throbber
Chapter 27
`
`Glucose Monitoring via Reverse Iontophoresis
`Neil Ackerman, Bret Berner, Jim Biegajski,
`Qiang Chen, Hilary Chen,
`Tom Conn, Hardip Dehal, Tim Dunn, Al Ewing, Steve Fermi,
`Russell Ford, Priya Jagasia,
`Priti Joshi,
`Yalia Jayalakshmi,
`Brian Kersten, Ronald Kurnik, Tim Lake, Matt Lesho, Jan-Ping Lin,
`David Liu, Margarita Lopatin, Lexa Mack, Heather Messenger,
`Sam Morley, Michelle Oliva, Norman Parris, Russell Potts1,
`Jeff Pudlo,
`Michael Reidy, Pravin Soni, Janet Tamada, Michael Tierney,
`Christopher
`Uhegbu, Prcma Vijayakumar,
`Charles Wei,
`Steve Williams, Don Wilson, and Christine Wu
`
`Cygnus, Inc., 400 Penobscot Drive, Redwood Cltyt CA 94063
`
`Background
`
`Frequent glucose monitoring is essential for people with diabetes to manage their
`blood glucose levels effectively. Present procedures for obtaining such information,
`however, are invasive and painful. Development of a painless approach would
`represent a significant improvement in the quality of life for people with diabetes. Jn
`addition, results from the DCCT £11, UKPDS lll and Kumamoto Trials l3l showed that a
`tight-control regimen, which uses aggressive
`therapy with frequent glucose
`measurements to guide the administration of insulin and oral agents, leads to a
`substantial decrease in the long-term effects of diabetes. Nevertheless, even as many
`as seven measurements per day were not sufficient to prevent an increase in
`hypoglycemic events in those patients who followed this aggressive therapy llJ. The
`GlucoWatch® biographer provides a means to obtain painless and non-invasive
`measurements for up to 12 hours using a single blood measurement for calibration. A
`monitoring system that provides automatic and frequent measurement of glucose
`could provide a warning of impending hypoglycemia, potentially making aggressive
`diabetes management safer.
`Each of the current techniques for measuring blood glucose concentrations
`
`has
`
`drawbacks. The most generally accepted method relies on extraction of small
`
`
`
`1Corresponding author.
`
`© 2000 American Chemical Society
`
`273
`
`Park and Mrsny; Controlled Drug Delivery
`ACS Symposium Series; American Chemical Society: Washington, DC, 2000.
`
`AGAMATRIX, INC. EXHIBIT NO. 1026
`Page 1 of 10
`
`

`

`274
`
`aliquots of blood, obtained via a fingerstick. It is painful and invasive however, often
`resulting
`in poor patient compliance to a glucose-monitoring program. The
`measurement frequency of a typical user is not sufficient to achieve tight control.
`Implantation of biosensors or m icrodialysis tubing to sample from the subcutaneous
`tissue or peritoneal cavity is capable of providing frequent measurements
`I4.s1.
`However, poor biocompatibility, due to protein deposition, for example, limits the
`life of these devices, and the invasiveness of the method prevents wide acceptance.
`Near infrared spectroscopy, a non-invasive technique for blood glucose measurement
`r6·Y1, is not commercially established and requires large, expensive equipment. As
`such, there is no commercial device that allows non-invasive, frequent measurement
`of blood glucose.
`
`Iontophoretic Glucose Extraction
`
`The non-invasive method described here extracts glucose through the skin using
`an applied potential (a process known as reverse iontophoresis)t and measures the
`lontophoresis is a
`extracted sample using an electrochemical/enzymatic sensor.
`technique whereby a constant, low-level electrical current {0.3mA/cm2 in these
`studies) is conducted through the skin between an anode and cathode. Due to the
`applied potential, sodium and chloride ions (from beneath the skin) migrate towards
`the cathode and anode, respectively t3•91_ Uncharged molecules (e.g., glucose} are also
`It is this
`carried along with the ions by convective (electroosmotic) transport.
`convective flow that causes interstitial glucose to be transported across the skin (toJ.
`The skin has a negative charge at neutral pH, and hence, there is greater net transport
`to the cathode. As a consequence, glucose is preferentially extracted at the cathode.
`Over the typical range of iontophoretic current densities (0-0.5 mA/cm2), glucose
`extraction is linear with current density and duration of iontophoretic current 1s.91.
`The feasibility of iontophoretic glucose extraction has been demonstrated both in
`vitro tio1 and in human subjects 1111.
`In the studies with human subjects, glucose
`extraction was measured by HPLC analysis. Changes in blood glucose levels
`correlated with glucose extracted into a buffer receiver solution over a 15 minute
`period of iontophoretic current application (current density of 0.3 mA/cm2).
`Calibration of the system was perfonned to account for possible biological variability
`in skin permeability. A single point calibration was found to compensate for this
`variability. The calibration was perfonned by taking a reading using a traditional
`blood glucose measurement method, and using this reading to calibrate all subsequent
`extraction readings. It was found that glucose transport correlates well with blood
`glucose in a linear fashion, however the sensitivity (i.e. the amount of glucose
`extracted compared to the blood glucose) varied among individuals and skin sites.
`The results of this feasibility study showed a mean correlation coefficient of0.92, and
`a mean absolute relative error of 13% for the comparison of extracted to blood
`glucose values. The extraction process, however, yields a glucose concentration
`which is about 0.1 % of that found in blood. Therefore, in contrast to the sensitivity
`
`Park and Mrsny; Controlled Drug Delivery
`ACS Symposium Series; American Chemical Society: Washington, DC, 2000.
`
`AGAMATRIX, INC. EXHIBIT NO. 1026
`Page 2 of 10
`
`

`

`275
`for glucose detection required for conventional fingerstick measurements, increased
`sensitivity is required to measure iontophoretically-extracted glucose.
`
`Glucose Detection of the Iontophoretic Extraction
`
`In the GlucoWatch® biographer the concentration of extracted glucose is
`measured by a biosensor. An amperometric, electrochemical-sensing chemistry was
`chosen as the most suitable for this application. The biological selectivity element in
`this biosensor is the enzyme glucose oxidase (GOx), which catalyzes the oxidation of
`glucose to gluconic acid. This enzyme is extremely selective towards glucose. To
`obtain a signal from this enzyme reaction, it must be coupled to the sensing
`electrodes. This is achieved by the direct detection of glucose oxidase-generated,
`hydrogen peroxide (H202).
`
`GOx
`glucose + 01 � gluconic acid + H102
`The H202 is detected via an electrocatalytic oxidation reaction at a Pt-containing
`working electrode in the sensor, producing an electric current, and regenerating 02•
`-
`+
`H202 ---)> 02 + 2 H + 2 e
`
`Thus. for every glucose molecule extracted, two electrons are transferred to the
`measurement circuit. The magnitude of the resulting electric current is correlated to
`the amount of glucose collected through the skin.
`The main challenge in developing the biosensor for measuring iontophoretically·
`extracted glucose is the small amount of glucose transported through the skin, and the
`resulting low concentration that must be accurately quantified. For example, at blood
`glucose level of 50 mg/dL, approximately 50 picomoles of glucose are extracted
`through the skin during three minutes of iontophoresis, resulting in a concentration at
`the biosensor of about 4 µM. This concentration is almost three orders of magnitude
`lower than the blood glucose concentration measured by typical fingerstick blood
`glucose monitors. The biosensor that has been developed for the GlucoWatch
`biographer has high sensitivity and low noise, resulting in an extremely low limit of
`detection for glucose. The operating principles of the electrochemicaVenzymatic
`sensor are described in detail elsewhere tt21.
`
`Operation of the Gluco Watch Biographer
`
`A miniaturized device for the combined extraction and detection of glucose (the
`GlucoWatch biographer) is shown schematically in Figure 1. The extraction and
`detection is achieved using two hydrogel pads placed against the skin. The side of
`each pad away from the skin is in contact with separate iontophoretic and sensing
`electrodes. Two such electrode assemblies are required to complete the iontophoretic
`
`Park and Mrsny; Controlled Drug Delivery
`ACS Symposium Series; American Chemical Society: Washington, DC, 2000.
`
`AGAMATRIX, INC. EXHIBIT NO. 1026
`Page 3 of 10
`
`

`

`276
`
`___'3-- Hydrogel
`�
`Pads
`Sensor
`
`diagram of the GlucoWatcH" biographer.
`Figure/, A schematic
`
`Park and Mrsny; Controlled Drug Delivery
`ACS Symposium Series; American Chemical Society: Washington, DC, 2000.
`
`AGAMATRIX, INC. EXHIBIT NO. 1026
`Page 4 of 10
`
`

`

`277
`
`circuit. During operation, one iontophoretic electrode is cathodic (negatively
`charged) and the other anodic (positively charged), enabling the passage of current
`through the skin. As a consequence, glucose is collected in the hydrogel during the
`iontophoretic extraction period. The iontophoretic time interval is adjusted to
`minimize skin irritation and power requirement.s, yet extract sufficient glucose for
`subsequent detection. It has been found that an optimal time for extraction of glucose
`is about three minutes. under the conditions described here.
`The hydrogel is composed of an aqueous salt solution in a crosslinked polymer
`containing the enzyme, glucose oxidase. As described above, this enzyme catalyzes
`the conversion of the glucose (in the presence of oxygen) to hydrogen peroxide and
`gluconic acid. The peroxide is subsequently detected at an electrochemical sensor.
`Glucose exists in two forms: a-glucose and p-glucose, which differ only in the
`position of the hydroxyl group at the C-1 position in the six membered ring (u.•41.
`These two forms (called anomers) are in a proportion of 37% and 63% for
`ex. and J3 forms at equilibrium, respectively. The same proportion of ex.- and f3-glucose
`is also found in blood and interstitial fluid. As glucose enters the hydrogel, it diffuses
`throughout, but only the J3-fonn of glucose reacts with the GOx enzyme. As the P­
`form is depleted, the cx.-fonn then converts (mutarotates) to the �-form to re-establish
`the equilibrium. The products of the GOx reaction (H202 and gluconic acid) also
`diffuse throughout the hydrogel.
`On the side of the hydrogel away from the skin, and adjacent to the annular
`iontophoretic electrode, is the sensing electrode (see Figure I). A sensing electrode is
`found at both the iontophoretic anode and cathode. Thus, there are two sensing
`electrodes, noted as sensor A and B. These circular sensing electrodes are composed
`of a platinum composite, and are activated by applying a potential of 0.3�0.8 V
`(relative to a Ag/AgCl reference electrode). At these applied potentials, a current is
`then generated from the reaction of H;i02 (generated from extracted glucose) which
`has diffused to the platinum sensor electrode. The measured current is proportional
`to the amount ofH202, and hence, extracted glucose.
`The current (mA) utilized in iontophoresis potentially interferes with detection of
`the low current (nA) generated at each electrochemical sensor. Consequently, the
`iontophoretic and sensing electrodes are not activated at the same time. Instead, a
`typical situation is to have the iontophoresis proceed for about three minutes to
`collect an adequate amount of glucose. During this period about lOng of glucose is
`typically extracted at the cathode.
`Iontophoresis is then stopped and the sensing
`electrodes are activated for typically seven minutes. This period of seven minutes is
`chosen so that all of the glucose (both a. and p) has been converted to Hi02, and that
`aH of the hydrogen peroxide has diffused to the platinum electrode, and subsequently
`oxidized, to generate a current. Thus, all extracted glucose and H:P2 are consumed
`during this cycle. The integrated current (or charge) over this seven minute interval
`is then proportional to the total amount of glucose that entered the hydrogel during
`the iontophoresis interval. The iontophoresis polarity is reversed and cycle is then
`repeated ris1. Thus, if sensor A is at the cathode during the first cycle, sensor B is the
`cathode during the second cycle. The combined cycle requires 20 minutes, and the
`
`Park and Mrsny; Controlled Drug Delivery
`ACS Symposium Series; American Chemical Society: Washington, DC, 2000.
`
`AGAMATRIX, INC. EXHIBIT NO. 1026
`Page 5 of 10
`
`

`

`278
`
`combined cathode sensor charge (A + B) is a measure of the glucose extracted. This
`20-minute cycle is repeated throughout operation of the Glucowatch biographer.
`An example of the sensor current vs. elapsed time during operation is shown in
`Figure 2. These results show that during iontophoresis the sensor current is zero.
`When the sensor circuits are then activated, H201 (converted from glucose) reacts
`with the platinum electrode to produce a current, which monotonically declines over
`the seven-minute detection cycle. During the initial period of sensor operation (t <90
`seconds), the current declines as r112, as predicted for diffusion to a planar electrode
`from a semi-infinite reservoir 1121. Note that current is generated at both electrodes,
`even though glucose is primarily collected at the cathode. The anode signal is due to
`ascorbic and uric acids, which migrate solely to the anode. Ascorbate and urate are
`known to react directly with a platinum electrode and produce a signal that interferes
`with conventional blood glucose monitoring devices. During iontophoresis these
`anions collect only at the anodet while glucose is found primarily at the cathode.
`Hence, the unique ion selective nature of the GlucoWatch biographer prevents the
`interference of the electroactive species in the measurement of iontophoretically­
`extracted glucose.
`
`Clinical Results
`
`Biographers were applied to the lower foreann of human subjects with diabetes
`requiring insulin injection. Subjects included Type l and 2 diabetics using insulin.
`All subjects were 18 years of age, or older, and consisted of both males and females
`from a broad ethnic cross-section. As many as three Gluco Watch biographer
`measurements were obtained per hour. In addition, subjects obtained two capillary
`blood samples per hour, and the glucose concentration was determined using a
`Hemocue® Blood Analyzer (HemoCue Inc., Mission Viejo, CA). The blood glucose
`measurement obtained at three hours was used as a single point calibration. This
`calibration value was used to calculate the extracted blood glucose for an subsequent
`GlucoWatch biographer measurements. Measurements were continued for 12 hours.
`yielding a maximum of23 paired measurements (not including the calibration point)
`comparing the GlucoWatch biographer and blood glucose values.
`Results obtained with one subject are shown in Figure 3. These results show
`close tracking of GlucoWatch biographer and blood glucose values throughout the
`study. An analysis of the results for 46 watches, with 897 paired data point is shown
`in the Table. These results show that the GlucoWatch biographer yields a mean
`absolute error (MAE =absolute value of [Biographer gJucose - Blood gJucose]/Blood
`glucose) of 15.6%. In addition, the paired data yielded a correlation coefficient of
`0.89. Finally, 96% of the data lie in the therapeutically relevant A+B region of the
`error grid analysis t16l. The values obtained in this study are similar to those obtained
`for conventional blood glucose measuring devices 1171.
`
`Park and Mrsny; Controlled Drug Delivery
`ACS Symposium Series; American Chemical Society: Washington, DC, 2000.
`
`AGAMATRIX, INC. EXHIBIT NO. 1026
`Page 6 of 10
`
`

`

`279
`
`urrent A {nA)
`
`CurrentB (nA)
`
`350
`
`300
`
`< =
`
`.� 200
`ISO ·
`
`250
`
`-';i
`
`rn
`
`ll
`t
`a
`
`-=
`
`100
`
`50
`
`0
`2:45
`
`3:15
`
`3:45
`
`4:1 s
`
`4:45
`
`Elapsed Time (b :mm)
`
`Figure 1. The current.from both sensors (alternative cathode and anode) during two
`hours of operation on a human subject.
`
`Park and Mrsny; Controlled Drug Delivery
`ACS Symposium Series; American Chemical Society: Washington, DC, 2000.
`
`AGAMATRIX, INC. EXHIBIT NO. 1026
`Page 7 of 10
`
`

`

`280
`
`-+-GlucoWatch
`
`-�,�-BG
`
`300
`
`250
`
`�
`
`"'
`
`3:00
`
`6;00
`
`9:00
`
`12:00
`
`15:00
`
`Elapsed Time (h:mm)
`
`:3
`t 200
`El
`
`en
`0
`c,,,
`::s
`
`-IV
`
`G
`
`150
`
`100
`
`50
`
`0
`0:00
`
`Figure 3. Glucose concentration vs. elaspsed time for one subject as measured by the
`GlucoWatch biographer and the reference blood (BG) method.
`
`Park and Mrsny; Controlled Drug Delivery
`ACS Symposium Series; American Chemical Society: Washington, DC, 2000.
`
`AGAMATRIX, INC. EXHIBIT NO. 1026
`Page 8 of 10
`
`

`

`281
`
`Conclusion
`
`In this study, the GlucoWatch biographer yields continuous measurements of
`glucose (3/hr) over a 12-hour period with accuracy and precision similar to existing,
`single-point blood measuring device. This non-invasive device holds promise to
`provide frequent glucose measurements to better guide insulin administration in
`diabetic subjects, and improved disease management.
`
`Table. The Statistical Summary for GlucoWatch Biographer Results from 46
`Subjects.
`
`MAE
`(%)
`15.6
`
`Correlation
`Coefficient
`0.89
`
`Error Grid Analysis
`A + B region (%)
`96
`
`References
`
`1. The Diabetes Control and Complication Trial Research Group New England J.
`1993, 329, 977.
`Medicine
`
`2. UK Prospective Diabetes Study (UKPDS) Group Lancet 1998, 352. 837-853.
`
`3. Ohkubo, Y.; Kishikawa, H.; Araki, E.; Miyata. T.; Isami, S.; Motoyoshi, S.;
`Kojima, Y.; Furuyoshi, N.; Shichiri, M. Diabetes Research & Clinical
`
`Practice
`1995, 28, 103* 17.
`
`4. Meyerhoff, C.; Bischof, F.; Sternberg, F.; Zier, E.; Pfeiffer, F. Diabetologia
`1992, 35, I087-1092.
`
`
`
`5. Moatti-Sirat, D.; et al. Diabetologia 1992, 35. 224-230.
`
`6. Arnold. M.A. Current Opinions in Biotechnology 1996, 7, 46-49.
`
`7.
`
`Jagemann, K.; Fischbacher, C.; Danzer, K.; Muller, U. A.; Mertes, B. Zietschrift
`fur Physicalische Chemie 1995, 191, 179-190.
`
`8. Pikal, M.J. Adv. Drug Del. Rev. 1992, 9, 201.
`
`9. Dinh, S. M.; Luo, C.-W.; Bemer, B. A/ChEJournal 1993, 39, 2011.
`
`10. Glikfeld, P.; Hinz, R. S.; Guy, R. H. Pharm. Res. 1989, 6, 988-990.
`
`Park and Mrsny; Controlled Drug Delivery
`ACS Symposium Series; American Chemical Society: Washington, DC, 2000.
`
`AGAMATRIX, INC. EXHIBIT NO. 1026
`Page 9 of 10
`
`

`

`282
`
`11. Tamada, J. A.; Bohannon, N. J. V.; Potts, R. 0. Nature Medicine 1995, I, 1198.
`
`12. Kurnik, R.T.; Bemert B.; Tamada, J.A.; Potts, R.O. J. Electrochem. Soc 1998,
`145, 4119.
`
`13. Fessenden, R. J.; Fessenden, J. S. Organic Chemistry; Willard Grant Press:
`Boston, MA, 1979.
`
`14. Morrison, R. T.; Boyd, R. N. Organic Chemistry; Allyn and Bacon, Boston, MA,
`1973.
`
`15. Kurnik, R. T.; Potts, R. O.; Tamada, J. A.; Tierney, M. J. PCT W097/24059,
`Cygnus Inc., Redwood City, CA; July JO, 1997.
`
`16. Clarke, W. L.; Cox, D. C.; Conder-Frederick, L. A.; Carter, W.; Pohl, S. L.,
`Diabetes Care 1987, 10, 622.
`
`Park and Mrsny; Controlled Drug Delivery
`ACS Symposium Series; American Chemical Society: Washington, DC, 2000.
`
`AGAMATRIX, INC. EXHIBIT NO. 1026
`Page 10 of 10
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket