throbber
For reprint orders, please contact:
`reprints@future-drugs.com
`
`I
`EXPERT
`I REVIEWS
`
`Insulin glargine and glulisine
`SoloSTAR® pens for the
`treatment of diabetes
`
`Expert Rev. Med. Devices 5(2), 113-123 (2008)
`
`Samita Garg,
`William Charles Kelly
`and Satish Gargt
`t Author for correspondence
`Barbara Davis Center for
`Childhood Diabetes, University
`of Colorado at Denver and
`Health Sciences Center,
`1775 North Ursula, Room
`M20-1323, Aurora,
`CO 80045, USA
`Tel.: +1 303 724 6713 I 6770
`Fax: + 1 303 724 6784
`satish.garg@uchsc.edu
`
`Insulin is an effective medication for lowering hemoglobin A 1c values and can be used for both
`basal and prandial coverage of hyperglycemia in Type 1 and Type 2 diabetes. Despite its
`effectiveness there is still reluctance by patients and physicians to add insulin into the
`treatment regimen for Type 2 diabetes when needed. One of the key barriers to initiating
`insulin therapy is the method of delivery. Insulin delivery pens are continually developed as a
`means to improve upon the vial and syringe and to make it easier for patients to incorporate
`insulin therapy into their lifestyles. The SoloSTAR® pen (Sanofi-Aventis, Paris, France) was
`developed to make insulin delivery easier and to help eliminate barriers to the initiation of
`insulin therapy. In this article, we discuss the features and characteristics of SoloSTAR that
`overcome existing unmet needs.
`
`KEYWORDS: diabetes • hypoglycemia • injection force • insulin • insulin dose • SoloSTAR® pen
`
`The increasing prevalence of diabetes in most
`populations has had a major impact on healthcare
`systems worldwide [I]. Global projections for dia(cid:173)
`betes are increasing at an alarming rate, with the
`total number of people with diabetes projected to
`rise from 171 million in 2000 to 366 million in
`2030 [2]. In the USA for example, crude preva(cid:173)
`lence in 1999-2002 of total diabetes was 6.3%
`(19.3 million, 2002 US population), consisting
`of 3.5% diagnosed and 2.8% undiagnosed [3].
`Currently, the prevalence of diabetes in the USA
`is approximately 7.0% (21 million people with
`diagnosed or undiagnosed diabetes) [IOI]. This
`rise in prevalence of diabetes is closely associated
`with an increasing prevalence of obesity across
`the globe (FIGURE I) [2,4,5,102,103].
`In healthy individuals, pancreatic ~ cells
`respond to changes in blood glucose by secreting
`insulin and increasing insulin synthesis. Diabe(cid:173)
`tes is characterized by progressive ~-cell failure,
`resulting in a decline in insulin secretion and
`hyperglycemia [6]. The therapeutic approach to
`diabetes commonly involves intensive insulin
`management (basal/bolus) to maintain normal
`glycemic levels, by replacing insulin as close to
`the physiological insulin secretion profile of
`healthy individuals as possible [6]. For patients
`with Type 1 or Type 2 diabetes, maintaining
`glycemic levels as close to the nondiabetic range
`
`as possible has been demonstrated to reduce the
`risk of developing diabetes-specific complica(cid:173)
`tions, including retinopathy, nephropathy and
`neuropathy [7-9]. Insulin is the most effective
`diabetes medication in lowering glycemia and,
`when used in adequate doses, can decrease any
`level of elevated hemoglobin (Hb)A1c [IO].
`The recently published American Diabetes
`Association (ADA) and European Association
`for the Study of Diabetes (EASD) consensus
`statement guidelines recommend the early addi(cid:173)
`tion of insulin therapy in patients who do not
`meet target goals (Box I) [IO].
`Previous data from intermittent self-monitor(cid:173)
`ing of blood glucose (SMBG) in Type 2 diabe(cid:173)
`tes indicated higher contributions to elevated
`HbA1c from fasting blood glucoses at higher
`levels (>9%), and significantly higher contribu(cid:173)
`tions to rise in HbA1c from postprandial blood
`glucose (PPBG) at lower HbA1c levels (<8%).
`More recent data from continuous glucose
`monitoring (CGM) systems indicate an inabil(cid:173)
`ity for patients to achieve normal fasting blood
`glucose, even amongst those with near normal
`HbA1c values (i.e., <6%; FrGuRE 2) [II]. In addi(cid:173)
`tion, a significant contribution to rises in HbA1c
`levels also comes from post-dinner elevations in
`blood glucose at higher HbA1c values, which in
`turn might contribute to higher fasting blood
`
`www.future-drugs.com
`
`10.2217/17434440.5.2. l 13
`
`© 2008 Future Drugs Ltd
`
`ISSN 1743-4440
`
`113
`
`Sanofi Exhibit 2128.001
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`Garg, Kelly & Garg
`
`glucose levels. Glucose variability, in part due to postprandial
`hyperglycemia, has been demonstrated to correlate with oxida(cid:173)
`tive stress markers [12]. Recent data from CGM also indicates
`loss of postprandial glucose control preceding fasting hyper(cid:173)
`glycemia with increasing duration of diabetes (FIGURE 3)
`[13].
`Thus, early focus on postprandial hyperglycemia may need to
`be considered (across all HbA1c levels), especially when
`attempting to achieve normal fasting glucose.
`
`Overview of the market: the impact of glucose control
`Tight blood glucose control has been shown to prevent, or delay,
`the development of diabetes-related microvascular and macro(cid:173)
`vascular complications. An epidemiologic analysis of data from
`the United Kingdom Prospective Diabetes Study (UKPDS)
`patient population demonstrated that for each 1 % reduction in
`mean HbA1c, there was a 37% reduction in risk for microvascu(cid:173)
`lar complications alongside a 14% reduction in macrovascular
`complications [14]. Similar data from the Diabetes Control and
`Complication Trial (DCCT) in Type 1 diabetes shows the bene(cid:173)
`fits of intensive insulin therapy resulting in significant reductions
`in both microvascular and macrovascular complications. Despite
`all of the available data, only approximately 30% of patients
`with Type 2 diabetes are on some sort of insulin treatment in the
`USA [15,IOl]. While this is, in part, due to some patients being
`well controlled with lifestyle management and oral antidiabetic
`drugs (OADs), patient and physician reluctance to start insulin
`therapy is also thought to be a contributor [16,17]. In order to
`achieve and maintain tight blood glucose control, insulin use in
`patients diagnosed with diabetes should be an integral compo(cid:173)
`nent of their management strategy. Most patients with Type 1
`diabetes, and increasingly more patients with Type 2 diabetes,
`use two different types of insulins to provide basal and prandial
`coverage for hyperglycemia; however, many still use premixed
`formulations. Indeed, it has been estimated that premixed insu(cid:173)
`lins account for 22% of the total volume of insulin sold world(cid:173)
`wide [15]. Premixed insulin usually contains a rapid-acting insu(cid:173)
`lin and an intermediate-acting insulin with an aim to mimic
`endogenous insulin secretion patterns [18]. However, the use of
`premixed insulins is declining as the use of basal and prandial
`insulin increases [15], which is in line with the ADA/EASD con(cid:173)
`sensus guidelines that only recommend the use of premixed
`insulin after a patient is stabilized on insulin and if their mix
`ratio is close to one of the available premixed insulin ratios [IO].
`A basal insulin supply, such as insulin glargine (Lantus®;
`Sanofi-Aventis, Paris, France), which has a relatively constant
`and peakless delivery over 24 h [19], can provide the steady,
`low-level insulin that is constantly present in the circulation to
`cover preprandial and overnight fasting periods. This is sup(cid:173)
`plemented with multiple preprandial injections of regular
`human insulin or rapid-acting insulin analogs, such as insulin
`glulisine (Apidra®; Sanofi-Aventis, Paris, France), which aim
`to normalize and maintain good glycemic control, reduced
`glucose variability and better HbA1c values.
`
`Box 1. Summary of the American Diabetes
`Association/European Association for the Study of
`Diabetes consensus algorithm for the
`management of Type 2 diabetes.
`
`• Step 1. Lifestyle modification and meformin
`- Lifestyle modification and metformin at diagnosis
`- Titrate metformin to maximum effective dose over
`1-2 months
`- Check HbA1 c every 3 months until <7%, and every 6 months
`thereafter
`• Step 2. Intensify therapy
`- Add further medications within 2-3 months if HbA1c remains
`>7%:
`- Insulin
`- Sulfonylureas
`- Glitazones
`- Choice of agent depends on HbA 1c level
`- Insulin is recommended if HbA1c remains >8.5%
`
`HbA1i Hemoglobin A1c
`From [IO]
`
`Insulin administration
`An important aspect of diabetes care and glycemic control is
`the delivery of insulin. The method by which insulin is
`administered has been shown to impact patient acceptability
`of insulin therapy and quality of life, and may serve as a key
`barrier to insulin initiation [20]. Previously, the predominant
`route of insulin administration for patients with diabetes was
`the syringe and vial. However, this method of administration
`has many disadvantages, including fear of injections [21,22],
`poor dose accuracy [23], lack of social acceptance [24], inaccu(cid:173)
`racy when self-mixing insulins [25] and possibly changing
`pharmacokinetics of both long- and rapid-acting insulins.
`While still an injection device, insulin pens help to overcome
`many of these barriers.
`Since the introduction of the first insulin pen, NovoPen®
`(Novo Nordisk, AS Bagsvaerd, Denmark) in 1985, insulin pens
`have continued to improve in design and usability features and
`address many of the barriers associated with administering
`insulin using a syringe and vial. Precision and accurate dosing is
`crucial for patients with diabetes, particularly for those on com(cid:173)
`plex treatment regimens. Previous studies have indicated that
`up to 80% of people with diabetes incorrectly administer their
`insulin when using a syringe [26,27]. Santiago et al. conducted a
`precision, accuracy and durability study of an insulin pen
`(NovoPen) that tested the pen at three preset doses under stress
`conditions (multiple thermal and vibration stress tests), which
`were intended to replicate daily use by patients [28]. The accu(cid:173)
`racy of the insulin pen was within 1 % of the preset dose after
`the stress and endurance tests, and the precision of the pen
`devices were likewise high (delivery-dose relative error was at
`most 0.8% of the intended dose) after thermal stress, vibration
`stress, free-fall testing or 5-year endurance testing.
`
`114
`
`Expert Rev. Med Devices 5(2), (2008)
`
`Sanofi Exhibit 2128.002
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`SoloSTAR® pen
`
`Device Profile
`
`--
`
`-
`
`-
`
`-
`
`-
`
`IB
`
`Male
`Female
`
`-
`
`-
`
`-
`
`-
`
`-
`
`-
`
`-
`
`-
`
`w
`w
`0
`
`-a.
`<I: -,
`
`a:
`u..
`<I:
`
`I
`
`<I: w
`
`(/)
`

`
`-
`
`---
`
`-
`
`-
`
`- - - - - - -
`
`2025: 333 milliom
`Increase 72%
`

`
`Figure 1. (A) Global projections for the diabetes epidemic; 2003-2025 (millions) and (B) increasing prevalence of obesity.
`AFR: Africa; CAV: China and Vietnam; CEE: Central and Eastern Europe; JAPI: Japan, Australia and Pacific Islands; LAC: Latin
`America/Caribbean; ME: Middle East; NAC: North America/Cuba; SEA: South-East Asia; WE: Western Europe.
`Part (A): Reprinted with permission from Macmillan Publishers Ltd [lJ.
`Part (B): Information from [2,4,5,101,102].
`
`Korytkowski et al. demonstrated
`that 82% of patients
`(n = 105) indicated greater confidence with dose setting using
`FlexPen ® (Novo Nordisk) versus 11 % of patients who preferred
`the syringe and vial method [29]. In addition, it was also demon(cid:173)
`strated that 73% of patients reported more confidence in the
`accuracy of the dose delivered with FlexPen compared with 19%
`of patients using a syringe and vial. Insulin pens are now the pre(cid:173)
`dominant form of insulin administration in many countries,
`accounting for over 50% of insulin use worldwide, especially in
`Europe and Asia. In the USA uptake of insulin pens is steadily
`increasing, but it lags behind that seen in Europe and Asia [30].
`
`Unmet needs
`While insulin pen devices have made it easier for users to adminis(cid:173)
`ter insulin, there remains scope for further development of insulin
`pens in response to unmet patient needs in relation to the develop(cid:173)
`ment ofSoloSTAR® (Sanofi-Aventis). Type 2 diabetes is character(cid:173)
`ized by obesity and insulin resistance and, coupled with the pro(cid:173)
`gressive nature of the disease, increasing doses of insulin are
`required over time. Accordingly, many patients need to administer
`doses of insulin exceeding 60 units, the maximum dose of many
`insulin pens, thus necessitating multiple injections. Limited joint
`mobility of the hand, commonly referred to as cheiroarthropathy,
`is frequently observed in patients with diabetes, particularly elderly
`patients, and is characterized by low grip strength and/or limited
`dexterity [31,32], which can impede the efficient administration of
`insulin, in such patients, using pen devices.
`Problems with visual acuity are common in patients with dia(cid:173)
`betes and occur primarily as a result of diabetic retinopathy [33,34].
`People with diabetes, particularly those with Type 1 diabetes,
`often use more than one type of insulin to manage basal and
`
`prandial insulin requirements, which can be provided by insulin
`glargine and insulin glulisine, respectively. The doses and
`pharmacodynamics of prandial and basal insulins differ; accord(cid:173)
`ingly, it is important that the delivery devices are sufficiently dif(cid:173)
`ferentiated to ensure low risk of users confusing the two insulin
`formulations. Patients with visual problems also place a greater
`reliance on non-visual modes when selecting dose. Dose setting
`and injections can be aided by audible recognition (the click
`sound), which occurs when a dose is dialed [35].
`
`The SoloSTAR pen
`An overview of the SoloSTAR pen & how it works
`The continual evolution of insulin devices has led to the Solo(cid:173)
`STAR pen, which is a prefilled, disposable insulin pen device
`designed for use once or several times daily. It is available for
`the administration of basal insulin glargine and prandial insulin
`glulisine for patients with either Type 1 or Type 2 diabetes,
`with two colors to differentiate the two pen devices. The insu(cid:173)
`lin glargine SoloSTAR pen is approved for use in both the EU
`and the USA, whilst the insulin glulisine pen is approved for
`use in the EU.
`The SoloSTAR pen is very easy to use. The user checks that they
`have the correct insulin pen. The user then attaches a new pen
`needle and performs a safety shot of 2 units to verify that the nee(cid:173)
`dle is working. The user then dials their dose and delivers the dose
`subcutaneously by pressing down on the injection button. The
`user will then remove the pen afrer counting to ten at the end of
`the injection to ensure the full dose is delivered. The needle is then
`taken off the pen and discarded safely. The pen cap is replaced
`and the pen can be stored until the next use. New, unopened
`
`www.future-drugs.com
`
`115
`
`Sanofi Exhibit 2128.003
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`Garg, Kelly & Garg
`
`® -
`
`200
`
`150
`
`Modal day by baseline HbA10 DISPLAY (periods 2 and 3)
`

`
`350
`
`300
`
`250
`
`200
`
`150
`
`Modal day by baseline HbA10 BLINDED (period 1)
`32 35 0 - t - - - - - - - - - - - - - - - - - - - - - - -=
`Cl
`,§,
`g> 300
`'6
`m 250
`0 rn
`C:
`GJ rn
`GJ rn
`0
`tJ
`::::J c,
`C: ca
`~ 100 +--------------------_____,, 100 + -~~~ -~ -~~ -~ -~~~~ -~ -~___ __ , ,
`
`12 am
`
`6 am
`
`Noon
`TI~~~
`
`6 pm
`
`12 pm
`
`12 am
`
`6 am
`
`Noon
`TI~~~
`
`6 pm
`
`12 pm
`
`Baseline HbA10
`.$ 6.0% (n = 5)
`-
`8.1-9.0% (n = 1) -
`-
`6.1-7.0% (n = 1) -
`-
`
`9.1-10.0% (n = 7)
`7.1-8.0% ( = 33)
`>10.0% (n = 5)
`
`Figure 2. Modal day by baseline HbA1c while subjects were blinded (A) to continuous glucose data or while subjects were
`given real-time access to continuous glucose values (B), trend graphs and high/low alerts.
`Significant postpradial elevations in blood glucose levels were observed across all HbA1c levels in both periods.
`Copyright© 2006 American Diabetes Association. Reproduced from [Ill with permission from The American Diabetes Association.
`
`SoloSTAR pens should be kept in the refrigerator (2-8°C). Once
`opened, the SoloSTAR pen should be kept at room temperature
`for up to 28 days in accordance with storage condition recom(cid:173)
`mendations, which differ in the EU (recommendations being
`below 25°C) and US (being below 30°C).
`
`Evaluation of the SoloSTAR pen & responding to
`unmet needs
`The SoloSTAR pen builds upon the strengths of current
`devices while including additional features, which have been
`ergonomically tested in order to establish their usability and
`effectiveness. Testing included collection of anthropometric
`data in intended user populations in order to recommend
`the most suitable user dimensions of the pen and develop
`the strength and robustness of the SoloSTAR pen, which
`compliments its user population.
`
`Sensitivity & specificity
`During development of the SoloSTAR pen, human factors
`were also considered, which led to a short-dial extension
`design facilitating easier grip during injection and enabling
`the user to administer even the maximum insulin dose with
`ease. This is an essential feature of the SoloSTAR pen when
`taking into account cheiroarthropathy, which is a significant
`problem in some patients with diabetes [31 ,32], with estimates
`that up to 58% of patients with diabetes have limited joint
`mobility of the hand [36] and significantly lower-grip strength
`compared with healthy controls [37]. The Solo STAR pen has
`
`a maximum dose administration of 80 units, which exceeds
`the maximum dose of other available pens, except OptiClik
`(also 80 units), but including FlexPen and Lilly pen, which
`both administer a maximum of 60 units (TABLE I). With higher
`doses of insulin required in obese patients especially with
`Type 2 diabetes, this is an important feature of the SoloSTAR
`pen, as it enables higher dose users to minimize the number
`of injections required.
`Also, under varying temperature conditions, the SoloSTAR pen
`successfully passed dose accuracy testing, ensuring consistent and
`reliable insulin dose accuracy in a laboratory setting (TABLE 2) [38].
`Moreover, the measured dose accuracy of SoloSTAR, when
`used by patients in a clinical setting, was within the limits of
`the International Organisation for Standardisation dose accu(cid:173)
`racy standard [39]. This reassures patients that, when used cor(cid:173)
`rectly, the SoloSTAR pen will deliver the dialed dose, which
`facilitates the titration of insulin dose without an increased risk
`of hypoglycemia or hyperglycemia. Injection force testing was
`performed to measure the force and force characteristics
`required to dispense a known volume of insulin (40 units)
`within a 4-s time period (TABLE 3) [38]. Findings from these tests
`showed that the SoloSTAR pen's highly efficient drive mecha(cid:173)
`nism translates into a lower injection force than that of the
`Flexpen and Lilly pen [38].
`To facilitate the differentiation of the insulin glargine pen
`from the insulin glulisine pen and to avoid the mistaken
`administration of basal instead of prandial insulin, or vice
`versa, the SoloSTAR pen is manufactured in two different
`
`116
`
`Expert Rev. Med Devices 5(2), (2008)
`
`Sanofi Exhibit 2128.004
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`Table 1. Features of the SoloSTAR®, FlexPen® and Lilly pen.
`
`SoloSTAR® pen
`
`SoloSTAR®
`(Sanofi-Aventis)
`
`FlexPen®
`(Novo Nordisk)
`
`80
`
`60
`
`Lilly pen
`(Eli Lilly and Company)
`
`60
`
`* Availability may vary.
`NPH: Neutral protamine Hagedorn.
`
`- Insulin glargine (LANTUS®)
`- Insulin glulisine (Apidra®)
`
`- Insulin detemir (Levemir®)
`- Insulin aspart (NovoLoi®)
`- NPH insulin (lnsulatard )
`- NovoMix® 30 (30% insulin aspart and 70% protaminated insulin aspart)
`
`- Insulin lispro (Humalo~®)
`- NPH insulin (Humulin )
`- Humalog Mix75/25™ (75% insulin lispro) protamine suspension,
`25% insulin lispro)
`- Humalog MixS0/50™ (50% insulin lispro protamine suspension,
`50% insulin lispro)
`
`colors: grey for insulin glargine and blue for insulin glulisine.
`This makes the SoloSTAR pen the first disposable insulin pen
`device to differentiate in pen body color.
`This feature was included after consultations and assistance
`from healthcare providers, who suggested it to further minimize
`any confusion between the two insulins, especially in patients
`with visual impairments (FIGURE 4). In addition, another difference
`between insulin glulisine and glargine SoloSTAR pens includes a
`tactile differentiation of a raised ring on the dose button of the
`insulin glulisine pen, and other differentiation features include
`different colors in the labels and packaging.
`
`Cost-effectiveness
`Results from a recent study demonstrated that, in patients with
`Type 2 diabetes treated in a managed care setting, conversion
`from insulin injection with a syringe and vial to administration
`with an insulin analog pen device was associated with significantly
`lower annual treatment costs (US$16,359 vs 14,769, respectively;
`p < 0.01) as a result of improved medication adherence, fewer
`hypoglycemic events and reduced emergency department and
`physician visits [40]. These reductions could be as a result of the
`increased accuracy in dosing and timing of injection when using
`the insulin pen, which leads to a lower risk of hypoglycemia.
`Medication adherence was significantly improved after conversion
`to the insulin pen device (from 62-69%; p < 0.01).
`A further potential cost saving for direct treatment could be
`made using pens when considering that insulin in vials is dis(cid:173)
`carded by physicians after 28 days as per US FDA guidelines.
`Accordingly, insulin pens could be more cost effective for chil(cid:173)
`dren and those taking small amounts of insulin. Since each insu(cid:173)
`lin pen only contains 300 units, there will be less wastage, main(cid:173)
`tainance of biological activity of insulin and greater likelihood
`to follow the FDA label in clinical practice.
`Recent results on the usability of the SoloSTAR pen reported
`by 65 healthcare professionals in clinical practice consider the
`SoloSTAR pen to be both easy to teach and easy to use for people
`
`with diabetes [41]. Of 65 healthcare professionals interviewed,
`most (n = 52; 80%) were able to spend less than 10 min train(cid:173)
`ing their patients to use SoloSTAR. This ease of use for both
`patients and healthcare professionals can translate into signifi(cid:173)
`cant cost savings in relation to the time and resources spent
`training users of the SoloSTAR pen.
`
`Use of the SoloSTAR pen in in-patient/hospital settings
`In an in-hospital setting, the accuracy of the dose delivered is a
`key factor when selecting an insulin delivery system. This is
`because lack of accuracy may increase the risk of hypo- or
`hyperglycemia, jeopardizing patient welfare and in turn increas(cid:173)
`ing diabetes-related treatment costs [38]. Used correctly, Solo(cid:173)
`STAR pens will accurately administer the dialed dose of insu(cid:173)
`lin, allowing reliable dose adjustment and minimizing the risk
`of resulting hypo- or hyperglycemia. In addition, color differ(cid:173)
`entiation of the insulin glargine and glulisine SoloSTAR pens
`reduces the potential for hospital staff to confuse the two
`devices [38]. Furthermore, due to the ease of use, the introduc(cid:173)
`tion of insulin therapy in the hospital setting is easier with
`SoloSTAR pens than with traditional syringes and vials and this
`may result in patients being more likely to continue insulin
`therapy when discharged from hospital [38].
`
`Clinical profile & post-marketing findings
`Haak et al. recently conducted a preference study across four
`countries
`(US, Germany, France and Japan)
`involving
`510 patients with Type 1 and Type 2 diabetes investigating the
`usability of the SoloSTAR pen, FlexPen and Lilly pen [42].
`Patients were assessed on their ability to correctly perform a
`number of tasks involved in using each pen (including getting
`started and removing the cap, attaching a needle, setting and
`delivering a safety dose and dialing and delivering a 40-unit
`dose) and their preference of pens. The assessed steps for the
`SoloSTAR pen and FlexPen devices were correctly completed
`by a similar proportion of patients: 94% for the SoloSTAR pen
`
`www.future-drugs.com
`
`117
`
`Sanofi Exhibit 2128.005
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`Garg, Kelly & Garg
`
`Table 2. Dose accuracy of the SoloSTAR® pen
`device (insulin glargine) at dialed doses of 1, 40
`and 80 units at temperatures of 5°C, ambient
`temperature (18-28°() and at 40°C.
`
`Recommended
`range according to
`ISO standards
`
`Cool temperature
`(5°C)
`
`Ambient
`temperature
`(18-28°C)
`
`Hot temperature
`(40°C)
`
`0-2
`
`38-42
`
`76-84
`
`1.22 ± 0.18
`
`39.85 ± 0.28 79.75 ± 0.29
`
`1.09 ± 0.15
`
`39.92 ± 0.34 79.82 ± 0.28
`
`1.15 ± 0.11
`
`39.87 ± 0.11 79.73±0.14
`
`Results are means± standard deviation; 30 pens were used for each dose, with
`two replicates per pen.
`ISO: International Organisation for Standardization.
`
`and 90% for FlexPen; however, fewer patients correctly com(cid:173)
`pleted the same steps with the Lilly pen (61 %), When patients
`were asked to rate their preference for each pen based on vari(cid:173)
`ous usability features, the feature 'easy/intuitive to figure out'
`was rated as 'best' most frequently for the SoloSTAR pen (55%
`of the time) followed by FlexPen (32% of the time) and least
`frequently for the Lilly pen (13% of the time),
`Similar findings were also observed in the usability subgroup
`analyses based on age, previous pen experience and visual/dex(cid:173)
`terity disabilities, A high proportion of patients aged 60 years or
`over correctly completed the assessed steps with the SoloSTAR
`pen (90%) and FlexPen (83%) compared with the Lilly pen
`(47%), A high proportion of patients with dexterity (91%) and
`visual (94%) impairments correctly completed all steps analyzed
`with the SoloSTAR pen, which was similar to that observed with
`FlexPen (84% of patients with dexterity and 89% of patients with
`manual impairment), Only half of all patients with either dexter(cid:173)
`ity (52%) or visual (52%) impairments correctly completed all
`analyzed steps with the Lilly pen,
`The ease of use of SoloSTAR has also been demonstrated in
`a single-center, open-label, single-arm sequential study, which
`investigated the usability of the SoloSTAR pen by patients
`with Type 1 or Type 2 diabetes, aged 21-78 years [39], After
`either face-to-face training (Part 1) or self-training (Part 2),
`subjects performed three dose-delivery repetitions into an
`injection pad using separate pens; pens were weighed before
`and after each dose delivery, The primary end point was the
`proportion of subjects delivering successful doses with all
`three repetitions, In Part 1, all 50 subjects delivered successful
`doses (100% success rate; 95% lower confidence bound
`[LCB]: 94,2%) and was within the limits of the ISO dose
`accuracy standard, In Part 2, 53 out of 54 validation subjects
`delivered successful doses (98%; 95% LCB: 91.5%), This
`study validates the SoloSTAR pen device for use by people
`
`with diabetes, with or without face-to-face trammg and
`showed that the SoloSTAR pen accurately delivered the dose
`that was dialed,
`In order to determine the usability and safety of the Solo(cid:173)
`STAR pen in clinical use, a 3-month observational study of
`2029 participants (1067 with Type 1 diabetes and 926 with
`Type 2 diabetes) was undertaken, with the primary end point
`defined as absence of serious adverse events directly related to a
`validated technical failure of the pen, Eight product technical
`complaints (PTC) were investigated and most were due to han(cid:173)
`dling errors, In total, 62 participants reported 77 adverse
`events, none of which were related to a PTC Overall, most
`(95.4%) patients reported that they were either 'very satisfied'
`or 'satisfied' with the SoloSTAR pen, and 96,8% of patients
`continued to use SoloSTAR at the end of the study [43],
`
`How does the SoloSTAR pen fit in the current treatment
`of diabetes?
`With evidence from the DCCT [7] and UKPDS [8] studies indi(cid:173)
`cating that improving glycemic control in patients with either
`Type 1 or Type 2 diabetes delays the onset of microvascular and
`macrovascular complications, intensive insulin treatment as a
`means of achieving and maintaining glycemic control has
`increased, The availability of the SoloSTAR pen for basal (insu(cid:173)
`lin glargine) and bolus (insulin glulisine) insulin provides both
`clinicians and patients alike with a simple approach for manag(cid:173)
`ing diabetes, With the gradual progression of Type 2 diabetes,
`the ADA recommend lifestyle approaches, including nutrition
`and exercise, as important components of diabetes manage(cid:173)
`ment followed by the introduction of one or more OADs, par(cid:173)
`ticularly metformin or sulfonylurea [44], In addition to the
`aggressive approach for achieving target HbA1c levels of less
`than 7,0%, step two of the ADA recommended treatment
`pathway (Box 1) now includes basal insulin, such as insulin
`glargine, as an option, Also, if HbA1c levels are above 8,5%
`with a first-line approach, it is wise to move to insulin therapy
`as additional oral medications are not likely to achieve target
`HbA1c values,
`This concept is supported by a wealth of clinical data,
`including trials such as the 36-week Lantus and Metformin
`(LANMET) study, comparing with NPH and metformin in
`Type 2 diabetes, conducted by Yki-Jarvinen et al, [45], which
`demonstrated that the early introduction of insulin glargine
`(with adequate titration of dose) to one OAD improved glyc(cid:173)
`emic control with a low risk of hypoglycemia (during the last
`12 weeks of this study, fasting plasma glucose [FPG] averaged
`5,75 ± 0,02 mmol/1 and mean HbA1c was 7,14 ± 0,12%),
`The next step in the treatment pathway should be the intro(cid:173)
`duction of one prandial dose of insulin, such as insulin gluli(cid:173)
`sine, A recent study on the basal plus approach demonstrated
`that the addition of insulin glulisine at breakfast or the main
`meal, for patients with Type 2 diabetes allows more patients to
`reach target HbA1c, while offering patients a flexible injection
`time (insulin glargine can be administered immediately after a
`
`118
`
`Expert Rev. Med Devices 5(2), (2008)
`
`Sanofi Exhibit 2128.006
`Mylan v. Sanofi
`IPR2018-01676
`
`

`

`SoloSTAR® pen
`
`Device Profile
`
`Fasting (nocturnal period)
`
`Post pradial (daytime period)
`
`11.5
`
`10.0
`
`8.4
`4.4
`
`0.7
`
`15
`
`14
`
`13
`
`12
`
`11
`
`10
`
`9
`
`8
`
`7
`
`6
`
`5
`
`-s
`
`0
`E
`.§.
`C:
`0
`
`~
`"E
`Q) u
`C:
`0 u
`Q)
`UI
`0 u
`:::J
`a
`
`0
`
`2
`
`4
`
`6
`
`8
`
`14
`12
`10
`Time of the day (h)
`
`16
`
`18
`
`20
`
`22
`
`24
`
`Figure 3. The 24-h recordings from continuous glucose monitoring in people with Type 1 diabetes.
`Curve 1 (blue): HbA1c < 6.5%; curve 2 (red): 6.5 to <7%; curve 3 (green): 7 to <8%; curve 4 (orange) 8 to <9%; curve 5 (purple):~ 9%.
`With increasing duration of Type 2 diabetes, post-prandial glucose elevations precede rise in fasting blood glucose levels.
`Copyright© 2006 American Diabetes Association. Reproduced from [13] with permission from The American Diabetes Association.
`
`meal, if necessary) [46]. This then offers a platform on which
`further prandial doses can be added, if patients are not reach(cid:173)
`ing target [47]. Patients using the SoloSTAR pen would benefit
`from not having to undertake additional learning to use a dif(cid:173)
`ferent type of pen when adding prandial insulin to their exist(cid:173)
`ing basal insulin regimen, and the different colors and tactile
`features mean that patients are at a low risk of confusing their
`insulin pens.
`The low injection force with SoloSTAR compared with Flex(cid:173)
`Pen and the Lilly pen [38] may offer advantages, particularly for
`people with limited manual dexterity, which may be age-related
`or as a result of other complications [31,32]. In the study by Haak
`et al. [42], 81 participants had dexterity impairments; of these
`91 % of participants completed a dose delivery with SoloSTAR
`versus 84% with FlexPen and 52% with the Lilly pen.
`Two devices, the Autopen® (Owen Mumford, Ltd, Oxford,
`UK) and lnnolet® (Novo Nordisk), both have low injection
`forces, thus making them potentially suitable for people with
`limited dexterity. Autopen is differentiated by body color and
`is a reusable cartridge-based pen, while the lnnolet has a large
`dose dial, with color differentiation on the injection button.
`However, t

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket