throbber
A sample automatic infusion printout is shown in Figure 4.4. Note that the
`first entry of the report contains status information. A normal infusion is
`indicated by the statement "INFUSION TERMINATED NORMALLY." If the
`infusion is unable to start or is interrupted in process, the detected error
`condition(s) is (are) printed out. Infusion error conditions are described in
`detail in the Section 6 of this manual.
`
`The next entries in the infusion report are the set-points selected by the
`operator for the infusion. Elution volume, patient volume, patient dose, and
`dose-rate threshold set-points are listed in the report. Additionally, the
`selected flow rate and the selected calibration factor are listed.
`
`The next entry in the infusion report is actual measured infusion data. The
`actual elution volume, patient volume, and patient dose are listed.
`Additionally, the activity present at end of infusion is listed. This activity is
`the activity that would be present in an infusion collection bottle at the end of
`an infusion. It is this activity value that is used for calibrations since on-going
`decay of the collected eluate is considered. Note that patient dose is the
`direct sum of activity leaving the patient port and does not consider on-going
`decay of the collected eluate. Finally, the activity present between detector
`and waste valve and the activity present between waste valve and patient
`are listed along with the corresponding saline volumes. This data is
`informational only and requires no user action as the infusion system
`automatically considers the effects of radioactivity stored in the tubing lines.
`
`Finally, the infusion report contains an activity profile listing in one-second
`intervals. Both the activity measured at the detector and the activity
`calculated at the patient port are displayed for each one-second interval.
`This data should be interpreted as the amount of activity passing the
`detector and patient port in successive one-second time intervals. Note that
`the transit-time delay between the detector and the patient port can be
`readily observed in the sample infusion report.
`
`19
`
`BRACCOITC00011042
`
`

`

`• The infusion
`operation is started
`by
`depressing the
`inject/sta rt/stop
`switch.
`
`• The saline pump is
`started.
`
`• Low-Level activity is
`directed to the
`waste bottle.
`
`• The running elution
`volume (volume
`through the
`generator) is
`displayed on the
`elution volume
`display.
`
`SALINE
`SUPPLY
`
`GENERATOR DETECTOR
`
`DIVERGENCE
`VALVE
`
`0
`
`op-
`PATIENT LINE
`
`° PUMP LIMIT
`
`HIGH PRESSURE
`
`VALVE FAILURE
`
`SYRINGE
`PUMP
`
`WASTE
`
`(A) Activity Build Up Cycle
`
`Figure 4.3.a Summary of the AUTOMATIC INFUSION Operation
`
`20
`
`BRACCOITC00011043
`
`

`

`SALINE
`SUPPLY
`
`\\/
`
`DIVERGENCE
`VALVE
`
`GENERATOR— DETECTOR
`
`PATIENT LINE
`
`z W
`
`ASTE
`
`o PUMP LIMIT
`
`°HIGH PRESSURE
`
`°VALVE FAILURE
`
`SYRINGE
`PUMP
`
`• Activity is directed
`to the patient line
`once the measured
`infusion dose rate
`exceeds the
`selected dose rate
`threshold.
`
`• The running patient
`volume is dispayed
`on the patient
`volume display.
`
`' The running patient
`dose is displayed
`on the patient
`volume display.
`
`•The running elution
`volume continues
`to be displayed in
`the elution volume
`display.
`
`(B) Patient Infusion Cycle
`
`Figure 4.3.b Summary of the AUTOMATIC INFUSION Operation (continued)
`
`21
`
`BRACCOITC00011044
`
`

`

`• The saline pump
`stops when the
`patient-dose limit is
`reached. The
`elution-volume and
`patient volume
`limits serve as
`backup limits.
`
`• The total elution
`volume, patient
`volume and patient
`dose is stored on
`the displays.
`
`• The generator
`activity is directed
`to the waste bottle
`and patient line is
`diconnected form
`the generator.
`
`• The printor begins
`to print out a
`report of the
`infusion.
`
`SALINE
`SUPPLY
`
`DIVERGENCE
`VALVE
`
`GENERATO
`
`DETECTOR
`
`PATIENT LINE
`
`0 PUMP LIMIT
`
`III
`SYRINGE
`PUMP
`
`°HIGH PRESSURE
`
`0 VALVE FAILURE
`
`WASTE
`
`(C) Infusion Ending Cycle
`
`Figure 4.3.c Summary of the AUTOMATIC INFUSION Operation (continued)
`
`22
`
`BRACCOITC00011045
`
`

`

`INFUSION TERMINATED NORMALLY
`
`SET-POINT VALUES:
`
`Elution Vol. = 40 ml
`Patient Vol. = 30 ml
`Patient Dose = 50.0 mCi
`Dose Rate Thld. = 1.0 mCi/sec
`Flow Rate = 50 ml/min
`Cal. Factor = 926
`
`ACTUAL INFUSION DATA:
`
`Elution Vol. = 26 ml
`Patient Vol. = 19 ml
`Patient Dose = 50.1 mCi
`
`Infused Activity Present
`At End Of Infusion = 45.4 mCi
`
`Activity Present Between
`Det. and Waste Valve - 0.674 mCi
`in Volume of 0.592 ml
`
`Activity Present Between
`Waste Valve and Pat. = 3.23 mCi
`in Volume of 2.75 ml
`
`4.4 Sample of Infusion Report Printout
`
`23
`
`BRACCOITC00011046
`
`

`

`ACTIVITY PROFILE
`
`Time
`sec
`
`Detector Patient
`mCi/sec mCi/sec
`
`1
`2
`3
`4
`5
`6
`7
`8
`9
`10
`11
`12
`13
`14
`15
`16
`17
`18
`19
`20
`21
`22
`23
`24
`25
`26
`27
`28
`29
`30
`31
`32
`
`0.000
`0.000
`0.000
`0.000
`0.000
`0.000
`0.003
`0.056
`0.377
`1.149
`1.937
`3.026
`3.558
`3.648
`3.624
`3.652
`3.650
`3.564
`3.385
`3.134
`2.848
`2.556
`2.277
`2.023
`1.797
`1.611
`1.450
`1.314
`1.204
`1.103
`1.020
`0.578
`
`0.000
`0.000
`0.000
`0.000
`0.000
`0.000
`0.000
`0.000
`0.000
`0.000
`0.016
`0.422
`1.191
`1.996
`3.023
`3.487
`3.548
`3.528
`3.558
`3.549
`3.457
`3.270
`3.024
`2.743
`2.462
`2.188
`1.946
`1.729
`1.552
`1.398
`1.268
`0.709
`
`4.4 Sample of Infusion Report Printout (Cont'd.)
`
`24
`
`BRACCOITC00011047
`
`

`

`5.0
`
`OPERATING PROCEDURE
`
`5.1 General Notes
`
`The sterility of each component must be preserved, be sure to wear
`protective gloves. Aseptic techniques must be strictly observed in all
`handling. Do not remove any luer-lock protective caps until instructed to
`do so. Additionally, do not allow any unprotected luer-lock fittings to make
`contact with anything other than its mating luer-lock fitting.
`
`As system pressure can approach 100 PSI it is important that luer-lock
`fittings be connected tightly without stripping or otherwise damaging the
`fittings. Improper luer-lock connections will result in saline leaks that may
`be difficult to find and correct.
`
`5.2 Rb-82 Generator and Sterile Components
`
`Compare the Rb-82 Generator sterile disposable tray label's parts list, list
`015100, with the actual components in the tray. If not correct, contact BDI
`Nuclear Medicine Customer Service at 1-800-447-6883.
`
`5.3 Rb-82 Infusion System Preparation
`
`Remove the syringe pump cover by loosening the three mounting thumb
`screws and carefully lifting the cover off the pump assembly. Activate the
`Rb-82 Infusion System by depressing the power switch and moving the
`mode switch to the Purge Generator to Waste position.
`
`Raise the lid of the hinged valve shield cover to expose the inside of the
`valve shield assembly. Visually make sure there are no used disposables
`present in the system. If any of the used disposables are present they
`must be disposed of to prevent possible confusion with the new set-up.
`Close the valve cover.
`
`Move the syringe pump to its upper end of the travel position by
`continuously depressing the purge switch while the system is in the
`Purge Generator to Waste mode. The pump will automatically stop
`when it reaches the upper end of the travel position. The LED pump limit
`light will come on.
`
`5.4 Installation of the Syringe Pump Components
`
`5.4.1 Pump Syringe Installation (Pump Syringe Package #1)
`
`Loosen the 4 thumb screws; upper and lower on the pump syringe
`carriage. Install the 140 mL pump syringe with its luer-lock fitting facing
`
`25
`
`BRACCOITC00011048
`
`

`

`upward into its carriage in the pump assembly. Be sure to rotate the
`syringe so that the graduations face toward the operator. Tighten the
`upper and lower thumbscrews to secure both the syringe body and
`plunger into the carriage. Do not remove the protective cap on this
`syringe until directed to do so.
`
`5.4.2 Pressure Sensor Syringe Installation (Inlet Assembly Package #2)
`
`Install the 10 cc pressure sensor syringe by placing it with the luer-lock
`fitting facing upward into the carriage in the pump assembly. This is done
`by tilting the syringe into place with the plunger contacting the bottom of
`the carriage followed by pushing the top of the syringe back until it's
`locked in place. Be sure to rotate the syringe so the graduations face
`toward the operator. Do not remove the protective cap on the syringe until
`directed to do so.
`
`5.4.3 Pump Cover Installation
`
`Re-install the pump cover by carefully sliding it over the top of the pump
`assembly with the pump syringe and pressure sensor syringe protruding
`through the pump cover. Be sure the protective caps on the syringes
`protect the syringe luer-lock fittings from contacting the pump cover.
`Verify that the pump cover is fully seated on the pump assembly.
`
`5.5 Generator Inlet Assembly Package
`
`Remove the tubing assembly from its package and connect the check
`valve to the pump syringe. Tighten the luer-lock fittings but be careful not
`to damage the pump syringe connection.
`
`5.5.1 Pressure Sensor Line Installation
`
`Connect the shortest line of the inlet tubing assembly to the 10 mL sensor
`syringe. Tighten the syringe to the luer-lock connection, but be careful not
`to damage.
`
`5.5.2 Generator Inlet Line Installation
`
`Open the Valve shield. Take the longest line of the generator inlet tubing
`assembly and insert it down through the tubing shield so that it appears in
`the generator shield. Remove the generator shield lead lid and retrieve
`the free end of the generator inlet line. Approximately three to six inches
`of this line will now be in the generator shield with the luer-lock fitting and
`the cap on. Place the saline supply hook assembly back onto the pump
`assembly.
`
`26
`
`BRACCOITC00011049
`
`

`

`5.5.3 Generator Inlet Line Sterilizing Filter Installation
`
`Take the filter from the generator inlet assembly package and connect
`the female luer-lock fitting of the Generator Inlet Sterilizing Filter to the
`male luer-lock fitting on the free end of the generator inlet line that is now
`located in the generator shield. Make sure that the luer-lock fitting on the
`other end of this filter remains in place until directed to do otherwise.
`
`5.6 Generator Outlet Assembly (Package #3)
`
`5.6.1 Divergence Valve Installation
`
`If it is not already open, raise the lid on the valve shield assembly located
`on the top of the Rb-82 Infusion System Cart and remove the valve
`retainer. Note: The lid cannot be raised if the Rb-82 Infusion System is
`unpowered or if the system detects activity within this shield. Take the
`divergence valve, (the white valve with the three arrows and the off sign)
`and orient it with the handle facing down toward the bottom of the Rb-82
`Infusion System Cart. Note: The shortest line of this connection will be
`facing towards the left end of the infusion system cart as you are looking
`at it. Finally, without changing the orientation of the valve, rotate the valve
`handle so that the valve can be dropped into its actuation carriage, to
`prevent it from being moved, with the "valve" arm fitting into the groove of
`the actuation carriage.
`
`5.6.2 Generator Outlet Line Installation
`
`Take the line with the red cap end and feed it through the hole in the
`valve shield assembly. Verify that the end of this line has entered the
`generator shield. There should now be two lines in the generator shield;
`the generator inlet line with its associated filter and the generator outlet
`line. Secure the end of the generator outlet line (located in the valve
`shield assembly) by inserting into the tubing slot near the detector and
`route the line past the detector; secure with detector cover.
`
`5.6.3 Generator Waste Line Installation
`
`Take the line from the outlet tubing assembly with the clear end and feed
`it down the opening of the valve shield assembly. There should now be
`three sets of tubing in the generator shield.
`
`27
`
`BRACCOITC00011050
`
`

`

`5.6.4 Generator Waste Line To Waste Shield
`
`Take the line in the generator shield with the clear cap and feed it through
`the hole inside the generator into the waste shield. Tape the tubing in the
`generator shield so that the generator will not crimp the tubing.
`
`5.6.5 Waste Bottle Preparation and Sterilizing Filter Installation
`
`From the outlet assembly package take the short line with the filter on the
`end and connect it to the waste line inside the waste bottle shield. Once
`this connection has been made, take the waste bottle and connect the
`waste line to the top of the bottle make sure you remove the cover from
`the needle under the bottle cap cover. Verify that the waste bottle
`connections are correct. Carefully lower the waste bottle into the waste
`shield and route the extra waste line tubing to prevent any kinking or
`obstruction in the tubing. Place the lid on top of the waste shield.
`
`The waste bottle should be emptied every morning prior to system
`use or at the end of system usage.
`
`5.7 Patient Line Installation
`
`Remove a patient line from its protective wrapping. Connect the female
`side of the luer-lock fitting, to the line on top of the infusion cart system
`which comes out the left side of the valve. This patient line contains its
`own sterilizing filter. This is the only part of the system that you must
`change as a new patient is being imaged. Make sure that the valve
`retainer is now over the valve and close the lid.
`
`5.8 Saline Supply Installation
`
`5.8.1 Installation of the Saline Supply
`
`Take the remaining free tubing of the generator inlet line, which was the
`first kit you hooked up, pinch off valve and then carefully insert the spike
`into your bottle or bag of preservative-free, normal saline. Make sure the
`spike is fully inserted. Hang the saline supply on the saline supply hook
`located on top of the pump assembly.
`
`28
`
`BRACCOITC00011051
`
`

`

`5.9 Rb-82 Generator Installation
`
`5.9.1 Installation Preparation
`
`Look inside the generator shield, and verify that there are two lines
`available for connection:
`
`• Generator inlet line and its attached filter with a male
`luer lock fitting.
`• Generator outlet line with a female luer-lock fitting
`
`Make sure the waste line that passes through the generator shield
`chamber is against the wall of the generator shield so that it does not
`interfere with the generator installation.
`
`5.9.2 Rb-82 Generator Installation
`
`Remove the generator from its shipping container, and unclip the outlet
`and inlet tubing, carefully lower it into the generator shield. Make sure
`that the generator does not interfere with the waste line that passes
`through the generator shield. Connect the inlet line on the generator
`(marked "inlet," and has a female luer-lock fitting on it) to the remaining
`fitting on the generator sterilizing filter and has a male luer-lock fitting on
`it. Connect the outlet line of the generator (marked outlet and has a male
`luer-lock fitting on it) to the remaining line in the generator shield which
`has a red cap on it. Carefully lower the generator shield lid making sure
`that none of the lines are restricted.
`
`5.10 Purge Operation
`
`5.10.1 Syringe Filling
`
`Position the syringe pump to its limit if not already there, by continuously
`depressing the PURGE switch (in the PURGE GENERATOR TO WASTE
`mode) until the pump automatically stops. Open the saline supply pinch
`valve and depress the REFILL switch once. The syringe pump should
`begin drawing in saline and a stream of air bubbles should appear in the
`saline supply bottle. The refill operation can be stopped at any time by
`depressing the REFILL switch a second time. Otherwise, the pump will
`continue refilling until it reaches its refill limit and automatically stops.
`Unless problems are encountered, allow the pump to refill until it
`automatically stops on its refill limit.
`
`29
`
`BRACCOITC00011052
`
`

`

`5.10.2 Waste Line Purging
`
`With the MODE switch in the PURGE GENERATOR TO WASTE position
`continually depress the PURGE switch to pump saline through the
`generator and into the waste bottle. A volume of 50 mL of saline should
`be sufficient to guarantee purging the waste lines of air. Note: The purge
`volume can be checked by observing the markings on the pump syringe.
`Lift the waste bottle shield cover and look for any signs of leakage.
`Correct any leaking connections before continuing to use the Rb-82
`infusion system.
`
`Note: It is recommended that leak testing be carried out by wipe test and
`radiation detection survey.
`
`5.10.3 Pressure Sensor Line
`
`Pinch off the saline bag valve, disconnect the pressure sensor line from
`the pressure sensor syringe and place it in a beaker. Depress the purge
`switch until liquid flows into the beaker. Reconnect the pressure sensor
`line to the pressure sensor syringe. Now unpinch the saline valve.
`
`5.10.4 Patient Line Purging
`
`Place the MODE switch in the PURGE GENERATOR TO PATIENT
`position. Connect the patient line to a shielded 50 mL vial and insert a
`venting needle. Continuously depress the PURGE switch until all the air
`has been expelled from the patient line into the vial. A volume of 20 mL
`should be sufficient to purge the patient line. The purge volume can be
`determined by observing the pump syringe markings.
`
`Lift the generator shield cover and look for any signs of leakage. Correct
`any leaking connections before continuing to use the Rb-82 Infusion
`System.
`
`Note: It is recommended that leak testing be carried out by wipe test and
`radiation detection survey.
`
`5.10.5 Volumetric Flow Rate Verification
`
`This is performed with the installation of a new generator.
`Set the controls on the Display/Control Panel as follows:
`
`• Mode Switch
`=
`• Elution Volume Limit =
`• Patient Volume Limit =
`
`Automatic Infusion
`99 mL
`50 mL
`
`30
`
`BRACCOITC00011053
`
`

`

`=
`• Patient Dose Limit
`• Dose Rate Threshold =
`• Flow Rate
`=
`
`99 mCi
`1.0 mCi/sec.
`50 mL/min.
`
`Check for a 50 mL/min. flow rate using a stop watch.
`
`5.11 Patient Administration
`
`PATIENT ADMINISTRATION MAY BE PERFORMED ONLY AFTER
`SUCCESSFUL COMPLETION OF DAILY CALIBRATION, SR-82/85
`BREAKTHROUGH PROCEDURES, AND FIRST WASH (ELUTION)
`DISPOSAL USING SAME SETTINGS AS SEEN ON SR-82/85
`BREAKTHROUGH SHEET PAGE.
`
`IN THE EVENT OF A POWER-FAILURE OR THE SYSTEM IS
`INADVERTENTLY SHUT DOWN, CALIBRATION SHOULD BE
`RECONFIRMED.
`
`5.11.1 Replace the patient administration set for each new patient.
`
`5.11.2 Purge all air out of the patient administration set and verify that all
`air is purged from the system. Refer to Installation Instructions, section
`5.10.4.
`
`5.11.3 Set the controls on the Display/Control panel as prescribed by the
`administering physician.
`
`5.11.4 Verify that the syringe pump has been filled with saline. The
`syringe pump volume must at least equal the selected elution volume-set
`point plus 20 mL. This volume will cover the 15 mL nominal dead volume
`in the syringe.
`
`5.11.5 Verify that the system printer is on and that approximately 1 inch
`of paper extends out of the printer. If the RS-232C port is being used for
`infusion data acquisition, the system printer can be turned off. Infusion
`report data is transmitted out the RS-232C port whether the system
`printer is powered or not.
`
`5.11.6 Attach the Rb-82 infusion system patient line to the patients
`intravenous line.
`
`5.11.7 Make sure that at least 10 minutes has elapsed following any
`purge or infusion operation. Start the infusion by depressing the INJECT
`START/STOP switch.
`
`NOTE: The infusion can be terminated at any time by depressing the
`INJECT START/STOP switch a second time.
`
`31
`
`BRACCOITC00011054
`
`

`

`The running elution volume, patient volume, patient dose and dose rate
`will be registered on the control/display panel. Once the infusion
`operation is completed the INJECT START/STOP switch will stop glowing
`red and the infusion report data will be printed on the system printer and
`echoed out the RS-232C port.
`
`5.11.8 Unless manually interrupted, the infusion will terminate when the
`number of mCi preset on the PATIENT DOSE LIMIT switch has been
`reached. As a safety precaution, the infusion will also terminate if the
`preset ELUTION VOLUME LIMIT or PATIENT VOLUME LIMIT are
`reached.
`
`NOTE: THE SR-82/RB-82 GENERATOR ELUATE IS RADIOACTIVE
`AND SHOULD BE HANDLED WITH PROPER RADIATION
`SAFETY PRECAUTIONS.
`
`NOTE: WAIT AT LEAST 10 MINUTES AFTER ANY PURGE OR
`INFUSION OPERATION BEFORE LIFTING THE GENERATOR
`SHIELD COVER, WASTE BOTTLE SHIELD COVER OR THE
`HINGED VALVE-SHIELD COVER. THIS WILL ALLOW ANY
`RB-82 ELUATE TO DECAY TO A SAFE LEVEL.
`
`NOTE: THE HINGED VALVE-SHIELD COVER AUTOMATICALLY
`LOCKS IN THE CLOSED POSITION DURING ANY PURGE OR
`INFUSION OPERATION. THE COVER REMAINS LOCKED
`UNTIL THE DETECTED RADIOACTIVITY DECAYS TO A
`SAFE LEVEL. NO PURGE OR INFUSION OPERATION CAN
`BE STARTED UNLESS THE COVER IS CLOSED.
`ADDITIONALLY, THE COVER CANNOT BE RAISED UNLESS
`THE RB-82 INFUSION SYSTEM IS POWERED.
`
`NOTE: ALWAYS WEAR GLOVES BEFORE TOUCHING ANY OF THE
`TUBING, SYRINGE PUMP, VALVE OR COLLECTION BOTTLE
`SYSTEM COMPONENTS.
`
`32
`
`BRACCOITC00011055
`
`

`

`co L L000011033VIE1
`
`FEMALE LUER LOCK
`
`CD
`
`MALE LUER LOCK
`
`(PACKAGE 3)
`FILTER SET
`WASTE LINE
`
`(PACKAGE D
`IOML
`SYRINGE
`
`UMP
`
`A
`
`WASTE BOTTLE
`
`, r
`
`SHIELD
`TUBING
`
`GENERATOR
`
`Rb-82
`
`INLET
`
`çi
`
`04
`
`(RED DOT)
`OUTLET
`
`SHEILD
`GENREATOR
`
`(PACKAGE 2)
`INLET FILTER
`GENERATOR
`
`(CUSTOMER PROVIDED)
`
`SALINE SUPPLY
`
`ASSEMBLY
`PUMP
`
`(PACKAGE 2)
`ASSEMBLY
`TUBING
`INLET
`GENERATOR
`
`(PACKAGE 2)
`IOML
`SYRINGE,
`SENSOR
`PRESSURE
`
`DETECTOR
`
`SHIELD
`VALVE
`
`PACKAGE 3)
`ASSEMBLY
`TUBING
`OUTLET
`GENERATOR
`
`(PACKAGE 4)
`SET
`ADMINISTRATION
`PATIENT
`
`weibuci buqni
`
`

`

`CardioGen-82 Generator
`Sr-82/85 Breakthrough Worksheet
`
`Date
`
`Infusion System Control Panel Settings
`
`Mode Switch:
`Elution Volume:
`Patient Volume:
`Patient Dose:
`Dose Rate:
`Flow Rate:
`
`Dose Calibrator Setting
`
`Automatic Infusion
`99 mL
`50 mL
`99 mCi
`1 mCi/sec.
`50 mL/min.
`
`Figure 5.1 Tubing Diagram
`
`504 (Capintec only) or
`Co-60 setting then divide reading obtained by 0.548
`
`1.
`
`Elute 50 mL of eluate into 50 mL vial using Infusion System Control Panel
`Settings as above. Note exact time at end of elution (EOE).
`
`Time when elution ended
`
`2. Set dose calibrator as above. Setting used:
`
`3. Measure Rb-82 activity in dose calibrator. Note exact time (minutes/seconds)
`when measurement is made.
`
`Rb-82 activity
`Time of measurement
`
`(mCi).
`
`4. Decay correct Rb-82 measurement to time when elution ended (EOE).
`
`(Note: If time between end of elution and measurement is allowed to be 150
`seconds (2.5 minutes), multiply dose calibrator reading by a factor of 4).
`
`Decay correction factor
`Rb-82 activity at EOE (mCi)
`
`5. Using same vial, let sample stand for 60 minutes to allow for complete decay
`of Rb-82.
`
`34
`
`BRACCOITC00011057
`
`

`

`6. Measure sample in dose calibrator.
`
`7. Calculate amount of Sr-82 in sample using the following equation:
`
`Sr-82 = dose calibrator reading (pCi)
`divisor (from sheet supplied with Generator)
`
`Example:
`
`Sr-82 = 0.5 =.216
`2.3
`
`Sr-82 = .216 µCi
`
`8. Determine Sr-82 Breakthrough by dividing tiCi of Sr-82 by the mCi of Rb-82
`at EOE
`
`Example:
`
`.216µCi Sr-82 = .004 µCi Sr-82/mCi Rb-82
`50 mCi Rb-82
`
`Allowable Limit = .02 liCi Sr-82/mCi Rb-82
`
`9. Determine Sr-85 Breakthrough by multiplying the result obtained in step 8 by
`the Sr-85/Sr-82 ratio from the data sheet supplied with the generator.
`
`Example:
`
`.004x 1.48 =.00592
`
`Allowable Limit = 0.2 pCi Sr-85/mCi Rb-82
`
`35
`
`BRACCOITC00011058
`
`

`

`Rb-82 INFUSION SYSTEM-CALIBRATION DATA SHEET
`
`Infusion System Control Panel Settings
`
`Mode Switch:
`Elution Volume:
`Patient Volume:
`Patient Dose:
`Dose Rate:
`Flow Rate:
`
`Automatic infusion
`99 mL
`50 mL
`60 mCi
`1 mCi/sec.
`50 mL/min.
`
`1. Date of Calibration or Calibration Verification:
`
`2. Initial Generator Sr-82 Potency on calibration date:
`
`3. Generator Sr-82 Potency on present date:
`
`4. Present Calibration Factor:
`
`mCi
`
`mCi
`
`5. Measured Rb-82 Activity from dose calibrator (corrected for decay):
` mCi
`
`6. Printed Rb-82 Activity Present at the end of infusion:
`
`mCi
`
`7. Ratio of Measured Rb-82 Activity and Printed Rb-82:
`
`Measured Rb-82 Activity
`Printed Rb-82 Activity =
`
`Ratio
`
`8. Is the difference between Measured Rb-82 Activity and Printed Rb-82 .95 -
`1.1?
` (If no, unit requires recalibration).
`
`9. Calculated New Calibration Factor (if needed):
`
`New
`
`Initial
`
`Measured Rb-82 Activity at End of
`Infusion
`
`Cal.
`Factor
`
`=
`
`Cal.
`Factor
`
`x
`Printed Rb-82 Activity Present at
`End of Infusion
`
`10. Is new Calibration factor within 5% of old Calibration factor?
`
` (If no, repeat calibration).
`
`36
`
`BRACCOITC00011059
`
`

`

`operation manual.
`** Calibration should be performed to 15% (Ratio=0.95 to 1.05). Daily calibration checks should be within ±10% (Ratio=0.9 to 1.1). See Section 5.2 of the
`* Use the dose calibrator reading directly if an RB-82 calibration setting is used or divide the reading by 0.548 if a Co-60 Calibration setting is used.
`
`Figure 5.2 Rb-82 Infusion System Calibration Log Sheet
`
`090 1.1-000011033VIE1
`
`FOS
`
`EGS
`
`N/A
`
`853
`
`Initials
`Operator
`
`Factor
`Calib.
`New
`
`J
`
`553
`
`1000
`
`Printer)
`Rb-82
`(from
`Factor
`Calib.
`
`I
`
`Ye:
`
`No
`
`1.10
`.90 to
`G =
`±10%?
`Calib.
`
`H "
`
`Yes
`
`No
`
`1.05
`.95 to
`G =
`t5%?
`Calib.
`
`CALIBRATION SHOULD ALWAYS BE DONE AT THE DESIRED PATIENT DOSE SETTINGS. THE FIRST ELUTION OF THE DAY MUST BE DISCARDED
`
`AND NOT USED FOR CALIBRATION. REFER TO SECTION 5.2 OF TI IE OPERATION MANUAL FOR INSTRUCTIONS.
`
`',d1
`
`, .. s. 7
`
`:“ 2
`
`13. 8'
`
`13.8
`
`.853
`
`57 7
`
`d3 8
`
`2:30
`
`12 2
`
`122
`
`(E/F)
`
`E to F
`Ratio
`
`G
`
`(mCi)
`Infusion
`End of
`Activity
`Printed
`
`F
`
`(mCi)
`Factor)
`
`(C x Decay
`Corrected
`Dose, Decay
`
`E
`
`(min:sec)
`
`Time
`Decay
`I )ose
`
`D
`
`(mCi)
`
`(B/0.548)
`(B/1) or
`
`Dose
`
`Corrected
`
`C*
`
`Calibrator
`From Dose
`
`(mCi)
`
`Reading
`Dose
`
`B
`
`8:12 AM
`
`Example: 9128/85
`
`8:00 AM
`
`Example: 9/28/85
`
`(Date & Time)
`
`Calibration Infusion
`
`A
`
`0 Sec
`45 Sec
`30 Sec
`15 Sec
`0 Sec
`
`5 Min
`3 Min
`2 Min
`1 Min
`0 Min
`
`Factor
`
`Measured 'rime
`
`Decay Factor Table
`
`3. Date
`
`mCi/sec
`
`2. Initial Gen. Potency
`
`1. Generator Lot Number
`
`nil/min
`
`mCi
`
`1.0
`
`50
`
`3. Patient Dose Threshold
`
`2. Flow Rate
`
`1. Patient Dose
`
`GENERATOR DATA
`
`INFUSION SYSTEM SETTINGS
`
`Rb-82 Infusion System Calibration Log Sheet
`
`Site Name
`
`K
`
`16
`
`1
`
`2
`
`4
`
`8
`
`

`

`Initials
`Operator
`
`L
`
`u Ci/mCi?
`
`<0.02
`Rb-82
`Sr-82 to
`Ratio
`
`K
`
`(u. Ci/mCi)
`
`(j.i Ci)
`
`I/E
`Rb-82
`Sr-82 to
`Ratio
`
`J
`
`G/H
`through
`Break-
`Sr-82
`
`I
`
`Data
`Gen.
`From
`Divisor
`through
`Break-
`
`H
`
`Corrected
`through
`Break-
`0'
`TRUCTIONS.
`
`(u Ci)
`F/0.548
`F/1 or
`
`(µ Ci)
`Calibrator
`from Dose
`Reading
`through
`Break-
`
`F
`
`(mCi)
`Factor)
`
`(C x Decay
`Corrected
`Decay
`Rb-82,
`
`E
`
`(min:sec)
`
`(mCi)
`
`(mCi)
`
`Time
`Decay
`Rb-82
`
`D
`
`(B/0.548)
`(B/1) or
`Rb-82
`
`Corrected
`
`C'
`
`Calibrator
`From Dose
`Reading
`Rb-82
`
`B
`
`THE FIRST ELUTION OF THE DAY MUST BE DISCARDED AND NOT USED FOR BREAKTHOUGH MEASUREMENT. REFER TO SECTION 5.3 OF THE
`
`I
`
`2
`
`4
`
`8
`
`Factor
`
`45 Sec
`30 Sec
`15 Sec
`0 Sec
`
`3 Min
`2 Min
`1 Min
`0 Min
`Measured Time
`
`Decay Factor Table
`
`3. Date
`2. Initial Gen. Potency
`1. Generator Lot Number
`
`ml
`ml
`nil
`
`GENERATOR DATA
`
`mCi/sec
`50
`50
`99
`mCi
`INFUSION SYSTEM SETTINGS
`
`1.0
`
`99
`
`5. Patient Dose Threshold
`4. Flow Rate
`3. Patient Volume
`2. Elution Volume
`1. Patient Dose
`
`Sr-82 Breakthrough Log Sheet
`CARDIOGEN® Rb-82 Generator
`
`Site Name
`
`* Use the dose calibrator reading directly if an RB-82 calibration setting is used or divide the reading by 0.548 if a Co-60 Calibration setting is used.
`
`Figure 5.3 CARDIOGENO-82 GENERATOR SR-82 Breakthrough Log Sheet
`
`1.901. 1.000011000VIE1
`
`W1.0
`
`Yes
`
`0.0026
`
`0.124
`
`6.43
`
`0.8
`
`0.8
`
`48.4
`
`2:30
`
`12.1
`
`12.1
`
`9:00 AM
`
`Example: 9/28/85
`
`(Date & Time)
`
`Breakthrough Infusion
`
`A
`
`

`

`6. TROUBLESHOOTING GUIDE
`
`The table below describes possible purge and infusion error conditions that
`are displayed on the Control/Display panel. These errors are also listed in
`the infusion report printout. Suggested corrective action is provided in the
`table for each error condition.
`
`This device has been tested for protection against electro-magnetic
`interference according to IEC 60601-1-2. However, the proximity of other
`devices that give off electro-magnetic radiation, such as X-ray machines, can
`interfere with the operation of this system.
`
`TROUBLESHOOTING TABLE
`
`Error Condition Suggested Action
`
`Automatic Infusion Errors (listed in infusion report printout)
`
`"Elution Vol. Setpt. = 0"
`"Patient Vol. Setpt. = 0"
`"Patient Dose Setpt. = 0"
`
`Select non-zero infusion set-points
`Select non-zero infusion set-points
`Select non-zero infusion set-points
`
`"Valve Shield Open"
`"Pump At Limit Position"
`"High Pressure Error"
`"Valve Error"
`
`Close the valve shield
`Refill syringe pump
`Look for possible tubing constriction
`Notify Bracco Diagnostics Inc.
`
`Automatic Infusion and Purge Errors (displayed on Control/Display panel)
`
`PUMP LIMIT
`
`HIGH PRESSURE
`VALVE ERROR
`Valve Shield Open
`(not displayed)
`
`Refill the syringe pump, if necessary.
`Note that this display is also active
`if the pump is on the refill limit.
`Look for possible tubing constriction
`Notify Bracco Diagnostics Inc.
`Close the valve shield for all purge and
`infusion operations
`
`39
`
`BRACCOITC00011062
`
`

`

`SYSTEM DIARY
`
`Date:
`Problem:
`
`Action Taken:
`
`Date:
`Problem:
`
`Action Taken:
`
`Date:
`Problem:
`
`Action Taken:
`
`Date:
`Problem:
`
`Action Taken:
`
`40
`
`BRACCOITC00011063
`
`

`

`Date:
`Problem:
`
`Action Taken:
`
`Date:
`Problem:
`
`Action Taken:
`
`Date:
`Problem:
`
`Action Taken:
`
`Date:
`Problem:
`
`Action Taken:
`
`Date:
`Problem:
`
`Action Taken:
`
`41
`
`BRACCOITC00011064
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket