throbber
tffD
`
`Main Menu
`
`&15
`
`,.,1-1
`
`l
`
`Generator Column Wash
`
`Generator column ,vash is complete.
`
`Time until next elution!
`
`9 Min 50 Sec
`
`Generator Column Wash
`
`Gen«.atorcolurnn wash in progress.
`
`•
`
`I
`
`FIGURE 6
`
`1702 of 1754
`
`

`

`'f-:fO
`
`Main Menu
`
`&npty\'41stit 91:»ttl,t
`
`[ ~1'411i)rCdllm \\\uh
`.
`-
`p
`9-kil\rough~
`
`l>4:tt Ope!'alioin
`
`Log.w.,t
`
`:::,13B
`
`~?5
`
`Breakthrough Test
`Press start button to begin breakthrough tQst
`
`'i:/4
`
`Breakthrough Test
`
`fnsertMWqc\iialinto~tution \lfal sb'-\iki
`PrQss 110ttto co~.
`
`I
`
`7-;t~
`
`Breakthrough Test
`Breakthrough E-1ution in progress.
`
`-
`
`-
`
`FIGURE 7A
`
`1703 of 1754
`
`

`

`Breakthrough Test
`~akthrough elution compl'1te.
`Enw rf>.t2 roding fr<im dOSQ calibrator
`
`mCI
`
`Tfme c·
`
`Enter time sinc-e end of elution
`
`'"'~--']
`
`I
`
`Breakthrough Test
`
`Br#3akthrough elution complete.
`
`TimQfrom end of elution:
`
`8 Min 50 Sec
`
`-
`
`91-9
`
`Breakthrough Test
`
`Breakthrough elution compl'1te,
`Pf'Qss dona to mtUrn·to main menu.
`
`• • •
`
`FIGURE 78
`
`1704 of 1754
`
`

`

`'9f>!
`
`Breakthrough Test
`Breakthrough elution tompkrt~.
`mCI 54.5 Tfme 2 Mn30 w:
`Enter breakthrough ieading at 60 min
`
`Enter bac:lqJrouoo radiation reading
`
`Time 5
`
`• • •
`
`UC(
`
`Breakthrough Test
`Breakthrough test compkrte.
`~tS!
`Sr-82/Rb-82 = 0.0005 uCi / mCi
`Sr-85/Rb~82 = 0.11 uCi / mCi
`
`Results should not ex:ceed :ailowable limits:
`Sr-s21m:.s2 < 0.02 ua I mCi
`Sr-ss/Rb-ai < 0,2 ua I ma
`
`•
`
`FIGURE 7C
`
`1705 of 1754
`
`

`

`e1t/
`
`Infusion System Calibration
`
`Insert MW q<: vial into ~ution vial sheid.
`Pl'Qss nQXt to continua .
`
`I
`
`g-:;-b
`
`Infusion System Calibration
`Prffl start button to l»;.tln calibration.
`
`Main Menu
`
`limptyW.m9onle
`
`013
`
`sJS
`
`Infusion System Calibration
`
`~OmCi
`fsoml
`f so mt/min (DEFAULT)
`
`j
`
`P.ltient Oose
`
`Patient Volume
`
`FlowRate
`DoseRata
`1hmshofd
`Elution \iolu!Tl(!f
`
`•
`
`•
`
`FIGURE SA
`
`1706 of 1754
`
`

`

`to-:fr
`
`B,¥-8
`
`Infusion System Calibration
`
`..... -.. \ I
`t
`
`J
`if
`
`,~
`
`\
`
`--~~t~--~--"'
`Otl:nm_,11,._.,,
`
`.f#'~"~.._.«;H~~., -~-4-\'1~·'<'~0 ...... t<
`
`-
`
`Infusion System Calibration
`
`Calibratkln elution complet~.
`
`TIO'IQfrom end of elution:
`
`8 Min 50 Sec
`
`I
`
`BB(J
`
`Infusion System Calibration
`C~ibratioo elutkln com~
`Entw tb-62 telding from dose calibr.ator
`
`ma(
`
`Ent«tiffiQ slna'! end of elution
`
`Time
`
`I
`
`•
`
`Infusion System Calibration
`
`Calibratkln artion comp!~.
`Result:
`calibration Ratio = 0.96
`tewgenfflt« callbmk!nsho~ b!wtthln0.95 ti:, u,s
`Dallycallbmk!nshookl beWlthlnO.tto L 1
`
`FIGURE 88
`
`1707 of 1754
`
`

`

`•
`
`992-
`
`Automatic Infusion
`
`Enterpati~t id .
`Press nextto continue.
`
`I
`
`91-1
`
`Automatic Infusion
`
`•
`
`P<'itiient Dost
`
`P<'itientVolu~
`
`Row Rate
`DoseRate
`Threshold
`Elution Volume
`
`FIGURE 9A
`
`tj?-3
`
`Automatic Infusion
`Attach Hew P.tient UM
`
`1708 of 1754
`
`

`

`
`“Bulweyy
`
`
`
`
`
`
`uoIsnjuySnewony
`
`
`
`wepadeaedaodenw
`
`t
`
`ANEpeMesdeIng
`“teeBRINHA
`th ;11
`·- .
`penedundype
`~ i
`C I»
`E ll I
`~ ,1 '
`
`
`
`1709 of 1754
`
`~
`
`~
`
` iil
`
`.
`--
`'2
`
`,;I: . . .
`
`Vl
`Q.
`
`E ::,
`
`Q.
`<U
`~
`
`g6agnosis
`
`-LL
`
`1709 of 1754
`
`
`
`
`

`

`Automatic Infusion
`
`Patient id: d430b
`
`====~-"-~•~M-,-,.__,_""'""'
`
`__ , --·
`
`Automatic Infusion
`
`Infusion complete,
`
`i11T11i? from end of infusion:
`
`8 Min 50 Sec
`
`•
`
`•
`
`FIGURE 9C
`
`l·., '"""\·t
`
`\
`
`i
`f
`
`-··-
`
`Automatic Infusion
`,'' :
`
`t:r.,"'•*"""
`··~»;;c~:;,~,lHo,o.«»1'~*•""-"""
`
`. . . . . ..,..._, _ _ lb;
`
`I
`
`C/ 1-!
`
`Automatic Infusion
`
`Infusion Summ.uy.
`
`Patient Dose
`P-itkmtV•fu~
`Butlon V•tu~
`
`52 mCi
`14ml
`28 ml
`
`Press done to ~um to the main menu.
`
`1710 of 1754
`
`

`

`1/10
`
`qs3
`
`t/B I
`qq/
`
`•
`
`Purge With Air
`
`Remove saline bag.
`Attach elution test vial to patient line.
`
`q15~
`
`Purge With Air
`
`Wamlng: You must bypass the generator
`if you intend to continue using it.
`
`•
`
`I
`
`Purge With Air
`
`FIGURE 10
`
`1711 of 1754
`
`

`

`SHIELDING ASSEMBLIES FOR INFUSION SYSTEMS
`
`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`RELATED APPLICATIONS
`
`[01] The present application is related to the following commonly assigned utility
`
`patent applications, all of which are filed concurrently herewith and all of which
`
`are hereby incorporated by reference in their entireties: Practitioner Docket No.
`
`56782.1.6, entitled: INFUSION SYSTEM CONFIGURATIONS; Practitioner
`
`Docket No. 56782.1.7, entitled: INFUSION SYSTEMS INCLUDING
`
`COMPUTER-FACILITATED MAINTENANCE AND/OR OPERATION; and
`
`Practitioner Docket No. 56782.1.8, entitled: CABINET STRUCTURES
`
`SUPPORTING INFUSION SYSTEMS.
`
`TECHNICAL FIELD
`
`[02] The present invention pertains to systems that generate and infuse
`
`radiopharmaceuticals, and, more particularly, to shielding assemblies thereof.
`
`BACKGROUND
`[03] Nuclear medicine employs radioactive material for therapy and diagnostic
`
`imaging. Positron emission tomography (PET) is one type of diagnostic imaging,
`
`which utilizes doses of radiopharmaceutical, for example, generated by elution
`
`within a radioisotope generator that are injected, or infused into a patient. The
`
`infused dose of radiopharmaceutical is absorbed by cells of a target organ, of the
`
`patient, and emits radiation, which is detected by a PET scanner, in order to
`
`generate an image of the organ. An example of a radioactive isotope, which may
`
`be used for PET, is Rubidium-82 (produced by the decay of Strontium-82); and
`
`an example of a radioisotope generator, which yields a saline solution of
`
`Rubidium-82, via elution, is the CardioGen-82® available from Bracco
`Diagnostics Inc. (Princeton, NJ).
`
`[04] Whether the half-life of a particular radioactive isotope, employed by a
`
`radiopharmaceutical, is relatively short or long, a patient undergoing a nuclear
`
`imaging procedure is not typically exposed to a significant amount of radiation.
`
`1
`
`1712 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`However those personnel, whose job it is to set up and maintain
`
`radiopharmaceutical infusion systems, and to administer doses therefrom, are
`
`subject to more frequent exposures to radiation. Therefore, shielding
`
`assemblies, which provide a radiation barrier to protect these personnel from
`
`excessive exposure to radiation sources, are an important component of
`
`radiopharmaceutical generators and infusion systems. These shielding
`
`assemblies are typically formed with lead sidewalls, the bulk and weight of which
`
`can pose difficulties for the personnel who regularly set up, maintain and use the
`
`systems. Thus, there is a need for improved shielding assemblies employed by
`
`systems that generate and infuse radiopharmaceuticals.
`
`BRIEF DESCRIPTION OF THE DRAWINGS
`
`[05] The following drawings are illustrative of particular embodiments of the present
`
`invention and therefore do not limit the scope of the invention. The drawings are
`
`not to scale (unless so stated) and are intended for use in conjunction with the
`
`explanations in the following detailed description. Embodiments of the present
`
`invention will hereinafter be described in conjunction with the appended
`
`drawings, wherein like numerals denote like elements.
`
`[06] Figure 1A is a first perspective view of an infusion system, according to some
`
`embodiments of the present invention.
`
`[07] Figure 1 B is another perspective view of a portion of a cabinet structure of the
`
`system shown in Figure 1A, according to some embodiments.
`
`[08] Figure 1C is a second perspective view of the system shown in Figure 1A,
`
`according to some embodiments.
`
`[09] Figure 1 D is a schematic of an infusion circuit, according to some embodiments
`
`of the present invention.
`
`[1 O] Figure 2A is a perspective view of a shielding assembly for an infusion system,
`
`such as that shown in Figures 1A-C, according to some embodiments of the
`
`present invention.
`
`2
`
`1713 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`[11] Figure 2B is a perspective view of a framework of the system, according to some
`
`embodiments, with an enlarged detailed view of a component of the system,
`
`according to some embodiments.
`
`[12] Figure 3A is another perspective view of the shielding assembly shown in Figure
`
`2A.
`
`[13] Figure 3B is a perspective view of the infusion circuit, shown in Figure 1 C,
`
`configured and routed, according to some embodiments.
`
`[14] Figure 3C is a perspective view of a disposable infusion circuit subassembly,
`
`according to some embodiments.
`
`[15] Figure 3D is a frame for the subassembly shown in Figure 3C, according to some
`
`embodiments.
`
`[16] Figure 4 is a main menu screen shot from an interface of a computer, which may
`
`be included in systems of the present invention, according to some
`
`embodiments.
`
`[17] Figure 5A is a schematic showing a first group of successive screen shots from
`
`the computer interface, according to some embodiments.
`
`[18] Figure 5B is a pair of screen shots from the computer interface, which provide
`
`indications related to eluant volume levels in a reservoir of the system, according
`
`to some embodiments.
`
`[19] Figure 5C is a schematic showing a second group of successive screen shots
`
`from the computer interface, according to some embodiments.
`
`[20] Figure 6 is a schematic showing a third group of successive screen shots from
`
`the computer interface, according to some embodiments.
`
`[21] Figures 7A-C are schematics showing a fourth group of successive screen shots
`
`from the computer interface, according to some embodiments.
`
`[22] Figures 8A-B are schematics showing a fifth group of successive screen shots
`
`from the computer interface, according to some embodiments.
`
`[23] Figures 9A-C are schematics showing a sixth group of successive screen shots
`
`from the computer interface, according to some embodiments.
`
`[24] Figure 10 is a schematic showing a seventh group of successive screen shots
`
`from the computer interface, according to some embodiments.
`
`3
`
`1714 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`DETAILED DESCRIPTION
`
`[25] The following detailed description is exemplary in nature and is not intended to
`
`limit the scope, applicability, or configuration of the invention in any way. Rather,
`
`the following description provides practical illustrations for implementing
`
`exemplary embodiments. Utilizing the teaching provided herein, those skilled in
`
`the art will recognize that many of the examples have suitable alternatives that
`
`can be utilized.
`
`[26] Figure 1A is a first perspective view of an infusion system 10, according to some
`
`embodiments of the present invention, wherein system 10 is shown supported by
`
`a cabinet structure, which includes a platform 113 (seen better in Figure 2B) and
`
`a shell 13; shell 13 extends upward from a skirt 11, that surrounds platform 113,
`
`to surrounds an interior space in which a portion of infusion system 10 is
`
`contained (-seen in Figure 1 C). Shell may be formed from panels of injection(cid:173)
`
`molded polyurethane fitted together according to methods known to those skilled
`
`in the art. Figure 1A illustrates the cabinet structure of system 10 including a grip
`
`or handle 14, which extends laterally from shell 13, in proximity to an upper
`
`surface 131 thereof, and a post 142, which extends upward from shell 13, and to
`
`which a work surface, or tray 16 and a computer 17 are, preferably, attached, via
`
`an ergonomic, positionable mount. According to some embodiments, computer
`
`17 is coupled to a controller of system 10, which is mounted within the interior
`
`space surrounded by shell 13, and a monitor 172 of computer 17 not only
`
`displays indications of system operation for a user of system 10, but also serves
`
`as a device for user input (e.g. touch screen input). However, according to
`
`alternate embodiments, another type of user input device, known to those skilled
`
`in the art, may be employed by computer 17. Other types of user input devices
`
`may included, for example, a keyboard, a series of control buttons or levers, a
`
`barcode reader (or other reader of encoded information), a scanner, a computer
`
`readable medium containing pertinent data, etc. The user input device may be
`
`mounted on the cabinet structure of system 10, as shown, or may be tethered
`
`thereto; alternatively the user input device may be remote from system 10, for
`
`example, located in a separate control room. According to some additional
`
`4
`
`1715 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`embodiments, another user input device, for example, in addition to a touch
`
`screen of computer 17, may be remote from system 10 and used to start and
`
`stop infusions. Operation of system 10, which is facilitated by computer 17, will
`
`be described below, in conjunction with Figures 4-9C.
`
`[27] Figure 1A further illustrates two pairs of wheels 121, 122, mounted to an
`
`underside of platform 113, to make system 10 mobile; handle 14 is shown
`
`located at an elevation suitable for a person to grasp in order to maneuver
`
`system 10, from one location for another, upon pairs of wheels 121, 122.
`
`According to some preferred embodiments, one or both pairs of wheels 121, 122,
`
`are casters, allowing for rotation in a horizontal plane (swivel), in order to provide
`
`additional flexibility for maneuvering system 10 in relatively tight spaces.
`
`[28] Figure 1 B is a perspective view of a portion of system 10, on a side 111 of the
`
`cabinet structure, which is in proximity to wheels 121. Figure 1 B illustrates a
`
`lever or pedal 125, which is located for activation by a foot of the person, who
`
`grasps handle 14 to maneuver system 10. In a neutral position, pedal 125 allows
`
`wheels 121, 122 to rotate, and, if embodied as casters, to swivel freely. Pedal
`
`125 may be depressed to a first position which prevents a swiveling of wheels
`
`122, according to those embodiments in which wheels 122 are casters, and may
`
`be further depressed to brake wheels 121, 122 from rolling and swiveling, upon
`
`reaching a desired location. Figure 1 B further illustrates a rear access panel
`
`17 4, for example, providing access to circuit boards of the aforementioned
`
`controller contained within the interior space surrounded shell 13, an optional
`
`lock 184, to secure panel 17 4, a power jack 118, for connecting system 10 to a
`
`power source, and a printer 117 for providing documentation of each patient
`
`infusion carried out by system 10, and of system quality control test results. In
`
`some embodiments, system 10 may further include one or more additional
`
`connectors, or ports (not shown), which allow system 10 to be coupled to, for
`
`communication with, other devices used for nuclear imaging procedures.
`
`[29] Figure 1A further illustrates upper surface 131 of shell 13 including several
`
`openings 133, 135, 139 formed therein. Figure 1 C is a partially exploded
`
`perspective view of system 10, wherein a removable access panel 132 is shown
`
`5
`
`1716 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`as a contoured portion of upper surface 131, which, when exposed, by lifting
`
`away a bin 18, that mates therewith, may be removed from another opening 137
`
`formed in upper surface 131. Figure 1 C also provides a better view of another
`
`panel 134 which may be lifted away from opening 139. According to the
`
`illustrated embodiment, openings 139 and 137 provide a user of system 10 with
`
`independent access to separate portions of infusion system 10, which are
`
`contained within shell 13, for example, to set up and maintain system 1 O; and
`
`openings 133 and 135 provide passageways for tubing lines to pass through
`
`shell 13. Figure 1C further illustrates an optional switch 102, which in case of an
`
`emergency, may be activated to abort function of system 10. With reference to
`
`Figures 1 A and 1 C, it may be appreciated that an arrangement of features
`
`formed in upper surface 131 of shell 13, in conjunction with bin 18, tray 16 and
`
`computer 17, provide a relatively ergonomic and organized work area for
`
`technical personnel who operate system 10.
`
`[30] Turning now to Figure 1 D, a schematic of an infusion circuit 300, which may be
`
`incorporated by system 10, is shown. Figure 1 D illustrates circuit 300 generally
`
`divided into a first part 300A, which includes components mounted outside shell
`
`13, and a second part 300B, which includes components mounted within the
`
`interior space surrounded by shell 13. (Parts 300A and 300B are delineated by
`
`dotted lines in Figure 1 D.) Figure 1 D further illustrates second part 300B of
`
`circuit 300 including a portion contained within a shielding assembly 200, which
`
`is designated schematically as a dashed line. Some embodiments of shielding
`
`assembly 200 will be described in greater detail, in conjunction with Figures 2A-B
`
`and 3A-B, below.
`
`[31] According to the illustrated embodiment, circuit 300 includes an eluant reservoir
`
`15, for example, a bag, bottle or other container, containing saline as the eluant,
`
`which is shown hanging from a post, or hanger 141 above upper surface 131 of
`
`shell 13 in Figure 1A; a syringe pump 33, for pumping the eluant from reservoir
`
`15, and a pressure syringe 34, for monitoring pumping pressure; a filter 37, which
`
`may also serve as a bubble trap, for the pumped eluant; a radioisotope generator
`
`21, through which the filtered eluant is pumped to create a radioactive eluate, for
`
`6
`
`1717 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`example an eluate carrying Rubidium-82 that is generated by the decay of
`
`Strontium-82, via elution, within a column of generator 21; and an activity
`
`detector 25, for measuring the activity of the eluate discharged from generator
`
`21, in order to provide feedback for directing the flow of the eluate, via a
`
`divergence valve 35WP, either to a waste bottle 23 or through a patient line
`
`305p, for example, to inject a dose of the radiopharmaceutical eluate into a
`
`patient. With reference back to Figure 1A, patient line 305p is shown extending
`
`out from shell 13, through opening 135, to a distal end thereof, which, according
`
`to some embodiments, includes a filter. Patient line 305p may be coupled to
`
`another line that includes a patient injection needle (not shown). Alternatively,
`
`patient line 305p may be coupled to another line (not shown), which extends
`
`from a source of another active substance, for example, a stress agent; the other
`
`line is coupled to the line that includes the patient injection needle, in order to
`
`permit injection of the additional active substance. Figure 1 D illustrates an eluant
`
`tubing line 301 coupled to reservoir 15 and to pump 33, and, with reference to
`
`Figures 1 A-B, it may be appreciated that opening 133 provides the passageway
`
`for tubing line 301 to enter the interior space surrounded by shell 13. According
`
`to some preferred embodiments, opening 133 includes a grommet-type seal that
`
`prevents leakage of eluant, which may spill from reservoir 15, into the interior
`
`space through opening 133, while allowing a user to assemble tubing line 301
`
`through opening 133. Likewise opening 135, which provides a passageway for
`
`patient line 305p, may include a grommet-type seal.
`
`[32] Figure 1 D further illustrates another eluant tubing line 302 coupled to pump 33
`
`and a divergence valve 35BG, which may either direct pumped eluant through a
`
`tubing line 304, to generator 21, or direct the pumped eluant through a by-pass
`
`tubing line 303, directly to patient line 305p. Divergence valve 35BG, as well as
`
`divergence valve 35WP, which directs eluate from an eluate tubing line 305
`
`either to a waste line 305w or to patient line 305p, may each be automatically
`
`operated by a corresponding servomotor (not shown), coupled to the controller
`
`(not shown) of system 10, which controller receives feedback from activity
`
`detector 25. When system 10 is operating for automatic infusion, to deliver a
`
`7
`
`1718 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`dose of radiopharmaceutical to a patient, for example, Rubidium-82 for
`
`diagnostic imaging, divergence valve 35BG is initially set to direct eluant to
`
`generator 21 and divergence valve 35WP is set to direct eluate from generator
`
`into waste bottle 23, until activity detector 25 detects the desired activity of the
`
`eluate, at which time the feedback from activity detector 25 causes the controller
`
`to direct the corresponding servo-motor to re-set valve 35WP for diverting the
`
`flow of eluate into patient line 305p. According to some embodiments, once a
`
`prescribed volume of the eluate has passed through patient line 305p, the
`
`controller directs the corresponding servomotor to re-set divergence valve 35BG
`
`for diverting the flow of eluant through by-pass line 303 and into patient line 305p
`
`in order to flush, or push any eluate remaining in patient line 305p into the
`
`patient. According to some embodiments, the controller may also direct the
`
`corresponding servomotor to re-set divergence valve 35WP back toward waste
`
`bottle 23, prior to the flush through by-pass line 303, in order to prevent back flow
`
`of eluant, through line 305, toward generator 21.
`
`[33] With further reference to Figure 1 D, it may be appreciated that shielding
`
`assembly 200 encloses those portions of circuit 300 from which radioactive
`
`radiation may emanate, with the exception of that portion of patient line 305p,
`
`which must extend out from shielding assembly 200 in order to be coupled to the
`
`patient for injection, or in order to be coupled to shielded sample vials, as will be
`
`described below. Thus, technical personnel, who operate system 10, are
`
`protected from radiation by shielding assembly 200, except at those times when
`
`an infusion is taking place, or when quality control tests require collection of
`
`eluate into sample vials. During infusions and quality control test sample
`
`collection, all technical personnel are typically in another room, or otherwise
`
`distanced from system 10, in order to avoid exposure to radiation during the
`
`infusion, and, according to some preferred embodiments of the present invention,
`
`system 10 includes at least one means for informing technical personnel that an
`
`infusion is about to take place or is taking place. With reference back to Figures
`
`1A and 1C, system 10 is shown including a light projector 100, mounted on post
`
`142. According to the illustrated embodiment, projector 100, projects a light
`
`8
`
`1719 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`signal upward, for maximum visibility, when pump 33 is pumping eluant and
`
`elution is taking place within generator 21, or at all times when pump 33 is
`
`pumping eluant. According to some embodiments, the light signal flashes on and
`
`off when the eluate is being diverted from generator 21 into waste bottle 23, and
`
`the light signal shines steadily when the eluate is being diverted through patient
`
`line 305p, or visa versa. According to other embodiments, a projector 100 shines
`
`a light having a first color, to indicate that eluate is being diverted to waste bottle
`
`23, and then shines a light having a second, different color, to indicate that eluate
`
`is being directed to patient line 305p for infusion. Light projector 100 may further
`
`project a more rapidly flashing light, for example, for approximately five seconds,
`
`once a peak bolus of radioactivity is detected in the eluate, to provide further
`
`information to technical personnel. Alternative means of informing technical
`
`personnel that an infusion is taking place may also be incorporated by system
`
`10, for example, including audible alarms or other types of visible or readable
`
`signals that are apparent at a distance from system, including in the control
`
`room.
`
`[34] When maintenance of system 10 requires the emptying waste bottle 23, relatively
`
`easy access to waste bottle 23 is provided through opening 139 in top surface
`
`131 of shell 13. It should be noted that technical personnel are preferably trained
`
`to empty waste bottle 23 at times when the eluate, contained in waste bottle 23,
`
`has decayed sufficiently to ensure that the radioactivity thereof has fallen below a
`
`threshold to be safe. Opening 139 is preferably located at an elevation of
`
`between approximately 2 feet and approximately 3 feet; for example, opening
`
`139 may be at an elevation of approximately 24 inches, with respect to a lower
`
`surface of platform 113, or at an elevation of approximately 32 inches, with
`
`respect to a ground surface upon which wheels 121, 122 rest. According to the
`
`illustrated embodiment, opening 139 is accessed by lifting panel 134; just within
`
`opening 139, a shielded lid or door 223 (Figure 2A) may be lifted away from a
`
`compartment of shielding assembly 200 that contains waste bottle 23. With
`
`further reference to Figure 1 C, it may be appreciated that opening 137 provides
`
`access to other portions of circuit 300 for additional maintenance procedures,
`
`9
`
`1720 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`such as changing out generator 21 and/or other components of circuit 300, as
`
`will be described below.
`
`[35] Figures 1A and 1C further illustrate a pair of relatively shallow external recesses
`
`190, which are formed in upper surface 131 of shell 13, for example, in order to
`
`catch any spills from infusion system; one of recesses 190 is shown located in
`
`proximity to post, or hanger 141, which holds reservoir 15, and in proximity to
`
`opening 133, through which tubing line 301 passes. Another recess 192 is
`
`shown formed in upper surface 131; a width and depth of recess 192 may
`
`accommodate storage of technical documentation associated with infusion
`
`system 10, for example, a technical manual and/or maintenance records, or
`
`printouts from printer 117 (Figure 1 B). With reference to Figure 1 C, upper
`
`surface 131 of shell 13 is shown to also include additional recesses 101, which
`
`are each sized to hold a shielded test vial, which contains samples from infusion
`
`system 10, for example, for breakthrough testing and/or calibration, which will be
`
`described in greater detail, below. Additional receptacles 180 are shown formed
`
`in bin 18, on either side of a handle 182, which facilitates removal of bin 18 away
`
`from shell 13. Technical personnel may, thus, conveniently transport bin 18 to a
`
`storage area for a collection of supplies, for example, sharps, gloves, tubing
`
`lines, etc ... , into one or more receptacles 180 thereof, and/or to a waste
`
`container where separate receptacles 180 of bin 18 may be emptied of waste,
`
`such as packaging for the aforementioned supplies, for example, deposited
`
`therein during infusion procedures.
`
`[36] Figure 2A is a perspective view of shielding assembly 200, according to some
`
`embodiments of the present invention. With reference to Figures 1 C and 2A,
`
`together, it may be appreciated that opening 137, in upper surface 131 of shell
`
`13, provides access to a lid or door 221 of a sidewall 201 of shielding assembly
`
`200, which sidewall 201 encloses a compartment sized to contain a radioisotope
`
`generator of system 10, for example, generator 21, previously introduced.
`
`According to the illustrated embodiment, opening 137 and door 221 are located
`
`at a lower elevation, for example, with respect to platform 113, than are opening
`
`139 and lid 223, which provide access to the compartment being formed by a
`
`10
`
`1721 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`sidewall 203 of shielding assembly 200 to contain waste bottle 23, as previously
`
`described. When panel 132 is separated from shell 13, and door 221 opened,
`
`generator 21 may be lifted out from an opening 231 (Figure 3A) which mates with
`
`door 221 of sidewall 201. A weight of generator 21, which includes its own
`
`shielding, may be between approximately 23 and approximately 25 pounds, thus,
`
`according to some preferred embodiments of the present invention, the elevation
`
`of each of openings 137 and 231, with respect to the lowermost portion of the
`
`cabinet structure, is between approximately 1 foot and approximately 2 feet, in
`
`order to facilitate an ergonomic stance for technical personnel to lift generator 21
`
`out from the compartment. According to an exemplary embodiment, when
`
`shielding assembly 200 is contained in the cabinet structure of Figure 1A,
`
`openings 137 and 231 are located at an elevation of approximately 12 inches,
`
`with respect to the lower surface of platform 113, or at an elevation of
`
`approximately 19 inches, with respect to the ground surface upon which wheels
`
`121, 122 rest. Figure 1 C further illustrates access panel 132 including a security
`
`lock 138, which mates with a framework 19 of system 10, shown in Figure 2B, in
`
`order to limit access to generator 21.
`
`[37] Figures 1 C and 2A further illustrate a lid or a door 225 of another sidewall 205
`
`(Figure 3A) of shielding assembly 200, which encloses another compartment that
`
`is accessible through opening 137 of shell 13, and which is located adjacent the
`
`compartment enclosed by sidewall 201. Each of doors 221, 225 are shown
`
`being attached by a corresponding hinge H, and another door 227 is shown
`
`attached to sidewall 203 by another hinge H. Figure 2A illustrates each of lid 223
`
`and doors 221, 225, 227 including a handle 232,212, 252 and 272, respectively,
`
`for moving lid 223 and doors 221, 225, 227, in order to provide access to the
`
`corresponding compartments, which can be seen in Figures 3A-B. Figure 2A
`
`further illustrates optional thumb screws 290, one securing lid 223 to sidewall 203
`
`and another securing door 221 to sidewall 201, or other means for securing the
`
`doors, which are known to those skilled in the art, may be incorporated. Each
`
`sidewall 201, 203, 205 and the corresponding lid/door 223, 221, 225, 227 thereof
`
`may be individually cast from 3% antimony lead, or from other known shielding
`
`11
`
`1722 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`materials, and then assembled together according to methods known to those
`
`skilled in the art.
`
`[38] According to the illustrated embodiment, doors 221, 225 are hinged to open in an
`
`upward direction, per arrows D and C, and, with reference back to Figure 1 C, a
`
`latch component 191 is provided to hold each of doors 221, 225 in an opened
`
`position, thereby, preventing doors 221, 225 from falling closed, which could
`
`pinch/crush fingers of technical personnel and/or tubing lines of circuit 300, when
`
`in the midst of a maintenance procedure. Figure 2B is a perspective view of
`
`framework 19 of the cabinet structure of system 10, according to some
`
`embodiments, to which latch component 191 is mounted; Figure 2B includes an
`
`enlarged detailed view of latch component 191, according to some embodiments.
`
`Figure 2B illustrates latch component 191 including a first pin 193, corresponding
`
`to door 225, and a second pin 195, corresponding to door 221; each pin 193, 195
`
`includes a lever end 193A, 193B, respectively, and a holding end 193B, 195B,
`
`respectively. An edge of each door 221, 225, upon opening of doors 221, 225,
`
`may push past the holding end 195B, 193B of the corresponding pin 195, 193, in
`
`a first direction, per arrow F, and then may rest against a respective side S95
`
`and S93 of each end 195B, 193B, until the corresponding lever end 195A, 193A
`
`is rotated in a counter-clockwise direction, per arrow cc, thereby moving the
`
`corresponding holding end 193B, 195B to make way for the closing of doors 221,
`
`225. Doors 221, 225 being held by latch component 191 in an open position may
`
`be seen in Figure 3A.
`
`[39] With further reference to Figure 2A, according to some preferred embodiments of
`
`the present invention, an edge of door 225 overlaps door 221 to prevent door
`
`221 from being opened, per arrow D, if door 225 is not opened, per arrow C; and
`
`an edge of door 227 overlaps an edge of door 225 to prevent door 225 from
`
`being opened if door 227 is not opened, per arrow B; and an edge of lid 223
`
`overlaps door 227 to prevent door 227 from being opened if lid 223 is not
`
`opened, per arrow A. Thus, access to the compartment enclosed by sidewall
`
`201 and containing generator 21 is only systematically allowed through a
`
`sequential opening of lid 223 and doors 227, 225, 221, since, when generator 21
`
`12
`
`1723 of 1754
`
`

`

`UTILITY PATENT APPLICATION
`Practitioner Docket No. 56782.1.5
`
`is replaced it is typic

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket