throbber
Exhibit 2001
`Exhibit 2001
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 1 of 17
`
`UNITED STATES DISTRICT COURT
`DISTRICT OF MINNESOTA
`
`Biomedical Device Consultants &
`Laboratories of Colorado, LLC,
`
`Plaintiff,
`
`v.
`
`TA Instruments - Waters, LLC,
`
`Defendants.
`
`Civil File No. 0:17-cv-03403
`
`DECLARATION OF MICHAEL
`GIRARD IN SUPPORT OF MOTION
`FOR PRELIMINARY INJUNCTION
`
`I, Michael J. Girard, hereby declare and state as follows:
`
`1.
`
`I have been retained by Plaintiff Biomedical Device Consultants &
`
`Laboratories of Colorado, LLC ("BDC") to offer technical analysis and opinions
`
`regarding various issues relevant to this action, including infringement and validity of the
`
`Patents-in-Suit, U.S. Patent Nos. 8,584,538 ("the '538 Patent"), 8,627,708 ("the '708
`
`Patent"), 9,186,224 ("the '224 Patent"), 9,237,935 ("the '935 Patent") (collectively, the
`
`"Patents-in-Suit"). I have personal knowledge of the facts herein, and if called as a
`
`witness, I could and would testify competently thereto.
`
`2. My education includes a Bachelor of Science in Civil (Structural)
`
`Engineering from the University of Illinois, and a Master of Business Administration
`
`from the University of St. Thomas.
`
`3.
`
`I am currently the President ofmy own consulting firm, Girard Technical
`
`Services, Inc. My firm provides research and development, engineering, and technical
`
`management consulting services in the medical device industry.
`
`4816-6498-1075\5
`
`Water Techs. Corp. v. Biomedical Device Consultants & Labs
`IPR2018-00498
`Ex. 2001
`
`Page 1 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 2 of 17
`
`4.
`
`I have thirty-seven years of experience in engineering, twenty-seven of
`
`which are in the medical device industry. I am a named inventor on 38 issued U.S.
`
`Patents with additional applications pending.
`
`5. My experience includes work with many cardiovascular devices and
`
`specifically includes 19 years of heart valve experience. I've worked for several medical
`
`device companies in the role of development and testing of both surgical and
`
`transcatheter delivered prosthetic heart valves. Testing of heart valves usually includes
`
`durability testing at accelerated rates with equipment like the systems produced by BDC
`
`and TA Instruments. Therefore, I am very familiar with durability test equipment and the
`
`requirements of such testing. My curriculum vitae is attached hereto as Exhibit A.
`
`A. Technology Overview
`
`6.
`
`This case concerns equipment used for durability or high cycle fatigue
`
`testing of heart valves. Before any medical device, such as a heart valve, is marketed it
`
`must meet certain regulatory standards. International bodies, such as the International
`
`Organization for Standardization ("ISO"), set certain standards, such as those for testing
`
`the durability of medical devices, including heart valves. The specific standards for heart
`
`valves are defined in ISO 5840.
`
`7.
`
`Prosthetic heart valves must be tested to ensure that they will function for
`
`the anticipated life of the patient by opening and closing the valve leaflets under flows
`
`and pressures that are present within the human vascular system. The normal human
`
`heart beats about 40 million times each year. The test requirements for evaluating
`
`prosthetic heart valves according to the standards require that the valves be able to
`
`2
`
`Page 2 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 3 of 17
`
`survive and function for hundreds of millions of cycles ( e.g., at least 5 years or 200
`
`million cycles). The valves must also be able operate over a specific range of opening
`
`and closing pressures that simulate physiological conditions, and therefore testing
`
`standards require that a specific pressure differential be generated across the valve when
`
`closed.
`
`8.
`
`Testing systems use a test fluid to mimic blood and pressurize the fluid to
`
`mimic the blood pressure in the human body. Testing requires fluid flow through the
`
`valve and creating a pressure differential across the test valve when closed at a certain
`
`minimum pressure for a certain length of the cycle. A "drive mechanism" such as a
`
`pump drives the test fluid into the test chamber in order to create the fluid flow and
`
`desired pressure conditions. In order to complete hundreds of millions of cycles in a
`
`commercially viable timeframe, durability testing is done on an "accelerated" basis. In
`
`other words, the speed of the cycles is faster than a normal human heartbeat ( a normal
`
`beat rate is 70 beats per minute - bpm). Using current technology at accelerated cycling
`
`of 800 bpm, testing takes approximately six months to simulate 200 million cycles.
`
`B. Problems in Prior Art Technology
`
`9.
`
`The Patents-in-Suit identify several problems with the prior art.
`
`10.
`
`Driving mechanisms in the prior art had limited control over closing rates
`
`and would often produce "pressure spikes" when the systems maintain pressure above the
`
`testing threshold for the amount of time required by testing standards. These pressure
`
`spikes are undesirable because they wear out valves during testing faster than they would
`
`be worn out in the human body, causing false test failures.
`
`3
`
`Page 3 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 4 of 17
`
`11.
`
`According to the Patents-in-Suit, prior art valve testing devices also
`
`experienced operational problem. For example, one prior art device used a flexible
`
`metallic bellows to pressurize. '538 Patent col.111.34-36. However, a higher load is
`
`required to drive the metallic bellows and can thus, impact the reliability of the test
`
`system and increase maintenance requirements which increases the already lengthy
`
`durability testing process.
`
`C. Overview of the Patents-in-Suit
`
`12.
`
`The Patents-in-Suit propose to solve these problems in the prior art.
`
`13.
`
`For example, the '935 Patent covers a device for accelerated testing of
`
`valved prosthetics with several components including a test chamber and an "excess
`
`volume area" that is connected to a return chamber. The excess volume area and return
`
`chamber within the test chamber are shown below:
`
`4
`
`Page 4 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 5 of 17
`
`Fluid return chamber
`
`~ - - - _ ;Holder for test sample of
`prosthetic valve
`
`126
`
`120
`
`111
`
`Fluid
`return
`conduit
`
`Pressure source
`
`14.
`
`Another patent, the '224 Patent, covers a method for operating a test system
`
`with an excess volume area. The '224 Patent describes a method that includes storing a
`
`volume of the test fluid (which approximates blood) in an "excess volume area" when the
`
`system is in a "drive" stroke of the system that opens the prosthetic valve and releases the
`
`fluid from the area during the "return" stroke of the system.
`
`15.
`
`The excess volume area improves the testing environment by minimizing
`
`unnatural and undesirable pressure spikes and provides advantages of speed and
`
`longevity in the drive system. When the system is driving test fluid through the
`
`5
`
`Page 5 of 24
`
`

`

`- ·-.~.· .. ,
`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 6 of 17
`
`prosthetic valve, the excess volume area is able to store the excess volume downstream of
`
`the prosthetic test valve. Tbis can alleviate some of the system pressure during the drive
`
`stroke. The excess volume area also provides compliance that controls the resistance and
`
`the forward flow pressure gradients across the valve, and minimizes unnatural and
`
`undesirable pressure noise or spikes. During the reverse stroke of the motor when the
`
`valve is closing, the excess volume area and compliance helps to build back pressure on
`
`the valve, return the downstream volume to the pump and minimize unnatural and
`
`undesirable valve closing pressure spikes, i.e., recoil, that can negatively impact the
`
`durability of the prosthetic test valve.
`
`16.
`
`The Patents-in-Suit offer potential advantages for heart valve durability
`
`testing compared to the prior art. The use of the excess volume area and the avoidance of
`
`pressure spikes allows the system to better comply with the rigors of the durability testing
`
`standards without exposing the test valves to undesirable, excess, and clinically irrelevant
`
`pressures.
`
`D. Infringement Analysis of DuraPulse Test System
`
`17.
`
`I understand that the accused device in this case is the DuraPulse Heart
`
`Valve Test Instrument ("DuraPulse"). To understand the components of the DuraPulse
`
`and their operation, I have reviewed the following: the TA Instruments website offering
`
`the DuraPulse for sale, the sales brochure for the DuraPulse attached to the Complaint in
`
`this case, pictures of the DuraPulse included in BDC's motion for preliminary injunction,
`
`a video of the operation of the DuraPulse available at
`
`https://www.youtube.com/watch?v=KgmpQCRrYpQ, and U.S. Patent No. 9,662,210
`
`6
`
`Page 6 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 7 of 17
`
`("the '210 Patent"), attached hereto as Exhibit B. The '210 Patent belongs to Defendant
`
`TA Instruments-Waters, LLC ("TA Instruments"). The '210 Patent is entitled "System
`
`for Testing Valves" and, based on the diagrams and description, appears to describe the
`
`DuraPulse. In addition, I understand that BDC's CEO Dr. Craig Weinberg has observed
`
`the DuraPulse and the '210 Patent describes that product.
`
`18.
`
`I understand that the patent infringement analysis involves two steps: ( 1)
`
`claim construction, and (2) comparison of the accused product to the construed claims. I
`
`understand that direct infringement of an apparatus claim requires that each and every
`
`limitation set forth in a claim appear in the accused products. I further understand that
`
`direct infringement of a method claim requires all steps of the claimed method to be
`
`performed by or attributable to a single entity.
`
`i.
`
`Claim Construction
`
`19.
`
`I understand that claim terms are to be given the ordinary and customary
`
`meaning of the term as evaluated from the perspective of one of ordinary skill in the art at
`
`the time of the invention. I further understand that the claim term is to be read in context
`
`of the claim itself and in the context of the entire patent, including the specification, but
`
`that limitations from the specification should not be read into the claims.
`
`20.
`
`Based on context, including both the claims and the specification, the
`
`person of ordinary skill would understand that the term "compliance chamber" to mean
`
`"a cavity or volume that functions to absorb some of the pressure in the system." This is
`
`explained directly in the specification of the Patents-in-Suit. See, e.g., '538 Patent col.8
`
`11.59-62. The specification explains that the chamber or chambers absorb some of the
`
`7
`
`Page 7 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 8 of 17
`
`pressure placed upon the fluid in the t~st chamber and can also impact the recoil.
`
`Likewise, the compliance chamber minimizes the effect of rapidly changing pressure
`
`gradients associated with accelerated testing. Finally, the compliance chamber is "a
`
`cavity or volume" because the specification notes that it may be air or another gas and
`
`may directly contact the fluid or may be separated from the fluid by a membrane.
`
`ii.
`
`Comparison of Claim Elements to DuraPulse
`
`21.
`
`It is my opinion that the DuraPulse infringes at least claims 1 and 9 of the
`
`'935 Patent. In addition, it is my opinion that use of the DuraPulse infringes at least
`
`claims 1 and 6 of the '224 Patent.
`
`22.
`
`The DuraPulse includes all the elements of the Patents-in-Suit. For
`
`example, an image of the DuraPulse is shown below, as noted it has a test chamber with
`
`an excess volume area:
`
`Test chamber
`
`23.
`
`The DuraPulse also has a drive motor that operates the system at an
`
`8
`
`Page 8 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 9 of 17
`
`accelerated rate:
`
`4--------- [A] Pressurizable test
`chamber
`
`1f::"'"------[B] Drive motor that
`1s a pressure source
`and operates at 900-
`1800 beats per minute
`
`24.
`
`As explained in the following claim charts, it is my opinion that due to
`
`these features, and the other features discussed in the charts, that the DuraPulse infringes
`
`claims of the Patents-in-Suit.
`
`25.
`
`The '935 Patent:
`
`Claim Limitation
`
`Claim
`Construction
`
`Presence of Limitation in DuraPulse System
`
`1. A device for
`accelerated cyclic
`testing of a
`valved prosthetic
`device
`..
`compnsmg
`
`The DuraPulse product is advertised by TA
`Instruments as a "Heart Valve Test Instrument."
`htt,g://www.tainstruments.com/heart-valve-
`durabiliiy-test-instrument/. A prosthetic heart
`valve is a valved prosthetic device.
`
`9
`
`Page 9 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 10 of 17
`
`a pressure source
`configured to
`drive a test
`system fluid
`cyclically within
`the device above
`a normal
`physiological
`rate, at an
`accelerated
`pulsed rate of
`greater than 200
`beats per minute
`within the device;
`and
`
`The DuraPulse product is also touted by TA
`Instruments as a device "for accelerated heart
`valve durability testing."
`http://www.tainstruments.com/heart-valve(cid:173)
`durability-test-instrument/ ( description tab,
`emphasis added). TA Instruments notes that it
`can be run at over 30Hz, which is well above the
`normal human heart rate. Further, the TA
`Instruments website states that the DuraPulse
`"provides testing to the ISO 5840 standard for
`heart valve durability assessment." This
`standard requires cyclic testing.
`
`Therefore, the DuraPulse is a system for
`"accelerated cyclic testing of a valved prosthetic
`device."
`
`The DuraPulse device is shown above. It
`includes a pressure source ( contained in the
`base/cylinders in the above image). The device
`operates at a frequency of 15-30+ Hz. See
`http://www.tainstruments.com/wp(cid:173)
`content/uploads/TA Heart Valve Tester.pdf at
`3. Hertz (Hz) is a unit of measurement for
`cycles per second. Therefore, mathematically,
`15-30 Hz frequency corresponds to a rate of 900-
`1800 beats per minute for the test system fluid
`within the device.
`
`10
`
`Page 10 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 11 of 17
`
`a pressurizable
`test chamber for
`containing the
`test system fluid
`and further
`compnsmg
`
`a fluid
`distribution
`chamber
`positioned on a
`first side of the
`valved prosthetic
`device and in
`fluid
`communication
`with the pressure
`source;
`
`The above image shows the DuraPulse test
`chamber. The red circle depicts the fluid line,
`showing that the test chamber contains test
`system fluid. In order for the DuraPulse to
`function as a heart valve testing system, the test
`chamber must be ressurizable.
`
`Fluid
`ion
`Distrib
`Chamb r
`
`Test sample hol
`prosthetic heart
`
`The above picture of the DuraPulse device
`shows a fluid distribution chamber positioned on
`one side of the prosthetic heart valve being
`tested. The clear cylinder below the fluid
`distribution chamber rovides fluid
`
`11
`
`Page 11 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 12 of 17
`
`communication between the fluid distribution
`
`a fluid return
`chamber
`positioned on a
`second side of the
`valved prosthetic
`device;
`
`etum
`hambe
`
`a fluid return
`conduit both
`structurally and
`fluidly
`connecting the
`fluid distribution
`chamber to the
`fluid return
`chamber; and
`
`On the other side of the prosthetic heart valve
`(the side opposite the fluid distribution
`chamber), is a fluid return chamber, shown with
`the red arrow.
`
`Small channels, one of which is marked with a
`red arrow, surround the test sample holder. The
`'210 Patent refers to these as "return flow
`orifices." '210 Patent col.911.49-53 & fig.6b.
`The video of the operation of the DuraPulse
`confirms that these channels are in the system
`and allow return flow of test fluid between the
`fluid return chamber and fluid distribution
`chamber.
`
`12
`
`Page 12 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 13 of 17
`
`an excess volume
`area capable of
`operating at the
`accelerated
`pulsed rate,
`wherein the
`excess volume
`area is in fluid
`communication
`with the fluid
`return chamber
`providing a
`volume for
`storing a volume
`of a test system
`fluid when the
`test system fluid
`is under
`compression.
`
`These conduits allow test fluid to move from the
`fluid return chamber back to the fluid
`distribution chamber. In other words, the
`conduits "both structurally and fluidly" connect
`the fluid distribution chamber and the fluid
`return chamber.
`
`xcess
`olume
`
`luid
`eturn
`chambe
`
`The excess volume area is indicated with a red
`arrow. As discussed above, the device operates
`at an accelerated rate. Therefore, the excess
`volume area is "capable of operating at the
`accelerated pulsed rate."
`
`Moreover, as shown, test fluid is able to move
`from the fluid return chamber to the excess
`volume area. In other words, the two are "in
`fluid communication." This is confirmed by the
`specification of the '210 Patent, which discusses
`the flow of test fluid into the excess volume area
`from the fluid return chamber. '210 Patent
`col.IO, 1.66-col.l l, 1.22.
`
`When the test system fluid is under compression
`it flows through the valve into the fluid return
`chamber. A portion of this fluid is stored in the
`excess volume area as shown below:
`
`13
`
`Page 13 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 14 of 17
`
`9. The device
`of claim 1,
`wherein the
`excess volume
`area comprises a
`compliance
`chamber defining
`a cavity within
`the fluid return
`chamber.
`
`"Compliance
`chamber" is a
`cavity or
`volume that
`functions to
`absorb some of
`the pressure in
`the system.
`
`Excess
`
`(
`
`As the image shows, the excess volume area is a
`cavity within the fluid return chamber.
`
`In the '210 Patent specification, it notes that this
`airspace is a "compliancy feature." '210 Patent
`col.7, 11.51-52. In other words, when the fluid is
`compressed, the airspace absorbs some of that
`pressure, working as a gas spring. '210 Patent
`col.11, 11.8-22,32-34. Therefore, the excess
`volume area com rises a com liance chamber.
`
`14
`
`Page 14 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 15 of 17
`
`26.
`
`The '224 Patent
`
`Claim Limitation
`
`1. A method for
`operating an
`accelerated cyclic
`test system for
`evaluating a
`valved prosthetic
`device comprising
`
`driving a test
`system fluid
`cyclically above a
`normal
`physiological rate,
`at an accelerated
`pulsed rate of
`greater than 200
`beats per minute
`within the test
`system;
`
`storing a volume
`oftest system
`fluid in an excess
`volume area
`during a system
`driving stroke that
`opens the valved
`prosthetic device;
`and
`
`Claim
`Construction
`
`Presence of Limitation in DuraPulse System
`
`As discussed above, the DuraPulse is a device
`used for accelerated heart valve durability
`testing. Therefore, any method for operating
`the DuraPulse is a method for operating an
`accelerated cyclic test system for evaluating a
`valved prosthetic device.
`
`The DuraPulse system has a linear motor that
`operates at between 15-30 Hz.
`http://www.tainstruments.com/wp-
`content/up 1 oads/ElectroF orce-Cardio-Test(cid:173)
`Instruments. pdf at 13. When in operation the
`linear motor drives the test fluid.
`Mathematically, 15-30 Hz corresponds to a
`cyclical pulse rate of between 900-1800 bpm.
`I understand that the ISO 5850 requirements
`define "normal physiological rate" as between
`30-200 beats per minute. Therefore, the
`DuraPulse device has a test system fluid
`cyclically above ·a normal physiological rate.
`
`,..._ ___ _i.:.Exces
`volu e
`area
`
`Fluid line
`
`15
`
`Page 15 of 24
`
`

`

`. . _,:-.. · .. · ..
`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 16 of 17
`
`releasing the
`stored volume of
`test system fluid
`during a return
`stroke that closes
`the valved
`prosthetic device.
`
`During the driving stroke, the DuraPulse
`system stores working fluid in an excess
`volume area shown in the image at the top of
`the left-hand chamber.
`
`As described in the specification of the '210
`Patent, as a bellows compresses the test fluid
`(i.e., during the drive stroke), fluid flows
`through the valve causing it to open, which
`causes some test fluid to then compress the air
`and thereby move into the previously occupied
`airspace (i.e., the excess volume area). '210
`Patent col.11, 11.12-22.
`
`The fluid
`from the
`excess
`volume
`area drains
`as fluid
`flows
`through
`return
`ports to th
`other side
`ofthe
`closed
`valve.
`
`L-------'"'~
`
`....._=..:L-....:.;..J1;...__;~ -=::....:
`
`During the return stroke, the valved prosthetic
`device closes and the stored volume is
`released as the working fluid flows out of the
`second chamber through a return path in the
`ports surrounding the valve.
`
`As described in the specification of the '210
`Patent, as the bellows is drawn back (i.e.,
`during a "return stroke"), fluid is released
`from the excess volume area causing an
`increase in the volume of the airspace. '210
`Patent col.11, 11.23-31. This is shown in the
`
`16
`
`Page 16 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41 Filed 11/22/17 Page 17 of 17
`
`video of the operation of the DuraPulse as
`well.
`
`6. The method
`of claim 1, further
`compnsmg
`compressmg a
`· volume of a
`compressible gas
`with the volume
`of test system
`fluid to provide a
`spring force
`counter to and in
`response to a
`pressure on the
`test system fluid
`when the volume
`of test system
`fluid is stored in
`the excess volume
`area.
`
`Air
`
`Fluid
`
`When the return fluid chamber fills with test
`fluid, it compresses the air in the excess
`volume area providing for storage of
`additional fluid volume, while also providing a
`spring force that is counter to and in response
`to the pressure on the test fluid. Indeed, the
`specification of the '210 Patent specifically ·
`notes that the airspace "functions as a gas
`spring." '210 Patent col.11, 1.34.
`
`I declare under the penalty of perjury that the foregoing is true and correct.
`
`Executed on
`
`i ( /z.:z.. , 201 7 in
`
`J- ',)..Io LA- I<&°$ J ,nµ'
`
`17
`
`Page 17 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41-1 Filed 11/22/17 Page 1 of 7
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`EXHIBIT A
`
`Page 18 of 24
`
`

`

`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`6318 White Owl Drive
`Lino Lakes, MN 55014
`(651) 484-4368 (home)
`(651) 442-4361 (mobile)
`mgirards@comcast.net
`
`
`MICHAEL J. GIRARD, P.E.
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`PROFESSIONAL PROFILE:
`Technical business leader with over 37 years of engineering and leadership experience, including the last 27 years in
`the medical device industry. Proven track record of developing products and technologies to drive business growth
`for US based and international organizations from start-up level to industry leaders. Demonstrated ability for
`creating technical strategic vision, implementing plans and building technical teams to support the strategy. Provided
`leadership and development for R&D groups of up to 25 people in highly cross-functional environments.
`
`CASE 0:17-cv-03403-DWF-SER Document 41-1 Filed 11/22/17 Page 2 of 7
`
`
`
`
`
`
`
`
`
`
`PROFESSIONAL EXPERIENCE:
`
`April 2015 to
`Present
`
`
`Girard Technical Services, Inc. – Lino Lakes MN
`President - Consultant
`
`Independent consulting services for R&D, engineering and technical management focusing on the medical device
`industry.
`
`
`January 2015 to HLT, Inc. – Bracco Group, Maple Grove MN
`April 2017
`
`
`
`Vice President, Research & Development – Reporting to General Manager
`
`
`April 2009 to
`January 2015
`
`
`
`
`HLT is a medical device company owned by the Bracco Group in Milan Italy that is developing products for
`Transcatheter Aortic Valve Implantation (TAVI). The company experienced deployment issues and halted a clinical
`feasibility study in late 2014 just prior to joining HLT. Initial responsibilities involved leading the redesign and
`development of the delivery system, but then took over the entire R&D group in mid-2015. Reorganized and
`developed the R&D department into a strong team of 20 engineers and technicians. Both the valve and delivery
`system were redesigned and design verification testing completed to demonstrate significant improvement in
`performance from the system that failed in 2014. The redesigned system resumed use in a clinical trial in late 2016
`and now has successfully implanted valves in five patients. In addition to leading the R&D efforts, the Product
`Development Process (PDP) was overhauled to be more effective, a short and long-term product development
`strategy was created and the patent portfolio was managed and expanded to protect the technology. The development
`strategy drove the initiation of several successful projects to fill the product pipeline with further improvements and a
`next generation system.
`
`JVT R&D Corporation/JenaValve, Irvine CA
`JenaValve Technology GmbH, Munich Germany
`
`Head of Valve Development – Reported to VP of R&D (11/12 to 1/15), Reported to CEO (4/09 to 11/12)
`
`JenaValve is a company developing and marketing products intended for the treatment of aortic valve disease based
`on catheter delivered prosthetic heart valve technologies. Started with the company at the early start-up stage
`providing technical expertise and leadership. Responsible for all valve engineering, development and manufacturing
`activities as the company grew from initial prototypes through the clinical trial and commercialization with two
`implantable valve product lines. Included extensive international travel to manage activities and interact with
`company personnel, outside development and contract manufacturing partners in multiple countries. Key
`responsibilities and accomplishments included the following:
`
`
`•
`
`Porcine root heart valve for the 32F transapical delivery approach:
`o
`Finalized the development and manufacturing processes
`o Coordinated in vitro valve testing and manufacturing of the prostheses to support First-In-Man
`(FIM) clinical studies started in 10/09
`o Coordinated all in vitro valve testing for Regulatory Submissions to obtain CE Mark approval
`obtained 9/11
`o Managed valve manufacturing and production activities with a UK based manufacturing partner
`through the commercial launch of the product in Europe.
`o Transferred manufacturing processes for the porcine root prosthesis to a 2nd source manufacturing
`partner in Canada to increase output and minimize production risks.
`Implemented manufacturing improvements to double manufacturing yields from 30% to over
`60%
`
`o
`
`Page 19 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41-1 Filed 11/22/17 Page 3 of 7
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
` Page 2 of 5
`
`
`MICHAEL J. GIRARD, P.E.
`
`
`
`•
`
`•
`
`•
`
`Pericardial heart valve for the 22F transapical and 18F transfemoral delivery approaches:
`Performed directly and/or managed all design, engineering and development for the unique 2nd
`o
`generation prostheses product line
`o Coordinated all in vitro valve testing and manufacturing of the prostheses to support FIM clinical
`studies started in 12/13 and Regulatory Submissions to initiate CE Mark clinical studies in 2014
`o Managed valve manufacturing of the pericardial valve prosthesis with a Canada based
`development and manufacturing partner
`o Transitioned manufacturing management activities to a new JenaValve manufacturing team
`based in Irvine
`Inventor on several key US and International patent applications to protect unique prosthesis design
`features.
`Led the establishment of a valve R&D group in Irvine, CA.
`
`ATS Medical, Inc., Maple Grove MN (now part of Medtronic, Inc.)
`
`Consultant – Reported to VP of R&D
`
`ATS Medical was a surgical cardiovascular medical device company. Provided expertise and services to support
`engineering and development activities. Responsibilities included cryogenic ablation product development and
`improvements for the surgical based Atrial Fibrillation (AF) treatment product line. Coordinated and led less
`experienced engineers with design, testing and implementation activities. Investigated new cryogenic technologies
`for a 2nd generation AF treatment product platform.
`
`St. Jude Medical, Inc., St. Paul MN (now part of Abbott Laboratories)
`
`Senior Director, Research & Development – Reported to VP of R&D, Cardiovascular Division (9/06 to
`4/08), Reported to Division President, Cardiac Surgery Division (4/05 to 9/06)
`
`Responsible for the management and leadership of all research and development activities for the cardiac surgery
`business. This included a department of 25 engineers, scientists and technicians with an annual expense budget of
`approximately $6M. Some of the key individual and group accomplishments and activities include the following:
`
`
`May 2008 to
`April 2009
`
`
`
`
`April 2005 to
`April 2008
`
`
`
`
`
`•
`
`• Responsible for developing business portfolio technology strategy.
`•
`Prepared and presented yearly operational plans and 5 year strategic plans to upper management for the
`cardiac surgery business.
`Launch of new products and line extensions of existing products within the cardiac surgery product
`portfolio. Products launched from 4/05 to 4/08 are expected to generate approximately 20% of Division
`revenue in 2008 ($60M).
`• Completed development and gained approval to start an IDE clinical trial for a completely new aortic tissue
`valve platform.
`• Developed technical strategy, negotiated outside collaborative partnerships and initiated development
`programs on two key future product platforms:
`o Transcatheter beating heart aortic valve replacement
`o
`Polymer valve
`• Worked with legal department and outside attorneys to manage extensive IP portfolio, including filing
`more than 20 key new patent applications for novel technologies.
`Primary technical interface for evaluation of new technologies and business opportunities with physicians
`and other outside organizations.
`Provided materials and analytical chemistry testing for entire Division
`
`•
`
`•
`
`
`October 2003 to Impres Medical, Inc., Edina MN
`April 2005
`
`
`
`
`Vice President, Research & Development – Reported to Chief Executive Officer
`
`Impres Medical was a small start-up company developing technologies in the area of gynecology. Responsibilities
`included directing all research and development strategies and activities focused on demonstrating concept feasibility
`and the development of an intrauterine implant device and delivery system to treat abnormal uterine bleeding.
`Coordinated the device design, manufacture and pathological evaluations of biological tissue response of the device
`in preclinical and clinical studies. Filed several key pieces of IP based on very unique device tissue response and
`performance. In addition, key support was provided to the CEO to help company fund raising activities.
`
`
`
`
`
`Page 20 of 24
`
`

`

`CASE 0:17-cv-03403-DWF-SER Document 41-1 Filed 11/22/17 Page 4 of 7
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
` Page 3 of 5
`
`
`MICHAEL J. GIRARD, P.E.
`
`
`May 1998 to
`October 2003
`
`Acorn Cardiovascular, Inc., St. Paul MN
`
`
`Director, Product Development – Reported to Chief Technology Officer (11/98 to 10/03), Reported to
`President/CEO (5/98 to 11/98)
`
`Acorn Cardiovascular was a start-up company developing a Class III implanted Cardiac Support Device (CSD) that
`provided mechanical support for the ventricular portion of the heart to treat congestive heart failure. Responsibilities
`included the overall management and leadership for all product development activities. Was the second hired
`employee in an instrumental role helping to build the staff, infrastructure and lead development efforts required to
`take the CSD from concept to a clinical and commercial product. This included defining and implementing the
`product development procedure framework to assure satisfying all QSR requirements.
`
`The CSD, packaging and accessories for implantation were developed and ready for first clinical use in only 12
`months, which exceeded initial company expectations. Key responsibilities, activities and challenges that were met
`included the following:
`
`
`
`• Hired staff to support product development and other operational functions.
`•
`Planned and helped coordinate build-out and equipment set-up for lab, prototype shop and pilot
`manufacturing in a new facility.
`• Created Product

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket