throbber
Durability / Wear Testing of Heart Valve Substitutes
`Helmut Reul, Klaus Potthast
`
`Helmholtz-Institute for Biomedical Engineering, Aachen, Germany
`
`Background and aims of the study: The current stan-
`dards for accelerated heart valve testing have consid-
`erable differences in test conditions. Another prob-
`lem arises from the fact that such test systems are not
`standardized at all. It was shown earlier that different
`
`test systems generate totally different valve loading,
`even if operating at standard conditions. The present
`study aimed to improve this unsatisfactory situation
`and to develop a new concept where actual loading of
`valves is measured either in vitro or in vivo under
`
`physiologic conditions and subsequently to repro-
`duce these conditions during accelerated testing.
`Methods: Integral loading forces at valve closure were
`measured for several valve types using a piezoelectric
`force ring within a real-time circulatory mock loop
`under physiologic conditions. This facilitated defini—
`tion of a physiologic loading range. Physiologic load-
`ing was subsequently reproduced in a single-chamber
`accelerated test system. Working conditions obtained
`in terms of stroke, bypass flow and compliance served
`as design criteria for a new test chamber and a com-
`
`plete 12-chamber accelerated testing system.
`Results: The integral loading obtained using the force
`ring showed a correlation with previous in vitro and
`in vivo results of strain-gauged valves. Loading
`forces for mechanical valves are about one order of
`
`magnitude higher than for bioprosthetic valves and
`are strongly related to cardiac output for both valve
`types. At physiologic loading, however, the differen-
`tial pressures across the valves are considerably
`below those given in FDA guidelines.
`Conclusions: This pilot study demonstrates that phys-
`iologic valve loading is reproducible over a wide
`range under appropriate testing conditions. It also
`showed that, at the back-pressures of the currentstan-
`dards, the loading forces during accelerated testing
`exceed the real-time loading forces by far and, thus,
`may provide unrealistically high valve loads. These
`initial findings indicate that amendments of the cur-
`rently valid standards may be need to be accorded.
`
`The Journal of Heart Valve Disease 1998;7t151-157
`
`In 1996 we published a critical review on the state of
`the art of heart valve wear and fatigue testing as repre-
`sented by the current standards of ISO, FDA and CEN
`(1). These standards have considerable differences in
`
`test conditions as far as pressure differences across the
`test valves and total cycle numbers are concerned.
`Moreover, since the testing devices themselves are not
`standardized, different devices generate totally differ-
`ent loading conditions on the valves, even if the tests
`are carried out under the same standard. This study
`clearly demonstrated that, under the current stan-
`~ dards, the conditions of actual, in vivo impact loading
`of a valve cannot be reproduced. Also, the two test
`devices which were compared, although both operat-
`ing at the given standards conditions, generate totally
`different loading conditions. For one tester loading
`decreases with increasing cycle rate, while for the other
`it increases. For all testers, time history of loading and
`
`Address for correspondence:
`Prof. Dr.-lng. H. Rcul, Helmholtzlnstitute for Biomedical Engineer-
`ing, Pauwelsstr. 20, D-52074 Aachen, Germany
`
`pressure difference across the valve are a function of
`test frequency, compliance and other factors.
`To improve this unsatisfactory situation it was
`strongly suggested that the actual loading conditions
`of each valve type be measured either in animal mod-
`els in vivo or within a circulatory mock loop which pro-
`vides physiologic loading conditions;
`this loading
`could then be subsequently reproduced during accel-
`erated wear testing. The present study is set out to
`investigate this concept.
`
`Measurements of valve loading
`under physiologic conditions
`
`Since most valves - and especially bioprostheses - are
`not suitable for the attachment of strain gauges due to
`the lack of appropriate fixation points which are repre-
`sentative for valve loading, a piezoelectric force mea—
`surement
`ring (Kistler,
`type 906 1A, Winterthur,
`Switzerland) (Fig. 1, top right) was used to assess load-
`ing at valve closure. This type of force transducer mea-
`sures the integral loading force acting on the valve.
`
`© Copyright by lCR Publishers 1998
`
`PAGE 1 OF 7
`
`WATERS TECHNOLOGIES CORPORATION
`
`EXHIBIT 1014
`
`WATERS TECHNOLOGIES CORPORATION
`EXHIBIT 1014
`
`PAGE 1 OF 7
`
`

`

`152 Durability/wear testing of heart valve substitutes
`H. Reul, K. Potthast
`
`J Heart Valve Dis
`Vol. 7. No. 2
`March 1998
`
`semiconductor
`pressure transducer
`
`( NF : 100 kHz )
`
`
`
`
`
`inductive
`displacement
`transducer
`
`elektrornagnetic
`vibrator
`W
`
`
`
`Figure 1: Schematic of single-chamber accelerated valve test
`system with piezoelectric force ring in the upper right.
`
`The next step was to correlate these integral loading
`forces with previously obtained in vivo and in vitro
`results (2,3) measured by a strain-gauged 29 mm Bjork—
`Shiley Convexo-Concave (BSCC) mechanical
`tilting
`disk valve. For this purpose, the strain—gauged BSCC
`valve was mounted on the force ring and inserted into
`the mitral position of a physiologic circulatory mock
`loop as previously described (4). Both, strut loads and
`integral loading force were measured simultaneously.
`For these measurements the left ventricular dp/dt
`(according to the FDA guidelines averaged over the
`last 20 ms before valve closure) of the model ventricle
`
`was varied between 500 and 1500 mmHg/s by increas-
`ing the stroke volume at a fixed test rate of 70 per min,
`a mean aortic pressure of 100 mmHg, and a systolic
`duration of 15%.
`
`The results are shown in Figure 2. The upper curve
`represents the force ring measurements which range
`from 25 to 110 N integral loading force at the corre-
`sponding dp/dt values. The two lower curves repre-
`sent the strut loading forces obtained by strain gauges
`in vivo and in vitro for the same valve; these range
`from 2 to 20 N strut loading force. The difference in
`measured forces between the two methods is obvious
`
`and is related to the fact that the strain gauges measure
`only loading of a single strut caused by strut deflection,
`while the force ring measures the integral total loading
`force acting on the closed valve. Nevertheless, both
`
`PAGE 2 OF 7
`
`120
`
`100
`
`z 80
`E2
`u? so
`
`40
`
`20
`
`Integral Loading Force in Vitro
`(Piezoelectric Force Ring)\
`
`Strut Loading in Vitro
`
`Strut Loading in Vivo
`
`0
`
`500
`
`2000
`1500
`1000
`Left Ventricular dp/dt/ mmHg/s —>
`
`2500
`
`Figure 2:111 vivo and in vitro loading of a 29 mm BSCC valve
`as a function of left ventricular dp/dt before valve closure.
`
`methods give a good correlation with left ventricular
`dp/dt and, thus, can as well be correlated among each
`other.
`
`Measurements under accelerated testing condi-
`tions
`
`‘
`
`After establishing the above correlation between inte-
`gral forces and left ventricular dp/dt the valve was
`inserted into a specifically developed single-chamber
`system for accelerated valve testing (also shown in Fig.
`1).
`
`An electromagnetic vibrator (Ling Dynamics, type
`409, Royston, UK) generates a sinusoidal flow through
`the test valve by compression and extension of a metal-
`lic bellows. When the valve is closed, fluid from the
`
`lower chamber flows to the upper chamber via an
`adjustable bypass. This bypass serves simultaneously
`to control the pressure difference across the valve,
`which is measured upstream and downstream of the
`valve by two semiconductor pressure transducers with
`a natural frequency of 100 kHz (Cobe Disposable
`Transducer, Lakewood, CO, USA). Test fluid is water
`at room temperature.
`The test rate was increased from 200 to 2000 per min
`and the integral loading forces were measured. The
`pressure difference across the valve was kept constant
`at 120 mmHg for all test rates. The results are present-
`ed in Figure 3.
`The shaded area represents the range of loading
`forces under all potential physiologic conditions as
`obtained in vivo and in vitro and as depicted in Figure
`2. The central curve shows the measured integral load-
`ing forces under accelerated testing conditions under
`observance of the FDA conditions for accelerated valve
`
`testing such as full opening and closing and 120 mmHg
`pressure difference. The minimally adjustable loads
`
`l
`r
`
`
`
`,
`
`PAGE 2 OF 7
`
`

`

`I Heart Valve Dis
`Vol. 7. No. 2
`March 1998
`
`140
`
`_. ND
`
`.n OO
`
`IntegralLoadingForce/N—> 88
`
`
`
`
`
`3.O
`
`ND
`
`Minimally Adjustable Loading
`
`Durability/wear testing of heart valve substitutes
`H. Reul, K. Potthast
`
`153
`
`
`
`2L/min
`
`7L/min
`
`Force/N——>
`
`
`
`2L/min
`
`7L/min
`
`Figure 4: Integral loading forces for a 27 mm SIM mechanical
`valve and a 25 mm ISP bioprosthesis measured under real—
`time conditions within a circulatory mock loop.
`
`This range of test conditions represents the normal
`physiologic range and can be used for the characteriza-
`tion of corresponding loading conditions. For each
`valve type the measured loading forces were averaged
`over 100 cycles.
`The results (Fig. 4) show that loading for the mechan—
`ical valve is about one order of magnitude higher than
`for the bioprosthetic valve, and loading increases with
`a factor of about two when cardiac output changes
`from 2 l/min to 71/min.
`
`Design of a new fatigue tester
`
`The principal design criteria for a valve test chamber in
`terms of stroke, displaced volume, bypass flow and
`compliance were obtained by means of the above-men-
`tioned single-chamber system (see Fig. 1). Based on
`these criteria a new test compartment for general use
`was designed (shown schematically in Fig. 5). For the
`new design, the electromagnetic vibrator was replaced
`by a swash plate with adjustable stroke. Otherwise, the
`operating principle was the same as already described
`above.
`
`Larger valves generate higher impact loading forces
`than smaller ones. Therefore, since the new test com-
`partment is intended for use with all kinds of valve
`sizes, an adjustable compliance chamber was added for
`additional control of
`loading forces. The desired
`impact force can then be adjusted by varying the fol-
`lowing four parameters:
`
`- stroke of the swash plate;
`—
`test rate;
`
`- air compliance; and
`- bypass throttle adjustment.
`
`The FDA guidelines require that for all valves, test-
`ing should be conducted on three of the largest, medi-
`um and smallest of each valve type. One equivalent tis-
`
`
`
`V
`
`o0
`
`200
`
`400
`
`600
`
`1000 1200 1400 1600 1800 2000 2200
`800
`Test rate / beats / min —>
`
`Figure 3: Integral loading forces within the single chamber
`accelerated test system obtained for a 29 mm BSCC valve at a
`hack—pressure of 120 mmHg (FDA—condition).
`
`are represented by the lower curve. Higher loads can
`be easily generated by adjusting stroke and bypass
`throttle of the test system. These results clearly show
`that physiologic valve loading can be reproduced over
`a wide range under appropriate accelerated testing
`conditions. It is also evident that, under FDA condi-
`tions, only a single loading force which increases with
`higher test rates can be generated for a defined test rate,
`whereas test conditions outside of the current standard
`
`in terms of pressure difference facilitate the adjustment
`of a range of physiologic loading conditions at any test
`rate. Thus, valve loading at resting conditions as well at
`exercise conditions can be adjusted, resulting in a much
`better simulation of the loading history of an implant-
`ed valve.
`
`Valve loading forces for alternative valve types
`
`So far, all presented results are only valid for one single
`valve type and size, a 29 mm diameter BSCC tilting
`disk valve. Therefore, in order to develop a generally
`valid concept, it is necessary to extend the above find-
`ings to other valve types and sizes. For this purpose, a
`St. Jude Medical (SIM) 27 mm mechanical bileaflet
`valve and an Ionescu—Shiley 25 mm pericardial bio—
`prostheses were selected as a first step. Both valves
`were inserted into mitral position of our circulatory
`mock loop and integral loading forces were measured
`under the following experimental conditions (accord-
`ing to FDA guidelines for pulsatile flow valve testing):
`
`Cardiac output: 2 l/ min and 7 l/min;
`Test rate: 70 per min;
`Mean aortic pressure: 100 mmHg;
`Mean atrial pressure: 10 mmHg; and
`Systolic duration: 35%.
`
`PAGE 3 OF 7
`
`PAGE 3 OF 7
`
`

`

`154 Durability/wear testing of heart valve substitutes
`H. Real, K. Potthast
`
`Test compartment of HlA-FT2
`
`7*" 7
`
`Observation Tube
`
`, / Test-Compartment
`
`.: '2» Pressure Taps
`
`, Compliance Chamber
`
`open to atmosphere '
`
`Throttte ,
`
`Testvalve
`Bypass
`
`.7
`
`Piston Rod
`
`
`
`
`)B r“ * Adjustablesn'oke
`Smash Plate
`
`Figure 5: Schematic cross-section of newly developed test com-
`partment for accelerated valve testing.
`
`sue annulus diameter of each type reference valve must
`be tested under identical conditions. This results in a
`
`total of 12 test valves. Accordingly, a 12-cylinder
`fatigue tester (HIA-FTZ) has been designed and manu—
`factured (Fig. 6).
`The first two of the above-listed parameters cannot
`be adjusted individually; they are the same for all 12
`test compartments. Thus, after the correct adjustment
`of one compartment the other 11 must be adjusted by
`variation of air compliance and bypass throttle flow.
`An example (Fig. 7) shows where the peak loading
`forces for a 27 mm SIM valve have been varied by
`changing the air compliance at a constant test rate,
`bypass throttle and stroke settings. Peak load decreas-
`es linearly with increasing compliance volume and can
`easily be adjusted for physiologic loading conditions.
`
`Verification of concept within new fatigue tester
`
`In a final step a 27 mm mechanical SIM valve and a 25
`mm Ionescu-Shiley pericardial valve (ISP) were inserted
`
`PAGE 4 OF 7
`
`I Heart Valve Dis
`Vol. 7. No. 2
`March 1998
`
`
`
`Figure 6: Top view of newly developed 12-chamber accelerated
`test system (HIA—FTZ).
`
`into the new test compartment and the loading forces
`together with the pressure difference across the valves
`were measured within the fatigue tester at a test rate of
`1000 per min for the SIM and 600 per min for the ISP,
`respectively. The results (Figs. 8 and 9) show that for
`both valve types the peak loading forces are within the
`previously determined physiologic range (compare Fig.
`4). The differential pressures, however, are considerably
`below the pressures given in the FDA guidelines, which
`should be adjusted to 120 mmHg for both valve types.
`An example illustrating this controversy for the SIM
`valve (Fig. 10) shows that, in this case, the pressure dif-
`ference across the valve was adjusted to 120 mmHg by
`reducing the test chamber compliance. Test
`rate,
`bypass flow and stroke were kept constant. As can be
`clearly seen, valve loading forces exceed 50 N and,
`thus, are far above physiologic loading.
`
`
`
`Discussion
`
`This pilot study demonstrates that physiologic valve
`loading can be reproduced under appropriate acceler-
`ated testing conditions. However, testing under these
`conditions requires special testing compartments with
`an increased number of control parameters for the
`adjustment of physiologic loads. It also requires a two—
`step testing approach: first, physiologic loading forces
`have to be determined within a real-time circulatory
`mock loop; and second, this loading has to be repro-
`duced within the accelerated tester. For proper transfer
`of physiologic real-time loading forces to accelerated
`testing the test valves have to be mounted within a cal-
`
`PAGE 4 OF 7
`
`

`

`Durability/wear testing of heart valve substitutes
`H. Reul, K. Potthast
`
`155
`
`Force/N—-
`
`U)a:O0
`
`D Ap/mmHg—>
`
`Figure 8: Integral loading force and pressure difference across
`valve for a 27 mm SIM valve at a test rate of 1000 per min.
`
`Durability testing of bioprosthetic heart valves
`requires some additional discussion. The FDA and ISO
`guidelines for the wear testing of tissue valves are very
`similar to those for mechanical valves. Valves are to be
`
`tested for an equivalent of five years at a peak back-
`pressure of at least 90 mmHg for aortic and 120 mmHg
`for mitral valves. The valves should open and close
`completely. For stentless valves, the aortic wall should
`be modeled by a compliant tube.
`Unlike mechanical valves, tissue valves may show
`severe damage as a result of the wear test and may
`become dysfunctional before completion of the experi-
`ment. Possible damage to tissue valves include tissue
`tear, holes, delamination, abrasion, prolapse and stent
`fracture (5,8,9). Table I lists commonly observed dam-
`age in tissue valves, their cause, and critical test para-
`meters that influence the damage. Wear in mechanical
`valves is also included as a reference.
`
`Tissue tear is typically caused by high tensile stress—
`es. Finite element modeling (6) indicates that the high-
`est tensile stresses are in the region of the commissures
`150
`150
`
`27 mm SJM
`
`50 -100
`
`ApImmHg—>
`
`s o
`
`na
`
`o -
`
`Force/N—>
`
`0,00
`
`0.04
`
`0.08
`
`Figure 10: Integral loading force and pressure difference across
`valve for a 27 mm SIM valve at a test rate of 1000 per min,
`but at a preselected pressure difference of 120 mmHg (FDA—
`condition) across the valve.
`
`i Heart Valve Dis
`Vol. 7. No. 2
`March 1998
`
`
`
`
`
`
`Physiologic Range
`according to FDA-Guidelines
`
`70
`
`100
`so
`so
`Compliance Volume / ml—>
`
`110
`
`Figure 7: Integral loading force as a function of air compliance
`volume measured for a 27 mm SIM valve within the new test
`compartment at constant test rate (1000 per min), bypass
`throttle and stroke settings.
`
`ibration compartment, equipped with a force ring
`transducer. Once physiologic loading is adjusted and
`the parameter settings in terms of stroke volume, com-
`pliance volume and bypass flow are obtained,
`the
`valves are mounted Within the geometrically similar
`final test compartments and the continuously mea-
`sured pressure difference serves as the single control
`parameter for long-term studies.
`It could also be shown that at the back-pressures of
`the current standards the loading forces during accel-
`erated testing exceed the real-time loading forces by far
`and thus may provide unrealistically high valve loads.
`These initial findings indicate that corresponding
`amendments of the currently valid standards may be
`necessary.
`
`25 mm ISP
`
`1.00
`
`Force/N—>
`
`Physiologic Range
`
`Ap/mmHg—>
`
`0.0
`
`0,1
`
`0,3
`0,2
`Tlme / s ->
`
`0,4
`
`0.5
`
`Figure 9: Integral loading force and pressure difference across
`valve for a 25 mm ISP bioprosthesis at a test rate of 600 per
`min.
`
`PAGE 5 OF 7
`
`PAGE 5 OF 7
`
`

`

`156 Durability/wear testing of heart valve substitutes
`H. Real, K. Potthast
`
`I Heart Valve Dis
`Vol. 7. No. 2
`March 1998
`
`
`
`
`Table I: Wear and damage observed in cardiac valve prostheses daring accelerated wear testing.
`
`Valve type
`Wear / damage
`Cause
`Critical physical
`Critical test
`
`parameters
`parameter
`
`Mechanical g
`Sliding wear
`Sliding motion
`Range of motion
`Full opening and heart
`
`valve
`closing
`
`
`Impact wear
`Inertia at impact
`Impact force
`Ap, dp/dt, impact force
`
`Ap
`Pressure force at
`High tensile stresses
`Tissue tear at
`Stented tissue valve
`
`commissure closure
`
`Tissue abrasion at
`commissure
`during valve opening
`tissue and stent during
`
`during valve opening
`Tissue deflection
`Contact between
`Abrasion at the base
`AP
`
`at closure
`of the stent
`. leaflets and stent
`
`Contact between
`
`Tissue deflection
`
`Maximum opening
`
`Delamination of
`High bending
`tissue
`stresses, tissue
`during cardiac cycle
`
`buckling
`
`Tissue deflection
`
`Ap, maximum opening
`
`Stentless tissue valve
`
`Tissue tear at
`Ap, pm, aortic wall
`commissure
`closure, aortic wall
`compliance
`
`motion
`
`High tensile stresses
`
`Pressure force at
`
`Contact between
`Ap, pm, aortic wall
`leaflets and inflow
`closure, aortic wall
`compliance
`of cusp
`
`rim
`motion
`
`Abrasion at the base
`
`Leaflet deflection at
`
`Delamination of
`High bending stresses,
`Ap, pm, maximum
`tissue
`during cardiac cycle
`tissue buckling
`opening, aortic wall
`
`compliance
`
`Tissue deflection
`
`Ap, ptm,
`Axial and radial
`High tensile stresses
`Tear in aortic wall
`
`pressure loading
`wall compliance
`AP: back-pressure across the leaflets; ptm: transmural pressure across the aortic wall.
`
`in response to the pressure loading at closure. It is
`therefore important to control the closing pressure
`accurately. In the case of stentless valves, the stresses in
`the leaflets may be redistributed by the flexible aortic
`wall which may act as a shock absorber. Proper model-
`ing of the aortic wall compliance and control of the
`transmural pressure gradients should therefore be con-
`sidered when testing stentless valves.
`tissue
`Tissue abrasion is caused by rubbing of
`against the stent, stent cover or tissue reinforcement. It
`most likely occurs close to the commissures or at the
`base of the cusps where excursion of the tissue is max—
`imum during the cardiac cycle. Maximum deformation
`of the leaflets in the cusp region occurs at maximum
`back-pressure, while maximum deformation of the
`leaflets at the commissure occurs during full opening.
`Thus, control of both peak back-pressure and maxi-
`mum leaflet opening is important when testing for
`leaflet abrasion. It should be noted that the high cycle
`
`rate in accelerated wear testers may not create the same
`leaflet deflection or stresses as in a real—time pulse
`duplicator. Vesely et al. (10) pointed out that porcine
`tissue is viscoelastic, i.e. the stress-strain relationship is
`time-dependent.
`Delamination in the tissue is due to internal shear
`
`stresses. Vesely and Boughner (7) have shown that
`porcine tissue loses some of its natural ability to shear
`when tanned. As a result, bending of tanned tissue
`leads to high internal shear stresses, and potential
`buckling of the tissue (8). Delamination is typically
`observed in the hinge area of porcine valves where the
`tissue is relatively thick.
`As indicated in Table 1, critical test parameters for the
`testing of tissue valves include the maximum back-
`pressure and the maximum valve opening. In some
`commercially available durability testers,
`the back-
`pressure is controlled by a throttle in the flow loop that
`adjusts the bypass flow during closure. As pointed out
`
`PAGE 6 OF 7
`
`PAGE 6 OF 7
`
`

`

`Durability/wear testing of heart valve substitutes
`H. Reul, K. Potthast
`
`157
`
`References
`
`1. Reul H, Eichler M, Potthast K, Schmitz C, Rau G. In
`vitro testing of heart valve wear outside of the manu-
`facturer’s laboratory-requirements and controversies.
`I Heart Valve Dis 1996;5(Suppl. I):S97—SlO4
`2. Rau G, Reul H, Eichler M, Schreck S, Wieting DW. The
`effect of left ventricular dp/dt on the in vitro dynam-
`ics of Bjork—Shiley convexo-concave mitral valves. I
`Heart Valve Dis 1995;4(Supp1. I):Sl 7-820
`3. Schreck S, Inderbitzen R, Chien H, et al. Dynamics of
`Bjork—Shiley convexo-concave mitral valves in sheep.
`I Heart Valve Dis 1995;4(Suppl. I):S 21-825
`4. Knott E, Reul H, Knoch M, Steinseifer U, Rau G. In
`
`vitro comparison of aortic heart valve prostheses. I
`Thorac Cardiovasc Surg 1988;96(6):S952—S961
`5. Walley VM, Keon WI. Patterns of failure in Ionescu—
`Shiley bovine pericardial bioprosthetic valves. I Tho—
`rac Cardiovasc Surg 1987;913:5925-8933
`6. Krucinski S, Vesely I, Dokainish MA, Campbell G.
`Numerical simulation of leaflet flexure in bioprosthet-
`ic valves mounted in rigid and expansile stents. I Bio—
`mech 1993;26(8):S929-S943
`
`7. Vesely I, Boughner DR. Analysis of the bending
`behaviour of porcine xenograft leaflets and of natural
`aortic valve material. Bending stiffness, neutral axis
`and
`shear
`measurements.
`I
`Biomech
`1989;22(6/ 7):3655-5671
`
`8. Vesely I., Boughner DR, Song T, Tissue buckling as a
`mechanism of bioprosthetic valve failure. Ann Thorac
`Surg 1988;416:5302—8308
`9. Thiene G, Bortolotti U, Valente M, et al. Mode of fail—
`
`ure of the Hancock pericardial xenograft. Am I Car-
`diovasc Surg 1989;63:5129-8133
`10. Vesely I, Boughner DR, Leeson-Dietrich I. Biopros-
`thetic valve tissue viscoelasticity:
`Implications on
`accelerated pulse duplicator testing. Ann Thorac Surg
`1995;60(2):S379-S382
`
`I Heart Valve Dis
`Vol. 7. No. 2
`March 1998
`
`earlier, tissue valves may experience notable damage
`during the experiments which often leads to increased
`backflow leakage. To maintain the required peak back—
`pressure, the bypass resistance has to be increased to
`compensate for leakage through the test valves. It is
`therefore desirable to have individual throttles for each
`
`valve in order to maintain the same back—pressure for
`all valves throughout the duration of the experiment.
`The second important test parameter is the degree of
`opening of tissue valves. According to the regulatory
`guidelines, full opening should be achieved in the
`durability tester.
`’Full opening’
`is somewhat
`ill-
`defined, as the degree of opening is a function of the
`flow rate. At Baxter Laboratories, for instance, the
`
`valve is first tested in a pulse duplicator at a cardiac
`output of 5 1/ min. The degree of opening during peak
`forward flow is recorded with a Video camera posi-
`tioned along the axis of the valve. When the valve is
`placed in the durability tester, the stroke of the actuator
`is adjusted to reproduce the same degree of opening as
`in the pulse duplicator. Video images of the valve in
`both systems are used to verify proper tuning.
`In conclusion, the above discussion addresses some
`of the basic considerations for the testing of tissue
`valves. Because of the complex biomechanical proper—
`ties of tissue, it is unclear if and how the accelerated
`
`cycling rate is modifying the stress and strain distribu-
`tion in the valves (10). Furthermore, biochemical
`
`degradation and in situ host response are not consid—
`ered. Biochemical degradation or host overgrowth may
`modify the tissue properties and alter the motion and
`stress distribution in the tissue. Results of accelerated
`
`wear testing of tissue valves should therefore be
`reviewed in the context of the limitations of the test.
`
`Additional knowledge about the viscoelastic behavior
`of fresh and aged tissue will be necessary to further
`refine the accelerated wear testing of tissue valves.
`
`-
`Acknowledgments
`The authors thank Stefan G. Schreck PhD, Director,
`
`Engineering Research and Standards at Baxter Health-
`care Corporation, Irvine, CA and Carlos Rios from the
`same group for their most valuable contribution to the
`discussion on specific aspects for accelerated testing of
`bioprostheses.
`
`PAGE 7 OF 7
`
`: I
`
`. l ,
`
`PAGE 7 OF 7
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket