throbber
Sept. 26, 1944.
`
`G. A. MILLIKAN
`
`OXYGEN METER
`
`2,358,992
`
`Filed June 28, 1941
`
`2 Sheets-Sheet 1
`
`
`
`5
`
`: MHALDT
`
`ZO
`
`3 VA
`SZ
`
`ir
`
`6 g/
`
`78
`
`nN
`

`g
`6
`ONE
`On
`_@
`on
`0
`Sar: ee =fI BS
`22
`
`iNf
`
`BO
`
`IZ
`
`TP
`
`INVENTOR
`14. ANMillikan
`
`Glent
`BY
`
`ATTORNEY
`
`IPR2018-00294
`Apple Inc. EX1015 Page 1
`
`IPR2018-00294
`Apple Inc. EX1015 Page 1
`
`

`

`Sept. 26, 1944.
`
`G. A. MILLIKAN
`OXYGEN METER
`
`2,358,992
`
`Filed June 28, 1941
`
`2 Sheets-Sheet 2
`
`
`>»
`
`
`
`cdSS
`
`AES
`
`
`
` LigA,
`
`Oxyhemoglobti
`
`ft
`
`
`LoLg3.
`
`
`
`aLightabsorption
`
`
`
`O7 Ligaytlhoftransinttledtight
` aeOE “INVENTOR 7.
`adie sue
`Glenn AMMlkan
`
`Wave length & GIVIPreCAIV Piro.
`
`
`ATTORNEY
`
`IPR2018-00294
`Apple Inc. EX1015 Page 2
`
`IPR2018-00294
`Apple Inc. EX1015 Page 2
`
`

`

`Patented Sept. 26, 1944
`
`2,358,992
`
`UNITED STATES PATENT OFFICE
`
`10
`
`18
`
`20
`
`25
`
`30
`
`2,858,992
`_
`OXYGEN METER
`Glenn A, Millikan, New York, N. Y.
`Application June 28,.1941, Serial No. 400,285
`(Cl. 88—14)
`11 Claims.
`scale which will suit that subject within a prac-
`This invention relates to colorimetry, and is
`tical degree of accuracy; and then by direct read-
`particularly useful in determining the amount of
`ings indicates on that scale the percentage of
`oxygen in the blood.
`It provides a means by
`oxygen content
`in the blood. The particular
`which the depletion of oxygen—as in the case of.
`scientific principles by which this is accomplished
`aviators flying at high altitudes or patients under
`will be described more in detail later in the speci-
`anesthesia—can be continuously observed,
`s0
`fication.
`that appropriate measures may be taken before
`In the drawings forming part of this specifica-
`the danger point is reached. Various other appli-
`tion, Fig. 1 is a.schematic view illustrating the
`cations in colorimetry will become apparent to
`general principles of a typical device.
`those skilled in the art. The objects of the inven-
`Fig. 2 is a side elevation view mostly in section
`tion are to provide a simple and practical device
`of a light sensitive cell and lamp unit which may
`for purposes such as those stated;
`to make it
`be clamped to the ear or other desired specimen.
`_ applicable to persons having different physical
`Fig. 3 is a front view of one form of the red
`characteristics:
`to permit frequent checking of
`and green filters used with the light sensitive
`the accuracy of the instrument; to protect the
`cell.
`person being tested from pain or discomfort; to
`Pig. 4 is a chart showing the comparative light
`give greater ease and accuracy of operation; and
`absorption of hemoglobin when high in oxygen
`in general to improve the performance, conven-
`and when reduced in oxygen, from which char-
`ience and utility of stich devices.
`acteristics the function of the green ‘Alter
`is
`The general principles upon which the sappa-
`deduced.
`ratus is based are as follows.
`It has long been
`Fig. 5 shows the logarithmic curves of trans-
`recognized that
`the hemoglobin of
`the blood
`mitted light and oxygen saturation for various
`changes color from red toward blue as the oxygen
`types of ears, labelled “thick ear,” “medium. ear”
`content diminishes.
`In recent years colorimeter
`and “thin ear’; from which the necessity for
`tests involving light passing thru a specimen to
`different scales for different types of ears will be
`fall on a light sensitive cell operating a galva-
`seen.
`nometer have been developed, so as to give some
`Similar reference numerals refer to similar
`means of determination independent of the mere
`parts thruout the various views.
`,
`judgment of the human eye. To calibrate the
`Referring first to Fig. 1, a light sensitive cell
`results laboratory analyses of blood samples in
`{ embodying separate units capabie of respond-
`each instance were required. In order to produce
`ing to different selected colors is placed in prox-
`a more direct reading method, and one adaptable
`imity to the specimen to be analyzed, such as the
`to persons of various physical characteristics, the
`human ear 2, thru which light from the electric
`present device uses a system of readings and
`light bulb 3 passes to energize the light sensitive
`scales by which coordinated results of sufficient
`cell f. To control the color of the light which is
`accuracy for most practical purposes can be
`to energize the celi {, color filters 4 and 4’, la-
`quickly obtained.
`belled R for red and G for green respectively, are
`A beam of light canbe passed thru a thin part
`interposed between the light 3 and the cell .
`of the body, suchas the ear or the web of a
`The light sensitive cell {
`is of the compound or
`finger, to fail upon 4 light sensitive cell operating
`multiple type, so that part of it responds only to
`a galvanometer.
`If all ears were alike, the prob-
`the light falling on it thru the green filter, and
`jem would present few difficulties. But the varia-
`another part responds only to the light falling on
`tions in thickness and texture of the tissues in
`it thru the red filter.
`‘These parts are labelled
`different individuals introduce variations in the
`R for red and G for green, and these letters are
`light transmitted quite independent of the color
`also applied to switch terminals, indicator lights,
`of the blood; so that different individuals with
`rheostats, and similar parts to be later described,
`equal proportions of oxygen in their blood may
`so that the red and green circuits can be readily
`produce quite different photo-electric readings.
`traced.
`:
`:
`The problem then is to provide a sound basis
`The cell { is therefore really two light sensitive
`for. calibration independent of such individual
`ces having a common ground wire § but other-
`peculiarities.
`:
`wise operating independently, one portion being
`In general the problem is solved in the present
`sensitive to green light and the other portion
`apparatus by providing a system which first de-
`being sensitive to red light.
`It is the well known
`termines a classification within which the indi-
`property of any light sensitive cell that when
`vidual subject falls, and selects a calibration
`
`ooen
`
`40
`
`45
`
`50
`
`55
`
`IPR2018-00294
`Apple Inc. EX1015 Page 3
`
`IPR2018-00294
`Apple Inc. EX1015 Page 3
`
`

`

`20
`
`25
`
`30
`
`45
`
`50
`
`55
`
`2 l
`
`2,858,992
`ight falls upon it an electro-motive force is gen-~
`circuit must be provided for the electric light
`erated, which is usually capable of operating a
`21 that shines on the galvanometer mirror. This
`measuring instrument such as a galvanometer,
`circuit is controlled by the switch 22, which really
`When the light thru the green filter 4’ falis on
`has only two positions, on and off; but for con-
`the corresponding portion of the cell {, the cir-
`venience of manufacture it may be provided with
`cuit is thru the conductor 6 to the terminal of
`contact points similar to the switches 1, 14 and
`the switch 1 marked G, then thru the conductor
`15, so that it may be operateg by the sameshaft:
`8 to the galvanometer 9, and back thru the
`but in the case of the switch 22 all the points
`ground wire 5 to the cell #.
`G, Z, and R are connected together so that the
`‘When the red portion of the cell | is in action,
`light 2! remains on regardless of the position of
`the light after passing thru the redfilter 4 falls
`the switch 22, unless it is in the “off” position.
`on, the corresponding portion of the cell i, gener-
`The switches 7, 14, 15 and 22, being of similar
`ating an electro-motive force in accord With the
`mechanical construction, are readily moved in
`intensity of the red light. When the switch 1 is
`unison by a single shaft 24 indicated by the broken
`turned to the terminal R,for the red circuit, the
`line, which is operated by the handle 25. When
`current flows thru the conductor 10, switch 1, and
`this handle 25 moves the switches 1, (4, (8 and 22
`conductor 8 to the galvanometer 9, and back thru
`from the “off” position to G (green), the gal-
`the ground wire § to the cell {. There are thus
`vanometer 9 is put in circuit with that portion of
`two simple alternative circuits, one energized by
`the ight sensitive cell 1 which is under the green
`red light and the other by green light, either of
`color filter 4’;
`the green indicator light
`12.
`is
`which may be read on the galvanometer §, by
`turned on; the light source 3 is turned up toa
`throwing the switch 7 to the appropriate position.
`higher intensity; and the galvanometer reading
`The point marked Z on the switch 7 is used for
`light 21 is turned on. When the handle 25 moves
`setting the zero reading of the galvanometer 9;
`the switches 7, 14, 15 and 22 to the “zero” posi-
`and the point marked O (for “oif”) is used when
`tion, the light sensitive cell circuits are discon-
`the apparatus is entirely shutoff. Since the cur-
`nected, as is the light source 3, but the gal-
`rent is generated by the direct action of the light
`vanometer light 21 remains on, so that the gal-
`falling on the light sensitive cell, no battery or
`vanometer can be set for zero reading, to give it
`other external power source is required for this
`the proper preliminary adjustment. When the
`part of the apparatus, these reading circuits being
`handle 25 moves the switches 1, 14, 15 and 22 to
`self-contained and independent of the supple-
`the R (red) position, the galvanometer is con-
`mentarycircuits now tobe described.
`:
`nected to that portion of the light sensitive cell
`To assist the operator in using the apparatus,
`energized by the light thru the red color filter 4;
`indicator lights 12 and {3 are Provided, the light
`the light source 3 is dimmed to a degree previ-
`{2 being on when the green Circuit is in use and
`ously set by the resistance 18; and the galvanom-
`the light [3 being on when the red circuit is in
`eter light 21 remains on to provide the reading
`use. These lights are operated by the switch 14,
`spot or indicator. Thus the zero setting, and the
`which has four switch points, O (off), R (red),
`readings produced by either the red or the green
`% (zero set), and G (green), similar to the switch
`1
`light are conveniently obtained, and their pres-
`ence indicated, by simultaneous switch movements
`The light source 3 which supplies the light to
`all operated by a single handle.
`the specimen 2 and light cell {, is preferably an
`The electricity for the indicator lights 12, 13,
`ordinary. light bulb of the miniature type, and
`the light source 3, and the galvanometer light 21,
`is controlled by the switch 15, which has four
`may be obtained from any suitable source indi-
`contact points similar to those of switches 7 and
`cated conventionally by the reference numeral 30,
`14, that is, marked O (off), G (green), Z (zero
`ordinary electric light lines being generally used
`set), and R (red). Light sensitive Cells are gener-
`when available.
`ally much more responsive to red rays: than to
`The light sensitive cell ( and color filters @ and
`green, and this would normally necessitate chang-
`4’ have been described as simply red and green
`ing thescale or resistance of the galvanometeror
`units set close together so as to utilize the same
`other measuring instrument.
`In order to take
`light source 3. The green filter andcell are made
`readings from both the red and green circuits
`larger in area than the red filter and cell, be-
`without changing the galvanometer, means are
`cause of the fact that-red light activates the light
`provided to reduce the intensity of the light source
`sensitive cell much more strongly than green, as
`3 when the red screen is in use; or conversely, to
`previously stated; and this difference in the green
`increase therelative intensity of the light source
`and red areas helps to bring the galvanometer
`3 when the green screenis in tse.
`readings into the same general order of magni-
`For this purpose, adjustable resistances iT and
`tude and avoids the necessity for changing the
`{8 are provided in the circuit of the light bulb
`galvanometer scale or
`resistance, particularly
`3, and are arranged so that the resistance {1,
`when this is combined with the alteration of the
`connected to the green circuit terminal G of the
`light source intensity above described.
`,
`switch 15 is less than the resistance '8, which is
`The difference between the red and green
`connected to the red circuit terminal R of the
`areas will best be seen in Fig. 2 and Fig. 3 which
`switch 15. This gives @ brighter light at 3 when
`show in greater detail a light cell unit more
`the switch 5 is at the point G of the green side,
`simply shown in Fig. 1. Referring now to Fig. 2,
`and a dimmerlight when the switch (8 is on the
`the housing 31 contains the color filters 4 and 4’
`point R of the red side. Conductors (9 lead to
`and the light sensitive cell. 1, which is conven-
`the light bulb 3. A voltmeter 20 is connected
`tionally shown in Fig. 2 and in somewhat greater.
`across the circuit of the light bulb 2 as an addi-
`detail in the front view of Fig, 3. The h
`tional means of checking the setting of that light
`31 is preferably made of hard rubber, plastic or
`if desired.
`similar smooth material so that it may be placed
`If the galvanometer: or other Treading instru-
`against the ear without discomfort; and is pro-
`ment § is of the type using a spot of light as its
`vided with an opening 32 to. admit light from
`indicator point, ag is generally the case, then a 76 the electric light bulb 3. This bulb 3 is sup
`
`-
`60
`
`65
`
`70
`
`IPR2018-00294
`Apple Inc. EX1015 Page 4
`
`IPR2018-00294
`Apple Inc. EX1015 Page 4
`
`

`

`10
`
`30
`
`ported on 8 hollow arm 33 extending from the
`housing 31, thru which the wires 19 pess to
`supply electricity to the bulb.3. The enc of the
`arm 33 extends back toward the housing 3{ and
`is provided with a threaded neck 38 on which
`is screwed a knurled ring 36 having smoothly
`rounded surfaces:on the side adjacent the hous-
`ing 31, so that it may be clamped on the human
`ear or other thin part of the body without dis-
`comfort.
`In use, the thin upper portion or shell
`of theearis slipped between the housing 3! and
`the ring 36, and the latter is then screwed up
`until it is gently clamped on the ear. The pres-
`sure should not be enough to impede the circu-
`lation, but only enough to hold the device lightly
`in place and exclude light other than that from
`the bulb 3. A gentle pressure is sufficient and
`is not uncomfortable.
`While the apparatus may be constucted with
`only one green light area and one red light area,
`it is generally desirable to construct the green
`light area in two parts connected together, with
`the relatively small red light area located in
`between, as shown in Fig. 3, where the green
`filter and cell areas are labelled G and the red
`filter and cell area is labelled R. This provides
`& more uniform distribution and guards against
`errors which might be ind.t¢-u by local varia-
`tions in brightness on one side. While such an
`arrangement is mechanically made up of three
`or more parts, in principle they are the equiva-
`lent of but two regions, since all the green areas
`are connected to the same conductor 6 and so
`act as one; and similarly, the red areas, if they
`were divided, would all act as one by being con-
`nected to the same conductor 10.
`The apparatus and the principles which have
`been described would suffice to determine blood
`color and thereby oxygen. content under most
`conditions were it not for the fact that varia-
`tions in the thickness and tissue of the ear in
`different individuals cause wide variations in the
`light transmitted thru the ear,
`irrespective of
`whether the blood in the ear is the same color.
`If all ears had the same thickness and texture, 4:
`relatively simple readings would suffice. But
`that not being the case, it is necessary to pro-
`vide means to overcome errors induced by such
`individual peculiarities.
`The present solution to this portion of the
`problem involves the following general, steps.
`(1) Providing a meansfor classifying the speci-
`mens or ears according to their physical char~
`acteristics of thickness and texture and other
`individual peculiarities affecting the transmis-
`sion of light, independent of the color. of the
`plood, so that, the classifications would not be
`affected by the oxygen content of the moment.
`(2) Having then segregated the ears into groups
`which act alike, the present invention provides
`separately calibrated scales for each group, on
`which can be correctly read the oxygen content
`of any individual in that group, at any particu-
`lar moment.
`The scientific principles on which this solution
`is based will be better understood after referring
`to Fig. 4, which is a typical chart. showing the
`light absorption of oxyhemoglobin, or blood high
`in oxygen—indicated in solid lines— as compared
`with reduced hemoglobin, or blood iow in oxygen,
`shown in broken lines, both curves being plotted.
`with light absorption as the ordinates and light
`wave length, A., that is color, as the abscissae.
`In the chart the spectrum runs from red on ‘the
`right to the blue violet on the left.
`It will be
`
`2,358,902
`seen that on the right the red ght is absorbed
`much more by the reduced hemoglobin than by
`the oxyhemoglobin, as would be expected from
`. the fact that the blood with more oxygen is
`redder in color and so transmits red light more
`freely. On the other hand, the left portions of
`the curves indicate that with the bluer light
`the oxyhemoglobin absorbs more, that is, is more
`resistant to the passage of blue light than blood
`reduced in oxygen, which is bluer.. While these
`characteristics are naturally to be expected, the
`interesting fact appears. in the-middle of. the
`curves that the two curves repeatedly crose sach
`other, with the oxyhemoglobin sometimes. above
`and sometimes below. Where they cross, the two
`curves have of course the same value; which:
`means that there are certain colors or wave
`lengths which are absorbed equally by. either
`oxyhemoblogin or reduced. hemoglobin; that is,
`that light of a certain color will be absorbed to
`the same degree regardless of whether the blood
`is high or low in oxygen. Since ‘with light of
`that color the oxygen content does not enter
`into the problem, that particular color can be
`used to measure the various other physical char-
`acteristics or peculiarities that retard the pass-
`age of light, such as thickness, texture of the
`tissues, color of the skin, etc.
`In general it may
`be said that the amount of the selected green
`light transmitted is determined almost entirely
`by the amount of blood in the ear, independent
`of how much oxygen it contains. The green
`light thus measures what may be termed the
`“blood thickness” of the ear.
`:
`While any one of
`the cross-over points of
`the two curves might be used, it is preferable
`with human subjects to use the cross-over. point
`at about 5900 A., which we have called the
`. “green”light. Using this color, which is absorbed
`40
`in the same degree with any oxygen content of
`the blood, we test out various individual ears
`and find some transmit the light rather freely,
`and others not so freely, according generally as
`they are thin or thick or vary in texture. The
`transmission of the green light, as indicated by
`the action of the light sensitive cell | on the
`galvanometer $, shows the general. resistance to
`light attributable to the personal characteristics
`of the individual, aside from the. oxygen content
`of his blood; and we use such a determination
`of the “blood thickness” of the ear to select the
`proper scale suitable for that type of person.
`Referring now to Fig. 5, which shows the oxy-
`gen saturation plotted against the logarithm of
`the transmitted light, for various types of ears
`labelled “Thick ear,” “Medium ear,” and “Thin
`ear”—it will be seen that the curves are not only
`spaced from each other, but are not parallel, that
`is, have different slopes. This means that the
`calibration curves for such different types of ears
`will be quite different; and any single scale on the
`galvanometer reading directly in terms of percent
`oxygen in the blood would not be accurate forall
`types of people.
`In practice it has been found
`that by using a reasonable numberof scales, pref-
`erably four, the oxygen content can be directly
`read with a sufficient degree of accuracy for most
`practical purposes.
`is provided
`Accordingly the galvanometer 9.
`with four separately calibrated scales labelled
`“Wafer,” ‘“Thin,”. “Medium,” and “Heavy,” as
`shown in Fig. 1, with curves running from 50% to
`100% oxygen content crossing them in a generally
`diagonal direction.
`In operating the apparatus
`these are used in the following manner.
`
`3
`
`IPR2018-00294
`Apple Inc. EX1015 Page 5
`
`35
`
`60
`
`65
`
`70
`
`75
`
`IPR2018-00294
`Apple Inc. EX1015 Page 5
`
`

`

`4
`
`'
`
`20
`
`_ 2,358,992
`First to warm up and properly dilate the blood
`changes in the lamp, voltage, lamp luminosity,
`vessels in the ear, the switches are turned to the
`and sensitivity of the light sensitive cell:
`It is
`green position, which it will be recalled makes
`advisable to check this incident “green” light
`the lamp 3 hotter than in the red position, This
`reading at the beginning and end ofa run to see
`green position is usually too hot for comfort if
`that it has not changed by more than a few per-
`Jeft.on continuously, and therefore during the
`cent,
`initial warming up period of from five to fifteen
`Beer's law, as applied to a mixture of two sub-
`minutes, the light is turned-on intermittently and
`stances, such as hemoglobin and oxyhemoglobin,
`left on each time until the ear begins to feel un-
`states that for a monochromatic light which is
`comfortable. Once the ear is properly vasodi-
`differently absorbed by them, the logarithm of the
`lated,as indicated by a steady reading on the gal-
`transmitted lightis linearly related to the frac-
`vanometer, the dimmerred light should be warm -
`tion of one substance in the mixture. Beer's law
`enough to keep it so, and the green need only be
`has been found to hold adequatelyeven for the
`flashed on occasionally—say at one minuteto five
`optically complex system of blood in the human
`minute intervals—to see that the blood thickness
`ear, so long as the amount of blood in the ear
`of the ear has not changed. This reading with
`remains fairly constant; as determined by the
`the switch on “green” shows how much blood
`“green” reading.
`It should be emphasized that
`there is between the lamp 3 and the photo-cell {,
`there are many reasons why Beer's law should not
`that ts, the “blood thickness” of the ear.
`hold in a system so far removed from that of a
`Whenthis green reading has arrived ata steady
`clear solution of pigment in’a parallel sided
`Value, we find it within one or another of the
`trough, and the happy validity ofthe law for the
`heavy black. blocks shown staggered along the
`human ear within the desired degree of accuracy
`four scales of the galvanometer. For example, if
`does not justify its extension to other tissues
`it were in the region of the figures 60-90 on the
`without independent check.
`:
`upper scale, it would fall within the block of the
`It follows
`from Beer’s law that.
`there ig a
`scale labelled “Wafer”; if the green reading were
`.Straight line relationship between the logarithm
`in the region of 50-60 on the upper scale, it would
`of the transmitted light and the percent satura-
`fall within the block on the scale labelled “Thin”;
`tion. This, however, is only true if. both the in-
`if in the region of 60-70 on the lower scale it would
`tensity of the incident light and the total pigment:
`lie in the block of the scale labelled “Medium”;
`concentration Chemoglobin plug oxyhemoglobin)
`and if in the region of 50-60 on the lower scale, it
`remain unchanged.
`If the light intensity is in-
`would lie in the block of
`the Scale labelled
`creased, this straight line is shifted to the right
`“Heavy.” With fewer or more scales the blocks
`in Fig. 5 without changeofslope, while if the Pig-
`would be different but the principle would be the
`same,
`.
`ment becomes more concentrated its slope is de-
`creased. The position of the line can be uniquely
`Since the blocks do not overlap, the galvanom-
`determined if its slope is known and if one point
`eter indication on the “green” setting is always
`on it is determined experimentally. The first
`within one block or another, and a glance at the
`quantity can be predicted from the “green” read-
`scales suffices to select
`the scale to ‘be used.
`ing, while the second datum can be obtained either
`Whichever scale has the block in which.
`the
`by forcing the saturation up.to 100% by breathing
`“green” reading falls is the scale on which the
`oxygen or by assuming the saturation is 96%
`Percent oxygen is read when the “red” setting is
`when air is breathed normally,
`used.
`The effective wave lengths depend both upon
`The proper scale having been so determined, the.
`the transmission characteristics of
`the filters
`switch is then moved to the “red” position. The
`used and upon the spectral Sensitivities of the
`four scales are so arranged that in each case 100%
`light sensitive cell. The choice of the color fil.
`saturation corresponds to the full scale deflection
`ters therefore depends to some extent on the par.
`of the measuring instrument.. The adjustment
`ticular photocell used. For the red light an ef-
`of the chosen scale is obtained by varying the in-
`fective wave length is desired which is very dif-
`‘ cident. “red”light intensity by meansof the rheo-
`ferently absorbed by reduced hemoglobin than by
`- stat [8 until the correct reading for a known oxy-
`oxyhemoglobin. A number of regions might be
`gen content (100% with oxygen or 96% with air)
`used, but for measurements on man the most
`is obtained for one point on the scale, Any other
`suitable region is from 6200 A. to 6600 A. For
`degree of oxygen can then be read directly on
`example, a “Wratten No, 29” filter having a con-
`that scale. The reading is quick, almost instan-
`trol wave length of 6400 A. was found to be sat-
`taneous; and by holding one’s breath, can be seen
`isfactory. For the green light an effective wave
`to drift down froma normal content of about
`length is desired which is equally absorbed by
`96% to say 80%, 70% or lower, depending on the
`oxyhemoglobin or reduced hemoglobin. Such a
`depletion of oxygen occurring. This rapid and
`region lies between 5200 A. and 6000 A., and vari-
`continuous reading of the oxygen content of the
`ous other points as indicated in Fig. 4. A band
`blood, without taking blood samples, is the chief
`in the neighborhood of 5900 A., obtained with
`advantage of the instrument.
`“Wratten No, 61” filter has been found satisfac-
`Care should be taken that carbon monoxide is
`tory. By keeping in mind the principles above
`not: present, as its effect on the color of the blood
`outlined various suitable selections can be made.
`is similar to oxygen.
`The apparatus has been described in the form
`In order to compare one ear with another, the
`of a single unit capable of handling one person
`incident light, in the “green” position of the
`at a time. Where it is desired to test a number
`switch, must always be the same,
`‘This is secured
`of persons simultaneously, multiple types can
`by initially placing a neutral filter of constant
`be made by the mere duplication or multiplica-
`transmission in the position of the ear and then
`tion of the corresponding parts.
`:
`adjusting the brightness of the lamp until the in-
`While I have in the foregoing described certain
`strument comes to a predesignated mark on the
`particular embodiments of the invention, it will
`calibration scale, determined by the original cali-
`be understood that they are merely for purposeg
`bration from known data. This method of ad-
`of illustration to make clear the Principles there-
`Justment automatically compensates for slow
`of, and that the invention is not limited to the
`
`5
`
`10
`
`2h
`
`ao
`
`AC
`
`45
`
`50
`
`55
`
`60
`
`65
`
`70
`
`if)
`
`IPR2018-00294
`Apple Inc. EX1015 Page 6
`
`IPR2018-00294
`Apple Inc. EX1015 Page 6
`
`

`

`5
`
`15
`
`20
`
`25
`
`35
`
`40
`
`30
`
`2,858,002
`‘strument operable by either of said cells, said in-
`particular form described, but is subject to vari-
`strument having different scales with marks at
`0US modifications and adaptations in different
`different locations thereon to indicate which type
`installations as will be apparent to those skilled
`scale is to be used, said indication being made by
`in the art without departing from the scope of
`operation of the first mentioned filter and cell,
`the invention as stated in the following claims.
`and different oxygen calibrations on said scales
`TI claim:
`1, In an oxygen meter, the combination of a —
`whereby the oxygen content may be directly
`read by operating the instrument by the second
`light source, 2 color filter and light sensitive cell
`color filter and cell and reading the oxygen con-
`for the selection of types operable by said light
`tent on the scale previously indicated, means for
`source and responsive to wave lengths which
`increasing thebrightness of the light source when
`are equally absorbed by oxyhemoglobin and re-
`the first color filter and cell is used and means
`duced hemoglobin, a second color filter and light
`for reducing the brightness when the second
`sensitive cell for oxygen determination operable
`filter and cell is used, a switch selectively con-
`by said light source and responsive to wave
`nected to eitherlight sensitivecell and the meas-
`lengths which are absorbed differently by reduced
`uring instrument, a second switch connected to
`hemoglobin than by oxyhemoglobin,: a measur-
`the means. for rendering operative the brightness
`ing instrument, a switch for connecting the
`control means of the light source according. to
`measuring instrument
`to either light sensitive
`which filter and cell is used, and means for op-
`cell.so that it is operable by either of said cells,
`erating said switches simultaneously.
`said instrument having different scales with
`5. In an oxygen meter, means for obtaining
`marks at different locations thereon to indicate
`electrical indications comprising in combination
`which type scale is to be used, said indication be-.
`@ light source, a red light filter and light sensi-
`ing made by operation of the first mentioned
`tive cell, a green light filter and light sensitive
`filter and cell, and different oxygen calibrations
`cell, said green filter and cell being of greater
`on said scales, whereby the oxygen content may
`area than the red filter and cell whereby effects
`be directly read by operating the instrument by
`of the same general order of magnitude are ob-
`the second color filter and cell and reading the
`tained when applied to oxygen determination in
`oxygen content on the scale previously indicated.
`the blood, means for attaching said light source,
`2. In a colorimeter, the combination of a light
`filters and cells to the ear, whereby they may re-
`filter and a light sensitive cell responsive to green
`spond to the color of the blood in the ear, and a
`light which is equally absorbed by: oxyhemo-
`measuring instrument operable by either of said
`globin and reduced hemoglobin, the type of the
`cells.
`specimen being determined by its absorption of
`6. In an oxygen meter, means for obtaining
`green light, a second light filter and light sensi-
`electrical indications comprising in combination
`tive cell responsive to red light, a light source for
`@ light source, a red light filter andlightsensi-
`said filters and cells, an electrical measuring in-
`tive cell, a green light filter and light sensitive
`strument operable by either of said cells, said in-
`cell, said green filter and cell being of greater
`strument having differently calibrated scales for
`area than the red filter and cell, a measuring in-
`different types of specimens, said calibrations in-
`strument operable by either of ‘said cells, and
`dicating the absorption by the specimen of the
`means for increasing the brightness of the light
`red light, said instrument also having type in-
`source when the green filter and cell is used and
`dicators on said scales for use with the green
`for reducing the brightness when the red filter
`light to determine which scale is to be used, said
`and cell is used, whereby the same generalscale
`types being independentof the oxygen content of
`of measurement may be used with either the red
`hemoglobin due to the equal light absorption,
`or the green light indications when applied to
`whereby direct readings of the red light absorp-
`oxygen determination in the blood.
`tion may be madeon the proper scale selected
`7. In an oxygen meter, the combination of a
`according to the green light absorption.
`light source, a color filter and light sensitive cell
`3. In an oxygen meter, the combination of a
`for the selection of types operable by said light
`light source, a green color filter and light sensl-
`source and responsive to wave lengths which are
`tive cell, a red color filter and light sensitive cell,"
`equally absorbed by oxyhemoglobin and reduced
`a measuring instrument for said cells, a rheo-
`hemoglobin, a second color filter and light sensi-
`stat in circuit with the light source to dim it
`tive cell for oxygen determination responsive to
`when the red filter and cel

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket