DIABETES TECHNOLOGY & THERAPEUTICS
`Volume 14, Number 9, 2012
`© Mary Ann Liebert, Inc.
`DOI: 10.1089/dia.2011.0298
`
`ORIGINAL ARTICLE
`
`Diabetes Technology & Therapeutics
`
`Study on the Dosing Accuracy of Commonly Used
`Disposable Insulin Pens
`
`Meike Krzywon, Dipl.-lng. Chemistr/ Thomas van der Burg, B.Sc.,2 Uwe Fuhr, M.D.,3
`Manfred Schubert-Zsilavecz, Mag. pharm., Ph.D.]·4 and Mona Abdel-Tawab, B.Sc., Ph.D)
`
`Abstract
`
`Background: Improved patient comfort and optimal glycemic control have led to the widespread use of insulin pens,
`particularly in Europe. Most of the former studies on the dose accuracy of insulin pens included only a small number of doses
`and pens. In extension to our previous large-scale study testing the dosing accuracy following a randomized dosing sequence
`with each pen, the present study was more directed toward the dose accuracy for one specific dose dispensed repeatedly with
`the same pen. This is the first study providing detailed comparative data on the accuracy of repeated dose delivery with
`prefilled disposable insulin pens at low, middle, and high doses, dispensed over the entire pen volume.
`Materials and Methods: In total, 15 previously unused insulin pens from two lots of each pen type (SoloSTAR® [sanofi(cid:173)
`aventis, Paris, France], FlexPen® [Novo Nordisk A/S, Bagsv<£rd, Denmark], Next Generation FlexPen [Novo Nordisk], and
`KwikPen™ [Eli Lilly, Indianapolis, IN]) were used to deliver 5-unit (low), 30-unit (middle), and 60-unit (high) doses, re(cid:173)
`spectively, dispensed four times from each pen in a nonrandomized manner. Actual doses were determined gravimetrically
`taking the density of the respective insulin into account and were evaluated according to the guidelines (DIN EN ISO 11608-
`1 :2000) of the International Organization for Standardization (ISO).
`Results: All tested insulin pens met the requirements for accuracy with none of the single values being outside the defined
`range of the ISO recommendations (1±1 units, 30±1.5 units, and 60±3 units, respectively).
`Conclusion: The present study demonstrated a consistent and accurate dose delivery at all dosage levels for all tested insulin
`pens, with no clinically relevant differences among the products.
`
`Introduction
`
`SINCE THE FIRST INSULIN PEN was introduced in 1985, in(cid:173)
`
`sulin pens have significantly contributed to improved
`patient comfort and optimal glycemic control.1 They offer
`several advantages over the traditional vial-and-syringe
`method, such as reduced pain during injection, discretion of
`use, easy portability, and fewer claims for hypoglycemic
`events.2-4 Because of these advantages the use of insulin pens
`is widespread, particularly in Europe. 1
`As patients rely heavily upon device accuracy to administer
`the correct amount of insulin, the accuracy of insulin pens is
`an important issue. Insulin pens are available in two types:
`reusable and disposable insulin pens. The most commonly
`used disposable pens include the SoloSTAR® (SR) (sanofi(cid:173)
`aventis, Paris, France), the FlexPen® (FP) and Next Generation
`FlexPen (NGFP) (Novo Nordisk A/S, Bagsv<£rd, Denmark),
`
`and the KwikPen™ (KP) (Eli Lilly, Indianapolis, IN). Several
`studies have been dedicated to comparing the dosing accu(cid:173)
`racy among disposable insulin pens. In general, these studies
`verified the accurate dosing of disposable insulin pens, even
`when compared with the vial and syringe. 5
`13 However, they
`-
`followed different study designs with regard to the number of
`pens and doses tested, yielded partly contradicting results,
`and, with the exception of the study of Krzywon et al., 11 did
`not extend over the whole dosing range but mainly concen(cid:173)
`trated on testing 10-unit and 30-unit doses. Therefore our first
`study was dedicated to comparing the dose accuracy of SR,
`FP, NGFP, and KP over a wide dosing range (1 unit, 10 units,
`30 units, 60 units, and 80 units) dispensed in random order. 11
`The dose regimen of patients with diabetes may require a
`roughly constant daily dose of their basal insulin. For that
`reason the following study was aimed to complement our
`previous study by investigating the dose accuracy of the SR,
`
`1Central Laboratory of German Pharmacists, Eschborn, Germany.
`2Medical Devices/Quality, Regulatory & Risk Management, Sanofi-Aventis Deutschland GmbH, Frankfurt am Main, Germany.
`3Clinical Pharmacology Unit, Department of Pharmacology, University of Cologne, Cologne, Germany.
`4Institute of Pharmaceutical Chemistry, J.W. Goethe-University, ZAFES, Frankfurt am Main, Germany.
`The results of the present study have been presented as a poster at the American Diabetes Association's 70th scientific sessions in Orlando,
`Florida, held June 25-29, 2010.
`
`804
`
`Sanofi Exhibit 2132.001
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`DOSING ACCURACY OF DISPOSABLE INSULIN PENS
`
`805
`
`FP, NGFP, and KP at the low (5-unit), middle (30-unit), and
`high (60-unit) dosage level when administered repeatedly
`with the same pen.
`
`Materials and Methods
`
`SR and FP were bought from German pharmacies. KP and
`NGFP were bought in pharmacies in the United States and
`France. An overview of the included insulin pens and corre(cid:173)
`sponding needles is given in Table 1. The needles were ap(cid:173)
`plied according to the manufacturers' recommendation.
`In total, 45 previously unused insulin pens from two lots of
`each pen type (15 pens of each pen type for each dosage level,
`respectively) was used for the study. Each of the low (5-unit),
`middle (30-unit), and high (60-unit) doses was dispensed four
`times from each pen in a nonrandomized manner according to
`the dose delivery scheme presented in Figure 1. Between these
`recorded doses intermediate doses were dispensed without
`recording the mass to ensure dose dispensing over the entire
`pen volume.
`The individual insulin pens were operated according to the
`manufacturers' instructions. Prior to starting the sequence of
`measurements, one to two priming doses of 2 units were
`discarded. If still no drops were seen at the top of the needle,
`the priming dose was repeated until this was the case. Fol(cid:173)
`lowing the manufacturers' recommendation the priming do(cid:173)
`ses of the KP were repeated until a visible stream of insulin left
`the needle. All measurements were performed by a single
`investigator to eliminate potential user variability. As per
`manufacturers' instructions the plunger was kept pressed
`down for 10 s, 6 s, and 5 sin the case of SR, both FP and NGFP,
`and KP, respectively, after each dose to ensure that all the
`dialed dose had been expelled. Each dose was deposited in a
`beaker containing a 0.5-1 cm layer of liquid paraffin, whereas
`the pen was held close to the surface of the paraffin layer. In
`case an insulin drop remained at the tip of the needle at the
`end of the relaxation time, this drop was stripped off at
`the paraffin surface, taking care that the needle did not strike
`the paraffin. Afterward the dose was weighed immediately
`using an analytical balance (model XP205/M; Mettler Toledo
`AG, Giefsen, Germany), which has an accuracy of 0.00001 g.
`The balance was zeroed before each dose of insulin was de(cid:173)
`posited and weighed. The weights were corrected for the
`specific density of each insulin formulation determined in
`the run-up to the study. For insulin glargine (SR), insulin
`detemir (FP and NGFP), and insulin lispro (KP), the density
`was determined to be 1.0036 g/ cm 3, 1.00798 g/ cm 3, and
`1.00447 g/ cm 3, respectively, using a DMA 4500 density meter
`(Anton Paar GmbH, Bruchkobel, Germany). For each dose
`
`Delivery of the 1st dose of 5U , 30U, and 60U,
`respectively, weighing and recording the mass
`
`Intermediate dose without recording the mass: 60 + 28U for
`the low dose test pens, 55U for the mid dose and 15U for
`the high dose test pens, respectively
`
`Delivery of the 2nd dose of 5U, 30U, and 60U ,
`respectively, weighing and recording the mass
`
`Intermediate dose without recording the mass: 60 + 28U for
`the low dose test pens, 55U for the mid dose and 15U for
`the high dose test pens, respectively
`
`Delivery of the 3rd dose of 5U, 30U, and 60U ,
`respectively, weighing and recording the mass
`
`Intermediate dose without recording the mass: 60 + 28U for
`the low dose test pens, 55U for the mid dose and 15U for
`the high dose test pens, respectively
`
`Delivery of the 4th 5U , 30U, and 60U dose,
`respectively, weighing and recording the mass
`
`FIG. 1. Dose delivery scheme for each pen.
`
`TABLE 1. INSULIN PENS AND CORRESPONDING NEEDLES INCLUDED IN THE STUDY
`
`Manufacturer
`
`Batch
`
`Insulin
`
`Needles
`
`40Ul42
`40U144
`VH 70047
`VH 70215
`VH 70007
`VH70235
`A 477063
`A 463790C
`
`Glargine
`
`BD Micro-Fine (0.25mm [31 gauge]x8)
`
`Detemir
`
`NovoFine (0.3 mm [30 gauge] x 8)
`
`Detemir
`
`NovoFine (0.3 mm [30 gauge] x 8)
`
`Lispro
`
`BD Micro-Fine (0.25mm [31 gauge]x8)
`
`Sanofi Exhibit 2132.002
`Mylan v. Sanofi
`IPR2018-01676
`
`Insulin pen
`
`SoloSTAR
`
`FlexPen
`
`sanofi-aventis
`
`Novo Nordisk
`
`Next Generation FlexPen
`
`Novo Nordisk
`
`KwikPen
`
`Eli Lilly
`
`

`

`806
`
`KRZYWON ET AL.
`
`-9-SoloSTAR
`
`- - •- - FlexPen
`
`•••• ,.. ••• NG FlexPen
`
`- t i - KwikPen
`
`5 U dose
`
`0
`
`2
`
`3
`
`4
`
`5
`
`6
`
`7
`8
`Pen(Number)
`
`9
`
`10
`
`11
`
`12
`
`13
`
`14
`
`15
`
`--e-SoloSTAR
`
`- - • - - FlexPen
`
`···•···· NG FlexPen ~ KwikPen
`
`30 U dose
`
`-
`.
`._ ,,
`• ""• ...... -'..c••···-.-..-=-....__
`............................ -·----------·-·· --............. ..
`
`-'2
`
`~
`QI
`U)
`0
`Cl
`:§
`
`;;i
`U)
`.E
`
`6.5
`
`6
`
`5.5
`
`5
`
`4.5
`
`4
`
`3.5
`
`32
`
`31.5
`
`- 31
`-·c
`
`~ 30.5
`QI
`U)
`0
`Cl
`.!:
`3
`Ill 29.5
`.E
`
`30
`
`2
`
`3
`
`4
`
`5
`
`6
`
`7
`
`8
`
`9
`
`10
`
`11
`
`12
`
`13
`
`14
`
`15
`
`Pen(Number)
`
`~ SoloSTAR
`
`- - .- - FlexPen
`
`•···A.···· NG FlexPen
`
`---.!l-
`
`KwikPen
`
`60 U dose
`
`29
`
`28.5
`
`28
`
`0
`
`-·2
`
`~
`QI
`U)
`0
`Cl
`:§
`;;i
`Ill
`.E
`
`64
`
`63
`
`62
`
`61
`
`60
`
`59
`
`58
`
`57
`
`56
`
`0
`
`2
`
`3
`
`4
`
`5
`
`6
`
`8
`7
`Pen(Number)
`
`9
`
`10
`
`11
`
`12
`
`13
`
`14
`
`15
`
`FIG. 2. Distribution of the average actual doses according to pen and dosage level (n = 4). Bold lines represent the Inter(cid:173)
`national Organization for Standardization limits, and the midline represents the target dose. NG, Next Generation.
`
`Sanofi Exhibit 2132.003
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`DOSING ACCURACY OF DISPOSABLE INSULIN PENS
`
`807
`
`application, a new injection needle was used, which was
`primed in accordance with the manufacturer's recommen(cid:173)
`dation before dose delivery. The whole study was carried out
`under Good Manufacturing Practice conditions and was ad(cid:173)
`ditionally documented by video.
`The evaluation of dose accuracy was based on the guide(cid:173)
`lines (DIN EN ISO 11608-1:2000) of the International Orga(cid:173)
`nization for Standardization (ISO). 14 The statistical tolerance
`interval [x ± (k • s)] (see Results) for each pen should lie within
`the upper and lower acceptance limits for each dosage level.
`According to the ISO guidelines the calculated statistical tol(cid:173)
`erance limit should not deviate from the target dose by more
`than 1 unit for the delivery of 5 units and not by more than 5%
`for the delivery of 30 units and 60 units. For the individual
`doses tested in this study the acceptance limits are 5 ± 1 units
`(4.0-6.0 units), 30 ± 1.5 units (28.5-31.5 units), and 60 ± 3 units
`(57-63 units). In addition, the arithmetic average of the actual
`doses, the SD, the coefficient of variation, and the average
`deviation in units and percentage from the target dose were
`calculated.
`
`Results
`
`According to the delivery scheme 60 doses were gravi(cid:173)
`metrically measured at each dosage level for each pen type,
`and the actual dose was calculated on the basis of the deliv(cid:173)
`ered masses and the density of the insulin solution. In general,
`the 5-unit doses were equally distributed around the target
`dose, whereas most of the 30-unit and almost all of the 60-unit
`doses were found to be below the target dose (Fig. 2). How(cid:173)
`ever, none of the tested insulin pens showed any trend toward
`increasing or decreasing doses with repetitive dosing.
`Despite the high number of doses measured (720 in total,
`180 per pen model) the study demonstrated a consistent and
`accurate dose delivery at all dosage levels for all tested insulin
`pens with none of the single values outside the specified limits
`recommended by ISO.
`The arithmetic average of the actual doses, the SD, and the
`coefficient of variation, as well as the average deviation in
`units and the average relative percentage deviation from the
`target dose, are summarized in Table 2.
`
`The average values determined for each pen at each dosage
`level were found to lie close to each other and to the target
`dose. For all tested insulin pens the average deviation of the
`actual dose from the target dose was below 2% at all tested
`dosage levels. The average deviations from the target dose
`ranged between -1.14% and -0.06% at 5 units, between
`- 1.01% and -0.11% at 30 units, and between -1.32% and
`- 0.55% at 60 units. Moreover, only marginal differences be-
`tween the tested insulin pens in the reproducibility of dose
`delivery were detected. The highest variation was observed at
`the lowest dosage level of 5 units with values ranging between
`3.1 % and 4.1 %, whereas at the higher dosage levels (30 units
`and 60 units) the coefficient of variation ranged between 0.5%
`and 0.9%.
`The statistical tolerance intervals determined for each pen
`at each dosage level are presented in Figure 3. All calculated
`tolerance intervals [x ± (k • s)] were found to lie within the
`acceptance range for each dosage level for each pen, where xis
`the average value of the actual doses for each pen at each
`dosage level, s = SD, and k is the tolerance limit factor, which
`was found to be 2.670 on the basis of the 95% confidence
`interval for n = 60.14
`
`Discussion
`
`Six studies examined the comparable dosing accuracy of
`disposable insulin pens.5
`11 In general, dosing accuracy was
`-
`found to be good in all studies. Two studies comparing the SR
`with the FP reported few doses outside the ISO limits for both
`pens.5
`6 However, the number of pens included in these
`'
`studies was very small (eight pens and three pens, respec(cid:173)
`tively, of each type) and did not meet the ISO recommenda(cid:173)
`tions. Therefore these data may not be considered robust. In
`fact, repeating the study of Asakura et al. 6 on a 10-fold higher
`number of pens (30 pens each) revealed comparable accuracy
`of the SR and the FP with no single value of the 2,280 doses
`measured being outside the ISO limits.8 Similar good results
`within the ISO limits were obtained for the NGFP, when
`comparing it with the FP.9 Comparing the SR with the NGFP
`revealed good performance of both pens but reported few
`doses outside the ISO limits for both pens. 10 All the above-
`
`TABLE 2. OVERVIEW ON THE AVERAGE OF THE ACTUAL DOSES, THE STANDARD DEVIATION, THE COEFFICIENT
`OF VARIATION, THE AVERAGE DEVIATION IN UNITS, AND THE AVERAGE RELATIVE PERCENTAGE DEVIATION
`FROM THE TARGET DosE FOR ALL PENS AT EACH DOSAGE LEVEL
`
`Target
`dose (units)
`
`Actual dose
`
`Average dose (units)
`
`SD (units)
`
`CV%
`
`Average
`deviation (units)
`
`Average relative
`deviation (%)
`
`5
`
`30
`
`60
`
`4.943
`4.972
`4.969
`4.997
`29.769
`29.698
`29.729
`29.968
`59.283
`59.247
`59.206
`59.667
`
`0.153
`0.202
`0.188
`0.160
`0.253
`0.275
`0.210
`0.215
`0.357
`0.324
`0.503
`0.517
`
`3.1
`4.1
`3.8
`3.2
`0.8
`0.9
`0.7
`0.7
`0.6
`0.5
`0.9
`0.9
`
`-0.057
`-0.028
`-0.031
`-0.003
`-0.231
`-0.302
`-0.271
`-0.032
`-0.717
`-0.753
`-0.794
`-0.333
`
`-1.14
`-0.57
`-0.62
`-0.06
`-0.77
`-1.01
`-0.90
`-0.11
`-1.19
`-1.26
`-1.32
`-0.55
`
`Pen
`
`SR
`FP
`NGFP
`KP
`SR
`FP
`NGFP
`KP
`SR
`FP
`NGFP
`KP
`
`CV, coefficient of variation; FP, FlexPen; KP, KwikPen; NGFP, Next Generation FlexPen; SR, SoloSTAR.
`
`Sanofi Exhibit 2132.004
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`808
`
`6
`
`-C 5
`
`::J
`
`4
`
`31.5
`
`-§30.0
`
`5 U dose
`
`5.51
`
`5.47
`
`5.42
`
`4.44
`
`4.47
`
`4.57
`
`-SR -
`
`FP -
`
`NGFP
`
`KP
`
`30 U dose
`
`30.44
`
`30.43
`
`30.54
`
`30.28
`
`29.40
`
`28.97
`28.5 ~ - - - - - - - - - - - - - - - - - - - -
`NGFP
`KP
`-SR-
`FP
`
`29.17
`
`60 U dose
`6 3 - . - - - - - - - - - - - - - - - - - - - - - -
`
`-§ 60
`
`61.04
`
`60.54
`
`58.34
`
`58.39
`
`58.30
`
`57.87
`57~ - - - - - - - - - - - - - - - - - - - -
`
`FIG. 3. Tolerance intervals for each pen at each dosage le(cid:173)
`vel. Bold lines represent the International Organization for
`Standardization limits, and the midline represents the target
`dose. FP, FlexPen; KP, KwikPen; NGFP, Next Generation
`FlexPen; SR, SoloSTAR.
`
`mentioned studies included only a small number of doses and
`pens. In terms of an overall assessment of the dosing accuracy
`we conducted the first comparable study on a large scale in(cid:173)
`cluding the most common insulin pens, SR, FP, NGFP, and
`KP, covering the low (1 unit), the middle (30 and 40 units), and
`the high (60 units) dosage levels for all pens in addition to the
`80-unit dosage level for the SR in line with ISO guidelines. 11
`
`KRZYWON ET AL.
`
`The 10-unit dose was also included because this dose was part
`of other comparative studies. This largest comparative study
`carried out so far revealed an excellent dosing accuracy for all
`tested insulin pens, with none of the single values (1,260 in
`total) at all dose levels being outside the defined range of the
`ISO recommendation. Previously reported dosing outside the
`6
`ISO limits5
`10 could not be verified in this study. 11
`'
`'
`Following the ISO recommendation, the 1-unit dosage level
`should have been chosen for the low dosage level for all pens
`and the 40-unit and 80-unit dose as middle and high doses,
`respectively, for SR. Our current study assessed the dosing
`accuracy at the 5-unit (low), 30-unit (middle), and 60-unit
`(high) dosage levels for all pens, to enable comparability with
`previously published studies and because of the greater
`clinical relevance of the 5-unit dose compared with the 1-unit
`dose. The results obtained represent the dosing accuracy in
`the hands of a professional and do not take into account
`variabilities that may be introduced by users with diabetes in
`daily practice.
`All tested insulin pens revealed an excellent repetitive
`dosing accuracy in the low, middle, and high dosage levels,
`thus assuring dose accuracy over the whole insulin pen. The
`tolerance limits defined by the ISO standards were met by all
`pens at all dosage levels. No single dose from SR, FP, NGFP,
`and KP at any dosage level was delivered outside the pre(cid:173)
`specified limits of the ISO guidelines. Again, previously
`published data that reported individual doses below the ISO
`limit for the SR, FP, 5
`6 and NGFP10 could not be verified. For
`'
`all pens at all dosage levels only minor deviations from the
`target dose not exceeding 1.32% were observed. Being well
`within the defined ISO limits, these deviations may be re(cid:173)
`garded negligible in daily practice.
`The comparable dosing accuracy of all insulin pens revealed
`in this study confirms the results of our earlier study using a
`randomized dosing sequence11 and the results of Penfomis and
`Horvat, 8 demonstrating comparable dosing accuracy of SR and
`FP at the 5-unit and 30-unit dosage level. The outcome of this
`study is also in line with a clinical study carried out by Frie(cid:173)
`drichs 15 demonstrating comparable dosing accuracy for both
`SR and FP when used by device-naive subjects.
`
`Conclusions
`
`The present study reveals similar repetitive dosing accu(cid:173)
`racy for the tested insulin pens SR, FP, NGFP, and KP. None of
`the four insulin pens delivered at the low, middle, and high
`dosage level doses outside the prespecified limits of the ISO
`guidelines. Existing marginal differences among the pens in
`the deviation of the actual doses from the target dose are
`negligible in daily practice.
`
`Acknowledgments
`
`The authors thank Stefan Kamlot for the fruitful discus(cid:173)
`sions and his valuable contribution to the manuscript. This
`study was sponsored by sanofi-aventis.
`
`Author Disclosure Statement
`
`T.v.d.B. is an employee of sanofi-aventis. M.K., M.S.-Z.,
`M.A.-T., and U.F. declare no competing financial interests.
`T.v.d.B. was involved in the design of the present study. M.K.,
`M.S.-Z., and M.A.-T. were responsible for the realization/
`
`Sanofi Exhibit 2132.005
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`DOSING ACCURACY OF DISPOSABLE INSULIN PENS
`
`809
`
`evaluation of the study, and U.F. was responsible for its sta(cid:173)
`tistical evaluation.
`
`References
`
`1. Thurman JE: Insulin pen injection devices for management
`of patients with type 2 diabetes: considerations based on an
`endocrinologists's practical experience in the United States.
`Endocr Pract 2007;13:672-678.
`2. Davis EM, Bebee A, Crawford L, Destache C: Nurse satis(cid:173)
`faction using insulin pens. Expert Opin Drug Deliv Diabetes
`Educ 2009;35:799-809.
`3. Robertson KE, Glazer NB, Campbell RK: The latest devel(cid:173)
`opments
`in
`insulin
`injection devices. Diabetes Educ
`2000;26:135-138, 141-146, 149-152.
`4. Lee WC, Balu S, Cobden D, Joshi AV, Pashos CL: Medication
`adherence and
`the associated health-economic
`impact
`among patients with type 2 diabetes mellitus converting to
`insulin pen therapy: an analysis of third-party managed care
`claims data. Clin Ther 2006;28:1712-1725.
`5. Haenel H, Weise A, Sun W, Pfi.itzner JW, Thome N, Pfi.itzner
`A: Differences in the dose accuracy of insulin pens. J Dia(cid:173)
`betes Sci Technol 2008;2:478-481.
`6. Asakura T, Seino H, Kageyama M, Yohkoh N: Dosing ac(cid:173)
`curacy of two insulin pre-filled pens. Curr Med Res Opin
`2008;24:1429-1434.
`7. Pfuetzner A, Asakura T, Sommavilla B, Lee W: Insulin de(cid:173)
`livery with FlexPen: dose accuracy, patient preference and
`adherence. Expert Opin Drug Deliv 2008;5:915-922.
`8. Penfornis A, Horvat K: Dose accuracy comparison between
`SoloStar and FlexPen at three different dose levels. Diabetes
`Technol Ther 2008;10:359-362.
`9. Pfi.itzner A, Reimer T, Hohberg C, Frokjaer LP, Jorgensen C:
`Prefilled insulin device with reduced injection force: patient
`
`perception and accuracy. Curr Med Res Opin 2008;24:
`2545-2549.
`10. Weise A, Pfi.itzner JW, Borig J, Pfi.itzner AM, Safinowski M,
`Hanel H, Musholt PB, Pfi.itzner A: Comparison of the dose
`accuracy of prefilled insulin pens. J Diabetes Sci Technol
`2009;3:149-153.
`11. Krzywon M, van der Burg T, Fuhr U, Schubert-Zsilavecz
`M, Abdel-Tawab M: Dosing accuracy of common dis(cid:173)
`posable insulin pens. Curr Med Res Opin 2010;26:901-
`905.
`12. Asakura T, Seino H, Nakano R, Muto T, Toraishi K, Sako Y,
`Kageyama M, Yohkoh N: A comparison of the handling and
`accuracy of syringe and vial versus prefilled insulin pen
`(Flex pen). Diabetes Technol Ther 2009;11:657-661.
`13. Luijf YM, DeVries JH: Dosing accuracy of insulin pens
`versus conventional syringes and vials. Diabetes Technol
`Ther 2010;12(Suppl 1):S73-S77.
`14. International Organization
`for Standardization: Pen(cid:173)
`Injectors for Medical Use-Part 1: Pen-Injectors-Requirements
`and Test Methods. ISO 11608-1:2000, Version 1. Geneva: In(cid:173)
`ternational Organization for Standardization, 2000.
`15. Friedrichs A: Dose accuracy of SoloSTAR and FlexPen as
`assessed in a clinical setting. Diabetes Technol Ther 2009;
`11:609-613.
`
`Address correspondence to:
`Mona Abdel-Tawab, B.Sc., Ph.D.
`Central Laboratory of German Pharmacists
`Carl-Mannich-Strafie 20
`65760 Eschborn, Germany
`
`E-mail: M.Tawab@zentrallabor.com
`
`Sanofi Exhibit 2132.006
`Mylan v. Sanofi
`IPR2018-01676
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket