throbber
BRITISH STANDARD
`
`BSEN
`62274:2005
`
`Medical electrical
`equipinent- Safety of
`radiotherapy record
`and verify systeins
`
`0 '
`
`' TI
`0 a.
`C
`
`The European Standard EN 62274:2005 has the status of a
`British Standard
`
`res 11.040.60
`
`NO COPYING WITHOUT BSI PERMISSION EXCEPT AS PERMITTED BY COPYRIGHT LAW
`
`British Standards
`
`JUBILANT EXHIBIT 1029
`Jubilant v. Bracco, IPR2018-01449
`
`

`

`National foreword
`
`This British Standard is the official English language version of
`EN 62274:2005. It is identical with IEC 62274:2005.
`The UK participation in its preparation was entrusted by Technical Committee
`CH/62, Electromedical equipment in medical practice, to Subcommittee
`CH/62/3, Equipment for radiotherapy, nuclear medicine and radiation
`dosimetry, which has the responsibility to:
`
`aid enquirers to understand the text;
`
`0
`7
`
`present to the responsible international/European committee any
`enquiries on the interpretation, or proposals for change, and keep UK
`interests informed;
`
`monitor related international and European developments and
`promulgate them in the UK
`
`A list of organizations represented on this subcommittee can be obtained on
`request to its secretary.
`Cross-references
`The British Standards which implement international or European
`publications referred to in this document may be found in the BSI Catalogue
`under the section entitled "International Standards Correspondence Index", or
`by using the "Search" facility of the BSI Electronic Catalogue or of British
`Standards Online.
`This publication does not purport to include all the necessary provisions of a
`contract. Users are responsible for its correct application.
`Compliance with a British Standard does not of itself confer immunity
`from legal obligations.
`
`Summary of pages
`This document comprises a front cover, an inside front cover, the EN title page,
`pages 2 to 18, an inside back cover and a back cover.
`The BSI copyright notice displayed in this document indicates when the
`document was last issued.
`
`Amendments issued since publication
`
`Amd. No.
`
`Date
`
`Comments
`
`BS EN 6227 4:2005
`
`This British Standard was
`published under the authority
`of the Standards Policy and
`Strategy Committee
`on 28 December 2005
`
`© BSI 28 December 2005
`
`ISBN O 580 47175 6
`
`

`

`EUROPEAN STANDARD
`
`NORME EUROPEENNE
`
`EN 62274
`
`EUROPAISCHE NORM
`
`June 2005
`
`ICS 11.040.60
`
`English version
`
`Medical electrical equipment-
`Safety of radiotherapy record and verify systems
`(IEC 62274:2005)
`
`Appareils electromedicaux -
`Securite des systemes d'enregistrement
`et de verification de radiotherapie
`(CEI 62274:2005)
`
`Medizinische elektrische Gerate -
`Sicherheit van Aufzeichnungs(cid:173)
`und Verifikationssystemen
`fur die Strahlentherapie
`(IEC 62274:2005)
`
`I
`
`0 '
`
`z
`0
`
`This European Standard was approved by CENELEC on 2005-06-01. CENELEC members are bound to
`comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
`Standard the status of a national standard without any alteration.
`
`Up-to-date lists and bibliographical references concerning such national standards may be obtained on
`application to the Central Secretariat or to any CENELEC member.
`
`This European Standard exists in three official versions (English, French, German). A version in any other
`language made by translation under the responsibility of a CENELEC member into its own language and
`notified to the Central Secretariat has the same status as the official versions.
`
`CENELEC members are the national electrotechnical committees of Austria, Belgium, Cyprus, Czech
`Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
`Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Slovakia, Slovenia, Spain, Sweden,
`Switzerland and United Kingdom.
`
`CENELEC
`
`European Committee for Electrotechnical Standardization
`Comite Europeen de Normalisation Electrotechnique
`Europaisches Komitee fOr Elektrotechnische Normung
`
`Central Secretariat: rue de Stassart 35, B - 1050 Brussels
`
`© 2005 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
`
`Ref. No. EN 62274:2005 E
`
`

`

`EN 6227 4:2005
`
`- 2 -
`
`Foreword
`
`The text of document 62C/381/FDIS, future edition 1 of IEC 62274, prepared by SC 62C "Equipment
`for radiotherapy, nuclear medicine and radiation dosimetry" des IEC TC 62 "Electrical equipment in
`medical practice", was submitted to the IEC-CENELEC parallel vote and was approved by CENELEC
`as EN 6227 4 on 2005-06-01.
`
`The following dates were fixed:
`
`-
`
`latest date by which the EN has to be implemented
`at national level by publication of an identical
`national standard or by endorsement
`
`(dop)
`
`2006-03-01
`
`latest date by which the national standards conflicting
`with the EN have to be withdrawn
`
`(dow)
`
`2008-06-01
`
`In this standard, the following print types are used:
`
`0 .,
`
`z
`0
`
`requirements proper: roman type;
`
`test specifications: italic type;
`
`notes and explanatory matter: small roman type;
`
`TERMS USED THROUGHOUT THIS PARTICULAR STANDARD THAT ARE DEFINED IN CLAUSE 3, OR IN OTHER
`STANDARDS: SMALL CAPITALS.
`
`-
`
`-
`
`Annex ZA has been added by CENELEC.
`
`Endorsement notice
`
`The text of the International Standard IEC 62274:2005 was approved by CENELEC as a European
`Standard without any modification.
`
`In the official version, for Bibliography, the following notes have to be added for the standards indicated:
`
`IEC 60601-1-6
`
`NOTE Harmonized as EN 60601-1-6:2004 (not modified).
`
`!EC 60601-2-11
`
`NOTE
`
`Harmonized as EN 60601-2-11:1997 (not modified).
`
`IEC 60601-2-17
`
`NOTE
`
`Harmonized as EN 60601-2-17:2004 (not modified).
`
`!EC 62083
`
`NOTE
`
`Harmonized as EN 62083:2001 (not modified}.
`
`

`

`EN 6227 4:2005
`
`t
`
`- 3 -
`
`CONTENTS
`
`INTRODUCTION ................................................................................................................... .4
`1
`Scope and object. ........................................................................................................... 5
`
`2
`
`3
`
`4
`
`5
`6
`
`7
`8
`
`9
`
`Normative references ..................................................................................................... 6
`
`Terms and definitions ..................................................................................................... 7
`
`General requirements for tests ........................................................................................ 7
`
`0
`7
`
`ACCOMPANYING DOCUMENTS ............................................................................................. 7
`Requirements for safety .................................................................................................. 8
`
`TREATMENT machine set-up verification ......................................................................... 11
`
`TREATMENT recording and reporting ............................................................................... 12
`
`Accuracy ...................................................................................................................... 12
`
`10 Abnormal operation and fault conditions ........................................................................ 12
`11
`Human errors in software design ................................................................................... 12
`
`12 Change in software versions ......................................................................................... 13
`
`13 Human errors in use ..................................................................................................... 13
`
`Bibliography ......................................................................................................................... 16
`
`Annex (normative) Hardware safety .................................................................................... 14
`
`Annex ZA (normative) Normative references to international publications with their
`corresponding European publications .................................................................................... 18
`Table 1 - Clauses and subclauses in this standard that require the provision of
`information in the ACCOMPANYING DOCUMENTS and the technical description .............................. 8
`
`

`

`EN 6227 4:2005
`
`-4-
`
`INTRODUCTION
`
`A RADIOTHERAPY RECORD AND VERIFY SYSTEM (RVS) is a PEMS (PROGRAMMABLE ELECTRICAL
`MEDICAL SYSTEM) or a subsystem that is used to help prevent erroneous set-up of a medical
`ELECTRON ACCELERATOR, GAMMA BEAM THERAPY EQUIPMENT, or other RADIOTHERAPY TREATMENT
`machine and to record all TREATMENT sessions. This is accomplished through verification of
`the set-up and preventing machine operation if the set-up does not match predetermined
`settings. Inaccuracies in the data or errors in the record and verify process may represent
`SAFETY HAZARDS to PATIENTS. This standard defines requirements to be complied with by
`MANUFACTURERS in the design and construction of an RVS
`in order to provide protection
`against the occurrence of such hazards.
`
`n
`0
`0 cc
`7
`
`n
`n:,
`0
`
`0
`7
`
`

`

`- 5 -
`
`EN 6227 4:2005
`
`MEDICAL ELECTRICAL EQUIPMENT -
`SAFETY OF RADIOTHERAPY RECORD
`AND VERIFY SYSTEMS
`
`1 Scope and object
`
`1.1 Scope
`
`This International Standard applies to the design and manufacture and some installation
`aspects of a RECORD AND VERIFY SYSTEM (RVS) for use in RADIOTHERAPY in human medical
`practice that
`
`a) provides, defines, or displays treatment machine set up data; imports data either through
`input by the OPERATOR or directly from other devices,
`b) may control the ability of that equipment to operate;
`c)
`records data of all TREATMENT sessions; and
`d)
`is intended to be:
`1)
`for NORMAL USE, under the authority of appropriately licensed or QUALIFIED PERSONS, by
`OPERATORS having the required skills and training;
`2) maintained in accordance with the recommendations given in the INSTRUCTIONS FOR
`USE; and
`3) used within the environmental and electrical supply conditions specified
`technical description.
`
`the
`
`in
`
`This standard does not address dynamic beam deliveries.
`
`NOTE This may be addressed in a future version of this standard.
`
`It does, however, address certain specific aspects of the connection of the RVS to a network
`or to other radiotherapy equipment, and the communication protocol to be used.
`
`1.2 Object
`
`This standard applies to any RVS and establishes the requirements for features, associated
`documentation, and testing of the software. Requirements for SAFETY of hardware are not
`included in this standard, as these vary with the nature of the hardware (see 1.3.1 and
`Annex A concerning hardware requirements).
`
`An RVS developed by a USER exclusively for the USER'S own use is not within the scope of this
`standard, but it is highly recommended that the principles of this standard be applied in its
`creation and use. If this technology is provided to another USER under conditions in which the
`developers do not directly control its use, and is represented as being suitable for use, then
`the developers assume the role of MANUFACTURER and this standard applies.
`
`

`

`z
`0
`
`0
`0
`
`EN 6227 4:2005
`
`- 6 -
`
`1.3 Relationship to other standards
`
`1.3.1
`
`Hardware SAFETY standards
`
`Requirements for SAFETY of hardware, such as for protection against electric shock, fire, and
`for ELECTROMAGNETIC COMPATIBILITY, are not included in this document. SAFETY requires that
`these subjects be addressed by the MANUFACTURER separately through compliance with an
`appropriate standard, depending upon the nature and environment of the hardware used for
`the RVS (see Annex A for hardware SAFETY standards).
`
`1.3.2
`
`Software SAFETY standards
`
`All clauses and subclauses of the collateral standard IEC 60601-1-4 (see 4.1) apply.
`
`In applying IEC 60601-1-4, the MANUFACTURER shall consider hazards that could occur due to
`use errors associated with the RVS.
`
`NOTE
`
`IEC 60601-1-6 describes a process for the risk management of use errors.
`
`1.3.3
`
`IEC 61217
`
`IEC 61217, Radiotherapy equipment - Co-ordinates, movements and scales, applies. The
`means of applying IEC 61217 are specified in appropriate subclauses of this standard.
`
`1.3.4 Other standards
`
`Because an RVS is likely to contain PATIENT-related information and other medical data,
`standards related to confidentiality of PATIENT information and security of electronic records
`may apply. Where there is an apparent conflict between any of these standards and a
`requirement of this standard, the MANUFACTURER shall resolve the conflict in a way that best
`meets the intent of each standard, and shall explain the implementation in the ACCOMPANYING
`DOCUMENTS.
`
`2 Normative references
`
`The following referenced documents are indispensable for the application of this document.
`For dated references, only the edition cited applies. For undated references, the latest edition
`of the referenced document (including any amendments) applies.
`
`IEC 60601-1, Medical electrical equipment - Part 1: General requirements for safety
`
`I EC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for safety -
`Collateral standard: Electromagnetic compatibility - Requirements and tests
`
`IEC 60601-1-4: 1996, Medical electrical equipment - Part 1-4: General requirements for safety
`- Collateral Standard: Programmable electrical medical systems
`Amendment 1 (1999)1)
`
`IEC 60601-2-29, Medical electrical equipment - Part 2-29: Particular requirements for the
`safety of radiotherapy simulators
`
`IEC TR 60788, Medical electrical equipment - Glossary of defined terms
`
`1) A consolidated edition 1.1 exists including IEC 60601-1-4:1996 and its Amendment 1 (1999).
`
`

`

`- 7
`
`EN 6227 4:2005
`
`IEC 60950-1, Information technology equipment- Safety- Part 1: General requirements
`
`IEC 61000 (all parts), Electromagnetic compatibility (EMC)
`
`IEC 61217, Radiotherapy equipment - Co-ordinates, movements and scales
`
`3 Terms and definitions
`
`For the purpose of this document, the terms and definitions given in IEC 60601-1, IEC 60601-1-2,
`IEC 60601-1-4, IEC 60601-2-29 and IEC 60788 and the following apply.
`
`0
`7
`
`3.1
`RECORD AND VERIFY SYSTEM
`RVS
`PROGRAMMABLE ELECTRICAL MEDICAL SYSTEM or subsystem including its associated peripherals,
`that is used to compare the set-up of a RADIOTHERAPY TREATMENT machine to predetermined
`set-up conditions prior to the start of a proposed RADIOTHERAPY TREATMENT and each
`TREATMENT session, and record actual TREATMENT sessions. It also provides a means of
`preventing the machine operation if the actual set-up is not the same as the pre-set intended
`set-up, within USER defined tolerances.
`
`4 General requirements for tests
`
`4.1 Testing during development
`
`Compliance with IEC 60601-1-4 requires identification of HAZARDS, assessment of their RISKS,
`and appropriate verification and validation of RISK controls. Demonstration of compliance with
`the requirements of this standard shall be included as part of the above processes, with
`explicit reference to each requirement. The MANUFACTURER shall retain compliance data as a
`permanent record. Each test shall include a protocol containing all the necessary input data,
`sufficient detail to provide for exact reproducibility, and the expected result.
`
`4.2 Testing during installation
`
`The MANUFACTURER shall provide an installation test document as part of the technical
`description that includes a demonstration that the RVS performs according to the operational
`description provided in the ACCOMPANYING DOCUMENTS as required in Clause 5.
`
`5 ACCOMPANYING DOCUMENTS
`
`The technical description and the INSTRUCTIONS FOR USE shall contain the information as
`required by this standard (see Table 1 for references).
`
`

`

`EN 6227 4:2005
`
`- 8 -
`
`Table 1 - Clauses and subclauses in this standard that require the provision of
`information in the ACCOMPANYING DOCUMENTS and the technical description
`
`Check reference
`
`Instructions for use
`
`Technical description
`
`6.5
`
`6.9
`
`6.9
`
`7.2
`
`7.3
`
`9
`
`10.1
`
`z
`0
`
`1
`
`2
`
`3
`
`4
`
`5
`
`6
`
`7
`
`8
`
`9
`
`10
`
`11
`
`12
`
`13
`
`14
`
`15
`
`16
`
`17
`
`18
`
`19
`
`20
`
`21
`
`22
`
`23
`
`6.3
`
`6.4 a)
`
`6.4 b)
`
`6.6
`
`6.7
`
`6.8
`
`7.4
`
`10.2
`
`11
`
`12 a)
`
`12 b)
`
`12 d)
`
`13
`
`A.1.4
`
`A.3
`
`NOTE The check reference is given as an aid for checking the availability of
`compliance documentation.
`
`6 Requirements for safety
`
`6.1 RADIATION quantities
`
`All values of RADIATION quantities requested, displayed or printed shall include their units.
`Units of RADIATION should conform to the SI convention. Units (e .g. "monitor units" (MU))
`describing dose delivery shall be consistent with those used by the TREATMENT machine.
`
`Compliance is checked by inspection of the DISPLAY and output information.
`
`6.2 Date and time
`
`When the date is displayed or printed, correct interpretation shall not depend upon the
`OPERATOR'S interpretation of format, and a DISPLAY of the year shall be in four digits.
`
`

`

`- 9 -
`
`EN 6227 4:2005
`
`Examples (acceptable): '03 Apr 2005', '03/04/2005 (dd/mm/yyyy)', '2005/04/03 (yyyy/mm/dd)'
`Examples (not acceptable): '03/04/05', 03/04/2005, '03 Apr 05'.
`
`When the time is requested, displayed or printed, it shall be represented on a 24-hour clock
`basis, or if a 12-hour clock is used it shall be unambiguously indicated whether it is a.m or
`p.m. Measurements of time shall include units (hr, min, sec.).
`
`When time is entered, displayed or printed, each denomination of time shall be labelled. To
`prevent confusion with numbers, single-letter abbreviations of time denomination shall not be
`used (e.g. h,m,s). Acceptable examples: 2,05 min; 1 hour 33 minutes; 1 :43:15 (hr:min:sec).
`
`0
`7
`
`Time-sensitive functions shall be performed correctly at transitions such as year boundaries,
`leap years, etc.
`
`Compliance is checked by testing and by inspection of the DISPLAY and output information.
`
`6.3 Coordinate systems and scales
`
`It shall be possible for the OPERATOR to perform all RVS functions with the scales and
`coordinates of RADIOTHERAPY TREATMENT EQUIPMENT displayed according to the IEC 61217
`convention. If, in addition, any convention other than IEC 61217 is employed for scales and
`coordinates, the conventions shall be identified. The units shall be the same as are used in
`the RADIOTHERAPY TREATMENT EQUIPMENT.
`
`The method and format of DISPLAY of scales shall be explained in the INSTRUCTIONS FOR USE.
`
`Compliance is checked by testing and by inspection of the DISPLAY, output information and
`ACCOMPANYING DOCUMENTS.
`
`6.4 Protection against unauthorized use
`
`Means shall be provided to prevent unauthorised changes. Where changes to the data are
`permitted by authorised persons, means shall be provided to prevent a person making
`changes he/she is not authorised to make.
`NOTE Password protection is a normal way of implementing this requirement. If password protection is used, then
`there should be means for setting a person's authority or individual authorisation for specific functions.
`Compliance is checked by testing, and by inspection of the ACCOMPANYING DOCUMENTS and
`INSTRUCTIONS FOR USE.
`
`Where network connection is permitted by the design, the following requirements apply.
`a) Access to the RVS shall be provided only to EQUIPMENT or individuals who are authorized
`(for example, by a password under the control of the USER).
`b) Access to TREATMENT prescriptions and other data containing the PATIENT identification
`information through the network shall be restricted to prevent unauthorized access.
`c) The MANUFACTURER shall recommend some means of virus protection in the INSTRUCTIONS
`FOR USE.
`Compliance is checked by testing and by inspection of the ACCOMPANYING DOCUMENTS.
`
`

`

`EN 6227 4:2005
`
`- 10 -
`
`6.5 Correctness of data transfer
`
`The MANUFACTURER shall identify RVS data transfer protocols in the technical description.
`
`Data transferred from and to the TREATMENT machines or other devices, excluding hardcopy
`devices, shall include a protection against data transmission errors. Means shall be provided
`to warn the OPERATOR that expected data have not been transferred.
`
`Examples: DICOM 3 or FTP, each of which includes error detection, or a proprietary format
`that includes a checksum of each input or output data set.
`
`Compliance is checked by inspection of the protocol specifications and by inspection of the
`ACCOMPANYING DOCUMENTS.
`
`6.6 Data acceptance
`
`Means shall be provided such that the TREATMENT machine set-up data and other patient
`TREATMENT data shall be available
`for TREATMENT use only after the OPERATOR has
`acknowledged that they have been reviewed for correctness and completeness.
`
`Compliance is checked by testing and by inspection of ACCOMPANYING DOCUMENTS.
`
`Where design allows, machine set-up data and other patient TREATMENT data shall be
`reviewed or approved by entry of an authorised identification:
`
`a) any modification to the data shall result in invalidation of the authorised identification;
`b) after modification of the approved data a new authorised identification shall be required;
`c)
`the RVS shall provide a means for preserving the history and the record of the authorised
`identification; and
`d) the INSTRUCTIONS FOR USE shall describe how these features are to be properly and safely
`used.
`
`Compliance is checked by testing and by inspection of ACCOMPANYING DOCUMENTS.
`
`6.7 Deleting and editing data
`
`Means shall be provided to restrict the ability to edit TREATMENT history data to persons who
`are authorised to carry out this function. A record of the change details shall be retained. The
`fact that the TREATMENT history has been modified shall be apparent to a person using it, e.g.
`by a visual indicator.
`
`Compliance is checked by testing.
`
`6.8 Backing up data
`
`Means shall be provided for backing-up data onto a separate medium from the primary
`storage, such that it can be restored in the case of a failure of the primary data storage device.
`
`NOTE Usually a backup provides a means to restore data in the case of system failure.
`
`Compliance is checked by testing and by inspection of ACCOMPANYING DOCUMENTS.
`
`0
`7
`
`0
`0
`
`~
`
`7
`rr
`0
`~
`rr
`
`

`

`0
`7
`
`0
`0
`
`C
`
`i!,
`D
`
`0
`~
`C
`C
`
`0
`0
`
`- 1 1 -
`
`EN 6227 4:2005
`
`6.9 Archiving data
`
`Means shall be provided for archiving sets of data for long term storage, such that the data
`can be accessed at a later date.
`
`NOTE Archiving is the process of moving or copying sets of data from the primary storage to a separate storage
`media. Standardising the archiving process is highly desirable. By using established standards such as DICOM or
`HL7, archiving would be vendor and media independent.
`
`Compliance is checked by testing and by inspection of ACCOMPANYING DOCUMENTS.
`
`7 TREATMENT machine set-up verification
`
`7 .1 Prevention of TREATMENT
`
`The RVS shall provide a means by which the operation of the TREATMENT machine shall be
`prevented in the event that the machine set-up does not correspond to the prescribed data
`within prescribed tolerances.
`
`Compliance is checked by testing.
`
`7 .2 Override
`
`If an override capability is provided, the USER shall:
`
`-
`-
`
`acknowledge the override parameters;
`provide authorised identification.
`
`The fact that an override has been made shall be recorded.
`
`Compliance is checked by testing
`
`7 .3 Transfer of prescribed TREATMENT data
`
`If the design of the RVS allows transfer of the prescribed TREATMENT parameters to the
`TREATMENT machine, the RVS shall prompt the OPERATOR to confirm that the transferred data
`are correct prior to the first use, and after any modification.
`
`NOTE This confirmation may be through not performing the download for the first TREATMENT session after entry
`or modification, so that the OPERATOR is required to set up the machine from a separate source of information. For
`TREATMENT session where this is not possible, confirmation may be implemented by requiring that the OPERATOR
`confirm, prior to the RVS allowing the TREATMENT to proceed, that a new or modified set-up has been verified by
`other means.
`
`Compliance is checked by inspection of the protocol specifications, and by inspection of the
`ACCOMPANY/NG DOCUMENTS.
`
`7 .4 Accompanying information
`
`The ACCOMPANYING DOCUMENTS shall include a caution to the OPERATOR that the correct
`operation of the RVS depends on the connection of the RVS to the radiotherapy system,
`which may
`include
`the RADIOTHERAPY TREATMENT PLANNING SYSTEM
`(RTPS) and
`the
`RADIOTHERAPY SIMULATOR, as well as the TREATMENT machine. Any change to the radiotherapy
`system to which the RVS is connected shall require testing to confirm correct operation.
`
`Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
`
`

`

`EN 6227 4:2005
`
`- 12 -
`
`8 TREATMENT recording and reporting
`
`For each patient the RVS shall provide a means by which the OPERATOR can retrieve and
`report all recorded TREATMENT machine parameters used in the previous TREATMENT sessions.
`
`NOTE The guides and contents of the record and report may be found in ICRU publications 50 and 62 for photon
`beam therapy and 58 for BRACHYTHERAPY.
`
`Compliance is checked by testing.
`
`9 Accuracy
`

`
`0
`7
`
`The MANUFACTURER shall state the accuracy of the RVS for all of the TREATMENT parameters
`recorded.
`
`Compliance is checked by testing as described in the ACCOMPANYING DOCUMENTS.
`
`10 Abnormal operation and fault conditions
`
`10.1 General hardware diagnostics
`
`Facilities shall be provided so that the correct function of the means referred to in 7 .1 and the
`communication between RVS and TREATMENT machine can be checked. The ACCOMPANYING
`DOCUMENTS shall contain an explanation of the function of these facilities.
`
`If these facilities require actions of the OPERATOR, such actions shall be described in the
`ACCOMPANYING DOCUMENTS together with a recommendation of the frequencies with which
`such checks should be made.
`
`Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
`
`10.2 Data and code
`
`Executable program code, TREATMENT machine set-up data and other patient TREATMENT data
`shall have a checksum or other equivalent protection that ensures that they will not be used if
`modified through a hardware fault, through a virus, accidentally during servicing, or by any
`other unauthorized manner. In the event that an error is detected the MANUFACTURER shall
`provide instructions to the OPERATOR for restoring correct operation, either on the DISPLAY or
`in the INSTRUCTIONS FOR USE.
`
`Compliance is checked by inspection of the DISPLAY or the ACCOMPANYING DOCUMENTS.
`
`11 Human errors in software design
`
`The requirements for software development process and RISK management that are defined in
`IEC 60601-1-4 shall apply. In applying the requirements of IEC 60601-1-4 the term PEMS
`shall include an RVS.
`
`is checked by testing and by examining system documentation
`Compliance
`requirements of /EC 60601-1-4.
`
`to
`
`the
`
`

`

`0
`7
`
`ci
`

`
`- 13 -
`
`EN 62274:2005
`
`The MANUFACTURER shall describe, in the INSTRUCTIONS FOR USE, a means by which the USER
`can report errors in the operation that are observed during use or testing.
`
`Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
`
`12 Change in software versions
`
`The following requirements apply when a new version of the software is provided to the USER
`by the MANUFACTURER.
`
`a)
`
`b)
`
`If the installation is not performed by the MANUFACTURER, the INSTRUCTIONS FOR USE shall
`provide clear instructions for the installation of a new version, and for any tests that are
`required to determine that the installation was successful.
`The INSTRUCTIONS FOR USE . shall advise the USER to maintain appropriate records of
`software updates. The MANUFACTURER shall keep records of software updates provided to
`or installed by each USER.
`If use of data from the previous version could cause incorrect results, either:
`1)
`the design shall convert the data to the new format; or
`2)
`the design shall prevent use of the data.
`If installation of a new version of the software may delete machine set-up or other patient
`TREATMENT data, the OPERATOR shall be warned and required to backup the data before
`proceeding with installation of the new version.
`If data created on the currently installed version of the software are to be used on the next
`successive version of software, the INSTRUCTION FOR USE shall provide instructions on how
`to restore data created with the current version onto the next version, unless these
`services are performed solely by the MANUFACTURER.
`e) The INSTRUCTIONS FOR USE shall provide instructions on how to restore the RVS to the
`condition it was in prior to the installation of a new version of software.
`
`c)
`
`d)
`
`Compliance is checked by testing (b and c), and by inspection of the ACCOMPANYING
`DOCUMENTS (a, b, C, d, e).
`
`13 Human errors in use
`
`The INSTRUCTIONS FOR USE shall provide comprehensive instructions to the USER of all
`information needed for safe operation, including, but not limited to, the specific information in
`other clauses and subclauses of this standard.
`
`The INSTRUCTIONS FOR USE shall provide a cautionary notice to the USER that the USER should
`ensure that individuals authorized to enter and accept machine set-up or other patient
`TREATMENT data are appropriately trained for the functions they perform.
`
`Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
`
`

`

`EN 62274:2005
`
`- 14 -
`
`Annex A
`(normative)
`
`Hardware safety
`
`A.1
`
`Hardware safety standards
`
`A.1.1
`
`General requirements
`
`This standard is concerned principally with operational features and other aspects of RVS
`software required for safe operation. It shall be supplemented by, or be supplemental to , an
`appropriate hardware SAFETY standard,
`to which
`the MANUFACTURER shall additionally
`demonstrate compliance . Below are general descriptions of some of the available standards,
`and some comments about their applicability. The list is not meant to be comprehensive, and
`it is the responsibility of the MANUFACTURER to identify and select appropriate standards,
`including
`their most recent revisions and amendments . The MANUFACTURER may use
`standards other than those listed where analysis shows them to be also appropriate.
`
`The MANUFACTURER shall state in the technical description all hardware SAFETY standards with
`which the RVS complies .
`
`A.1.2
`
`IEC 60950-1: SAFETY of INFORMATION TECHNOLOGY EQUIPMENT
`
`This standard applies to a range of INFORMATION TECHNOLOGY EQUIPMENT as identified in its
`Subclause 1.1.1, EQUIPMENT covered by this standard. If an RVS uses general-purpose
`computer hardware and peripherals, and is not used with direct connections to a PATIENT, then
`IEC 60950-1 is a pertinent standard.
`
`A.1.3
`
`IEC 60601-1: MEDICAL ELECTRICAL EQUIPMENT - Part 1: General requirements
`for SAFETY
`
`IEC 60601-1 is the General Standard for the Safety of medical electrical equipment. If the
`RVS hardware is used in the presence of patients , or is integrated with hardware used in the
`presence of patients , then IEC 60601-1 may be an appropriate standard to use for hardware
`safety considerations .
`
`A.1.4
`
`IEC 61000 series of standards:
`Requirements for ELECTROMAGNETIC COMPATIBILITY (EMC) testing
`
`This is a family of generic standards containing general requirements for ELECTROMAGNETIC
`COMPATIBILITY e.g. immunity, emissions, environment, testing and measurement.
`
`A.1.5
`
`IEC 60601-1-2 (collateral standard to 60601-1):
`ELECTROMAGNETIC COMPATIBILITY - Requirements and tests
`
`This standard address ELECTROMAGNETIC COMPATIBILITY test requirements and/or methods for
`INFORMATION TECHNOLOGY EQUIPMENT and for MEDICAL ELECTRICAL EQUIPMENT. Applicability will
`depend on the nature of the hardware and the environment in which it is to be used . In most
`cases, an RVS will use general-purpose commercial computer hardware for use in an
`environment appropriate to such EQUIPMENT, and the computer MANUFACTURER may have
`
`0 -,
`
`z
`0
`
`r
`r
`
`"
`
`

`

`- 15 -
`
`EN 6227 4:2005
`
`certified the EQUIPMENT to one of these standards. If the computer is of custom construction,
`or is integrated with EQUIPMENT that has connections to a PATIENT, further analysis will likely
`be required to determine which standards apply.
`
`Compliance is checked by testing and inspection as required by the appropriate standards,
`and by inspection of the ACCOMPANYING DOCUMENTS for identification of the standards.
`
`A.2 Completeness of hardware safety
`
`0
`7
`
`Demonstration of hardware safety shall include, but not necessarily be limited to, the following
`potential hazards: electric

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket