throbber
July 2, 1963
`
`w. E. TOLLES
`
`3,095,872
`
`BLOOD PRESSURE MEASUREMENT
`
`'
`
`Filed Nov. 27, 1959
`
`25$
`
`2 ,_ _\
`
`YE. ._
`\- _
`
`_'
`
`*
`
`— \\
`
`‘
`
`ll
`
`I0
`
`l3
`I
`Oscillator
`Driver
`
`l2
`
`Brachial
`l,\\
`Artery
`:
`\
`'. x '4
`mm
`T ‘
`
`gle
`Ampli'fier
`1 and Phase
`Detector
`
`l5
`
`in
`’
`7 Indicator
`
`A FIG 2
`
`22
`
`Time —->
`
`tllllll'l'lvl'l l'l
`"I I H
`I
`
`Time _,_
`
`Time " "
`
`23
`
`(A)
`
`(B)
`
`(C)
`
`o
`
`0
`
`0
`
`o
`
`
`
`Volts —> Pressure
`
`
`
`Pressure -—:>
`
`O
`
`24
`
`‘
`
`FIG. 3
`
`3|
`
`32
`
`g
`_.)_
`Ampel‘lfler N Limiter ru
`Filter
`
`|
`
`l
`
`32)
`3| 3 ‘
`Ampel‘lfler N; Llmner 11/
`Filter
`
`w_—__
`33-“, Phase
`Detector
`l
`Indicator
`
`l7
`
`INVENTOR
`Walter E. Tolles
`BY
`9‘
`
`ATTORNEYS
`
`1
`
`

`

`United States iiatent
`
`1
`
`ice
`
`3,095,872
`Patented July 2., I963
`
`10
`
`15
`
`20
`
`25
`
`2
`of an individual may be expected not. to change marked
`ly in a short time, measurement of the velocity of
`propagation of the heart pulse wave has been proposed
`as a measurement of blood pressure.
`Direct measurement of the time of travel of heart
`pulse waves between two spaced points along an artery
`has been proposed, but is subject to considerable inac
`curacies. The heart pulse wave is quite complex, and
`it is di?icult to select precisely the same point on the
`Wave as it passes two spaced points so as to get an ac
`curate measurement. Also, the outputs of transducers
`used to pick up the heart pulse wave may be expected
`to vary in amplitude due to changes in the intimacy of
`contact with the artery over which they are placed,
`slight displacements thereof, etc., unless the subject is
`immobilized. This further impairs accuracy of meas
`urement.
`Proposals have also been made to apply single impact
`pulses, or a train of impact pulses, to the artery and
`measure the time of travel thereof between spaced points
`along the artery. However, considerable distortion of
`the pulses occurs as they pass along the artery, so that
`measuring dii?culties arise similar to those present in
`measuring the heart pulse waves. Also, changes in the
`intimacy of contact and location of both the transmitting
`and receiving transducers cause amplitude variations,
`thereby impairing measurement accuracy.
`In accordance with the present invention, alternating,
`substantially symmetrical continuous-wave pressure var
`iations are impressed on the arterial blood stream at a
`frequency substantially higher ‘than the frequency of
`hear-t contraction, and variations in the time of travel
`or phase of the continuous-wave pressure varia ions be
`tween points spaced along the arterial blood stream are
`measured. Due to the changing propagation character
`istics of the artery as a function of the blood pressure
`therein, the continuous-wave pressure variations become
`phase-modulated (or frequency-modulated, since frequen
`cy is the time derivative of phase) as a function of the
`blood pressure. In effect, the applied continuous-wave
`pressure variation is a carrier which is phase-modulated
`by the blood pressure variations. Accordingly a signal
`picked up by a suitable transducer mounted over the
`artery may then be amplitude-limited so as to eliminate
`errors due to varying amplitude, While at the same time
`preserving the necessary information as to blood pres
`sure variations. The phase-modulation may then be
`detected to obtain a signal varying with the blood pres
`sure over a heart beat cycle.
`Advantageously, changes in phase of the applied pres
`sure wave between two spaced points along the arterial
`blood stream are measured. With a ?xed-frequency
`applied pressure variation, changes in phase between two
`?xed points are inversely proportional to changes in ve
`locity of propagation, and hence give ‘an indication of
`changes in blood pressure.
`In a preferred embodiment, a transmitting transducer
`is mounted over an artery and energized with a sinusoidal
`wave having ‘a frequency substantially higher than the
`frequency of heart contraction. The transducer pro
`duces corresponding sinusoidal pressure variations in
`the arterial blood stream. These pressure variations are
`advantageously small compared to the normal arterial
`blood pressures. A pair of receiving transducers are
`mounted at spaced points along the arterial blood stream,
`preferably at a su?icient distance from the transmitting
`transducer so that pickup of pressure waves transmitted
`through the ?esh rather than the artery is negligible.
`Accordingly, the signal outputs of the pair of receiving
`transducers will correspond to the pressure variations
`in the arterial blood stream ‘at the respective spaced
`points. A phase detector is then connected to receive‘
`
`1
`
`3,695,872
`BLOGD PRESSURE MEASUREMENT
`Walter E. Tolles, Oyster Bay, N.Y., ‘assignor to Cutler
`Hammer, Inc, Milwaukee, Wis, a corporation of Dela
`ware
`
`Filed Nov. 27, 1959, Ser. No. 855,699
`9 Claims. ({1}. 123-2415)
`
`This invention relates to a method of and apparatus
`for measuring arterial blood pressure. The invention
`enables blood pressure to be measured continuously,
`and the apparatus is capable of automatic operation.
`In many ?elds of medicine there is a need for appara
`tus capable of continuously measuring arterial blood
`pressure in a simple and convenient manner. The need
`is particularly keen in hospital operating rooms and the
`like, and in many experimental ?elds of medicine, for
`example in space ‘medicine. In the latter ?eld partic
`ularly, there is great need for continuous blood pressure
`measuring apparatus which will allow moderate activity
`by the subject and will not produce any general disco-In
`fort.
`In conventional sphygmomanometry, the subject must
`remain relatively quiet for many seconds and withstand
`periodic circulatory occlusion. Furthermore, the meas
`urement gives only two pressure values (i.e. systolic and
`diastolic) over a considerable number of pulse cycles,
`which results in the discarding of a substantial amount
`of information.
`Automatic methods of measuring blood pressure are
`known, based on the sphygmomanometer and digital
`plethysmograph principles. In addition to the limited
`information given, these methods are unsatisfactory for
`monitoring active subjects and the periodic occlusion of
`the blood circulation causes discomfort. The weight
`and bulk of the required auxiliary air supply, and equip—
`ment to calibrate for environmental pressure, are also
`disadvantages.
`Continuous measurement of blood pressure is possible
`by arterial cannulation, but serious drawbacks are pres
`ent. In addition to the trauma of initial arterial punc
`ture, movement of an active subject may cause sec
`ondary traumata and there is danger that the cannula
`may be dislodged from the artery.
`Accordingly, it is an object of the present invention
`to provide a method and apparatus for continuously
`measuring arterial blood pressure without interfering
`with the blood circulation or unduly impairing the physi
`cal activity and general comfort of the subject.
`It is a further object of this invention to provide ap
`paratus which ‘continuously and automatically records
`arterial blood pressure without the need for constant
`supervision by an attending physician.
`The invention is based on the fact that the velocity of
`propagation of pressure waves in the blood stream varies
`with changes in arterial blood pressure. Although the
`relationship is not linear, it is a single-valued function
`and hence changes in velocity of propagation can be
`used to indicate changes in blood pressure.
`As is well known, the beating of the heart causes heart
`pulse pressure waves to travel down the arteries. Since
`the arterial walls are elastic, expansion and contraction
`of the walls take place as the pressure changes during
`a heart beat cycle. This results in transverse waves
`which propagate down the arterial blood stream with a
`velocity of propagation which varies with the blood pres-,
`sure and the elastic condition of the arterial walls.
`Commonly, velocities of propagation of tens of feet per
`second are encountered in humans, as contrasted with
`velocities of propagation of longitudinal compression
`waves of thousands of feet per second common in liquids.
`Inasmuch as the elastic condition ‘of the arterial walls
`
`40
`
`45
`
`50
`
`55
`
`65
`
`2
`
`

`

`3,095,872
`
`4
`F of the impressed wave and the wavelength A by the
`following equation:
`
`(1)
`c=hF
`Thus, for a constant frequency F, a measurement of
`wavelength will give the velocity of propagation.
`The velocity of propagatilon can also be expressed
`as:
`
`(2)
`CZD/f
`wher D is the distance between the two points of meausre
`ment and t is the time required for the wave to travel
`from one point to the other. It is also possible to ex
`press the velocity of propagation as:
`
`10
`
`15
`
`the outputs and respond to changes in the phase angle
`therebetween.
`‘By employing continuous-Wave excitation and measur
`ing changes in phase angle of the pressure variations at
`spaced points, important advantages may be obtained.
`Some well-known phase detectors are inherently insensi
`tive to amplitude variations, and where necessary limit
`ers may be employed in the output circuits of the re
`ceiving transducers, so that the phase measurement is
`substantially independent of amplitude variations. Thus,
`variations in input signal amplitude, different amounts
`of attenuation between the transducers and the artery,
`different amounts of attenuation along the arterial blood
`stream, etc. will not markedly affect the accuracy of
`the measurement. Furthermore, since only a single
`frequency sinusoidal wave is required for measuring pur
`poses, ?ltering may be employed to eliminate the effect
`of ambient environmental pressures which might give
`rise to extraneous outputs from the receiving transducers.
`Also adverse effects due to frequency distortion in the
`artery are eliminated.
`The ‘invention will be more fully understood by ref
`erence to the following description of speci?c embodi
`ments thereof, taken in conjunction with the drawings,
`in which:
`FIG. 1 is a block diagram of a system embodying the
`principles of the present invention;
`FIGS. 2A through 2B are waveforms explanatory of
`the system of FIG. 1; and
`~
`'
`FIG. 3 shows a modi?cation of the receiver portion
`of FIG. 1.
`'
`'
`Referring to FIG. 1, a transmitting transducer 10 is
`strapped or otherwise a?ixed to an arm 11 of a subject
`over the brachial artery '12. The transducer may be of
`any suitable type, for example a crystal or magnetic
`transducer. The brachial artery is particularly suitable
`since it is sufficiently close to the surface to permit pres
`sure waves to be impressed thereon and responses to
`be obtained therefrom. However, other arteries such
`as those in the leg or neck may be used if desired.
`Transducer 10 is energized from oscillator 13, the
`output of which is advantageously sinusoidal in wave
`form and substantially constant in frequency. The me
`chanical vibrations of the transduceryapply an alternating
`force to the artery wall which in turn produces alternat
`ing pressure variations in the arterial blood stream.
`These constant-frequency pressure variations are super
`posed on the normal blood pressure variations and are
`advantageously maintained at a low amplitude level com
`pared to the blood pressures expected to be encountered,
`for example an amplitude variation corresponding to
`a few millimeters of mercury.
`A pair of pressure-sensitive receiving transducers 14
`and 15 are strapped or otherwise af?xed to the arm over
`the brachial artery downstream from transducer 10, in
`order to respond to the incremental pressure waves pro
`duced by the transmitting transducer and yield corre~
`spending outputs. The receiving transducers may be
`small pressure-sensitive microphones of the crystal, mag
`netic or capacity types, or the like. They are advanta
`geously mounted in a unit with ?xed separation there
`between.
`The transmitting transducer, in impressing pressure
`variations on the artery, may also produce pressure
`waves in and near the skin. Consequently it is ad
`vantageous to place the receiving transducers su?iciently
`far away from the input transducer to avoid responses
`to such pressure waves.
`The pressure waves impressed on the arterial blood
`stream by transducer 10 will travel along the artery at
`a velocity of propagation which varies with the blood
`pressure, and accordingly will become phase-modulated
`'thereby. It is possible to measure changes in the veloc
`ity of propagation in several ways. As is well known,
`the velocity of propagation c is related to the frequency
`
`where (p is the phase angle between the sinusoidal waves
`at the two points.
`It is preferred to obtain an indication of the blood
`pressure by measuring the phase angle between the out
`puts of the two receiving transducers ‘14, 15. vTo this
`end an ampli?er and phase detector 16 is supplied with
`the outputs of the receiving transducers and yields an
`output which varies with the phase angle therebetween.
`The ‘output of the phase detector is supplied to a suitable
`indicator 17, which maybe a meter, recorder, etc, as
`desired.
`The operation of the apparatus of FIG. 1 is illustrated
`by the waveforms shown in FIG. 2. Here, FIG. 2A
`shows a continuous plot of the arterial blood pressure as a
`function of time covering approximately one heart pulse
`cycle. The maximum (systolic) pressure developed
`during the heart contraction is shown at 21, and the
`minimum (diastolic) pressure present during heart ex—
`pansion is shown at 22. This waveform will repeat at
`the frequency of heart contraction, normally of the order
`of 72 cycles per minute.
`FIG. 2B illustrates the constant-frequency sinusoidal
`pressure variation produced in the arterial blood stream
`by the transmitting transducer 10. The frequency is
`substantially higher than the frequency of heart con
`traction, for example, of the order of 50 to 300 cycles
`per second. This sinusoidal pressure variation is super
`posed on the normal arterial pressure variation, giving a
`resultant pressure variation illustrated in FIG. 2C. FIGS.
`2A and 2B are drawn to different scales for clarity.
`As indicated in FIG. 2C, the impressed sinusoidal pres
`sure variation has an amplitude which is small compared
`to the normal arterial blood pressure. Thus, the instan
`taneous blood pressure measured at any time is substan
`tially unchanged by the presence of the superimposed
`continuous-wave pressure variation.
`FIG. 2D illustrates the output of the phase detector
`16. As the blood pressure rises to the maximum or
`systolic pressure, the velocity of propagation increases
`and accordingly the phase angle between the outputs of
`receiver transducers 14 and =15 decreases. A number of
`types of phase detectors are known in the art and may
`be employed as meets the requirements of a given ap
`plication. It is here assumed that the phase detector
`yields a maximum output at zero phase angle, and lower
`outputs as the phase angle increases. Consequently, the
`output varies in a manner similar to the blood pressure
`variation.
`It should be understood that the output of the phase
`detector will not necessarily be linearly proportional
`to the blood pressure. However, variations in blood
`pressure will be readily apparent, and the magnitude
`may be determined by suitable calibration of the indica
`tor.
`It is advantageous to supply the output of the
`phase detector to a continuous recorder so that a com
`plete record of the subject’s blood pressure may be
`obtained. The recorder output may be calibrated in
`units of pressure as indicated in FIG. 2E, where the
`
`25
`
`30
`
`35
`
`40
`
`45
`
`50
`
`55
`
`60
`
`65
`
`70
`
`75
`
`3
`
`

`

`3,095,872
`
`5
`systolic pressure 23 and diastolic pressure 24- are shown
`for one heart cycle.
`Calibration may be accomplished in any convenient
`manner. For example, the subject’s blood pressure may
`be determined by conventional techniques at the same
`time it is obtained by the apparatus of the invention, and
`the indicator calibrated accordingly.
`It will be noted that by making the impressed fre
`quency high compared to the heart beat frequency, the
`manner in which the blood pressure varies during a
`single heart beat cycle may be determined. The fre
`quency components in the blood pressure variation will
`also be considerably lower than the impressed frequency,
`and may readily be eliminated in the ampli?er so as not
`to interfere with the measurement.
`It may often be desired to measure a subject’s blood
`pressure in an environment where ambient pressure
`variations are present. There may also be amplitude
`variations in the outputs of the receiver transducers due
`to many factors. For example, even though the output
`of oscillator 13 is maintained constant, the pressure
`variation impressed on the arterial blood stream by trans
`ducer iii may vary due to variations in the attenuation
`from the skin to the artery. There may also be a
`varying attenuation along the arterial blood stream, and
`between the artery and transducers l4 and 15. It is ‘ac
`cordingly highly desirable that the indicated blood pres
`sure be made independent of environmental conditions
`and insensitive to amplitude variations. ‘In such cases
`the embodiment shown in FIG. 3 may be employed.
`Inasmuch as a substantially ?xed applied sinusoidal
`freqeuency is employed for measuring purposes, it is
`possible to amplify and sharply ?lter the outputs of the
`receiving transducers so as to prevent response to other
`than the desired frequency, thereby eliminating the
`e?ects of environmental pressure changes at other fre
`quencies. Accordingly, the outputs of transducers 14
`and 15 are supplied to respective ‘ampli?ers and ?lters
`31, '31’. The ?lters are tuned to the frequency of
`oscillator 13 (FIG. 1) so that frequencies substantially
`di?erent therefrom will be discriminated against.
`The outputs of ampli?er-?lters 31, 31’ are then sup
`plied to respective limiters 32, 32' so as to substantially
`eliminate amplitude variations. With suf?cient ampli?
`cation and limiting, the outputs of limiters 32, 32’ may be
`substantially square waves, whose relative phase is the
`same as that of the input sinusoidal waves. These square
`waves are then supplied to a phase detector 33 and thence
`to indicator 17. Suitable circuits for measuring the phase
`between square waves are well known in the art. In the
`event that a phase detector is employed which is in
`herently insensitive to amplitude variations, limiters 32,
`32’ may be omitted.
`As illustrated in the drawings, it is preferred to em—
`ploy a pair of receiving transducers spaced from the
`transmitting transducer, and utilize Variations in the phase
`angle between the receiver transducer outputs in order
`to obtain an indication of blood pressure. However,
`it is possible within the broad scope of the invention to
`employ only a single receiving transducer and measure
`variations in the time of travel of the continuous~wave
`pressure variations between transmitting and receiver
`transducers. Here also, it is preferred to measure changes
`in phase between the impressed pressure variation and
`the received pressure variation.
`The invention has been described in connection with
`preferred embodiments thereof. It will be understood,
`however, that variations and modi?cations of the ar
`rangements described may be made within the spirit and
`scope of the invention.
`I claim:
`1. Apparatus for continuously measuring arterial blood
`pressure which comprises transmitting transducer means
`and driving ‘source for externally impressing an alternating
`substantially symmetrical continuous-wave pressure varia
`
`10
`
`15
`
`20
`
`25
`
`30
`
`35
`
`40
`
`45
`
`50
`
`55
`
`60
`
`65
`
`tion on the arterial blood stream at a frequency substan
`tially higher than the frequency of heart contraction and
`with an amplitude which is substantially smaller than
`the normal arterial blood pressures, whereby the im
`pressed continuous-wave pressure variation is phase-modu
`lated in accordance with the arterial blood pressure over
`the heart beat cycle as the continuous-wave pressure varia
`tion travels along the artery, receiver transducer means
`externally positioned over the arterial blood stream and
`spaced from the transmitting transducer means and re
`sponsive to said phase-modulated continuous-wave pres
`sure variation to yield an output corresponding thereto,
`and detecting means insensitive to amplitude variations
`for detecting the phase-modulation of the output of the
`receiver transducer means.
`2. Apparatus in accordance with claim 1 including
`?lter means between the receiving transducer means and
`the detecting means for passing outputs having the fre
`quency of the continuous-wave pressure variation and dis
`criminating against frequencies substantially different
`therefrom.
`3. Apparatus for continuously measuring arterial blood
`pressure which comprises transmitting transducer means
`and driving source for externally impressing an alternating
`substantially symmetrical continuous-wave pressure varia
`tion on the arterial blood stream at a frequency substan
`tially higher than the frequency of heart contraction and
`with an amplitude which is substantially smaller than
`the normal arterial blood pressures, whereby the im
`pressed continuous-wave pressure variation is phase
`modulated in accordance with the arterial blood pres
`sure over the heart beat cycle as the continuous~wave
`pressure variation travels along the artery, 1a pair of
`receiving transducers externally positioned over the arte
`rial blood stream at spaced points therealong and re
`sponsive to said continuous-wave pressure variation to
`yield respective outputs, and detecting means insensitive
`to amplitude variations for detecting changes in the rela
`tive phase of the outputs of the receiving transducers.
`'4. Apparatus in accordance with claim 3 including
`means for mounting said pair of receiving transducers
`with substantially ?xed separation therebetween.
`5. Apparatus in accordance with claim 3 including
`?lter means between the receiving transducers ‘and the de
`tecting means for passing outputs having the frequency
`of the continuous-wave pressure Variation and discrimi
`iiating against frequencies substantially different there
`rom.
`6. Apparatus for continuously measuring arterial blood
`pressure which comp-rises transmitting transducer means
`and driving source for external-1y impressing an ‘alternat
`ing substantially symmetrical continuous-wave pressure
`variation on the arterial blood stream at a frequency sub
`stantially higher than the frequency of heart contraction
`and with an amplitude which is substantially smaller
`than the normal arterial [blood pressure, whereby the
`impressed continuous-wave pressure variataion is phase
`modulated in accordance with the arterial blood pressure
`over the heart best cycle as the continuous-wave pressure
`variation travels along the artery, 1a pair of receiving
`transducers externally positioned over the arterial blood
`stream at spaced points therealong and responsive to said
`continuous-wave pressure variation to yield respective
`outputs, means for limiting the outputs of the receiving
`transducers to respective predetermined amplitude levels,
`and means ‘for detecting changes in the relative phase of
`said amplitude-limited outputs. '
`7. Apparatus in accordance with claim 6 including
`?lter means between the receiving transducers and the
`means for limiting the outputs thereof for passing out
`puts having the frequency of the continuous-wave pres
`sure variation and discriminating against frequencies
`substantially different therefrom.
`8. The method of measuring arterial blood pressure
`which comprises externally impressing an alternating sub
`
`70
`
`75
`
`4
`
`

`

`3,095,872
`
`7
`stantially symmetrical continuous-wave pressure varia
`tion on the ‘arterial blood stream having a frequency sub
`stantially higher than the frequency of heart contraction
`and an amplitude which is substantially smaller than the
`normal arterial blood pressures, whereby the impressed
`continuous-wave pressure variataion is phase-modulated
`in accordance ‘with the arterial blood pressure over the
`heart beat cycle as the continuous-Wave pressure variation
`travels along the ‘artery, externally deriving a signal corre—
`sponding to said phase-modulated continuous-Wave pres
`sure variation at a point over the arterial ‘blood stream
`spaced from ‘the point of impressing thereof, ‘and detect
`ing the phase-modulation of said signal substantially in
`dependent of the amplitude thereof.
`9. A method in accordance with claim 8 including ex
`
`5.
`
`15
`
`8
`ternally deriving signals corresponding to the eonti11u—
`ous-wave pressure variataion at a pair of spaced points
`along the arterial blood stream, limiting ‘the amplitudes
`of the signals, and measuring the relative phase of the
`amplimde-limited signals.
`
`References Cited in the ?le of this patent
`UNITED STATES PATENTS
`Grabau _____________ __ Aug. 30, 1949
`Henning _____________ __ July 18, 1950
`Sheer ______________ __ Nov. 10, 1953
`Uemura et val. ________ __ May 27, 1958
`Brown ______________ __ July 14, 1959
`Barnett et a1 ___________ __ July 12, 1960
`
`2,480,646
`2,515,221
`2,658,505
`2,836,173
`2,894,595
`2,944,542
`
`5
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket