throbber
W Taylor & Francis
`
`~
`
`Taylor&FrannsGroup
`
`Current Medical Research and Opinion
`
`ISSN: 0300-7995 (Print) 1473-4877 (Online) Journal homepage: https://www.tandfonline.com/loi/icmo20
`
`Prefilled insulin device with reduced injection
`force: patient perception and accuracy
`
`A. Pfutzner, T. Reimer, C. Hoh berg, L. P. F. Fr0kjrer & C. J0rgensen
`
`To cite this article: A. PfUtzner, T. Reimer, C. Hohberg, L. P. F. Fr0kja:ir & C. J0rgensen (2008)
`Prefilled insulin device with reduced injection force: patient perception and accuracy, Current
`Medical Research and Opinion, 24:9, 2545-2549, DOI: 10.1185/03007990802329264
`
`To link to this article: https://doi.org/10.1185/03007990802329264
`
`la Published on line: 29 Jul 2008.
`
`~~ Submit your article to this journal C?
`
`I 1111 Article views: 261
`
`~ Citing articles: 37 View citing articles C?
`
`Full Terms & Conditions of access and use can be found at
`https://www.tandfonline.com/action/journal1nformation?journa1Code=icmo20
`
`Sanofi Exhibit 2175.001
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`CURRENT MEDICAL RESEARCH AND OPINION®
`VOL.24,NO.9,2008,2545-2549
`
`© 2008 lnforma UK Ltd.
`
`0300-7995
`doi:10.1185/03007990802329264
`
`All rights reserved: reproduction in whole or part not permitted
`
`ORIGINAL ARTICLE
`Prefilled insulin device with
`reduced injection force: patient
`perception and accuracy
`A. PfUtznera, T. Reimera, C. Hohberga,
`L. P. F. Fr0kjffirb and C. J0rgensenc
`a Institute for Clinical Research and Development (ikfe) GmbH, Parcusstr. 8,
`
`55116 Mainz, Germany
`
`bNovo Nordisk A/S, Brennum Park, DK-3400 Hillerod, Denmark
`
`0 Novo Nordisk A/S, Hummeltoftevej 49, 2830 Virum, Denmark
`
`Address for correspondence: Christina Jorgensen, Hummeltoftevej 49, 2830 Virum, Denmark.
`Tel.: +45 4443 2884; Fax: +45 4442 1848; cjrg@novonordisk.com
`
`Key words: Dose accuracy - FlexPen - Injection force
`
`ABSTRACT
`
`Objective: The injection force and the patient perception of
`the Next Generation FlexPen* (NGFP) with design modifi(cid:173)
`cations aimed at reducing injection force was assessed.
`The accuracy and precision of the NGFP was also tested
`under standard conditions.
`Research design and methods: Dosing accuracy was
`tested (according to ISO 11608 requirements) at 1 IU, 30 IU
`and 60 IU doses (acceptable limits were 1 ± 1 IU (0-2 IU),
`30 ± 1.5 IU (28.5-31.5 IU), and 60 ± 3 IU (57-63 IU)). Pens
`were tested at reference conditions (18-28°C and relative
`humidity 25-75%). Delivered doses were measured on
`a sensitive balance and corrected for the specific density
`of the insulin aspart used (according to ISO 11608-1).
`Precision was calculated from the variance around the
`mean delivered dose. The injection force of NGFP was
`measured, and user-preference of NGFP and FlexPen
`(FP) were compared in 50 patients with type 2 diabetes.
`
`Results: The mean injection force with NGFP and FP
`was 12.57±1.81 N and 17.90±1.51 N (p<0.001),
`respectively. Almost twice as many patients rated the
`injection force as 'good' or 'very good' with NGFP (80%,
`72% and 38% when delivering 20 IU, 40 IU and 60 IU,
`respectively) compared with FP (48%, 32% and 20% when
`delivering 20 IU, 40 IU and 60 IU, respectively) and 76% of
`patients rated NGFP as superior, in terms of simplicity and
`comfort, to FP. NGFP accurately delivered the set doses
`(means [SD] were 0.98 [0.06] IU, 29.98 [0.18] IU, and
`59.93 [0.24] IU for the 1 IU, 30 IU and 60 IU doses,
`respectively).
`Conclusions: These results show that NGFP has a 30%
`reduction in injection force compared with FP and was
`rated as 'more simple and comfortable to use' by patients.
`Furthermore, NGFP was as accurate and as precise as FP.
`
`Introduction
`
`The use of insulin injections for the management of
`diabetes can be perceived as inconvenient and trau(cid:173)
`matic by patients. Many patients fear injections, find
`traditional vial and syringes inconvenient, and lack
`confidence in delivering accurate doses of insulin;
`these factors limit patients' overall confidence to self(cid:173)
`2
`manage their diabetes with insulin 1
`,
`.
`
`Partly because of injection fears and partly because
`the use of vial and syringe can lead to inaccurate
`dosing3
`,4, injection pens are now the predominant
`devices for insulin delivery, especially in Europe. Pens
`overcome the patient's fear of injection, are convenient
`to use and increase confidence in dose delivery, and
`therefore, potentially improve the quality of life of
`10
`the user 5
`. Because it can have consequences for gly(cid:173)
`-
`caemic control, the improved accuracy and precision of
`
`*FlexPen is a registered trade name of Novo Nordisk A/S, Bagsva:rd, Denmark
`
`Article 4634/333093
`
`2545
`
`Sanofi Exhibit 2175.002
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`insulin pens compared with vials and syringes is very
`13
`important11
`. The ability of pens to deliver more
`-
`accurate insulin doses than syringes is particularly
`obvious when low doses (1-5 IU) have to be adminis(cid:173)
`tered 11· 12, but the impact this may have on the
`frequency of hypoglycaemia or even more serious long(cid:173)
`term consequences of poor glycaemic control has not
`been demonstrated.
`FlexPen* (FP) is a prefilled insulin pen that is used by
`4
`more than 3 million people with diabetes 1
`. The dosing
`accuracy and consistency of FP has been confirmed in
`18
`several studies 15
`. It is known that some insulin pens
`-
`can, on rare occasions, block if the push button is
`pushed at an oblique angle, and the manufacturer has
`received occasional reports that patients found the
`push-button hard to push down. To overcome these
`issues and to reduce the injection force, a Next
`Generation FP (NGFP) has been designed. The NGFP
`has changes to the ratchet, the clutch and the push(cid:173)
`button to reduce the injection force and to eliminate
`blocking with FP. The other features of NGFP are
`exactly the same as FP. As with any new insulin pen,
`existing pens that undergo design changes must be
`shown to deliver accurate and reliable insulin doses
`during rigorous testing. Here we report on the dosing
`accuracy of NGFP compared with FP. The injection
`force of NGFP and FP was also tested in our study. In
`addition, we present the results of a questionnaire
`survey of user preference between NGFP and FP.
`
`qualified and experienced in conducting such tests
`according to ISO standards, and standard methodology
`was used for the measurement of dose accuracy as
`9
`described in the ISO standards for insulin pen devices 1
`.
`All doses were delivered in a random order. The accuracy
`of dosing was measured by delivering 1 IU, 30 IU or 60 IU
`60 times (180 doses in all) with each pen (according to
`ISO 11608-1). Each test was carried out with a new and
`unused pen, and before delivering each dose the pen was
`prepared according to ISO 11608-1 instructions. Doses
`were delivered onto a precision balance and weighed.
`Weights were corrected for the specific density of insulin
`aspart (1.008 g/ml at S°C, 1.005 g/ml at 20°C and
`0.999 g/ml at 40°C). Dose accuracy was measured at
`reference conditions: pens were exposed to 18-28°C
`and relative humidity (RH) 25-75% for at least 4
`hours before measuring dose accuracy.
`Acceptable limits for dose accuracy were based on
`ISO regulations (ISO 11608-1): 1 ± 1 IU (0-2 IU),
`30±1.SIU (28.5-31.SIU), and 60±3IU (57-631U).
`The mean value and standard deviations of the 60 doses
`delivered at each of the three settings (1 IU, 30 IU and
`60 IU) were calculated. In addition, minimum, maxi(cid:173)
`mum and inter-quartile range values were recorded.
`The difference between mean absolute deviation in
`the delivered dose of each pen was calculated using
`Stigma Stat v3.5. and the Levenes test was used to
`test for homogeneity of variance as a measure of pen
`precision.
`
`Methods
`
`All dose accuracy test methods were conducted in an
`identical manner with both NGFP and FP.
`
`Materials
`
`The NGFPs tested for injection force and performance
`were all drawn from the same batch (lot TG 70313/ A)
`and were filled with NovoRapidt (insulin aspart) (batch
`TZ60322). For the tests on FP, pens were drawn from
`batch SPS0497 and filled with NovoRapid (insulin
`aspart) (batch RQ50794). Pens were picked randomly
`from these lots to perform each test. The same batches
`of pens were used for the user preference study.
`
`Calculation of dose accuracy
`
`Accuracy was measured in the Novo Nordisk NS
`Quality Assurance laboratory, all testers were suitably
`
`Injection force measurements
`
`Injection force was measured under standard condi(cid:173)
`tions with 20 previously unused pens with the cartridge
`and NovoFinei 30G 8 mm needle attached and at
`a flow rate of 10 IU/s.
`
`User preference questionnaire
`
`In this part of the study NG FP and FP were com pared for
`injection force perception by users. The survey recruited
`50 patients with type 2 diabetes. The study was handled
`in accordance with local legal and ethical requirements
`and each patient signed an informed consent form before
`taking part in the study. All participants were receiving
`insulin at the time and used an insulin pen at least once
`per day. Test injections with insulin aspart (NovoRapid)
`were made into an injection pillow. Pens were fitted with
`30G 8 mm needles, and each patient used a previously
`unused pen of each type.
`
`*FlexPen is a registered trade name of Novo Nordisk A/S, Bagsva:rd, Denmark
`tNovoRapid is a registered trade name of Novo Nordisk NS, Bagsva:rd, Denmark
`tNovoFine is a registered trade name of Novo Nordisk NS, Bagsva:rd, Denmark
`
`2546 Preti/led insulin device with reduced injection force
`
`© 2008 lnforma UK Ltd - Curr Med Res Opin 2008; 24(09)
`
`Sanofi Exhibit 2175.003
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`Each patient delivered one dose of 20 IU, 40 IU and
`60 IU with each pen into the injection pillow, in
`a randomised order. After delivering each dose with
`both pens, each patient answered Questions 1, 2 and 3,
`respectively (see Appendix). After delivering all doses
`and answering Questions 1-3, each patient answered
`Question 4 on pen 'simplicity and comfort' of use.
`For the answers to Questions 1-3, data were mod(cid:173)
`elled by logistic regression for ordinal response with
`pen and patient as fixed effects. A likelihood-ratio
`Chi-square test was used to test if pen type had a sig(cid:173)
`nificant effect on answers. The responses to Question 4
`were tested for equal 'simplicity and comfort' using
`a two-sided chi-square test. JMP version 7 .0.1 was
`used for these analyses.
`
`Results
`Dose accuracy
`
`Dose accuracy was well within the acceptable limits for
`NGFP and FP (Table 1 ). No single dose delivered by any
`pen (at 1 IU, 30 IU or 60 IU) was below or above the
`pre-specified acceptable ranges. The mean absolute
`deviation from the set doses for FP were 0.06 IU (6% ),
`0.23 IU (0.8%) and 0.36 IU (0.6%) for 1 IU, 30 IU and
`60 IU, respectively. The corresponding values for NGFP
`were 0.05 IU (5% ), 0.14 IU (0.5%; p < 0.05 compared
`with FP) and 0.19IU (0.3%; p<0.01 compared with
`FP) for 1 IU, 30 IU and 60 IU, respectively. The analysis
`of precision also showed that NGFP was significantly
`more precise when delivering doses of 30 IU (p < 0.05)
`and 60IU (p<0.05) than FP. Stress tests of NGFP at
`different
`temperatures and humidity
`( conducted
`according to ISO methodology) showed that there
`was no significant change in the dose accuracy under
`various environmental conditions (results not shown).
`
`Injection force
`
`The mean injection force with NGFP was 12.57 N and
`this is considerably lower than the injection force with
`
`FP (17 .90 N measured with the same methodology).
`This equated to a 29 .8% relative reduction in the
`injection force of NGFP compared with FP (p<0.001;
`t-test assuming normally distributed measurements).
`The range of injection force measured with FP was
`15.35-19.99N, so all FPs had injection force higher
`than the mean injection force with NGFP.
`
`User preference survey
`
`After delivering three doses with each pen, more
`patients rated the injection force with NGFP as 'good'
`or 'very good' than with FP (Figure 1). The difference
`in responses after using each pen was significant
`(p < 0.0001) for all three doses. An injection force
`rating of 'very good' or 'good' was given to NGFP
`by 80% of patients when delivering 20 IU, 72% when
`delivering 40 IU and 38% when delivering 60 IU - the
`corresponding proportions for FP were 48, 32 and 20%,
`respectively.
`In response to Question 4, 7 6% of the patients stated
`that NGFP was 'simpler and more comfortable' to use
`than FP, and only 24% stated that FP was 'simpler and
`more comfortable' to use than NGFP. The hypothesis
`of equal
`'simplicity and comfort' is rejected and
`shows that a significantly greater proportion of patients
`found NGFP more 'simple and comfortable' to use
`(p=0.0002).
`
`Discussion
`
`This study demonstrated that NGFP delivers insulin in
`an accurate and precise manner and is more 'simple and
`comfortable' to use than FP. The design modifications
`to NGFP have not adversely affected any of the pre(cid:173)
`18 but have
`viously demonstrated attributes of FP 15
`-
`reduced the injection force by 30% compared with
`FP. NGFP delivered doses were more accurate and pre(cid:173)
`cise (significantly for 1 IU) than FP delivered doses.
`
`Table 1. Accuracy and precision of NGFP and FP under standard conditions*
`
`Dose of insulin delivered with NGFP
`
`Dose of insulin delivered with FP
`
`Value
`
`Set to 1 IU
`
`Set to 30IU
`
`Set to 60IU
`
`Set to 1 IU
`
`Set to 30IU
`
`Set to 60IU
`
`Mean
`Standard deviation
`Min
`25th percentile
`50th percentile
`75th percentile
`Max
`Range
`
`0.98
`0.06
`0.83
`0.95
`0.97
`1.02
`1.10
`0.27
`
`*60 tests for each set dose
`
`29.98
`0.18
`29.65
`29.89
`29.98
`30.09
`30.78
`1.13
`
`59.93
`0.24
`59.43
`59.80
`59.94
`60.10
`60.43
`1.00
`
`1.04
`0.06
`0.91
`1.00
`1.03
`1.08
`1.23
`0.32
`
`29.93
`0.31
`29.06
`29.71
`29.99
`30.10
`30.52
`1.46
`
`59.90
`0.45
`59.00
`59.59
`59.96
`60.21
`60.82
`1.82
`
`© 2008 lnforma UK - Curr Med Res Opin 2008; 24(09)
`
`Preti/led insulin device with reduced injection force Prntzner et al. 2547
`
`Sanofi Exhibit 2175.004
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`FP 20 IU
`
`FP 40 IU
`
`5
`::::.,
`Q) NGFP 20 IU
`U)
`0
`"lJ
`.!:
`::i
`U) NGFP 40 IU
`.!:
`"lJ
`C:
`Ctl
`C:
`Q)
`
`FP 60 IU
`
`D.. NGFP 60 IU
`
`22%
`
`6%
`
`24%
`
`22%
`
`4%
`
`46%
`
`2%
`
`14%
`
`1•
`
`..
`
`48%
`
`22%
`
`10%
`
`32%
`
`24%
`
`■ . . ■ = . ■
`
`36%
`
`12%
`
`, ..
`
`- 48%
`
`14% 6%
`
`....
`
`30%
`
`8%
`
`100
`
`50
`0
`50
`Percentage of patients rating as postive or negative
`
`100
`
`*p<0.0001; tdata missing for one patient
`
`Figure 1. Patient perception of FP and NG FP injection force when injecting 20 JU, 40 JU and 60 JU of insulin aspart.
`*p < 0.0001 between pens; responses compared using likelihood ratio chi-square test
`
`One aim of this study was to confirm that the design
`changes had reduced the injection force of NGFP .
`The mean injection force for all NGFPs was 30%
`lower than the mean value for FPs. The maximum
`and minimum injection forces measured with 20
`NGFPs (10.63-18.57 N) were consistently lower than
`the maximum and minimum injection forces measured
`with 20 FPs (15.35-19.99 N), therefore NGFP does
`have the desired reduced injection force. A lower injec(cid:173)
`tion force is generally preferred by patients simply
`because the lower the force required for injection, the
`easier it is to inject the insulin dose. This is particularly
`important in patients with impaired manual dexter(cid:173)
`ity20-22. To test if the reduction in injection force
`with NGFP had an actual benefit to the end-user, the
`results from the user-preference study were included
`here. In this user-preference study, patients had
`a better perception of the injection force of NGFP
`than FP at all doses tested ( nearly twice as many patients
`rated the injection force as 'good' or 'very good' with
`NGFP compared to FP). Not only does this suggest that
`the reduced injection force is perceived as an improve(cid:173)
`ment by patients, but it also contributed to three out of
`four patients finding NGFP 'simpler and more comfor(cid:173)
`table' to use. However, a limitation of this study is the
`fact that NGFP has not been tested in routine clinical
`practice, and whether this preference is maintained in
`everyday use will need to be investigated.
`With the increased use of insulin pens23, a wider vari(cid:173)
`ety of patients will gain more experience with a range of
`pens. Different patient populations will have different
`specific demands and the importance of various pen
`features will vary from patient to patient. Several stu(cid:173)
`dies have investigated the reasons for patient preference
`• l
`d
`.
`l
`.
`s-10 24-27 F
`wr pens over via an syringe
`'
`. or examp e, m
`1:
`a study of insulin users and non-users, social acceptabil(cid:173)
`ity and ease of use were the most important factors for
`preference for pens10. In a study of FP versus vial and
`
`syringe, patients preferred FP due to improvements in
`convenience, flexibility, perceived clinical efficacy and
`quality of life26. The changes to NGFP are designed to
`enhance the preferred features of FP further and this
`report suggests that NGFP is rated as 'simpler and
`more comfortable' to use than FP, because it delivers
`the same accurate and reliable doses of insulin at high
`and low doses under a variety of everyday conditions,
`but with a considerably lower dose force.
`
`Conclusion
`
`The modifications made to NGFP have resulted in
`a significant 30% reduction in injection force compared
`with FP. These changes have contributed to NGFP
`being considered more 'simple and comfortable' to
`use by patients. Importantly, the dose accuracy of
`NGFP is at least as good as FP, and our study suggests
`that NGFP may actually be significantly more accurate
`and precise than FP - deviation from set doses and the
`variance in the delivered dose were small with both
`pens but smallest with NGFP. Although the NGFP
`has not been tested in everyday use, this study suggests
`that the changes made are likely to enhance patient
`comfort and aid accurate insulin dosing.
`
`Acknowledgements
`
`Declaration of interest: The preference study was sup(cid:173)
`ported by Novo Nordisk and performed by the Institute
`for Clinical Research and Development (ikfe) GmbH,
`Mainz, Germany. Publication was supported by Novo
`Nordisk NS (Bagsv~rd, Denmark) with editorial assis(cid:173)
`tance from ESP Bioscience (Sandhurst, UK).
`Dr Andreas Pfiitzner is a part of the advisory board of
`Novo Nordisk A/S. C. forgensen and L. P. F. Frokj~r
`are employees of Novo Nordisk A/S. T. Reimer and
`C. Hohberg are employees of ikfe and assisted in the
`
`2548 Preti/led insulin device with reduced injection force
`
`© 2008 lnforma UK Ltd - Curr Med Res Opin 2008; 24(09)
`
`Sanofi Exhibit 2175.005
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`practical performance of the study. The authors have
`approved the final draft of this manuscript and take full
`responsibility for its contents.
`
`References
`1. Bashoff EC, Beaser RS. Insulin therapy and the reluctant
`patient. Overcoming obstacles to success. Postgrad Med 1995;
`97 :86-90, 3-6
`2. Korytkowski M, Niskanen L, Asakura T. FlexPen: addressing
`issues of confidence and convenience in insulin delivery. Clin
`Ther 2005;27(Suppl B):S89-100
`3. Puxty JA, Hunter DH, Burr WA. Accuracy of insulin injection
`in elderly patients. Br Med J (Clin Res Ed) 1983;287:1762
`4. Kesson CM, Bailie GR. Do diabetic patients inject accurate
`doses of insulin? Diabetes Care. 1981;4:333
`5. Bohannon NJ. Insulin delivery using pen devices. Simple(cid:173)
`to-use tools may help young and old alike. Postgrad Med
`1999;106:57-8, 61-4, 8
`6. Chantelau E, Schiffers T, Schutze J, et al. Effect of patient(cid:173)
`selected intensive insulin therapy on quality of life. Patient
`Educ Couns 1997;30:167-73
`7. Kadiri A, Chraibi A, Marouan F, et al. Comparison of NovoPen
`3 and syringes/vials in the acceptance of insulin therapy in
`NIDDM patients with secondary failure to oral hypoglycaemic
`agents. Diabetes Res Clin Pract 1998 ;41: 15-23
`8. Albano S. Assessment of quality of treatment in insulin-treated
`patients with diabetes using a pre-filled insulin pen. The
`ORBITER Study Group. Acta Biomed 2004;75:34-9
`9. Lombardo F, Salzano G, Messina MF, et al. Compliance and
`administration methods in management of type 1 diabetes.
`Acta Biomed 2005;76(Suppl 3):66-9
`10. Summers KH, Szeinbach SL, Lenox SM. Preference for insulin
`delivery systems among current insulin users and nonusers. Clin
`Ther 2004;26:1498-505
`11. Keith K, Nicholson D, Rogers D. Accuracy and precision of
`low-dose insulin administration using syringes, pen injectors,
`and a pump. Clin Pediatr (Phila) 2004;43:69-74
`12. Lteif AN, Schwenk WF. Accuracy of pen injectors versus insulin
`syringes in children with type 1 diabetes. Diabetes Care 1999;
`22:137-40
`13. Coscelli C, Lostia S, Lunetta M, et al. Safety, efficacy, accept(cid:173)
`ability of a pre-filled insulin pen in diabetic patients over
`60 years old. Diabetes Res Clin Pract 1995;28: 173-7
`
`Appendix: Questionnaire
`
`14. IMS Health. Estimated number of patients using FlexPen®,
`based on worldwide sales in number of packs sold, IMS world
`wide data Q4'07 and Daily Defined Dosage (DDD) for insulin
`as issued by WHO 2007.
`15. Asakura T, Seino H. Insulin dosing accuracy with FlexPen®
`versus syringe by healthcare professionals and healthy volun(cid:173)
`teers. Diabetes 2005;54 (Suppl l):A498-9
`16. Asakura T. Comparison of the dosing accuracy of two insulin
`injection devices. J Clin Res 2005;8:33-40
`17. Pfutzner A, Weise A, Sun H, et al. Differences in dose accuracy
`of insulin pens.
`in 43rd EASD Annual Meeting, 2007.
`Amsterdam, The Netherlands.
`18. Asakura T, Seino H, Kageyama M, et al. Dosing accuracy of two
`insulin pre-filled pens. Curr Med Res Opinion 2008;24: 1429-34
`19. Pen-injectors for medical use - Part 1: Pen-injectors - require(cid:173)
`ments and test methods. ISO l l 608-1:2000-2001.
`20. Ziegler D, Gries FA, Spuler M, et al. The epidemiology of dia(cid:173)
`betic neuropathy. DiaCAN Multicenter Study Group. Diabet
`Med 1993;(10 Suppl 2):82-6S
`21. Schady W, Abuaisha B, Boulton AJ. Observations on severe
`ulnar neuropathy in diabetes. J Diabetes Complicat 1998;
`12:128-32
`22. Savas S, Koroglu BK, Koyuncuoglu HR, et al. The effects of the
`diabetes related soft tissue hand lesions and the reduced hand
`strength on functional disability of hand in type 2 diabetic
`patients. Diabetes Res Clin Pract 2007; 77: 77-83
`23. Costa SD, Brackenridge B, Hicks D. A comparison of insulin pen
`use in the United States and the United Kingdom. Diabet Educ
`2002; 1 :52-6
`24. Thurman JE. Insulin pen injection devices for management
`of patients with type 2 diabetes: considerations based on an
`endocrinologist's practical experience in the United States.
`Endocr Pract 2007; 13:672-8
`25. Bohannon NJ, Ohannesian JP, Burdan AL, et al. Patient and
`physician satisfaction with the Humulin/Humalog Pen, a new
`3.0-mL prefilled pen device for insulin delivery. Clin Ther
`2000;22: 1049-67
`26. Rubin RR, Peyrot M. Quality of life, treatment satisfaction,
`and treatment preference associated with use of a pen device
`delivering a premixed 70/30 insulin asp art suspension ( asp art
`protamine suspension/soluble aspart) versus alternative treat(cid:173)
`ment strategies. Diabetes Care 2004;27:2495-7
`27. Stocki K, Ory C, Vanderplas A, et al. An evaluation of
`patient preference for an alternative insulin delivery system
`compared to standard vial and syringe. Curr Med Res Opin
`2007;23: 133-46
`
`Question 1. How do you perceive the injection force at 20 IU?
`Possible answers: 'very bad', 'bad', 'neutral', 'good', and 'very good'
`Question 2. How do you perceive the injection force at 40 IU?
`Possible answers: 'very bad', 'bad', 'neutral', 'good', and 'very good'
`Question 3. How do you perceive the injection force at 60 IU?
`Possible answers: 'very bad', 'bad', 'neutral', 'good', and 'very good'
`Question 4. Which pen, in your opinion, is simpler and more comfortable to use?
`
`CrossRef links are available in the online published version of this paper:
`http://www.cmrojournal.com
`Paper CMRO-4634_ 4, Accepted for publication: 8 July 2008
`Published Online: 29 July 2008
`doi: 1 O. l l 85/03007990802329264
`
`© 2008 lnforma UK - Curr Med Res Opin 2008; 24(09)
`
`Preti/led insulin device with reduced injection force Prntzner et al. 2549
`
`Sanofi Exhibit 2175.006
`Mylan v. Sanofi
`IPR2018-01676
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket