throbber
Anal. Chem. 1994, 66, 4127-4132
`
`Precolumn Reactions with Electrophoretic
`Analysis Integrated on a Microchip
`Stephen C. Jacobson, Roland Hergenroder, Alvin W. Moore, Jr., and J. Michael Ramsey*
`Chemical and Analytical Sciences Division, Oak Ridge National Laboratory, P.O. Box 2008, Building 4500S, MS 6142,
`Oak Ridge, Tennessee 37831-6142
`
`A glass microchip was constructed to perform chemical
`reactions and capillaiy electrophoresis sequentially. The
`channel manifold on the glass substrate was fabricated
`using standard photolithographic, etching, and deposition
`techniques. The microchip has a reaction chamber with
`a 1 nL reaction volume and a separation column with a
`separation length. Electrical control of the
`15.4 mm
`buffer, analyte, and reagent streams made possible the
`precise manipulation of the fluids within the channel
`manifold. The microchip was operated under a continu-
`ous reaction mode with gated injections to introduce the
`reaction product onto the separation column with high
`reproducibility (<1.8% rsd in peak area). The reaction
`and separation performances were evaluated by reacting
`amino acids with o-phthaldialdehyde to generate a fluo-
`rescent product which was detected by laser-induced
`fluorescence. Control of the reaction and separation
`conditions was sufficient to measure
`reaction kinetics and
`variation of detection limits with reaction time. Half-times
`of reaction of 5.1 and 6.2 s and detection limits of 0.55
`and 0.83 fmol were measured for arginine and glycine,
`respectively.
`
`The microfabrication of analytical instrumentation will poten-
`tially enable the laboratory to be transported to the samples rather
`than vice versa. Miniature chemical instrumentation can be based
`on conventional laboratory approaches to chemical measurement
`problems. Many laboratory-based procedures require sample
`manipulation prior to the actual measurement, and many analyses
`are fully automated to circumvent operator bias. Similarly, a
`portable instrument should have these sample preparatory steps
`incorporated into the design and function of the instrument.
`In
`instruments as opposed to
`addition, micromachined chemical
`single-analyte chemical sensors should be able to identify and
`quantify all members of a desired class of compounds using a
`single device. The approach taken is to micromachine a channel
`manifold in a monolithic device that includes all desired fluid
`manipulation for complete chemical analysis. Chemical separation
`techniques including capillary electrophoresis,1-6 free-flow elec-
`
`(1) Harrison, D. J.; Manz, A.; Fan, Z.; Liidi, H.; Widmer, H. M. Anal. Chem.
`1992, 64, 1926.
`(2) Manz, A; Harrison, J.; Verpoorte, E. M. J.; Fettinger, J. C.; Paulus, A; Liidi,
`H.; Widmer, H. M. J. Chromatogr. 1992, 593, 253.
`(3) Seiler, K; Harrison, D. J.; Manz, A Anal. Chem. 1993, 65, 1481.
`(4) Harrison, D. J.; Fluri, K; Seiler, K; Fan, Z.; Effenhauser, C. S.; Manz, A.
`Science 1993, 261, 895.
`(5) Jacobson, S. C.; Hergenroder, R; Koutny, L. B.; Warmack, R J.; Ramsey,
`J. M. Anal. Chem. 1994, 66, 1107.
`
`0003-2700/94/0366-4127$04.50/0 &copy; 1994 American Chemical Society
`
`trophoresis,1 2345*7 *open channel electrochromatography,8 and capillary
`electrophoresis with postcolumn derivatization9 have been dem-
`onstrated using microfabricated devices.
`To demonstrate a reactor/analyzer microchip, we chose the
`relatively simple reaction of amino acids with o-phthaldialdehyde
`(OPA), which yields a fluorescent product.10 For capillary elec-
`trophoresis, both pre- and postseparation labeling using OPA have
`been studied.11 The reaction is moderately fast (a half-time of
`s 12), but the
`reaction for alanine at room temperature of
`fluorescent product can be short lived («10 min for glycine13).
`This reaction is fast enough to perform postcolumn labeling when
`the analysis time is long compared to the reaction time, but as
`analysis times decrease, the postcolumn labeling becomes less
`attractive. Both detection limits and separation efficiency suffer
`due to “slow” product formation when high-speed analysis is
`performed.9 Analysis times on microchip electrophoresis devices
`can be significantly faster than the OPA reaction times; separation
`times for microchip electrophoresis as short as 150 ms have been
`reported.6 Precolumn derivatization becomes the desired method
`if the electrophoretic analysis can be performed before the product
`degrades and without hindering the quality of the separation.
`In this paper, a simple channel manifold was fabricated on a
`glass microchip to perform on-line precolumn reactions coupled
`with electrophoretic analysis of the reaction products. Here, the
`reactor is operated continuously with small aliquots introduced
`periodically onto the separation column to be analyzed. The
`operation of the microchip consists of three elements: derivati-
`zation of the amino acids with OPA, injection of the sample onto
`the separation column, and separation/detection of the compo-
`nents of the reactor effluent. Each of these elements is evaluated.
`
`EXPERIMENTAL SECTION
`The microchips were fabricated using standard photolitho-
`graphic, wet chemical etching and bonding techniques.14 A
`
`(6) Jacobson, S. C.; Hergenroder, R; Koutny, L. B.; Ramsey, J. M. Anal. Chem.
`1994, 66, 1114.
`(7) Raymond, D. E.; Manz, A; Widmer, H. M. Anal. Chem. 1994, 66, 2858.
`(8) Jacobson, S. C.; Hergenroder, R; Koutny, L. B.; Ramsey, J. M. Anal. Chem.
`1994, 66, 2369.
`(9) Jacobson, S. C.; Koutny, L. B.; Hergenroder, R; Moore, A. W., Jr., Ramsey,
`J. M. Anal. Chem. 1994, 66, 3472.
`(10) Roth, M. Anal. Chem. 1971, 43, 880.
`(11) For example: Rose, D.J.; Jorgenson, J. W.J. Chromatogr. 1988,447,117.
`Pentoney, S. L, Jr.; Huang, X.; Burgi, D. S.; Zare, R N. Anal. Chem. 1988,
`60, 2625. Nickerson, B.; Jorgenson, J. W.J. Chromatogr. 1989, 480, 157.
`Albin, M.; Weinberger, R; Sapp, E.; Moring, S. Anal. Chem. 1991, 63,
`417.
`(12) Butchner, E. C.; Lowry, O. H. Anal. Biochem. 1976, 76, 502.
`(13) Lindroth, P.; Mopper, K Anal. Chem. 1979, 51, 1667.
`(14) For example: Ko, W. H.; Suminto, J. T. In Sensors; Gopel, W., Hasse, J.,
`Zemmel, J. N., Eds.; VCH: Weinhein, Germany, 1989; Vol. 1, pp 107-168.
`Analytical Chemistry, Vol. 66, No. 23, December 1, 1994 4127
`
`Downloaded via UNIV OF VIRGINIA on November 6, 2018 at 17:53:41 (UTC).
`
`See https://pubs.acs.org/sharingguidelines for options on how to legitimately share published articles.
`
`Agilent Exhibit 1283
`Page 1 of 6
`
`

`

`Figure 1. Schematic of the microchip with integrated precolumn
`reactor. The reaction chamber is 2 mm long, and the separation
`column is 15.4 mm long.
`
`photomask was fabricated by sputtering chrome (50 nm) onto a
`glass slide and ablating the microchip design (Figure 1) into the
`chrome film using a CAD/CAM excimer laser machining system
`(ArF, 193 nm; Resonetics, Inc.). The column design was then
`transferred onto the substrates using a positive photoresist
`(Shipley 1811). The channels were etched into the substrate in
`a dilute, stirred HF/NH4F bath. To form the closed network of
`channels, a cover plate was bonded to the substrate over
`the
`etched channels by use of a direct bonding technique.5 Cylindrical
`glass reservoirs were affixed on the substrate using epoxy.
`Platinum electrodes provided electrical contact from the power
`supply (Spellman CZE1000R) to the solutions in the reservoirs.
`The reaction chamber was designed to be wider than the
`separation column to give lower electric field strengths in the
`reaction chamber and thus longer residence times for the
`reagents. Figure 2a shows a scanning electron microscope (SEM)
`image of the injection cross and the lower portion of the reaction
`chamber prior to bonding of the cover plate. The image has the
`same orientation as the microchip schematic in Figure 1. The
`reaction chamber is 96 /im wide at half-depth and 6.2 /<m deep,
`and the separation column is 31 /4m wide at half-depth and 6.2
`«m deep. Figure 2b shows an SEM image of the cross section of
`the separation channel. The microchip was broken perpendicular
`to the separation column ~1 mm downstream from the injection
`cross. The isotropic etch of the glass substrate, i.e., uniform etch
`in all directions, is evident in the trapezoidal geometry of the cross
`section.
`Column performance and separations were monitored on-
`microchip using a single-point detection scheme via laser-induced
`fluorescence (LIF). An argon ion laser
`(351.1 nm, 10 mW;
`Coherent Innova 90) was used for excitation and focused to a spot
`onto the microchip using a lens (100 mm focal
`length). The
`fluorescence signal was collected using a 20x objective lens,
`followed by spatial filtering (0.6 mm diameter pinhole) and spectral
`filtering (440 nm bandpass, 10 nm bandwidth), and measured
`using a photomultiplier tube (PMT; Oriel 77340). The schematic
`
`4128 Analytical Chemistry, Vol. 66, No. 23, December 1, 1994
`
`(a, top) Scanning electron microscope (SEM) image of
`Figure 2.
`is the lower portion of the
`injection cross. The broader channel
`is 6.2 /<m deep and 31
`reaction chamber. The separation channel
`//m wide, and the reaction chamber is 6.2 nm deep and 96 /<m wide.
`The SEM is in the same orientation as Figure 1. (b, bottom) SEM
`section of separation column after bonding of
`image of cross
`coverplate to the substrate.
`
`in Figure 3a shows the data acquisition/voltage switching ap-
`paratus, which is computer controlled using programs written in-
`house in Labview 3.0 (National Instruments). The compounds
`used for the experiments were arginine (0.48 mM), glycine (0.58
`mM), and o-phthaldialdehyde (5.1 mM; Sigma Chemical Co.). The
`buffer in all of the reservoirs was 20 mM sodium tetraborate with
`2% (v/v) methanol and 0.5% (v/v) 2-mercaptoethanol. 2-Mercap-
`is added to the buffer as a reducing agent for the
`toethanol
`derivatization reaction.10
`Figure 3b shows the ratio of the potentials at each of the
`reservoirs for a given potential applied to the system. This
`configuration allowed the lowest potential drop across the reaction
`chamber (25 V/cm for 1.0 kV applied to the microchip) and the
`highest across the separation column (300 V/cm for 1.0 kV applied
`to the microchip) without significant bleeding of the product into
`the separation column. The voltage divider used to establish the
`potentials applied to each of the reservoirs had a total resistance
`of 100 with 10 MQ divisions. The sample and reagent are
`electroosmotically pumped into the reaction chamber with a
`the solutions from the
`volumetric ratio of 1:1.06. Therefore,
`analyte and reagent reservoirs are diluted by a factor of %2. Buffer
`
`Agilent Exhibit 1283
`Page 2 of 6
`
`

`

`Figure 3.
`(a, top) Diagram of the high-voltage switching apparatus
`and detection/data acquisition system, (b, bottom) Block diagram of
`flow pattern for precolumn reactor microchip with the relative potentials
`applied at each reservoir (arrows depict direction of flow).
`
`was simultaneously pumped by electroosmosis from the buffer
`reservoir toward the waste and analyte waste reservoirs. This
`buffer stream prevents the newly formed product from bleeding
`into the separation column. A gated injection scheme, described
`previously,9 is used to inject effluent from the reaction chamber
`onto the separation column. The potential at the buffer reservoir
`is simply floated (opening of the high-voltage switch in Figure
`3a) for a brief period of time (0.1—1.0 s), and sample migrates
`into the separation column as in an electrokinetic injection.15 To
`break off the injection plug, the potential at the buffer reservoir
`is reapplied (closing of the high-voltage switch in Figure 3a). The
`length of the injection plug is a function of both the time of the
`injection and the electric field strength. With this configuration
`of applied potentials, the reaction of the amino acids with the OPA
`continuously generates fresh product to be analyzed.
`
`RESULTS AND DISCUSSION
`Fluid manipulation, including sample injections, is performed
`by simply controlling the potentials applied to the reservoirs; i.e.,
`no valves or pumps are required. This allows reagents to be
`mixed accurately and samples to be injected reproducibly onto
`the separation column. Using the gated injection scheme de-
`scribed above, Figure 4 shows four injections of glycine derivatized
`with OPA for increasing injection times of 0.2, 0.4, 0.6, and 0.8 s
`with an injection field strength of 0.6 kV/cm. The injection field
`strength is the electric field strength in the separation column
`during the injection; i.e., the high-voltage switch in Figure 3a is
`open. Similar profiles were obtained for arginine. The injections
`were observed in the separation column 0.05 mm downstream
`from the injection cross using the single-point detection scheme.
`The peak fronts coincide as expected, and the durations of the
`injection profiles correspond to the injection times.
`
`(15) Jorgenson, J. W.; Lukacs, K D. Anal. Chem. 1981, 53, 1298.
`
`time [s]
`Figure 4. Series of injection profiles for injection times of (a) 0.2,
`(b) 0.4, (c) 0.6, and (d) 0.8 s with £j„j = 0.6 kV/cm and Ubs = 0.05
`
`20
`
`15
`
`T3
`
`5
`
`2 % -> 
`0
`
`0.0
`
`0.2
`
`1.0
`
`0.6
`0.8
`0.4
`injection time [s]
`Figure 5. Reproducibility of injections as percent relative standard
`deviation (% rsd) of the peak area with injection time for arginine ( )
`and glycine (•) at £jnj = 1.2 kV/cm and for arginine (a.) and glycine
`( ) at £jnj = 0.6 kV/cm with Ubs = 0.05 mm. Dashed line represents
`2% rsd.
`
`A significant shortcoming of many capillary electrophoresis
`experiments has been the poor reproducibility of the injections.
`Here, because the microchip injection process is computer
`controlled, and the injection process involves the opening of a
`single high-voltage switch, the injections can be accurately timed
`events. Figure 5 shows the reproducibility of the amount injected
`(percent relative standard deviation, % rsd, for the integrated areas
`for both arginine and glycine at
`injection field
`of the peaks)
`strengths of 0.6 and 1.2 kV/cm and injection times ranging from
`0.1 to 1.0 s. For injection times of >0.3 s, the percent relative
`standard deviation is below 1.8%. This is comparable to reported
`values for commercial, automated capillary electrophoresis instru-
`ments.16 However, injections made on the microchip are «100
`times smaller in volume, e.g., 100 pL on the microchip versus 10
`nL on a commercial instrument. Part of this fluctuation is due to
`the stability of the laser, which is «0.6%. For injection times of
`
`(16) For example, Hewlett-Packard reports 1-2% rsd for peak area measure-
`ments.
`
`Analytical Chemistry, Vol. 66, No. 23, December 1, 1994
`
`4129
`
`Agilent Exhibit 1283
`Page 3 of 6
`
`

`

`Figure 6. Variation of injection widths with injection time for arginine
`( ) and glycine (•) at Enj = 1.2 kV/cm and for arginine (a) and glycine
`( ) at Enj = 0.6 kV/cm with Lobs = 0.05 mm. Lines represent best fit
`for injection times of >0.2 s for arginine and >0.3 s for glycine.
`
`>0.3 s, the error appears to be independent of the compound
`injected and the injection field strength.
`In Figure 6, the injection plug lengths at half-height are plotted
`versus the injection time. For injection times of >0.2 s for arginine
`and >0.3 s for glycine at the two selected injection field strengths,
`the lengths of the injected plugs are a linear function of the
`injection times and injection field strengths. Because the injection
`is electrophoretically biased, and arginine has a higher electro-
`phoretic mobility than glycine, injection widths are linear at shorter
`injection times for arginine. The injection widths are not propor-
`to the injection times for shorter injection times. The
`tional
`region has a finite width, and once filled with
`injection cross
`analyte, the area must be swept clean after the injection. Lines
`with a y-intercept equal to zero are fitted to the data for injection
`times of >0.2 s for arginine and >0.3 s for glycine. The slopes of
`the lines are 3.13 and 1.51 mm/s for arginine with injection field
`strengths of 1.2 and 0.6 kV/cm, respectively, and 1.44 and 0.692
`mm/s for glycine with injection field strengths of 1.2 and 0.6 kV/
`cm, respectively. For both arginine and glycine, the slopes for
`the injection field strength of 1.2 kV/cm are 2 times greater than
`at 0.6 kV/cm. For the experiments described below, the injector
`was operated in the reproducible (Figure 5) and linear ranges
`(Figure 6), and the injection times were scaled to give comparable
`peak widths and, thus, injection volumes for separations performed
`at different separation field strengths. The electrophoretic bias
`in the injections requires the injection time to be long enough to
`inject accurately the compound with the lowest electrophoretic
`mobility, e.g., glycine. This, in turn, means compounds with a
`higher electrophoretic mobility, e.g., arginine, are injected in
`relative excess.
`Figure 7 shows the overlay of three electrophoretic separations
`of arginine and glycine after on-microchip precolumn derivatization
`with OPA with a separation field strength of 1.8 kV/cm and a
`separation length of 10 mm. The separation field strength is the
`electric field strength in the separation column during the
`separation; i.e., the high voltage switch in Figure 3a is closed.
`The field strength in the reactor is 150 V/cm. The reaction times
`
`4130 Analytical Chemistry, Vol. 66, No. 23, December 1, 1994
`
`time [s]
`Figure 7. Overlay of
`three electropherograms of arginine and
`glycine using precolumn derivatization with o-phthaldialdehyde with
`Esep = 1.8 kV/cm and Lsep = 10 mm.
`
`for the analytes are inversely related to their mobilities; e.g., for
`arginine the reaction time is 4.1 s and for glycine the reaction
`time is 8.9 s. The volumes of the injected plugs were 150 and 71
`pL for arginine and glycine, respectively, which correspond to 35
`and 20 fmol of the amino acids injected onto the separation
`column. The gated injector allows rapid sequential injections to
`In this particular case, an analysis could be performed
`be made.
`every 4 s. The observed electrophoretic mobilities for the
`compounds are determined by a linear fit to the variation of the
`linear velocity with the separation field strength. The slopes were
`29.1 and 13.3 mm2/ (kV-s) for arginine and glycine, respectively.
`No evidence of Joule heating was observed, as indicated by the
`field strength data. A linear fit
`linearity of the velocity versus
`produced correlation coefficients of 0.999 for arginine and 0.996
`for glycine for separation field strengths from 0.2 to 2.0 kV/cm.
`With increasing potentials applied to the microchip, the field
`strengths in the reaction chamber and separation column increase.
`This leads to shorter residence times of the reactants in the
`reaction chamber and faster analysis times for the products. By
`the reaction
`varying the potentials applied to the microchip,
`kinetics can be studied. The variation in amount of product
`generated with reaction time is plotted in Figure 8. The response
`is the integrated area of the peak corrected for the residence time
`in the detector observation window and photobleaching of the
`product (discussed below). The offset between the data for the
`arginine and the glycine in Figure 8 is due primarily to the
`difference in the amounts injected, i.e., different electrophoretic
`mobilities, for the amino acids. A 10-fold excess of OPA was used
`to obtain pseudo-first-order reaction conditions. The slopes of the
`lines fitted to the data correspond to the rates of the derivatization
`reaction. The slopes are 0.13 s_1 for arginine and 0.11 s_1 for
`glycine, corresponding to half-times of reaction of 5.1 and 6.2 s,
`respectively. These half-times of reaction are comparable to the
`4 s previously reported for alanine.12 We have found no previously
`reported data for arginine or glycine.
`The reaction time varies with the potential applied to the
`microchip. For lower field strengths in the reaction chamber, the
`reaction time is longer and should yield more product and larger
`In Figure 9, the mass detection limits (signal-to-noise
`signals.
`ratio (S/N) equal to 2) are estimated for a series of reaction times
`for arginine and glycine. For longer reaction times, the mass
`detection limits improve as expected. Detection limits of 0.55 fmol
`
`Agilent Exhibit 1283
`Page 4 of 6
`
`

`

`0.0
`
`S = kIoFNe~At/r
`
`(1)
`
`where k is a geometrical factor, I is the intensity of the laser (~400
`W/cm2), oF is the fluorescence cross section, N is the number of
`molecules in the detection volume, Af is the detector residence
`time (0.01-0.1 s for arginine and 0.02-0.25 s for glycine), and r
`is the photochemical lifetime, k, /, oF, and N are constant for the
`experiments. The residence time of the product in the detection
`volume is known, and thus, the photochemical lifetime, r, of the
`reaction product can be estimated by use of a linear fit of the
`variation of In S with At. The photochemical lifetimes are 51 ms
`for arginine and 58 ms for glycine under these experimental
`conditions. The products are known to have poor stability,13 and
`not surprisingly, the photochemical lifetimes are short compared
`to the longer observation times. Accurate quantitative measure-
`ments must take account of the photobleaching effect.
`To improve the peak capacity of the separation,
`the band
`dispersion needs to be minimized. For electrophoretic separations
`the total plate height can be expressed as
`^total = ^inj + #det + #diff
`
`(2)
`
`where Hmi, Ha&, and Hm are the contributions to the plate height
`from the injection plug length, the detector observation length,
`and axial diffusion, respectively. The contributions from the
`injection plug length and the detector observation length are time
`independent, and the contribution to the plate height from axial
`diffusion is time dependent. Effects such as Joule heating are
`not considered because the variation of the linear velocity with
`the separation field strength is linear, as mentioned above.
`The contributions to the plate height from the injection plug
`length and the detector observation length are17
`
`Hh =
`
`(jnj2/ (16Lsep)
`
`tfdet =
`
`/det2/(16LSep)
`
`(3)
`
`(4)
`
`where lm) and ldet are the lengths of the injection plug and detector
`observation, respectively, and Lsep is the separation length. The
`contributions from the injection plug length (see injection profiles
`in Figure 4) and detector observation length are written as
`Gaussian functions. The lengths of the injection plug and the
`If these
`detector observation are constant for all experiments.
`time-independent contributions predominate in their contribution
`to the plate height, then the total plate height decreases as the
`separation length increases.
`The contribution to the plate height from axial diffusion is18
`Hm = 2DJ{uE)
`
`(5)
`
`where pi is the effective electrophoretic mobility of the analyte, E
`is the electric field strength, and Dm is the diffusion coefficient of
`the analyte in the buffer. The contribution from axial diffusion
`to the plate height is reduced by increasing the separation field
`strength and, consequently, reducing the analysis time.
`
`(17) Sternberg, J. C. Adv. Chromatogr. 1966, 2, 205.
`(18) Giddings, J. C. Dynamics of Chromatography, Part I: Principles and Theory,
`Marcel Dekker: New York, 1965; Chapter 2.
`
`Analytical Chemistry, Vol. 66, No. 23, December 1, 1994
`
`4131
`
`reaction time [s]
`Figure 8. Variation of product formation (In [response]) with reaction
`time for arginine (•) and glycine ( ) derivatized with o-phthaldialde-
`hyde. Lines represent linear fits. Error bars are ±o for three runs.
`
`Figure 9. Variation of detection limits (S/N = 2) with reaction time
`for arginine (•) and glycine ( ) derivatized with o-phthaldialdehyde.
`Error bars are ±o for three runs.
`
`for arginine and 0.83 fmol for glycine are reached for reaction
`times of 36 and 80 s, respectively. The difference in detection
`limits between arginine and glycine is assumed to be differences
`in relative fluorescence and in separation efficiency. The source
`of the limiting background for this experiment is fluorescence
`from the glass substrate. Detection limits could be improved by
`using fused silica substrates, which would minimize the back-
`ground fluorescence.
`Photobleaching can affect quantitative results in fluorescence
`measurements. The degree of photobleaching was estimated by
`determining the photochemical lifetimes of the fluorescent prod-
`ucts. The fluorescence signal was measured as a function of the
`residence time in the detector observation window by running a
`frontal electropherogram of the individual products. The fluo-
`rescence signal, S, for the continuous stream measurements is
`
`Agilent Exhibit 1283
`Page 5 of 6
`
`

`

`especially at the channel walls. As seen in the SEM image in
`Figure 2a, the separation channel walls have a noticeable rough-
`ness. These etch defects derive primarily from the imperfections
`in the photomask. The photomask that was generated using UV
`ablation of a thin chrome film had an edge smoothness of ± 1
`wide line. This roughness in the mask is then
`fxm on a 25
`transferred to the photoresist during exposure and, subsequently,
`In Figure 10, the plate height increases
`etched into the substrate.
`If related to the surface
`with increasing electric field strength.
`roughness, the magnitude of such a contribution is difficult to
`quantitate since the exact nature of the channel surface is not
`known. Prior work at comparable separation field strengths
`yielded more favorable results and agrees with eq 2.6 However,
`in that work, an e-beam written photomask, which had edge
`smoothness more than 1 order of magnitude better, was used for
`the microchip fabrication. The threshold of roughness that can
`be tolerated for the photomask and thus the channel wall is under
`further investigation.
`
`CONCLUSION
`Chemical reactions followed by analysis of the products have
`been demonstrated on a single microchip. Devices can be tailored
`to accommodate any such reaction to be performed in a miniatur-
`ized environment. The reactor volumes and separation lengths
`can be optimized accordingly. Electroosmotically driven flow
`enables buffer, analyte, and reagent streams to be controlled
`precisely in the channel manifold of the microchip without the
`use of valves or pumps. The gated injector allows continuous
`loading of sample and rapid sequential injections to be made. Low
`dead volume connections between channels can be easily fabri-
`cated allowing pre- and postseparation reactions to be incorporated
`on-microchip. Having an electrode placed at the injection cross
`would allow independent control of the potentials in the reaction
`chamber and separation column. To have a long reaction time
`for the chemical reaction,
`the field strength in the reaction
`chamber should be minimized, but for capillary electrophoresis,
`the resolution between compounds increases with increasing
`electric field strength. With the current design, a compromise
`must be reached with the potentials applied to the microchip so
`that the OPA has sufficient time to react with the amino acids
`without seriously impeding the quality of the separation.
`
`ACKNOWLEDGMENT
`This research was sponsored by the U.S. Department of
`Energy (DOE), Office of Research and Development. Oak Ridge
`National Laboratory is managed by Martin Marietta Energy
`Systems, Inc. for the U.S. Department of Energy under contract
`DE-AC05-840R21400. Also, this research was sponsored in part
`for S.C.J.
`to the Alexander Hollaender
`by an appointment
`Distinguished Postdoctoral Fellowship Program sponsored by the
`U.S. DOE and for A.W.M. and R.H. to the ORNL Postdoctoral
`Research Associates Program. These postdoctoral programs are
`administered by the Oak Ridge Institute for Science and Education
`and ORNL.
`
`Received for review August 1, 1994. Accepted September
`26, 1994.®
`
`Abstract published in Advance ACS Abstracts, October 15, 1994.
`
`separation field strength [kV/cm]
`Figure 10. Variation of plate height with the separation field strength
`for arginine (•) and glycine ( ) derivatized with o-phthaldialdehyde
`with Lsep =
`10 mm. The contribution to the plate height from the
`injection and detection is calculated from eqs 3 and 4 for arginine
`(a) and glycine ( ). Error bars are ±o for three runs.
`
`In Figure 10, the total plate height versus
`the electric field
`strength is plotted along with the calculated contributions from
`the injection plug length and the detector observation length. The
`time-independent contributions differ for arginine and glycine
`because the injections have an electrophoretic bias, and conse-
`quently, compounds with a greater electrophoretic mobility will
`have a larger contribution to the plate height. The injection plug
`lengths at half-height were 0.80 and 0.37 mm for arginine and
`glycine, respectively, which correspond to
`equal to 11.6 and
`2.5 /<m. These contributions are substantial, but in order to run
`the injector in the linear, reproducible regime for all separation
`field strengths, longer injection times were required. The detector
`observation length, i.e., the laser spot size, was «50 /xm, which
`In the range of separation
`corresponds to Hi# equal to 21 nm.
`field strengths used, the contribution from axial dispersion is small
`relative to the other contributions to the total plate height and is
`not discussed further. Unfortunately, as the field strength
`the total plate height increases. This behavior is
`increases,
`reminiscent of liquid chromatography, where mass transfer in the
`mobile and/or stationary phases dominates at high mobile phase
`velocities. The contribution from mass transfer to the plate height
`is a linear function of the mobile phase velocity:
`Hmt = Cu
`
`(6)
`
`where
`
`(7)
`u=
`fxE
`u is the linear velocity of the mobile phase and C is the coefficient
`of mass transfer. Typically, in capillary electrophoresis, contribu-
`tions to the plate height from mass transfer are neglected because
`there should be no interaction of the analyte with the walls of the
`capillary. However, in Figure 10 the magnitude of this contribu-
`tion is surprisingly large especially for small molecules which tend
`to have little or no interaction with the walls.
`An alternative explanation might be that the source of band
`broadening is due to inhomogeneity of the channel surface,
`
`4132 Analytical Chemistry, Vol. 66, No. 23, December 1, 1994
`
`Agilent Exhibit 1283
`Page 6 of 6
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket