`Up to Speed on PCR
`
`
`Real-time systems enable
`fast, quantitative analyses
`By Desoran A. FrrzceERALD
`
`
`
`
`revolutionary impact—
`its
`forget
`
`enables the specific amplification and
`
`detection ofaslittle as a single copy of a
`
`particular nucleotide sequence. However,
`
`PCR has the potential to be used not just
`
`for the detection of specific sequences,
`
`but also for their quantification, because
`
`of the quantitative relationship between
`
`the amountofstarting target sequence and
`
`the amount of PCR product at any given
`
`cycle that falls within the reaction’s expo-
`
`nential range.
`
`The theory is straightforward, but a
`number of technical caveats are associat-
`ed with the use of conventional end-point
`
`methodologies for quantitative PCR.!2
`
`In these techniques, PCR results are
`
`monitored after a given number of
`
`cycles, by which point factors such as
`
`limiting reagent concentrations and side
`
`reactions may have played a significant
`
`role in affecting final product concentra-
`
`tion. Quantitative competitive PCR was
`
`developed in response to some of these
`
`difficulties. In this approach, the starting
`
`
`amount of target is calculated based on
`
`the ratio of target
`to competitor after
`
`amplification. However, quantita-
`
`live competitive PCR is
`
`cumbersome,and it can be
`associated with a number
`
`has
`commercialized
`the
`requisite
`eral general types of fluorescent probes?
`for detection.
`of drawbacks including a
`Teagents in its well-known TaqMan®
`limited dynamic range and
`Several different basic types of fluo-
`product
`line. The fluorogenic probe is
`
`
`the need to screen multiple
`rescent probes are used for real-time PCR
`complementary to the target sequence,
`
`dilutions. }-2
`applications. Some assays employ general
`and initially contains both reporter and
`Someofthelimitations
`
`probesthat bind preferentially to double-
`quencher moieties. When the probe is
`
`of end-point PCR have
`stranded DNA (as opposed tothe single-
`not bound to template DNA,its reporter
`
`been assuagedin real-time
`stranded template). Others use target
`and quencherdyes are in close proximi-
`
`PCR systems, a number of
`ty, and the reporter’s fluorescent emis-
`sequence-specific reagents such as
`which are now on the market.
`
`exonuclease probes, hybridiza-
`sion is quenched.
`
`These systems offer many
`tion probes, or molecular bea-
`The probe is designed to anneal
`
`general technical advantages,
`cons (hairpin probes). Although
`specifically between the forward and
`
`including reduced probabili-
`more
`expensive,
`sequence-
`reverse primer
`sites of
`the
`target
`
`ties of variability and con-
`specific probes add specificity to
`sequence.If a template bearing the target
`
`tamination, as well as
`the assay, and enable multiplex-
`sequenceis present, the probe anneals to
`
`online monitoring and
`ing applications.
`it. During PCR, the nuclease activity of
`
`the lack of need for
`Taq polymerase cleaves the reporter dye
`
`First ON LINE
`postreaction analy-
`from the probe. The reporter dye, now
`ses. Further, some of
`
`Applied Biosystems of
`separated from the quencher, emits a flu-
`these systems were
`Foster City, Calif., was
`orescent signal. Thus, fluorescent signal
`
`the first to commercial-
`developed with con-
`is emitted only after the probe binds
`
`ize real-time PCR.This
`temporary
`applications
`template DNAandis cleaved during the
`
`such as quantitative PCR,
`company offers
`the
`course of PCR. The fluorescent signal is
`
`ABI PRISM® 7700
`multiplexing, and high-throughput
`monitored at every cycle as additional
`
`Sequence Detection
`(HT) analysis in mind.? In real-time
`reporter dye molecules accumulate. After
`quantitative PCR techniques, signals
`System, which inte-
`the signal rises above background,
`its
`
`grates a closed-tube
`(generally fluorescent) are monitored as
`rate of increase is tracked during a num-
`they are generated and are tracked after
`avail-
`PCR-based
`assay
`ber of linear cycles before the reaction
`
`able real-
`with instrumentation and
`they rise above background but before
`reaches a plateau. These data are then
`
`the reaction reaches a plateau.
`Initial
`time
`PCR
`software that enable real-time quantita-
`used to calculate initial template levels.
`template levels can be calculated by ana-
`systems
`are
`tive PCR. This system is optimized for
`The ABI PRISM 7700 was designed
`overviewedin this article and/or summa-
`lyzing the shapeofthe curve or by deter-
`use of a fluorogenic 5’ exonuclease assay
`with HTapplications in mind. Reactions
`mining whenthe signal rises above some
`rized in the accompanyingtable. Each of
`that employs the Thermus aquaticus
`are performed in 96-well microplates,
`these systems employs either one of sev-
`threshold value.!2 Several commercially
`(Taq) polymerase. Applied Biosystems
`and it takes merely 15 seconds forthis
`
`CR—a technique so common in
`
`Pes laboratoriesthatitis easy to
`
`The Scientist 3 1
`November 27, 2000
`
`Agilent Exhibit 1225
`Page 1 of 4
`
`Agilent Exhibit 1225
`Page 1 of 4
`
`
`
`LabConsumer
`
`Real-time PCR Systems
`
`Appuep Biosystems
`(800) 345-5224, www.appliedbiosystems.com
`Bio-Rap
`(800) 4-BIORAD, www.bio-rad.com
`
`ABI PRISM 7700 Sequence Detection System
`
`Microplate (96)
`
`37
`
`24
`
`27
`
`320Ibs.
`
`Yes
`
`iCycler iQ Real-Time PCR Detection System
`(thermal cycler plus optional optical module)
`
`Microplate (96), Tubes (96)
`
`
`
`cieenceain ee , :
`
`
`
`
`
`
`
`
`
`* distributed by Fisher Scientific, (800) 766-7000, waw.fishersci.com
`1 dimensions given in inches
`
`system to collect one round of fluores-
`cent emission data from each well. The
`reaction tube’s transparent
`lid allows
`laser light, which is carried on an array
`of opticalfibers, to be distributed to each
`well. The laser light excites the reporter
`dye molecules to fluoresce, and the
`resultant signals are carried by the optic
`fibers to a charge-coupled device (CCD)
`camera
`for detection. Applications
`include allelic discrimination, melting
`curves, and quantification of single or
`multiplexed targets.
`GOING LIGHT
`Roche
`LightCycler™ from
`The
`Molecular Biochemicals of Indianapolis
`harnesseslight technology licensed from
`Idaho Technology Inc. of Idaho Falls,
`and then adds some additional bells and
`whistles, such as a built-in microvolume
`fluorimetric
`detection
`system that
`enablesreal-time quantitative PCR. This
`system employs thin-walled glass capil-
`lary tubes, and typical PCR experiments
`can be performed in less than 30 min-
`utes. Currently, this system supports two
`fluorescence-based methods
`for
`the
`detection of amplification products: the
`“general” DNA stain SYBR Green I (a
`product of Molecular Probes of Eugene,
`Ore.), or sequence-specific hybridization
`probe pairs.
`SYBR GreenI exhibits very little flu-
`orescence when free in solution; emis-
`sion is greatly enhanced whenit binds to
`the minor groove of the DNA double
`helix. Prior to amplification, the reaction
`mixture contains the denatured DNA,the
`ptimers, and the dye. The low-level
`background fluorescence signal generat-
`ed by the unbound dye molecules is sub-
`tracted during computer analysis. After
`annealing of the primers, a few dye mol-
`ecules can bind to the double strand.
`During elongation, more and more dye
`molecules bind to the newly synthesized
`DNA,resulting in dramatically increased
`light emission. If the reaction is moni-
`tored continuously, this increase in fluo-
`rescence can be viewed in real
`time.
`After denaturation of the DNA during
`
`The sequences of the two oligonu-
`cleotides are selected so that
`they
`hybridize to the amplified DNA frag-
`ment
`in a head-to-tail arrangement.
`When the oligonucleotides hybridize in
`this orientation, the two fluorescent dyes
`are positioned in close proximity to each
`other. The first dye (fluorescein) is excit-
`ed by the LightCycler’s light emitting
`diode (LED) filtered light source, and
`emits green fluorescent light at a slightly
`longer wavelength. When the two dyes
`are in close proximity, the emitted ener-
`gy excites the dye attached to the second
`hybridization probe, which subsequently
`emits red fluorescent
`light at an even
`longer wavelength. This energy transfer,
`referred to as
`fluorescence resonance
`energy transfer (FRET), occursefficient-
`ly only when the dyes are in close
`proximity (a distance between
`1-5
`nucleotides). Thus,in this type of assay,
`fluorescent
`intensity measurements are
`made after
`the annealing steps. The
`increasing amount of emitted fluores-
`cence is proportional
`to the increasing
`amount of DNA generated duringthelin-
`ear phase of the ongoing PCR process.
`Core Mopute
`The “heart” of the Smart Cycler® System
`from Cepheid of Sunnyvale,Calif., is the
`I-CORE™ (Intelligent Cooling/Heating
`Optical Reaction) module. According to
`companyliterature, the I-CORE module
`incorporates state-of-the-art microfluidic
`and microelectronic design. Each Smart
`Cycler processing block contains
`16
`independently programmable I-CORE
`modules, each of which performs four-
`color, real-time fluorometric detection. A
`wide variety of different multiplex or
`simplex fluorescent tags can used in con-
`junction with this system,
`including
`FAM, TET, TAM, ROX, SYBR Green,
`Cy3, Alexa, and Texas Red.
`Samples are amplified and measured in
`proprietary, sealable reaction tubes that are
`designed to optimize rapid thermal transfer
`and optical sensitivity. The Smart Cycler
`software enables single or multiple opera-
`tors to define and simultaneously carry out
`
`es
`ak UE
`The LightCycler System from Roche
`
`the next heating cycle, the dye molecules
`are released and the fluorescence signal
`falls. A fluorescence measurementis per-
`formed at the end of the elongation step
`of every PCR cycle to monitor
`the
`increasing amount of amplified DNA.
`The hybridization probe
`format
`employs
`two
`specially
`designed,
`sequence-specific oligonucleotides lab-
`
`eled with fluorescent dyes. One oligonu-
`cleotide probe carries a fluorescein label
`at its 3' end; the other probecarries a dif-
`ferent label (LC Red 640 or LC Red 705)
`at its 5' end. The chemical nature of the
`hybridization probes prevents
`their
`extension: one probe contains fluores-
`cein at the 3’ end, whereasthe 5'-labeled
`probe contains a 3' phosphate moiety.
`
`
`32 The Scientist
`November 27, 2000
`
`Agilent Exhibit 1225
`Page 2 of 4
`
`Agilent Exhibit 1225
`Page 2 of 4
`
`
`
`multiple separate experiments, each with a
`uniqueset of cycling protocols. In addition,
`thermal and optical data from each andall
`sites can be monitored in real time, and
`graphs of temperature, growth curves, and
`melt curves can be charted as the data are
`collected. The Smart Cycler Starter System
`includes a processing block, Windows-
`compatible computer and monitor, soft-
`ware, mini-centrifuge, tube racks, and a
`cooling block specifically designed to
`accommodate Smart Cycler reaction
`tubes (25 or 100 pl).
`
`Cominc Soon
`Stratagene of LaJolla, Calif., plans to
`Jaunch
`the Mx4000™ Multiplex
`Quantitative PCR System at the end of
`this year. The Mx4000 combines the
`capabilities of a microplate fluores-
`cence reader with a PCR thermal cycler
`into a single real-time detection system
`and allows detection of multiple fluo-
`rescent PCR chemistries. It includes an
`integrated, proprietary thermal system
`and a multiple-fluorophore detection
`system. The Mx4000 Multiplex
`Quantitative PCR System has extended
`excitation (350 to 750 nm) and detec-
`tion (350 to 830 nm)ranges, allowing
`researchers to choose fluorophores
`with little or no spectral overlap. Each
`of the four scanning fiber-optic heads
`independently excites and detects dyes,
`reading upto fourdyes in a single tube.
`Optimized interference filters are used
`to block out unwanted crosstalk from
`spectrally adjacent fluorophores.
`Researchers can choose from FAM,
`HEX, TAMRA, Cy5, Cy3, Texas
`Red/ROX,and TETfilter sets, and cus-
`tom sets for other fluorophores also are
`available. Reactions are performed in
`96-well microtiter plates, but
`res-
`earchers can opt 10 analyze reactions in
`a subsetof the wells rather than always
`reading the entire plate. The system
`includes PCR application software,
`including a feature that enables real-
`time amplification plots. Thus, the user
`can monitor progress of an experiment
`at any time during thermal cycling,
`rather than waiting until the end of the
`run. Data can be viewed in a variety of
`forms,
`including amplification plots,
`scatter plots, sample value screens,flu-
`orescence intensity screens, melting
`curves, annealing ranges, and text
`reports.
`The
`system
`employs
`Stratagene’s new solid-state heating
`and cooling technology. According to
`the company’s promotional literature,
`the low thermal mass of the sample
`temperature control] block promotes
`rapid changes in temperature, resulting
`in speedy thermal
`ramp rates and
`greater temperature uniformity.
`The instruments mentioned here
`enable “closed tube” PCR analysis in real
`time. Results are available during and
`after PCR, with no additional purification
`or analyses. This reducesthe likelihood of
`introducing variability or contaminants
`and allows researchers to say goodbye to
`postamplification analytical gels. This
`article focused primarily on products
`
`LabConsumer
`
`from companies offering fairly comprehen-
`sive quantitative PCR systems that include
`notjust the thermal cycler and optical equip-
`ment, but the requisite computer hardware
`and software as well. The accompanying
`table contains summary information for
`these systems, as well as for additional real-
`
`time PCR instruments that employ user-
`supplied computer hardware.
`Deborah Fitzgerald
`(df@ sciwriters.com) is a freelance
`science writer in Birmingham, Ala.
`References
`
`1.
`
`PE Biosystems, “DNA/RNA Real-Time Quant-
`itative PCR,” www.appliedbiosystems.com/
`molecularbiology/about/per/sds/5700_sds/pdf/
`dnarna.pdf, 1999.
`E. Zubritsky, “Pinning down PCR: Widespread
`interest in gene quantitation and high-throughput
`assays are putting quantitative PCR backin the spot-
`light,” Analytical Chemistry, 71:1914-SA, 1999.
`
`
`
`
`Software Specifically Designed for
`Genomic Expression Experiments
`
`Easy to Use
`
`© Wizard imports from any data source
`
`* Saves your analysis as HTMLfiles with full
`
`annotation
`
`© Weblinks to other data sources
`
`for fast and easy mining
`
`¢ Instantly makes gene
`
`lists from keywords,
`EC numbers,
`Easy to Visualize
`Function, etc.
`* Visually displays physical position,
`classification, trees/dendrograms, graph
`array layouts, biochemical pathways,
`Venn Diagram, orderedlists, scatter plots,
`gene to gene comparisons and more
`* Simple zooming and panning
`° Vivid display ofvexperiment,
`function, or parameter by color
`
`Easy to Analyze
`Hierarchical and non-hierarchical clustering, principal
`componentanalysis, Self-Organizing Maps, multiple
`correlations, quantitative restrictions and more
`e Automatic annotation from GenBank and LocusLink
`
`* Annotation-derived genelists
`
`
` ¢ Regulatory sequence search
`
`
`
`PLUS GenEx™:A database that allows scientists working with any organism to inter-
`nally distribute and visualize gene expression data from micro-arrays, Affymetrix
`chips and related technologies. GenEx works seamlessly with GeneSpring.
`
`Call or visit our website for
`FREEfully-functional demo
`650-367-9600
`‘~S
`
`Silicon
`Genetics
`www.sigenetics.com
`Circle No. 15 on Reader Service Card
`
`The Scientist 33
`November 27, 2000
`
`Agilent Exhibit 1225
`Page 3 of 4
`
`Agilent Exhibit 1225
`Page 3 of 4
`
`
`
`Please Note: The contents of the attached document are
`copyrighted works and may not be shared. You have secured
`permission to use this document for the requested purpose below:
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)(cid:82)(cid:101)(cid:103)(cid:117)(cid:108)(cid:97)(cid:116)(cid:111)(cid:114)(cid:121)(cid:32)(cid:83)(cid:117)(cid:98)(cid:109)(cid:105)(cid:115)(cid:115)(cid:105)(cid:111)(cid:110)
`
`To the extent that a publisher or other appropriate rights-holder has
`placed additional terms and conditions on your use of this
`document, such terms and conditions are specified herein under
`“Copyright Terms”.
`
`This coversheet may not be removed from the article without
`permission from the Information Center. For permission to share
`this article please contact the Information Center at
`infocenter@thermofisher.com or 815-987-4625.
`