throbber
IN THE UNITED STATES PATENT AND TRADEMARK OFFICE
`
`In Re the Application of:
`
`Inventors:
`
`Benjamin McCloskey etal.
`
`Confirmation No:
`
`3190
`
`Serial No.:
`
`14/523,104
`
`Group Art Unit:
`
`2855
`
`Filed:
`
`24 October 2014
`
`Examiner:
`
`Philip L. Cotey
`
`Title:
`
`FATIGUE TESTING SYSTEM
`FOR PROSTHETIC DEVICES
`
`
`Docket No.:
`
`P201384.US.05
`
`AMENDMENT B AND RESPONSE TO OFFICE ACTION
`
`MAIL STOP AMENDMENT
`Commissioner for Patents
`PO. Box 1450
`
`Alexandria, VA 22313-1450
`
`Dear Commissioner:
`
`In response to the Office Action dated 20 March 2015, please amend the above-
`
`identified application as follows:
`
`Claim Amendments begin on page 2 of this paper.
`
`Remarks/Arguments begin on page 5 of this paper.
`
`4822-5681-6933
`
`1
`
`PAGE 1 OF 12
`
`WATERS TECHNOLOGIES CORPORATION
`
`EXHIBIT 1007
`
`WATERS TECHNOLOGIES CORPORATION
`EXHIBIT 1007
`
`PAGE 1 OF 12
`
`

`

`Docket No. P201384.US.05
`
`CLAIM AMENDMENTS
`
`The following listing of claims replaces all prior versions and listings of claims in this
`
`application. Additional terms are presented in underline text and deleted terms are indicated in
`
`strikethreugh text or are enclosed in [[double brackets]].
`
`1.
`
`(Currently Amended) A method for operating an accelerated cyclic test system for
`
`evaluating a valved prosthetic device comprising
`
`driving a test system fluid cyclically above a normal physiological rate, at an accelerated
`
`pulsed rate of greater than 200 beats per minute within the test system;
`
`storing a volume of test system fluid in an excess volume area during a system driving
`
`stroke that opens the valved prosthetic device; and
`
`releasing the stored volume of test system fluid during a return stroke that closes the
`
`valved prosthetic device.
`
`2.
`
`(Original) The method of claim 1, wherein the excess volume area enlarges in
`
`response to a pressure on the test system fluid during the driving stroke and decreases during
`
`the return stroke.
`
`3.
`
`(Previously Presented) The method of claim 2, wherein the excess volume area
`
`provides a spring force counter to and in response to the pressure on the test system fluid.
`
`4.
`
`(Original) The method of claim 3 further comprising altering a spring factor of the
`
`spring force provided by the excess volume area through selection of a material forming at least
`
`a portion of a boundary of the excess volume area.
`
`5.
`
`(Original) The method of claim 4, wherein the material is an elastomeric material
`
`that expands and contracts in response to the pressure on the test system.
`
`6.
`
`(Previously Presented) The method of claim 1, further comprising compressing a
`
`volume of a compressible gas with the volume of test system fluid to provide a spring force
`
`counter to and in response to a pressure on the test system fluid when the volume of test
`
`system fluid is stored in the excess volume area.
`
`7.
`
`(Original) The method of claim 6 further comprising altering a spring factor of the
`
`spring force provided by the excess volume area by adjusting the volume of the compressible
`
`gas.
`
`4822-5681-6933
`
`PAGE 2 OF 12
`
`PAGE 2 OF 12
`
`

`

`Docket No. P201384.US.05
`
`8.
`
`(Withdrawn) A device for accelerated cyclic testing of a valved prosthetic device
`
`comprising
`
`a pressurizable test chamber for containing test system fluid and further comprising
`
`a fluid distribution chamber positioned on a first side of the valved prosthetic
`
`device and in fluid communication with a pressure source;
`
`a fluid return chamber positioned on a second side of the valved prosthetic
`
`device;
`
`a fluid return conduit both structurally and fluidily connecting the fluid distribution
`
`chamber to the fluid return chamber; and
`
`an excess volume area in fluid communication with the fluid return chamber
`
`providing a volume for storing a volume of a test system fluid when the test system fluid is under
`
`compression.
`
`9.
`
`(Withdrawn) The device in claim 8 further comprising
`
`a drive motor; and
`
`a fluid displacement member connected with and driven by the drive motor to provide
`
`the pressure source that increases and decreases a pressure on the test system fluid in the test
`
`chamber.
`
`10.
`
`(Withdrawn) The device of claim 8, wherein the excess volume area enlarges in
`
`response to compression of the test system fluid and decreases during depressurization of the
`
`test system fluid.
`
`1 1.
`
`(Withdrawn) The device of claim 8 further comprising an elastomeric material
`
`that forms at least a portion of a boundary of the excess volume area and that expands and
`
`contracts in response to changes in pressure on the test system fluid within the test chamber.
`
`12.
`
`(Withdrawn) The device of claim 8, wherein the excess volume area further
`
`contains a volume of a compressible gas that is compressed by the volume of the test system
`
`fluid to provide a spring force when the volume of the test system fluid is stored in the excess
`
`volume area.
`
`13.
`
`(Withdrawn) The device of claim 8, wherein the excess volume area comprises a
`
`compliance chamber defining a cavity within the fluid return chamber.
`
`4822-5681-6933
`
`PAGE 3 OF 12
`
`PAGE 3 OF 12
`
`

`

`Docket No. P201384.US.05
`
`14.
`
`(Withdrawn) The device of claim 13 further comprising an elastomeric
`
`membrane separating at least a portion of the compliance chamber from fluid in the fluid return
`
`chamber.
`
`15.
`
`(Withdrawn) The device of claim 13 further comprising a porous material at least
`
`partially filling the compliance chamber.
`
`16.
`
`(Withdrawn) The device of claim 13, wherein the compliance chamber provides
`
`a volume for holding a gas or elastomeric material that compresses under a pressure placed
`
`upon the test system fluid in the test chamber and allows the test system fluid in the test
`
`chamber to occupy a portion of the volume in the compliance chamber.
`
`17.
`
`(Withdrawn) The device of claim 8, wherein
`
`the test chamber defines a first port on a first side of the valved prosthetic device and a
`
`second port on a second side of the valved prosthetic device; and
`
`the first port and the second port are configured to receive one or more sensor devices.
`
`18.
`
`(Withdrawn) The device of claim 9, wherein the drive motor is configured to
`
`operate cyclically, acyclically, or a combination of both, to provide cyclic and acyclic fluid
`
`pressures within the test chamber.
`
`19.
`
`(Withdrawn) The device of claim 9, wherein the drive motor comprises a linear
`
`motor.
`
`20.
`
`(Withdrawn) The device of claim 9, wherein the fluid displacement member
`
`further comprises a flexible rolling bellows connected to a shaft of the drive motor.
`
`4822-5681-6933
`
`PAGE 4 OF 12
`
`PAGE 4 OF 12
`
`

`

`Docket No. P201384.US.05
`
`REMARKS
`
`This Response is considered fully responsive to the Office Action mailed 20 March 2015.
`
`Claims 1-20 are pending in the application. Claims 8-20 are withdrawn. Claims 1-7 stand
`
`rejected.
`
`In this Response, claim 1
`
`is amended. Reexamination and reconsideration are
`
`requested.
`
`Interview Summary
`
`Applicant thanks the Examiner and Supervisory Examiner Lisa Caputo for their time on 7
`
`May 2015 and participation in a telephone interview with the undersigned and Craig Weinberg,
`
`Ph.D., one of the inventors.
`
`Dr. Weinberg discussed the differences between the invention claimed in claims 1-77
`
`and the prior art references of record, namely, Pickard and Lundell et aI. Dr. Weinberg noted
`
`that Pickard discloses a “real-time” test system (e.g., operating at physiologic rates on the order
`
`of 72 beats per minute, or 1.2 Hz) for hydrodynamic performance testing of heart valves to
`
`characterize and define their anticipated fluid mechanical performance post implantation Dr.
`
`Weinberg noted that Pickard’s disclosure is thus not an accelerated durability testing system
`
`(e.g., operating at rates 2 3.5 Hz or 200 beats/cycles per minute) like the presently claimed
`
`invention and trying to cycle the Pickard system faster would frustrate the purpose of the test it
`
`is trying to perform (i.e., characterizing valve performance in a simulated circulatory system
`
`under which the valve is to be used) while not being able to perform the accelerated durability
`
`wear testing of the claimed invention. Dr. Weinberg also discussed the different purpose of
`
`system compliance between the claimed durability test system and the hydrodynamic
`
`performance system of Pickard. Dr. Weinberg noted that in the Pickard system, compliance is
`
`used to shape the systemic pressure waveform and modify the systemic pressures to mimic the
`
`response of the circulatory system distensiblity (e.g., arterial vascular compliance) and create a
`
`physiological relevant environment to characterize the prosthetic heart valve performance.
`
`In
`
`contrast, Dr. Weinberg noted the purpose of the excess volume area as claimed in the method
`
`of independent claim 1
`
`is to prevent or minimize the kinetic energy of fluid flow generated by the
`
`system driver from translating into high static fluid pressure in the test system during the
`
`accelerated frequency testing.
`
`The Examiner expressed concern that the term “accelerated” in the preamble of claim 1
`
`was insufficient to differentiate the types of test systems disclosed in the cited prior art from the
`
`test system in which the claimed method operates. While Applicant disagreed with this
`
`analysis, the Applicant and Examiner discussed possible amendments to provide the clarity
`
`4822-5681-6933
`
`PAGE 5 OF 12
`
`PAGE 5 OF 12
`
`

`

`Docket No. P201384.US.05
`
`sought by the Examiner. Applicant noted that the ISO 5840 Standard provides guidelines for
`
`the accelerated durability testing of heart valves and that the industry (Le, a person of ordinary
`
`skill in the art) recognizes an “accelerated” valve test system to mean a system that cycles
`
`faster than a normal physiological rate.
`
`It was discussed that the typical upper end of a normal
`
`physiological rate is above 200 beats per minute. The Examiner agreed that a limitation
`
`describing the system environment as being greater than 200 beats per minute would be
`
`sufficient to address this concern with the claim.
`
`Continuation Application
`
`Applicant asserts that no arguments or disclaimers made in the parent application no.
`
`12/718,316 apply to this continuation/divisional application and, consequently, it is asked that
`
`the Examiner review the present set of claims in view of all of the prior art of record and any
`
`search that he deems appropriate. Applicant presumes the Examiner has considered the file
`
`history in parent application no. 12/718,316 and has determined any rejections therein not
`
`expressly made in the prosecution of the present application to be immaterial to the present
`
`application.
`
`Rejections Under 35 U.S.C. §102
`
`The Examiner has rejected claims 1-3 and 6-7 under 35 U.S.C. §102(b) as being
`
`anticipated by Pickard (U.S. Patent No. 4,682,491). Applicant respectfully traverses this
`
`rejection for at least the following reasons.
`
`Pickard discloses a system and method for testing heart valves “prior to implant in the
`
`human body” as a “mock circulatory loop.” (See Pickard, Abstract.) Pickard states that “the
`
`present invention is particularly useful for mimicking the human circulatory system so that a
`
`heart valve may be placed therein, tested and observed for determining the suitability of the
`
`valve for actual implantation.” (See, Pickard, 1:9-14.) Pickard “seeks to provide a mechanical
`
`analog for the human circulatory system, including the heart, arteries, veins and capillaries so
`
`that a prosthetic valve may be tested and observed prior to use in the human body.” (Pickard
`
`7:8-13; see also 3:9-57.) Thus, Pickard discloses a system and related methodologies for real-
`
`time testing of cardiac valves to examine fluid mechanical performance in a facsimile circulatory
`
`system, not a method in an accelerated system as claimed for the purposes of evaluating leaflet
`
`wear and implant durability over hundreds of millions of cycles in the expected lifetime of a
`
`valve.
`
`As is well known to persons of ordinary skill in the art of cardiac valve testing, real-time
`
`and accelerated testing have completely different requirements and use completely different
`
`4822-5681-6933
`
`PAGE 6 OF 12
`
`PAGE 6 OF 12
`
`

`

`Docket No. P201384.US.05
`
`methodologies for testing. For example, in Section 6.2.1 of the ISO 5840 Standard, Table 1
`
`specifies the “physiological parameters of the intended patient population” including the “Heart
`
`Rate: 30 beats/min to 200 beats/min.” These parameters define the physiological conditions of
`
`operation of heart valve devices for the intended patient population.
`
`5.2.1 Operationalspecifications
`
`
`
`Table 1—Heart valve substitute operational environment
`
`Pa ra m eter Description
`
`
`1g the pri
`nment if? \v
`
`
`
`Surrounding medium:
`
`
`
`Temperature:
`
`Heart rate:
`38 Enema-him to 23$} Seatsvtmiri
`
`
`Ceres-ac eutput: 3 sfmiz‘i to {5 limit}
`
`Sistine L-‘alume:
`
`
`Differential pressur
`
`
`
`
`
`
`
`
`
`
`
`Aorta: 3,3,,
`mm Hg
`
`83
`
`'35
`
`{3:3
`
`3.0
`
`5i:
`
`CC:
`
`
`MB is: ti:
`5‘38 to 99
`323.
`
`
`
`‘lGGte list?
`
`338
`
`Extreme
`presmre
`
`'3 iii i=1
`
`>
`
`163
`
`3'35
`
`:03
`
`2:; ‘~
`
`35
`
`2":
`
`SJC
`
`ANSI/AMMl/ISO 5840, “Cardiovascular Implants—Cardiac valve prostheses,” American National
`
`Standards Institute, Inc. (2005).
`
`Section 7.2.4.2 of ISO 5840 specifically discusses the number of anticipated test cycles
`
`i.e., at least 400 million cycles for stiff or artificial valve leaflets or at least 200 million cycles for
`
`tissue valve leaflets, for durability testing. An excerpt is presented below.
`
`4822-5681-6933
`
`PAGE 7 OF 12
`
`PAGE 7 OF 12
`
`

`

`Docket No. P201384.US.05
`
`TEA? Device durability assessment
`
`An 55855551551: oi“ the durability of the head wisesub:‘t’i‘iuteis‘ shaii be perftiimiad in 5839i :0 555555 minimum
`fusribiien 055'!” 5 reesonabie iifetime Unieszs the:5:351:85: for 5 {iaiiicu-rl.5: deal-rim isnciudas 5i? exrilib‘l‘1 state-nicer
`'
`'
`
`.
`anticipated ii:- Vii-'5 device iifetiiiie, testing: ghali be meriormed :5 denierstrata 55536:ebie assurance the: rig
`waive gubetitute5 wiii remain funrtonal f9: 43% miiiien (235355 and :he: tiexibie heed Vaisre substiiiiteg wisii remain
`
`f:initioriai f0? 2203 mi!iim cycles. itmeeiaiweii
`~' far a Liam-2:315 r device inciudee en explicit etaiemen: about flnfiClflatefi
`is.i‘in iiéiiii:25e‘liieti'me testing shat: be pars‘mn'ieij is flipper: tha- iateiiziig iiiaim.
`
`
`T55~mg shaiihe5 , perfmned m ail-5551:3555 555i": 5:»? the iaigesat medi um and 5n51i395:5i:5-=5f east.h:3~55 {afflict am
`mitral‘ 0: heart val 55 \bbk‘itituie One equiv:aient size r5ieiei1ce wise siiaii D5:esteiij(rider identical imediums far
`
`each ualv5size , ed.
`
`Tests; 5iiali be Qe-rs'nmiedi at a defined difierenhei priesisure corisisiieist with nermmensi‘oie conditieiss Specified iris Tabs-e
`i. Diiiim the duiabiiiisi ”53*:ng the defin5d tan 5: peak difierentiai {3:855:55 acmss :he dissed val-ire 5ha§l be
`
`
`maint-ais‘ied f5: 55 ‘tfn er more of aii :he test 0;
`'
`'
`ach :55: 555:9 shaii e erienm a d=iff5r5ntiei ’prezssui‘e equai to (is
`
`
`
`greater than the defined diffiei'ezntiai pressure
`3: er aware of the dun-3
`n at each cyci5 if mitts and ii "
`' heart
`Veiire stitisfimteis 555 ideizticei
`in d55§gn 5x55
`:5; it:5. 55iwii‘g cuff testing need minis-r be peifm‘nes'i under the
`
`differientiai pressure mnditima 555555 for the mitmi mice
`
`i.g:5 rates used for ac:eieiated a:d quasi res:it time 55:55: :esstizi1g Eéhtflilij be justified free~.~'i :he resuit: 5f :335 3133':
`
`anal:~y“1_ {_03:55:5er; Sammie be given :c- the b5haVi.r {if nmaieisondertiatiina.3 when Qalecrm andsizu.etif'gii1g
`aggirsoimate £15535 53:55
`
`Id.
`
`If such cycle testing for durability were performed at normal heart rates, the testing would
`
`take between 6 and 12 years. Moreover, the third paragraph demonstrates there are no
`
`requirements on the pressure waveform shape during testing, to wit: “Each test valve shall
`
`experience a differential pressure equal to or greater than the defined differential pressure for
`
`5% or more duration of the cycle.” Therefore the waveform shape is not intended to be
`
`physiological, rather it simply has to meet this requirement for differential pressure loading and
`
`thus is not designed to mimic the circulatory system as is the goal in Pickard. The 4th
`
`paragraph of this section demonstrates that the term ‘accelerated’ is common vernacular to
`
`those versed in the art of heart valve development and testing and, since it will need to be
`
`‘justified,’ it assumes that it is within a condition that is not standard physiological conditions.
`
`Additionally, ISO 5840, Annex L, describes the requirements for real-time test systems
`
`(e.g., the Pickard system) and defines that testing should be conducted in a “pulse duplicator
`
`that produces pressures and flow waveforms that approximate physiological conditions....” Id.
`
`Notably, this is not a requirement for durability testing as set forth in ISO 5840, Annex M. Id.
`
`Thus, while the concept of using a compliance chamber to store excess volume of test
`
`fluid in a real-time, physiologically accurate, cardiac valve test system is well known for the
`
`purpose of substituting for the arteries of the human circulatory system (in fact, compliance is
`
`required by ISO 5840 in Annex L and detailed guidelines for compliant chambers are provided
`
`in Annex F), a method in an accelerated cyclic test system that uses an excess volume area is
`
`entirely new. As described in the specification of this application, “the compliance chambers
`
`135 assist in minimizing the effects of large and quickly changing pressure gradients (i.e.,
`
`pressure loading or pressure spikes) across test samples 130 placed within the test chamber
`
`4822-5681-6933
`
`PAGE 8 OF 12
`
`PAGE 8 OF 12
`
`

`

`Docket No. P201384.US.05
`
`106.” (See, 11 0061.)
`
`In methodologies of prior commercial accelerated test systems, there were
`
`no specific design elements or system features to address this phenomenon of pressure spikes,
`
`which were therefore viewed as an accepted drawback associated with testing at accelerated
`
`rates. As described in the present application, the compliance chamber may also be used to
`
`fine-tune the pressure gradient across the valve sample being tested. The prior art of record
`
`fails to teach the claimed method in the context of an accelerated cyclic test system. None of
`
`these concerns, goals, or solutions addressed and achieved by the claimed method are
`
`contemplated or recognized in the prior art of record. Thus, Pickard cannot be held to anticipate
`
`the invention of claim 1 because a method performed within an accelerated cyclic test system
`
`was not even a consideration in the context of the disclosure of Pickard.
`
`Further, with respect to Pickard, a piston pump is disclosed to drive the disclosed
`
`system. A piston pump may be adequate for a real-time (i.e., ~1.2 Hz) valve performance test
`
`system. However, it has significant drawbacks in the context of an accelerated test system
`
`such as claimed in claim 1. For example, recall the ISO 5840 standard requirement that the
`
`valved prostheses be loaded through 200,000,000 test cycles for biologic prostheses or
`
`400,000,000 test cycles for synthetic prostheses. This extremely high cycle requirement
`
`prohibits the use of a standard piston with seal(s) as a driver since because O-rings, cup seals,
`
`and other standard seal structures wear out before the completion of a single test run. (See
`
`Declaration of Craig Weinberg, Ph.D., 11 10.)
`
`(Note: This declaration was submitted during the
`
`prosecution of the parent application to this continuation application.) Further, the friction
`
`caused by the interference between the piston, seal, and chamber generate heat within the test
`
`system and additional wear on the components and driver (ld.).
`
`Dependent claims 2, 3, 6, and 7 depend upon and contain all the limitations of
`
`independent claim 1. For at least the reasons discussed above in connection with independent
`
`claim 1, the Applicant respectfully submits that dependent claims 2, 3, 6, and 7 are allowable.
`
`The Applicant makes this statement without waiving any independent bases for patentability in
`
`claims 2, 3, 6, and 7. The Applicant reserves the right to separately argue the patentability of
`
`dependent claims 2, 3, 6, and 7 in a subsequently filed response, if necessary.
`
`The Applicant therefore requests reconsideration and withdrawal of the rejection of
`
`claims 1-3, 6, and 7 under 35 U.S.C. § 102.
`
`4822-5681-6933
`
`PAGE 9 OF 12
`
`PAGE 9 OF 12
`
`

`

`Docket No. P201384.US.05
`
`Rejections Under 35 U.S.C. §103
`
`The Examiner has rejected claims 4 and 5 under 35 U.S.C. §103(a) as being
`
`unpatentable over Pickard, in view of Lundell et al. (U.S. Publication No. 2002/0116054 A1).
`
`Applicant respectfully traverses the rejection.
`
`Initially, Applicant asserts that dependent claims 4 and 5 depend upon and contain all
`
`the limitations of independent claim 1. For at least the reasons discussed above in connection
`
`with independent claim 1, the Applicant respectfully submits that dependent claims 4 and 5 are
`
`allowable. The Applicant makes this statement without waiving any independent bases for
`
`patentability in claims 4 and 5.
`
`As noted, the deficiencies of Pickard with respect to claim 1 are equally applicable to
`
`claims 4 and 5 which depend therefrom. Lundell et al. fails to remedy the deficiencies of
`
`Pickard. Lundell et al. also discloses a real time system (e.g., 50 to 120 bpm; see Lundell et al.,
`
`11 0109) for testing medical devices “to more closely approximate natural biological conditions in
`
`which pulsed flow circulates fluid... without use of a pump that directly applies pulsatile forces to
`
`the fluid.” (Lundell et al., 11 0044.) “Conditions in the flow system can be adjusted to mimic the
`
`conditions in a patient’s cardiovascular system... [and] can also be used to test and evaluate
`
`cell attachment and proliferation in association with a prosthesis or cell culture support matrix.”
`
`(See Lundell et al., 11 0047; see also 11 0052 (“pulsed flow similar to in vivo conditions”).) Lundell
`
`et al. is thus not an accelerated testing system (e.g., cycles >= 2.5 Hz) like the presently claimed
`
`invention. Cycling the Lundell et al. system faster would frustrate the purpose of the test it is
`
`trying to perform (i.e., “circulating blood, cell culture medium or other fluids containing viable
`
`cells...for seeding biocompatible materials with viable cells...to produce prostheses with
`
`associated cells”; see Lundell et al., 11 0046) while not being able to perform the accelerated
`
`durability wear testing of the claimed invention.
`
`The design shown in Figs. 12 and 13 of Lundell et al. depicts a test system using a
`
`shaker table to provide translational motion of fluid (110) across the valves (264, 266) in tubes
`
`(252, 254). The ends of the tubes are connected to reservoirs (256, 258), which, in one
`
`embodiment, may expand and contract to accommodate changing fluid volume as the fluid
`
`moves from one side of the valves to the other.
`
`In this embodiment, the reservoirs must be able
`
`to accommodate this excess volume because one of the valves will be closed depending upon
`
`direction of the movement of the system and the fluid will back up behind it.
`
`The device of Lundell et al. cannot have a “driving stroke” or “return stroke” as required
`
`by claim 1 because Lundell et al. expressly disavows use of a pulsatile pump after describing
`
`problems with such a pump for its real-time physiologic testing purposes. (See Lundell et al.,
`
`4822-5681-6933
`
`PAGE 10 OF 12
`
`1 0
`
`PAGE 10 OF 12
`
`

`

`Docket No. P201384.US.05
`
`11 0044 (“standard pulsed pumps tend to damage cells”).) The embodiment identified by the
`
`examiner uses, as noted, a shaker table to mechanically translate fluid flow from one side of the
`
`device to the other.
`
`In general, Lundell et aI. states that it causes flow by moving the test
`
`system with respect to the fluid within the system (i.e., it uses a “swashplate” or a “shaker table”)
`
`or it uses a constant flow pump.
`
`In contrast, the claimed method expressly operates by driving
`
`a pulsed flow with a driving stroke and a return stroke.
`
`Further, the combination of Lundell with Pickard et al. is improper and the purported
`
`motivation to combine is fa||acious. Pickard expressly requires a pulsatile pump. Lundell et aI.
`
`expressly disavows the use of a pulsatile pump. Thus, there would be no motivation
`
`whatsoever to attempt to achieve “laminar flow” in Pickard as argued by the Examiner. Laminar
`
`flow would only be achievable by using the swash plate or shaker table of Lundell et al.
`
`The purpose of the excess volume areas (i.e., compliance chamber(s)) as claimed in the
`
`method of independent claim 1
`
`is “to act as a resilient spring force to dampen the effects of
`
`large, quickly changing pressure gradients within the test chamber” in the test system during the
`
`accelerated frequency testing. (See Specification, 10046.) As with Pickard, this is not an issue
`
`considered by Lundell et aI. nor do the reservoirs (256, 258) provide such a function. The
`
`reservoirs merely allow fluid to continue to flow across the open test valve when the fluid head
`
`encounters the closed opposing test closed test valve in the system. There is no teaching in
`
`Lundell et al. of using the reservoirs to alter a spring factor of the spring force provided by the
`
`compliance chamber as recited in claim 4.
`
`The Applicant therefore requests reconsideration and withdrawal of the rejection of
`
`claims # under 35 U.S.C. § 103.
`
`Conclusion
`
`Claims 1-20 are currently pending in the application; claims 8-20 are presently
`
`withdrawn. Applicant has fully responded to each and every objection and rejection in the Office
`
`action dated 20 March 2015, and believes that claims 1-7 are in condition for allowance.
`
`Applicant therefore requests that a timely Notice of Allowance be issued in this case.
`
`The Applicant believes no other fees or petitions are due with this filing. However,
`
`should any such fees or petitions be required, please consider this a request therefor and
`
`authorization to charge Deposit Account No. 04-1415 as necessary.
`
`If the Examiner believes any issues could be resolved via a telephone interview, the
`
`Examiner is invited to contact the undersigned at the telephone number listed below.
`
`4822-5681-6933
`
`PAGE 11 OF 12
`
`1 1
`
`PAGE 11 OF 12
`
`

`

`Respectfully submitted this 17th day of June 2015 by
`
`Docket No. P201384.US.05
`
`
`
`Brad' J. Hesttes'nhach
`Registratiér—‘i‘ No. 42,642
`USPTO Customer No. 20686
`
`Dorsey & Whitney LLP
`1400 Wewatta St., Suite 400
`Denver, Colorado 80202
`Tel: 303-629-3400
`Fax: 303-629-3450
`
`4822-5681-6933
`
`PAGE 12 OF 12
`
`1 2
`
`PAGE 12 OF 12
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket