throbber
Hermann Mayer*
`Department of Robotics and
`Embedded Systems
`Technische Universita¨t Mu¨nchen
`D-85748 Garching bei Mu¨nchen
`and Department of
`Cardiovascular Surgery
`German Heart Center Munich
`D-80636 Mu¨nchen
`
`Istvan Nagy
`Alois Knoll
`Department of Robotics and
`Embedded Systems
`Technische Universita¨t Mu¨nchen
`D-85748 Garching bei Mu¨nchen
`
`Eva U. Braun
`Robert Bauernschmitt
`Ru¨ diger Lange
`Department of Cardiovascular
`Surgery
`German Heart Center Munich
`D-80636 Mu¨nchen
`
`Haptic Feedback in a Telepresence
`System for Endoscopic Heart
`Surgery
`
`Abstract
`
`The implementation of telemanipulator systems for cardiac surgery enabled heart
`surgeons to perform delicate minimally invasive procedures with high precision un-
`der stereoscopic view. At present, commercially available systems do not provide
`force-feedback or Cartesian control for the operating surgeon. The lack of haptic
`feedback may cause damage to tissue and can cause breaks of suture material. In
`addition, minimally invasive procedures are very tiring for the surgeon due to the
`need for visual compensation for the missing force feedback. While a lack of Carte-
`sian control of the end effectors is acceptable for surgeons (because every move-
`ment is visually supervised), it prevents research on partial automation. In order to
`improve this situation, we have built an experimental telemanipulator for endo-
`scopic surgery that provides both force-feedback (in order to improve the feeling
`of immersion) and Cartesian control as a prerequisite for automation. In this article,
`we focus on the inclusion of force feedback and its evaluation. We completed our
`first bimanual system in early 2003 (EndoPAR Endoscopic Partial Autonomous Ro-
`bot). Each robot arm consists of a standard robot and a surgical instrument, hence
`providing eight DOF that enable free manipulation via trocar kinematics. Based on
`the experience with this system, we introduced an improved version in early 2005.
`The new ARAMIS system (Autonomous Robot Assisted Minimally Invasive Surgery)
`has four multi-purpose robotic arms mounted on a gantry above the working
`space. Again, the arms are controlled by two force-feedback devices, and 3D vision
`is provided. In addition, all surgical instruments have been equipped with strain
`gauge force sensors that can measure forces along all translational directions of the
`instrument’s shaft. Force-feedback of this system was evaluated in a scenario of ro-
`botic heart surgery, which offers an impression very similar to the standard, open
`procedures with high immersion. It enables the surgeon to palpate arteriosclerosis,
`to tie surgical knots with real suture material, and to feel the rupture of suture ma-
`terial. Therefore, the hypothesis that haptic feedback in the form of sensory substi-
`tution facilitates performance of surgical tasks was evaluated on the experimental
`platform described in the article (on the EndoPAR version). In addition, a further
`hypothesis was explored: The high fatigue of surgeons during and after robotic op-
`erations may be caused by visual compensation due to the lack of force-feedback
`(Thompson, J., Ottensmeier, M., & Sheridan, T. 1999. Human Factors in Telesurgery,
`Telmed Journal, 5 (2) 129 –137.).
`
`Presence, Vol. 16, No. 5, October 2007, 459 – 470
`© 2007 by the Massachusetts Institute of Technology
`
`*Correspondence to mayerh@in.tum.de
`
`Mayer et al. 459
`
`Ethicon Exhibit 2017.001
`Intuitive v. Ethicon
`IPR2018-01254
`
`

`

`460 PRESENCE: VOLUME 16, NUMBER 5
`
`tering, and a stereo vision interface at the input console.
`Surgeons can now operate using a surgical mechatronic
`assistant in a comfortable, dextrous, and intuitive man-
`ner (Falk, Jacobs, Gummert, & Walther, 2003; Falk,
`Jacobs, Gummert, Walther, & Mohr, 2003; Falk,
`Mintz, Grunenfelder, Fann, & Burdon, 2001). Despite
`the obvious potential advantages of robot-assisted, en-
`doscopic surgery, most researchers and surgeons in this
`area agree that the lack of haptic feedback is the most
`important drawback of currently available systems (Mit-
`suishi, Tomisaki, Yoshidome, Hashizume, & Fujiwara,
`2000). The inability of the operator to sense the applied
`forces causes increased tissue trauma and frequent su-
`ture material damage. The systems are telemanipulators
`with no Cartesian position control (the control loop is
`implicitly closed by the visual surveilance of the surgeon).
`Both features are important in order to move the surgeon
`up in the control hierarchy, that is, to implement “shared
`control” or “partial autonomy.”
`In order to overcome these deficiencies, two crucial
`issues have to be solved. One is inclusion of force sen-
`sory and feedback, and the other is the implementation
`of full Cartesian control of the end effector. The latter is
`indispensable for calculating exact directions of forces in
`a known coordinate frame. Therefore, one of our main
`research interests is the construction and evaluation of
`force sensory/feedback in realistic scenarios of robotic
`surgery. In particular, we focus on instrumental suturing
`and knot-tying tasks, which are very time consuming if
`performed via telemanipulation. Our working hypothe-
`sis was that the handling of telemanipulated surgical
`systems can be significantly improved by the inclusion of
`force-feedback. Therefore, we focus below on hardware
`and software for force-feedback in endoscopic surgery
`and present an evaluation of its quality.
`
`2
`
`Related Work
`
`Telemanipulators for endoscopic surgery are al-
`ready commercially available. Systems, such as daVinci
`(Guthart & Salisbury, 2000), offer comfortable user
`interface and are used in daily practice to perform even
`delicate operations. However, they offer no force-
`
`Figure 1. Location of the instrument and camera port.
`
`1
`
`Introduction
`
`Recently, minimally invasive surgery (MIS) has
`become a promising option for a great number of surgi-
`cal interventions (such as heart surgery) and has had
`significant impact on both patients and surgeons. Mini-
`mally invasive and endoscopic cardiac surgery not only
`minimizes the collateral surgical trauma, but it also re-
`sults in quicker recovery. The length of the hospital stay
`and the infection rate can be reduced (Morgan et al.,
`2004). Therefore, patients can profit significantly from
`this treatment option. On the other hand, surgeons
`have to cope with increasingly complex working condi-
`tions. Since endoscopic surgery is performed through a
`small port or keyhole in the patient’s chest (Figure 1),
`surgeons must learn to operate with unfamiliar and of-
`ten awkward surgical instruments. Hence, the tech-
`niques of endoscopic surgery have been applied infre-
`quently, particularly in the field of heart surgery. An
`important step in developing this technology was the
`introduction of telemanipulation, which was especially
`designed to enable delicate interventions with high sur-
`gical precision. The surgeon no longer controls the in-
`struments directly, but they are controlled by a special
`device with a Cartesian user interface that surgeons can
`handle as usual, that is, like instruments for open sur-
`gery. They offer as much freedom of movement as the
`surgeon’s own hand would in conventional open sur-
`gery, thus providing 6 DOF instead of the 4 DOF that
`conventional endoscopic instruments have. In addition,
`they assist the surgeon with motion scaling, tremor fil-
`
`Ethicon Exhibit 2017.002
`Intuitive v. Ethicon
`IPR2018-01254
`
`

`

`Mayer et al. 461
`
`Figure 2. Master console and magnified view of a customized PHANToM stylus.
`
`feedback and their control loop is closed visually by the
`surgeon. In addition to these commercial systems, a
`variety of devices for robotic surgery do exist, imple-
`mented by other research groups. At the University of
`California, Berkeley, a robotic system was developed
`that has already been used to perform certain surgical
`tasks, such as suturing and knot-tying (Cavasoglu, Wil-
`liams, Tendick, & Sastry, 2003). The Korean Advanced
`Institute of Science and Technology developed a micro-
`telerobot system that provides force-feedback (Kwon,
`Woo, Song, Kim, & Cho, 1998). In Germany, two sys-
`tems for robotic surgery were built at the Research Fa-
`cility in Karlsruhe (Voges, Holler, Neisius, Schurr, &
`Vollmer, 1997) and at the DLR in Oberpfaffenhofen
`(Konietschke, Ortmaier, Weiss, & Hirzinger, 2004).
`While the first system provides no force-feedback, the
`latter is equipped with PHANToM devices for haptic
`display. There is also some work available dealing with
`analysis of knot-tying. At Johns Hopkins University,
`Kitagawa, Okamura, Bethea, Gott, and Baumgartner
`(2002) have evaluated forces occurring during knot-
`tying. They did not measure forces directly on the in-
`struments or during realistic operations, but with a con-
`trivance that was especially designed for these
`measurements. Wagner, Vasilyev, Perrin, del Nido, and
`Howe (2006) have proposed an instrument with force-
`feedback in order to provide additional modality for
`ultrasound-guided interventions. Tavakoli, Patel, and
`Moallem (2005) developed a novel endoscopic instru-
`ment with exchangeable head and force-feedback. Their
`
`evaluation showed that surgical tasks profit from force
`feedback, regarding performance time and reduction of
`fatigue. Cao, MacKenzie, and Payandeh (1996) ana-
`lyzed a variety of surgical tasks (among other things,
`knot-tying) and broke them down into subtasks. They
`did not include force measurement. However, the ne-
`cessity of haptic feedback in robotic surgery has been
`discussed controversially by surgeons and haptic engi-
`neers (Bethea et al., 2004; Fager, 2004; Hu, Tholey,
`Desai, & Castellanos, 2002; MacFarlane, Rosen, Han-
`naford, Pellegrini, & Sinanan, 1999). At the Rensselaer
`Polytechnic Institute, a robotic system capable of knot-
`tying was developed (Kang & Wen, 2001), but they
`mainly focused on force control and have used dedi-
`cated instruments for knot-tying. At the German Heart
`Center in Munich, a daVinci system has been installed.
`Surgeons often ponder the question of whether perfor-
`mance could be improved if force-feedback is included.
`Since this problem cannot be evaluated directly within
`the daVinci system, we decided to take the original in-
`struments and incorporate them into an experimental
`setup that can be used for evaluation.
`
`3 Materials and Methods
`
`Like typical systems for robotic surgery, our setup
`consists of an operator-side master console (Figure 2)
`for in-output and a patient-side robotic manipulator
`that directly interacts with the operating environment.
`
`Ethicon Exhibit 2017.003
`Intuitive v. Ethicon
`IPR2018-01254
`
`

`

`462 PRESENCE: VOLUME 16, NUMBER 5
`
`Figure 3. New system ARAMIS with four ceiling mounted robots.
`
`3.1 Robotic Telemanipulator
`
`The ARAMIS system consists of four small robots
`mounted on the ceiling (Figure 3). The robots have 6
`DOF. Since the rotation of the robot’s flange and the
`rotation of the instrument share one axis, our system
`ultimately ends up having 8 DOF under the restriction
`of trocar kinematics (Mayer, Nagy, & Knoll, 2004).
`The crucial part of the master console is two PHANToM
`haptic interfaces. They are used to control the instruments
`and to feed back forces. Particular advantages of this
`setup with multipurpose robots are high precision and
`stiffness, moderate costs, and an advanced dynamic be-
`havior. The latter could be exploited to perform sophis-
`ticated tasks in motion compensation, for example, sup-
`port for beating heart surgery as it was proposed in
`Ortmaier, Groeger, Boehm, Falk, and Hirzinger (2005),
`or compensation for respiratory motion of the ribs.
`We have also developed adapters that are attached to
`the robotic arms and can be equipped with either an
`instrument or a stereo endoscope. For security reasons
`and better handling, we have equipped all flange adapt-
`ers with magnetic security couplings (Figure 4c). Those
`will disengage if forces beyond a certain level are exerted
`that might cause harm to the instruments or chest
`mockup. Each of the surgical instruments has 3 DOF. A
`microgripper at the distal end of the shaft can be ro-
`
`Figure 4. Sensorized instrument, magnetic adapter, and servos.
`
`tated, and adaptation of pitch and yaw angles is possi-
`ble. All movable parts of the gripper are driven by steel
`wires. Their motion is controlled by four driving wheels
`at the proximal end of the instrument, one for each de-
`gree of freedom (two for the yaw of the fingers). In or-
`der to control the joints of the instruments, we have
`flanged servomotors to each driving wheel by means of
`an Oldham coupling (Figure 4c). This guarantees in-
`strument movement that is free of jerk. The servocon-
`trollers can be connected via serial lines or a CANbus to
`a multiport card. The basic idea of endoscopic surgery is
`that only small openings have to be made into the sur-
`face of the patient’s body (so-called keyholes, Figure 1).
`Therefore, translational movements of the instruments
`are essentially restricted by shifts and rotations within
`these holes. In order to provide the surgeon with a
`comfortable user interface, it is desirable to map move-
`ments of the stylus of the input device directly onto in-
`strument motions. Considering these requirements, we
`have implemented the inverse kinematics of our system
`as shown in Figure 1.
`A desired position of the instrument is given by the
`position of the input stylus. Arbitrary positions of the
`
`Ethicon Exhibit 2017.004
`Intuitive v. Ethicon
`IPR2018-01254
`
`

`

`Mayer et al. 463
`
`Figure 5. Endoscopic stereo camera with customized magnetic adapter.
`
`instrument’s tip are mapped on corresponding positions
`of the motors that control the 8 DOF. The input data is
`represented by a homogenous transform matrix. Since
`the position of the instrument’s shaft is restricted by the
`port (the position of the keyhole), only one possibility
`for aligning the instrument exists. The angle of the cor-
`responding joints of the instrument can be found by
`geometric calculations, which are explained in detail in
`Mayer et al. (2004). In order to speed up this complex
`computation, the complete inverse kinematics of the
`robot can be performed on the graphical processing unit
`(GPU) of a graphics card. In order to prepare this com-
`putation, the desired Cartesian positions are transferred
`to a texture on the graphics card. A so-called fragment
`shader, which is written in the OpenGL Shading Lan-
`guage and contains the algorithm for inverse kinematics,
`is applied to the texture. The results (the joint angles of
`the robot) are transferred to the framebuffer of the card
`and can be moved back to the CPU. The calculated an-
`gles can be directly applied to the robot. Employment
`of a GPU (NVIDIA GeForce 5200 graphics card) leads
`to a computation time that is 20 times faster than a
`modern CPU (Athlon64 2200 MHz).
`As mentioned above, the master console (the work-
`station of the surgeon) consists of custom made modifi-
`cations of PHANToM force feedback devices (see below).
`
`3.2 Optical System
`
`Due to the extremely high-precision requirements
`in modern minimally-invasive surgery (arteries in heart
`surgery have a diameter between 1 and 2 mm), it is im-
`perative to provide an accurate 3D view.
`
`Most of the other systems mentioned above come
`with magnifying three-dimensional endoscopes and ap-
`propriate display devices. This is also an essential prereq-
`uisite for high accuracy machine vision. The ARAMIS
`system is equipped with a 3D endoscope providing two
`separate optical (fiberglass) channels and two synchro-
`nized CCD cameras. A magnetic coupling mechanism
`(Figure 5) allows quick mounting and dismounting
`without losing the calibration against the rest of the sys-
`tem. We have precisely determined the intrinsic and ex-
`trinsic parameters of the camera system in order to
`achieve the necessary calibration for scene reconstruc-
`tion. Like the instruments, the endoscopic camera can
`be moved by means of trocar kinematics and can either
`be actively controlled by the operator or automatically
`guided by the system (following the region of interest,
`i.e., where the instruments are placed). Images taken
`from the stereo camera system are displayed on a 3D
`device, which provide differentially-polarized images for
`right and left eye. Therefore, the surgeon has to wear
`polarized lenses in order to get a 3D view.
`
`3.3 Force Feedback
`
`Positions and orientations of the instruments are
`controlled by two PHANToM devices (Figure 2). This
`device is available in different versions with different
`capabilities. We have chosen the version PHANToM
`Premium 1.5. It has a working space of approximately
`20 ⫻ 25 ⫻ 40 cm, which is appropriate for performance
`of surgical procedures. The user controls a stylus pen
`that is equipped with a switch that can be used to open
`
`Ethicon Exhibit 2017.005
`Intuitive v. Ethicon
`IPR2018-01254
`
`

`

`464 PRESENCE: VOLUME 16, NUMBER 5
`
`and close the microgrippers. We have extended the
`plain stylus with sleeves for thumb and forefinger in or-
`der to increase the user’s comfort (Figure 2).
`The most interesting feature of the employed
`PHANToM devices is the capability to display forces
`which are fed back by small servomotors incorporated
`into the device. They are used to steer the stylus pen in
`a certain direction. This creates the impression of occur-
`ring forces, while the user is holding the pen in a certain
`position. Our version of the PHANToMTM device can
`display forces in all translational directions, while no
`torque is fed back. In order to be able to display realistic
`forces during operation, we have equipped the instru-
`ments with force sensors.
`Since the shaft of the surgical instrument is made of
`carbon fiber, force sensors have to be very sensitive and
`reliable. Therefore, we decided to apply strain-gauge
`sensors, which are employed in industrial force registra-
`tion (Figure 4b).
`Strain-gauge sensors are usually made of silicon,
`which provides by far the best sensitivity with respect to
`elongations. Its piezoelectric resistance increases under
`dilation and decreases under compression. The resulting
`resistance R is linearly dependent on the axial stress
`(␴ ⫽ F/A), the relative elongation (␧ ⫽ ⌬l/l) of the re-
`sistor and its transversal contraction (⌬r ⫽ ⫺␮r␧),
`where ␮ is the Poisson ratio, a material-dependent con-
`stant. For small, reversible elongations (␧ ⬍ 10⫺3) the
`following equation holds:
`
`␴␧册␧ ⫽K 䡠 ␧
`
`⌬␴
`
`⌬R
`R
`
`⫽
`
`⌬l
`l
`
`⫺
`
`2⌬r
`r
`
`⫹
`
`⌬␴
`␴
`
`⫽冋1 ⫹ 2␮⫹
`
`(1)
`
`Therefore, resistance depends only on the elongation ␧
`and a material-dependent constant K, which denomi-
`nates the sensitivity of the material and accounts for a
`dimension of 100 for silicon. These strain gauges are
`very compact, cheap, and have fast reactions. We ap-
`plied the sensor gauges at the distal end of the instru-
`ment’s shaft, that is, near the gripper. This results in an
`inevitable disadvantage that sensor readings are flawed
`by the tension of the driving wires. This is particularly
`an issue of measurements in axial direction when the
`
`gripper is opened and closed. For that reason, we dis-
`abled this axis for our experiments and only took into
`account lateral movements. Due to their layout as full
`bridges (see below), they are not negatively impacted by
`the closing state of the gripper. However, since disturb-
`ing forces still can exceed an acceptable limit during
`extensive movements (e.g., knot-tying), we provided a
`switch to reduce/remove force levels in situations when
`advanced sensitivity is demanded. As mentioned above,
`the sensors were directly attached to the instruments by
`a heat-resistant, two-component adhesive and have been
`coated with shrink tubes. Therefore, this technology is
`waterproof and heat-resistant, in order to survive the
`medical autoclave. Since we did not know the exact ma-
`terial parameters of the instruments, we calibrated the
`sensor system with standard weights.
`In order to make sensor readings more stable, we em-
`ployed four sensor gauges interconnected as a Wheat-
`stone full bridge for each translational direction. Nor-
`mally, there are three possibilities to arrange the sensor
`gauges as a full bridge. We have chosen the layout de-
`picted in Figure 6 (left side) with all four gauges mea-
`suring elongations along the main axis. Sensor readings
`are independent of thermal fluctuations, since resulting
`changes of single sensors eliminate each other. In addi-
`tion, the chosen layout has the advantage that the re-
`sulting resistance is still linearly correlated with elonga-
`tions. Other possibilities presume that two of the
`sensors are arranged perpendicular to the axial direction
`in order to measure lateral strains. But they are either
`nonlinear or not resistant to changes in temperature.
`For measurement of x- and y-forces the bridge parts
`for dilation (1 and 3 in Figure 6) and compression (2
`and 4 in Figure 6) are attached on opposite areas of the
`shaft (Figure 7 left). Therefore the forces measured in
`these directions are independent from axial elongations
`as they may occur during the opening and closing of the
`gripper (induced by the steel wires inside the shaft). For
`digitizing the sensor readings, we have used GSV3CAN
`measurement amplifiers from ME-Messsysteme GmbH.
`The blank with the main components is very compact
`(Figure 7: blank size 30 ⫻ 55 mm) and can be attached
`directly onto the instruments. It also provides a CAN-
`bus interface which can transmit 1220 16 bit values per
`
`Ethicon Exhibit 2017.006
`Intuitive v. Ethicon
`IPR2018-01254
`
`

`

`Mayer et al. 465
`
`Figure 6. Layout of strain gauge sensors: four active strain gauges (left), two active and two oblique
`strain gauges in nonlinear arrangement (middle), two active and two oblique strain gauges in linear
`arrangement (right).
`
`Figure 7. Arrangement of the sensor gauges (left), measurement amplifier (right).
`
`second. This is necessary since haptic inter-action with
`stiff contacts requires a sampling rate of at least 1000
`Hz. For internal processing, the amplifier provides a
`sampling rate of 10 kHz. Therefore, it calculates the
`medium of 8 sensor evaluations and transmits them
`thereafter. Since we know the position and orientation
`of the grippers, we can transform the forces into the
`coordinate frame of the PHANToMTM devices. In addi-
`tion, forces can be arbitrarily amplified by our control soft-
`ware and recorded to data files for subsequent evaluation.
`
`3.4 Evaluation of Force-Feedback
`
`3.4.1 Human Participants and Haptic Levels.
`In order to test the quality of force-feedback and the
`overall handling of the system, we have conducted a
`comprehensive evaluation of our setup. The subjects of
`
`this study were 25 heart surgeons of varying levels of
`surgical training and age (Table 1). We divided them
`into three groups regarding their skill level.
`The study group consisted of one group of eight
`young surgeons, another group of twelve experienced
`surgeons, and a third group of five surgeons with ro-
`botic experience. The study included basic surgical and
`cardiac surgical procedures. A psychophysical scaling
`pretest of static force was performed beforehand, to
`evaluate the alignment of the implementation of force-
`feedback in the telepresent robotic system to human
`perception. This pretest demonstrated that due to indi-
`vidual differences, two different haptic levels—1:1 and
`1:2 haptics— had to be transferred to and used in the
`evaluation. The order of haptic conditions (type of tasks
`and haptic level) was completely balanced to avoid
`learning effects and was double blinded.
`
`Ethicon Exhibit 2017.007
`Intuitive v. Ethicon
`IPR2018-01254
`
`

`

`466 PRESENCE: VOLUME 16, NUMBER 5
`
`Table 1. Random Sample of the Human Subjects
`
`All participants
`n⫽25
`
`Young surgeons
`n⫽8
`
`Experienced
`surgeons
`n⫽12
`
`Robotic surgeons
`n⫽5
`
`7.4 (⫾6.9)
`
`0.4 (⫾0.3)
`
`11.5 (⫾7.6)
`
`9.0 (⫾5.2)
`
`36 (⫾8)
`
`29 (⫾7)
`
`39 (⫾8)
`
`39 (⫾5)
`
`0
`
`0
`
`0
`
`2.4 (⫾2.0)
`
`Surgical experience (years)
`mean value ⫾ standard
`deviation
`Age (years) mean value ⫾
`standard deviation
`Robotic experience (years)
`mean value ⫾ standard
`deviation
`
`3.4.2 Surgical Tasks. 3.4.2.1 Knot-tying. The
`human subjects had to tie surgical knots with two surgi-
`cal instruments equipped with haptic feedback. The sur-
`geons had ten minutes to perform as many knots as pos-
`sible in alternating manner (left and right tied knots).
`The total number of knots, the applied forces, the break-
`age of suture material, and the speed of motion during
`knot-tying have been recorded.
`
`3.4.2.2 Breaking of Suture Material. The breaking
`of suture material represents the amount of telepresence
`and immersion of the robotic system for the surgeons,
`and was considered an indicator for the technical quality
`of force feedback. The surgeons had to strain the thread
`until an assumed breaking point was reached at which
`point they had to indicate imminent breaking. The dif-
`ference of force between the indicated point and the
`actual breaking point of the thread was measured in
`Newtons. The surgical suture material used, ProleneTM
`6-0 (Ethicon Inc., Somerville, NJ, USA), is a common
`and frequently used nonabsorbable thread used for
`heart surgery.
`
`3.4.2.3 Detection of Arteriosclerosis. The subjects
`had to detect a possible stenosis with one haptic instru-
`ment (dominant hand), in artificial arteries made from
`polymer (Figure 8), precisely and at the same time rap-
`idly. Detection error, regarding the existence or size of
`
`the stenoses, have been counted. In addition, the ap-
`plied forces during detection have been recorded and
`the time needed for detection in seconds.
`
`3.4.3 Workflow and Critical Flicker Fusion
`Frequency (CFF). The critical flicker fusion frequency
`(CFF) is an individual part of the Wiener Testsystem
`(Schuhfried GmbH, Mo¨dling, Austria) analyzing the
`progression of fatigue during the evaluation (Wiemeyer,
`2002). The CFF is regarded as an indicator of the cen-
`tral-nervous function capacity, the activation level, and
`the progression of fatigue during practical tasks (Johans-
`son & Sandstroem, 2003). The CFF is defined as me-
`dian of the fusion frequency in the ascending and de-
`scending process.
`
`3.4.4 Training Skills. Before performing the
`actual tasks (knot-tying, breaking, and detecting), the
`surgeons underwent a short training program to be-
`come familiar with the robotic system. They had to
`move small objects with the grippers and thread a
`rubber band through narrow eyelets (Figure 9). Af-
`terwards, all tasks were repeated with three different
`levels of haptic feedback: absence of feedback, feed-
`back with real measured forces, and amplified force-
`feedback (factor two). Each surgeon had to perform
`each task at all force levels (Table 2). Since tasks had
`to be performed in dedicated task-sequences with
`
`Ethicon Exhibit 2017.008
`Intuitive v. Ethicon
`IPR2018-01254
`
`

`

`Mayer et al. 467
`
`Figure 8. Mockup for arteriosclerosis detection.
`
`Figure 9. Workplate with eyelets for training.
`
`double blinding, the sequence of force levels for each
`task was chosen randomly.
`
`4
`
`Results
`
`With the setup described above, we had already
`performed automized procedures like knot-tying (Nagy,
`Mayer, Knoll, Schirmbeck, & Bauernschmitt, 2004;
`Bauernschmitt et al., 2004). The focus of this article is
`on the evaluation of the impact of force measurement
`and feedback on surgical procedures. Our hypothesis,
`which has been proven correct (see below), was that
`haptic feedback contributes to better performance of
`systems for robotic surgery by preventing force-induced
`damage.
`
`4.1 Evaluation of the System
`
`The experiments which were conducted with the
`surgeons from the German Heart Center in Munich
`have shown the following results: increasing the amplifi-
`cation of force feedback during knot-tying reduces the
`applied forces significantly (p ⬍ .05, Figure 10a), but
`does not cause significant change in the performance
`time.
`During breaking of suture material, the correspond-
`ing differences of forces were calculated as described
`above. Increasing the magnitude of haptic feedback
`leads to a decrease in this difference (p ⬍ .05, Figure
`10b). This militates in favor of the precision of the esti-
`mated force when the thread was breaking. And it
`
`Table 2. Workflow of the Evaluation: (*) Randomly Presented Sequence of Force-Feedback Levels
`
`*Force-feedback levels were absence of feedback, feedback with real measured forces, and amplified force-feedback.
`
`Ethicon Exhibit 2017.009
`Intuitive v. Ethicon
`IPR2018-01254
`
`

`

`468 PRESENCE: VOLUME 16, NUMBER 5
`
`Figure 10. Forces while knot-tying (left); with increasing haptic feedback the applied forces decrease significantly (right: *p ⬍ .05). The
`difference of the assumed and the real force while breaking a surgical suture decreases significantly with haptic feedback (*p ⬍ .05).
`
`Figure 11. Comparison of surgeons with different skill levels in detection of arteriosclerosis and the surgeon’s fatigue.
`
`shows that the telemanipulator system provides a high
`level of immersion. Similar results have been found for
`detection of arteriosclerosis. Haptic feedback signifi-
`cantly influences the amount of applied forces: an in-
`crease in the amplification of force-feedback leads to a
`
`significant decrease of applied forces. However, haptic
`feedback has no significant influence on detection errors
`(p ⫽ .05, Figure 11a) and performance time. And lastly,
`with regard to fatigue, operating with haptic feedback
`leads to a significant decrease (p ⬍ .05, Figure 11b).
`
`Ethicon Exhibit 2017.010
`Intuitive v. Ethicon
`IPR2018-01254
`
`

`

`Mayer et al. 469
`
`5
`
`Discussion and Conclusion
`
`The goal of these experiments was to examine
`claims about the necessity of force-feedback for robot-
`assisted surgical procedures in cardiac surgery. We
`present an approach of evaluating haptic feedback with a
`novel robotic system for minimally invasive and endo-
`scopic surgery. Haptic feedback is needed for surgical
`tasks since less force is applied by the surgeon. The
`number of tied knots does not increase, but the experi-
`ments have shown that haptic feedback can be em-
`ployed to prevent the surgeon from potentially harmful
`mistakes, such as the breakage of suture material and
`the loss of the surgical needle. The fatigue level of the
`surgeon decreases and the perception of telepresence by
`the surgeons increases. The safety level of the patient
`increases when being operated on by a telemanipulator
`with integrated haptic. These results lead to the conclu-
`sion that haptic feedback is an indispensable feature for
`surgical telemanipulators, in particular for operations
`with delicate suture material, such as is used in heart
`surgery. To further research and confirm the findings of
`these experiments, the next step would be to examine
`an animal model in order to further define surgical quality.
`
`Acknowledgments
`
`This work is supported by the German Research Foundation
`(DFG) within the Collaborative Research Center SFB 453-I4
`on “High-Fidelity Telepresence and Teleaction” and by the
`German Heart Center Munich, Department of Cardiovascular
`Surgery.
`
`References
`
`Bauernschmitt, R., Schirmbeck, E. U., Knoll, A., Mayer, H.,
`Nagy, I., Lange, R., et al. (2004). Towards robotic heart
`surgery: Introduction of autonomous procedures into an
`experimental surgical telemanipulator system. The Interna-
`tional Journal of Medical Robotics and Computer Assisted
`Surgery, 1(3), 74 –79.
`Bethea, B. T., Okamura, A. M., Kitagawa, M., Fitton, T. P.,
`
`Cattaneo, S. M., Gott, V. L., et al. (2004). Application of
`haptic feedback to robotic surgery. Journal of Laparoendo-
`scopic Advances in Surgical Technique, 14(3), 191–195.
`Cao, C., MacKenzie, C., & Payandeh, S. (1996). Task and
`motion analyses in endoscopic surgery. Proceedings of ASME
`Dynamic Systems and Control Division (Fifth Annual Sympo-
`sium on Haptic Interfaces for Virtual Environment and Tele-
`operator Systems), 583–590.
`Cavasoglu, M., Williams, W., Tendick, F., & Sastry, S. S.
`(2003). Robotics for telesurgery: Second generation Berke-
`ley/UCSF laparoscopic telesurgical workstation and looking
`towards the future applications. Industrial Robot, Special
`Issues on Medical Robotics, 30(1), 22–29.
`Fager, P. J. (2004). The use of haptics in medical applications.
`International Journal of Medical Assisted Surgery, Robotic
`Computer 1(1), 36 – 42.
`Falk, V., Jacobs, S., Gummert, J., & Walther, T. (2003). Ro-
`botic coronary artery bypass grafting (CABG)—The Leipzig
`experience. Surgical Clinician North America, 83(6), 1381–
`1386.
`Falk, V., Jacobs, S., Gummert, J., Walther, T., & Mohr, F. W.
`(2003). Computer-enhanced endoscopic coronary artery
`bypass grafting: The da Vinci experience. Seminal Thoracic
`Cardiovascular Surgery, 15(2), 104 –111.
`Falk, V., Mintz, D., Grunenfelder, J., Fann, J. I., & Burdon,
`T. A. (2001). Influence of three-dimensional vision on sur-
`gical telemanipulator performance. Surgical Endoscopy,
`15(11), 1282–1288.
`Guthart, G. S., & Salisbu

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket