DIABETES TECHNOLOGY & THERAPEUTICS
`Volume 10, Number 4, 2008
`© Mary Ann Liebert, Inc.
`DOI: 10.1089/dia.2008.0035
`
`Dosing Accuracy with a Novel Pen Device (SoloST AR®)
`as Performed by Patients with Diabetes in a Clinical Setting
`
`Norbert Hermanns, Ph.D., Bernhard Kulzer, Ph.D., and Thomas Haak, M.D.
`
`Abstract
`
`Background: Pen devices can help to overcome some of the barriers associated with insulin therapy. The pres(cid:173)
`ent study evaluated the accuracy of dose delivery by people with diabetes using the novel prefilled, dispos(cid:173)
`able SoloSTAR® device with insulin glargine (Lantus®) and insulin glulisine (Apidra®) (all from sanofi-aven(cid:173)
`tis, Paris, France).
`Methods: People with type 1 or type 2 diabetes (insulin users or insulin naive) were eligible to participate in
`this randomized, single-center, open-label study. Each participant delivered six separate insulin doses into a
`sponge using SoloSTAR (three with glargine [10, 40, and 80 units] and three with glulisine [5, 15, and 30 units]).
`Pens were weighed before and after each test dose to determine the dose delivered. Thresholds for dosing ac(cid:173)
`curacy were calculated according to the 2000 International Organization for Standardization (ISO) recommen(cid:173)
`dations (Guideline 11608-1).
`Results: All doses of glargine and glulisine delivered (60 participants; 360 individual doses) were within the
`ISO limits. Mean (standard deviation) glargine doses delivered were 9.87 (0.24), 39.63 (0.36), and 79.02 (0.62)
`units for 10, 40, and 80 units, respectively. Insulin glulisine doses delivered were 4.98 (0.20), 14.87 (0.29), and
`29.67 (0.34) units for 5, 15, and 30 units, respectively.
`Conclusions: The SoloST AR pen allows people with diabetes to achieve a dosing accuracy with glargine and
`glulisine similar to that achieved in laboratory conditions. The dosing accuracy and ease of use of SoloST AR
`may provide greater confidence in the precision and accuracy of the device while titrating glargine and/ or
`glulisine to goal.
`
`Introduction
`
`A LTHOUGH THE EFFICACY of insulin therapy is well estab(cid:173)
`
`lished in terms of reducing the risk of diabetes-related
`complications, 1,2 many barriers to its successful use by pa(cid:173)
`tients with diabetes exist, including hypoglycemia, social ac(cid:173)
`ceptability, and the individual's perceived ability to self(cid:173)
`manage his or her treatment with injections that may be
`difficult and painful to administer. 3
`The introduction of pen devices has increased patients' ac(cid:173)
`ceptance of insulin therapy, by helping to overcome the fear
`of injections associated with using a standard syringe.4,5 Pens
`can also increase accuracy of dosing (and, therefore, glyce(cid:173)
`mic control) in people with diabetes. 6- 8 However, the abil(cid:173)
`ity of patients to accurately administer the correct insulin
`dose can vary between pen devices. 6
`Accurate dosing with an easy-to-use device can increase
`a patient's confidence in his or her ability to self-manage di-
`
`abetes,6,9 leading to improved treatment adherence. 10,11
`SoloSTAR® (sanofi-aventis, Paris, France) is a disposable,
`prefilled insulin pen device for administration of insulin
`glargine (Lantus®; sanofi-aventis) or
`insulin glulisine
`(Apidra®; sanofi-aventis). Unlike other disposable insulin
`pens, the body of the SoloST AR pen is color coded to aid
`with the correct identification of basal and bolus insulins,
`even in patients with impaired vision. 12 In addition, the
`SoloST AR pen requires a lower injection force and can de(cid:173)
`liver a higher maximum dose than some other disposable
`pens. 13
`Laboratory studies have shown that using the SoloSTAR
`pen to deliver insulin glargine is highly accurate for three
`different doses, 13 according to procedures and standards de(cid:173)
`fined by the International Organization for Standardization
`(ISO) (Guideline 11608-1).14 However, accuracy in the lab
`does not necessarily reflect the real-life setting in which pa(cid:173)
`tients administer their own injections. Therefore, the present
`
`FIDAM GmbH, Forschungsinstitut Diabetes-Akademie, Bad-Mergentheim, Germany.
`
`322
`
`Sanofi Exhibit 2130.001
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`SOLOSTAR DOSE ACCURACY BY PATIENTS
`
`323
`
`TABLE 1. PATIENT CHARACTERISTICS
`
`Characteristic
`
`Total number of patients
`Mean ± SD age (years)
`Female (n [%])
`Type of diabetes (n [%])
`Type 1
`Type 2
`Mean ± SD duration of diabetes (years)
`Diabetes therapy (n [%])
`Oral antidiabetic drugs
`Oral antidiabetic drugs + insulin
`Insulin
`Previous insulin pen experience (n [%])
`Mean ± SD duration of insulin therapy
`(years)
`Moderate motor impairment (n [%])
`Moderate visual impairment (n [%])
`
`study was designed to determine the dosing accuracy of
`SoloSTAR, when used by patients with diabetes to deliver
`clinically relevant doses of insulin glargine (10, 40, and 80
`units) and insulin gluisine (5, 15, and 30 units).
`
`Research Design and Methods
`
`Study design
`
`This was a prospective, open-label, single-center study,
`which was conducted over a period of 1 week at our dia(cid:173)
`betes research clinic (Forschungsintitut Diabetes GmbH,
`Bad-Mergentheim, Germany).
`
`Patients
`
`A minimum of 54 patients was considered sufficient to
`demonstrate validity of the study; considering that 10% of
`dose deliveries could be non-evaluable, the target sample
`size was 60 patients. Men and women 18-79 years old with
`
`60
`55.9 ± 13.0
`24 (40)
`
`11 (18)
`49 (82)
`14.0 ± 9.5
`
`6 (10)
`22 (37)
`32 (53)
`53 (88)
`8.1 ± 6.4
`
`7 (12)
`9 (15)
`
`• • •• • •• • . .........
`• • ... •
`....
`•
`
`♦
`
`...
`
`•
`...
`
`~
`
`T
`
`A
`
`'ii,
`~
`C:
`2.
`e Q)
`"C
`. 2:
`Q)
`"C
`Q)
`VI
`0
`C
`
`11 .5
`
`11 .0
`
`10.5
`
`10.0
`
`9.5
`
`9.0
`
`8.5
`0
`
`B
`
`... T.
`
`I
`
`. ..........
`
`•
`
`•
`•
`•
`• •
`•
`•
`•
`
`50
`
`60
`
`I
`
`I
`
`• •• •
`
`Upper limit (11.0 units) according to ISO 11608-1
`...
`•• •
`•
`••
`• •
`•
`
`...
`Lower limit (9 .0 units) according to ISO 11608-1
`
`10
`
`20
`
`30
`Dose number
`
`40
`
`Upper limit (42 .0 units) according to ISO 11608-1
`
`. ..
`
`...
`
`♦ ♦
`
`...
`
`♦
`♦
`
`j
`43.0
`C:
`2. 42.0
`'g 41.0
`... Q)
`40.0
`.2:
`Q)
`"C
`~ 38.0
`0
`C
`
`37.0
`
`39.0
`
`'ii, 86.0
`~
`C:
`2.
`e Q)
`"C
`~
`Q)
`"C
`Q)
`VI
`0
`C
`
`84.0
`
`82.0
`
`80.0
`
`78.0
`
`76.0
`
`74.0
`
`... •
`•
`•••• ••
`... ••
`...
`T
`
`♦
`
`•
`...
`...
`....
`♦ • ••••
`• • ♦ • • • ••
`
`...
`:
`•• •••
`
`♦
`♦
`
`♦
`♦ :
`
`Lower limit (38.0 units) according to ISO 11608-1
`
`0
`
`10
`
`20
`
`30
`Dose number
`
`40
`
`Upper limit (84.0 units) according to ISO 11608-1
`
`•
`
`60
`
`50
`
`I
`
`I
`
`C
`
`......
`
`T
`
`.. ••• ♦
`
`•
`♦ . . . . . • • • • • • ♦
`
`♦ •
`
`♦
`
`♦ ...
`
`•
`
`♦
`
`♦ ♦
`
`•
`
`•
`
`• T···· • • ♦ • • • T • • • ♦ T .
`........
`I
`
`Lower limit (76 .0 units) according to ISO 11608-1
`
`0
`
`10
`
`20
`
`30
`Dose number
`
`40
`
`50
`
`60
`
`FIG. 1. Accuracy of the SoloSTAR pen when delivering (A) 10-, (B) 40-, and (C) 80-unit doses of insulin glargine. The up(cid:173)
`per and lower limits, as defined by the ISO standard, 11608-1, are indicated for each dose. Each data point represents the
`dose delivered by an individual person.
`
`Sanofi Exhibit 2130.002
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`324
`
`HERMANNS ET AL.
`
`a confirmed diagnosis of type 1 or type 2 diabetes were eli(cid:173)
`gible to participate in the study. Patients were required to
`be current or past users of injectable insulin or to be insulin
`naive and considered to be candidates for initiation of in(cid:173)
`sulin therapy. Current insulin users with experience of dis(cid:173)
`posable pens (including SoloSTAR) and/or reusable pens
`were eligible to participate in the study. Patients were re(cid:173)
`quired to speak and read German.
`Exclusion criteria included the presence of a substance use
`disorder, severe visual or motor impairment, or a diagnosis
`of dementia, as recorded in the patient's medical files, which
`were reviewed after an initial screening visit. Patients with
`mental illnesses rendering them unable to understand the re(cid:173)
`quirements and implications of the study and those consid(cid:173)
`ered unlikely to adhere to the study protocol were also ex(cid:173)
`cluded. All patients provided written informed consent prior
`to study entry.
`A total of 63 patients participated in the study (Table 1).
`The majority of patients had type 2 diabetes (82%), the mean
`disease duration was 14 years, and 88% had previous expe(cid:173)
`rience with insulin pens. Seven participants had moderate
`motor impairment, and nine had moderate visual impair(cid:173)
`ment. All participants were instructed in the use of the
`SoloSTAR pens prior to the study. Results from three par(cid:173)
`ticipants were excluded. In one case, 15 units of insulin gluli(cid:173)
`sine was injected twice instead of one 5-unit and one 15-unit
`injection, in another case, a single patient was excluded for
`providing a second set of measurements, and in the third
`case, a patient with severe motor impairment was incorrectly
`included.
`
`Study design
`
`Patients were required to deliver six separate insulin in(cid:173)
`jections into a sponge. Study nurses demonstrated the use of
`SoloSTAR pen. Patients were asked if they had any ques(cid:173)
`tions regarding the use of the SoloSTAR pen. After that pa(cid:173)
`tients were requested to give a test injection of 10 units into
`a sponge.
`Each participant dialed and delivered three doses with the
`insulin glargine SoloSTAR pen (10, 40, and 80 units) and
`three doses with the insulin glulisine SoloSTAR pen (5, 15,
`and 30 units); the order of pen usage and dose delivery was
`randomized. Between the study doses, the pens were primed
`with an insulin dose of 2 units. The randomization list was
`generated by the lead investigator using SYSTAT (Systat
`Software, Inc., Chicago, IL) before the start of the study with
`allocations sealed in envelopes, which were opened by the
`nurse during the participant's visit.
`The SoloSTAR pens were weighed by the investigators fol(cid:173)
`lowing the priming doses and again following the test in(cid:173)
`jections delivered by the patients, using a precision balance
`(P 205 DR/M, Mettler Toledo GmbH, Giessen, Germany).
`Pens were fitted with 0.25-mm (31-gauge) X 8-mm needles
`(Becton Dickinson GmbH, Heidelberg, Germany) for all
`tests.
`
`Statistical analysis
`
`Dosing accuracy was assessed by comparing pre- and
`post-injection weights of the SoloSTAR pens. Mean± stan(cid:173)
`dard deviation (SD) of the delivered doses were calculated
`for each test dose for both insulins. The level of dosing ac-
`
`curacy was assessed by applying ISO analyses. 14 In brief, ISO
`11608-1, which is applicable for insulin pen devices, provides
`a set of criteria against which the dosing accuracy of an in(cid:173)
`sulin pen is assessed. Results are normally presented as mil(cid:173)
`liliters; for insulin formulations of 100 units/mL, results can
`be presented in units for easier interpretation (as has been
`done here).
`The ISO limits are as follows:
`
`• For doses of 2-20 units, the tolerance is ±5% (e.g., 38-42
`units for the 40-unit dose; Eq. 1)
`• For doses <20 units, the tolerance is ±0.01 mL (e.g., 4-6
`units for the 5-unit dose), as defined in the ISO recom(cid:173)
`mendations (ISO 11608-1)
`
`Further calculations were performed according to ISO
`11608-1, as shown in Eq. 2, to confirm whether the data sets
`were within the threshold limits. This calculation is valid for
`k values of 2.667, for a P value of 0.975, and with a 95% con(cid:173)
`fidence interval (CI).
`Upper threshold value = Vset + Vset · 5%
`
`Lower threshold value = Vset - Vset · 5%
`
`(1)
`
`where Vset = volume set (i.e., 40 units [0.4 mL]).
`x + (k · s) -,c:: upper threshold value
`
`x -
`
`(k · s) 2- lower threshold value
`
`(2)
`
`where s = standard deviation and x = mean dose delivered.
`The k value was derived from a lookup table, which was
`based on the number of samples and a P value of 0.975 with
`a 95% CI (using the Clopper-Pearson method). 14 The k value
`used was 2.667.
`A third evaluation was the proportion of doses delivered
`by the patients that were within the ISO ranges, with a suc(cid:173)
`cess criterion of 2-90%.
`
`TABLE 2. DOSE ACCURACY FOR INSULIN GLARGINE
`DELIVERED USING SowSTAR
`
`Target dose
`
`10 units (ISO limit, ±1 unit: 9, 11 units)
`Observed dose
`Lowest dose delivered
`Highest dose delivered
`ISO calculated minimuma
`ISO calculated maximuma
`40 units (ISO limit, ±5%: 38, 42 units)
`Observed dose
`Lowest dose delivered
`Highest dose delivered
`ISO calculated minimuma
`ISO calculated maximuma
`80 units (ISO limit, ±5%: 76, 84 units)
`Observed dose
`Lowest dose delivered
`Highest dose delivered
`ISO calculated minimuma
`ISO calculated maximuma
`
`Values for observed dose are mean:+: SD.
`"Calculated using Eq. 2.
`
`Dose (units)
`
`9.87 ± 0.24
`9.15
`10.54
`9.23
`10.50
`
`39.54 ± 0.36
`38.50
`40.33
`38.58
`40.49
`
`79.02 ± 0.62
`76.83
`80.31
`77.35
`80.69
`
`Sanofi Exhibit 2130.003
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`SOLOSTAR DOSE ACCURACY BY PATIENTS
`
`325
`
`For statistical analysis the SYSTAT version 10.2 software
`(Systat Software, Inc.) was used.
`
`Results
`
`Accuracy of doses delivered
`
`For insulin glargine, the mean delivered dose in each test
`was marginally below the target dose (between -0.9 and
`-1.3%). For all three doses (10, 40, and 80 units), 100% of in(cid:173)
`dividual deliveries were within the established tolerance lim(cid:173)
`its according to the ISO standard (Fig. 1 and Table 2). When
`the data sets were analyzed, the 95% CI (Eq. 2) for each dose
`was within the tolerance limits, thus passing the ISO stan(cid:173)
`dards.
`The mean delivered dose in each test for insulin glulisine
`was marginally below the target dose (between -0.4 and
`-1.1 %). For all three doses (5, 15, and 30 units), 100% of in-
`
`dividual deliveries were within the established tolerance lim(cid:173)
`its according to the ISO standard (Fig. 2 and Table 3). When
`the data sets were analyzed, the 95% CI (Eq. 2) for each dose
`was within the tolerance limits, thus passing the ISO stan(cid:173)
`dards.
`The predefined success rate of ::>:90% of delivered doses
`within the ISO ranges was also achieved for both insulin
`glargine and insulin glulisine.
`
`Discussion
`
`The results of this study are consistent with those reported
`previously in a clinical setting, when used by patients with
`diabetes, and also in a laboratory setting, when delivered by
`trained technicians. As required for regulatory approval,
`and as previously demonstrated, 13 in a laboratory setting
`with doses delivered by technicians, the SoloSTAR device
`
`A
`
`I
`
`Upper limit (6 .0 units) according to ISO 11608-1
`
`I
`
`...
`~•·
`••••
`•• &
`.
`• •
`•
`•
`
`•
`
`•
`
`• • •
`•
`....
`y • • • •• •••
`•
`
`~
`
`~
`
`•
`•
`.... • •••••• y
`
`&
`
`•
`
`y
`
`"ii,
`
`-'i::
`
`2.
`'O
`e!
`Cl)
`. :?
`Cl)
`'O
`Cl)
`Ill
`0
`C
`
`6.5
`
`6.0
`
`5. 5
`
`5.0
`
`4.5
`
`4.0
`
`3.5
`
`•• • •
`•
`• •• . y • •
`
`50
`
`60
`
`I
`
`I
`
`Lower limit (4 .0 units) according to ISO 11608-1
`
`20
`
`30
`Dose number
`
`40
`
`Upper limit (16.0 units) according to ISO 11608-1
`
`I
`
`I
`
`0
`
`10
`
`B
`
`16.5
`
`16.0
`
`15.5
`
`15.0
`
`~
`
`y
`
`•
`
`.... y
`
`y
`
`"ii,
`:!::
`C:
`2.
`'O
`Cl) ... Cl)
`.:?
`ai
`'O
`Cl)
`Ill
`0
`C
`
`13.5
`0
`
`C
`
`• .
`•
`
`TY
`
`• •
`
`y
`
`14.5
`
`14.0
`
`. .. •
`•
`.
`.
`•
`•
`
`...
`
`....
`•
`
`•
`
`•
`• ....... •
`•
`•
`• ~•
`• •
`•••• • •
`••
`....
`....
`•
`
`•
`
`...
`
`...
`•
`•
`
`I
`
`Lower limit (14.0 units) according to ISO 11608-1
`
`10
`
`20
`
`30
`Dose number
`
`40
`
`50
`
`60
`
`I
`
`u, 32 .5
`:!::
`32.0
`C:
`2. 31 .5
`31 .0
`'O
`e!
`30.5
`Cl)
`30.0
`.:?
`29 .5
`ai
`29 .0
`'O
`28.5
`Cl)
`Ill
`28 .0
`0
`C
`27 .5
`0
`
`• ·~
`
`.
`• ....
`•
`• •
`
`I
`
`~
`
`Upper limit (31 .5 units) according to ISO 11608-1
`
`I
`
`-
`
`•
`.... ... ....
`
`• •
`.... •• & . . . . . . .
`··••~
`••
`• •
`....
`
`y
`
`• • • • . . . . . . . . &
`
`....
`
`•
`•
`•. • ....
`•
`
`•
`
`.....
`
`....
`
`....
`
`....
`
`•
`I
`
`Lower limit (28.5 units) according to ISO 11608-1
`
`•
`I
`
`10
`
`20
`
`30
`Dose number
`
`40
`
`50
`
`60
`
`FIG. 2. Accuracy of the SoloSTAR pen when delivering (A) 5-, (B) 15-, and (C) 30-unit doses of insulin glulisine. The up(cid:173)
`per and lower limits, as defined by the ISO standard, 11608-1, are indicated for each dose. Each data point represents the
`dose delivered by an individual person.
`
`Sanofi Exhibit 2130.004
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`326
`
`HERMANNS ET AL.
`
`TABLE 3. DOSE ACCURACY FOR INSULIN GLULISINE
`DELIVERED UsrNG SowSTAR
`
`Target dose
`
`5 units (ISO limit, ±1 unit: 4, 6 units)
`Observed dose
`Lowest dose delivered
`Highest dose delivered
`ISO calculated minimuma
`ISO calculated maximuma
`15 units (ISO limit, ±1 unit: 14, 16 units)
`Observed dose
`Lowest dose delivered
`Highest dose delivered
`ISO calculated minimuma
`ISO calculated maximuma
`30 units (ISO limit, ±5%: 28.5, 31.5 units)
`Observed dose
`Lowest dose delivered
`Highest dose delivered
`ISO calculated minimuma
`ISO calculated maximuma
`
`Values for observed dose are mean± SD.
`"Calculated using Eq. 2.
`
`Dose (units)
`
`4.98 ± 0.20
`4.50
`5.55
`4.44
`5.52
`
`14.87 ± 0.29
`14.07
`15.48
`14.09
`15.65
`
`29.67 ± 0.34
`28.93
`30.69
`28.76
`30.57
`
`achieved repeated doses and accuracy well within the ISO
`standards. In the clinical setting, people who had not previ(cid:173)
`ously used an insulin pen device were able to accurately de(cid:173)
`liver repeated doses of insulin glargine with the SoloSTAR
`pen. 15 The majority of participants in the present study were
`experienced in the use of insulin pens; therefore, this may
`be considered a true test of dosing accuracy rather than us(cid:173)
`ability of the SoloSTAR device.
`In the present study, we evaluated the inter-subject vari(cid:173)
`ability in dose accuracy. For future studies, it would also be
`of interest to evaluate the intra-subject variability to confirm
`whether people with diabetes can consistently deliver the re(cid:173)
`quired dose. In addition, future studies could also evaluate
`the dose accuracy of SoloSTAR compared with alternative
`pen devices or the vial and syringe, to investigate if other
`devices achieve the same standards when used by patients
`as shown here for SoloST AR.
`This study confirms that people with diabetes can achieve
`a level of accuracy with the SoloSTAR device similar to that
`of technicians in a controlled lab environment that is con(cid:173)
`sistently within ISO standards across the dosing range of the
`pen.
`
`Conclusions
`
`When considered together, the dosing accuracy and ease
`of use associated with the SoloSTAR device16,17 may en(cid:173)
`courage patients and healthcare professionals alike to strive
`for optimal glycemic control with greater confidence in the
`precision and accuracy of the device while titrating insulin
`glargine and/ or insulin glulisine to treatment goals.
`
`Acknowledgments
`
`This study was supported by sanofi-aventis. Editorial sup(cid:173)
`port was provided by the Global Publications Group of
`sanofi-aventis.
`
`Author Disclosure Statement
`
`N.H. is a member of the Global Diabetes educator advi(cid:173)
`sory board for Eli Lilly. B.K. is a member of an advisory
`board for Roche Diagnostics. T.H. is a member of the Global
`Advisory Board for sanofi-aventis.
`
`References
`
`1. Diabetes Control and Complications Trial Research Group:
`The effect of intensive treatment of diabetes on the devel(cid:173)
`opment and progression of long-term complications in in(cid:173)
`sulin-dependent diabetes mellitus. The Diabetes Control and
`Complications Trial Research Group. N Engl J Med 1993;
`329:977-986.
`2. Nathan DM, Cleary PA, Backlund JY, Genuth SM, Lachin
`JM, Orchard TJ, Raskin P, Zinman B: Intensive diabetes treat(cid:173)
`ment and cardiovascular disease in patients with type 1 di(cid:173)
`abetes. N Engl J Med 2005;353:2643-2653.
`3. Polansky WH, Fisher L, Guzman S, Villa-Caballero L, Edel(cid:173)
`man SV: Psychological insulin resistance in patients with
`type 2 diabetes: the scope of the problem. Diabetes Care
`2005;28:2543-2545.
`4. Kadiri A, Chraibi A, Marouan F, Ababou MR, el Guermai
`N, Wadjinny A, Kerfati A, Douiri M, Bensouda JD, Belkhadir
`J, Arvanitis Y: Comparison of NovoPen 3 and syringes/vials
`in the acceptance of insulin therapy in NIDDM patients with
`secondary failure to oral hypoglycaemic agents. Diabetes
`Res Clin Pract 1998;41:15-23.
`5. Stocki K, Ory C, Vanderplas A, Nicklasson L, Lyness W,
`Cobden D, Chang E: An evaluation of patient preference
`for an alternative insulin delivery system compared to
`standard vial and syringe. Curr Med Res Opin 2007;23:
`133-146.
`6. Fox C, McKinnon C, Wall A, Lawton SA: Ability to handle,
`and patient preference for, insulin delivery devices in visu(cid:173)
`ally impaired patients with type 2 diabetes. Pract Diabetes
`Int 2002;19:104-107.
`7. Gnanalingham MG, Newland P, Smith CP: Accuracy and re(cid:173)
`producibility of low dose insulin administration using pen(cid:173)
`injectors and syringes. Arch Dis Child 1998;79:59-62.
`8. Korytkowski M, Bell D, Jacobsen C, Suwannasari R: A mul(cid:173)
`ticenter, randomized, open-label, comparative, two-period
`crossover trial of preference, efficacy, and safety profiles of
`a prefilled, disposable pen and conventional vial/ syringe for
`insulin injection in patients with type 1 or 2 diabetes melli(cid:173)
`tus. Clin Ther 2003;25:2836-2848.
`9. Rubin RR, Peyrot M: Factors affecting use of insulin pens
`by patients with type 2 diabetes. Diabetes Care 2007;31:430-
`432.
`10. Graff MR, McClanahan MA: Assessment by patients with
`diabetes mellitus of two insulin pen delivery systems ver(cid:173)
`sus a vial and syringe. Clin Ther 1998;20:486-496.
`11. Lee WC, Balu S, Cobden D, Joshi AV, Pashos CL: Medica(cid:173)
`tion adherence and the associated health-economic impact
`among patients with type 2 diabetes mellitus converting to
`insulin pen therapy: an analysis of third-party managed care
`claims data. Clin Ther 2006;28:1712-1725.
`12. Austin F, Cox G: Lantus SoloStar and Apidra SoloStar pen
`colors contributes to the differentiation by users with nor(cid:173)
`mal vision and by users with impaired color vision. J Dia(cid:173)
`betes Sci Technol 2007;1(Suppl 1):A4.
`13. Clarke A, Spollett G: Dose accuracy and injection force dy(cid:173)
`namics of a novel disposable insulin pen. Expert Opin Drug
`Deliv 2007;4:165-174.
`
`Sanofi Exhibit 2130.005
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`SOLOSTAR DOSE ACCURACY BY PATIENTS
`
`327
`
`14. Pen-Injectors for Medical Use-Part 1: Pen-Injectors-Re(cid:173)
`quirements and Test Methods (ISO 11608-1:2000). Geneva:
`International Organization for Standardization, 2000.
`15. Schwartz SL, Vlajnic A: Validation of the SoloStar insulin
`pen [abstract]. J Diabetes Sci Technol 2007;1(Suppl 1):A159.
`16. Haak T, Edelman S, Walter C, Lecointre B, Spollett G: Com(cid:173)
`parison of usability and patient preference for the new dis(cid:173)
`posable insulin device Solostar versus Flexpen, lilly dispos(cid:173)
`able pen, and a prototype pen: an open-label study. Clin
`Ther 2007;29:650-660.
`17. Roberts A, Thornley S: Health care professional-reported us(cid:173)
`ability of SoloStar in a 3-month observational survey in
`
`everyday clinical practice. J Diabetes Sci Technol 2007;
`l(Suppl 1):A151.
`
`Address reprint requests to:
`Professor Norbert Hermanns, Ph.D.
`FIDAM GmbH
`Forschungsinstitut Diabetes-Akademie
`Theodor-Klotzbiicher-Strafle 12
`97980 Bad-Mergentheim, Germany
`
`E-mail: hermanns@diabetes-zentrum.de
`
`Sanofi Exhibit 2130.006
`Mylan v. Sanofi
`IPR2018-01676
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket