throbber
IN THE UNITED STATES PATENT AND TRADEMARK OFFICE
`
`Shelton, IV
`In re Patent of:
`
`8,479,969
`U.S. Pat. No.:
`July 9, 2013
`Issue Date:
`Appl. Serial No.: 13/369,609
`Filing Date:
`Feb. 9, 2012
`Title:
`DRIVE INTERFACE FOR OPERABLY COUPLING A
`MANIPULATABLE SURGICAL TOOL TO A ROBOT
`
`
`
`Attorney Docket No.: 11030-0049IP9
`
`
`Mail Stop Patent Board
`Patent Trial and Appeal Board
`U.S. Patent and Trademark Office
`P.O. Box 1450
`Alexandria, VA 22313-1450
`
`
`
`
`PETITION FOR INTER PARTES REVIEW OF U.S. PATENT NO. 8,479,969
`PURSUANT TO 35 U.S.C. §§ 311–319, 37 C.F.R. § 42
`
`
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`TABLE OF CONTENTS
`
`I.
`INTRODUCTION ........................................................................................... 3
`II. MANDATORY NOTICES UNDER 37 C.F.R § 42.8 .................................... 7
`A. Real Parties-In-Interest Under 37 C.F.R. § 42.8(b)(1) ............................... 7
`B. Related Matters Under 37 C.F.R. § 42.8(b)(2) ........................................... 7
`C. Lead And Back-Up Counsel Under 37 C.F.R. § 42.8(b)(3) ...................... 7
`D. Service Information .................................................................................... 7
`PAYMENT OF FEES – 37 C.F.R. § 42.103 ................................................... 8
`III.
`IV. REQUIREMENTS FOR IPR UNDER 37 C.F.R. § 42.104 ............................ 8
`A. Grounds for Standing Under 37 C.F.R. § 42.104(a) .................................. 8
`B. Challenge Under 37 C.F.R. § 42.104(b) and Relief Requested ................. 8
`SUMMARY OF THE ’969 PATENT ............................................................. 9
`V.
`VI. SUMMARY OF THE PRIOR ART .............................................................. 11
`A. Prisco ........................................................................................................ 11
`B. Cooper ...................................................................................................... 22
`C. Wallace ..................................................................................................... 24
`D. Tierney ...................................................................................................... 26
`VII. PROSECUTION HISTORY ......................................................................... 27
`VIII. PRIORITY DATE ......................................................................................... 28
`IX. CLAIM CONSTRUCTION UNDER 37 C.F.R. § 42.104(B)(3) .................. 28
`X.
`THERE IS A REASONABLE LIKELIHOOD THAT AT LEAST
`ONE CLAIM OF THE ’969 PATENT IS UNPATENTABLE .................... 29
`A. Ground 1: Claims 23-26 are Anticipated under § 102(e) by Prisco ......... 29
`B. Ground 2: Claims 23-26 Would Have Been Obvious Under § 103
`over Prisco in View of Cooper ................................................................. 74
`C. Ground 3: Claims 23-26 Would Have Been Obvious over Prisco in
`View of Cooper and Tierney .................................................................... 77
`D. Ground 4: Claims 25-26 Would Have Been Obvious Under § 103
`over Prisco in View of Cooper and Wallace, and, If Necessary,
`Tierney ...................................................................................................... 79
`XI. CONCLUSION .............................................................................................. 87
`
`
`
`
`1
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`EXHIBITS
`
`U.S. Pat. No. 8,479,969 to Shelton, IV (“the ’969 Patent”)
`
`Prosecution History of the ’969 Patent (Serial No. 13/369,609)
`
`Declaration of Dr. Bryan Knodel (Prisco as Primary Reference)
`
`Reserved
`
`Reserved
`
`U.S. Patent No. 8,545,515 to Prisco et al. (“Prisco”)
`
`U.S. Patent No. 6,817,974 to Cooper et al. (“Cooper”)
`
`U.S. Patent No. 6,699,235 to Wallace et al. (“Wallace”)
`
`U.S. Patent No. 6,331,181 to Tierney et al. (“Tierney”)
`
`Reserved
`
`Reserved
`
`Reserved
`
`Reserved
`
`U.S. Patent App. No. 2008/0167672 to Giordano et al.
`
`(“Giordano”)
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`IS1001
`
`IS1002
`
`IS1003
`
`IS1004
`
`IS1005
`
`IS1006
`
`IS1007
`
`IS1008
`
`IS1009
`
`IS1010
`
`IS1011
`
`IS1012
`
`IS1013
`
`IS1014
`
`
`
`2
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`
`
`I.
`
`INTRODUCTION
`Intuitive Surgical, Inc. (“Petitioner”) petitions for Inter Partes Review
`
`(“IPR”) of claims 23-26 of U.S. Patent 8,479,969 (“the ’969 Patent”). The ’969
`
`Patent is entitled “Drive Interface for Operably Coupling a Manipulatable Surgical
`
`Tool to a Robot.” Drive interfaces for surgical robots were well-known in the prior
`
`art. In fact, the ’969 Patent incorporates by reference, and largely copies, the prior
`
`art drive interfaces designed by Petitioner and disclosed in its prior art patents:
`
`[T]he tool arrangement described above may be well-suited for use with
`those robotic systems manufactured by Intuitive Surgical, Inc. of
`Sunnyvale, Calif., U.S.A., many of which may be described in detail in
`various patents incorporated herein by reference. The unique and novel
`aspects of various embodiments of the present invention serve to utilize
`the rotary output motions supplied by the robotic system to generate
`specific control motions….
`IS1001 at 31:52-59.1 Not surprisingly, the robotic surgical system described in the
`
`’969 Patent is uncannily similar to the prior art robotic surgical system described in
`
`Petitioner’s patents, including, for example, Petitioner’s prior art “Tierney” patent
`
`(U.S. Patent No. 6,331,181). IS1002 at 280-284; IS1009 (“Tierney”):
`
`
`1 Emphasis added in quotations throughout.
`
`3
`
`

`

`’969 Patent
`
`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`Tierney Prior Art
`
`Robotic Controller
`
`
`
`
`
`
`
`Robotic Manipulator
`
`
`
`4
`
`

`

`’969 Patent
`
`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`Tierney Prior Art
`
`Surgical Tool With Proximal Tool Holder
`
`
`Tool Drive Assembly
`
`
`
`
`
`
`IS1001, FIGs. 23, 25-27; IS1009, FIGs. 3A, 4, 7A, 7J, 8B.
`
`The grandparent application to the ’969 Patent was directed to a hand-held
`
`5
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`surgical instrument. Essentially, the ’969 Patent simply adds a surgical robot and
`
`attempts to patent the obvious combination of a surgical instrument adapted for a
`
`robot. Moreoever, the ’969 Patent does not add just any surgical robot, but
`
`specifically the prior art surgical robot of Petitioner, Intuitive Surgical, Inc.
`
`Not surprisingly, the Examiner found that the original claims of the ’969
`
`Patent read directly on Petitioner’s prior art patents. Specifically, the originally-
`
`filed independent claims were rejected as anticipated and obvious over Tierney.
`
`The applicant then amended the claims, adding details it contended were not
`
`disclosed in Tierney. However, the details supposedly absent from Tierney were
`
`well-known in Petitioner’s other robotic surgical patents, as exemplified by Prisco,
`
`Cooper, and Wallace. IS1003, ¶28. Indeed, the Prisco reference, which serves as
`
`the primary reference for all grounds in this petition, is a prior art patent assigned
`
`to Petitioner that discloses a robotic surgical system of the type described in the
`
`’969 Patent. In addition, the system disclosed in the Prisco prior art patent is
`
`specifically designed to interface with a variety of surgical instruments, including
`
`those of the type described in the ’969 Patent, such as surgical scissors, tissue
`
`graspers, and needle drivers. IS1006, 6:7-37; 8:34-44; 13:38-48; FIGs. 1A-3; 9A-
`
`9E.
`
`Prisco (alone or in combination with other references) renders at least the
`
`challenged claims invalid as anticipated and/or obvious. Petitioner requests IPR of
`
`6
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`the challenged claims on Grounds 1-4 below.
`
`II. MANDATORY NOTICES UNDER 37 C.F.R § 42.8
`A. Real Parties-In-Interest Under 37 C.F.R. § 42.8(b)(1)
`Intuitive Surgical, Inc. is the real party-in-interest. No other party had
`
`access to the Petition, and no other party had any control over, or contributed to
`
`any funding of, the preparation or filing of the present Petition.
`
`B. Related Matters Under 37 C.F.R. § 42.8(b)(2)
`The ’969 Patent is the subject of Civil Action No. 1:17-cv-00871-LPS, filed
`
`on June 30, 2017, in the United States District Court for the District of Delaware.
`
`Concurrently with this petition, Petitioner is filing two more IPR petitions related
`
`to the ’969 Patent directed to different sets of claims, different statutory bases,
`
`and/or different primary references.
`
`C. Lead And Back-Up Counsel Under 37 C.F.R. § 42.8(b)(3)
`Petitioner provides the following designation of counsel.
`
`LEAD COUNSEL
`Steven R. Katz, Reg. No. 43,706
`3200 RBC Plaza, 60 South Sixth Street
`Minneapolis, MN 55402
`Tel: 617-542-5070 / Fax: 877-769-7945
`
`BACK-UP COUNSEL
`John C. Phillips, Reg. No. 35,322
`Tel: 858-678-5070
`Ryan O’Connor, Reg. No. 60,254
`Tel: 858-678-5070
`
`D.
`Service Information
`Please address all correspondence to the address above. Petitioner consents
`
`to electronic service by email at IPR11030-0049IP9@fr.com (referencing No.
`
`7
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`11030-0049IP9 and cc’ing PTABInbound@fr.com, katz@fr.com, phillips@fr.com,
`
`and oconnor@fr.com).
`
`III. PAYMENT OF FEES – 37 C.F.R. § 42.103
`Petitioner authorizes the Office to charge Deposit Account No. 06-1050 for
`
`the petition fee set in 37 C.F.R. § 42.15(a) and for any other required fees.
`
`IV. REQUIREMENTS FOR IPR UNDER 37 C.F.R. § 42.104
`A. Grounds for Standing Under 37 C.F.R. § 42.104(a)
`Petitioner certifies that the ’969 Patent is available for IPR, and Petitioner is
`
`not barred or estopped from requesting IPR.
`
`B. Challenge Under 37 C.F.R. § 42.104(b) and Relief Requested
`Petitioner requests IPR of claims 23-26 of the ’969 Patent on the grounds
`
`listed below. A declaration from Dr. Bryan Knodel (IS1003) is provided in
`
`support.
`
`Grounds
`
`Claims
`
`Ground 1
`Ground 2
`
`23-26
`23-26
`
`Ground 3
`
`23-26
`
`Ground 4
`
`25-26
`
`Basis for Rejections under
`35 U.S.C. § 102 and §103
`Anticipated by Prisco (U.S. 8,545,515)
`Obvious over Prisco in view of Cooper (U.S.
`6,817,974)
`Obvious over Prisco in view of Cooper and
`Tierney (U.S. 6,331,181)
`Obvious over Prisco in view of Cooper and
`Wallace (U.S. 6,699,235), and, if necessary,
`Tierney
`
`As explained in greater detail in Section VIII, infra, the ’969 Patent is
`
`8
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`entitled to a priority date no earlier than May 27, 2011 (filing date of the parent of
`
`the ’969 Patent). The application that resulted in the Prisco patent was filed Nov.
`
`13, 2009, and claims priority to a provisional application filed Sep. 23, 2009.
`
`Prisco therefore qualifies as prior art under at least pre-AIA 102(e). Prisco was not
`
`cited during prosecution. See IS1002. Wallace issued March 2, 2004, and
`
`therefore qualifies as prior art under at least pre-AIA 102(b). Cooper issued on
`
`Nov. 16, 2004, and therefore qualifies as prior art under at least pre-AIA 102(b).
`
`Cooper was made of record during prosecution as part of an 82-page IDS filed
`
`after a notice of allowability that listed over 2,000 references, but Cooper was not
`
`substantively addressed or cited in any office action during prosecution. IS1002,
`
`357-438; 280-285. The specific combinations applied in this petition were not
`
`considered by the Examiner.
`
`V.
`
`SUMMARY OF THE ’969 PATENT
`The ’969 Patent discloses subject matter related to hand-held surgical
`
`instruments and to robotic surgical instruments, but the claims all relate to the
`
`robotic embodiments. As the title states, the ’969 Patent is directed to a “drive
`
`interface for operably coupling a manipulatable surgical tool to a robot.” IS1001,
`
`Title. The disclosed robotic surgical system includes the typical and expected
`
`components, for example, a “master controller and robotic arm cart” and a “tool
`
`drive assembly” that control the surgical tools. IS1001, 23:50-62; 24:62-25:29;
`
`9
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`FIGs. 26-27. As explained above, the robotic surgical system disclosed in the ’969
`
`Patent was copied from Petitioner’s prior art. Various embodiments of the claimed
`
`features are found in the prior art, for example, (1) a tool mounting portion; (2) an
`
`end effector (such as a surgical stapler); (3) a shaft assembly for coupling the end
`
`effector to the tool mounting portion; (4) an articulation joint; and/or (5) a “tube
`
`gear segment 5114” on the shaft of the instrument, which is used to rotate the shaft
`
`and end effector relative to the tool mounting portion:
`
`
`
`10
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`End Effector
`
`IS1001, FIGs. 26, 102; 25:1-26:56; 27:19-47; 30:26-64; 65:32-64; 82:42-83:23.
`
`
`
`
`
`None of these features were novel as of the filing of the ’969 Patent or its
`
`parent application (filed on May 11, 2011) to which the ’969 Patent claims priority.
`
`In fact, these features were found in petitioner’s own prior art patents as described
`
`below.
`
`VI. SUMMARY OF THE PRIOR ART
`A.
`Prisco
`Prisco describes various embodiments of a surgical tool for use with a
`
`robotic surgical system. IS1006 at 6:7-22; 8:4-44. “FIG. 1A is a front elevation
`
`view of the patient side cart component 100 of [Petitioner’s prior art] da Vinci®
`
`11
`
`

`

`Surgical System.” IS1006, 6:14-15.
`
`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`Instrument
`Arm -
`
`Surgical
`Instrument
`
`IS1006, FIG. 1A.
`
`Patient Side Cart
`
`
`
`“FIG. 1A further shows interchangeable surgical instruments 110a, 110b, 110c
`
`mounted on the instrument arms 106a, 106b, 106c . . . .” IS1006, 6:26-28. “FIG.
`
`1B is a front elevation view of a surgeon’s console 120 component of [Petitioner’s
`
`prior art] da Vinci ® Surgical System. The surgeon’s console is equipped with left
`
`and right multiple DOF [(degree of freedom)] master tool manipulators (MTMs)
`
`122a, 122b . . . that are used to control the surgical tools[.]” IS1006, 6:38-42.
`
`- Surgeon’s Console
`
`12
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`IS1006, FIG. 1B.
`
`
`
`As shown below in FIG. 2A, each “[instrument] arm [106a, 106b, 106c] is
`
`divided into two portions. The first portion is the ‘set-up’ portion 202, in which
`
`unpowered joints couple the links. The second portion is powered, robotic
`
`manipulator portion 204 (patient side manipulator; ‘PSM’) that supports and
`
`moves the surgical instrument.” IS1006, 8:2-6, Fig. 2A.
`
`13
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`- Set-Up Portion
`
` - “Patient Side
` Manipulator” (PSM)
`
`“FIG. 2B [(below)] is a perspective view of a manipulator [PSM (patient side
`
`manipulator) 204] with an instrument [110] mounted.” IS1006, 3:4-5, FIG. 2B.
`
`
`
`14
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`Mounting
`Carriage
`
`Actuators
`
`Surgical
`Instrument
`
`- PSM
`
`Force Transmission Assembly
`
`Shaft
`- Wrist
`
`- End Effector
`
`
`
`PSM (patient side manipulator) 204 transmits rotary motion to the removable
`
`surgical instrument via “force transmission disks.” A driving set of force
`
`transmission disks is mounted on “mounting carriage 212” of PSM 204 and a
`
`driven set of force transmission disks is mounted on the “force transmission
`
`assembly 230” of the surgical instrument (which is the “tool mounting portion” of
`
`Prisco). Prisco explains:
`
`Matching force transmission disks in mounting carriage
`212 and force transmission assembly 230 couple actuation
`forces from actuators 232 in PSM 204 to move various
`
`15
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`parts of instrument 110 in order to position, orient, and
`operate instrument end effector 234. Such actuation forces
`may typically roll instrument shaft 218 (thus providing
`another DOF [degree of freedom] through the remote
`center), operate a wrist 236 that provides yaw and pitch
`DOFs, and operate a movable piece or grasping jaws of
`various end effectors. . . .
`
`IS1006, 8:34-44.
`
`Figure 4A of Prisco provides another view of the mounted surgical
`
`instrument, this time with a flexible shaft for use in a curved cannula. In this
`
`figure, the interface disks 414a “couple actuation forces from servo actuators in
`
`PSM 204a to move instrument 402a components” (e.g., to open and close end
`
`effector 408a):
`
`
`
`16
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`- Interface Discs
`
`- Force Transmission Mechanism
`- Mounting Carriage
`
`PSM -
`
`Instrument -
`
`- Shaft
`
`- Cannula
`
`- End Effector
`
`
`
`IS1006, 10:31-41, FIG. 4A. Prisco explains:
`
`FIG. 4A is a schematic view of a portion of a patient side
`robotic manipulator [(PSM)] that supports and moves a
`combination of a curved cannula [416a] and a passively
`flexible surgical instrument [402a]. As depicted in FIG.
`
`17
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`4A, a telerobotically operated surgical instrument 402a
`includes a force transmission mechanism 404a, a passively
`flexible shaft 406a, and an end effector 408a. Instrument
`402a is mounted on [a mounting carriage] 212a of a PSM
`204a (previously described components are schematically
`depicted for clarity). Interface discs 414a couple actuation
`forces from servo actuators in PSM 204a to move
`instrument 402a components.
`
`IS1006, 10:31-41, FIG. 4A.
`
`Figures 5 and 6 of Prisco illustrate a surgical instrument with a flexible shaft
`
`in both curved and straight positions, respectively. “FIG. 5 is a diagrammatic view
`
`of an illustrative flexible instrument 500 used with a curved cannula. Instrument
`
`500 includes a proximal end force transmission mechanism 502, a distal end
`
`surgical end effector 504, and a shaft 506 that couples force transmission
`
`mechanism 502 and end effector 504.” IS1006, 12:8-13, FIG. 5.
`
`Shaft -
`
`Force -
`Transmission
`Mechanism
`
`
`“FIG. 6B is a diagrammatic view that illustrates aspects of a push/pull
`
`End -
`Effector
`
`18
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`[surgical] instrument design. As shown in FIG. 6B, a force transmission
`
`mechanism 620 is coupled to a grip type end effector 622 by a flexible shaft body
`
`624.” IS1006, 14:29-32, FIG. 6B.
`
`- Flexible Shaft Body
`Instrument -
`Force Transmission Assembly
`
`- End Effector
`
`
`
`Figure 9E provides a close up view of one of Prisco’s end effectors. In
`
`particular, “FIG. 9E is a diagrammatic view that illustrates a push/pull type end
`
`effector that may be at the distal end of the flexible shaft instruments (an
`
`illustrative clip applier end effector is shown).” IS1006, 19:42-45, FIG. 9E.
`
`19
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`
`It is typically necessary to rotate the shaft of a surgical instrument to
`
`reposition the end effector. Like the ’969 Patent, Prisco uses a tube gear to rotate
`
`the shaft. Prisco specifically discloses that the shaft is rotated by a “helical drive
`
`gear” that drives a tube gear called the “shaft roll gear,” as shown in FIG. 7C:
`
`20
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`
`
`IS1006, FIG. 7C; 15:36-16:7.
`
`
`
`Prisco also discloses articulation of the end effector via a “wrist” of the
`
`surgical instrument. For the wrist mechanism, Prisco incorporates Cooper, which
`
`is another surgical robotic reference: “A wrist to provide one or more end effector
`
`DOF’s [Degrees of Freedom] (e.g., pitch, yaw; see e.g., U.S. Pat. No. 6,817,974
`
`(filed Jun. 28, 2002) (disclosing ‘Surgical Tool Having Positively Positionable
`
`Tendon-Actuated Multi-Disk Wrist Joint’), which is incorporated herein by
`
`reference) is optional and is not shown.” IS1006, 10:43-48. This statement
`
`incorporates at least Cooper’s wrist mechanism as if it were set out expressly rather
`
`than through incorporation. IS1003, ¶42.
`
`Finally, Prisco discloses gear-driven actuation of the end effector. For
`
`21
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`example, Prisco discloses a “rack gear 784” that operates the grasping jaws of
`
`various end effectors. Pinion drive gears 782 cause rack gear 784 to move back
`
`and forth along the shaft’s longitudinal axis, thus pushing and pulling a drive rod
`
`that runs through the shaft and is coupled to the end effector jaws. IS1006, 16:13-
`
`23; FIG. 7D.
`
`Rack Gear
`
`Pinion Drive
`Gears
`
`
`
`B. Cooper
`As noted above, Prisco incorporates by reference at least the wrist
`
`mechanism disclosed in Cooper. Much like Prisco, Cooper discloses “a surgical
`
`instrument 400 having an elongate shaft” configured to releasably couple to “a
`
`robotic arm or system.” IS1007, 17:26-50. As shown in FIG. 36, the instrument
`
`400 includes an end effector 406 and a base 410, which couples the instrument to a
`
`robotic system. A shaft connects the end effector to the base and the shaft rotates
`
`in either direction to rotate the end effector, “as indicated by arrows H.”
`
`22
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`
`
`IS1007, FIG. 36; 17:35-50; see also FIG. 64; 24:1-23.
`
`Furthermore, Cooper’s surgical instrument includes an articulation system
`
`comprising a proximal disk 412 and a distal disk 416 coupled via grip support 420
`
`to an end effector 406. IS1003, ¶45. As shown in FIG 37, there are several pivot
`
`points between the proximal disk 412 and the distal disk 416 that supports end
`
`effector 406. Together these form an articulation joint (a wrist):
`
`23
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`Pivot Point
`
`Pivot Point
`
`Pivot Point
`
`Pivot Point
`
`
`
`IS1007, 5:54-6:3; 13:22-49, 17:25-64; FIGs. 14-21, 36-39, 51-56; see also 14:14-
`
`59 (“hinge mechanisms disposed on opposite sides of the disks guide the disks in
`
`pitch and yaw rotations to produce, for instance, the 90° pitch of the wrist 140”),
`
`21:49-22:41, FIGS. 17-21 (disclosing another embodiment having proximal and
`
`distal disks).
`
`C. Wallace
`Wallace discloses “a robotic surgical tool for use in a robotic surgical
`
`system” that, much like the surgical instrument of Prisco, includes a shaft for
`
`supporting a surgical end effector, the shaft supported by a “tool base 62 [that]
`
`24
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`includes an interface 64 which mechanically and electrically couples the tool 50 to
`
`a manipulator on the robotic arm cart” as shown in FIGs. 1 and 2A:
`
`End
`Effector
`
`Wrist
`Joint
`
`
`
`
`
`IS1008, FIGs. 1, 2A; Abstract; 7:33-56; IS1003, ¶46. Wallace further discloses an
`
`articulation assembly comprising a “wrist joint or mechanism” that is actuated by
`
`“a plurality of rods” that are driven by a gear assembly including “sector gears
`
`312” and “gears 400” to advance and retract rods to cause articulation of the
`
`articulation wrist joint. IS1008, 7:57-65; 13:6-14:15, FIG. 30; IS1003, ¶47.
`
`
`
`25
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`Wrist Articulation
`Gear Assembly
`
`Articulation Rods
`
`
`
`D. Tierney
`Prisco broadly and unequivocally incorporates Tierney by reference.
`
`IS1006, 15:17-20, IS1009. This incorporates all of Tierney into Prisco as if it were
`
`set out expressly rather than through incorporation. Harari v. Lee, 656 F.3d 1331,
`
`1335 (Fed. Cir. 2011) (holding that “the broad and unequivocal language” stating
`
`that “[t]he disclosures of the two applications are hereby incorporate[d] by
`
`reference” incorporated the entire disclosures of the two applications);2 Advanced
`
`
`2 See also Biscotti Inc. v. Microsoft Corp., No. 2:13-CV-01015-JRG-RSP, 2017
`
`U.S. Dist. LEXIS 144164, at *12 (E.D. Tex. May 11, 2017) (confirming that
`
`Harari, which addressed incorporation by reference in the context of written
`
`description, also applies to anticipation because “[t]he incorporation by reference
`
`doctrine . . . does not vary across different applications of the doctrine.”).
`
`26
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`Display Sys., Inc. v. Kent State Univ., 212 F.3d 1272, 1282 (Fed. Cir. 2000)
`
`(“Material not explicitly contained in [a] single, prior art document may still be
`
`considered for purposes of anticipation if that material is incorporated by reference
`
`into the document.”); IS1003, ¶48. As noted above, the robotic surgical system
`
`described in the ’969 Patent appears to be the prior art robotic surgical system
`
`described in Petitioner’s Tierney patent. See Section I.
`
`VII. PROSECUTION HISTORY
`During prosecution, the USPTO issued a single office action rejecting the
`
`broad independent claims, but indicating that two independent picture claims and a
`
`variety of dependent claims contained allowable subject matter. IS1002, 280-284.
`
`The broad claims were rejected over Petitioner’s Tierney patent. IS1002, 280-284.
`
`The applicant subsequently amended the independent claims to include subject
`
`matter deemed allowable and added new dependent claims containing the
`
`allowable subject matter of original dependent claim 7 (issued claim 19), claim 9
`
`(issued claim 21), claim 12 (issued claim 23), and claim 14 (issued claim 24).
`
`IS1002, 311; 304-310. The examiner then issued a notice of allowance. Rather
`
`than allowing the patent to issue, the applicant filed an RCE and submitted an IDS
`
`listing over 2,000 references. IS1002 at 328-333; 357-483. A notice of allowance
`
`promptly followed, and the ’969 Patent issued on July 9, 2013. IS1002, 547-552;
`
`IS1001, Face.
`
`27
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`
`VIII. PRIORITY DATE
`The ’969 Patent issued from U.S. Application No. 13/369,609, which is a
`
`continuation of U.S. Application No. 13/118,259, filed on May 27, 2011, which is
`
`a continuation-in-part of U.S. Application No. 11/651,807, which issued as U.S.
`
`Patent No. 8,459,520. The grandparent ’807 application (IS1014) does not provide
`
`support for any of the challenged claims and therefore the ’969 Patent is entitled
`
`only to the priority date of the ’259 application: May 27, 2011. For example, each
`
`of the challenged independent claims (23-24) recites “[a] surgical tool for use with
`
`a robotic system that has a tool drive assembly that is operatively coupled to a
`
`control unit of the robotic system.” Each of the challenged independent claims
`
`further recites that the surgical tool has a tool mounting portion “being configured
`
`to operably interface with the tool drive assembly” that has at least one “rotatable
`
`body portion.” IS1001, claims 23-24. The grandparent ’807 application provides
`
`no support for these recitations. IS1003, ¶¶29-31. Rather, the ’807 application is
`
`directed toward handheld “endoscopic surgical instrument[s]” with only a passing
`
`reference to “robotic-assisted surgery.” IS1014, ¶¶15, 89, FIGs. 1-2.
`
`IX. CLAIM CONSTRUCTION UNDER 37 C.F.R. § 42.104(B)(3)
`For the purposes of IPR only, Petitioner submits that the terms of the ’969
`
`Patent are to be given their broadest reasonable interpretation as understood by a
`
`28
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`POSITA in view of the specification (“BRI”).3 37 CFR § 42.100(b).
`
`X. THERE IS A REASONABLE LIKELIHOOD THAT AT LEAST ONE
`CLAIM OF THE ’969 PATENT IS UNPATENTABLE
`As detailed below, claims 23-26 of the ’969 Patent are anticipated by Prisco
`
`or obvious over Prisco in view of Cooper, Tierney, and/or Wallace.
`
`A. Ground 1: Claims 23-26 are Anticipated under § 102(e) by Prisco
`[23.P] A surgical tool for use with a robotic system that has a tool drive
`assembly that is operatively coupled to a control unit of the robotic system
`that is operable by inputs from an operator and is configured to provide at
`least one rotary output motion to at least one rotatable body portion
`supported on the tool drive assembly, said surgical tool comprising:
`If the preamble of claim [23] is deemed to be limiting, then Prisco discloses
`
`it. IS1003, ¶¶52-59. In addition, Prisco incorporates Tierney, which the PTO has
`
`already found discloses the preamble of claim 23. IS1006, 15:16-20; IS1002, 282-
`
`83 (rejecting original claim 1 as anticipated by Tierney).
`
`
`3 Petitioner acknowledges that the Office has proposed to change from the BRI
`
`standard to the standard applied in District Courts. See 83 Fed. Reg. 21221 (pro-
`
`posed May 9, 2018). Petitioner submits that the prior art discussed herein invali-
`
`dates the challenged claims under either standard. If the Office changes the rule
`
`after the filing of the Petition and applies the new standard to this proceeding, then
`
`due process requires the Office afford Petitioner an opportunity to provide addi-
`
`tional argument and evidence on that issue.
`
`29
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`“A surgical tool for use with a robotic system”
`
`Prisco discloses a surgical tool (e.g., surgical instruments 402a, 500) for use
`
`with a robotic system (Petitioner’s prior art da Vinci® Surgical System). IS1003,
`
`¶¶52- 53; IS1006, 6:14-15, 6:26-28, 6:38-42, 10:31-41, FIGs., 1A, 1C, 4A, 5. For
`
`example, FIG. 4A depicts surgical instrument 402a mounted to the PSM (patient
`
`side manipulator) 204a portion of an instrument arm via carriage 212a:
`
`- Interface Discs
`
`- Force Transmission Mechanism
`- Mounting Carriage
`
`PSM -
`
`Instrument -
`
`FIG. 4A
`
`- End Effector
`
`
`
`30
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`FIGs. 1A and 1B depict the robotic system on which the instrument is
`
`mounted. FIG. 1A shows the patient side cart 100 to which the surgical
`
`instruments are mounted, and FIG. 1B shows the surgeon’s console 120 used to
`
`control movement of the surgical instruments:
`
`
`
`IS1006, 6:14-15 (“FIG. 1A is a front elevation view of the patient side cart
`
`component 100 of the da Vinci ® Surgical System.”), 6:38-39 (“FIG.1B is a front
`
`elevation view of a surgeon’s console 120 component of the da Vinci ® Surgical
`
`System.”), Fig. 1A, 1B.
`
`31
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`The robotic system “has a tool drive assembly”
`
`
`
`Prisco’s robotic system contains a tool drive assembly (e.g., the combination
`
`of “actuators 232” and “mounting carriage 212”). “An illustrative surgical
`
`instrument 110 is shown mounted at an instrument mounting carriage 212” in FIG.
`
`2B of Prisco. IS1006, 8:15-17, FIG. 2B; IS1003, ¶¶54-55.
`
`Mounting
`Carriage
`
`Actuators
`
`- PSM
`
`Surgical
`Instrument
`
`Prisco explains that “actuators 232 in PSM [(patient side manipulator)] 204 . . .
`
`move various parts of instrument 110 in order to position, orient, and operate
`
`instrument end effector 234.” IS1006, 8:34-38; see also 10:37-41 (“Instrument
`
`
`
`32
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`402a is mounted on [a mounting carriage] 212a of a PSM 204a . . . . Interface
`
`discs 414a couple actuation forces from servo actuators in PSM 204a to move
`
`instrument 402a components.”).
`
`
`
`In addition, Prisco’s incorporation of Tierney discloses the same tool drive
`
`assembly as the ’969 Patent:
`
`’969 Patent
`
`Tierney
`
`Rotatable Bodies 1250
`Adapter
` |
`
` |
`Adapter
`
`Rotatable
`- Bodies
`
`Tool Drive -
`Assembly
` |
`Tool Holder
`
`Tool Holder -
`
`Drive
`Elements -
`
`Drive
`Elements
`1271
`
`
`
`
`Compare IS1001, FIG. 27 with IS1009, FIG. 7A, 7J, 7F; IS1003, ¶50; see also
`
`IS1009, 11:33-35, 4:33-35, 7:65-8:7, 10:12-15, 11:3-6, FIGs. 3A, 8A, 8B, 9FIGs.
`
`7F-7M.
`
`33
`
`

`

`IPR of U.S. Pat. No.: 8,479,969
`Attorney Docket No. 11030-0049IP9
`The tool drive assembly “is operably coupled to a control unit”
`
`Prisco’s tool drive assembly (the combination of “actuators 232” and
`
`“mounting carriage 212”) is operably coupled to a control unit (“surgeon’s console
`
`120” alone or in combination with “patient side cart 100”). IS1003, ¶56. “The
`
`surgeon’s console [120] is equipped with . . . master tool manipulators (MTMs)
`
`122a, 122b . . . that are used to control the surgical too

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket