Journal of Diabetes Science and Technology
` Volume 4, Issue 3, May 2010
` © Diabetes Technology Society
`
`SYMPOSIUM
`
`Practical Aspects of Insulin Pen Devices
`Teresa L. Pearson, M.S., R.N., CDE
`
`Abstract
`Insulin pen devices have several advantages over the traditional vial-and-syringe method of insulin delivery,
`including improved patient satisfaction and adherence, greater ease of use, superior accuracy for delivering
`small doses of insulin, greater social acceptability, and less reported injection pain. In recent years, pens have
`become increasingly user-friendly, and some models are highly intuitive to use, requiring little or no
`instruction. Despite this progress, uptake of these devices in the United States has not matched that in many
`other areas of the world. There is a need for improved awareness of the current characteristics of insulin
`pen devices among United States health care professionals. Knowledge of the design improvements that have
`been incorporated into pens, both to address patient needs and as a result of the improved technology behind the
`device mechanics, is essential to promoting the use of insulin pen devices. This review highlights some of the
`practical aspects of pen use and discusses the factors to be considered when selecting among different insulin pens.
`J Diabetes Sci Technol 2010;4(3):522-531
`
`Introduction
`
`Numerous studies have shown that insulin pen
`
`devices have several advantages over the traditional
`vial-and-syringe method of insulin delivery, including
`improved patient satisfaction and adherence, greater ease
`of use, and superior dosing accuracy.1–7 Despite these
`advantages, the use of insulin pen devices in the United
`States remains low compared with other developed
`countries.8 About two-thirds of insulin prescriptions in
`Europe and about three-quarters in Japan are for pen
`devices.9 In contrast, in the United States, only 15% of
`patients are thought to use insulin pens.10
`
`Possible reasons for the low adoption rates in the United
`States include lack of awareness among health care
`providers of the advantages of pens compared with the
`vial and syringe.8,11 An additional issue is the greater
`
`prescription cost of insulin cartridges and prefilled insulin
`pens compared with insulin vials, although the cost to the
`patient may be the same depending on their coverage;
`in fact, if they have one copay per box of pens, the cost
`to the patient may actually be less per unit of insulin.
`It should be noted, however, that despite the higher
`unit cost of insulin in pen devices versus vials, several
`studies have found that overall diabetes-related treatment
`costs are lower with pen devices than with vial and
`syringe.1,2,12 In addition, most pen devices have good
`formulary coverage. For example, the FlexPen® prefilled
`pen is covered on more than 90% of managed care plans.13
`Therefore, in theory, costs should not prevent the use of
`these devices.14 However, many smaller health maintenance
`organizations and state Medicaid plans may require prior
`authorization for insulin pens.
`
`Author Affiliation: Fairview Health Services, Minneapolis, Minnesota
`Keywords: diabetes, insulin, insulin pen, practical use
`Corresponding Author: Teresa L. Pearson, Director, Diabetes Care, Fairview Health Services, Minneapolis, MN 55436; email address
`tpearson9@comcast.net
`
`522
`
`Mylan Ex. 1045
`Mylan v. Sanofi - IPR2018-01676
`
`

`

`Practical Aspects of Insulin Pen Devices
`
`Pearson
`
`Given the clinical, practical, and potential health economic
`advantages of insulin pens, there is a need for improved
`awareness of insulin pen devices among U.S. health care
`professionals. A study of patients with type 2 diabetes
`in the United States found that the physician’s role in
`presenting the pen as an option and recommending
`pen use was a critical factor in patients’ use of pens.15
`Physicians, nurses, and diabetes educators should therefore
`become familiar with the various insulin delivery devices
`available so that they can discuss the potential benefits
`of these devices with patients and offer advice on
`which device best meets an individual patient’s needs.
`Nurses, diabetes educators, and pharmacists have a
`particularly important role in educating patients on how to
`use insulin pens because incorrect use can affect pen
`performance and thus the accuracy of the administered
`dose.16 This review describes the practical aspects of
`insulin pen devices,
`including considerations when
`selecting among different insulin pens.
`Advantages of Pen Devices Compared
`with Vial and Syringes
`The traditional route of subcutaneous insulin adminis-
`tration has been the vial and syringe. However, this
`method of administration has many disadvantages,
`including fear of injections, poor dose accuracy, lack of
`social acceptance, lengthy training time, and difficulty of
`transportation. These factors can all act as barriers to
`insulin therapy, impacting on lifestyle flexibility and
`negatively influencing treatment adherence, patient self-
`management behavior, and achievement of euglycemia.17,18
`Insulin pens have been developed to help address
`these issues, with resulting improvements in portability,
`dosing accuracy, mealtime flexibility, and convenience
`of delivery.4–6,19 Increased patient preference, treatment
`satisfaction, and quality of life have been reported for
`pen devices compared with the vial and syringe;3,7,20,21
`these benefits may be particularly important due to their
`demonstrated impact on patient adherence. Other studies
`have shown that pen devices are associated with improved
`costs of care, less reported injection pain, and improved
`patient self-management behaviors, including adherence
`to treatment, compared with the vial and syringe.1,22–24
`Because of the greater ease of use of insulin pens,
`people with visual impairment or reduced dexterity may
`especially benefit from using an insulin pen rather than
`a vial and syringe.
`insulin analog vials, cartridges,
`Once
`in use, most
`and prefilled pens must be discarded after 28 days.
`This means that many patients who use a 10-ml vial end
`up either wasting insulin or using insulin beyond its
`
`recommended discard date. This is rarely a problem for
`patients using either a 3-ml prefilled pen or a reusable
`pen containing a 3-ml insulin cartridge. Exceptions to
`the 28-day discard date are insulin detemir (vials and
`pens), which can be kept for up to 42 days once in use;
`biphasic insulin aspart 70/30 prefilled pens, which must
`be discarded after 14 days; and cartridges and prefilled
`pens containing biphasic insulin lispro premixes, which
`must be discarded after 10 days.
`
`Two open-label, randomized, crossover studies have
`examined whether patients have greater dose confidence
`with a prefilled insulin pen or with a vial and syringe;
`both studies found that patients preferred the pen
`over the vial and syringe in this regard.3,25 In the first
`of these studies, 73% of patients felt more confident in
`the accuracy of the insulin dose delivered with the pen
`(original FlexPen) compared with 19% for the vial/syringe.3
`In the other study, 88% of patients had greater confidence
`that they were taking the right dose with the pen
`(KwikPen®) than with the vial and syringe.25 A more
`detailed discussion on the benefits of pen devices over
`vial and syringe can be found in the article by Selam26
`in this issue of Journal of Diabetes Science and Technology.
`Available Insulin Pen Devices
`Two types of insulin pen are available: prefilled disposable
`(Table 1).27–44 Most
`pens and refillable pens
`insulin
`pens are proprietary devices, manufactured by Eli Lilly,
`Novo Nordisk, and sanofi-aventis, and are designed to
`work with specific insulins from the same manufacturer.
`All currently available pens either are prefilled with 3 ml
`(300 units) of insulin or are refillable pens designed for use
`with 3-ml insulin cartridges (pens taking 1.5 ml insulin
`were formerly available). Insulin cartridges or prefilled
`disposable pens are available for all insulin analogs
`(rapid-acting, long-acting, and premixed) and for most
`human insulins (Table 1).
`
`For all insulin pen devices, a separate prescription for
`pen needles is required. Pen needles are available from
`various manufacturers (Allison Medical, BD, Can-Am Care,
`Delta Hi-Tech, Medical Plastic Devices, Novo Nordisk,
`Owen Mumford, UltiMed) and come in gauges ranging
`from 29 to 32 gauge and in lengths from 5 to 12.7 mm.45
`More recent developments have resulted in the intro-
`duction of safety needles with protective shields that
`not only reduce needle-stick injuries but may also allay
`patient anxieties about needle use.46 These are discussed
`in more detail in the section entitled “Individualizing
`Insulin Treatment with Pen Devices.”
`
`J Diabetes Sci Technol Vol 4, Issue 3, May 2010
`
`523
`
`www.journalofdst.org
`
`Mylan Ex. 1045
`Mylan v. Sanofi - IPR2018-01676
`
`

`

`Practical Aspects of Insulin Pen Devices
`
`Pearson
`
`Table 1.
`Insulin Pen Devices Currently Available in the United States.a, 27–44
`Dose range
`Increment
`Recommended
`needle typeb
`(unit)
`(unit)
`
`Product name
`
`Insulin type
`
`Dialing feature
`
`Other features
`
`Prefilled disposable insulin pens
`
`Humalog®
`KwikPen®
`
`Original
`Humalog® pen
`
`Humalog®
`Mix75/25™ pen
`Humalog®
`Mix50/50™ pen
`
`Humulin® N pen
`
`Humulin® 70/30 pen
`
`Maximum
`dose
`60 units
`
`Maximum
`dose
`60 units
`
`Insulin lispro
`and insulin
`lispro protamine
`suspension
`
`Insulin lispro
`
`Insulin lispro
`protamine
`suspension/insulin
`lispro injection
`
`Human insulin
`isophane
`suspension
`
`Maximum
`dose
`60 units
`
`Human insulin
`isophane
`suspension/human
`insulin injection
`
`Lantus® SoloSTAR®
`
`Insulin glargine
`
`Apidra® SoloSTAR®
`
`Insulin glulisine
`
`Levemir®FlexPen®c
`
`Insulin levemir
`Insulin aspart
`
`Novolog® FlexPen®c
`Novolog® Mix 70/30
`FlexPen®c
`
`Insulin aspart
`protamine
`suspension/insulin
`aspart
`
`Maximum
`dose
`80 units
`
`Maximum
`dose
`60 units
`
`1
`
`1
`
`1
`
`1
`
`1
`
`BD Ultra-
`Fine™ needles
`
`BD Ultra-Fine
`needles
`
`BD Ultra-Fine
`needles
`
`BD Ultra-Fine
`needles
`
`Dial-back feature
`Cannot dial more than
`the amount of insulin
`remaining
`
`Dial-back feature
`After full dose is
`delivered, an arrow or
`diamond is centered
`in dose window
`Cannot dial more than
`the amount of insulin
`remaining
`
`Dial-back feature
`After full dose is
`delivered an arrow or
`diamond is centered
`in the dose window
`Cannot dial more than
`the amount of insulin
`remaining
`
`Dial-back feature
`Cannot dial more than
`the amount of insulin
`remaining
`
`NovoFine or
`NovoTwist
`needles
`
`Dial-back feature
`Cannot dial more than
`the amount of insulin
`remaining
`
`Refillable pens
`
`Autopen® classic
`
`Insulin lispro
`protamine
`suspension/insulin
`lispro injection
`
`Maximum
`dose
`21 or 42
`units
`
`1 or
`2 unit
`models
`available
`
`Compatible
`with all pen
`needles
`
`Dial back not possible
`
`Autopen 24®
`
`Insulin glargine
`Insulin glulisine
`
`HumaPen®
`LUXURA™ and
`LUXURA™ HDd
`
`Insulin lispro
`Insulin lispro
`protamine
`suspension/insulin
`lispro injection
`
`Maximum
`dose
`21 or 42
`units
`
`Maximum
`dose
`60 units
`or 30 units
`(HD)
`
`1 or
`2 unit
`models
`available
`
`1 or 0.5
`(HD)
`(minimum
`dose 1
`unit)
`
`Compatible
`with all pen
`needles
`
`Dial back not possible
`
`BD Ultra-Fine
`needles
`
`Dial-back feature
`
`Light and portable
`
`The two pens are
`distinguished by
`color, and the Apidra
`SoloSTAR has a raised
`dot on the push button
`
`Little force required,
`dose delivery is
`confirmed by audible
`click, different insulins
`color coded, only pen
`with all three types of
`insulin analogs
`Raised dot on push
`button
`
`Side-mounted release
`button that is pushed
`the same minimal
`distance to inject,
`regardless of the dose
`size and the needle
`gauge
`
`The two dosing models
`are distinguished by
`color
`
`Relatively heavy; long
`reach needed for large
`doses
`
`4
`Continued
`
`J Diabetes Sci Technol Vol 4, Issue 3, May 2010
`
`524
`
`www.journalofdst.org
`
`Mylan Ex. 1045
`Mylan v. Sanofi - IPR2018-01676
`
`

`

`Practical Aspects of Insulin Pen Devices
`
`Pearson
`
`Table 1. Continued
`
`I
`
`1
`
`Dose range
`(unit)
`
`Insulin type
`
`1
`
`Increment
`(unit)
`
`1
`
`1
`
`I
`
`Other features
`
`Dialing feature
`
`Recommended
`needle typeb
`
`Product name
`
`Refillable pens
`
`HumaPen® Memoir™
`
`NovoPen® 3
`
`As HumaPen
`LUXURA
`
`Maximum
`dose
`60 units
`
`Insulin aspart
`Insulin levemir
`Insulin aspart
`protamine
`suspension/insulin
`aspart
`Human insulin
`isophane suspension
`
`1
`
`1
`
`1
`
`BD Ultra-Fine
`needles
`
`Dial-back feature
`
`NovoFine
`needles
`
`NovoPen 3 dial
`back requires partial
`disassembly
`
`Digital display and
`memory of last
`16 doses, including
`priming doses
`(i.e., eight injections)
`Relatively heavy; long
`reach needed for large
`doses
`
`NovoFine
`needles
`
`NovoPen 4 has easy
`dial back and cannot
`dial more than the
`amount of insulin
`remaining
`
`Large easier-to-read
`numbers, reduced dose
`force, dose delivery
`confirmed by audible
`click
`
`0.5
`(minimum
`dose 1
`unit)
`
`1
`
`NovoFine
`needles
`
`Dial back requires
`partial disassembly
`
`Ypsomed
`Clickfine™
`needles or
`BD Ultra-Fine
`needles
`
`Dial-back feature
`Cannot dial more than
`the amount of insulin
`remaining
`
`Distinguished from
`NovoPen 3 by raised
`circle on push button
`
`Dose displayed for
`2 minutes
`
`Maximum
`dose
`70 units
`
`Maximum
`dose 60
`units
`
`Maximum
`dose 35
`units
`
`Maximum
`dose, 80
`units
`
`NovoPen® 4
`
`As NovoPen 3
`
`NovoPen® Junior
`
`As NovoPen 3
`
`OptiClik®
`
`Insulin glargine
`Insulin glulisine
`
`a Cartridge delivery capacity is 300 units, unless otherwise stated.
`b Needle types shown are recommended by the insulin pen device manufacturer. BD Ultra-Fine and Ypsomed Clickfine needles also fit all
`current insulin pens.
`c Also known as the improved FlexPen.
`d Half dose.
`
`How to Use an Insulin Pen
`In a study assessing patient and physician acceptability
`of the prefilled Humulin®/Humalog® insulin pen device,
`88% of the 33 physicians who completed questionnaires
`at the end of the study thought that it took less time
`to teach patients to use the pen and 73% thought that
`it took less time to initiate insulin therapy with the pen
`compared with a vial and syringe.47 Teaching patients
`how to use an insulin pen can be summarized in the
`five main steps shown in Figure 1. It should be noted
`that because of the mechanics of pen devices, insulin can
`still be flowing out of the pen for several seconds after
`the button is fully depressed.11 Patients must therefore
`keep the device in place with the button pressed in
`for 5–10 seconds (the stipulated time varies between
`
`package inserts of the various insulin pens). For example,
`SoloSTAR® recommends a longer in situ time (10 seconds)
`compared with the FlexPen (6 seconds).37,38,41,42 The easiest
`way to ensure this is to instruct the patient to count to 5
`(or 10, if using the SoloSTAR) before removing the needle.
`If the patient is using more than 50 units of insulin per
`dose, a good rule of thumb might be to instruct them to
`count to 10 regardless of the pen they are using to ensure
`complete absorption of the insulin.
`
`If patients are using a pen that contains an insulin
`suspension (neutral protamine Hagedorn insulin or an
`insulin premix), they must carefully roll or tip the pen
`for the recommended number of times according to the
`
`J Diabetes Sci Technol Vol 4, Issue 3, May 2010
`
`525
`
`www.journalofdst.org
`
`Mylan Ex. 1045
`Mylan v. Sanofi - IPR2018-01676
`
`

`

`Practical Aspects of Insulin Pen Devices
`
`Pearson
`
`1. Attach the needle to the pen
`
`1
`
`2. Prime the pen (also referred to as "doing a safety test"
`or "performing an air shot")a
`
`3. Select the dose by dialing
`
`4. Inject the dose by depressing the buttonb
`
`5. After the injection, remove the needle from the pen
`
`Figure 1. The five steps of insulin pen use. aThis is performed by
`instructing the patient to dial up 2 units and to inject these units into
`the air. bThe button needs to be pressed and the needle held in the
`skin for 5–10 seconds to ensure complete delivery of insulin dose.
`The easiest way to ensure this is to instruct the patient to count to 5
`(or 10, if using the SoloSTAR®) before removing the needle.
`
`package insert to ensure even mixing of the insulin
`suspension before attaching the needle. This issue may
`be especially important for patients who have used the
`vial-and-syringe method, as vials have a greater diameter
`than cartridges and so need to be tipped less often.48
`
`Pens must be primed before each injection, and the needle
`removed immediately after each use.11 This is performed
`by instructing the patient to dial up 2 units and inject these
`units into the air (also called an “air shot”). This will
`displace any air in the needle and ensure an accurate
`injection. This air shot may need to be repeated when
`using a new pen or cartridge until a steady stream of
`insulin is observed. Insulin pens are manufactured
`with enough extra insulin to account for this air shot.
`An insulin pen must never be used by more than one
`individual, even if the pen needle is changed, because
`sharing of insulin pens can result in the transmission
`of hepatitis viruses, human immunodeficiency virus, or
`other blood-borne pathogens.49
`
`Patients and health care workers also need to be aware
`of the different procedures associated with insulin pens
`in case of accidentally dialing a dose larger than required.
`In the case of the Autopen®, the side push-button design
`makes it impossible to dial back, and care should be
`taken not to dial past the required dose. With older
`versions of NovoPen® and NovoPen® Junior, the dose
`can be adjusted back down by partial disassembly of
`
`the pen. This involves pulling the mechanical section
`and the cartridge holder apart and pressing the dial-
`up button back to zero. With other pens, namely the
`SoloSTAR, the improved FlexPen, Humalog KwikPen,
`original Humalog/Humulin pens, HumaPen® LUXURA™,
`HumaPen® LUXURA™ HD, HumaPen® Memoir™,
`OptiClik®, and NovoPen 4, easy dial back is possible by
`simply reversing the dial-up action.
`
`Prior to first use, the insulin cartridge or pen should be
`stored in the refrigerator. The pen should be warmed to
`room temperature [below 86°F (30°C) for most insulin
`analogs] before use. After use, the pen should remain
`at room temperature below 86°F (30°C) in order to avoid
`producing air bubbles, which can form when the pen
`mechanism and the insulin expand/contract during a
`temperature change. As with all types of insulin, pens
`in use should be kept from extremes in temperature,
`keeping them as close as possible to room temperature
`below 86°F (30°C) at all times. Insulin glulisine has a
`narrower temperature range for storage than the other
`insulin analogs: Once in use, insulin glulisine must be
`stored below 77°F (25°C). In some buildings, for example,
`schools, the air conditioning is turned off at night,
`which may result in the room temperature rising above
`77°F (25°C) or 86°F (30°C). Insulated storage packs are
`recommended in such conditions.
`
`If a patient is switching from one type of insulin pen to
`another, it is important to check whether the procedure
`used for the previous pen also applies to the new pen.50
`Individualizing Insulin Treatment with
`Pen Devices
`Health care practitioners should work with the patient to
`select insulin delivery devices that are compatible with
`their insulin regimen, lifestyle, and personal preferences.
`A regimen that causes the least disruption to the patient’s
`day-to-day life is much more likely to be used. Pens are
`more than just a matter of convenience, though; their ease
`of use allows patients to take better care of their own
`condition.15 As discussed by Selam26 and summarized
`earlier, insulin pens can provide many potential benefits
`over vial-and-syringe delivery.
`
`Patients across all age groups often have concerns regarding
`insulin therapy, and many of these concerns can be
`addressed effectively through choosing an insulin pen
`device rather than a vial and syringe.51 In particular,
`adolescents and children may find insulin pens more
`socially acceptable because of the pens’ greater portability
`
`J Diabetes Sci Technol Vol 4, Issue 3, May 2010
`
`526
`
`www.journalofdst.org
`
`Mylan Ex. 1045
`Mylan v. Sanofi - IPR2018-01676
`
`

`

`Practical Aspects of Insulin Pen Devices
`
`Pearson
`
`and discreetness. The NovoPen Junior, for example, has
`been developed specifically for use by children and
`teenagers or others requiring the ability to adjust their
`insulin in half-unit increments.33 The HumaPen LUXURA
`HD can also deliver insulin in half-unit increments.
`However, both these pens administer a minimum dose
`of 1 unit, meaning that no pen is yet available that can
`deliver a dose of one-half unit. This can be problematic
`for young children, especially at school. With proper
`training, loading insulin and attaching the needle should
`be simple enough for grade school-aged kids.
`
`Numbers on insulin pen dials are also larger than those
`on syringes, making it easier for those with visual
`impairment to select the correct dose. Some devices also
`incorporate audible clicks that notify the user of the
`number of units injected. Patients with impaired manual
`dexterity may find an insulin pen easier to use because
`it eliminates a step in the injection procedure, i.e., filling
`the syringe with insulin from the vial.
`
`Advances in pen devices have also been made to improve
`needle safety and potentially reduce any needle anxiety.
`Needle-stick injuries are a common occupational hazard
`for health care professionals, particularly nurses. Use of
`the NovoFine® Autocover® safety pen needle has been
`shown to reduce the incidence of needle-stick injuries
`among nurses.46 Although safety needles are not readily
`available outside the hospital setting, they may be a
`consideration for secondary caregivers to avoid needle-
`stick injuries. The other currently available safety pen
`needle is the BD AutoShield™. Both the NovoFine
`Autocover and the BD AutoShield conceal the needle, thus
`also potentially reducing needle anxiety. Two injection
`aids for insulin pens are available that also conceal
`the needle: NeedleAid™ and NovoPen 3 PenMate®.45
`Concealing the needle using the NovoPen 3 PenMate has
`been shown to reduce pain perception.52 The NeedleAid
`is an attachment designed to help visually impaired
`patients self-administer insulin.
`
`Insulin pens are not without their limitations, and it is
`important that patients and health care workers are
`aware of these to ensure maximal outcomes. The maximum
`dose with most insulin pens is 60 to 80 units, but with
`a syringe it is 100 units. Patients cannot mix their own
`insulin formulations for use in a single injection given
`by an insulin pen. Despite their ease of use, pens are
`mechanically more complex than syringes, and some cases
`of malfunction have been reported in the literature.53,54
`Therefore, patients using an insulin pen should have a
`backup pen available if traveling.
`
`In terms of costs, there should be few reimbursement
`issues with most private health plans,9 although coverage
`by third-party payers may be under a tier that does
`not pay out in full.11 However, costs to the patient for
`prefilled pens are minimized, as the pack dispensed by
`the pharmacy under one copayment contains five pens.11
`Most pens are now covered by Medicare.9,14 Some pens
`are associated with specific reimbursement issues.11
`Considerations When Choosing among
`Insulin Pens
`In U.S. medical practice, the choice of insulin pen will be,
`to a large extent, determined by the choice of insulin,
`as particular insulins are specific to certain makes of insulin
`pen. Anecdotal reports suggest that many patients prefer
`prefilled disposable pens to refillable pens because
`disposable pens are typically lighter and smaller and are
`also simpler to use, as there is no requirement to load
`new insulin cartridges.55 However, certain refillable pens
`have features, such as a memory function or the ability
`to dial in half-unit increments, that are not available
`with prefilled pens (Table 1). For example, the HumaPen
`LUXURA HD and NovoPen Junior are the only pens
`that allow the dose to be selected in half-unit increments.
`This may be important in children or in those sensitive
`to insulin. The OptiClik and SoloSTAR pens have a
`larger maximum dose (80 units) than the other insulin
`pens and therefore may be preferable in patients who
`take large doses of insulin.
`
`The recently introduced improved FlexPen has a push-
`button mechanism that has been modified to reduce
`injection force while maintaining dose accuracy56,57 and
`retaining the ability to dial back. It therefore provides
`significantly less discomfort when injecting, which may
`be particularly suitable for patients with impaired manual
`dexterity or conditions such as arthritis. The improved
`FlexPen also has color-coded cartridges and packaging that
`clearly differentiate among insulin types.19 Color coding
`has also been incorporated into the SoloSTAR Lantus
`and Apidra pens to distinguish among insulin types.
`
`Another initiative is for pens to supply auditory feedback,
`particularly since patients with type 2 diabetes often suffer
`from impairments in vision as well as manual dexterity.58
`The OptiClik, NovoPen 4, and improved FlexPen provide a
`confirmatory click when the correct dose has been delivered.
`In addition, the recently introduced NovoTwist® needle,
`for use with the improved FlexPen, has an easy twist-on
`action with an audible click when the needle is in place.
`
`J Diabetes Sci Technol Vol 4, Issue 3, May 2010
`
`527
`
`www.journalofdst.org
`
`Mylan Ex. 1045
`Mylan v. Sanofi - IPR2018-01676
`
`

`

`Practical Aspects of Insulin Pen Devices
`
`Pearson
`
`Both the OptiClik and the HumaPen Memoir devices
`have liquid crystal display screens powered by non-
`replaceable batteries. Therefore, when the battery runs
`out, patients have no visual confirmation of the dose
`selected, and the pen must be replaced. However, the
`OptiClik pen can be obtained only through a physician’s
`office, making replacement more difficult. According to
`the manufacturers, the expected battery life for each of
`these pens is 3 years.
`Clinical Studies Comparing Insulin Pens
`Several studies have investigated dosing accuracy among
`pens. Generally, dosing accuracy is good, but the OptiClik
`pen can be subject to underdosing.59,60 One study found
`that, because of underdosing, 6.8% of 10-unit doses and
`13.9% of 30-unit doses with the OptiClik pen were outside
`the International Organization for Standardization limits
`(±1 unit for a 10-unit dose, and ±1.5 units for a 30-unit
`dose).59 Another separate study found that OptiClik
`underdosed for 17.1% of doses at 10 units and for 28.9%
`of doses at 30 units.60 To avoid this, it may be necessary
`to carry out multiple priming before injecting with the
`OptiClik pen. Other studies have examined the dosing
`accuracy of SoloSTAR compared with FlexPen (improved
`and original), Humulin/Humalog pen, and OptiClik
`pen59,61–64 (Table 2). The largest of these comparative studies
`investigated 2280 doses of the SoloSTAR and the original
`FlexPen devices and showed that pens had comparable
`accuracy over the standard doses of 5, 10, and 30 units.53
`
`Another study also found that SoloSTAR and the
`original FlexPen were similarly accurate when used
`by device-naive individuals to deliver 20-unit doses of
`insulin.65 One study compared the improved FlexPen
`with SoloSTAR: both pens showed very good accuracy,
`and the improved FlexPen was even more accurate than
`the SoloSTAR (p < 0.05)64(Table 2).
`
`Insulin pens also differ in the force required to inject
`an insulin dose, and this feature has been investigated
`in several studies,56,63,66–68 with results summarized in
`Table 3. In general, differences in the injection force
`among insulin pens are relatively small; however, the
`improved FlexPen does seem to show benefits when
`compared with SoloSTAR and OptiClik pens.56,67
`
`Data on ease of use and patient preference for different
`types of pen have been assessed in a number of clinical
`studies and in clinical practice. Many open-label studies
`have obtained information on patient preference, and
`results show that newer designs of pens are increasingly
`user-friendly and intuitive to use, requiring little or no
`instruction.54,69–71 For example, an open-label, crossover
`study showed that the improved FlexPen was faster to
`teach, simpler to use, and more trusted by patients
`compared with OptiClik.69 Similarly, in an open-label,
`crossover study, the Novolog® Mix 70/30 FlexPen was
`associated with significantly greater intuitiveness and a
`shorter injection time compared with the HumaPen
`LUXURA device.70 A separate study also showed a
`
`Table 2.
`Studies of Dose Accuracy of Insulin Pens with Comparable End Points
`Improved
`HumaPen
`FlexPen
`LUXURA
`
`Original FlexPena
`
`SoloSTAR
`
`OptiClik
`
`Reference
`
`Deviation from
`10-IU dose (mean)
`
`Deviation from
`30-IU dose (mean)
`
`Mean delivered dose
`of 5 IU
`
`Mean delivered dose
`of 10 IU
`
`Mean delivered dose
`of 30 IU
`
`1.64 ± 0.84%
`
`NAb
`
`1.10 ± 0.20%
`
`2.61 ± 0.92%
`
`4.78 ± 3.31%
`
`NA
`
`1.63 ± 0.84%c
`
`NA
`
`2.11 ± 0.92%
`
`NA
`
`0.83 ± 0.26%
`
`NA
`
`0.62 ± 0.19%
`
`1.70 ± 0.84%
`
`2.97 ± 2.48%
`
`NA
`
`1.23 ± 0.76%d
`
`5.07 ± 0.15
`
`4.95 ± 0.19
`
`9.87 ± 0.16
`
`9.61 ± 0.27
`
`29.70 ± 0.38
`
`29.70 ± 0.34
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`1.54 ± 0.84%
`
`5.03 ± 0.21
`
`4.86 ± 0.39
`
`9.83 ± 0.14
`
`9.27 ± 0.52
`
`29.45 ± 0.25
`
`28.73 ± 0.47
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`NA
`
`59
`
`64
`
`59
`
`64
`
`61
`
`62
`
`61
`
`62
`
`61
`
`62
`
`a No longer available commercially.
`b Not assessed.
`c p < 0.001
`d p < 0.05 for improved FlexPen vs SoloSTAR.
`
`J Diabetes Sci Technol Vol 4, Issue 3, May 2010
`
`528
`
`www.journalofdst.org
`
`Mylan Ex. 1045
`Mylan v. Sanofi - IPR2018-01676
`
`

`

`Practical Aspects of Insulin Pen Devices
`
`Pearson
`
`Reference
`
`KwikPen
`
`Table 3.
`Studies of Injection Force of Insulin Pen Devices: FlexPen, Lilly Pen, SoloSTAR, and KwikPen
`Injection speed
`Original
`Improved
`FlexPena
`mm/s and insulin
`FlexPen
`
`Lilly Pen
`
`SoloSTAR
`
`Injection force using BD
`Micro-Fine 31-gauge
`needle (N)
`
`Injection force using
`NovoFine 32-gauge tip
`
`3.3
`
`5
`
`8.3
`
`4
`
`6
`
`8
`
`3.3
`
`5
`
`8.3
`
`4
`
`6
`
`8
`
`40 units rapid-acting
`insulin analog
`in 4 seconds
`
`40 units basal insulin
`analog in 4 seconds
`
`40 units NPHb insulin
`in 4 seconds
`
`30 units
`
`60 units
`
`17.2
`
`16.3
`
`17.7
`
`5.36
`
`5.62
`
`Maximum glide force (lb)
`
`a No longer available commercially.
`b Neutral protamine Hagedorn.
`c p < 0.05 for the improved FlexPen versus SoloSTAR or KwikPen.
`d p < 0.0001 for KwikPen versus original FlexPen.
`
`8.1 ± 0.7c
`
`10.7 ± 1.4c
`
`15.6 ± 0.9c
`
`5.7 ± 0.4c
`
`8.2 ± 0.7c
`
`12.7 ± 0.5c
`
`8.3 ± 0.6
`
`12.0 ± 0.9
`
`16.2 ± 1.3
`
`9.2 ± 0.5
`
`12.5 ± 1.6
`
`13.3 ± 0.8
`
`16.9 ± 1.2
`
`56
`
`20.7 ± 2.4
`
`24.5 ± 2.6
`
`12.9 ± 0.8
`
`20.5 ± 1.3
`
`29.6 ± 1.9
`
`67
`
`6.7 ± 0.3
`
`9.1 ± 1.3
`
`10.4 ± 2.1
`
`13.1 ± 0.8
`
`56
`
`16.3 ± 1.1
`
`21.6 ± 2.0
`
`10.2 ± 1.1
`
`15.7 ±1.5
`
`22.3 ± 1.1
`
`10.3
`
`11.3
`
`25.3
`
`24.2
`
`67
`
`63
`
`66
`
`3.42d
`
`3.61d
`
`greater patient preference with fewer problems for the
`Novolog Mix 70/30 FlexPen compared with the Humalog
`Mix75/25 Pen.71 Results of other studies have shown an
`overall preference for SoloSTAR (53%) rather than for the
`original FlexPen (31%) or Humulin/Humalog pen (15%),72
`with the SoloSTAR and original FlexPen found to be
`more user-friendly than the Eli Lilly disposable pen.73
`To date, there are no studies comparing patient preference
`for the improved FlexPen compared with the SoloSTAR
`or Humulin/Humalog Pen.
`
`Yakushiji and colleagues74 assessed the preferences of
`22 Japanese men and women for various insulin pens
`when self-injecting and when administering an injection
`to another person. Insulin injections were administered
`into prosthetic skin attached either to the participant
`or to a mock patient. The FlexPen was rated as the best
`
`device for self-injection, whereas the OptiClik, perhaps
`because of its larger size, was rated as the best device for
`administration of an injection to another person.
`Conclusions
`In conclusion, the use of insulin pens offers many options
`to allow insulin delivery to be tailored to the individual
`patient. Newer designs of pens are increasingly user-
`friendly and are intuitive to use, requiring little or no
`instruction. This means that patients can quickly be
`taught how to use an insulin pen, which should go hand
`in hand with educating the patient on the importance
`of achieving accurate dosing. Thorough training of patients
`in the practical aspects and aims of insulin injection
`remains important, as errors in insulin delivery can result
`in incorrect dose administration and thus affect

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket