throbber
Anal. Chem. 1995, 67, 4184-4189
`
`Microchip Separations of Neutral Species via
`Micellar Electrokinetic Capillary Chromatography
`Alvin W. Moore, Jr., Stephen C. Jacobson, and J. Michael Ramsey*
`Chemical and Analytical Sciences Division, Oak Ridge National Laboratory, P. 0. Box 2008, Building 4500S, MS 6142,
`Oak Ridge, Tennessee 37831-6142
`
`Micellar electrokinetic capillary chromatography (MECC)
`of three neutral coumarin dyes was performed on glass
`microchips. Manifolds of channels for analyte injection
`and separation were machined into one surface of the
`glass substrates using standard photolithographic, etch-
`ing, and deposition techniques. Coverplates were then
`these channels to form capillaty
`directly bonded over
`networks, with fluid flow in these networks controlled by
`varying the applied high-voltage potentials at the outlets.
`The separation capillaty was 16.5 cm long for a serpentine
`channel chip and 1.3 cm long for a straight channel chip.
`Detection of analyte zones was accomplished by laser-
`induced fluorescence using the UV lines (~350 nm) of
`an argon ion laser. At low applied electric field strengths,
`MECC analyses with on-chip injections gave high repro-
`ducibilities in peak areas and migration times (<1% for
`two of the three coumarins) and near constant separation
`efficiencies throughout the analysis. At high fields (> 400
`V/cm), analysis times were shorter, but separation ef-
`ficiency decreased at later migration times. These peaks
`showed significant broadening, consistent with mass
`transfer effects.
`
`Many of today’s applications of analytical chemistry involve
`collection of samples from some remote site, transportation of
`these samples back to the laboratory, and analysis using benchtop
`instrumentation. These situations range from detection of pol-
`lutants in water supplies to process monitoring for quality control.
`In the typical analytical laboratory, a universal detection system
`such as a UV absorption detector and a separation technique such
`as liquid chromatography (LC) or capillary electrophoresis (CE)
`are used to quantify the analyte of interest. The separation
`method provides the selectivity between components and classes
`If the analysts’
`of components that the universal detector lacks.
`needs change, such that they must quantify other species, the
`selectivity can be easily modified by changing the separation
`conditions such as the pH of the mobile phase or buffer, the
`organic solvent content, or the elution gradient. The enhanced
`versatility of this method comes at the expense of requiring that
`samples be collected and transported to the lab.
`An attractive means of saving time and expense would be to
`take the instrument to the sample. Micromachining technology
`may allow the ultimate miniaturization of chemical measurement
`instrumentation. Liquid phase separation devices are particularly
`amenable to miniaturization because analytical separation perfor-
`mance often improves with decreasing size of the components
`involved. CE is a relatively new separation method which has
`4184 Analytical Chemistry, Vol. 67, No. 22, November 15, 1995
`
`used microcolumn capillaries from its inception. Because CE
`gives high-efficiency separations but requires no high-pressure
`pump or gas supply, it is especially well suited for miniaturization.
`Substantial progress has already been made in this area.1 23™4
`A problem with CE for general analysis is its inability to
`separate uncharged species. All neutral species in a particular
`sample will have zero electrophoretic mobility and thus the same
`migration time. Previously, our group has investigated open-
`channel electrochromatography (OCEC) on a microchip as a
`means of separating neutrals.5 In this previous work, sample
`components were separated by their partitioning interaction with
`a stationary phase coated on the channel walls. The mobile phase
`was driven not by a conventional pump but by electroosmotic flow.
`Micellar electrokinetic capillary chromatography (MECC) is a
`operational mode of CE which was developed by Terabe et al.6 7to
`address the separation of neutrals by CE. A surfactant such as
`is added to the CE buffer in
`sodium dodecyl sulfate (SDS)
`sufficient concentration to form micelles in the buffer.
`In a typical
`experimental arrangement, the micelles move much more slowly
`toward the cathode than does the surrounding buffer solution.
`The partitioning of solutes between the micelles and the sur-
`rounding buffer solution provides a separation mechanism similar
`to that of LC.
`In this work, MECC was performed in a capillary etched into
`the surface of a glass chip. All of the fluidic manipulation
`necessary for sample injection and analysis was done through a
`manifold of channels, with flow controlled by the voltage applied
`the end of each channel. Laser-induced
`to the reservoir at
`fluorescence detection was used to monitor the separations. Some
`results of this implementation of MECC will be shown, and some
`of its advantages and disadvantages for chip-based microinstru-
`mentation will be discussed.
`
`OVERVIEW
`An in-depth treatment of the theory of MECC has been given
`by others6™8 and is beyond the scope of this paper. Only a brief
`
`(1) Harrison, D. J.; Manz, A; Fan, Z.; Liidi, H.; Widmer,  . M. Anal. Chem.
`1992, 64, 1926-1932.
`(2) Seiler, K; Harrison, D. J.; Manz, A Anal. Chem. 1993, 65, 1481-1488.
`(3) Jacobson, S. C.; Hergenrtider, R.: Koutny, L. B.; Warmack, R J.; Ramsey, J.
`M. Anal. Chem. 1994, 66, 1107-1113.
`(4) Jacobson, S. C.; Hergenrtider, R; Koutny, L. B.; Ramsey, J. M. Anal. Chem.
`1994, 66, 1114-1118.
`(5) Jacobson, S. C.; Hergenrtider, R; Koutny, L. B.; Ramsey, J. M. Anal. Chem.
`1994, 66, 2369-2373.
`(6) Terabe, S.; Otsuka, K; Ando, T. Anal. Chem. 1985, 57, 834-841.
`(7) Sepaniak, M. J.; Powell, A C.; Swaile, D. F.; Cole, R 0. In Capillary
`Electrophoresis: Theory and Practice·, Grossman, P. D„ Colburn, J. C., Eds.;
`Academic Press, Inc.: San Diego, CA 1992; pp 159—89.
`(8) Sepaniak, M. J.; Cole, R. 0. Anal. Chem. 1987, 59, 472-476.
`
`0003-2700/95/0367-4184$9.00/0 &copy; 1995 American Chemical Society
`
`Downloaded via UNIV OF VIRGINIA on November 6, 2018 at 18:16:15 (UTC).
`
`See https://pubs.acs.org/sharingguidelines for options on how to legitimately share published articles.
`
`Agilent Exhibit 1277
`Page 1 of 6
`
`

`

`review will be given here. For neutral solutes, the separation
`mechanism is essentially chromatographic and can therefore be
`described with modified chromatographic relationships. The
`capacity factor, k', for a given solute and experimental conditions
`is the ratio of the total moles of solute in the stationary phase
`(here, the pseudostationary phase of the micelles) to those in the
`is modified to account for the
`mobile phase. For MECC, k'
`movement of the micelles,
`
`k' =
`
`^o(l ""
`
`*rAji)
`
`= TO
`
`(i)
`
`where ír is the solute retention time, k is the “void time” (retention
`time for a solute moving at the rate of the electroosmotic flow),
`tm is the micelle retention time (i.e„ retention time of a completely
`retained solute), Kis the partition coefficient, 7S is the volume of
`the micellar phase, and 7m is the volume of the mobile phase. If
`the micelles were indeed stationary, tm would become infinite and
`the equation would reduce to its conventional form for chroma-
`tography.
`Because to is the retention time for a solute moving with the
`electroosmotic flow, and tm is that of a solute completely retained
`in the micelles, neutral solutes must elute between to and tm.
`Resolution can be improved by increasing the difference between
`to and tm, thus increasing the “window” of time over which sample
`components can elute. Resolution can also be improved by
`changing the k' values of the solutes, which affects both the
`retention and the selectivity of the separation. This can be done
`in many of the same ways it is done in liquid chromatography,
`such as by adjusting the temperature, buffer concentration, or pH,
`or by the addition of organic modifiers. Organic solvents such
`as methanol and acetonitrile can have multiple effects in that they
`may modify the electroosmotic flow in the system or alter the
`hydrophobic interactions between solute and micelle, but they
`also affect micelle structural interactions and so alter the partition-
`ing kinetics. Changes in the organic solvent content of the buffer
`also modify the electroosmotic flow in the system.
`
`EXPERIMENTAL SECTION
`The microchips were constructed using standard photolitho-
`graphic, wet chemical etching and bonding techniques described
`previously.9 An ordinary soda lime glass microscope slide was
`used as a substrate, into which was etched a network of open
`channels (see Figure 1). The serpentine channel geometry
`(Figure la) allowed use of a longer separation capillary within a
`small area, while the straight channel geometry (Figure lb)
`allowed experiments with high applied electric field strengths.
`closed with thin coverplates,
`These channel networks were
`to form capillary networks.
`directly bonded to the substrate,
`Cylindrical glass reservoirs were then bonded with epoxy to the
`capillary outlets. The lengths of the capillary channels may vary
`with the placement of the coverplate. Figure 1 gives the channel
`lengths for the two chips used in this work. Use of the wet
`chemical etch on a glass substrate results in an isotropic etch.
`That is, the glass etches at the same rate in all directions, and
`the resulting channel geometry is trapezoidal.5 The channel cross
`the same for both chips used in this
`section dimensions were
`
`(9) Jacobson, S. C.; Koutny, L. B.; Hergenroder, R.; Moore, A W„ Jr.; Ramsey,
`J. M. Anal. Chem. 1994, 66, 3472-3476.
`
`Figure 1. Schematic diagrams of microchip geometry. The large
`outer boxes show the outlines of the glass substrates.
`Individual
`capillary lengths are measured from the point of Injection to the point
`where the capillary passes beyond the cover plate Into the respective
`reservoir, (a) The serpentine channel geometry used for the majority
`lengths as follows: analyte channel, 9.2
`of the work, with channel
`mm; buffer channel, 6.6 mm; analyte waste channel, 7.7 mm; waste
`(b) The straight channel geometry
`(separation) channel, 171 mm.
`used for the high-field experiments, with channel lengths as follows:
`analyte channel, 8.0 mm; buffer channel, 4.6 mm; analyte waste
`channel, 9.8 mm; waste (separation) channel, 17.2 mm.
`
`work. The channels were 10 µ   deep, 60 µ   wide at the bottom,
`and 80 µ   wide at the top. These dimensions were measured
`with a profilometer (Alpha-Step 200, Tencor Instruments, Moun-
`tain View, CA) after bonding the coverplate but before attaching
`the reservoirs.
`Separation of the analyte zones was monitored via laser-induced
`fluorescence using an argon ion laser (351.1-363.8 nm, all lines;
`~50 mW; Coherent Innova 90, 5 W; Palo Alto, CA) for excitation.
`The UV beam from the laser was prefiltered with a Coming 7-37
`colored glass filter (blue, narrow bandpass; transmittance, 30% at
`360 nm) to reduce plasma emissions and then focused at the
`desired point on the capillary with a planoconvex fused silica lens
`(focal length, 100 mm; Newport Corp., Irvine, CA). The laser
`impinged on the chip at an angle of 45° with the surface. The
`fluorescence signal was collected from below the chip by a 21 x
`microscope objective (Bausch & Lomb Opt. Co., Rochester, NY),
`filtered with a Coming 3-73 colored glass filter (yellow, sharp cut-
`on at 426 nm), and detected with a photomultiplier tube (PMT;
`Oriel 77340, Stratford, CT). The PMT current was amplified and
`converted to a proportional voltage with a Keithley 617 electrom-
`eter (Keithley Instruments Inc., Cleveland, OH). The analog
`voltage output of the electrometer was measured with a multi-
`function interface board (AT-MIO-16X, National
`Instruments,
`Austin, TX) controlled by software written in-house using Lab-
`VIEW 3.0 for Windows (National Instruments, Austin, TX) on a
`PC compatible computer. Separation efficiencies were obtained
`by calculation of peak statistical moments, also in LabVIEW.
`Platinum electrodes in each reservoir provided electrical
`contact between the buffer solutions and the CE high-voltage
`power supply (CZE1000R, Spellman Inc., Plamview, NY). A
`
`Analytical Chemistry, Vol. 67, No. 22, November 15, 1995
`
`4185
`
`Agilent Exhibit 1277
`Page 2 of 6
`
`

`

`voltage divider/relay apparatus described earlier10 was used to
`set the relative voltages applied to each reservoir and to switch
`between run and inject modes under computer control. For both
`chip geometries, the separation voltage was applied over the entire
`length of the capillary channel for all analyses. However, for
`individual groups of analyses, the length of the separation channel
`used was set by the location of the point of detection.
`The reservoirs at the end of each capillary channel had a
`volume of only 200 µ . To avoid problems with changes in buffer
`concentration due to evaporation, the reservoirs were sealed with
`thin rubber septa held in place with parafilm. The platinum
`electrodes could be inserted through the septa if the septa was
`initially pierced with a syringe needle. The reservoirs could be
`flushed and filled a in similar manner with a syringe of buffer or
`In use, the seal formed around the electrodes was almost
`sample.
`gas-tight. The resolution between peaks could be maintained from
`run to run for several hours before significant evaporation effects
`were noticed. Only the data in Figure 3, as mentioned below,
`were acquired with open buffer reservoirs.
`The analytes used in these experiments were the neutral dyes
`coumarin 440 (C440), coumarin 450 (C450), and coumarin 460
`Individual stock solutions of each dye were
`(C460, Exciten Inc.).
`prepared in methanol and then diluted in the analysis buffer before
`use. The concentration of each dye was ~50 µ  unless indicated
`otherwise. The substrate is glass rather than quartz, so there is
`substantial background fluorescence where the UV laser strikes
`the chip. No attempt was made to measure limits of detection
`with the present system. The MECC buffer was composed of 10
`mM sodium borate (pH 9.1), 50 mM SDS, and 10% (v/v) methanol.
`The methanol aids in solubilizing the coumarin dyes in the
`aqueous buffer system and also affects the partitioning of some
`of the dyes into the micelles. Due care must be used in working
`with coumarin dyes as the chemical, physical, and toxicological
`properties of these dyes have not been fully investigated.* 11
`Sample Injection. The microchips were operated in a
`“pinched injection” mode described previously.3 The voltages
`applied to the reservoirs are set to either an “inject” (sample
`In the inject mode,
`loading) or a “run” (separation) configuration.
`a frontal chromatogram of the solution in the analyte reservoir is
`pumped electroosmotically through the intersection and into the
`analyte waste reservoir. Voltages applied to the buffer and waste
`reservoirs also cause weak flows into the intersection from the
`sides and then into the analyte waste reservoir. These flows serve
`to confine the stream from the analyte reservoir to give a well-
`defined plug of sample in the intersection. The chip remains in
`this mode until the slowest moving component of the sample has
`passed through the intersection. At this point, the sample plug
`in the intersection is representative of the analyte solution, with
`no electrokinetic bias.
`An injection is made by switching the chip to the run mode,
`which changes the voltages applied to the reservoirs such that
`buffer now flows from the buffer reservoir through the intersec-
`tion, into the separation column, and eventually into the waste
`reservoir. The plug of sample which was previously in the
`intersection is swept onto the separation column. Proportionately
`lower voltages are applied to the analyte and analyte waste
`reservoirs to cause a weak flow of buffer from the buffer reservoir
`
`(10) Jacobson, S. C.; Hergenroder, R; Moore, A. W„ Jr.; Ramsey, J. M. Anal.
`Chem. 1994, 66, 4127-4132.
`(11) Exciton, Inc., Dayton, OH, 1991.
`
`4186 Analytical Chemistry, Vol. 67, No. 22, November 15, 1995
`
`Figure 2. Microchip MECC analysis of a mixture of coumarin dyes.
`The small peak at 4.6 min is an unidentified impurity in one of the
`dyes. Concentration of each dye was ~50 µ . Other analysis
`conditions are given in the text.
`
`into these channels. These flows ensure that the sample plug is
`cleanly “broken off” from the analyte stream and that no excess
`analyte leaks into the separation channel during the analysis. This
`sample loading method is time-independent (after the initial time
`necessary to pump all components through the intersection),
`nonbiased, and reproducible.
`
`RESULTS AND DISCUSSION
`The results of an MECC analysis of a mixture of C440, C450,
`and C460 are shown in Figure 2. The peaks were identified by
`individual analyses of each dye. The migration time stability of
`the first peak, C440, with changing methanol concentration (see
`below) was a strong indicator that this dye did not partition into
`the micelles to a significant extent. Therefore, it was considered
`an electroosmotic flow marker with migration time to. The last
`peak, C460, was assumed to be a marker for the micellar migration
`time, tm. Using these values of to and tm from the data in Figure
`is 0.43. This agrees well
`2, the calculated elution range, to/tm,
`with a literature value of to/tm = 0.4 for a similar buffer system8
`and supports our assumption. This analysis was done in the
`serpentine chip with a separation length of 21.3 mm and an applied
`electric field strength of 47 V/cm.
`Buffer Evaporation. Figure 3 shows several successive
`analyses of the same three component sample over a period of
`30 mb. These analyses were done with an applied field of 96
`V/cm and a separation length of 17 mm. Notice that from parts
`a to c of Figure 3, the C440 peak and the C450/C460 peak pair
`tend to move closer together (the elution range decreases), and
`the resolution between the C450 and C460 peaks gradually
`decreases. The high voltage was switched off for 5 min between
`parts b and c of Figure 3, so that if gradual heating of the MECC
`buffer m the chip were responsible for the lost resolution, it should
`have been partially restored in Figure 3c. Obviously, this was
`Immediately before the analysis shown in Figure
`not the case.
`3d, the buffer and sample solutions were replaced. Notice that
`the resolution between the last two peaks has been restored,
`indicating some change in the buffer solutions during the course
`of the analyses shown in Figure 3a-c.
`For the data in Figure 3, the reservoirs at the end of each
`capillary channel were open to the surrounding air. When the
`
`Agilent Exhibit 1277
`Page 3 of 6
`
`

`

`Table 1. Peak Parameters*
`
`migration
`time (s)
`
`209.1
`1.94
`0.93
`
`384.6
`3.27
`0.85
`
`438.1
`5.76
`1.31
`
`peak
`area
`
`C440
`2.82
`0.03
`0.91
`C450
`4.92
`0.05
`0.92
`C460
`7.62
`0.17
`2.21
`
`theoretical
`plates (A)
`
`HETP
`(µ  )
`
`3523
`60
`1.71
`
`3855
`80
`2.09
`
`3705
`101
`2.73
`
`6.05
`0.10
`1.71
`
`5.53
`0.12
`2.09
`
`5.75
`0.16
`2.78
`
`av (  = 5)
`SD
`%RSD
`
`av (  = 5)
`SD
`%RSD
`
`av {  = 5)
`SD
`%RSD
`
`“ Data for five replicate analyses with applied electric field of 47
`V/cm, separation length of 21.3 mm.
`
`Table 2. Ratio Peak Parameters*
`peak area ratio
`migration time ratio
`C460/
`C440/
`C440/
`C460/
`C450
`C450
`C450
`C450
`
`capacity
`factor (kr),
`C45Ó
`
`av (  = 5)
`SD
`%RSD
`
`0.544
`0.01
`1.57
`
`1.14
`0.01
`0.54
`
`0.574
`0.01
`1.11
`
`1.55
`0.02
`1.51
`
`6.89
`0.05
`0.78
`
`“ Ratio values and capacity factor for C450 calculated from the data
`used in Table 1.
`
`generally even higher. Here, for the C440 and C450 peaks, peak
`the %RSD values for
`area %RSD values are <1%. For C460,
`migration times and peak areas are 1.3 and 2.2%, respectively. Such
`low values are presumably due to the on-chip injection scheme;
`all fluid handling for sample injection and analysis is done on-
`chip with no moving parts (other than the buffer itself). Thus,
`of variance associated with benchtop
`many of
`the sources
`instruments, such as those resulting from moving the separation
`capillary between vials, are minimized.
`To better correct for environmental variances, peak migration
`time and peak area ratios may be calculated relative to an internal
`standard peak. These ratios help to account for differences in
`flow rate and injection volume from run to run. The %RSD for
`ratio data is usually <2%.7 In Table 2, we have calculated ratio
`values for C440 and C460 considering C450 an internal standard,
`based on the same raw data used to construct Table 1. As
`expected, our %RSD values for these ratios are also <2%.
`Alternately, some have calculated capacity factors, k', for the
`analytes of interest. Because the partitioning into the micelles is
`similar to the partitioning found in liquid chromatography, the k'
`values are distinctive for a given compound under given experi-
`mental conditions.
`If we consider the migration time of C440 to be that of the
`electroosmotic flow, and that of C460 to be that of the micelles,
`then a k' value for C450 can be calculated for each analysis. The
`%RSD for the k' value of C450 is also below 1%, which compares
`well to a value of 0.5% found in the literature.13
`For a given applied electric field (47 V/cm, as above), and
`consequently, a constant plate height, the number of theoretical
`plates should increase as the length of the separation capillary
`injection to point of detection)
`(the distance from point of
`increases. Figure 4 is a plot of efficiency data gathered from
`
`Analytical Chemistry, Vol. 67, No. 22, November 15, 1995
`
`4.187
`
`Figure 3. Effects of methanol evaporation in microchip MECC. The
`times indicated in the figure are times between the ends of the
`analyses, immediately after acquisition when the data were saved.
`Peaks are identified as in Figure 2.
`
`buffers were entirely aqueous, buffer evaporation was not a
`significant problem. However, the MECC buffer contained 10%
`(v/v) methanol, so methanol evaporation might be significant. A
`decrease in the methanol concentration would cause the C450 to
`prefer the free buffer solution less and partition to a greater extent
`into the micelles. That is, the partition coefficient of the C450
`would increase, and the selectivity between C450 and C460 would
`decrease, lowering the resolution between them. Also, addition
`of methanol is known to extend the elution range in MECC,7 so
`the decrease seen in Figure 3a-c agrees well with a decrease in
`methanol concentration due to evaporation. Weinberger and
`Lurie12 have suggested that similar evaporation problems with
`MECC buffers containing organic modifiers contribute to poor
`run-to-run reproduciblity.
`Reproducibility, The results for five replicate analyses of the
`three coumarin dyes mix as in Figure 2 are shown in Table 1.
`The migration times, peak areas, and peak variances were
`obtained from the best-fit values for Gaussian curves fitted to the
`experimental data. The percent
`relative standard deviation
`(%RSD) of the migration times and peak areas are slightly lower
`than expected for MECC. Some authors have reported %RSD of
`less,712 with some as high as 10% over
`migration times of 2% or
`the course of a day,13 while %RSD values for peak area are
`
`(12) Weinberger, R.: Lurie, I. Anal. Chern. 1991, 63, 823—827.
`(13) Northrop, D. M.; Matire, D. E.; MacCrehan, W. A Anal. Chem. 1991, 63,
`1038-1042.
`
`Agilent Exhibit 1277
`Page 4 of 6
`
`

`

`14-
`
`Capillary Length (mm)
`Figure 4. Plot of peak efficiency versus
`length of separation
`capillary. Each point represents the average of three analyses, with
`error bars above and below the average at 1 SD. See text for further
`description.
`
`triplicate analyses at four different capillary lengths. The dotted
`lines are the best-fit lines for each of the components (C440, C450,
`and C460). As expected, at each capillary length, the measured
`peak efficiencies for each of the three compounds are similar. The
`linear increase in efficiency as the separation length increases
`demonstrates the homogeneity of the capillary walls over
`the
`length of the separation.
`MECC at High Applied Field Strengths. The straight
`channel chip (Figure lb) was used to study the effects of applied
`electric field strength on the MECC separations. Because the
`separation capillary in the straight chip is much shorter, signifi-
`cantly higher applied fields can be obtained for the same total
`In our present experimental arrangement,
`applied voltages.
`applied voltages of >4 kV begin to give arcing between the buffer
`reservoirs, so studies of higher applied field strengths are best
`done with shorter separation capillaries.
`A single analyte, C450, was used for the range of applied fields
`studied. C450 was chosen because it is partially retained by the
`micelles, and thus its behavior should be indicative of free solution
`and micellar effects.
`In Figure 5a, average analyte velocity is
`plotted versus applied electric field strength, with a best-fit line
`through the four points at the lowest field strengths. Theoreti-
`cally, the analyte velocity should increase linearly with applied
`field. The curve upward seen in the actual data indicates Joule
`heating effects at the higher applied fields. This is not unexpected,
`considering the high ionic strength of the MECC buffer. Analyses
`also done at 1340 V/cm, but the C450 migration time
`were
`continually drifted to earlier times. The precision is greatest for
`the points at low field. This is due both to the lack of heating
`its point of
`thermal
`effects, so that
`the capillary is nearer
`equilibrium, and to the longer migration times, any variations in
`which will be a smaller percentage of the average value.
`In Figure 5b, the average plate height for these analyses is
`plotted against the average analyte velocity. Error bars are shown
`at the 1 standard deviation (SD) points for both dimensions
`because both are calculated from the experimental data. The data
`give a Van Deemter-like plot, similar to that found in LC. The
`plate height at 0.9 mm/s is high relative to the trend of the other
`points and is probably the result of a systematic error
`in the
`analyses at that point. With this point omitted, a curve fit of the
`Van Deemter equation to the experimental data gave an Is- value
`of 0.9990.
`The general form of the Van Deemter equation is
`
`4188 Analytical Chemistry, Vol. 67, No. 22, November 15, 1995
`
`Figure 5.
`(a) Plot of analyte velocity versus applied electric field
`strength. Each point represents the average of three analyses, with
`error bars above and below the average at 1 SD. The fitted data are
`calculated from a linear regression fit of the four data points at the
`lowest applied fields. See text for further description, (b) Plot of plate
`height versus analyte velocity. Each point represents the average of
`three analyses, with error bars In both dimensions at plus and minus
`1 SD. The point at 0.9 mm/s Is omitted from the data fit to the Van
`Deemter equation. The fit equation is of the form  
`= A + B/u + Cu,
`where HK = A, H\ = B/u, and (Hmc + Hep) = Cu, to give hh0, = Hec
`+ H\ + Hmc + Hep. From the curve fit, A = 2.32 µ \, B = 0.141 («m2V
`s, and C = 0.651 s. For this fit,  2 = 0.0063 and r2 = 0.9990. See
`text for further description.
`H = A + B/u + Cu
`
`(2)
`
`where H is the theoretical plate height, u is the velocity of the
`mobile phase, A represents band broadening effects independent
`of u (such as eddy diffusion in LC), B represents band broadening
`effects inversely dependent on u (such as longitudinal diffusion),
`and C represents band broadening effects directly dependent on
`u (such as the mass transfer term in LC).
`For MECC, the major causes of on-column band broadening
`have been studied by Terabe et al.14 and Sepaniak and Cole.8 The
`mobile phase velocity represented as u above is given by the
`electroosmotic flow velocity veo in MECC. The A term can include
`all constant additions to the plate height, such as extracolumn
`effects of the injector and detector, because with free solution CE
`or MECC, there is no eddy diffusion term. This is written as HK
`and is independent of the electroosmotic flow velocity. The B
`term is a longitudinal diffusion term as in LC which is inversely
`proportional to the electroosmotic flow velocity, written as H\. The
`C term for MECC includes two terms, Hmc and Hev. The sorption/
`desorption kinetics term, Hmc, is the plate height due to the rate
`at which solutes move into and out of the micelles.
`In most cases,
`
`(14) Terabe, S.; Otsuka, K; Ando, T. Anal. Chem. 1989, 61, 251-260.
`
`Agilent Exhibit 1277
`Page 5 of 6
`
`

`

`microchip which employed a serpentine column geometry.5 For
`the same three dyes were analyzed here on a
`comparison,
`microchip of the serpentine configuration as well as on a straight
`channel chip. As would be expected, the elution order was the
`same as that found with OCEC.
`In Figure 2, the three dyes are
`baseline resolved in 8 min. The longer analysis time results from
`the lower applied field strength for the long serpentine channel.
`In Figure 6, with the straight channel and resultant higher applied
`fields, the analysis time is shorter than for OCEC. The MECC
`resolution is better than that of the OCEC, and the peak shape is
`significantly improved. At low applied fields,
`the separation
`the course of the analysis is relatively constant
`efficiency over
`for MECC. As mentioned above, at high fields, peaks for later-
`eluting sample components show more broadening due to the
`higher capacity factors of those components. Even with these
`high-field effects, however, the efficiency of the later peaks does
`not decrease as rapidly as in the OCEC analysis.
`The MECC has the advantage that
`the stationary phase
`(pseudostationary phase) is continuously replaced as fresh buffer
`is brought into the chip.
`In addition, the capacity factor of solutes
`can be varied by changing the micelle concentration.7 This
`In OCEC, the
`effectively changes the “stationary phase loading”.
`stationary phase is bonded to the capillary surface before use, so
`the stationary phase loading is fixed. Also, as with other bonded
`LC columns, the performance of the bonded stationary phase in
`time with repeated use.
`the microchip degrades over
`As a microchip technique, MECC gives better than average
`reproducibility for neutral species and is relatively simple to
`implement. The same chip used for CE can be used for MECC
`with only a change in buffer. Analysis times can be decreased
`by using high applied fields, while further addition of organic
`modifiers such as methanol or acetonitrile would increase the rate
`of solute exchange kinetics and so sharpen later peaks. Use of
`higher micelle concentrations would limit the contribution of
`micelle polydispersity and thus would also help sharpen later
`peaks. Higher micelle concentrations would also give higher
`electrophoretic currents, and thus more problems with Joule
`heating of the buffer, but these could be circumvented by the
`use of narrower
`channels in the chip. Finally, evaporation
`problems would be eliminated in an actual portable instrument
`by using sealed reservoirs.
`
`ACKNOWLEDGMENT
`This research was sponsored by U.S. Department of Energy
`(DOE), Office of Research and Development Oak Ridge National
`Laboratory is managed by Lockheed-Martin Energy Systems, Inc.,
`for the U.S. Department of Energy under Contract DE-AC05-
`840R21400. Also, this research was sponsored in part by an
`appointment for AW.M. to the ORNL Postdoctoral Research
`Associates Program. These postdoctoral programs are adminis-
`tered by the Oak Ridge Institute for Science and Education and
`ORNL.
`
`Received for review June 21, 1995. Accepted August 18,
`1995.®
`AC950629Y
`
`® Abstract published in Advance ACS Abstracts, October 1, 1995.
`
`Analytical Chemistry, Vol. 67, No. 22, November 15, 1995
`
`4189
`
`Figure 6. High-field MECC analysis of a mixture of coumarin dyes.
`Analysis conditions are given in the text. As described in the text,
`the analysis time is much shorter, but the separation efficiency is no
`longer constant over
`the course of the analysis. The number of
`theoretical plates calculated for each peak was as follows: C440, N
`= 8800; C450, N = 3900; C460, N = 3200.
`
`fíne is small, but it can be significant if the sorption/desorption
`kinetics are very slow, such as with ionic interactions between
`the solute and the micelle. The electrophoretic dispersion term,
`fíep, is the plate height due to the variance in micelle size (micelle
`polydispersity). The micelle solution gives a distribution of micelle
`sizes with a standard deviation of as much as 20% from the
`average. Both fímc and fíep increase with increasing veo. Thus
`fíec + fí, + fímc + fíep
`
`fíTot =
`
`(3)
`
`where HTot is the total plate height.
`From the fit equation in Figure 5b, fíec is the major fraction of
`the total plate height This indicates a large constant contribution
`to the plate height, such as from extracolumn effects. A detailed
`examination of these effects was

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket