throbber
Available on line at www.sciencedirect.com
`
`SCIENCE@DIRECT•
`
`ELSEVIER
`
`Diabetes Research and Clinical Practice 63 (2004) 27-35
`
`llllll&.!Blll!'ll'IE~ mll!~Willl.lCIHI
`.I.\W!D)
`WNITICAll. !PlIU\all<ClB
`
`www.clscvicr.com/locatc/diabres
`
`Preference and resource utilization in elderly patients:
`InnoLet® versus vial/syringe
`
`John Shehnet a,*, Sherwyn Schwartz b , John Cappleman c, Gregory Peterson d,
`Soren Skovlund e, Lene Lytzen e, Lars Nicklas son r, John Liang r, William Lyness r,
`InnoLet® Study Group 1
`
`a 3131 Princeton Pike, Building 2B, Suite #104, Lawrenceville, NJ 08648, USA
`b Diabetes & Glandular Disease Clinic, 5107 Medical Drive, San Antonio, TX 78229, USA
`c CEC Inc., 10,000 West Colonial Drive, Suite 187, Ocoee, FL 34761, USA
`d Diagnostic & Critical Care Medicine PC, 411 Laurel Street, Suite 3275, Des Moines, IA 50314, USA
`e Novo Nordisk A/S, Sorgenfri, Denmark
`f Novo Nordisk Inc., Princeton, NJ 08540, USA
`
`Received 20 March 2003; received in revised form 8 August 2003; accepted 20 August 2003
`
`Abstract
`
`InnoLet® is a disposable insulin injection device with a large easy-to-read dial, large push button for injection, and audible
`clicks for each unit injected. This clinical trial assessed patient preference, satisfaction, and utilization of healthcare resources
`(estimated nursing care) for InnoLet and vial/syringe. Patients with diabetes mellitus (N = 79, mean age 68.2 ± 8.6 years,
`duration of diabetes 16.5 ± 10.9 years) having visual and/or motor disabilities and having difficulty (or required caregiver assis(cid:173)
`tance) for previous injections by vial/syringe were randomized to use of either InnoLet or vial/syringe for 6 weeks, then switched
`to the alternate regimen for 6 weeks. At the end of the study, utilization of healthcare resources was assessed in terms of the
`caregiver time required to assist in preparation, storage, and disposal of each device. For vial/syringe, 60% of patients required
`assistance in drawing up the appropriate dosage in the syringe, and 36% of patients required assistance when injecting insulin.
`A major portion of the patients (53%) could independently conduct injections (without nursing/caregiver assistance) during use
`of InnoLet, versus 20% for vial/syringe. As a result, mean daily nursing costs associated with the injection regimen were US$
`114 for the InnoLet device, and US$ 196 for vial/syringe ( P < 0.001 ). A majority of patients (82%) indicated a preference for
`the InnoLet device (P < 0.001).
`© 2003 Elsevier Ireland Ltd. All rights reserved.
`
`Keyword1·: Insulin delivery systems; Healthcare costs; Quality oflife; Patient independence; Patient acceptance
`
`* Corresponding author. Tel.: + 1-609-896-8050; fax: + 1-609-896-3053.
`E-mail address: jjsdiabetes@aol.com (J. Shelmet).
`1 InnoLet® Study Group: Bruce Bode, Edward Gillie, Sam Lerman, Serigo Mather, Diane Normandin, Richard Sievers and James West.
`
`0168-8227/$ - see front matter© 2003 Elsevier Ireland Ltd. All rights reserved.
`doi: I 0.10 l 6/j.diabres.2003.08.013
`
`Sanofi Exhibit 2141.001
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`28
`
`J. She/met et al. I Diabetes Research and Clinical Practice 63 (2004) 27-35
`
`1. Introduction
`
`The Diabetes Control and Complications Trial
`demonstrated
`that
`intensive
`insulin
`therapy can
`significantly reduce the incidence of late diabetic
`complications, and delay the progression of existing
`conditions in type 1 diabetes [1]. The United Kingdom
`Prospective Diabetes Study reported similar findings,
`and concluded that intensive therapy reduces the risk
`of many complications of type 2 diabetes (retinopa(cid:173)
`thy, nephropathy, and neuropathy) [2]. Significant
`costs are accrued by patients with diabetes who have
`poor glycemic control with insulin [3]. Diabetes is
`associated with blindness and end stage renal disease.
`Coronary artery disease is more severe in patients
`with diabetes.
`Many patients with diabetes using insulin injection
`may have poor treatment compliance. Lack of dia(cid:173)
`betes education is often a major factor. Other compli(cid:173)
`ance considerations may be poverty, fear of needles,
`denial, and lifestyle. In elderly patients, diabetes is of(cid:173)
`ten complicated by debilitating co-morbidities (vision
`problems, neuropathy, etc.) that aggravate the difficul(cid:173)
`ties of self-injection and increase the risk of dosing er(cid:173)
`rors. Any treatment approach that allows the patient to
`more easily self-inject insulin may have implications
`regarding the financial burden of diabetes complica(cid:173)
`tions, deterioration of quality of life, and requirements
`for nursing assistance.
`The InnoLet® insulin injection device was engi(cid:173)
`neered for ease of use. The disposable device has a
`large, easy-to-read dial, which aids in accurate insulin
`dose selection by the patient. The large injection button
`requires less physical effort for a patient with motor
`impairment (stroke, Parkinson's disease). InnoLet has
`been previously studied in vision-impaired patients:
`patients with visual acuities of 20/100 to 20/200 were
`able to correctly dispense an insulin dose with Inno(cid:173)
`Let, while 15 and 36% were unable to perform this
`task with a Humulin pen or vial/syringe, respectively
`[ 4].
`This study was designed to determine whether the
`InnoLet injection device provides benefit to elderly
`patients in terms of nursing resources, treatment
`satisfaction, and patient preference as compared to
`the vial/syringe method of insulin injection. Con(cid:173)
`tinued quality of care (glycemic control) was also
`monitored.
`
`2. Methods
`
`2.1. Study design
`
`This trial was performed in accordance with the
`Declaration of Helsinki, and written informed consent
`was obtained from all patients. This was a multicenter,
`randomized, open-label, two-period crossover trial in
`patients with type 1 or type 2 diabetes mellitus con(cid:173)
`ducted at 11 sites in the United States. The enrolled
`patients were currently injecting their own insulin (ei(cid:173)
`ther Novolin N or Novolin 70/30), but had difficul(cid:173)
`ties (requiring some assistance of a nurse or caregiver)
`due to motor dysfunction (arthritis, familial tremor,
`Parkinson's disease, stroke-induced partial paralysis)
`and/or visual problems (partially blind, cataracts, vi(cid:173)
`sual field defects, blurred vision). Enrolled patients
`were familiar with either a pen and/or vial and sy(cid:173)
`ringe method of insulin injection. Patients were ran(cid:173)
`domly assigned use oflnnoLet with N ovoFine® 30G x
`8 mm needles or the conventional disposable 0.5 cc
`Becton-Dickinson insulin syringe with MicroFine 30G
`permanently attached needle for the first period of 6
`weeks in duration, and switched to the alternate device
`for the second period of 6 weeks. The insulin dose
`was predetermined based on the patient's prior daily
`insulin requirements, and was changed over the course
`of study based on the health needs and/or diet of the
`patient. In this study, the costs of training patients in
`the use oflnnoLet® were negligible [ 4].
`
`2.2. Healthcare resource utilization
`
`Healthcare resource utilization for vial/syringe or
`device use was examined at the end of each study
`period: a questionnaire was utilized to determine the
`aspects of use that required assistance (i.e. what per(cid:173)
`centage of patients required to draw up the appro(cid:173)
`priate volume, to inject themselves, dispose of the
`vial/syringe, etc.). For each means of injection, the
`nursing/caregiver resource utilization was estimated as
`minutes per day necessary for assisting injection ( from
`injection preparation to disposal): 0 min was indicated
`if the patient was capable of performing all the activi(cid:173)
`ties by themselves. The resource utilization (in dollar
`amounts) was also calculated in terms of the number
`of visits per day that nurses/caregivers had to make
`to the patient in order to assist with insulin injections.
`
`Sanofi Exhibit 2141.002
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`J. She/met et al. I Diabetes Research and Clinical Practice 63 (2004) 27-35
`
`29
`
`The time necessary for each component (preparation
`to disposal) was individually tabulated by treatment
`and study period. If a patient required assistance with
`any aspect of preparation or injection, they were con(cid:173)
`sidered as requiring a nursing visit. Cost calculations
`were based upon an assumption of $80 per visit (mini(cid:173)
`mum of 1 hr per visit). Comparison of the two devices
`and the differences in dollar values were made using
`general linear models for the crossover design.
`
`2.3. Preference and acceptance questionnaires
`
`Patient preference was assessed by a questionnaire
`at the end of the study: "If you were given the choice
`of using one of the two systems (vial and syringe or
`InnoLet), which would you prefer?" Ease of use was
`also assessed: "Compared with the vial and syringe
`you were using, injection of insulin with the InnoLet
`system you are now using is easier, the same or more
`difficult?" A patient handling and acceptance ques(cid:173)
`tionnaire were provided at the end of each 6-week
`treatment period.
`
`2. 4. Efficacy and safety assessments
`
`The glycemic control was assessed by measur(cid:173)
`ing the serum fructosamine levels at the time of
`screening and at the end of each treatment period.
`Safety assessments were based on adverse events
`and adverse device effects, hypoglycemic episodes,
`vital signs, physical examination and the physicians
`review of the blood glucose (BG) diaries. Patients
`recorded meter-measured BG values and symptoms
`of hypoglycemia associated with blood glucose meter
`readings.
`Hypoglycemia was defined as minor when the pa(cid:173)
`tient had a symptom of hypoglycemia (i.e., palpita(cid:173)
`tions, tiredness, sweating, strong hunger, dizziness,
`tremor, etc.) confirmed by blood glucose meter read(cid:173)
`ing <50 mg/dl, and was able to deal with the episode
`without assistance. A major or severe hypoglycemic
`episode was an event that had a blood glucose meter
`reading <50 mg/dl and required third-party assistance.
`
`val was bigger than 50%, the null hypothesis was re(cid:173)
`jected, and it was concluded that the patients preferred
`the InnoLet device. A hypothesis of equal treatment
`median was tested for two treatments, versus the hy(cid:173)
`pothesis that they were not equal. A two-sided test was
`used when analyzing the comparison. The score data
`of the questionnaires was not assumed to be a nor(cid:173)
`mal distribution; therefore a non-parametric analysis,
`Wilcoxon Signed Rank test [5] was used for analysis
`of resource utilization (in dollars), and diabetes treat(cid:173)
`ment satisfaction. A binomial test was used for patient
`preference.
`
`3. Results
`
`3.1. Demographic and other baseline
`characteristics
`
`Seventy-nine patients (age 47-85, inclusive) with
`type 1 or type 2 diabetes and requiring ::::;50 U of in(cid:173)
`sulin per injection were randomized to the treatment
`sequence InnoLet ➔ vial/syringe or vial/syringe ➔
`Innolet. A total of73 (92%) patients out of the 79 com(cid:173)
`pleted the study. Five (6%) patients discontinued In(cid:173)
`noLet use due to adverse events, non-compliance with
`protocol, and change of insulin to Lantus. One (1 % )
`patient discontinued use of syringe due to unavailabil(cid:173)
`ity for follow-up. The demography of randomized pa(cid:173)
`tients is summarized in Table 1.
`
`3.2. Insulin regimen
`
`The insulin dose was based upon prior daily insulin
`requirements, and was changed as necessary over the
`course of the study. For all patients who completed
`study treatment, the mean± S.E.M. total daily in(cid:173)
`sulin dose at screening was 44.9 ± 0.14 IU/day (range
`4-100 IU) for patients using the Irmo Let device, and
`46. 7 ± 0.12 IU/day (range 10-100 IU) for patients us(cid:173)
`ing the vial/syringe.
`
`3.3. Healthcare resource utilization
`(assistance needed)
`
`2.5. Statistical analysis
`
`For preference data, a two-sided 95% confidence
`interval was calculated. If the lower limit of the inter-
`
`The resource utilization questionnaire was com(cid:173)
`pleted at the end of each trial period. In response to the
`question "When using a vial and syringe to administer
`
`Sanofi Exhibit 2141.003
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`30
`
`J. She/met et al. I Diabetes Research and Clinical Practice 63 (2004) 27-35
`
`Table I
`Demographics and baseline characteristics
`
`Number randomized, N
`Age (year), mean ± S.D.
`
`Sex, n (%)
`Male
`Female
`
`Race, n (%)
`Caucasian
`Black
`Hispanic
`
`BM! (kg/m2
`), mean ± S.D.
`AIC (%), mean ± S.D.
`
`Years since diagnosis
`Mean± S.D.
`
`Type of diabetes, n (%)
`Type I
`Type 2
`
`Overall
`
`79
`68.2 ± 8.6
`
`33 (42)
`46 (58)
`
`65 (82)
`6 (8)
`8 (10)
`33 ± 6.5
`7.5 ± 1.4
`
`16.5 ± 10.9
`
`2 (3)
`77 (97)
`
`insulin the patient needs my assistance", 60% of the
`patients required assistance in drawing up the appro(cid:173)
`priate volume, and 36% patients required assistance
`in injecting themselves with insulin. In response to
`the question "When using a vial and syringe to ad(cid:173)
`minister insulin the patient is totally self-sufficient",
`25% of patients were self-sufficient (able to prepare,
`store, and dispose insulin without nursing assistance)
`using vial/syringe, but using various injection aids
`(magnifying glass etc.).
`
`3.4. Resource utilization-time in minutes for
`nursing/caregiver resources
`
`The time (minutes) necessary for assisting injection
`(from injection preparation to disposal) was assessed
`by treatment and study period. The mean time spent
`by nurses or caregivers while assisting in injection
`preparation was less for patients using the InnoLet
`device (4.2 ± 8.1 min) than for vial/syringe (5.8 ±
`8.9 min).
`
`3.5. Resource utilization-number of visits for
`nursing/caregiver resources
`
`ing/caregiver assistance (Table 2). Some patients in
`the study required as many as three visits per day at
`US$ 80 per visit. For vial and syringe users, a major(cid:173)
`ity of the patients required such assistance (three daily
`visits). Only 20% of the patients with vial/syringe
`reported independence (required no assistance). The
`mean daily cost was US$ 196 per day for vial/syringe.
`By comparison, 53% of the patients were independent
`of nursing/caregiver assistance for injections with In(cid:173)
`noLet use, and the mean daily cost of their injections
`was US$ 114. Overall, the mean daily nursing cost
`of the InnoLet was significantly less than vial and
`syringe (US$ 114 versus US$ 196, P < 0.001).
`
`3.6. Patient preference, handling, and acceptance of
`the insulin delivery system
`
`Overall patient preference is summarized in Fig. 1.
`Patient preference was assessed by the question: "If
`you were given the choice of using one of the two
`systems (vial and syringe and InnoLet), which would
`you prefer?" A majority of patients (82% P-value <
`0.001) indicated a preference for the InnoLet device;
`only 8% indicated no preference in response to this
`question.
`Patient handling and acceptance of the devices was
`based upon the patient handling and acceptance ques(cid:173)
`tionnaire. Patient assessment of their handling expe(cid:173)
`rience with InnoLet is summarized in Fig. 2. Patients
`were asked, "Given your disability, how would you
`rate the InnoLet compared to vial and syringe: easier
`
`lnnolet
`(82%)
`
`The resource utilization was tabulated as a per(cid:173)
`centage of patients requiring/not requiring nurs-
`
`Fig. 1. Overall patient preference: "If you were given the choice
`of using one of the two systems (vial/syringe and lnnoLet), which
`would you prefer?".
`
`Sanofi Exhibit 2141.004
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`J. She/met et al. I Diabetes Research and Clinical Practice 63 (2004) 27-35
`
`31
`
`Table 2
`Resource utilization: proportion of patients requiring nursing/caregiver assistance
`
`No. of patients requiring assistance" (%)
`
`lnnolet (N = 72)
`
`Vial/syringe (N = 76)
`
`Steps of insulin injection
`Store supplies
`Prepare individual dosing
`Deliver individual dosing
`Attach needle to device
`Eliminate air from the needle
`Monitor subject's drawing up/dialing up of correct dose
`Prepare the site for injection (alcohol swab, pinch skin, etc.)
`Insure the patient has dosed themselves properly
`Return of supplies to the refrigerator
`Dispose of needles (lnnoLet)
`Dispose of syringe
`Dispose of vial
`Dispose of lnnoLet
`
`Overall independence
`Patients requiring nursing/caregivers time
`Patients not requiring nursing/caregivers time
`
`17 (24)
`16 (22)
`13 (18)
`15 (21)
`9 (13)
`19 (26)
`11 (15)
`16 (22)
`18 (25)
`8 (11)
`NIA
`NIA
`17 (24)
`
`34 (47)
`38 (53)
`
`18 (24)
`46 (61)
`37 (49)
`NIA
`33 (43)
`23 (30)
`13 (17)
`13 (17)
`19 (25)
`NIA
`24 (32)
`18 (24)
`NIA
`
`61 (80)
`15 (20)
`
`N/A: not applicable.
`a Times required for each step were collected. Patients having an estimated caregiver time that did not equal "0" min for that step are
`tabulated.
`
`to use, the same or harder to use?" A majority of the
`patients (82%) rated InnoLet easier to use. Only 4%
`found InnoLet harder to use, whereas 12% found In(cid:173)
`noLet and the vial/syringe to be the same.
`Patient acceptance of InnoLet is summarized in
`Fig. 3. Patients were asked: "Just after changing to
`InnoLet, how did you find managing the practical
`aspects (dosing and injecting) of the new insulin
`system?" About 86% of the patients reported that
`
`100
`
`~
`
`::R
`~ 80
`2
`co 60
`Cl:'.
`OJ
`C 40
`=o
`C co 20
`0
`
`I
`
`Easier
`
`Same
`
`Harder
`
`Fig. 2. Patient handling questionnaire. At the end of each study pe(cid:173)
`riod, patients were asked to complete a patient handling question(cid:173)
`naire. "Given the disability, rate lnnoLet compared to vial/syringe."
`Possible responses: (1) easier to use, (2) the same, and (3) harder
`to use.
`
`they found "managing the practical aspects" of in(cid:173)
`sulin injection ( e.g. dosing and injecting) by InnoLet
`to be "easy" or "very easy" when initiating insulin
`treatment, whereas 1 % of the patients found the use
`of InnoLet "difficult" or "very difficult" (Fig. 3A).
`When patients judged the practical aspects of insulin
`administration after switching to InnoLet, 97% of the
`patients considered the use of the InnoLet to be "easy"
`or "very easy," while 3% of the patients found the use
`of InnoLet to be "difficult" or "a little difficult" to use
`(Fig. 3B).
`When asked about the reliability of insulin injec(cid:173)
`tion, 62% rated Innolet as being more reliable than
`vial/syringe, 34% considered the treatment devices to
`be "about the same." Only 3% of patients considered
`InnoLet to be less reliable (Fig. 3C).
`Patients were asked: "Were you able to inject your(cid:173)
`self with less nursing assistance when using InnoLet?"
`A majority (84%) of patients indicated that they re(cid:173)
`quired less assistance when using InnoLet, 3% re(cid:173)
`quired more assistance, and 8% required about the
`same nursing assistance (Fig. 3D).
`Patients showed a positive response to the question:
`"How did you find the setting of the insulin dose with
`
`Sanofi Exhibit 2141.005
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`32
`
`J. She/met et al. I Diabetes Research and Clinical Practice 63 (2004) 27-35
`
`100
`
`80
`
`Cl)
`
`c
`Q) 60
`~
`a..
`~ 40
`
`0
`
`Very
`Easy
`
`Easy
`
`Little Difficult Very
`Difficult
`Difficult
`
`(B)
`
`20
`
`0
`
`Very
`Easy
`
`Easy
`
`Little Difficult Very
`Difficult
`Difficult
`
`100
`
`100
`
`80
`
`Cl)
`
`c
`Q) 60
`~
`a..
`~ 40
`
`0
`
`20
`
`0
`
`(A)
`
`100
`
`Q)
`
`Cl) 80
`c
`~ 60
`a..
`~
`0
`40
`
`20
`
`0
`
`80
`
`Cl)
`
`Q)
`
`c
`~ 60
`a..
`~
`0
`
`40
`
`20
`
`0
`
`Less
`
`Same
`
`More
`
`(C)
`
`More Reliable Same
`
`Less Reliable
`
`(D)
`
`100
`
`80
`
`.l!l
`C
`Q)
`
`~ 60
`a..
`~
`0
`
`40
`
`20
`
`0
`
`100
`
`.l!l 80
`
`C
`Q)
`~ 60
`a..
`~ 0
`
`40
`
`20
`
`0
`
`(E)
`
`Easy
`
`Difficult
`
`Very Difficult
`
`(F)
`
`No
`
`Sometimes Not Aware
`
`Fig. 3. Patient acceptance questionnaire. Answers to questions were scaled 0-100 (representing the least and most acceptance, respectively).
`(A) Initial experience-managing practical aspects: "Just after changing to InnoLet, how did you find managing the practical aspects
`( dosing and injecting) of the new insulin system?" Possible responses: (I) very easy and I learned the procedure very quickly, (2) easy,
`(3) a little difficult but I got used to it, (4) difficult, or (5) very difficult. (B) Current experience-managing practical aspects: "How do
`you now find the practical aspects of insulin administration after switching to InnoLet?" Possible responses: (I) very easy, (2) easy, (3)
`little difficult, (4) difficult, or (5) very difficult. (C) Reliability injecting insulin: "In terms of reliability of injecting your insulin, rate the
`lnnoLet." Possible responses: (I) more reliable than vial and syringe, (2) about the same, (3) less reliable. (D) Nursing assistance: "Were
`you able to inject yourself with less nursing assistance when using Irmo Let?" Possible responses: (I) required less nursing assistance, (2)
`required about the same nursing assistance, (3) required more nursing assistance. (E) Setting insulin dose: "How do you find setting of
`the insulin dose with InnoLet?" Possible responses: (I) easy, (2) difficult but I can manage, (3) very difficult and caused some problems
`for me. (F) Drawing incorrect insulin dose: "Did you ever draw up an incorrect insulin dose?" Possible responses: (I) no, never, (2) yes,
`sometimes, (3) I am not aware of it. (G) Pain of injection: "How did you find the injections?" Possible responses: (I) no pain at all, (2)
`slightly painful, (3) painful, (4) very painful, or (5) severely painful, hardly tolerable. (H) Self-injection: "Injecting myself .... " Possible
`responses: (I) has become a part of my daily life, (2) I dislike it but I do it without problems, (3) I do it with reluctance.
`
`Sanofi Exhibit 2141.006
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`J. She/met et al. I Diabetes Research and Clinical Practice 63 (2004) 27-35
`
`33
`
`100
`
`80
`
`Cl)
`
`c
`Q) 60
`~
`a..
`~ 40
`
`0
`
`100
`
`80
`
`Cl)
`
`c
`Q) 60
`~
`a..
`~ 40
`
`0
`
`20
`
`0
`
`(G)
`
`No
`Pain
`
`Slight Painful
`Pain
`
`Very Severely
`Painful Pain
`
`(H)
`
`20
`
`0
`
`Become part
`of life
`
`Dislike but
`do it with with
`no problem
`
`Do with
`reluctance
`
`Fig. 3. (Continued).
`
`Innolet?" Setting the insulin dosage with InnoLet was
`considered to be "easy" by 96% of patients (Fig. 3E).
`Patient acceptance of InnoLet was likewise high for
`the question: "Did you ever draw up an incorrect in(cid:173)
`sulin dose?" A majority of the patients (86%) reported
`that they had "never" drawn an incorrect dose, as com(cid:173)
`pared to 10% who sometimes drew an incorrect dose
`(Fig. 3F).
`When judging the pain of injections using InnoLet,
`a majority (73%) of the patients reported no pain at all,
`as compared to 26% who found Innolet to be slightly
`painful (Fig. 3G).
`When asked to evaluate their reaction to self(cid:173)
`administering daily injections, most patients (81 %)
`said that self-administration had "become a part of
`their daily life." Twenty percent patients said they
`"dislike it, but had no problem with the delivery
`system" (Fig. 3H).
`
`3. 7. Glycemic control
`
`During treatment of enrolled patients, no change
`in the glycemic control was observed during the en(cid:173)
`tire study period, as measured by fructosamine levels.
`For all patients who completed study treatment, the
`mean± S.D. for the serum fructosamine at screening
`was 268 ± 74.8 µmol/1, and at the end of the study
`was 269 ± 69 .4 µmol/1.
`
`3.8. Safety evaluation
`
`Forty-six treated patients reported 97 treatment(cid:173)
`emergent adverse events. Two (3%) patients of78 dur-
`
`ing InnoLet use reported four serious adverse events
`( chronic obstructive airway disease, cardiac failure,
`coagulation disorder, hypoglycemic coma) and 4 (5%)
`patients of 76 during vial/syringe treatment reported
`four serious adverse events (pneumonia, heart mur(cid:173)
`mur, hematemesis/ketosis, hyponatremia). There were
`no events that were considered by the investigator to
`be related to the study treatment.
`
`3.9. Adverse device effects
`
`Two adverse device effects were reported during the
`study (malfunctioning of the InnoLet device). These
`events were not associated with any adverse events.
`Two patients reported drippage from the needle fol(cid:173)
`lowing injection, possibly related to injection tech(cid:173)
`nique. No adverse device events were reported for use
`of vial/syringe.
`
`4. Discussion
`
`The difficulties of self-injection can be an obstacle
`to initiating insulin therapy at any age. Such difficul(cid:173)
`ties may be physical as well as psychological, with
`many people with type 2 diabetes being introduced
`to insulin self-injection at an advanced age. Some pa(cid:173)
`tients are uncomfortable with injection devices that
`appear technical or complex and some conditions such
`as poor eyesight or manual difficulties may pose phys(cid:173)
`ical obstacles to patient use of most injection devices
`previously available [6]. The likelihood of arthritis or
`tremors increases with advancing age [7] , adding to
`
`Sanofi Exhibit 2141.007
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`34
`
`J. She/met et al. I Diabetes Research and Clinical Practice 63 (2004) 27-35
`
`the difficulty of manually operating handheld devices.
`Clinical studies have shown that patients using a vial
`and syringe have more risk of drawing and injecting
`an inaccurate dose of insulin. In a study of patients
`with diabetes over the age of 40 years using tradi(cid:173)
`tional vial and syringe, Kesson and Bailie [8] found
`a relative error of about 19% in the accuracy of the
`doses drawn. The inaccuracy associated with patients
`injecting insulin can be a significant clinical problem,
`especially in the elderly. Puxty et al. [9] studied a
`population of syringe users over the age of 55 years
`(mean 66 years), their results showed a mean varia(cid:173)
`tion in dose of 12% when drawing up and expelling
`a dose of 20U. Sammler [10] found that 44% of sy(cid:173)
`ringe users over the age of 65 years injected doses of
`insulin that were more than 25% different from that
`prescribed. If errors in dose selection should occur,
`these errors may have profound consequences for pa(cid:173)
`tient safety. Hence, a large, easy-to-read dial could be
`an advantage. If patients believe that assistance from
`caregivers will be necessary, a reluctance to become
`dependent upon others may be an additional barrier to
`initiating insulin therapy [6].
`Complex insulin injection procedures, possibility of
`patient errors in drawing the insulin dose prescribed
`by the physician, and compliance with the treatment
`are perhaps the key points of insulin treatment safety
`in elderly patients with diabetes. The InnoLet insulin
`delivery system was developed to enable more pa(cid:173)
`tients to safely self-administer insulin, by providing
`a device that would be as intuitive and easily manip(cid:173)
`ulated as possible. The disposable InnoLet injection
`device provides a large "oven timer" scale, a large
`easily-handled injection button. The utility of such
`a simplified injection device was best assessed in a
`subset of patients having the most difficulties with
`self-injection. The aim of the present study was there(cid:173)
`fore to compare Innolet with vial/syringe in terms of
`utilization of healthcare resources, patient preference,
`and patient acceptance in elderly patients with dia(cid:173)
`betes who had difficulty in taking insulin injections
`without assistance.
`This clinical trial indicated that for the elderly dis(cid:173)
`abled patient population of this study (where about
`60% needed assistance in drawing up appropriate vol(cid:173)
`ume, 36% needed assistance making self-injections), a
`major portion of the patients (53%) did not need assis(cid:173)
`tance if they were provided with an Innolet injection
`
`system. This increased patient self-sufficiency com(cid:173)
`pared very favorably to the 20% who could manage
`alone with vial/syringe, and resulted in a significant
`reduction in costs of caregiver assistance (from US$
`196 to US$ 114 a day). In view of the fact that 82%
`of patients preferred the InnoLet device and consid(cid:173)
`ered it easier to use, InnoLet appeared to have quality
`of life implications as well.
`Fox et al. [4] recently evaluated patients with in(cid:173)
`sulin nai:ve type 2 diabetes and visual impairment
`(20/40 to 20/200) in terms of their preference and
`ability to operate three different insulin delivery sys(cid:173)
`tems: InnoLet, Humulin® NPH Pen (Eli Lilly), or
`vial/syringe (Becton-Dickinson). These investigators
`found that 87% of the patients had an overall prefer(cid:173)
`ence for use oflnnoLet, 13% preferred Humulin Pen
`(P < 0.001), and no patients preferred the syringe.
`In the United States, poor treatment compliance by
`patients with diabetes may often be due to a lack of
`educational support and the time constraints in the of(cid:173)
`fice of the primary care provider. Use of InnoLet could
`facilitate the initiation of insulin in more patients.
`Insulin delivery systems designed to simplify accu(cid:173)
`rate, reliable insulin delivery for people with visual
`impairment can improve the ability of such patients
`to repeatedly set and self-deliver the correct insulin
`dose. This study is additional evidence that an insulin
`injection delivery device can provide greater positive
`patient feelings than the traditional vial/syringe.
`
`5. Conclusions
`
`The InnoLet insulin injection device could reduce
`the cost of treatment by fostering independence in
`patients who had difficulties in self-injection using
`vial/syringe. Patients showed a preference for use of
`the InnoLet insulin injection device over the vial and
`syringe, with a majority of the patients reporting more
`confidence that the correct amount of insulin was
`injected using the InnoLet device. Elderly patients
`showed a higher treatment acceptance during use of
`InnoLet.
`
`Acknowledgements
`
`This study was supported by Novo Nordisk Pharma(cid:173)
`ceuticals Inc., Princeton, NJ. The authors wish to thank
`
`Sanofi Exhibit 2141.008
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`J. She/met et al. I Diabetes Research and Clinical Practice 63 (2004) 27-35
`
`35
`
`the site coordinators for their assistance during the
`trial: Kelly McCulloch, Karen Jones, Carrie-Ann Sil(cid:173)
`via, Joyce Stathopoulos, Stacy Frerichs, Debbie Kee,
`Billie Jo Shelmet, Paula Wells and Cindy Nelson.
`
`References
`
`[1] The Diabetes Control and Complications Trial Group,
`The effect of intensive
`treatment of diabetes and the
`development and progression of long-term complications in
`insulin-dependent diabetes mellitus, N. Engl. J. Med. 329
`(1993) 977-986.
`[2] United Kingdom Prospective Diabetes Study Group,
`Intensive blood-glucose control with sulphonylureas or
`insulin compared with conventional treatment and risk of
`complications in patients with type 2 diabetes, Lancet. 1998,
`352:837-853.
`[3] R.J. Rubin, W.A. Altman, D.N. Mendelson, Health care
`expenditures for people with diabetes mellitus, 1992, J. Clin.
`Endocrinol. Metab. 78 (1994) 809A-809F.
`
`[4] C. Fox, C. McKinnon, A. Wall, S.A. Lawton, Ability to
`handle, and patient preference for, insulin delivery devices in
`visually impaired patients with type 2 diabetes, Pract. Diab.
`Int. 19 (4) (2002) 104-107.
`[5] W.J. Conover, Practical nonparametric statistics, in: Wiley
`Series in Probability and Statistics: Applied Probability
`and Statistics Section, third ed., Wiley, New York, 1999,
`pp. 352-369.
`[6] E. Bashoff, R. Beaser, Insulin therapy and the reluctant
`patient, Postgrad. Med. 97 (1995) 86-90.
`[7] M. Frey, E. Barrett-Connor, P. Sledge, et al., The effect of
`non-insulin dependent diabetes mellitus on the prevalence of
`clinical osteoarthritis. A population based study, J. Rheumatol.
`23 (1996) 716-722.
`[8] C.M. Kesson, G.R. Bailie, Do patients with diabetes
`inject accurate doses of insulin? Diabetes Care 4 (1981)
`333.
`[9] J.A.H. Puxty, D.H. Hunter, W.A. Burr, Accuracy of insulin
`injection in elderly patients, Br. Med. J. (Clin. Res. ed.) 287
`(1983) 1762.
`[10] A. Sammler, Probleme alterer, insulinspritzender diabetiker
`be! deer ambulanten behandlung, Versiccherungsmedizin 42
`(1990) 59-64.
`
`Sanofi Exhibit 2141.009
`Mylan v. Sanofi
`IPR2018-01676
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket