throbber

`
`
`
`
`
`The touch screen uses intuitive color-coding: green indicates normal operation and red indicates an
`error, warning or that some intermediate step is needed. Blue indicates the user's menu selection and
`also indicates non-radioactive operation; the yellow radiation symbol indicates radioactive operation so
`the user knows when the unit is generating ionizing radiation.
`
`
`
`When the unit is in radioactive operation, users should move away from the unit
`to a distance that allows them to avoid unnecessary exposure, in accordance
`with ALARA principles.
`
`
`
`
`
`
`
`935 of 1085
`
`21
`
`

`

`
`
`
`
` LOADING PRINTING LABELS 5.3
`
`The system ships with one packet of labels for use
`with the RUBY Rubidium Elution System. Additional
`labels must be ordered from Jubilant DraxImage
`Inc.
`
`To load the labels, open the printer access door on
`the back of the system (see Fig. 16, Printer Access
`Door). The label should face out when it is unfolded
`up
`toward
`the printer
`(see Fig. 17, Label
`Orientation). Thread the leading edge of the label
`up through the opening (see Fig. 18, Label Feed).
`Make sure the edges of the label are parallel to the
`edges of
`the print mechanism. The
`labels
`automatically feed into the printer. Close the printer
`access door before proceeding.
`
`
`
`
`
`
`Figure 16, Printer Access Door
`
`Figure 17, Label Orientation
`
`Figure 18, Label Feed
`
`22
`
`
`
`936 of 1085
`
`

`

`
`
`
`
` CONFIGURING THE SYSTEM 5.4
`
`At installation, the system needs to be configured to the site's specification. The Jubilant DraxImage
`representative will configure the unit according to each site's needs by selecting Settings on the top
`row of the task bar. The admin level PIN may also access the Settings section that includes a choice of
`units, time zones, site-specific identification (name of clinic or hospital) and personal identification
`numbers (PINs) (see Fig. 19, Settings Control Panel Screen).
`
`These five settings must be defined:
`
`
`1. Manage PINs (The system's default PINs are User: 12345; Administrator: 11111)
`2. Constancy Standard Setup
`3. Site Setup
`4. Elution Setup
`5. USB Transfer File Setup (This should be performed prior to report transfer)
`
`After the settings have been modified, press Save Changes, and return to the main software page by
`pressing Home at the bottom of the screen.
`
`
`
`
` MANAGING PERSONAL IDENTIFICATION NUMBERS (PINS) 5.5
`
`The RUBY Rubidium Elution System has two permission categories: administrator (admin) and user.
`The admin level has authority to select important site parameters determined by the medical staff. The
`user level allows the nuclear medicine technologist to perform normal set-up procedures and use the
`RUBY Rubidium Elution System. To modify a PIN, the administrator clicks on the appropriate
`permission level, then enters and confirms the new PIN. The administrator may reset the admin level
`PIN, or the PIN of the user (see Fig. 20, Managing PINs Screen).
`
`
`Figure 19, Settings Control Panel Screen
`
`Figure 20, Managing PINs Screen
`
`
`
`
`
`937 of 1085
`
`23
`
`

`

`
`
`
`
`There is one admin PIN for all administrators and one user-specific PIN for all
`users. An admin level PIN must be entered to modify any setting.
`
`
`
`
`
` CONSTANCY STANDARD SETUP 5.6
`
`The RUBY Rubidium Elution System is equipped with an on-board dose calibrator. The daily quality
`assurance procedure recommended by the manufacturer and by regulatory bodies includes a daily
`constancy check of the dose calibrator. The check is performed by using a calibrated standard (for
`example, Cs-137) of long half-life and checking against it every day. The calibrated standard
`parameters must be entered in all the fields shown (see Fig. 21, Constancy Standard Setup Screen).
`
`
`
`
` SITE SETUP 5.7
`
`The RUBY Rubidium Elution System has a built-in report system. To customize reports and labels with
`site information, fill in all the fields on the screen (see Fig. 22, Site Setup Screen). The information
`entered here appears on all reports generated by the system (Daily Quality Control and Patient
`Infusion reports). The site setup information is also a part of a system setup.
`
`
`
`Changing the system time, time zone and/or date could adversely affect future elutions. These parameters
`are set at initial installation. The Admin level PIN is required to adjust the system time for the daylight saving
`time period.
`
`
`
`Press Save Changes to keep the changes. Then go to the Main Menu to select the next
`Settings screen or the Home screen. To restore default values, select Restore Defaults and
`Save Changes to reset the system to the factory assigned values.
`
`
`
`
`
`938 of 1085
`
`24
`
`

`

`
`
`
`
`Figure 21, Constancy Standard Setup Screen
`
`Figure 22, Site Setup Screen
`
`
`
`
`
` ELUTION SETUP 5.8
`
`The RUBY Rubidium Elution System has four configurable elution parameters: (see Fig. 23, Elution
`Setup Screen)
`
`
`1. Activity/Patient Weight Ratio determines the amount of activity to inject per kilogram or
`pound. This ratio is multiplied by the weight of the patient to determine the total activity for
`the procedure.
`2. Extra Flush to Patient sets the volume of saline that is injected after the total activity has
`reached the patient to lead the isotope to the heart. The default value is 5 mL. This volume
`is included in the 60 mL limit of total volume infused.
`3. Labels Printed per Elution allows the user to determine how many copies of each label
`need to be printed after each daily QC and patient infusion. The default value is 1.
`
`
`
`
`
`939 of 1085
`
`25
`
`

`

`
`
` Figure 23, Elution Setup Screen
`
`
`
`
`
` USB FILE TRANSFER SETUP 5.9
`
`After each Patient Infusion or Daily QC, data is saved on the system's hard drive and accessible
`through the USB Port (see Fig. 24, USB Port). Report data can be transferred using the USB key
`provided with the system, or saved by the report application, which can upload to a web-based
`application for storage on one dedicated personal computer that is synced with the elution system. To
`configure a report for USB transfer, follow the steps on USB File Transfer Setup screen (see Fig. 25,
`USB File Transfer Setup Screen).
`
`Only one computer can be configured to run the Reports Application for each RUBY Rubidium Elution
`System. This ensures secure handling of report data. Refer to the Report Application instructions for
`information on how to authenticate the USB key and use the Report Application.
`
`
`
`If a USB other than the formatted USB key is used, it will erase all information on that USB key.
`
`
`
`
`
`
`
`940 of 1085
`
`26
`
`

`

`
`
`Figure 24, USB Port
`
`
`
`
`
`
`
`Figure 25, USB File Transfer Setup Screen
`
`
`
`The RUBY Rubidium Elution System has a scenario mode that should be turned off
`during normal use of the system. This is disabled by a JDI representative during
`installation. This can also be performed by the administrator.
`
`
`
`
`
`
` SYSTEM SHUTDOWN 5.10
`
`To shutdown the RUBY Rubidium Elution System, select shutdown from the Home Screen. Select OK
`to continue with shutdown, or Cancel to cancel the shutdown. The RUBY Rubidium Elution System
`must be connected to a power outlet and the main on/off switch near the power cord at the back base
`of the system must be in the ON position to allow for continuous charge to the battery. For a complete
`shutdown, Press on the main on/off switch near the power cord at the back base of the system to the
`off position. Turn off the UPS by opening the printer access door and pressing the power button.
`
`
`
`
`
`941 of 1085
`
`27
`
`

`

`
`6. OPERATING THE SYSTEM
`
` DOSE CALIBRATOR QUALITY ASSURANCE 6.1
`
`The RUBY Rubidium Elution System comes with an integrated dose calibrator. This dose calibrator is
`fully controlled using the system's software. To access the quality assurance mode, press Dose
`Calibrator Quality Assurance on the second row of the task bar and enter the required password. The
`control screen is displayed (see Fig. 26, Dose Calibrator Quality Assurance Screen). To perform the
`required quality assurance tests, such as Linearity, Geometry and Accuracy, an independent control
`screen is included to enable the operation and/or configuration of the dose calibrator in conventional
`mode (Table 2, Dose Calibrator Quality Assurance & Required Frequency of Performance). JDI
`personnel will complete all four quality assurance tests upon installation of the elution system. Users
`are responsible for performing Constancy daily (integrated into Daily Quality Control); Linearity
`quarterly and Accuracy annually.
`
`Dose Calibrator Quality
`Assurance Test
`Constancy
`
`Required Frequency
`
`Upon Installation; Daily (integrated into Daily Quality Control procedure);
`following repair
`Upon installation; quarterly; following repair
`Linearity
`Upon installation; annually; following repair
`Accuracy
`Upon installation; following repair
`Geometry
`Table 2, Dose Calibrator Quality Assuarance & Required Frequency of Performance
`
`Required frequency of dose calibrator quality assurance tests may vary according to local
`regulations. Daily Dose Calibrator Constancy check is integrated into Daily Quality Control tab.
`
`
`
`
`
`
`
` PERFORMING READINGS ON THE DOSE CALIBRATOR 6.2
`
`To perform dose calibrator readings to complete quality assurance testing the user will follow the
`steps below on the Dose Calibrator Quality Assurance screen (see Fig. 26, Dose Calibrator Quality
`Assurance Screen).
`
`
`1. Click Stop Continuous Reading
`2. Enter the isotope to be assayed (for example: Cs-137 or TC-99) by entering the isotope
`into the box displayed on the screen labeled Isotope
`3. Click on Change Isotope
`4. Ensure that the dose calibrator chamber is empty and press Zero Background
`5. Insert radioactive sample to be assayed
`6. Press Perform Reading or press Start Continuous Reading to see the stabilization of the
`reading or to monitor the decay automatically
`7. When finished, select Home Screen on the touchscreen monitor to return the Home
`Screen
`
`
`
`942 of 1085
`
`28
`
`

`

`
`
`
`
`
`
`Figure 26, Dose Calibrator Quality Assurance Screen
`
`
`
`
`
` GENERATOR INSTALLATION & SETUP 6.3
`
`Generator Installation is performed by the user by accessing Generator Installation & Setup on the
`second row of the task bar. Press Generator Installation & Setup and enter your PIN (see Fig. 27,
`Generator Installation & Setup Screen). From there, six task screens detail each step required to safely
`remove an expired generator and accessories and install a new generator and accessories.
`
`Some safety precautions are necessary when working with the generator and its shipping container.
`Be sure to follow these steps:
` (cid:135) The generator is heavy; work slowly and deliberately when handling the shipping container.
`(cid:120) Ensure correct placement of the generator handle to the generator container before moving the
`generator.
`(cid:135) Lift and lower the generator slowly into the generator well to avoid damaging the lead shielding.
`
`
`
`
`
`943 of 1085
`
`29
`
`

`

`
`
`Figure 27, Generator Installation & Setup Screen
`
`
`
`
`
` REMOVAL OF EXPIRED COMPONENTS 6.4
`
`Before installing a new Generator, the expired Generator and RUBY Tubing Sets must be removed. An
`expired Generator still contains a significant quantity of radioactive strontium (Sr-82, Sr-85) and must
`be discarded or stored according to local procedures.
`
`To remove the expired Generator and consumables, perform the first two tasks identified in the left
`column on the screen: Remove Generator and Remove RUBY Tubing. Click Next after each task.
`When you select Next, the system considers those tasks to be complete.
`
`
`tubing components are radioactive. All
`The Generator and expired
`procedures for disposal of radioisotopes should be followed.
`
`
`
`
`
`
`
`944 of 1085
`
`30
`
`

`

`
`
`
`
` 6.5
`
`INSTALLING THE RUBY SET
`
`If any part of the RUBY SET becomes contaminated or damaged before or during installation, a new
`RUBY SET must be installed by the user.
`
`
`1. Obtain new RUBY SET and visually inspect packaging for damage.
`2. Remove from packaging. Do not remove the protective caps yet from any of the connectors
`at the end of each line.
`Insert plastic tube holder containing RUBY SET into place on the RUBY Rubidium Elution
`System (see Fig. 28, Tubing Holder).
`
`3.
`
`Figure 28, Tubing Holder
`
`Figure 29, Flow Regulator
`
`
`
`
`
`4. Set Flow Regulator to 250 mL/h (see Fig. 29, Flow Regulator).
`5. Clip Pressure Sensor into holder (#1 & #2 positions on the RUBY Rubidium Elution System)
`(see Fig. 30, Pressure Transducer).
`6. Plug in #3 connector (see Fig. 31, Pressure Sensor Connector).
`
`
`
`
`
`
`
`Figure 31, Pressure Sensor Connector
`
`
`
` Figure 30, Pressure Transducer
`
`Check expiry dates on all consumables prior to use
`
`
`7. Insert line #1 into pinch valve 2nd from the right (see Fig. 32, RUBY SET Installation
`Diagram, or the Installation Diagram on the RUBY Rubidium Elution System, section 9.5)
`
`31
`
`
`
`945 of 1085
`
`

`

`
`
`
`
`
`
`
`
`946 of 1085
`
`8. Insert line #2 into pinch valve furthest to the right.
`9. Insert waste section line into pinch valve with red circle around it.
`10. Insert patient section line into pinch valve with blue circle around it.
`11. Remove the protective cap and attach the Waste portion of tubing (next to the Red Trash
`Icon) to the Waste Bottle.
`12. Insert line section between #4 and #5 into groove of the PMT, ensuring PMT door may be
`closed properly after tube placement (See Fig. 33, Installation of PMT portion of RUBY
`SET).
`
`Be sure to properly connect Waste portion of RUBY SET to waste container with
`each RUBY SET installation and after each time the waste container is emptied. The
`waste container holds 1 L of fluid and must be emptied daily.
`
`
`
`PInch
`Valves
`
`Figure 32, RUBY SET installation diagram
`
`
`
`
`
`PMT
`
`Figure 33, Installation of PMT portion of RUBY SET
`
`
`
`
`
`
`
`32
`
`

`

`
`
`11. Using aseptic technique, open the sterilized package of RUBY CONNECTORS. Connect IN to
`the female RUBY CONNECTOR and OUT to the male RUBY CONNECTOR (see Fig. 34,
`Female RUBY CONNECTOR to IN, Fig. 35, Male RUBY CONNECTOR to OUT).
`12. Using aseptic technique, connect both RUBY CONNECTORS to each other to maintain sterility
`until Generator has been installed (see Fig. 36, Connect RUBY CONNECTORS attached to
`each other).
`
`
`
`
`
`
`
`Figure 34, Female RUBY CONNECTOR to
`IN
`
`
`
`Figure 35, Male RUBY CONNECTOR to
`OUT
`
`
`
`
`
`Figure 36, RUBY CONNECTORS attached
`
`
`
`Aseptic technique must be used when connecting the RUBY SET to the RUBY
`CONNECTORS. Do not use RUBY CONNECTORS if packaging is damaged.
`
`
`It is recommended to have a spare set of RUBY CONNECTORS and RUBY SET on site at all
`times.
`
`
`
`
`
`33
`
`
`
`947 of 1085
`
`

`

`
`
`
`
` 6.6
`
`INSTALLING THE NEW GENERATOR
`
`Using the Generator handle, install a new Generator by inserting it carefully into the Generator well
`(see Fig. 37, Installing the New Generator). Remove the Generator handle, then the cover. Using
`aseptic techniques, remove the metal caps on the Generator and connect both RUBY CONNECTORS
`to the quick connects on the Generator, mating a male to a female quick connect. Make sure the
`tubing lines are in the grooves of the Generator lead well and close the Generator well lid promptly to
`avoid unnecessary exposure. The Generator packing slip is shipped with the Generator in a plastic
`pouch inside of the shipping container. Retrieve the packing slip, and on the Generator Setup screen,
`step 5 under the Generator Installation and Setup (Fig. 38, Generator Setup Screen) enter the lot
`number, strontium ratio (Sr-85/Sr-82), calibration date and expiration date.
`
`
`
`
`Figure 37, Installing the New Generator
`
`
`
`
`
`Figure 38, Generator Setup Screen
`
`
`
`Save the Generator handle and cover with shipping container for use when removing
`the expired Generator and metal caps to recap the generator.
`
`The Generator and expired components are radioactive. All procedures for disposal of
`radioisotopes should be followed.
`
`
`
`
`
`The Generator is heavy. Use proper lifting technique whenever lifting or moving the
`Generator.
`
`It is very important that the Generator information be entered correctly.
`
`
`
`
`
`
`
`
`
`948 of 1085
`
`34
`
`

`

`
`
`
`
` 6.7
`
`INSTALLING THE 0.9% SODIUM CHLORIDE (ADDITIVE FREE) INJECTION, USP
`(SALINE SUPPLY) & RUBY SALINE LINE
`
`A 0.9% sodium chloride (additive free) injection USP saline bag, RUBY SALINE LINE and RUBY IV
`LINE must be installed by the user to perform the Setup Validation in Daily Quality Control tab on the
`touchscreen monitor.
`
`1. Visually inspect the RUBY SALINE LINE packaging for damage. Discard if damaged.
`2. Using aseptic techniques, remove protective cap from saline bag and spike saline bag with the
`spike end of the RUBY SALINE LINE.
`3. Ensure that the roller clamp is in the closed position on the RUBY SALINE LINE and hang the
`saline bag on the Saline hook.
`4. Insert the RUBY SALINE LINE into the RUBY Rubidium Elution System via the keyhole on the right
`hand side of the system (see Fig. 39, RUBY SALINE LINE keyhole).
`5. Open the pump by pulling lever and install the RUBY SALINE LINE into the pump (see Fig. 40,
`Installing RUBY SALINE LINE into Pump). Close the pump using the lever.
`6. Using aseptic techniques, remove the protective cap of the RUBY SALINE LINE at point A and
`connect to the A portion of the RUBY SET (see Fig. 41, Attaching RUBY SALINE LINE to RUBY
`SET).
`7. Open roller clamp on the RUBY SALINE LINE.
`8. On the touch screen, select Pre-set Saline Volume (500mL or 1000ml) or Customized Saline
`Volume. The 0.9% USP saline bag must contain more than 50mL.
`9. Change RUBY SALINE LINE with each new install of saline supply.
`
`
`
`
`
`
`
`
`
`
`
` Figure 39, RUBY SALINE LINE Keyhole
`
`
`
`Figure 40, Installing RUBY SALINE LINE into Pump
`
`
`
`Only 0.9% sodium chloride (additive free) injection, USP may be used with the Generator.
`
`
`
`Make sure no clips are pinching the tubes before moving to the next step.
`
`
`
`35
`
`
`
`949 of 1085
`
`

`

`
`
`
`
` 6.8
`
`INSTALLING THE RUBY IV LINE
`
`1. Visually inspect the RUBY IV LINE packaging for damage. Discard if damaged.
`2. Using aseptic techniques, remove cap from B end of RUBY IV LINE and connect to B end of RUBY
`SET (see Fig. 42, Installing RUBY IV LINE).
`3. A new RUBY IV LINE must be used for each patient.
`4. The patient end of the RUBY IV Line (yellow sticker with patient icon) is connected to the patient.
`5. Between Rubidium-82 Chloride injections (rest & stress) and if the RUBY IV LINE is disconnected
`from the patient, engage the clamp to close the line.
`6. After each patient, and at the end of the day, remove and discard used RUBY IV LINE from the
`RUBY SET and discard according to local procedures for potentially biohazardous materials.
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
` Figure 41, Saline Line to RUBY SET Figure 42, Installing RUBY IV Line
`
`Remove and discard used RUBY IV LINE after each patient and at the end of day. Discard
`according to local procedures for potentially biohazardous materials.
`
`
`
`Engage the clamp to the closed position on the RUBY IV LINE if the line is detached from the patient
`and between infusions.
`
`
`
`
`
`950 of 1085
`
`36
`
`

`

`
`7. DAILY QUALITY CONTROL
`
`Daily Quality Control must be performed after Generator Installation and every day prior to patient use.
`Daily Quality Control must be successfully performed before the Patient Infusion tab is available on the
`RUBY Rubidium Elution System (see Fig. 43, Daily Quality Control Needed Screen). The Daily Quality
`Control is accessed through the second row of the task bar. Once a PIN is entered, the following steps
`are required:
`
`
`Figure 43, Daily Quality Control Needed Screen
`
`It is very important to perform a Daily Quality Control immediately after the installation
`of a new RUBY SET and/or Generator to confirm the correct performance of the
`system. RUBY tubing and Generator installation takes approx. 20 minutes to
`complete.
`
`
`
`If the RUBY SET needs to be replaced following the Generator and Tubing installation, leave the
`RUBY CONNECTORS attached the Generator, attach a new set of sterile caps (provided with each
`generator) to the RUBY CONNECTORS to maintain sterility of the Generator until new RUBY SET
`is installed.
`
`
`
`
`
`951 of 1085
`
`37
`
`

`

`
`
`
`
` EMPTY WASTE CONTAINER 7.1
`
`To empty the liquid waste bottle, open the shielded lid and disconnect the tube from the bottle and
`discard the solution according to local regulations. Reinstall the waste bottle; making sure the tube is
`correctly installed to avoid leaks into the well. Close the lid and press Next on the Empty Waste
`Screen (see Fig.44, Empty Waste Screen).
`
`
`
`
`
`Discard the waste solution according to local regulations and procedures for radioactive
`waste.
`
`Failure to empty the waste daily could cause the waste bottle (1L fluid limit) to overflow into the
`waste well. If this occurs, remove the liner and clean waste well per site-specific procedures.
`
`
`
`Figure 44, Empty Waste Screen
`
`
`
`
`
`
`
`952 of 1085
`
`38
`
`

`

`
`
`
`
` 7.2
`
`INSTALL SALINE SUPPLY
`
`Refer to INSTALLING THE 0.9% SODIUM CHLORIDE (ADDITIVE FREE) INJECTION, USP (SALINE
`SUPPLY) & RUBY SALINE IV LINE, section 6.7 (see Fig. 45, Install Saline Supply Screen).
`
`Figure 45, Install Saline Supply Screen
`
`
`
`Make sure no clips are pinching the tubes before moving to the next step.
`
`Only 0.9% Sodium Chloride (additive free), USP Saline may be used with the generator.
`
`Change RUBY SALINE LINE with each new saline supply, or at least daily, and perform Setup
`Verification.
`
`
`
`
`
`
`
`39
`
`
`
`
`
`
`
`
`
`
`
`
`953 of 1085
`
`

`

`
`
`
`
` SETUP VERIFICATION 7.3
`
`After the RUBY SET and Generator installation are complete, the software automatically displays the
`Setup Verification Screen (see Fig. 46, Setup Verification Screen) to check three components for
`normal operation:
`
`
`Pinch Valves: Each valve is activated by the system. Clicks will be heard as each valve is
`checked; this is normal. Check that all tubing remains correctly inserted into each pinch
`valve before proceeding. If an error occurs, contact Jubilant DraxImage Inc.
`
`Pump Flow Rate: Obtain the provided graduated cylinder. The system dispenses a test
`volume with continuous flow rate to calibrate the pump and verify that it is working correctly.
`Install a RUBY IV LINE (Installing RUBY IV LINE; refer to section 6.8), and a sterile 20G
`needle and place tip inside a graduated cylinder to collect the pumped volume, and then
`press Verify Pump. Enter the volume collected in the volume field (mL) on the screen. Press
`Confirm Volume. The % value obtained is normally close to 0. If the error % is >10% the
`user should repeat pump validation. If the volume entered is too far from the targeted
`value, an error is displayed. If an error occurs the user needs to repeat pump validation. If
`the error persists, contact Jubilant DraxImage.
`
`Pressure Transducer Range: Press Verify Pressure and wait for completion. After 10
`seconds the pump runs with increasing speed. If there is an error, the user should check
`that the pressure transducer is connected, verify proper installation of the RUBY tubing and
`repeat the process by pressing Validate Pressure transducer. If the error persists, contact
`Jubilant DraxImage.
`
`
`
`Figure 46, Setup Verification Screen
`
`
`
`The blue bar at the bottom of the screen indicates the progress of each procedure (see Fig. 46,
`Setup Verification Screen). On screen, a green check appears next to each step when each
`verification is complete.
`
`
`
`
`
`
`40
`
`1.
`
`2.
`
`
`
`3.
`
`
`
`954 of 1085
`
`

`

`
`
`
`
`If the supplied graduated cylinder is lost the replacement cylinder must be at least 30ml.
`
`
`
`
`
`
` DAILY DOSE CALIBRATOR CONSTANCY CHECK 7.4
`
`Verify the dose calibrator using a calibrated source. Note that before the first verification, the source
`must be entered in the system as the dedicated reference source (see Settings, see refer to Section
`5.4).
`
`To perform the verification:
`1. Make sure no exterior source is near the elution system to minimize background noise and
`ensure dose calibrator is empty (no radioactive sources).
`2. Press Zero Background on the Check Constancy Screen (see Fig. 47, Check Constancy
`Screen) and wait until the green check mark appears. This can take up to 90 seconds,
`depending on background noise.
`3.
`Insert the calibrated standard in the dose calibrator and wait until the reading stabilizes.
`4. Press Start. A screen should appear to indicate if the verification was successful (see Fig. 48,
`Check Constancy Screen (Successful). If an error occurs, check that the correct calibration
`standard is being used and repeat the constancy check.
`5. Remove the calibrated standard from the dose calibrator.
`
`
`
`
`
`
`
`41
`
`Figure 48, Check Constancy Screen (Successful)
`
`
`
`Figure 47, Check Constancy Screen
`
`
`
`955 of 1085
`
`

`

`
`
`
`
` FLUSH, CALIBRATION & BREAKTHROUGH 7.5
`
`The last step within the Daily Quality Control tab is preparing for the Flush, Calibration, and
`Breakthrough (see Fig. 49, Flush, Calibration, and Breakthrough Screen).
`
`The flush process pumps a precise amount of saline solution through the Generator to rinse off any
`impurities from the column or the lines. The solution is directed to the waste container and can be left
`there to decay in the shielded enclosure. After this procedure, the Generator needs time to fully
`recharge before beginning the calibration process. The user should verify that no leak occurred after
`the flush (during the Generator recharge) and stop the process if needed.
`
`As soon as the system stops directing saline through the Generator, the Generator recharge begins.
`The recharge progress is displayed in the upper right corner of the screen (see Fig. 50, Performing
`Daily Quality Control), The percentage value is the recharge percentage for Rubidium-82 in terms of
`half-life. The time value until recharge is displayed below that. A complete recharge takes ~15
`minutes. The time remaining is also displayed. Calibration does not begin unless a full recharge has
`been reached.
`
`The calibration process runs a measured amount (35mL) of saline through the Generator at a flow rate
`of 20 mL/min to deliver a calibrated sample. This calibrated sample is collected in a vial in the dose
`calibrator and measured to determine the activity. This activity will be used by the system to determine
`the activity available at this point in the life of the Generator and to measure the activity delivered in
`patient infusions.
`The breakthrough test uses the sample produced during the calibration process. This portion of the
`Daily Quality Control takes 30 minutes, during which the Rb-82 decays completely, and the system
`assays the calibration sample and measures the amount of strontium-82 and strontium-85 present in
`the sample. This information is saved in the calibration report (refer to section 8.1, Reports).
`
`Follow these steps before initiating Start for the Flush, Calibration and Breakthourgh.
`
`
`1. Obtain a 50 ml sealed glass vial (with rubber stopper)
`2. Aseptically install a RUBY IV LINE on the B needleless injection port of the RUBY SET
`3. Aseptically install a sterile needle (20G) on the end of the RUBY IV LINE and insert into rubber
`stopper of glass vial
`4. Insert a sterile vent needle (20G) into rubber stopper of glass vial
`5. Place the vial into dose calibrator dipper and lower into the dose calibrator chamber
`6. Press Start to begin the 50-minute procedure (see Fig. 49, Flush, Calibration, and
`Breakthrough Screen).
`
`
`The word Successful (in green) appears next to each item as it is completed. This information is
`saved in the Daily QC report. If an error occurs, a red 'X' appears and that process is aborted. If this
`occurs, the user needs to repeat this process before proceeding to the next step (see Fig. 50,
`Performing Daily Quality Control).
`
`
`
`
`956 of 1085
`
`42
`
`

`

`
`
`
`
`Figure 49, Flush, Calibration, and Breakthrough Screen
`
`
`Figure 50, Performing Daily Quality Control (Flush and Calibration
`are completed and the Breakthrough Check is in progress)
`
`The yellow progress bar indicates that a radioactive procedure is in progress and that radioactive
`solution is flowing through unshielded tubes. The user should remain at a safe distance from the
`elution system.
`
`
`
`When the breakthrough check is complete, a label with the daily QC values automatically prints as
`configured in the settings. The user can either press Reprint Labels or press Finish.
`
`When the user selects Finish, the Ready for Patient Infusion Screen appears (see Fig. 51, Ready for
`Patient Infusion Screen). This screen displays important information about the state of the elution
`system at a glance. The most important information displayed on this screen is:
`
`
`(cid:120) Breakthrough: Indicated as a percentage of the USP limit* (see Table 3), this field indicates the
`level of impurities calculated in the Daily Quality Control.
`(cid:120) Breakthrough History: The graph tracks the breakthrough levels for each day of use of the
`installed generator.
`(cid:120) Below the yellow line (<20% of USP limit*, see Table 3), the system can be used without any
`restriction on the number of infusions per Quality Control.
`(cid:120) An alert limit is reached when the reading is above the yellow line ((cid:149) 20%) and below
`the red line (<50% of the USP limit*, see Table 3). Alert limits may be triggered by the
`following:
`
`o 20 L of saline eluted through the Generator.
`o or an eluate Sr-82 level of (cid:149)0.004 μCi/mCi Rb-82. and (cid:148) 0.01 μCi/mCi Rb-82.
`o or an eluate Sr-85 level of (cid:149)0.04 μCi/mCi and (cid:148) 0.1 μCi/mCi Rb-82.
`
`
`
`
`
`
`
`957 of 1085
`
`
`
`43
`
`

`

`
`PASS
`< 20% of USP limits*
`(Green)
`
`Breakthrough level is low.
`
`ALERT
`(cid:149) 20% and <50% of USP limits*
`OR 20L volume limit
`(Yellow)
`
`Breakthrough level is increased.
`
`FAIL
`(cid:149) 50% of USP limits*
`OR 30L volume limit
`(Red)
`Breakthrough level is approaching the
`allowable limit.
`The Daily Quality Control (automated
`breakthrough test) does not allow a
`sufficient margin of safety to continue the
`elutions (scans).
`The use of the RUBY-FILL® Rubidium Rb
`Repeat an automated Daily Quality Control
`82 Generator must be discontinued
`after every 4 patients (8 scans) and record
`immediately.
`the results
`Contact Jubilant DraxImage:
`Contact Jubilant DraxImage:
`1-888-633-5343
`1-888-633-5343
`*USP limits: <0.02μCi of Sr-82/mCi of Rb-82; <0.2μCi of Sr-85/mCi of Rb-82
`Table 3: Strontium Breakthrough Results
`
`The Daily Quality Procedure (automated
`breakthrough test) is valid for a 24 hour
`period.
`
`The Daily Quality Procedure (automated
`breakthrough test) is valid for 4 patients only.
`
`Proceed with use
`
`
`
`
`
`
`
`
`
`The system can be used with four (4) patients before a Quality Control procedure must be performed if
`the breakthrough reaches an alert limit. If the user repeats the flush, the system counts this as a patient
`infusion.
`
`
`(cid:120) (cid:149) 50% of the USP limit* (see Table 3), the system does not allow the user to perform a patient
`infusion.
`(cid:120) Refer to Table 3 below for instructions to follow on Strontium Breakthrough results.
`
`
`
`Remove and discard the calibration
`vial according to local procedures
`and regulations. Use a clean vial for
`each Daily Quality Control.
`
`After the initial Daily Quality Control
`following Generator
`installation,
`visually inspect the RUBY SET and
`all needleless injection ports for any
`leaks.
`
`Figure 51, Ready for Patient Infusion Screen
`
`
`
`
`
`Generator should not be used if any of the following are reached: (1) 50% of Sr-82/ Sr-85 USP limits,
`(2) 30 L volume limit or (3) Generator expiration date
`
`44
`
`
`
`
`
`958 of 1085
`
`

`

`
`
`
`
` PREPARING FOR PATIENT INFUSIONS 7.6
`
`Upon completion of Daily Quality Control, in preparation for Patient Infusions, close the front door and
`top panel of the RUBY Rubidium Elution System. The patient portion of the RUBY SET is placed
`outside of the RUBY Rubidium Elution System through the notch in the top panel for accessibility to
`change the RUBY IV LINE.
`
`An absorbency pad may be placed on top of the RUBY Rubidium Elution System to place the RUBY IV
`LINE and other materials in preparation for patient setup.
`
`
`
`
`
`959 of 1085
`
`Do not use the top of the elution system as a work desk.
`
`
`
`45
`
`

`

`
`8. PATIENT INFUSIONS
`
`To access the patient infusion mode, press Patient Infusion on the second row of the task bar. Enter
`PIN and

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket