`Development of the SoloSTAR® insulin pen device:
`design verification and validation
`
`Andreas Bode
`
`To cite this article: Andreas Bode (2009) Development of the SoloST AR® insulin pen device:
`design verification and validation, Expert Opinion on Drug Delivery, 6:1, 103-112, DOI:
`10.1517/17 425240802636187
`
`To link to this article: https://doi.org/10.1517 /17425240802636187
`
`ti Published on line: 10 Dec 2008.
`
`~~ Submit your article to this journal C3'
`
`I 1111 Article views: 189
`
`~ Citing articles: 5 View citing articles C3'
`
`Full Terms & Conditions of access and use can be found at
`https://www.tandfonline.com/action/journal1nformation?journa1Code=iedd20
`
`Sanofi Exhibit 2116.001
`Mylan v. Sanofi
`IPR2018-01675
`
`

`

`ExP.ert
`Opinion
`
`1.
`
`Introduction
`
`2. Case studies
`
`3. Discussion
`
`4. Conclusions
`
`5.
`
`Expert opinion
`
`informa
`
`healthcare
`
`Technology Evaluation
`
`Development of the SoloST AR®
`insulin pen device: design
`verification and validation
`Andreas Bode
`Device Design & Development, sanoji-aventis Deutsch/and GmbH, Frankfort, Germany
`
`Background: SoloSTAR® (SOL; sanofi-aventis, Deutschland, GmbH) is a new,
`disposable insulin injection pen device for use by people with type 1 or
`type 2 diabetes to administer long- or short-acting insulin. Objectives: To
`discuss factors that have underlined the design process of the SOL device. In
`addition, to highlight the studies that shaped the direction of its develop(cid:173)
`ment, such as addressing the unmet needs of people with diabetes, which
`included a need for better differentiation features and a lower injection
`force compared with existing prefilled disposable pen devices. Results: The
`development of the SOL pen device was an iterative process involving both
`patients and the design team, which has lead to a manufacturable, tailor-made
`pen device. Patients' needs have been taken into account in the pen design;
`there are numerous differentiators on the device, which avoids confusion
`between insulin types. Furthermore, the SOL device has a lower injection
`force compared with other marketed pen devices. Finally, studies have
`shown that the SOL device is more accurate, easier to use and is preferred
`by patients over other pens on the market. Conclusions: The SOL device has
`undergone rigorous user and laboratory testing, which has captured evolving
`improvements to better meet the needs of people with diabetes.
`
`Keywords: design, development, device validation, device verification, diabetes mellitus,
`engineering, insulin pen device, long-acting insulin, short-acting insulin
`
`Expert Opin. Drug Deliv. (2009) 6(1):103-112
`
`1. Introduction
`
`SoloSTAR® (SOL; sanofi-aventis, Deutschland, GmbH) was developed for use
`by people with type 1 or type 2 diabetes for the administration of long- or
`short-acting insulin. Both SOL devices were approved in Europe in 2006 and for
`the long-acting insulin pen, approval was given in the United States in 2007.
`SOL is a disposable insulin pen device with a 3 ml capacity (300 units of insulin)
`designed for use once or several times a day (Figure 1). Studies have found that these
`products are easy to use [I-3], easy to teach [41, they dose accurately and have a
`lower injection force than both the FlexPen® (FP; Novo Nordisk, Bagsvaerd,
`Denmark) and the Lilly disposable pen device (LP; Humalog®/Humulin® pen;
`Eli Lilly and Company, Indianapolis, United States) [5,6].
`The SOL pen device was effectively developed from the ground up and the
`development process took into consideration not only laboratory testing (design
`verification) and user testing (design validation), but also extensive human and
`ergonomic factors, which will be discussed here using two case studies to illustrate
`the stages in the process. Design verification and validation were fed into an iterative
`design process at every stage of development from initial concept design through
`proof of principle and proof of concept. As a result, the SOL pen device is an
`intuitive, easy to use device [Il with a similar user interface (i.e., common mode
`of operation) as other pen devices, but also fulfils patients' needs to a degree that
`
`10.1517/17425240802636187 © 2009 lnforma UK Ltd ISSN 1742-5247
`All rights reserved: reproduction in whole or in part not permitted
`
`103
`
`Sanofi Exhibit 2116.002
`Mylan v. Sanofi
`IPR2018-01675
`
`

`

`Design verification and validation of SoloSTAR (SOL)
`
`Pen cap
`
`Pen needle (not included)
`
`Pen body
`
`(
`
`Outer
`needle
`cap
`
`Inner
`needle
`cap
`
`Needle
`
`Dosage I I Injection
`
`selector
`
`button
`
`Figure 1. Schematic diagram of the Lantus® SOL (insulin glargine) and Apidra® SOL (insulin glulisine) products.
`
`advances SOL close to the ideal mechanical disposable device.
`Indeed, some of the features such as addressing the unmet
`needs of the patient, which included better differentiation
`features and a lower injection force compared with existing
`devices that will be discussed, were identified during the
`development programme in user testing and subsequently
`incorporated into the end product.
`
`1.1 What are design verification and validation?
`The core of design verification and validation
`is best
`described by two simple questions: 'Did I design the product
`right?' must be answered positively to pass design verification;
`'Did I design the right product?' needs to be evaluated in
`design validation. Design verification is a laboratory-based
`exercise and involves the assessment of device function,
`including individual components, from a technical perspective.
`International standards have to be fulfilled and compliance
`proven. In addition to parameters set by the International
`Organization for Standardization (ISO), other parameters
`are defined as part of the design brief and, therefore, are
`implicit in the verification process. One such factor is the
`injection force of the device. Testing operating forces and
`torques has become state of the art and has been included to
`verify the end product against the design brief The aim of
`design verification is to quantitatively ensure the individual
`components and the device fulfils the technical requirements.
`For design validation it has to be demonstrated that the
`user can operate the device and that it answers their needs,
`and that the collection of objectives (i.e., design brief) is
`achieved. There are many ways to understand the degree
`of overlap of user requirements and device functionality.
`Ergonomics and human factors specialists can be consulted,
`as well as medical advisors who oversee large numbers of
`patients. However, to best validate a product, it has to be
`bought to the user. User surveys and studies in clinical
`settings, which were performed continuously during the
`development, are appropriate means to get direct feedback
`and understand the degree of overlap, and demonstrate that
`
`the end product fulfils the needs of the users as reflected in
`the design brief
`
`1.2 Why was a new insulin pen device needed?
`The use of the vial and syringe is still relatively common in
`some regions, particularly in the US. It had been estimated
`that in the US, only 14% of patients using insulin were
`using insulin pen devices (prefilled pens or cartridges) as a
`percentage of total insulin use, whereas in Europe, 92% of
`patients were using insulin pen devices [7]. However, it has
`been demonstrated that switching from the vial and syringe
`to insulin pens results in increased medication adherence
`and reduced treatment costs [SJ. There are several prefilled
`and reusable
`insulin pen devices now on the market,
`although availability may vary in some regions/ countries.
`Each device offers the patient specific advantages compared
`with the other pen devices. However, despite the multitude
`of pens available, there remains scope for further development
`of insulin pen devices in response to unmet patient needs.
`Some of these unmet needs will be discussed here, in relation
`to the development of the SOL pen device.
`
`1.3 Original development requirements/design brief
`The original design specification of the SOL pen device was
`based on the feedback from users with respect to existing
`devices plus research to understand the basic needs of
`customers (2001/2002). Here, users are considered to be
`patients as they inject insulin using the pen, as well as
`doctors, nurses and pharmacists as they prescribe, train and/or
`advise the patients on the pen. In addition, human factors
`analysis by means of a literature search provided basic
`requirements. The intent was to provide a pen device with
`better characteristics than the FP and the LP, as those were
`the most commonly used prefilled pen devices on the market
`at the time. Factors such as maximum length, diameter
`and injection force provided an integral part of the initial
`specification. Refinement of the requirements was done on
`the basis of human factor and ergonomic analyses.
`
`104
`
`Expert Opin. Drug De/iv. (2009) 6(1)
`
`Sanofi Exhibit 2116.003
`Mylan v. Sanofi
`IPR2018-01675
`
`

`

`Bode
`
`1.4 Guidelines and standards for insulin pen devices
`Insulin pen devices are subject to several regulatory guidelines
`developed by the national/international medical regulatory
`bodies, for example, FDA and the European Agency for the
`Evaluation of Medicinal Products (EMEA). Before approval
`from the FDA and EMEA can be sought, pen devices and
`related materials must also meet several criteria specified by
`ISO, in particular, ISO 11608-1 for insulin pen devices [9].
`The guidelines for insulin pen devices cover not only
`specific aspects of device use, such as dose accuracy or visibility
`of the selected dose, but also that the pen device doses
`correctly afrer storage in a range of environmental conditions
`(e.g., temperature and humidity), functions properly afrer
`being dropped from a height of 1 m at various orientations
`and that labels or other distinguishing marks are durable
`during use. However, factors such as injection force and
`design features such as colour and size are not covered by
`the ISO standards. Thus, for the design verification and
`validation of the SOL pen device, tests were performed to
`ensure it met the ISO guidelines and that it met the more
`stringent targets that were set internally. Furthermore, user
`testing was carried out to ensure that the SOL pen device
`was intuitive to use by the intended population.
`
`1.4.1 Unmet needs
`1.4.1.1 Insulin dose
`Increasing doses of insulin are required over time to overcome
`the insulin resistance and relative insulin deficiency. Indeed,
`many patients need to administer individual doses of insulin
`> 60 units, the maximum dose of many insulin pen devices,
`thus necessitating several injections. As a result, the design
`brief of the SOL pen device included the recommendation
`of a maximum dose of 80 units.
`
`1.4.1.2 Hand function and injection force characteristics
`Limited joint mobility of the hand, commonly referred to as
`cheiroarthropathy, is frequently observed in patients with
`diabetes, particularly elderly patients, which may occur as a
`result of connective tissue disorders or diabetic neuropathy,
`and is characterised by low grip strength and/or limited
`dexterity [10-15]. As a result, the recommendation in the
`design brief was for the SOL pen device to have a lower
`injection force than other prefilled devices available at the
`time, as well as a short dial extension length to reduce the
`mechanical strain on
`the user's
`thumb.
`Indeed, one
`would anticipate that a short dial extension with low force
`requirements would be easier to use for most of the patients.
`
`1.s Overview of the SOL pen design verification and
`validation process
`Numerous concepts were initially investigated; all could fulfil
`the design brief, but used different mechanical principles.
`Complex designs, such as an odometer mechanism for
`displaying the dose, were investigated, as well as toothed rod
`type mechanisms and simple tampo-printed dose scales. Of
`
`the proposals mentioned earlier, very few were selected afrer
`mechanism concept evaluation, which used quality function
`deployment techniques (QFD). These techniques ensured
`end user requirements were the main focus of selection. Extra
`parameters, such as complexity, technical risk or perceived
`patent infringement risk, were taken into consideration.
`The selected concepts were then progressed through the
`design verification and validation processes. As described
`earlier, the verification and validation of the SOL pen device
`followed an iterative process: at each stage of the process,
`studies were done to assess
`technical aspects (i.e.,
`the
`mechanical/physical properties) and user aspects (i.e., feedback
`from the intended user population). Additionally, structured
`risk assessment was done at each stage, with the results used
`for risk management of the device. Failure mode and effects
`analysis (FMEA) and user task analyses served as tools for
`risk assessment and provided an approach from two different
`directions. The FMEA helped us
`to understand which
`component failure or feature malfunction could lead to critical
`loss in performance and the user task analyses highlighted
`potential ambiguity leading to reasonably foreseeable misuse.
`Studies done at each stage of the process assessed not only
`the pen device itself (block models, proof of principle rigs,
`proof of concept prototypes and eventually the industrialised
`pen), but also individual components and features (including
`dial display, pen colours, label size and format, dose knob,
`pen cap and clip, overall dimensions). Results of these studies
`were fed back into the iterative design process, as summarised
`in Figure 2, to ensure the ongoing developments in pen
`device design and function continued to meet not only the
`original design specification, but also subsequent suggestions
`and recommendations leading to an updated design brief to
`further meet the patients' needs.
`As shown in this figure, there are two key areas that govern
`the design and development process. On the one hand, it
`is important to understand the patient's needs through a
`combination of literature research, ergonomics studies and
`user testing. On the other, it is important to respond to
`these needs with rational design and adaptation to ensure a
`solution is found before the next stage of development can be
`entered. User testing and laboratory-based testing performed
`at each stage of the development cycle helps ensure that the
`design is verified and validated.
`
`1.6 Objectives
`The design validation process involved nine user studies,
`which were done with a total of> 2,300 participants, including
`health-care professionals (nurses and physicians) and people
`with diabetes. Moreover, 12 ergonomics analyses were done in
`addition to numerous meetings with health-care professionals
`(nurses and physicians), which are part of an advisory board
`that allows medical experts to provide feedback on most
`aspects of product development. Technical tests of all
`products and components were performed in advance of each
`user/ergonomics study to ensure the product and component
`
`Expert Opin. Drug De/iv. (2009) 6(1)
`
`105
`
`Sanofi Exhibit 2116.004
`Mylan v. Sanofi
`IPR2018-01675
`
`

`

`Design verification and validation of SoloSTAR (SOL)
`
`Initial
`des;go bdef
`
`\
`
`~ Specify
`
`design criteria
`
`Proceed to next step
`in development
`process
`
`patient needs
`
`Realise
`patient needs
`
`\
`0
`
`~
`
`Final product
`for manufacturing
`
`J
`
`Continuous
`improvement
`programme
`
`Figure 2. A simplified diagram illustrating the feedback loops for the input of technical testing and user testing results on
`the design and development of the SOL pen device.
`
`met the design brief and ISO guidelines. It is not the aim of
`this paper to discuss all details of the extensive test programme
`(in excess of several 10,000 pen devices and components were
`tested). Instead, we have focused on two key events that
`helped shape the development of the SOL pen device; these
`two events are exemplary for the development process and systems
`applied for SOL. An overview of the development method
`is shown in Figure 3. In brief, each stage of the development
`process involved an iterative approach, and the process was
`guided by user research as well as laboratory testing.
`First, we wish to discuss the rationale for developing the
`SOL pen devices with distinct body colours specific for the
`delivery of long- or short-acting insulin devices. Second, we
`wish to discuss the impact of mechanics on the function of
`the dose dial extension and injection force. For both charac(cid:173)
`teristics, we will present the results of studies in which these
`characteristics were identified, provide a summary of the impact
`on the verification process and the final validation testing.
`
`2. Case studies
`
`2.1 Colour differentiation
`Problems with visual acuity are relatively common in diabetes,
`and may be either age related, such as macular degeneration
`or cataract formation [16,17], or diabetes related, with onset
`and progression of diabetic retinopathy [18-20]. People with
`
`diabetes, particularly those with type 1 diabetes, but also
`some with type 2, often use more than one type of insulin
`to manage the basal and prandial insulin requirements,
`which can be provided by insulin glargine and insulin
`glulisine, respectively.
`Owing to the differences in typical dose and pharmaco(cid:173)
`dynamic characteristics of basal and prandial insulins, it is
`important that the delivery devices (pen device or vial) are
`sufficiently differentiated to ensure a low risk for confusing
`the two insulin formulations and to minimise the risk of
`hypoglycaemia. Typically,
`this may involve some colour
`applied to the label and dose button of the device along
`with text and potentially tactile features.
`
`2.1.1 Colour deficiencies
`Colour deficiencies in people with diabetes occur primarily
`as a result of retinopathy [21], which is associated with altered
`colour perception owing to a reduction of light falling on
`the retina and the death of cones where the oxygen supply
`is restricted, or maculopathy, such as age-related macular
`degeneration, which is associated with an accumulation of
`fluid in the cone-rich area of the fovea, leading to distorted
`vision along with altered colour perception [16]. People with
`poorly controlled diabetes (type 1 or type 2) are at increased
`risk of developing retinopathy, whereas risk factors for
`maculopathy include aging, smoking and poor glycaemia
`
`106
`
`Expert Opin. Drug De/iv. (2009) 6(1)
`
`Sanofi Exhibit 2116.005
`Mylan v. Sanofi
`IPR2018-01675
`
`

`

`User input
`
`Risk
`management
`
`Bode
`
`SoloStar
`design process
`
`Design
`requirements
`-Intended use
`-TPP
`-PDS
`
`Evaluation/
`verification
`
`Survey
`
`Survey
`
`Survey
`
`Survey
`
`Design
`validation
`study
`
`1st risk
`analysis
`
`Concepts
`
`Proof of principle
`prototype
`
`Proof of concept
`prototype, V1
`
`Proof of concept
`prototype, V2
`
`Design
`evaluation
`
`Design
`evaluation
`
`Design
`verfication
`
`1st FMEA
`
`Revised
`FMEA
`
`Design ready for 00-series
`
`Revised risk
`analysis 1-------
`
`00-series
`prototype
`
`Design
`verfication
`
`Revised risk
`1------..
`an a I y sis
`
`>--------
`
`Design ready for production
`
`Patient
`trial
`(Australia)
`
`Initial post
`market
`surveillance
`
`Design
`verfication
`
`Production
`devices
`
`Launch
`
`Continuous
`improvement
`programme
`
`Figure 3. A summary of the key stages of the SOL pen device development process and tests done at each stage.
`FMEA: Failure mode and effects analysis; PDS: Product design specification; TPP: Target product profile.
`
`Expert Opin. Drug De/iv. (2009) 6(1)
`
`107
`
`Sanofi Exhibit 2116.006
`Mylan v. Sanofi
`IPR2018-01675
`
`

`

`Design verification and validation of SoloSTAR (SOL)
`
`control [22]. In addition, colour perception is altered with
`aging; as the lens of the eye ages, it becomes less transparent
`and absorbs more blue light. This results
`in a slight
`blue-weak colour deficiency, particularly in adults aged > 30.
`Inherited colour deficiency, although not related to diabetes,
`is a significant concern; as many as 8% of Caucasians are
`thought to be colour deficient, with deuteranopia (impaired
`middle wave receptor; confusion between red and green)
`being the most common [23].
`
`2.1.2 Colour discrimination
`Discrimination of colour is based on three dimensions:
`
`1. Hue: the perceptual quality of light of different wavelengths.
`2. Saturation: the purity of a hue, which is reduced as more
`black, grey or white, is introduced into the hue.
`3. Brightness: the amount oflight reflected by a surface.
`
`People with normal colour vision are unable to differentiate
`hue in low lighting levels, whereas people with colour vision
`deficiencies are restricted in their perception of hues, even at
`normal lighting levels. As a result, to aid colour discrimination,
`colours should differ sufficiently on all three dimensions.
`
`2.1.3 Implications for the treatment of diabetes
`Traditionally, insulin vials and pen devices have relied on
`text for differentiation of the type of insulin, although colour
`swatches were also used. The International Diabetes Federation
`(IDF) introduced a colour coding scheme to help identify
`certain insulin formulations.
`
`2.1.4 Study: review of colour pairings based on
`human factors for two insulin pen devices
`The development of the SOL pen device was initially planned
`to provide one pen device body colour, with differentiation
`between the long- and short-acting insulins provided by colours
`on the label and the dose button, as well as the text on the label.
`Results of some user and ergonomics/human factors studies
`indicated that the differentiation provided is sufficient and in
`line with common practice, but that further means would be
`beneficial. The results of these earlier studies suggested that there
`was potential to improve the differentiation of the two insulin
`pen devices; two distinct colour schemes should be used for
`pens that deliver the long- or short-acting insulin. A range of body
`colour options were considered to help differentiate between the
`two insulins, as well as differentiate within the sanofi-aventis
`portfolio and relative to pen devices already available on the
`diabetes market. Therefore, a study was designed to determine
`the optimal pairings of colours for the two SOL pens.
`The team investigated potential options for further colour
`differentiation; the largest area of the pen device is the pen body
`itself; hence, it was decided to differentiate the two insulins
`through distinct pen body colours. A variety of colour options
`was developed from a technical perspective (what is technically
`feasible) and analysed as part of the design validation [24],
`and the optimal colour pairing is shown in Figure 4.
`
`2.1.s Discussion
`In terms of insulin pen differentiation, the structured develop(cid:173)
`ment process has proven to be a successful tool to detect user
`needs (i.e., the need for differentiating features), highlight
`room for improvement (i.e., introduce body colour as well
`as labels for differentiation), make the best choices between
`available options (i.e., optimal colour pairs) and confirm that
`the optimal solution has been identified (i.e., user testing).
`As a result of the studies described earlier, the SOL pen
`device is available in two body colours to aid differentiation
`between the long- or short-acting insulins (Figures 1 and 4),
`which have been validated in studies with patients with poor
`visual acuity or colour blindness. Furthermore, the SOL pen
`device contains several extra features that can help the user
`discriminate between a SOL pen device that delivers long- or
`short-acting insulin, including label text and colours, dose
`injection button colour and tactile features. The use of two
`differentiated body colours has also been validated in user
`studies, and the SOL pen device was associated with low
`risk of confusion in terms of insulin type as well as pen
`devices from other manufacturers (unpublished observations).
`
`2.2 Ease of injection
`The design brief for the SOL pen device required a low
`injection force and short dial extension to allow easy injection.
`However, these two factors are inextricably linked in terms
`of mechanics of operation. Reducing the injection force or
`the dial extension is relatively simple; however, development
`of earlier injection devices indicated that reducing the one
`will normally increase the other. Thus, for the SOL pen
`device, the challenge was to inject large doses on short
`strokes at low forces.
`To achieve this, the team developed a totally new concept.
`As the low injection force and short dial extension are
`conflicting criteria, detailed mathematical modelling of the
`mechanism concepts was required, as well as stress and
`material analyses such as Finite Element Analysis or Moldflow.
`Careful selection of materials was needed both for low
`friction and high stress resistance.
`During the continuous user studies, the benefit of low
`and consistent injection force was identified to be a key
`differentiating factor with regard to ease of use. As a result,
`the injection mechanism was exposed to several stages of the
`iterative design verification (from a technical perspective)
`and validation (from a user perspective) process, to ensure
`the injection force characteristics were as good, or better,
`than both the FP and LP.
`
`2.2.1 Study: dose injection force
`The design brief for
`the SOL pen device defined an
`injection force lower than FP and LP. In user studies, a variety
`of injection forces were tested to determine the impact of
`the different forces on the perception of users; the data
`helped the team to have a set goal to design an appropriate
`mechanism to achieve this.
`
`108
`
`Expert Opin. Drug De/iv. (2009) 6(1)
`
`Sanofi Exhibit 2116.007
`Mylan v. Sanofi
`IPR2018-01675
`
`

`

`Bode
`
`Lantus®
`
`Figure 4. The Lantus® SOL (insulin glargine) and Apidra® SOL (insulin glulisine) products showing the colour pairing
`(with European labels).
`
`In addition to the earlier-mentioned user studies, the dose
`injection force characteristics of SOL versus FP and LP have
`been assessed in two separate studies, and all force characteristics
`were lower with SOL, particularly peak injection force and
`mean injection force [5,6].
`The mean injection force required to dispense 40 units in
`4 sec for the SOL, FP and the LP in these two studies was
`in the range of 10.7 - 10.8 Newtons [NJ, 15.5 - 17.1 N
`and 23.5 - 24.9 N, respectively [5,6]. Peak dose force for the
`SOL, FP and the LP was in the range of 14.4 - 14.7 N,
`22.9 - 25.0 N and 30.9 - 31.8 N, respectively [5,6]. The mean
`injection force was similar across the two studies; SOL had
`- 40% lower value than LP and - 30% lower than that
`measured for FP [5,6]. These laboratory-based studies are
`consistent with the subjective assessment by people with
`diabetes, as more people rated the effort to dispense a
`40 unit dose with SOL (63% of participants) as best compared
`with either the FP (19%) or LP (16%) [25].
`
`2.3 Patient usability of insulin pens
`Finally, it is important to consider the end result of the
`design process; the impact the pen has on patient usability.
`An open-label study done by Haak et al. across four countries
`(United States, Germany, France and Japan) compared four
`pen devices (SOL, LP, FP and Pen X [a prototype pen]) in
`510 patients with diabetes [25]. A significantly higher proportion
`of patients correctly prepared the pen and performed an
`injection into a receptacle with SOL (94%) compared with
`the FP (90%; p < 0.05), LP (61 %; p < 0.05) and Pen X
`(64%; p < 0.05). Similar results were seen regardless of
`previous pen use, age and manual/visual impairments. In
`addition to usability, a higher proportion of patients had an
`overall preference for SOL (53%) compared with FP (31 %;
`p < 0.05) or LP (15%; p < 0.05) [25].
`
`3. Discussion
`
`Here, we have provided a brief overview of the design and
`developmental process to bring the two SOL insulin pen
`devices that deliver either long- or short-acting insulin from
`a concept on paper to mass-produced insulin pen devices.
`The processes involved in the design and development of
`
`the SOL pen device can be likened to those involved in
`developing pharmaceutical drugs, which requires preclinical
`(i.e., concept design and selection) and Phase I - IV studies
`(i.e., validation and verification). However, there are significant
`differences with respect to the iterative process, in which a
`prototype is tested from both a technical and user perspective,
`and the resulting data are used to change the design as
`needed to then go into testing (user and technical); this
`approach introduces an element of consumer products into
`the development process. Only a lively dialog with several
`iterations between users and the development/device design
`team allowed the SOL pen device to be tailored to suit
`the patient needs and be manufacturable. The primary
`considerations in the design brief were to develop an easy to
`use insulin pen device with an injection force lower than
`that of other pens on the market. Within a laboratory
`setting, the SOL pen device has a lower and smoother injection
`force than both FP and LP [5l.
`The ease of use of the SOL pen device has also been
`demonstrated
`in a single-centre,
`two-group sequential
`study [3]. In that study, after a training session on the day
`before assessment, patients were able to correctly administer
`three doses of insulin.
`As demonstrated in the Haak et al. study, more patients
`correctly used and preferred SOL and FP than the LP and a
`prototype pen [25]. Similar ease of use and patient preference
`for the SOL were seen regardless of age, manual dexterity or
`visual impairments, factors that may have particular importance
`when considering elderly patients with diabetes, or patients
`that have peripheral neuropathy, a common complication
`of diabetes [26,27].
`the dose
`injected
`the users
`In
`these
`three studies,
`into a receptacle or injection pad; in the case of the
`force measurement
`[5l, dose deliveries were performed
`by a force stand. As these studies do not provide evidence
`of the usability of the SOL pen device in a clinical setting
`(i.e., at least once each day for a duration of time),
`an observational survey of everyday clinical practice was
`done in Australia to evaluate the usability and safety
`of SOL, which showed that most of the participants
`(95%) were very satisfied or satisfied with using SOL to
`inject insulin [Il.
`
`Expert Opin. Drug De/iv. (2009) 6(1)
`
`109
`
`Sanofi Exhibit 2116.008
`Mylan v. Sanofi
`IPR2018-01675
`
`

`

`Design verification and validation of SoloSTAR (SOL)
`
`4. Conclusions
`
`Here, we have presented data from two aspects of the design
`verification and validation of the SOL pen device and have
`shown how these have impacted the final production model,
`which was approved for use in a clinical setting to administer
`either long- or short-acting insulin. We have presented
`selected aspects from a wealth of data that we collected
`through the iterative development process. This comprehensive
`design verification and validation process with rigorous
`testing has resulted in a final product that is easy to use
`with a dose injection force lower than both FP and LP and,
`in a nonclinical environment, is preferred by more people
`than the FP and LP. Finally, within a clinical setting, the
`use of SOL delivering long-acting insulin in 5,983 patients
`with type 1 or type 2 diabetes was associated with a low
`incidence of adverse events (0.3%) and adverse reactions
`(0.15%; such as skin reactions, hypoglycaemia, pruritis),
`in addition to a low rate of reported technical problems
`(0.6%; such as leakage, defective dose knob or incomplete
`dose delivery) [28].
`
`s. Expert opinion
`
`Insulin therapy becomes a necessary part of a patient's disease
`management in order to reach appropriate glycaemic control
`in type 1 diabetes mellitus and following disease progression
`in type 2 diabetes mellitus. Common stressors of injection
`based insulin therapy affecting people with both types of
`diabetes include the psychological aspects of injecting an
`accurate dose and the reproducibility of doing so;
`this
`can lead to a decrease in patient compl

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket