an implant would protrude from either the anterior or
`posterior aspect of the spine resulting in great harm to the patient.
`Id., 2:3-12. For a lateral implant, as taught by Michelson, this means that the width
`
`of the implant needs to be narrower than the depth of the vertebrae. Id., 3:11-17
`
`(“[S]uch implants are necessarily limited by the depth, measured from front to
`
`back of the vertebrae”).
`
`C. Frey’s Implants Are Designed to Fit Within the Annulus Fibrosis
`30.
`I understand that the Petition asserts that Frey’s implants should be
`
`long enough to span the entire bony surface of the midline of the vertebral body as
`
`-12-
`
`

`

`
`
`measured by Berry. I disagree. Frey itself discloses that its implants are smaller
`
`than the vertebral body and that they are designed to fit within the annulus fibrosis.
`
`EX1040, Figs. 47, 53, 65.
`
`
`
`31. The commercial embodiment of Frey’s implant is consistent with
`
`Frey’s disclosure, describing the largest boomerang implant as 36 mm in length
`
`(M-L dimension) and 13 mm wide (A-P dimension). EX2031 at 2.
`
`
`
`
`
`VI. A PERSON OF ORDINARY SKILL IN THE ART WOULD NOT
`HAVE SEQUENTIALLY INSERTED LATERAL IMPLANTS
`A. Michelson Does Not Disclose Sequential Insertion of a Lateral Implant
`or Modular Members
`
`-13-
`
`

`

`
`
`32.
`
`I understand that Dr. Branch asserts that Michelson teaches a modular
`
`approach where components are inserted piece-by-piece into the interbody space
`
`and assembled together following insertion. I disagree. Michelson has no such
`
`teaching.
`
`33. Michelson does not disclose inserting implants into the disc space
`
`piece-by-piece. In fact, such an insertion method would be contrary to what
`
`Michelson describes as a benefit of the inventive implants—that is, the invention
`
`permits using a single implant inserted in a single procedure. EX1032, 3:46-49
`
`(“The dimensions of the translateral spinal fusion implant of the present invention
`
`permits a single implant to be inserted by a single procedure into the spine and to
`
`engage more of the adjacent vertebrae.”).
`
`34. Michelson describes the modular and non-modular embodiments as
`
`similar and discloses that the modular implant is assembled prior to insertion.
`
`The spinal fusion implant 1000 is similar to the spinal fusion implant
`900, but has a narrower width such that more than one spinal fusion
`implant 1000 may be combined in a modular fashion for insertion
`within the disc space D between the adjacent vertebrae.
`EX1032, 10:50-53 (emphasis added). A person of ordinary skill in the art would
`
`not have understood Michelson as teaching serial insertion of modular members.
`
`Instead, a skilled artisan would have understood Michelson to disclose what is
`
`stated expressly—the implant is assembled prior to insertion thereby retaining the
`
`benefits of using a single implant inserted through a single procedure.
`
`-14-
`
`

`

`
`
`B. Brantigan Does Not Disclose Sequential Insertion of a Lateral Implant
`or Modular Members
`35.
`I understand that Dr. Branch asserts that Brantigan discloses the same
`
`modularity concept as Michelson. This statement is true to the extent that both
`
`Michelson and Brantigan disclose modular embodiments that are assembled prior
`
`to insertion into the patient. However, as I explain above, Dr. Branch incorrectly
`
`describes the “modularity concept” of Michelson as comprising sequential
`
`insertion of modular members in the disc space. As with Michelson, Brantigan has
`
`no such disclosure. Furthermore, I understand that Dr. Branch conflates
`
`Brantigan’s disclosure regarding the vertically stacked modular embodiment with
`
`the non-modular posterior insertion embodiment.
`
`36. For example, Brantigan discloses an alternate embodiment for
`
`replacement of vertebral bodies—not an interbody fusion implant embodiment—
`
`where devices are assembled in a vertically stacked configuration. EX1007,
`
`Abstract, 2:34-43. Brantigan discloses that multiple devices may be “stacked to the
`
`exact height required.” EX1007, 2:35-38; see also id., 2:41-43 (“An average
`
`grafting height is 30 mm after corpectomy and this can be achieved by stacking,
`
`for example, three 10 mm high oval implants.”), Figs. 1, 3, 4 (shown below).
`
`-15-
`
`

`

`
`
`
`
`37.
`
`I have firsthand knowledge of the development of Brantigan’s
`
`vertebral body replacement embodiments, as I was involved in their development.
`
`These stacked devices are assembled and held together by connecting bar 15 that
`
`conforms to the height of the stack and locks the pieces of the stack together via
`
`slot 14. EX1007, Abstract (“The implants can be provided in sets of different
`
`thicknesses and are internally grooved to receive an upstanding connecting bar to
`
`bind together the individual stacked implants into a stable unit.”), 1:59-61 (“When
`
`stacked, an interior connecting bar can be provided to lock the components in fixed
`
`relation and cooperate with interfitting ridges.”), 3:25-28 (“FIG. 3 is a top and side
`
`perspective view of a connecting bar fitting the illustrated grooves in the devices of
`
`FIGS. 1 and 2 to hold a plurality of the devices in stacked relation”), 3:29-31
`
`(“FIG. 4 is a top and side perspective view of a stack of the devices of FIG. 1 with
`
`the connecting bar of FIG. 3 in place”), 4:23-27 (describing Fig. 1 as having
`
`“vertical grooves 14 … for mounting a rectangular connecting bar 15 shown in
`
`-16-
`
`

`

`
`
`FIG. 3.”), 4:38-49 (“The connecting bar 15 has a height conforming with the total
`
`height of a stack 17 of plugs 11 shown in FIG. 4 or with only a single plug 11 if a
`
`stack of plugs is not necessary. As shown in FIG. 4 three plugs 11 are stacked
`
`together with the ridges 12 of the intermediate plug nested in and interdigitating
`
`with the ridges of top and bottom plugs. These ridges interfit to provide a stable
`
`stack and the connecting bar 15 seated in the aligned grooves 14 of the three plugs
`
`will prevent shifting of the stack.”), 5:18-21 (“The lengths or heights of the
`
`connecting bars 15 can also be varied to suit conditions or can be ground down at
`
`the time of the operation to match the stack.”).
`
`38. A person of ordinary skill in the art would have understood that the
`
`stacked device of Brantigan are assembled prior to insertion in the patient. It would
`
`be physically impossible to insert the connecting bar 15 into slot 14 after placing
`
`the individual pieces into the patient.
`
`39.
`
`I understand that Dr. Branch also points to the posterior lumbar
`
`implant embodiment of Brantigan as being similar to the modular embodiment of
`
`Michelson. Brantigan discloses “[t]wo such hemi-oval rings can be used in the
`
`posterior lumbar area in side-by-side relation since the dural sac and nerve roots
`
`must be retracted to each side in turn as the implant is placed on the opposite side.”
`
`EX1007, 2:7-11. This portion of Brantigan is describing implants for posterior
`
`insertion—not a “modular” assembly. Because posterior access to the center of the
`
`-17-
`
`

`

`
`
`disc space is blocked by the spinal canal typically two smaller implants are used.
`
`This concept is well illustrated in the Telamon Guide. See generally EX1011; see
`
`also id., 8 (reproduced below, showing the arrangement of two smaller implants).
`
`
`
`40. That is, Brantigan is not here disclosing a modular implant, but rather
`
`disclosing that two implants should be inserted along two posterior surgical paths
`
`into the disc space.
`
`41.
`
`I was further asked to comment on the hypothetical placement of two
`
`Brantigan implants side-by-side in the disc space. Brantigan does not teach placing
`
`two oval implants side-by-side and doing so would run contrary to Brantigan’s
`
`teaching.
`
`42. The shape of the Brantigan implant is poorly suited for side-by-side
`
`placement in the interbody space. Brantigan instructs that the implant should have
`
`-18-
`
`

`

`
`
`“dimensions in the same ratio as normal vertebral bodies” and should “conform
`
`with the general outline perimeter of the vertebrae.” EX1007, 1:18-21, 2:2-4. The
`
`proposed side-by-side placement would leave unsupported gaps on the lateral
`
`aspects of the two implants, as shown below.
`
`
`
`The gaps on the lateral aspects of the two implants would be expected to result in
`
`an unstable platform for the spinal column. Moreover, such an arrangement would
`
`fundamentally alter the implant design such that it no longer conforms to the
`
`general outline perimeter of the vertebrae. In sum, such a configuration would
`
`result in a less stable platform versus retaining Brantigan’s original implant design.
`
`43. This modification also would disregard Brantigan’s instruction to
`
`make implants with the same width-to-length ratio as normal lumbar vertebral
`
`bodies. Brantigan specifically used “the ratio of A-P width to M-L width of
`
`-19-
`
`

`

`
`
`vertebrae” reported by Berry in sizing his implants. EX2029 at 8; see also id.
`
`(disclosing using the average width to length ratio for L5 (0.66) as reported by
`
`Berry). Brantigan believed, as did I, that “the ratio of dimension of the cages
`
`should duplicate the known average dimensions of lumbar vertebrae” and that “the
`
`different sizes should maintain about the same ratio.” EX2029 at 8; EX1007, 1:18-
`
`21 (“The rings are bottomed on the opposing end faces of adjoining vertebrae, are
`
`preferably oval shaped with medial-lateral and anterior-posterior dimensions in the
`
`same ratio as normal vertebral bodies…”). Our implants were “oval shaped to
`
`conform with the general outline perimeter of the vertebrae.” EX1007, 2:2-4.
`
`Placing two Brantigan implants side by side would abandon this ratio and shape,
`
`contradict the teachings of Brantigan and Berry, and result in an inferior implant.
`
`VII. THE PROPOSAL TO INSERT FREY’S IMPLANTS THROUGH A
`HOLLOW TUBE IS NONSENSICAL
`44.
`I understand that Dr. Branch proposes inserting Frey’s implants
`
`through a hollow tube. Inserting Frey’s implants through a hollow tube would be
`
`disfavored at least because it (1) disregards the surgical approach for which Frey
`
`was specifically designed, and (2) would have increased the risk of damage to the
`
`implant.
`
`45. The Frey implant is designed for insertion in a transforaminal lumbar
`
`interbody fusion (i.e., TLIF) surgery. Spinal fusion implants are made to fit tightly
`
`in the disc space and to withstand the force needed to push the implant into
`
`-20-
`
`

`

`
`
`position. E.g., EX1032, 8:20-22 (“The implant driver instrument may then be
`
`impacted by a mallet, or similar device…”); EX1040, ¶139 (“[T]he surgeon
`
`controls inserter 1100 with one hand while the other hand provides a pushing or
`
`impaction force on implant 1000 with pusher 1200.”). As such, Frey’s implants
`
`were designed to withstand the forces associated with a curved path through the
`
`disc space.
`
`46.
`
`Inserting Frey’s implants through a hollow tube would require
`
`substituting a linear insertion pathway into the disc space instead of the curved
`
`path Frey was designed for. The curved insertion path requires radial movement of
`
`the insertion tool (EX1040, ¶¶148,149) and this movement is not possible if the
`
`insertion tool must fit within the hollow tube described by Michelson. The
`
`insertion pathway would thus have to be linear.
`
`47. Applying an impaction force on the boomerang implant of Frey in a
`
`linear direction would subject the implant to forces that it was not designed to
`
`withstand. Specifically, applying force linearly would place a great deal of stress at
`
`the center of the implant which would increase the risk of implant failure.
`
`VIII. BRANTIGAN’S INTERBODY FUSION IMPLANTS DO NOT HAVE A
`LONGITUDINAL APERTURE LENGTH PARALLEL TO THE
`LONGITUDINAL LENGTH OF THE IMPLANT
`48. The petition materials allege that the first fusion aperture limitation is
`
`met by the implants exemplified in Figures 1 (implant 11) and 6 (implant 30).
`
`-21-
`
`

`

`
`
`However, neither implant 11 nor 30 has a longitudinal aperture length parallel to
`
`the longitudinal length of the implant.
`
`
`
`49. The difference between implants 11 and 30 is whether the reinforcing
`
`bar 32 is integral to the implant. As shown above, the reinforcing bar 32 separates
`
`the interior portion of the implant 30 into two side-by-side apertures. EX1007,
`
`5:40-44 (“The hollow interior 23 of the plug 31 is thus bisected by an integral
`
`internal partition 32 forming a pair of side-by-side apertures through the plug
`
`adapted to receive bone graft material.”). Likewise, the connecting bar 15 separates
`
`the interior portion of implant 11 into two side-by-side apertures. EX1007, 4:38-41
`
`(“The connecting bar 15 has a height conforming with the total height of a stack 17
`
`of plugs 11 shown in FIG. 4 or with only a single plug 11 if a stack of plugs is not
`
`necessary.”), 4:50-54 (“The central aperture 11d of each plug 11 is separated by
`
`the bar 15 into two side-by-side chambers which are easily packed with bone graft
`
`-22-
`
`

`

`
`
`material…”), (“Instead of providing a separate bar or plate 15, as shown in FIG. 6,
`
`a modified device 30 of this invention is a plug 31 of the same oval shape as the
`
`plug 11 of FIGS. 1 and 4 but the reinforcing bar 32 of this plug is integral with its
`
`side walls 31a.”).
`
`50. The “longitudinal length” is the longest dimension of the implant. The
`
`challenged claims require that the “longitudinal length [of the implant]…extend[s]
`
`from a proximal end of said proximal wall to a distal end of said distal wall.” That
`
`is, the longest dimension of the implant must be the length between the proximal
`
`and distal walls. Furthermore, the longitudinal length of the aperture must also
`
`“extend[] generally parallel to the longitudinal length of said implant.” Therefore,
`
`the longitudinal length of the aperture must be generally parallel to the longest
`
`dimension of the implant.
`
`51.
`
` The first fusion aperture of implant 30 does not have a “longitudinal
`
`aperture length extending generally parallel to the longitudinal length of said
`
`implant, and a lateral aperture width extending between said first sidewall to said
`
`second sidewall, wherein the longitudinal aperture length is greater than the lateral
`
`aperture width.” The longest dimension of the first aperture is not parallel to the
`
`long axis of the implant—it is orthogonal.
`
`52.
`
`Implant 11 also does not have a “longitudinal aperture length
`
`extending generally parallel to the longitudinal length of said implant, and a lateral
`
`-23-
`
`

`

`
`
`aperture width extending between said first sidewall to said second sidewall,
`
`wherein the longitudinal aperture length is greater than the lateral aperture width.”
`
`Figure 1 depicts an unassembled implant 11 lacking connecting bar 15. When
`
`connecting bar 15 is inserted into slot 14, two side-by-side apertures are created in
`
`implant 11 with the same dimensions as the apertures of implant 30.
`
`53. Accordingly, as with implant

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket