throbber
Anal. Chem. 1994, 66, 1114-1118
`
`High-Speed Separations on a Microchip
`Stephen C. Jacobson, Roland Hergenroder, Lance B. Koutny, and J. Michael Ramsey*
`Chemical and Analytical Sciences Division, Oak Ridge National Laboratory,
`Oak Ridge, Tennessee 37831-6142
`
`Fast, efficient separations are sought in liquid-phase analyses
`which incorporate a nondiscriminatory sample injection scheme
`and can implement a variety of detection modes. A glass
`microchip device for free solution electrophoresis was fabricated
`using standard photolithographic procedures and chemical wet
`etching. Separations were performed at several separation
`lengths from the injector to the detector with electric field
`strengths from 0.06 to 1.5 kV/cm. For a separation length of
`0.9 mm, electrophoretic separations with baseline resolution
`are achieved in less than 150 ms with an electric field strength
`of 1.5 kV/cm and an efficiency of 1820 plates/s. For a
`separation length of 11.1 mm, a minimum plate height of 0.7
`jtm and a maximum number of plates per second of 18 600
`were achieved.
`
`Microfabricated chemical instruments show great promise
`for the laboratory and as advanced chemical sensors. Micro-
`fabrications of chemical separation techniques have received
`noticeable attention over
`the past several years and have
`included the techniques of gas chromatography,1 liquid
`3456and capillary electrophoresis.3-7 Micro-
`chromatography,1 2
`electronic devices have been able to achieve even faster response
`times in part due to miniaturization. Similar benefits may
`also accrue from miniaturization of some chemical measure-
`ment techniques. The response times of chemical measure-
`ments can often be an important issue, in particular in the
`case of chemical sensing.
`With microfabrication, the performance of many liquid
`separation techniques improves, especially capillary electro-
`phoresis.8-10 For capillary electrophoresis, smaller column
`dimensions enable the power generated to be dissipated more
`efficiently, and as a direct result, separation devices can be
`operated at higher electric field strengths. The efficiency of
`the separations, therefore, improves in two areas. First, Joule
`heating, which leads to thermal gradients within the channel
`and ultimately contributes to the dispersion of the analyte
`
`(1) Terry, S. C.; Jerman, J. H.; Angell, J. B. IEEE Trans. Electron Devices 1979,
`26, 1880.
`(2) Manz, A.; Miyahara, Y.; Miura, J.; Watanabe, Y.; Miyagi, H.; Sato, K. Sens.
`Actuators 1990, Bl, 249.
`(3) Manz, A.; Harrison, J.; Verpoorte, E. M. J.; Fettinger, J. C.; Paulus, A.; Ludi,
`H.; Widmer, H. M. J. Chromatogr. 1992, 593, 253.
`(4) Harrison, D. J.; Manz, A.; Fan, Z.; Ludi, H.; Widmer, H. M. Anal. Chem.
`1992, 64, 1926.
`(5) Seiler, K.; Harrison, D. J.; Manz, A. Anal. Chem. 1993, 65, 1481.
`(6) Harrison, D. J.; Fluri, K.; Seiler, K.; Fan, Z.; Effenhauser, C. S.; Manz, A.
`Science 1993, 261, 895.
`(7) Jacobson, S. C.; Hergenroder, R.; Koutny, L. B.; Warmack, R. J.; Ramsey,
`J. M. Anal. Chem., preceding paper in this issue.
`(8) Mikkers, F. E. P.; Everaerts, F. M.; Verheggen, Th. P. E. M. J. Chromatogr.
`1979, 169, 11.
`(9) Jorgenson, J. W.; Lukacs, K. D. Anal. Chem. 1981, 53, 1298.
`(10) Jorgenson, J. W.; Lukacs, K. D. J. Chromatogr. 1981, 218, 209.
`1114 Analytical Chemistry, Vol. 66, No. 7, April 1, 1994
`
`band, is minimized.11 The second and primary advantage is
`that higher field strengths enable separations to be performed
`faster, thereby minimizing the contribution of axial diffusion
`to the dispersion of the analyte band.9 10*To take advantage of
`these two properties, on-column optical gating12 was used to
`injection plugs for high-speed separations.
`define narrow
`However, this application is limited because the analyte must
`be tagged with a fluorophore that can be photobleached to
`implement such a scheme.
`Planar glass microchips have been constructed for elec-
`trophoretic separations using a cross-column geometry.3-6 The
`cross-column geometry is not an elaborate design, but it
`effectively illustrates the power of a microchip separation
`device. For these fast separations, we have fabricated a glass
`microchip which has a cross-column geometry and utilizes a
`nondiscriminatory injection procedure and a fluorescence
`detection scheme. To evaluate the microchip, a series of
`separations were performed at several separation lengths and
`over a range of field strengths employing a binary analyte
`mixture.
`
`EXPERIMENTAL SECTION
`The channels on the microchip substrate were fabricated
`using standard photolithographic techniques and chemical
`wet etching. The micromachined substrate and the cover plate
`were joined using a direct bonding technique as described
`previously.7 8Figure 1 shows a schematic of the column design
`implemented for the experiments. The cross-column geometry
`is the simplest structure that allows injection and separation
`to occur quickly and efficiently. The enclosed column lengths
`are 3.8 mm from the buffer reservoir to the injection cross,
`7.0 mm from the analyte reservoir to the injection cross, 9.8
`mm from the analyte waste reservoir to the injection cross,
`and 17.8 mm from the waste reservoir to the injection cross.
`The dimensions of the channel are 15 pm deep, 94 pm wide
`at the top, and 64 pm wide at the bottom. The dimensions
`were obtained using a profilometer (Alpha-Step 200, Tencor
`Instruments).
`The separations were monitored on-microchip via fluo-
`rescence using an argon ion laser (514.5 nm, Coherent Innova
`90) for excitation and a photomultiplier tube (PMT; Oriel
`77340) to collect the fluorescence signal. The PMT with
`collection optics was situated below the microchip with the
`optical axis perpendicular to the microchip surface. The laser
`was operated at ~20 mW, and the beam impinged upon the
`microchip at a 45° angle from the microchip surface and a
`45° angle from the separation channel. The laser beam and
`
`(11) Hjerten, S. Chromatogr. Rev. 1967, 9, 122.
`(12) Monnig, C. A.; Jorgenson, J. W. Anal. Chem. 1991, 63, 802.
`
`0003-2700/94/0366-1114$04.50/0
`© 1994 American Chemical Society
`
`Downloaded via UNIV OF VIRGINIA on November 6, 2018 at 17:52:07 (UTC).
`
`See https://pubs.acs.org/sharingguidelines for options on how to legitimately share published articles.
`
`Agilent Exhibit 1280
`Page 1 of 5
`
`

`

`Figure 2. CCD Image ot
`(shaded area).
`
`'sample loading" mode for rhodamine B
`
`stability of the injection plug.
`Ideally, when the analyte plug
`is injected into the separation column, only the analyte in the
`injection cross would migrate into the separation column. From
`Figure 2, the volume of the injection plug in the injection
`cross is — 120 pL with a plug length of 130 /am. A portion
`of the analyte in the analyte channel and the analyte waste
`channel is drawn into the separation column. Following the
`switch to the separation mode, the volume of the injection
`plug is —250 ± 24 pL with a plug length of 208 ± 19 pm.
`These dimensions are estimated from a series of CCD images
`taken immediately after the switch is made to the separation
`mode.
`One particular advantage to the planar microchip design
`is that with laser-induced fluorescence the point of detection
`can be placed anywhere along the separation column. The
`electropherograms are detected at separation lengths of 0.9,
`I. 6, and 11.1 mm from the injection cross.
`The 1.6- and
`II. 1 -mm separation lengths were used over a range of electric
`field strengths from 0.06 to 1.5 kV/cm, and the separations
`this range. At an electric field
`had baseline resolution over
`strength of 1.5 kV/cm,
`rhodamine B and
`the analytes,
`fluorescein, are resolved in less than 150 ms for the 0.9-mm
`separation length (Figure 3a), in less than 260 ms for the
`1.6-mm separation length (Figure 3b), and in less than 1.6 s
`for the 11.1-mm separation length (Figure 3c). Due to the
`trapezoidal geometry of the channels, the upper corners make
`it difficult to cut
`the sample plug away exactly when the
`potentials are switched from the sample loading mode to the
`separation mode. Thus, the injection plug has a slight tail
`associated with it, and we believe that this effect accounts for
`the tailing observed in some of the separated peaks.
`Applying these separations to real samples depends largely
`on the capacity of the separation and the types of samples.
`The peak capacity with a resolution of 1.0 can be calculated;1}
`I + (TV05/ 4) lnd^/V^)
`
`nc =
`
`(1)
`
`where N is the number of theoretical plates and Kma, and Fmj„
`are the largest and smallest volumes in which zones can be
`eluted and detected. An experimentally viable value for the
`ratio of Vmax to Kmjn for capillary electrophoresis is 5. For
`Figure 3a-c, the number of plates and the peak capacities for
`rhodamine B and fluorescein at the three separation lengths
`are listed in Table 1. Although the peak capacities of 5.5 for
`
`(13) Giddings, J. C. Anal. Chem. 1967, 39, 1027.
`
`Analytical Chemistry, Vol. 66, No. 7, April 1, 1994
`
`1115
`
`Figure 1. Schematic of the microchip with a cross-column geometry.
`The large circle represents the cover plate that forms the closed
`channels.
`
`PMT observation axis were separated by a 135° angle.
`Labview (National Instruments) software was used to monitor
`the ADC board, which was connected to the output of the
`PMT. A charge-coupled device (CCD; Princeton Instruments,
`Inc. TE/CCD-512TKM) camera was used to determine the
`sample plug length. The CCD camera was mounted on a
`stereomicroscope (Nikon SMZ-U), and the microchip was
`illuminated using the argon ion laser operating at 1.5 W with
`the beam expanded to ~20 mm.
`The analytes used for the
`rhodamine B and disodium fluorescein
`experiments were
`(Exciton Chemical Co., Inc.) at 20 #<M. A sodium tetraborate
`buffer (4 mM, pH 9.2) was the buffer in all experiments.
`The microchip was operated under a “pinched sample
`loading” mode and a “separation" mode as described previ-
`ously.7 During the separation mode,
`the analyte and the
`analyte waste reservoirs are maintained at 43% of the potential
`applied to the buffer reservoir.
`
`RESULTS AND DISCUSSION
`The sample loading scheme does not discriminate against
`an analyte because of its net charge as do the electromigration9
`and the photolytic12 injections used for capillary electro-
`phoresis. By loading the sample using a frontal electrophero-
`gram, the injection plug is stoichiometrically representative
`of the solution being analyzed.
`In Figure 2, a CCD image
`displays the flow pattern of the analyte (shaded area) and the
`buffer (white area) through the region of the injection cross.
`By pinching the flow of the analyte, the volume of the analyte
`time.7 The slight asymmetry of the plug
`plug is stable over
`shape is due to the different electric field strengths in the
`buffer channel (470 V/cm) and the separation channel (100
`V/cm) for 1.0 kV applied to the analyte, buffer, and waste
`reservoirs and with the analyte waste reservoir grounded.
`However, the different field strengths do not
`influence the
`
`Agilent Exhibit 1280
`Page 2 of 5
`
`

`

`Table 1. Plata Numbers, Peak Capacities, and Plata* par
`Second for the Electric Field Strength of 1.5 kV/cm*
`plate
`separation
`peak
`no., N
`length, mm
`capacity, nc
`5.5
`120
`0.9
`130
`450
`450
`15100
`11900
`
`compound
`rhodamine B
`fluorescein
`rhodamine B
`fluorescein
`rhodamine B
`fluorescein
`
`1.6
`
`11.1
`
`plates/s
`1820
`1070
`3820
`2080
`18600
`7900
`
`9.5
`
`47.7
`
`0 For electropherograms, see Figure 3.
`
`separations the total plate height can be expressed as
` ^total = H'mj + #det + ^diff
`
`(2)
`
`where Hinj, H&tt, and Hm are the contributions to the plate
`height from the injection plug, the detector path length, and
`the axial diffusion, respectively. The contributions from the
`injection plug length and the detector path length are time
`independent, and the contribution to the plate height from
`axial diffusion is time dependent. Effects such as Joule heating
`were not considered because the power dissipation was well
`below 1 W/m for all experiments.12
`The contributions to the plate height from the injection
`plug length and the detector length are15
`H-mj = jy (12L)
`#det = PdJWL)
`
`(3)
`
`(4)
`
`where /j„j and ket are the lengths of the injection plug and of
`the detector path, respectively, and L is the separation length.
`The lengths of the injection plug and the detector path are
`If
`for all experiments.
`constant
`these time-independent
`contributions predominate in their contribution to the plate
`height, then the total plate height decreases as the separation
`length increases.
`The contribution to the plate height from axial diffusion
`is16
`
`J^diff= 2Dm/u
`
`(5)
`
`where u is the linear velocity of the analyte and Dm is the
`diffusion coefficient of
`the analyte in the buffer. The
`contribution from axial diffusion to the plate height is reduced
`by large linear velocities for the analyte.
`The total plate height can be rewritten17
`= A + B/u
`
`(6)
`
`where A is the sum of the contributions from the injection
`plug length and the detector path length and B is equal to
`2Dm.
`
`(15) Sternberg, J. C. Adv. Chromatogr. 1966, 2, 205.
`(16) Giddings, J. C. Dynamics of Chromatography, Part I: Principles and Theory,
`Marcel Dekker: New York, 1965; Chapter 2.
`(17) van Deemter, J. J.; Zuiderweg, F. J.; Klinkenberg, A. Chem. Eng. Sci. 19S6,
`5, 271.
`
`0
`
`:
`
`.
`
`300
`
`400
`
`100
`
`200
`time [ms]
`.............
`
`rhodamine B
`
`fluorescein 7
`

`
`I
`
`j
`I l
`
`L.
`
`1500
`
`2000
`
`0
`
`500
`
`1000
`time [ms]
`Figure 3. Electropherograms of rhodamine B and fluorescein with a
`separation field strength of 1.5 kV/cm and a separation length of (a,
`top) 0.9, (b, middle) 1.6, and (c, bottom) 11.1 mm.
`the 0.9-mm separation length and 9.5 for the 1.6-mm
`in less
`separation length are not high, the separations occur
`than 150 and 260 ms, respectively. This is appealing when
`only a few species require analysis or when high-speed capillary
`electrophoresis is used as the second dimension to a two-
`dimensional separation.14 If the time duration of the first
`dimension is on the order of tens of seconds to minutes, having
`such a rapid second step enables information to be extracted
`without compromising the resolution obtained from the first
`step.
`In order to improve the peak capacity of the separation,
`the band dispersion needs to be minimized. For electrophoretic
`
`(14) Lemmo, A. V.; Jorgenson, J. W. Anal. Chem. 1993, 65, 1576.
`
`1116 Analytical Chemistry, Vol. 66, No. 7, April 1, 1994
`
`Agilent Exhibit 1280
`Page 3 of 5
`
`

`

`the contribution of the axial dispersion to the total plate height
`is negligible relative to the contributions from the injection
`and detector. The average values obtained from the data in
`Figure 4 (Table 2, term A) compare well with theory for the
`1.6-mm separation length, 3.46 versus 3.42 Mm, but are 50%
`higher than predicted for the 11.1-mm separation length, 0.76
`versus 0.50 Mm. How the column dynamics change down-
`stream from the 1.6-mm observation point is not understood.
`Also, the 11.1-mm separation length is 6.9 times longer than
`the 1.6-mm separation length, and the average time-inde-
`pendent contributions to the plate height should differ by 6.9
`times, but a factor of 4.6 is observed between 3.46 and 0.76
`Mm (see Table 2). Consequently, the separations at the 11.1 -
`mm separation length are not as efficient as expected. The
`efficiencies of the separations at the 0.9- and 1,6-mm separation
`lengths are limited by the lengths of the injection plug and the
`Improvements can be made on both counts by
`detector path.
`channels and by using smaller laser spot sizes.
`using narrower
`The time-dependent contribution of axial diffusion to the
`plate height is the dominant contribution at linear velocities
`below 3 mm/s (see Figure 4). From the B coefficient (see
`Table 2 and eq 6), the diffusion coefficients for each analyte
`can be calculated at
`the two separation lengths. The
`for
`experimentally determined diffusion coefficients, Dm,
`rhodamine B are 1.3 X 10-5 and 8.9 X 10-6 cm2/s for the 1.6-
`and 11.1-mm separation lengths, respectively, and for fluo-
`rescein are 1.2 X 10~5 and 7.8 X 10-6 cm2/s for the 1.6- and
`1.1-mm separation lengths, respectively. The values for each
`analyte at the two separation lengths differ by as much as
`50%, and this error can be attributed to an insufficient number
`of points in the lower liner velocity range. However, the values
`for two analytes at the same separation length compare well,
`1.3 X 10-5 cm2/s for rhodamine B and 1.2 X 10~5 cm2/s
`fluorescein at the 1.6-mm separation length and 8.9 X KH
`cm2/s for rhodamine B and 7.8 X 10-6 cm2/s for fluorescein
`at the 11.1 -mm separation length. The diffusion coefficients
`obtained at the 11.1 -mm separation length are consistent with
`diffusion coefficients for molecules of similar mass.
`An important measure of the utility of a separation system
`is the number of plates generated per unit time:18
`N/t = L/(Ht)
`
`(7)
`
`Combining eqs 6 and 7 and substituting u = pE where p is
`the effective electrophoretic mobility of the analyte and E is
`the electric field strength, the plates per unit time can be
`expressed as a function of the electric field strength:
`N/t = (pEf/iApE + B)
`
`(8)
`
`At low electric field strengths, when axial diffusion is the
`dominant form of band dispersion, the term ApE is small
`relative to B, and consequently, the number of plates per second
`increases with the square of the electric field strength. As the
`electric field strength increases, the plate height approaches
`a constant value, and the plates per unit time increase linearly
`with the electric field strength because B is small relative to
`ApE.
`
`(18) Giddings, J. C. Sep. Sci. 1969, 4, 181.
`
`Analytical Chemistry, Vol. 66, No. 7, April 1, 1994
`
`1117
`
`Figure 4. Variation of the plate height as a function of the linear
`velocity for rhodamine B at separation lengths of 1.6 (circle) and 11.1
`mm (square) and for fluorescein at separation lengths of 1.6 (diamond)
`and 11.1 mm (triangle). Lines are eq 6 fitted to the experimental data
`with the resulting coefficients listed In Table 2.
`
`Table 2. Plate Height Coefficients Determined Using Equation 6
`Fitted to the Data in Figure 4
`separation
`length, mm
`1.6
`
`A, mm
`0.003 46
`0.003 46
`0.000 716
`0.000 802
`
`B, mm2/s
`0.002 59
`0.002 47
`0.001 79
`0.001 56
`
`compound
`rhodamine B
`fluorescein
`rhodamine B
`fluorescein
`
`11.1
`
`In Figure 4, the plate height is plotted versus
`the linear
`velocity of the analyte. As expected, the plate height decreases
`with increasing linear velocity until a constant value is obtained.
`Equation 6 is fitted to the experimental data, and the resulting
`coefficients are listed in Table 2. The time-independent
`contributions to the plate height, A, reach average constant
`values of 3.46 and 0.76 *im for the 1.6- and 11.1-mm separation
`lengths, respectively. The longer separation length of 11.1
`mm is expected to have a lower plate height contribution from
`the time-independent events because the only variable is the
`separation length (see eqs 3 and 4). Also, no adsorption of
`the analytes to the column wall
`is observed. Otherwise, a
`linear increase in the plate height with increasing linear
`velocities would be seen due to a resistance to mass transfer.
`For these experiments the injection plug length, /inj, was
`208 Mm and the detector path length, i.e., laser spot size, kel,
`was 150 Mm. The contributions to the plate height for the
`injection plug length calculated using eq 3 are 4.81 Mm for the
`0.9-mm separation length, 2.25 Mm for the 1,6-mm separation
`length, and 0.33 Mm for the 11.1-mm separation length. The
`contributions to the plate height for the detector path length
`calculated using eq 4 are 2.50 Mm for the 0.9-mm separation
`length, 1.17 Mm for the 1.6-mm separation length, and 0.17
`pm for the 11.1-mm separation length. The sums of these
`time-independent contributions are 7.31, 3.42, and 0.50 Mm
`for the 0.9-, 1.6-, and 11.1 -mm separation lengths, respectively.
`The average plate height for the separation at the 0.9-mm
`separation length in Figure 3a is 7.25 Mm, which is comparable
`to the calculated value for the time-independent contributions
`of 7.31 Mm. For analyte linear velocities greater than 3 mm/
`the time-independent contributions to the plate height
`s,
`predominate as expected from eq 6 and seen in Figure 4; i.e.,
`
`Agilent Exhibit 1280
`Page 4 of 5
`
`

`

`format of plate height versus linear velocity, small deviations
`between the experimental and fitted data are exaggerated
`when presented in the format of plates per unit time versus
`electric field strength.
`
`CONCLUSIONS
`This work demonstrates advantages of microfabricating
`capillary electrophoresis devices. Micromachining allows four
`channels to be joined at an intersection with essentially no
`dead volume. The spatial extent of the injection plug is limited
`to the intersection volume by electroosmotic flow in all four
`channels. These well-defined injections have permitted
`extremely high speed separations with the greatest number of
`plates per second yet reported by capillary electrophoresis.19
`This injection scheme, and thus high-speed separation, is not
`limited to fluorescent species; any type of sample amenable
`In addition,
`to electrophoretic separation can be analyzed.
`this injection approach does not introduce mobility-based bias
`as with the optically gated approach. The results presented
`indicate that microchip capillary electrophoresis devices can
`provide very fast response times for sensor applications.
`In
`addition, microfabricated devices such as demonstrated here
`incorporated into two-dimensional separation strategies would
`appear attractive.
`
`ACKNOWLEDGMENT
`This research was sponsored by the U.S. Department of
`Energy (DOE), Office of Research and Development. Oak
`Ridge National Laboratory is managed by Martin Marietta
`Energy Systems, Inc., for the U.S. Department of Energy
`under Contract DE-AC05-840R21400. Also, this research
`was sponsored in part by an appointment for S.C.J. to the
`Alexander Hollaender Distinguished Postdoctoral Fellowship
`Program sponsored by the U.S. DOE, for L.B.K. to the DOE
`Laboratory Cooperative Postgraduate Research Training
`Program, and for R.H. to the ORNL postdoctoral research
`associate program. These programs are administered by the
`Oak Ridge Institute for Science and Education and ORNL.
`The authors acknowledge many useful discussions with Drs.
`Roswitha Ramsey and Georges Guiochon.
`
`Received for
`1994.®
`
`review October 5, 1993. Accepted January 10,
`
`•Abstract published in Advance ACS Abstracts, February 15, 1994.
`
`electric field strength [kV/cm]
`Figure 5. Variation of the number of plates per unit time as a function
`of the electric field strength for rhodamine B at separation lengths of
`1.6 (circle) and 11.1 mm (square) and for fluorescein at separation
`lengths of 1.6 (diamond) and 11.1 mm (triangle). Lines are calculated
`from eq 8 using the coefficients listed in Table 2.
`
`The values of the plates per second are approximately twice
`as high for rhodamine B than for fluorescein. For the electric
`field strength of 1.5 kV/cm (see Figure 3 for electrophero-
`grams), the number of plates per second for rhodamine B and
`for fluorescein is listed in Table 1. The number of plates for
`the two analytes is nearly the same, but the retention time of
`the rhodamine B is approximately half that of the fluorescein.
`In Figure 5, the number of plates per second for the 1.6- and
`11.1 -mm separation lengths is plotted versus the electric field
`strength, and the number of plates per second quickly becomes
`a linear function of the electric field strength, because the
`plate height approaches a constant value (see Figure 4). The
`symbols in Figure S represent the experimental data collected
`for the two analytes at the 1.6- and 11.1-mm separation lengths.
`The lines are calculated using eq 8 and the coefficients from
`Table 2. A slight deviation is seen between the experimental
`data and the calculated numbers for rhodamine B at the 11.1 -
`mm separation length. This is primarily due to experimental
`error. Also, when eq 6 is fitted to the data in Figure 4 in the
`
`(19) Since submission of the manuscript, the following article appeared, producing
`a greater number of plates per second using the optical gating injection
`technique: Moore, A. W.; Jorgenson, J. W. Anal. Chem. 1993, 65, 3550.
`
`1118 Analytical Chemistry, Vol. 66, No. 7, April 1, 1994
`
`Agilent Exhibit 1280
`Page 5 of 5
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket