throbber

`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`
`
`Jubilant DraxImage Inc.,
`
`Petitioner
`
`v.
`
`Bracco Diagnostics Inc.
`
`Patent Owner
`
`
`
`
`
`
`U.S. Patent No. 9,299,467
`
`
`Inter Partes Review Case No.: 2018-01449
`
`PETITIONER’S REPLY BRIEF
`
`
`Mail Stop: PATENT BOARD
`Patent Trial and Appeal Board
`U.S. Patent & Trademark Office
`P.O. Box 1450
`Alexandria, VA 22313-1450
`
`
`
`

`

`TABLE OF CONTENTS
`
`IPR2018-1449
`
`Page
`Introduction ...................................................................................................... 1
`Ground 1: Klein Anticipates the Challenged Claims. ..................................... 2
`A. Klein Discloses a Computer That Prevents Patient Elutions in
`Response to Failed Strontium Breakthrough Tests. .............................. 3
`The ITC Rejected Bracco’s Expert Witness Testimony and
`Arguments. .......................................................................................... 11
`III. Ground 3: Bracco Distributed the Bracco Manual to Its Customers with
`No Confidentiality Markings. ........................................................................ 13
`IV. Bracco’s Attacks on Dr. Murthy’s Testimony Are Unjustified
`Distractions from the Remaining Issues in Dispute. ..................................... 15
`Conclusion ..................................................................................................... 18
`
`B.
`
`I.
`II.
`
`V.
`
`
`
`
`
`
`ii
`
`
`
`

`

`IPR2018-1449
`
`TABLE OF AUTHORITIES
`
`Cases
`Cordis Corp. v. Boston Scientific Corp.,
`561 F.3d 1319 (Fed. Cir. 2009) ..................................................................... 14, 15
`CRFD Research, Inc. v. Matal,
`876 F.3d 1330 (Fed. Cir. 2017). ............................................................................. 7
`Garrett Corp. v. United States,
`422 F.2d 874 (Ct. Cl. 1970) ................................................................................. 13
`Gemtron Corp. v. Saint-Gobain Corp.,
`572 F.3d 1371 (Fed. Cir. 2009) ............................................................................ 14
`Icon Health & Fitness, Inc. v. Strava, Inc.,
`849 F.3d 1034 (Fed. Cir. 2017) ............................................................................ 14
`In re Bond,
`910 F.2d 831 (Fed. Cir. 1990) ................................................................................ 7
`In re Reynaud,
`331 F.2d 625 (C.C.P.A. 1964) ............................................................................. 10
`Nobel Biocare Services AG v. Instradent USA, Inc.,
`903 F.3d 1365 (Fed. Cir. 2018) ............................................................................ 15
`ParkerVision, Inc. v. Qualcomm Inc.,
`903 F.3d 1354 (Fed. Cir. 2018) ............................................................................ 10
`Purgess v. Sharrock,
`33 F.3d 134 (2d. Cir. 1994) .................................................................................. 13
`Other Authority
`Manual of Patent Examining Procedure § 2145 (9th ed. Nov. 2015) ..................... 13
`
`
`
`
`
`
`iii
`
`
`
`

`

`IPR2018-1449
`
`EXHIBIT LIST
`
`EXHIBIT
`
`DESCRIPTION
`
`1001
`
`U.S. Patent No. 9,299,467
`
`1002
`
`1003
`
`Complaint, Bracco Diagnostics Inc. v. Jubilant DraxImage Inc., Case
`No. 3-18-cv-04422 (D.N.J.)
`
`Complaint,
`the Matter of Certain Strontium-Rubidium
`In
`Radioisotope Infusion Systems, and Components Thereof Including
`Generators, Inv. No. 337-TA 3303 (U.S. I.T.C.)
`
`1004
`
`Order Granting Joint Motion to Stay Pending Resolution of ITC
`Proceedings (D.I. 9) Case No. 3-18-cv-04422 (D.N.J.)
`1005 Waiver of Service, Bracco Diagnostics Inc. v. Jubilant DraxImage
`Inc.
`
`1006
`
`1007
`
`1008
`
`1009
`
`1010
`
`1011
`
`1012
`
`1013
`
`Prosecution History, U.S. application s.n. 12/137,356
`
`Prosecution History, U.S. application s.n. 12/137,377
`
`Prosecution History, U.S. application s.n. 12/137,363
`
`Prosecution History, U.S. application s.n. 12/137,364
`
`Prosecution History, U.S. application s.n. 12/808,467
`
`Prosecution History, U.S. application s.n. 14/455,623
`
`Alvarez-Diez, “Manufacture of strontium-82/rubidium-82 generators
`and quality control of rubidium-82 chloride for myocardial perfusion
`imaging in patients using positron emission tomography,” Applied
`Radiation and Isotopes, v. 50, pp. 1015-23 (1999)
`
`Klein et al, “Precision Control of Eluted Activity from a Sr/Rb
`Generator for Cardiac Positron Emission Technology,” Engr. in Med.
`and Biology Soc. 26th Annual International Conference of the IEEE,
`vol. 1, pp. 1393-1396 (2004)
`
`
`
`
`
`iv
`
`
`
`

`

`IPR2018-1449
`
`EXHIBIT
`
`DESCRIPTION
`
`1014
`
`1015
`
`1016
`
`1017
`
`1018
`
`1019
`
`1020
`
`1021
`
`1022
`
`1023
`
`Ran Klein, “Precise 82Rb Infusion System for Cardiac Perfusion
`Measurement Using 3D Positron Emission Tomography,” Ottawa-
`Carleton Institute for Electrical and Computer Engineering (Feb.
`2005)
`
`Declaration of Robert T. Stone, Ph.D.
`
`Curriculum Vitae of Robert T. Stone, Ph.D.
`
`Declaration of Ventakesh Murthy, M.D.
`
`Barker, et al., U.S. Patent 4,585,009, granted April 29, 1986
`
`U.S. Pharmacopeia 23 National Formulary 18 (1995)
`
`Declaration of Andy Adler, Ph.D. with Attachments A-C
`
`Bracco CardioGen-82® Infusion System User’s Guide, Rev. 07 (July
`20, 2004)
`
`Tate, et al., U.S. Patent Publication No. 2008/0177126, filed October
`31, 2007 and published July 24, 2008
`
`ISO 13485:2003 – Medical Devices – Quality Management Systems
`– Requirements for Regulatory Purposes (July 2003)
`
`1024
`
`21 CFR Part 820.1 (2005)
`
`1025
`
`1026
`
`1027
`
`1028
`
`Chatal, et al., “Story of rubidium-82 and advantages for myocardial
`perfusion PET Imaging,” Frontiers in Medicine, v. 2, art. 65, pp. 1-7
`(Sept. 11, 2015)
`
`Bracco CardioGen-82® Rubidium Rb 82 Generator, Rev. 43-8200
`(May 2000)
`
`10 CFR Part 20 (10 CFR 20.1001-2)
`
`10 CFR Part 20 (10 CFR 20.1003)
`
`
`
`
`
`v
`
`
`
`

`

`IPR2018-1449
`
`EXHIBIT
`
`DESCRIPTION
`
`1029
`
`1030
`
`1031
`
`1032
`
`1033
`
`1034
`
`1035
`
`EN 62274:2005 – Medical Electrical Equipment – Safety of
`Radiotherapy Record and Verify Systems (December 28, 2005)
`
`The Chemical Rubber Co., Handbook of Radioactive Nuclides, Yen
`Wang ed., 1969
`
`Complainant Bracco Diagnostics Inc.’s Responses to Respondent’s
`Fourth Set of Interrogatories (No. 68), In the Matter of Certain
`Strontium-Rubidium Radioisotope
`Infusion
`Systems,
`and
`Components Thereof Including Generators, Inv. No. 337-TA-1110
`(U.S.I.T.C.)
`
`Declaration of Carol Wadke
`
`deKemp, U.S. Patent Publication No. 2007/0213848
`
`deKemp, U.S. Patent Publication No. 2007/0140958
`
`Initial Determination on Violation of Section 337 and Recommended
`Determination on Remedy and Bond), In the Matter of Certain
`Strontium-Rubidium Radioisotope
`Infusion
`Systems,
`and
`Components Thereof Including Generators, Inv. No. 337-TA-1110
`(U.S.I.T.C. Aug. 1, 2019)
`
`1036
`
`U.S. Patent No. 9,814,826
`
`1037
`
`2001
`
`2002
`
`2003
`
`2004
`
`2005
`
`Exhibit 1002 from Dr. Murthy’s Deposition in IPRs 2018-01448, -
`01449, and -01450 (Bracco CardioGen-82® Infusion System User’s
`Guide, Rev. 11 (July 3, 2007))
`
`U.S. Patent Application Publication No. 2011/0178359, filed
`December 28, 2007
`
`[RESERVED]
`
`Declaration of Dr. Norbert Pelc
`
`ITC Deposition of Dr. Robert Stone, October 9, 2018 (Redacted)
`
`ITC Deposition of Dr. Robert Stone, October 10, 2018 (Redacted)
`
`
`
`
`
`vi
`
`
`
`

`

`IPR2018-1449
`
`EXHIBIT
`
`DESCRIPTION
`
`2006
`
`2007
`
`2008
`
`2009
`
`2010
`
`2011
`
`2012
`
`2013
`
`2014
`
`2015
`
`2016
`
`ITC Trial Testimony of Dr. Robert Stone, April 15, 2019 (Public
`Version)
`
`ITC Trial Testimony of Dr. Robert Stone, April 16, 2019 (Public
`Version)
`
`Errata of Dr. Robert Stone regarding Ex. 1015 of IPRs 2018-01448,
`-01449, and -01450
`
`Exhibit from Deposition of Dr. Robert Stone in IPRs 2018-01448, -
`01449, and -01450
`
`Deposition testimony of Dr. Robert Stone in IPRs 2018-01448, -
`01449, and -01450
`
`ITC Corrected Expert Report of Dr. Norbert Pelc (Redacted)
`
`ITC Trial Testimony of Dr. Norbert Pelc (Public Version)
`
`ITC Trial Exhibits of Dr. Norbert Pelc (Redacted)
`
`CAPINTEC CRC-15R User Manual, Nov. 2004
`
`CAPINTEC CRC-15R User Manual, July 2007
`
`Deposition testimony of Dr. Venkatesh Murthy in IPRs 2018-01448,
`-01449, and -01450
`
`2017
`
`Exhibit 1001 from Dr. Murthy’s Deposition (corrections to Ex. 1017)
`
`
`
`
`
`vii
`
`
`
`

`

`IPR2018-1449
`
`I.
`
`INTRODUCTION
`The Petition demonstrates that Klein discloses all claimed components of the
`
`’467 Patent’s infusion system, including a shielding assembly with a generator (red),
`
`a computer (green), and a dose calibrator (blue) that perform breakthrough testing
`
`on an eluate sample (pink):
`
`Ex. 1014, FIG 2-2 (annotated)
`
`
`
`The Petition also shows that the prior art teaches every limitation of the Challenged
`
`Claims, that Klein anticipates certain Challenged Claims, and that the modifications
`
`of Klein to practice the other Challenged Claims would have been obvious to a
`
`POSITA.
`
`The issues remaining for the PTAB to decide are narrow. There are no claim
`
`construction issues in dispute. POR § II.B. Although the parties define POSITA
`
`
`
`
`
`1
`
`
`
`

`

`IPR2018-1449
`
`differently, neither party argues that the differences between them are material to the
`
`outcome of this IPR. Id. § II.C. Bracco makes no challenge specific to Ground 2.
`
`Thus, the Board need not revisit these issues.
`
`Bracco’s limited counter-arguments are unpersuasive. On Ground 1, Bracco
`
`argues only that Klein does not disclose that its computer is “configured to prevent
`
`a patient infusion procedure if a breakthrough test result exceeds an allowable limit.”
`
`On Ground 3, Bracco disputes that its own product manual qualifies as a printed
`
`publication. As a last recourse, Bracco takes issue with clerical errors in, but not the
`
`substance of, Dr. Murthy’s declaration.1 These counter-arguments are inapt, as
`
`discussed below.
`
`II. GROUND 1: KLEIN ANTICIPATES THE CHALLENGED CLAIMS.
`On Ground 1, Bracco raises a single distinction against the prior art, namely,
`
`that Klein does not disclose a computer system that prevents patient infusions when
`
`high breakthrough tests are detected. POR § IV. As discussed below, Klein’s
`
`
`1 Bracco takes issue with the Bracco Manual and Murthy Declaration in all three
`
`related IPRs, but the roles of this evidence vary among these IPRs. For example, Dr.
`
`Murthy’s testimony is relied upon only to discuss the background technology in this
`
`proceeding.
`
`
`
`
`
`2
`
`
`
`

`

`IPR2018-1449
`
`computer disables patient elutions each night. Ex. 1014 at 19. Bracco’s expert, Dr.
`
`Pelc, agrees that this operation locks out patient elutions. Ex. 2012 at 981:9-10.
`
`Klein further discloses that the computer re-enables the patient elutions only after
`
`completing a morning breakthrough test with low breakthrough levels. Ex. 1014 at
`
`18, 29, 44. While Bracco’s expert disputes that Klein performs this latter operation,
`
`his opinion is contradicted by clear, straightforward disclosure from Klein.
`
`Therefore, the Board should adopt Ground 1 and cancel the affected claims.
`
`A. Klein Discloses a Computer That Prevents Patient Elutions
`in Response to Failed Strontium Breakthrough Tests.
`The Board should reject Bracco’s argument that Klein does not disclose the
`
`lockout feature, recited in the claims as follows:
`
`Claim 1:
`
`the computer is further configured to prevent a patient
`infusion procedure if a breakthrough test result exceeds
`an allowable limit.
`
`Claim 13:
`
`preventing, with the computer, a patient infusion
`procedure if a breakthrough test result exceeds an
`allowable limit.
`
`Klein discloses a computer-controlled infusion system. Ex. 1014 at 19, 30-
`
`31; see Ex. 1015 § IX.B.3; Pet. at 14-17. The computer provides a touch-screen user
`
`interface through which operators control the system. Ex. 1014 at 30-31. The
`
`computer “limit[s] user input to ensure validity” through the user interface by adding
`
`
`
`
`
`3
`
`
`
`

`

`IPR2018-1449
`
`buttons “only at relevant states” and removing them immediately thereafter. Id.; Ex.
`
`1015 § IX.B.3.
`
`Klein provides an exemplary user interface screen showing how the computer
`
`selectively enables and disables elutions (yellow) based on operational state. This
`
`example relates to test runs:
`
`Klein, FIG. 3-12(c) (annotated)
`
`
`
`Ex. 1014 at 53-54 (system requires proper user ID to enable test runs). Thus, Klein’s
`
`computer locks out different types of elutions—it enables and disables buttons for
`
`them on the user interface—based on the system’s operational state. Ex. 1015
`
`§ IX.B.3.
`
`With respect to patient infusions, Klein’s computer disables Patient Elution
`
`Runs each night:
`
`
`
`
`
`4
`
`
`
`

`

`IPR2018-1449
`
`Klein, FIG. 2-1
`
`
`
`Ex. 1014 at 19; Ex. 1015 § IX.B.5. Bracco’s expert, Dr. Pelc, agrees that this
`
`operation locks the system out until a strontium breakthrough test is performed:
`
`The Klein thesis discloses that the system is locked out until you do
`the calibration run. It's absolutely clear that Klein intends his system
`to require that that calibration run be done before patient elutions are
`done for the day, and then at midnight, that permission expires, and you
`have to redo the calibration run.
`
`Ex. 2012 at 981:9-14 (emphasis added); see also Ex. 1014 at 18 (calibration run
`
`includes strontium breakthrough test). Dr. Pelc distinguished the claims based on
`
`his incorrect inference that, in Klein, operators determine whether strontium
`
`breakthrough test results are sufficiently low to resume patient infusions, not the
`
`computer. Ex. 2012 at 1078:8-1079:12. Dr. Pelc’s opinion is contradicted by
`
`overwhelming disclosure from Klein.
`
`
`
`
`
`5
`
`
`
`

`

`IPR2018-1449
`
`Klein discloses that the computer does not re-enable patient infusions unless
`
`a calibration run is performed that yields strontium breakthrough values lower than
`
`a predetermined limit:
`
`Only after a calibration run with low Sr breakthrough has been
`successfully completed can patient elutions be carried out. …
`Once the daily protocol has been completed successfully, patient
`elutions are enabled until the end of the day.
`
`Ex. 1014 at 18 (emphases added).
`
`The 82Rb infusion system software must ensure that the protocol is
`followed (i.e. that each run is enabled only after the prerequisites have
`been completed successfully).
`
`Id. at 44 (emphasis added).
`
`The system must ensure compliance with the daily protocol described
`in the previous chapter. A flush followed by a calibration run and
`successful breakthrough measurement must be completed in order to
`enable patient elutions for the remainder of the day.
`
`Id. at 29 (emphases added).
`
`The software ensured that the daily protocol (discussed below) is
`followed and generated a recording of each elution.
`
`Id. at 17 (emphasis added). Contrary to Dr. Pelc’s opinions, Klein assigns
`
`responsibility for re-enabling patient infusions to the computer’s software. Thus, a
`
`POSITA would recognize from this disclosure that Klein’s computer operates as
`
`
`
`
`
`6
`
`
`
`

`

`IPR2018-1449
`
`claimed: It “prevent[s] a patient infusion procedure if a breakthrough test result
`
`exceeds an allowable limit.” Ex. 1001 at claims 1, 13; see also Ex. 1015 §§ XI.A.7,
`
`XI.J.5.
`
`Bracco’s counter-arguments are unpersuasive, and they should be rejected.
`
`First, Bracco argues that Klein’s disclosure does not “expressly state” that a lockout
`
`occurs. POR at 12-13. Anticipation, however, does not demand word-for-word
`
`correspondence between prior art and the claims. In re Bond, 910 F.2d 831, 832
`
`(Fed. Cir. 1990) (no ipsissimus verbis test). Instead, the relevant question is whether
`
`“one skilled in the art would reasonably understand or infer from the prior art
`
`reference’s teaching that every claim [limitation] was disclosed in that single
`
`reference.” CRFD Research, Inc. v. Matal, 876 F.3d 1330, 1338 (Fed. Cir. 2017).
`
`A POSITA would recognize immediately from Klein’s disclosure that the computer
`
`“prevent[s] a patient infusion procedure if a breakthrough test result exceeds an
`
`allowable limit.”
`
`Bracco’s second argument
`
`that Klein’s discussion of breakthrough
`
`experiments somehow suggests the computer does not prevent patient infusions as
`
`claimed (POR at 14), should be rejected. Klein disables Patient Elutions (orange),
`
`the only type of elution that qualifies as a “patient infusion procedure” under the
`
`claims:
`
`
`
`
`
`7
`
`
`
`

`

`IPR2018-1449
`
`Klein, FIG. 3-15 (annotated)
`
`
`
`The “experiments” to which Bracco refers involve a Calibration Run (green) not a
`
`“patient infusion procedure”:
`
`Daily calibration samples are used to test for the breakthrough of Sr
`activity; if significant Sr activity is detected, the generator cannot be
`used on humans. … We continued to use the generator on the
`development system past the specifications and experienced first
`breakthrough after approximately 40 L were eluted through the
`generator. Once breakthrough appeared, it was tracked over two
`months.
`
`
`
`
`
`8
`
`
`
`

`

`IPR2018-1449
`
`POR at 14-15 (quoting Ex. 1014 at 122) (emphasis added). Klein distinguishes these
`
`experiments and patient procedures. Id. at 53 (test runs have no clinical application).
`
`Klein never suggests that the breakthrough experiments involve any “patient
`
`infusion procedure.” To the contrary, Klein confirms that “if significant Sr activity
`
`is detected, the generator cannot be used on humans.” Id. at 122 (emphasis added);
`
`Ex. 1015 §§ XI.A.7, XI.J.5. Bracco’s claims impose no requirement that the
`
`computer must disable all operation, as Bracco suggests, only “patient infusion
`
`procedure[s].” Therefore, Bracco’s argument should be rejected.
`
`Bracco misreads Klein to argue that its discussion of manual entry of
`
`breakthrough readings somehow indicates that a lockout does not occur. POR at 15.
`
`Klein discloses that manual readings are the exception, not the norm, and automated
`
`breakthrough testing suffices in many cases:
`
`The integral activity recorded from the dose calibrator is used to
`calibrate the activity counter and verify that the calibration constant is
`within tolerance from previous records. If the dose calibrator
`sensitivity is sufficiently high, breakthrough measurements are
`conducted after 20 minutes from completion of the elution. If the dose
`calibrator is not sufficiently sensitive to measure breakthrough, the
`activity can be entered manually after measurement in a more sensitive
`device.
`
`Ex. 1014 at 43 (emphasis added); see also id. at 64 (The computer “read[s]
`
`breakthrough from the dose calibrator” at the conclusion of the “Wait for
`
`
`
`
`
`9
`
`
`
`

`

`IPR2018-1449
`
`Breakthrough Reading” state). Klein’s computer operates as claimed when it
`
`performs automated breakthrough readings, which suffices for present analysis.2 A
`
`prior art reference may meet a limitation if the reference discloses subject matter that
`
`is reasonably capable of operating so as to meet the limitation, even if it does not
`
`meet the limitations in all modes of operation. ParkerVision, Inc. v. Qualcomm Inc.,
`
`903 F.3d 1354, 1361 (Fed. Cir. 2018) ); In re Reynaud, 331 F.2d 625, 628 (C.C.P.A.
`
`1964) (prior art may be relied upon for all that it teaches).
`
`Bracco’s argument that responsibility for preventing patient elutions falls on
`
`the operator, not the computer, (POR at 16-18) is inconsistent with a plain reading
`
`of Klein. This argument relies on a tortured interpretation of FIG. 3-15 (above), in
`
`which Dr. Pelc concluded a calibration run is “successful” when it completes, even
`
`if high strontium breakthrough levels were detected. Ex. 2012 at 981:22-983:9 (“I
`
`interpret this to say that a successful calibration run in Klein means you went through
`
`these steps.”). Dr. Pelc’s opinion is not credible.
`
`
`2 Manual entry is also consistent with Klein’s disclosure that the computer performs
`
`lockout. No matter whether the computer receives breakthrough measurements
`
`electronically from the dose calibrator or manually via human data entry, the
`
`software ensures that the daily protocol is followed. Ex. 1014 at 17, 33 (supra).
`
`
`
`
`
`10
`
`
`
`

`

`IPR2018-1449
`
`Klein’s system does not deem strontium breakthrough tests to be “successful”
`
`merely when a calibration run completes. They are successful only when test results
`
`show low strontium breakthrough:
`
`Only after a calibration run with low Sr breakthrough has been
`successfully completed can patient elutions be carried out.
`
`Ex. 1014 at 18. Moreover, it is the computer’s “software”—not its operator—that
`
`causes Patient Elution runs to be “enabled only after the prerequisites [e.g., the low
`
`Sr breakthrough] have been completed successfully.” Id. at 44. The computer
`
`disables patient elutions each night and re-enables them only after the system
`
`confirms that breakthrough activity falls within Health Canada limits. Id. at 29, 33.
`
`Dr. Pelc’s opinion contradicts this clear disclosure, and it should be rejected.
`
`B.
`
`The ITC Rejected Bracco’s Expert Witness Testimony and
`Arguments.3
`Bracco relies heavily on Dr. Pelc’s ITC testimony involving patents with the
`
`same disclosure that recite the lockout function. POR at 2-3, 15-18. For example,
`
`U.S. Patent No. 9,814,826, which is at issue at the ITC, recites an analogous claim
`
`element:
`
`
`3 Petitioner cites the ALJ’s reasoning as persuasive authority, recognizing that the
`
`Board will assess the evidence for itself.
`
`
`
`
`
`11
`
`
`
`

`

`IPR2018-1449
`
`configuring a computer…to not allow a patient infusion if the strontium
`breakthrough test result is greater than or equal to an allowed limit.
`
`Ex. 1036, claim 1; Ex. 1035 at 2, 39-40.
`
`In the ITC, the ALJ heard the parties’ arguments regarding the lockout
`
`function, observed both experts testify, and ultimately rejected Bracco’s arguments.
`
`Ex. 1035 at 49-53. Overall, and on the lockout feature specifically, the ALJ found
`
`Petitioner’s expert, Dr. Stone, to be more credible than Dr. Pelc. See id. at 52-53
`
`(declining to adopt Pelc’s opinions regarding lockout), 48, 57, 79, 85, 96 (crediting
`
`Stone’s testimony).
`
`On the lockout feature, the ALJ recognized a basic fallacy in Bracco’s
`
`interpretation of Klein:
`
`Klein expressly states the system will only “enable patient elutions”
`with a successful breakthrough measurement. Klein at .000039
`(emphasis added); see also id. at .000028. If there were no system
`lock-out, as Bracco contends, then there would be no need to
`“enable” patient elutions. See also id. at .000029 (system disables
`patient elusion [sic] runs each night at midnight, when the validation
`from the calibration run expires).
`
`Id. at 51 (emphasis added). The ALJ rejected Bracco’s arguments—and Dr. Pelc’s
`
`opinions —because they drew illogical inferences from the evidence. Id. at 49-53.
`
`The Board should do the same here.
`
`
`
`
`
`12
`
`
`
`

`

`IPR2018-1449
`
`III. GROUND 3: BRACCO DISTRIBUTED THE BRACCO MANUAL TO
`ITS CUSTOMERS WITH NO CONFIDENTIALITY MARKINGS.
`The Bracco Manual (Ex. 1021) qualifies as a printed publication. Bracco
`
`admits that it distributed the printed manual to an interested segment of the public—
`
`users of its commercially-available CardioGen system. POR at 6, 19. These users
`
`are the very people that Bracco identifies as having ordinary skill in the art. POR at
`
`3-4 (clinical experience with PET imaging).
`
`Bracco’s admissions demonstrate the manual is a printed publication.
`
`Comcast Cable Comm’ns. v. Promptu Sys. Corp., IPR2018-00343, Paper 56, at 73
`
`(P.T.A.B. 2019) (“A court can appropriately treat statements in briefs as binding
`
`judicial admissions of fact.”) (quoting Purgess v. Sharrock, 33 F.3d 134, 144 (2d.
`
`Cir. 1994)); Manual of Patent Examining Procedure § 2145 (9th ed. Nov. 2015)
`
`(“Attorney argument is not evidence unless it is an admission, in which case, an
`
`examiner may use the admission in making a rejection.”); see Fed. R. Evid.
`
`801(d)(2). Bracco admits that it distributed Bracco Manuals to an interested segment
`
`of the public (i.e., users of Bracco’s CardioGen system). Bracco did so for a product
`
`that has been commercially available since 1989, roughly two decades before the
`
`earliest priority date claimed by the ’467 Patent. See Ex. 1002 ¶¶ 30-33; Garrett
`
`Corp. v. United States, 422 F.2d 874, 878 (Ct. Cl. 1970) (“distribution to commercial
`
`companies without restriction on use does [constitute publication]”). Moreover, the
`
`
`
`
`
`13
`
`
`
`

`

`IPR2018-1449
`
`manual (Ex. 1021) contains no confidentiality markings or other language restricting
`
`its use.
`
`Bracco argues that the manual does not qualify as a printed publication solely
`
`because “all such customers were under an obligation of confidentiality during the
`
`relevant time period.” POR at 19. Bracco, however, provided no evidence to
`
`substantiate this allegation.4 Its contention is mere attorney argument and should be
`
`disregarded. Icon Health & Fitness, Inc. v. Strava, Inc., 849 F.3d 1034, 1043 (Fed.
`
`Cir. 2017) (vacating PTAB decision relying on attorney argument that was
`
`unsupported by evidence) (citing Gemtron Corp. v. Saint-Gobain Corp., 572 F.3d
`
`1371, 1380 (Fed. Cir. 2009) (“[U]nsworn attorney argument...is not evidence and
`
`cannot rebut...other admitted evidence....”)).
`
`Bracco’s reliance on Cordis (POR at 58) is misplaced. There, evidence
`
`demonstrated that an inventor requested—and the receiving parties maintained—
`
`confidentiality of a disputed monograph. Cordis Corp. v. Boston Scientific Corp.,
`
`561 F.3d 1319, 1334 (Fed. Cir. 2009). Here, Bracco “points to no evidence that [it]
`
`
`4In its Institution Decision, the Board recognized that Bracco “likely possess[es]
`
`information relevant to public availability of the Bracco manual” which might
`
`support its contention. Paper 7 (Dec. to Inst.) at 17. Significantly, Bracco does not
`
`dispute this point in its Patent Owner Response.
`
`
`
`
`
`14
`
`
`
`

`

`IPR2018-1449
`
`ever distributed the [Bracco Manual] with an expectation that it would be kept
`
`confidential or not disseminated.” Nobel Biocare Services AG v. Instradent USA,
`
`Inc., 903 F.3d 1365, 1377 (Fed. Cir. 2018) (citing Cordis, 561 F.3d at 1333-34)
`
`(disregarding attorney arguments of confidentiality). Furthermore, the Bracco
`
`Manual (Ex. 1021) “bears no designations, such as ‘draft’ or ‘confidential,’ that
`
`might suggest that it was not intended for public distribution.” 903 F.3d at 1377.
`
`Having admitted to distributing the manual, Bracco confirms the Bracco Manual
`
`qualifies as a printed publication.
`
`IV. BRACCO’S ATTACKS ON DR. MURTHY’S TESTIMONY ARE
`UNJUSTIFIED DISTRACTIONS FROM THE REMAINING ISSUES
`IN DISPUTE.
`Bracco’s attacks on Dr. Murthy’s declaration are immaterial to the issues
`
`remaining in this IPR. Bracco disputes only that Klein discloses a computer lockout
`
`function and that the Bracco Manual is a printed publication. POR §§ IV, V. Bracco
`
`waived any other patentability argument. Paper 8 (Sched. Order) at 6. Neither the
`
`Petition nor Dr. Stone’s declaration rely on Dr. Murthy’s declaration for these issues.
`
`With respect to the lockout function, the Petition and Dr. Stone’s declaration
`
`both refer to Klein—not to Dr. Murthy’s declaration—to demonstrate how Klein’s
`
`disclosure matches claims 1 and 13. See Pet. at 32-33, 47-48 (discussing Ex. 1014);
`
`Ex. 1015 §§ IX.B.5, XI.A.7, XI.J.5 (same). The Petition and Dr. Stone’s declaration
`
`
`
`
`
`15
`
`
`
`

`

`IPR2018-1449
`
`do not rely on Dr. Murthy’s declaration to analyze the Bracco Manual. See Pet. at
`
`64-67; Ex. 1015 §§ IX.D, XIII. Thus, no matter whether the Board affords Dr.
`
`Murthy’s declaration substantial weight, it does not affect the issues that the Board
`
`must decide.
`
`Dr. Murthy’s declaration provides context for Dr. Stone’s declaration.
`
`Although Dr. Stone is an engineer with substantial experience both working with
`
`radioactive control systems and designing medical devices (Ex. 1015 ¶¶ 2-10), he is
`
`not a medical doctor. He is not an expert on topics involving medical uses of
`
`rubidium-82. Thus, Dr. Stone looks to Dr. Murthy, a board-certified physician with
`
`a decade of experience working with rubidium infusion systems (Ex. 1017 ¶¶ 2-6),
`
`for an explanation of these topics. Id. ¶¶ 79-95, 122-125. There is nothing improper
`
`in doing so.
`
`Bracco criticizes Dr. Murthy’s declaration for two reasons: 1) a clerical error,
`
`his declaration refers to a 2007 version of Bracco’s Manual (Ex. 1037) rather than
`
`the 2004 version that became Ex. 1021; and 2) the timing of his experience with
`
`Bracco’s CardioGen system. As discussed, neither challenge is material to the
`
`Grounds in the Petition.
`
`On the first issue, the 2004 and the 2007 versions of the Bracco Manual are
`
`identical in all material respects. See Exs. 1021, 1037. Bracco offers no evidence
`
`otherwise.
`
`
`
`
`
`16
`
`
`
`

`

`IPR2018-1449
`
`On the second issue, Dr. Murthy testified that he trained on the CardioGen
`
`system in July 2008, a mere month after the June 2008 date to which Bracco’s patent
`
`claims priority. Ex. 1017 ¶ 5. As part of that training, Dr. Murthy reviewed journal
`
`articles and product documentation describing rubidium infusions systems. Id.; see
`
`also Ex. 2016 at 30:13-31:12 (testifying that his review of literature relating to the
`
`CardioGen system also included review of images, clinical records, and infusion
`
`records on many scans dating back to 2004). This experience provides a solid basis
`
`on which to describe prevailing practices in 2008 for rubidium-82 elution systems.
`
`Bracco does not challenge the substance of Dr. Murthy’s declaration. Bracco
`
`provides no evidence to suggest that his declaration is unreliable in any way. Most
`
`importantly, Bracco identified no issue remaining in dispute to which Dr. Murthy’s
`
`declaration is material. Bracco’s criticism of Dr. Murthy’s declaration is a red
`
`herring.
`
`
`
`
`
`17
`
`
`
`

`

`V. CONCLUSION
`Claims 1-4, 6-16, and 18-22 from the ’467 Patent should be canceled.
`
`IPR2018-1449
`
`
`
`
`
`
`
`
`
`
`
`Date: August 22, 2019
`
`
`
`
`
`
`
`Respectfully submitted,
`
`
`
`By:
`
`/Robert L. Hails/
`Robert L. Hails, Reg. No. 39,702
`T. Cy Walker, Reg. No. 52,337
`Theresa Weisenberger, Reg. No. 65,559
`BAKER & HOSTETLER LLP
`Washington Square, Suite 1100
`1050 Connecticut Avenue, N.W.
`Washington, D.C. 20036-5304
`Telephone: 202-861-1500
`Facsimile: 202-861-1783
`Jubilant-Baker@bakerlaw.com
`
`Attorneys for Petitioner
`JUBILANT DRAXIMAGE INC.
`
`
`
`
`
`
`18
`
`
`
`
`
`
`
`
`
`
`
`
`

`

`IPR2018-1449
`
`CERTIFICATION
`The undersigned certifies that the foregoing Petitioner’s Reply Brief,
`
`exclusive of the Table of Contents, Table of Authorities, and Mandatory Notices,
`
`has a word count of less than 5,600 words as computed by Microsoft Word.
`
`
`
`
`
`
`
`By:
`
`/Robert L. Hails/
`Robert L. Hails, Reg. No. 39,702
`
`
`
`19
`
`
`
`
`
`

`

`IPR2018-1449
`
`CERTIFICATE OF SERVICE
`The undersigned hereby certifies that on August 22, 2019, a copy of this
`
`Petitioner’s Reply Brief and all supporting exhibits have been served in their entirety
`
`via electronic mail by emailing Patent Owner’s counsel at:
`
`Barry J. Schindler
`SchindlerB@gtlaw.com
`Heath J. Briggs
`BriggsH@gtlaw.com
`Vimal Kapadia
`KapadiaV@gtlaw.com
`Stephen Ullmer
`UllmerS@gtlaw.com
`Lennie Bersch
`BerschL@gtlaw.com
`bracco-iprs@gtlaw.com
`
`as provided for by Patent Owner’s Service Information in Patent Owner’s
`
`Mandatory Disclosures.
`
`
`
`
`
`
`
`By:
`
`/Robert L. Hails/
`Robert L. Hails, Reg. No. 39,702
`
`20
`
`
`
`
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket