throbber
Case IPR2014-00362
`U.S. Patent No. 6,306,141
`
`Paper No.____
`Filed on April 29, 2014
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`____________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`____________
`
`EDWARDS LIFESCIENCES CORPORATION
`Petitioner
`v.
`
`MEDTRONIC, INC.
`Patent Owner
`____________
`
`Case IPR2014-00362
`U.S. Patent 6,306,141
`____________
`
`MEDTRONIC, INC.’S PRELIMINARY RESPONSE
`
`Mail Stop PATENT BOARD
`Patent Trial and Appeal Board
`United States Patent and Trademark Office
`P.O. Box 1450
`Arlington, VA 22313-1450
`
`COOK
`Exhibit 1006-0001
`
`

`

`TABLE OF CONTENTS
`
`Page
`
`2.
`
`2.
`
`3.
`
`I.
`II.
`
`B.
`
`C.
`
`INTRODUCTION..........................................................................................1
`FACTUAL BACKGROUND ........................................................................3
`A.
`The Technology of the ’141 Patent......................................................3
`B.
`The Petition Alleges Nothing More than Previously Considered
`Art, and in Some Cases, Exactly the Same Grounds of
`Invalidity ..............................................................................................9
`Asserted References...........................................................................10
`C.
`III. CLAIM CONSTRUCTION .........................................................................14
`IV.
`THERE IS NO REASONABLE LIKELIHOOD THAT PETITIONER
`WILL PREVAIL AS TO GROUNDS #1-#5 OF THE PETITION.............14
`A.
`Petitioner Is Precluded from Raising ODP in this Proceeding
`(Ground #5)........................................................................................14
`Petitioner Is Not Reasonably Likely to Prevail on Its
`Anticipation Arguments as to Cragg I and Cragg II (Grounds #1
`and #2)................................................................................................16
`1.
`Neither Cragg I nor Cragg II Expressly Discloses the
`SIM Limitation.........................................................................17
`Petitioner’s Argument that Cragg I and Cragg II
`Inherently Disclose the SIM Limitation Is Based Upon
`Expert Opinion that Should Be Accorded No Weight ............18
`Petitioner Is Not Reasonably Likely to Prevail on Its
`Obviousness Arguments (Grounds #2, #3, and #4) ...........................21
`1.
`The Combination of Cragg I and Cragg II Fails, Because
`Neither Reference Teaches the SIM Limitation (Ground
`#2) ............................................................................................21
`The Combinations of (1) Cragg I and Miyauchi and
`(2) Cragg I and Fannon Fail, Because None of the
`References Teach the SIM Limitation (Grounds #3 and
`#4) ............................................................................................22
`Petitioner’s Combinations of References Fail, as Those
`References Teach Away from the Claimed Invention.............24
`
`COOK
`Exhibit 1006-0002
`
`

`

`TABLE OF CONTENTS
`(continued)
`
`Page
`
`D.
`
`Petitioner Has Failed to Amend Its Claim Charts to Remove
`Attorney Argument in Contravention of the Board’s Order,
`Thereby Prejudicing Medtronic .........................................................25
`CONCLUSION ............................................................................................26
`
`V.
`
`-ii-
`
`COOK
`Exhibit 1006-0003
`
`

`

`Ex. #
`
`2001
`
`2002
`
`2003
`
`2004
`
`2005
`
`2006
`
`2007
`
`2008
`
`2009
`
`EXHIBIT LIST
`
`Exhibit Description
`
`D. Stoeckel, “The Shape Memory Effect – Phenomenon,
`Alloys and Applications,” Proceedings: Shape Memory Alloys
`for Power Sys. EPRI, 1-13 (2005)
`
`Biscarini et al., “Enhanced Nitinol Properties for Biomedical
`Applications,” Recent Patents on Biomed. Eng’g, No. 1, 180-
`196 (2008)
`
`Medtronic, Inc. v. AGA Med. Corp., Case No. 07-00567 MMC
`(Trial Exhibit)
`
`Medtronic, Inc. v. AGA Med. Corp., Case No. 07-00567 MMC,
`2009 WL 1163976 (N.D. Cal. Apr. 28, 2009)
`
`Medtronic, Inc. v. AGA Med. Corp., Case No. 07-00567 MMC,
`(N.D. Cal. Aug. 6, 2009) (Verdict Form)
`
`Medtronic, Inc. v. W.L. Gore & Assoc., Inc., Case No. 06-
`04455 JSW, 2008 WL 5191846 (N.D. Cal. Dec. 9, 2008)
`
`Press Release, Medtronic and W.L. Gore & Associates Settle
`Patent Dispute, Sept. 16, 2009
`
`M. Wu, “Fabrication of Nitinol Materials and Components,”
`Proceedings of Int’l Conference on Shape Memory &
`Superelastic Techs., 285-92 (2001)
`
`U.S. Publication No. 2004/0193257 to Wu and Poncet (filed
`Mar. 2004; published Sept. 20, 2004)
`
`-iii-
`
`COOK
`Exhibit 1006-0004
`
`

`

`I.
`
`INTRODUCTION
`
`Patent Owner Medtronic, Inc. (“Medtronic”) respectfully submits this
`
`Preliminary Response to the Petition filed by Edwards Lifesciences Corporation
`
`(“Petitioner”) seeking inter partes review of Claims 1-22 of U.S. Patent No.
`
`6,306,141 (the “’141 patent”). The Petition for inter partes review of the ’141
`
`patent should be denied, as it fails to demonstrate a reasonable likelihood of
`
`prevailing as to any claim on any of Petitioner’s five asserted Grounds.
`
`As for obviousness-type double patenting (“ODP”) (Ground #5), this
`
`argument is not a permissible Ground on which to base inter partes review. Such
`
`proceedings may only be based on patents or printed publications under 35 U.S.C.
`
`§§ 102 or 103. The ODP doctrine, by contrast, is judicially-created and rooted in
`
`35 U.S.C. § 101.
`
`As for anticipation, Grounds #1 (Cragg I) and #2 (Cragg II) fail for at least
`
`two reasons. First, Cragg I and Cragg II fail to expressly disclose a key limitation
`
`recited in all of the claims, namely the stress-induced martensite limitation. That is
`
`because those references exclusively teach medical devices made of shape memory
`
`alloys (“SMA devices”) that are deployed in the body by temperature-induced
`
`martensite (“TIM”), while the ’141 patent claims SMA devices deployed in the
`
`body by stress-induced martensite (“SIM”). Second, Petitioner’s argument that
`
`Cragg I and Cragg II inherently teach the SIM limitation is based upon an expert
`
`1
`
`COOK
`Exhibit 1006-0005
`
`

`

`opinion that should be given no weight. Dr. Ming Wu calculated transition
`
`temperatures for the disclosed SMA devices without any reference to a crucial
`
`factor in generating such temperatures, namely the specific chemical composition
`
`of the Nitinol used in the disclosed SMA devices. Dr. Wu’s failure to account for
`
`this factor is particularly striking given that Dr. Wu himself has noted the
`
`importance of the specific chemical composition of Nitinol in calculating transition
`
`temperatures, and even Cragg I teaches that point.
`
`With respect to obviousness, Grounds #2 (Cragg I and Cragg II), #3 (Cragg I
`
`and Miyauchi) and #4 (Cragg I and Fannon) likewise fail for at least two reasons.
`
`First, as with anticipation, all of the asserted references fail to disclose the SIM
`
`limitation. Second, those references teach away from the claimed invention, as
`
`they exclusively teach the use of TIM to introduce SMA devices into the body, not
`
`SIM.
`
`Lastly, Petitioner impermissibly incorporated attorney argument into its
`
`claim charts in contravention of the Board’s rules. Petitioner has yet to comply
`
`with the Board’s instruction three months ago to correct its Petition, thereby
`
`prejudicing Medtronic’s ability to prepare its Response through circumvention of
`
`the page limit.
`
`2
`
`COOK
`Exhibit 1006-0006
`
`

`

`For all of these reasons, Petitioner has failed to show a reasonable likelihood
`
`of prevailing with respect to any claims on any of the five asserted Grounds, such
`
`that the Petition for inter partes review of the ’141 patent should be denied.
`
`II.
`
`FACTUAL BACKGROUND
`
`A.
`
`The Technology of the ’141 Patent
`
`The ’141 patent is directed to medical devices made of shape memory alloys
`
`(“SMAs”) such as Nitinol, a term which refers to a range of alloys that can be
`
`made of varying amounts of nickel and titanium. (Ex. 1001 at col. 1, ll. 21-23; Ex.
`
`1004 at 1-2 (defining Nitinol and discussing Nitinol compositions).) SMAs can
`
`“remember” their original shape and revert back to it after they have been
`
`deformed. (Ex. 1001 at col. 1, ll. 26-34.) This property has the advantage of
`
`enabling medical devices made with SMAs (“SMA devices”) to adopt a smaller
`
`configuration for insertion into a placement device, e.g., a catheter, and in turn,
`
`deployment into the body. (Id. at col. 2, ll. 23-28.) Once deployed, an SMA
`
`device can remember, and thereby return to, its original, larger configuration. (Id.)
`
`The ability of an SMA to remember its original shape is a consequence of its
`
`ability to exist in two different states (or crystal structures) that occur at the atomic
`
`level, namely “austenite” and “martensite.” (See, e.g., id. at col. 1, ll. 35-38.) The
`
`transformation between austenitic and martensitic states can be effected one of two
`
`ways: by temperature or mechanical stress. In the case of temperature, martensite
`
`3
`
`COOK
`Exhibit 1006-0007
`
`

`

`forms at relatively lower temperatures, and austenite forms at relatively higher
`
`temperatures. (Id.) In the case of stress, martensite forms through the application
`
`of stress, and austenite forms when that stress is relieved. (Id. at col. 1, ll. 52-65.)
`
`Significantly, the prior art references cited in the Petition are concerned with
`
`effecting this transformation by way of temperature.
`
`By way of example, in the asserted Cragg I reference, the desired SMA
`
`device was a coiled wire to be used as an aortic stent. (Ex. 1004 at 1.) A Nitinol
`
`wire was shaped into a coil and heated to its “annealing temperature” to confer
`
`memory. (Id.) In this coiled configuration, the SMA device was in its “austenitic”
`
`state. By decreasing its temperature via submersion in sterile ice water, the coiled
`
`wire was transformed into its martensitic state, allowing it to be deformed into a
`
`straight wire, so that it could fit into a catheter for deployment (Id.) Once deployed
`
`from the catheter and into the aorta, the increased temperature in the aorta
`
`transformed the temperature-induced martensitic state (straight wire) back to its
`
`austenitic state (coil). (Id. at 2.)
`
`As seen in Cragg I and the other references asserted by Petitioner, the
`
`medical community
`
`initially employed
`
`temperature
`
`to control SMA
`
`transformations between austenite and martensite, thereby exploiting the property
`
`known as temperature-induced martensite (“TIM”). (Ex. 1001 at col. 2, ll. 16-28.)
`
`As pertinent here, the temperature at which austenite begins transforming to
`
`4
`
`COOK
`Exhibit 1006-0008
`
`

`

`martensite is the martensite start temperature (Ms); the temperature at which the
`transformation is complete is the martensite finish temperature (Mf). (Id. at col. 1,
`ll. 44-46.) Corresponding temperatures for the martensite to austenite conversion
`
`are austenite start (As) and austenite finish (Af). (Id. at col. 1, ll. 46-51.) These
`transition temperatures may be graphically depicted as follows:
`
`(Ex. 2001, D. Stoeckel, “The Shape Memory Effect – Phenomenon, Alloys and
`
`Applications,” Proceedings: Shape Memory Alloys for Power Sys. EPRI, 1, 2
`
`(2005). As will be explained infra, these transition temperatures are highly
`
`dependent on the chemical composition of the SMA. (See Section IV.B.2.) In the
`
`case of Nitinol, for example, transition temperatures can vary dramatically based
`
`upon the ratio of nickel and titanium in the alloy. (Id.) As such, simply knowing
`
`that an SMA device is made of Nitinol reveals nothing with respect to the
`
`temperatures at which the foregoing phase transitions will occur. (Id.)
`
`Significantly, the TIM approach had at least three problems. First, the TIM
`
`approach required very precise temperature control for deploying an SMA device.
`
`5
`
`COOK
`Exhibit 1006-0009
`
`

`

`(Ex. 1001 at col. 2, ll. 37-41.) This process included using low temperatures for
`
`converting the device to its martensitic state for insertion into a placement device,
`
`followed by using higher temperatures to transform the device into its austenitic
`
`state for deployment. Because such temperature manipulation would partly occur
`
`in the patient’s body, tissue damage could result. (Id. at col. 2, ll. 43-48.) Second,
`
`transition temperatures could vary greatly due to the variability of the chemical
`
`composition and heat treatment processing of the SMA. (Id. at col. 2, ll. 31-37.)
`
`Lastly, the need for temperature manipulation was a source of substantial
`
`inconvenience for physicians. (Id. at col. 2, ll. 41-43.) Indeed, as demonstrated by
`
`the asserted references, temperature manipulation would involve such cumbersome
`
`techniques as the use of ice baths, cooling elements, and heating elements during
`
`surgical procedures. (Ex. 1004 at 1; Ex. 1005 at 3; Ex. 1006 at 4.)
`
`The invention of the ’141 patent overcame these problems by controlling
`
`SMA transformations between austenite and martensite by applying or removing
`
`stress to an SMA device, thereby exploiting the property known as stress-induced
`
`martensite (“SIM”). (Id. at col. 2, l. 64 to col. 3, l. 4.) The application of stress to
`
`an SMA device at a temperature above its As transforms austenite to martensite,
`and removal of the stress transforms the SMA device back to its original austenitic
`
`state, irrespective of the temperature of the SMA device. (Id. at col. 4, ll. 6-11.)
`
`Thus, while the prior art is exclusively concerned with the temperature-dependent
`
`6
`
`COOK
`Exhibit 1006-0010
`
`

`

`creation and reversion of martensite for the deployment of SMA devices, the ’141
`
`patent sets forth a stress based approach that is controllable, safe and much more
`
`convenient to use. (Id. at col. 2, ll. 31-54.)
`
`Claim 1 recites as follows, with pertinent limitations highlighted:
`
`1. A medical device for insertion into a mammalian body, the
`device comprising
`
`(a) a hollow placement device;
`
`(b) a memory alloy element formed at least partly from
`pseudoelastic shape-memory alloy, the alloy displaying reversible
`stress-induced martensite at about body temperature such that it
`has a stress-induced martensitic state and an austenitic state, the
`memory alloy element having (i) a deformed shape when the alloy
`is in its stress-induced martensitic state and (ii) a different
`unstressed shape when the alloy is in its austenitic state; and
`
`(c) a guide wire;
`
`[(d)] the memory alloy element being within the hollow placement
`device, and the placement device being guidable by the guide wire,
`the hollow placement device stressing the memory alloy element
`at a temperature greater than the As of the alloy so that the
`memory alloy element is in its deformed shape,
`
`[(e)] wherein the memory alloy element can be extruded from the
`hollow placement device by the guide wire at a temperature greater
`than the As of the alloy to transform at least a portion of the alloy
`from its stress-induced martensitic state so that the memory alloy
`
`7
`
`COOK
`Exhibit 1006-0011
`
`

`

`element transforms from its deformed shape to its unstressed
`shape, and wherein the alloy is selected so that the transformation
`can occur without any change in temperature of the placement
`device or the memory alloy element.
`
`(Id. at col. 10, l. 60 to col. 11, l. 20 (emphases and bracketing added).)
`
`Thus, the claimed invention provides a combination of an SMA device and a
`
`hollow placement device such that the hollow placement device has a structure that
`
`stresses the SMA device into its deformed shape. (Id.) Because the SMA device
`
`displays reversible SIM at about body temperature and the placement device keeps
`
`the SMA device in its stress-induced martensitic state, the SMA device can
`
`transform its shape without a change in temperature when deployed from the
`
`placement device into the body. (Id. at col. 2, l. 64 to col. 3, l. 4.)
`
`It should be noted that the ’141 patent’s approach of employing SIM to
`
`deploy SMA devices was a transformational invention in the medical device
`
`industry, leading to widespread application and industry-wide adoption. As
`
`explained in one article:
`
`A great number of medical devices are based on stress
`induced martensite. The most commercially important
`patents now pending
`for
`the medical devices
`incorporating SIM alloys include the Jervis’s series
`which is listed in Jervis patent [37].
`
`8
`
`COOK
`Exhibit 1006-0012
`
`

`

`(Ex. 2002, Biscarini et al., “Enhanced Nitinol Properties for Biomedical
`
`Applications,” Recent Patents on Biomed. Eng’g, No. 1, 180, 183 (2008).) Indeed,
`
`many of the major players in the medical device industry have taken licenses to the
`
`’141 patent, including Johnson & Johnson, Boston Scientific, U.S. Surgical, and
`
`Lombard Medical. (Ex. 2003, Medtronic, Inc. v. AGA Med. Corp., Case No. 07-
`
`00567 MMC, Trial Exhibit (providing partial list of licensees).)
`
`B.
`
`The Petition Alleges Nothing More than Previously Considered
`Art, and in Some Cases, Exactly the Same Grounds of Invalidity
`
`The ’141 patent underwent extensive prosecution, and because of its seminal
`
`nature and adoption by others in the medical device industry, it was the subject of
`
`two hotly contested litigations in the Northern District of California. Petitioner’s
`
`present challenge raises nothing more previously considered art, and some cases,
`
`exactly the same invalidity grounds.
`
`With respect to prosecution of the ’141 patent, the Examiner considered two
`
`of the four asserted references, namely Cragg I and Fannon. (Ex. 1025, Form
`
`PTO-1449 at 447.)
`
` In fact, both references feature prominently in the
`
`specification. (Ex. 1001 at col. 6, l. 58 to col. 7, l. 5; col. 9, ll. 41-57.)
`
`With respect to litigation, the ’141 patent was challenged in Medtronic, Inc.
`
`v. AGA Medical Corp., Case No. 07-00567 MMC (N.D. Cal.). In that case, AGA
`
`asserted what is now Petitioner’s anticipation argument in Ground #2 (§ 102(a) in
`
`view of Cragg II) and obviousness argument in Ground #3 (§ 103(a) based on
`
`9
`
`COOK
`Exhibit 1006-0013
`
`

`

`combination of Cragg I and Miyauchi). Judge Chesney denied AGA’s motion for
`
`summary judgment of invalidity on both grounds, and the jury likewise refused to
`
`invalidate the claims on those grounds. (Ex. 2004, Medtronic, Inc. v. AGA Med.
`
`Corp., No. 07-00567 MMC, 2009 WL 1163976, at *2-*5 (N.D. Cal. Apr. 28,
`
`2009); Ex. 2005, Verdict Form at 4-5.)
`
`The ’141 patent was also challenged in Medtronic, Inc. v. W.L. Gore &
`
`Associates, Inc., Case No. 06-04455 JSW (N.D. Cal.). W.L. Gore asserted that the
`
`’141 patent was obvious in light of certain prior art references, including Cragg I.
`
`Judge White denied summary judgment, holding that Cragg I taught away from
`
`deploying an SMA device by way of SIM. (Ex. 2006, Medtronic, Inc. v. W.L.
`
`Gore & Assoc., Inc., Case No. 06-04455 JSW, 2008 WL 5191846, at *9 (N.D. Cal.
`
`Dec. 9, 2008).) The W.L. Gore case settled shortly after the AGA verdict in favor
`
`of Medtronic. (Ex. 2007, Press Release.)
`
`C.
`
`Asserted References
`
`As described below, each of the four asserted references discloses the same
`
`temperature-dependent approach for deploying an SMA device into the body.
`
`Stated simply, this approach involves (1) cooling the SMA device to its martensitic
`
`state, which allows the SMA device to be deformed from its original configuration,
`
`usually resulting in a smaller configuration that enables it to fit into a placement
`
`device, e.g., a catheter, and (2) deploying the SMA device from the placement
`
`10
`
`COOK
`Exhibit 1006-0014
`
`

`

`device into the body where it is warmed, causing the SMA device to revert to its
`
`austenitic state and reform its original configuration. This approach, which relies
`
`exclusively on TIM to deploy the SMA device, is in marked contrast to the ’141
`
`patent, which relies exclusively on SIM to deploy the SMA device.
`
`Cragg I. Cragg et al., Nonsurgical Placement of Arterial Endoprostheses: A New
`
`Technique Using Nitinol Wire, 147 Radiology No. 1, 261-63 (April 1983)
`
`Cragg I discloses an endovascular prosthesis, i.e., a coiled wire made from
`
`Nitinol, for deployment into the aorta of four canine subjects. (Ex. 1004 at 1.) To
`
`deploy the coil via a catheter, the coiled wire was straightened by cooling “in a
`
`bath of sterile ice water,” which converted the Nitinol wire to its martensitic state.
`
`(Id.) The wire was maintained in the straightened configuration i.e. in its
`
`martensitic state, by flushing “a cold saline solution (10o C)” through the catheter.
`
`(Id.) Once the wire was deployed from the catheter into the aorta, body
`
`temperature converted the straightened wire back to the austenitic state, causing
`
`the wire to resume its original coiled shape. (Id. at 2.) Cragg I thus discloses a
`
`temperature-dependent approach for deploying a Nitinol coiled wire into a canine
`
`aorta.
`
`Cragg II. Cragg et al., A New Percutaneous Vena Cava Filter, 141 Am. J.
`
`Radiology, 601-04 (Sept. 1983)
`
`11
`
`COOK
`Exhibit 1006-0015
`
`

`

` Cragg II discloses a percutaneous filter for embolic capture in the canine
`
`inferior vena cava. (Ex. 1005 at 2.) The filter consists of a Nitinol wire formed
`
`into the shape of a helical spiral. (Id.) To allow the helically spiraled filter to be
`
`deployed via catheter, Cragg II follows the same basic approach as Cragg I: the
`
`filter was straightened by cooling it in sterile ice water (saline solution), thereby
`
`converting the filter to its martensitic state. (Id. at 3.) And as with Cragg I, the
`
`straightened wire was then deployed into the body (inferior vena cava) via catheter,
`
`where body temperature converted the straightened wire back to the austenitic
`
`state, causing the wire to resume its original helical spiral shape. (Id.) Thus, like
`
`Cragg I, Cragg II exclusively discloses a temperature-dependent approach for
`
`deploying an SMA device into the body.
`
`Miyauchi. Japanese Patent Publication No. S58-46923 entitled “Endoscope”
`
`Miyauchi discloses an endoscope equipped with a “manipulation tool” for
`
`use in surgical applications. (Ex. 1006 at 2.) The manipulation tool can be an
`
`SMA wherein the reverse transformation temperature is at or about body
`
`temperature. (Id. at 3.) To insert the tool into the endoscope, the manipulation tool
`
`is “temporarily deform[ed] into a shape long in the lengthwise direction.” (Id. at
`
`4.) As in Cragg I and Cragg II, Miyauchi teaches that the tool be deformed to its
`
`martensitic state by being “cooled in advance” (Id.), and further discloses that the
`
`tool may be maintained in that state “by cooling the surroundings of the tube-
`
`12
`
`COOK
`Exhibit 1006-0016
`
`

`

`shaped path 16 with a water line, etc.” (Id.) Restoring the shape of the tool to its
`
`austenitic state is accomplished by either body temperature as in Cragg I and Cragg
`
`II, or by a heating means. (Id.) Thus, like Cragg I and Cragg II, Miyauchi
`
`exclusively discloses a temperature-dependent approach for deploying an SMA
`
`device into the body.
`
`Fannon. U.S. Patent No. 3,620,212 entitled “Intrauterine Contraceptive Device”
`
`Fannon discloses an intrauterine contraceptive device (“IUD”) comprised of
`
`a Nitinol alloy. (Ex. 1007 at 3.) To insert the device into the uterus, the IUD is
`
`deformed into a straight strand or a compact fold configuration and inserted into a
`
`cannula for insertion. (Id. at 3-4.) As with Cragg I, Cragg II, and Miyauchi,
`
`Fannon teaches that this deformation is accomplished by keeping the IUD at a
`
`relatively lower temperature, in this case below 70° F. (Id. at 4.) And in a manner
`
`similar to those references, once the deformed IUD is inserted into the body, it
`
`returns to the austenitic state, and assumes its original shape. (Id.; see also
`
`Abstract (“The device will gradually resume its free shape as it is heated to
`
`temperature near the human body temperature.”).) Thus, Fannon, like Cragg I,
`
`Cragg II, and Miyauchi, once again exclusively discloses a temperature-dependent
`
`approach for deploying an SMA device into the body.
`
`As may be observed from the foregoing, all of the asserted references
`
`exclusively disclose a temperature-dependent approach for deploying an SMA
`
`13
`
`COOK
`Exhibit 1006-0017
`
`

`

`device into the body. For that reason, the Board should deny the Petition given
`
`that those references fail to anticipate or render the claimed invention obvious, as
`
`the claims exclusively employ a stress-based approach for employing an SMA into
`
`the body. (See Section IV, infra.)
`
`Should the Board decide to institute a trial, Medtronic urges the Board to
`
`limit the Grounds upon which it proceeds given the highly duplicative disclosures
`
`of the references upon which those Grounds are based. See, e.g., Amkor Tech., Inc.
`
`v. Tessera, Inc., Case IPR2013-00242, Paper No. 37 at 33 (PTAB Oct. 11, 2013)
`
`(denying institution of inter partes review as to Grounds based upon cumulative
`
`prior art).
`
`III. CLAIM CONSTRUCTION
`
`Medtronic believes that no claim construction issues impact the arguments
`
`raised in this Preliminary Response, but reserves the right to raise issues of claim
`
`construction in the event that trial is instituted.
`
`IV. THERE IS NO REASONABLE LIKELIHOOD THAT PETITIONER
`WILL PREVAIL AS TO GROUNDS #1-#5 OF THE PETITION
`
`A.
`
`Petitioner Is Precluded from Raising ODP in this Proceeding
`(Ground #5)
`
`Ground #5 seeks to invalidate Claims 1-22 of the ’141 patent for
`
`obviousness-type double patenting (“ODP”) in view of U.S. Patent No. 5,597,378
`
`14
`
`COOK
`Exhibit 1006-0018
`
`

`

`(the “’378 patent”). (Petition at 50-60.) A trial should not be instituted on this
`
`Ground, as there is no statutory basis for raising ODP in this proceeding.
`
`The America Invents Act strictly limits the grounds on which a petitioner
`
`can seek inter partes review to those “that could be raised under section 102 or 103
`
`and only on the basis of prior art consisting of patents or printed publications.” 35
`
`U.S.C. § 311(b); see also 37 C.F.R. § 42.104(b)(2). Petitioner’s ODP challenge is
`
`impermissible for two reasons: (1) it is based on a judicially-created doctrine
`
`stemming from Section 101, and (2) the ’378 patent is not prior art to the ’141
`
`patent, because both patents are in the same family.
`
`ODP is a non-statutory, judicially-created doctrine rooted in public policy.
`
`Apple Inc. v. SightSound Techs., LLC, CBM2013-00021, Paper No. 13 at 24
`
`(PTAB Oct. 8, 2013). While non-statutory, the origins of ODP may be traced to 35
`
`U.S.C. § 101, which prohibits patenting of the same invention twice. Eli Lilly &
`
`Co. v. Teva Pharm. USA., Inc., 619 F.3d 1329, 1341 (Fed. Cir. 2010)
`
`(“Nonstatutory double patenting [ODP] was borne out of 35 U.S.C. § 101, not §
`
`103.”). ODP thus cannot be the basis for inter partes review, because it is the
`
`wrong sort of claim rejection (based on judicial doctrine, rather than statute), and
`
`derives from the wrong statute (Section § 101). See Apple, CBM2013-00021,
`
`Paper No. 13 at 25 (dismissing ODP ground in CBM petition, reasoning that ODP
`
`“is a judicially created, policy based doctrine” rather than a statutory one).
`
`15
`
`COOK
`Exhibit 1006-0019
`
`

`

`Moreover, it is well established that “the patent principally underlying the
`
`double patenting rejection is not considered prior art.” M.P.E.P. § 804(II)(B)(1)
`
`(8th ed. Rev. 7, July 2008). In this case, the ’378 patent and the ’141 patent both
`
`claim ultimate priority to the same patent application, Appl. Ser. No. 06/541,852,
`
`filed October 14, 1983. Thus, the ’378 patent is not prior art to the ’141 patent.
`
`Because ODP is a non-statutory, non-prior-art-based argument, it is not a
`
`permissible basis for inter partes review under 35 U.S.C. § 311(b). Therefore,
`
`review of Claims 1-22 should not be instituted based on Ground #5.
`
`B.
`
`Petitioner Is Not Reasonably Likely to Prevail on Its Anticipation
`Arguments as to Cragg I and Cragg II (Grounds #1 and #2)
`
`Grounds #1 and #2 seek to invalidate Claims 1-22 of the ’141 patent under
`
`§ 102(a) in view of Cragg I or Cragg II. (Petition at 13-17.) A trial should not be
`
`instituted on those Grounds, as both references teach SMA medical devices
`
`deployed by way of TIM, while the ’141 patent claims SMA devices deployed by
`
`way of SIM. (Ex. 1004 at 2; Ex. 1005 at 3.) As such, Petitioner cannot show that
`
`Cragg I or Cragg II teaches the SIM limitation recited in Claims 1-22, either
`
`expressly or inherently. This deficiency is fatal to Petitioner’s anticipation
`
`arguments. See, e.g., Atofina v. Great Lakes Chem. Corp., 441 F. 3d 991, 999
`
`(Fed. Cir. 2006) (“Anticipation requires a showing that each limitation of a claim is
`
`found in a single reference, either expressly or inherently.”); In re Oelrich, 666
`
`F.2d 578, 581 (C.C.P.A. 1981) (“Inherency, however, may not be established by
`
`16
`
`COOK
`Exhibit 1006-0020
`
`

`

`probabilities or possibilities. The mere fact that a certain thing may result from a
`
`given set of circumstances is not sufficient.”).
`
`1.
`
`Neither Cragg I nor Cragg II
`Expressly Discloses the SIM Limitation
`
`As an initial matter, Cragg I and Cragg II both fail to expressly disclose the
`
`SIM limitation recited in Claims 1-22. As explained in detail in Section II.C,
`
`supra, both references employ the same temperature-dependent deployment
`
`approach: (1) cooling the SMA device to its martensitic state, thereby allowing for
`
`the deformation of the device so it to fits into a placement device (Ex. 1004 at 1;
`
`Ex. 1005 at 3), and (2) deploying the SMA device into the body where it is
`
`warmed, causing it to revert to its austenitic state and reassume its original shape,
`
`(Id.). Both references exclusively teach the use of temperature to transform the
`
`SMA device (austenite to martensite), and then deploy the SMA device into the
`
`body (martensite to austenite). The application of stress to create martensite and
`
`the release of stress to create austenite is nowhere taught or suggested. As such,
`
`neither reference provides an express teaching of the SIM limitation, and Petitioner
`
`has failed to point to any such disclosure. See AGA Med., 2009 WL 1163976 at *2
`
`(“A review of the Cragg II reference reveals that the display of stress-induced
`
`martensite is not expressly disclosed therein.”).
`
`17
`
`COOK
`Exhibit 1006-0021
`
`

`

`2.
`
`Petitioner’s Argument that Cragg I and Cragg II Inherently
`Disclose the SIM Limitation Is Based Upon Expert Opinion
`that Should Be Accorded No Weight
`
`Petitioner appears to argue that Cragg I and Cragg II inherently disclose the
`
`SIM limitation. (Petition at 13-17.) In support of this position, Petitioner relies
`
`upon Dr. Ming Wu’s declaration, which contains a series of transition temperature
`
`calculations that purportedly demonstrate the inherent disclosure of the SIM
`
`limitation in Cragg I and Cragg II. Notably absent from his analysis is any
`
`disclosure of a crucial factor for performing these calculations, namely the ratio of
`
`nickel to titanium in the Nitinol SMA devices disclosed in those references. As
`
`Dr. Wu himself has acknowledged, this factor is essential to calculating transition
`
`temperatures. (See, e.g., Ex. 1002, ¶ 26.) Accordingly, his opinion is unsupported
`
`and fatally defective, and Petitioner’s arguments based upon it should be rejected.
`
`Section 42.65(a) provides that “[e]xpert testimony that does not disclose the
`
`underlying facts or data on which the opinion is based is entitled to little or no
`
`weight.” 37 C.F.R. § 42.65(a). The Board has thus made clear that expert
`
`affidavits accompanying petitions “must disclose the underlying facts or data upon
`
`which the opinion is based.” Office Patent Trial Practice Guide, 77 Fed. Reg.
`
`48,756, 48,763 (Aug. 14, 2012) (emphasis added).
`
`As applied to this case, Dr. Wu provides purported calculations for the As,
`Ms, and Mf temperatures of the Nitinol alloys employed in the Cragg I and Cragg II
`
`18
`
`COOK
`Exhibit 1006-0022
`
`

`

`SMA devices. (Ex. 1002, ¶¶ 100, 102, 114, 116-117.) What is glaringly absent
`
`from those calculations is any discussion of the ratio of nickel to titanium in those
`
`Nitinol compositions. Dr. Wu fails to account for these critically important ratios
`
`for the simple reason that he cannot: neither Cragg I nor Cragg II discloses the
`
`ratio of nickel and titanium used to make the disclosed SMA devices.
`
`The importance of these ratios in calculating transition temperatures cannot
`
`be understated. As Dr. Wu himself has admitted, transition temperatures of Nitinol
`
`alloys are governed in part by the relative proportions of nickel and titanium in the
`
`composition, and even small changes in the ratio may lead to drastic changes in the
`
`transition temperatures. (Id., ¶ 26 (“[

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket