`Robert J Barsottt, PhD
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`The measurement of ammonia, now known to be a normal constituent of all
`
`
`
`
`
`
`
`
`
`
`
`
`bodyfluids, is franght with problems. An elevated ammonialevelin blood
`
`
`
`
`
`
`
`
`
`
`
`
`
`(100 pmol/L or higher)is an indicator of an abnormality in nitrogen home-
`
`
`
`
`
`
`
`
`
`
`ostasis. The collection, handling, storage, and analysis of blood samples,
`
`
`
`
`
`
`
`
`
`
`their limitations, and potential sources oferror are discussed. Newtech-
`
`
`
`
`
`
`
`
`
`
`niques that permit continuousorreal-time estimates of systemic ammonia
`
`
`
`
`
`
`
`
`
`
`
`
`
`levels over a broad rangeare also discussed. The aim should always be to
`
`
`
`
`
`
`
`
`
`
`
`minimize the release of ammonia from the collected sample before analysis.
`
`
`
`
`
`
`
`
`
`
`Recommendations are made on the collection and processing of blood sam-
`
`
`
`
`
`
`
`
`
`
`
`
`ples, for it is by standardization and rigid adherence to these techniques
`
`
`that the reliability of the test results will be improved. (J Pediatr 2001;
`
`
`
`
`
`
`
`
`
`
`
`
`138:511-S20)
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`monia levels are approximately 30
`Umol/L, and levels exceeding 1 mmol/L
`
`
`
`
`
`
`
`
`
`occur underconditions of acute hyper-
`ammonemia.”° Overthe past 100years,
`
`
`
`
`
`
`numerous methods have been devel-
`
`
`
`
`
`
`
`
`
`
`oped to measure ammonialevels in var-
`
`
`
`
`
`ious bodyfluids including blood,plas-
`
`
`
`
`
`ma, erythrocytes, saliva, sweat, and
`
`urine.® This brief review considers a
`
`
`
`
`
`few of the most common methods cur-
`
`
`
`
`
`
`
`
`
`
`
`
`rently used to measure ammonia in
`
`
`blood: alkalization-diffusion, enzymat-
`
`
`
`
`
`
`ic, ion exchange, and electrode. Sample
`
`
`
`
`
`collection, handling, storage, and some
`ofthe limitations and potential sources
`
`
`
`
`
`
`oferrors associated with these methods
`
`
`
`
`
`
`
`
`
`
`
`are discussed. Finally, some promising
`
`
`
`
`
`novel
`techniques for the continuous
`
`
`
`
`
`
`monitoring of ammonialevels that may
`be used clinically in the near future are
`
`
`
`
`
`
`
`
`also described.
`
`
`
`SIFT
`
`
`
`
`PREANALYTICAL
`METHOD For
`
`
`HANDLING BLOOD
`
`AMMONIA
`
`DETERMINATION
`SAMPLES
`
`
`
`
`Ammonia is nowconsidered a normal
`
`
`
`
`
`
`
`
`
`
`itive proof was provided by studies
`
`
`
`
`
`
`
`
`
`
`
`showing that the enzyme glutaminede-
`constituentofall body fluids, resulting
`from the metabolism of amino acids.
`
`
`
`
`
`
`
`
`
`hydrogenase specifically reacts with
`ammonia, O-ketoglutarate, and a coen-
`However, doubts of the presence of
`
`
`
`
`
`
`
`
`
`zyme, reduced nicotinamide-adenine
`free ammoniain biologic solutions per-
`
`
`
`
`
`
`
`dinucleotide, to form glutamic acid, !
`
`
`
`
`
`
`
`
`
`sisted until approximately 1960, Before
`This reaction has becomethe basis for
`this time the various methods available
`
`
`
`
`
`
`
`
`
`
`
`
`
`GLDH=Glutamine dehydrogenase
`
`
`
`for the determination of ammonia con-
`
`
`
`
`
`
`
`
`
`
`the most commonly used blood ammo-
`NADH—Reduced nicotinamide-adenine
`
`
`
`dinucleotide
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`tent in biologic fluids relied on the sep-
`nia assayin clinical chemistry. One of
`
`NADPH Reduced nicotinamide adenine
`aration or release of ammonia from the
`the attractive features ofthis assay is
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`dinucleotide phosphate
`
`that ammoniais determined directly at
`
`
`
`
`
`
`
`
`
`
`
`fluid sample byvolatilization alter the
`Selected-ion flow tube
`
`
`
`addition ofan alkaline solution. With
`
`
`
`
`
`
`
`
`
`physiological pH without previous
`these methods, increased levels of am-
`
`
`
`
`
`
`
`
`
`
`
`
`treatment of the sample with either an
`
`acid or a base.”"5
`
`
`
`
`
`
`
`monia had been reported in the blood
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`of patients with severe hepaticfailure.
`Today, an elevated ammonia blood
`
`
`
`
`
`
`
`
`
`
`
`
`Despite such reports, doubts about the
`level is considered a strong indicator of
`presenceoffree ammoniain blood con-
`
`
`
`
`
`
`
`
`
`
`
`an abnormality in nitrogen homeostasis,
`
`
`
`
`
`
`
`
`
`
`
`
`the most commonrelated to liver dys-
`tinued, mainly as a result of the con-
`
`
`
`
`
`
`
`
`
`
`
`
`cern that free ammonia measured by
`function. In excess, ammoniais a potent
`these methods resulted from its libera-
`
`
`
`
`
`
`
`
`
`
`toxin, principally of central nervous
`
`
`
`
`
`
`
`
`
`
`
`
`
`tion from labile amides in blood during
`system function. In the venous blood of
`incubation in alkaline solutions. Defin-
`healthy adults and children, blood am-
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Frant the Department of Pathology, Thomas Jefferson Univeraity, Philadelphia, Pennayloanta.
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Reprint requests: Robert J. Barsotti, PhD, Associate Professor, Department of Pathology, Anato-
`myand Cell Biology, Thomas Jefferson University, 1020 Locust St, Philadelphia, PA 19107.
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Copyright © 2001 by Mosby,Inc.
`0022-3476/2001/335.00 + 0 9/0/111832
`
`
`
`
`doi:10,.1067/mpd.2001.111832
`
`
`
`
`
`
`
`
`
`
`
`Most methods recommendcollecting
`
`
`
`
`
`
`from patients who havefasted for at
`least 6 hours with the use of a verified
`
`
`
`
`
`
`
`
`
`
`
`ammonia-free heparin as an anticoagu-
`
`
`
`
`
`lant. Heparin is the preferred anticoag-
`ulant, because it has been shown to
`
`
`
`
`
`
`
`
`
`
`
`reduce red bloodcell ammonia produc-
`
`
`
`
`
`
`
`SII
`
`
`
`Par Pharmaceutical, Inc. Ex. 1022
`Par v. Horizon, IPR of Patent Nos. 9,254,278, 9,095,559, and 9,326,966
`Page 1 of 10
`
`

`

`BARSOTTI
`
`
`
`
`
`
`
`
`
`
`pH Dependenceof the Fraction of the Ionized Form
`
`
`
`
`of Ammonia in Solution
`
`
`
`
`
`0.60
`
`
`
`e3
`
`
`
`
`
`
`NH,"/NH,Ratio é
`
`
`
`
`
`8 7
`
`
`
`
`
`.0
`
`
`
`8.0
`
`
`
`10.0
`
`
`
`11.0
`
`
`
`9.0
`
`pH
`
`
`
`
`
`
`
`THE JOURNAL OF PEDIATRICS
`
`
`
`
`
`JANUARY 2001
`
`
`ANALYTICAL METHODS
`
`FOR THE
`
`
`DETERMINATION OF
`
`AMMONIA
`
`
`
`
`
`
`
`
`
`
`
`
`Manyof the procedures for ammonia
`
`determination involve 2 general steps:
`
`
`
`
`
`
`
`
`
`
`
`the release of ammonia gas or capture
`of ammonium ions from the sample
`
`
`
`
`
`
`and the quantitation ofthe liberated
`
`
`
`
`
`
`
`
`
`
`
`
`
`gas or captured ions. Over the years
`the most common methods used to
`
`
`
`
`
`
`volatilize ammonia have been distilla-
`
`
`
`
`tion and aeration/microdiffusion, ion-
`
`
`
`exchange chromatography, and blood
`
`
`
`or plasma deproteinization.
`
`
`
`
`
`
`Properties ofAmmonia
`
`
`
`
`
`
`
`
`In attempting to understand the ra-
`tionale used to measure ammonia,it is
`
`
`
`
`
`
`
`important to review some ofthe physi-
`
`
`
`
`
`
`cal properties of the compound. Am-
`
`
`
`
`
`monia (NH;) is a colorless, acrid-
`
`
`
`
`
`smelling gas at room temperature and
`
`
`
`
`
`
`
`
`
`
`
`
`pressure. It easily dissolves in water
`andionizes to form NH," asfollows:
`
`
`
`
`
`
`
`
`
`NH, + H,O @ NH,‘+ OH"
`
`An increase in the pH or tempera-
`
`
`
`
`
`
`ture ofthe solution increases the level
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`of the ionized form. Fig 1 shows the
`ratio that exists in plasma between the
`
`
`
`
`
`
`
`ionized or NH,‘
`form versus the
`
`
`
`
`
`
`gaseous or NH, form as a function of
`
`
`
`
`
`
`
`
`pH. Thus in plasma at pH 7.4,
`the
`
`
`
`
`
`
`
`
`NH,* form represents approximately
`
`
`
`
`
`
`
`
`
`
`
`
`98%of the total ammonia. Manyofthe
`
`
`
`
`
`approaches used to estimate ammonia
`levels in body fluids involve volatiliza-
`
`
`
`
`
`tion of the NH,’ form of ammonia into
`
`
`
`
`
`
`
`
`its gaseous form, NH, ,byalkalization
`
`
`
`
`
`
`
`
`
`
`
`
`of the sample to a pH +10.
`
`
`
`
`Distillation
`
`
`
`
`
`
`
`
`Oneofthe earliest techniques for the
`measurement of ammonia involves the
`
`
`
`
`addition olan alkaline buller to a sam-
`
`
`
`
`
`
`
`
`
`
`
`
`
`ple of blood followed by in vacuo dis-
`
`
`
`
`
`tillation. The released ammonia gasis
`
`
`
`
`
`collected in
`trap containing an
`a
`
`
`
`
`
`aliquot ofdilute acid, which converts
`
`
`
`
`
`
`
`
`
`
`Fig 1. Ratic between ionized and free gas form of ammonia in plasma as function of pH.
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`tion. EDTA can also be used. Donors’
`
`
`
`
`
`
`
`
`
`
`
`
`arms should be as relaxedas possible,
`because muscle exertion may increase
`
`
`
`
`venous ammonia levels.’
`Blood is
`
`
`
`
`drawn into a chilled, heparinized vacu-
`
`
`
`
`um tube that is immediately placed on
`
`
`
`
`
`
`
`
`
`
`
`
`ice, and plasmais separated within 15
`
`
`
`
`
`
`minutes. It
`is crucial to keep blood
`samples cold after collection, because
`
`
`
`
`the ammonia concentration ofstanding
`
`
`
`
`blood and plasma increases sponta-
`
`
`
`
`
`
`
`
`
`
`neously. Most ofthis increase has been
`
`
`
`
`
`attributed to the generation and re-
`lease of ammonia from red bloodcells
`
`
`
`
`
`
`
`and the deamination of amino acids,
`
`
`
`
`
`
`
`particularly glutamine.®!! Plasma am-
`
`
`monia levels of whole blood main-
`
`
`
`
`
`tained at 4°Care stable for <1 hour.
`
`
`
`
`
`
`
`
`
`
`
`
`
`When promptly separated from blood,
`
`
`
`
`
`
`plasma ammonia levels are stable at
`4°C for 4 hours and for 24 hours if
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`stored frozen at —20°C. To put the
`problem ofrising ammonia levels in
`
`
`
`
`
`
`perspective, the total nitrogen concen-
`
`
`
`
`
`
`
`
`
`
`tration in venous plasma ofhealthy
`
`
`
`
`
`adults exceeds ] mol/L and represents
`a potential pool of free blood ammo-
`
`
`
`
`
`
`nia.'* In normal healthy adults, homeo-
`
`
`
`
`
`stasis maintains free ammonia levels at
`
`
`
`
`
`
`
`
`approximately 30 [mol/L.
`
`
`
`
`
`
`
`It is important to note that the crite-
`ria for sample stability and the meth-
`
`
`
`
`
`
`ods for the measurement of ammonia
`
`
`
`
`
`levels in blood were established almost
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`exclusively with blood specimens from
`
`healthy subjects.® Significantly more
`
`
`
`difficult but much more constructive
`
`
`
`
`
`would be the establishment ofsimilar
`
`
`
`
`
`
`criteria for blood ammonia measure-
`
`
`
`
`ments from patients with metabolic or
`
`
`
`
`
`
`
`
`
`
`liver pathologic conditions. Standing
`
`
`
`
`
`
`blood or plasma samples from these
`patients contain numerous elevated
`
`
`
`
`
`
`
`
`
`sources of ammonia, resulting in in-
`creases in the rate and amount of am-
`
`
`
`
`
`
`
`monia formed. Some ofthese sources
`
`
`
`
`
`
`
`
`
`
`
`include elevated levels of circulating
`
`
`
`deaminase, Y-glutamyltransferase, an
`enzyme that may deaminate [ree amino
`
`
`
`
`
`
`
`
`
`
`acids, particularly glutamine in blood
`and plasma samples, resulting in over-
`
`
`
`
`
`estimates of blood ammonia levels.
`
`
`
`
`
`
`
`
`
`
`Fast, careful handling and preparation
`
`
`
`
`
`
`of blood samples is required, especially
`
`
`
`
`
`
`from patients with metabolic or liver
`pathologic conditions, to minimize pre-
`
`
`
`
`analysis increases in ammonia concen-
`
`
`
`
`tration until other assay techniques or
`
`
`
`
`
`
`
`
`
`
`
`
`methods are developed. Until this is
`accomplished, measurements in this
`
`
`
`
`patient population, in which blood am-
`
`
`
`
`
`
`
`
`
`
`monia levels require the closest moni-
`toring, will continue to be the most un-
`
`
`
`
`
`
`
`
`
`
`
`
`
`reliable. To date, the easiest and most
`
`
`
`
`cost-effective methodis the stringent
`
`
`
`
`and diligent maintenance of blood
`
`
`
`
`
`samples on ice before, during, and
`
`
`
`alter plasma separation.
`
`
`
`
`
`
`$12
`
`
`
`Par Pharmaceutical, Inc. Ex. 1022
`Par v. Horizon, IPR of Patent Nos. 9,254,278, 9,095,559, and 9,326,966
`Page 2 of 10
`
`

`

`THE JOURNAL OF PEDIATRICS
`
`
`
`
`
`
`Votume 138, Numeer |
`
`
`
`
`
`
`
`
`
`
`the gas into the nonvolatile ammonium
`
`
`
`
`
`ion. This approach is rather cumber-
`some and slow, particularly when
`
`
`
`
`
`
`
`
`many samples require analysis, al-
`thoughinitially, speed was the primary
`
`
`
`
`
`
`
`
`
`
`
`
`advantage of this approach over the
`
`
`
`early microdiffusion techniques de-
`scribed in the following text. This ad-
`
`
`
`
`
`
`
`
`
`
`
`vantage, however, was temporary, and
`was lost with the development of
`
`
`
`
`
`
`smaller microdiffusion vessels. The last
`
`
`
`
`
`
`
`
`
`
`
`reported use ofdistillation was by
`
`Burg and Mook! in 1963.
`
`
`
`
`
`BARSOTTI
`
`
`
`
`Image available
`
`
`
`in print only
`
`
`
`
`
`Aeration/Microdiffusion
`Techniques
`
`
`
`
`
`
`
`Anotherearly technique thatis still
`in use relies on liberation of free am-
`
`
`
`
`
`
`
`
`
`
`
`
`monia byalkalization by the addition
`
`
`
`
`
`
`of a strong base to the specimen. The
`
`
`
`
`released ammonia diffuses through an
`air- or nitrogen-filled gap and is
`
`
`
`
`trapped in acid within the same appa-
`
`
`
`
`
`
`
`
`
`
`
`ratus. This approach, introduced by
`
`Conwayand Berne!* in 1933, uses a
`
`
`
`
`
`
`
`
`
`
`
`glass container resembling a Petri dish,
`within which a smaller second cham-
`
`
`
`
`
`ber is centered. The wall of this inner
`
`
`
`
`
`
`
`
`chamber is approximately half ofthat
`
`
`
`
`
`
`
`
`
`
`
`
`
`of the outside wall. An aliquot ofa
`standard acid solution or ammonia in-
`
`
`
`
`
`
`
`
`
`
`
`dicator such as bromocresol green is
`
`
`
`
`
`
`
`placed into the inner chamber, and the
`sample is added to the outer chamber
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`(Fig 2). A measured quantityofbaseis
`
`
`
`
`
`
`
`added to the sample, and the “petri”
`dish is sealed and gently rotated to mix
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`the sample and base in the outer cham-
`
`
`
`
`
`
`ber and then left at room temperature
`
`
`
`
`
`
`for 90 minutes. This same principle is
`used in the Blood Ammonia Check-
`
`
`
`
`
`er!®!® and in the Kodak Ektachem
`
`
`
`
`
`
`
`dry-film method.'” In these systems
`
`
`
`
`
`the diffusion distance for ammonia is
`
`
`
`
`
`
`
`
`
`
`
`
`significantly reduced from those in the
`
`
`
`
`
`original diffusion apparatus of Conway
`
`
`
`
`
`
`
`and Berne, requiring only 5 minutes to
`
`
`complete (Fig 3).
`
`Distillation and microdiffusion both
`
`
`
`
`
`
`represent purification procedures that
`
`
`
`
`
`isolate ammonia from the manyother
`
`
`
`
`
`constituents of blood and plasma, re-
`
`
`
`
`
`
`ducing the possible effects of other
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Fig 2. Early microdiffusion apparatus for determination of blood ammonia. (Reproduced with per-
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`mission from Conway E, Byme A. An absorption apparatus for the micro-determination of ammonia.
`
`
`
`
`
`Biochern J. 1993:27:4 19-29, © the Biochemical Society,)
`
`
`
`
`Current Micro-diffusion Apparatus
`
`Sample compartment
`
`
`
`
`Polyethylenefilm
`
`
`
`
`
`
`
`
`
`
`
`
`
`Tablet (Alkaline Buffer)
`Cc
`over
`
`
`>~—Polyethylene spacer
`
`Color pHindicator es
`ii—Window
`
`
`
`
`
`
`
`
`
`
`
`
`Probe Light Beam
`
`Detector
`
`
`
`
`
`
`\\lustration of micrediffusion technique used in Blood Ammonia Checker (Reproduced with
`Fig 3.
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`permission from Tada Kk, Okuda K,Watanabe Kk, et al. A new method for screening for hyperam-
`
`
`
`
`
`
`
`
`
`
`
`
`monemia. Eur | Pediatr: 1979;| 30;105-10.)
`
`
`
`
`
`
`
`
`
`
`
`constituents and drugs on the ammo-
`
`
`
`
`
`
`
`nia assay. One main drawback ofthese
`procedures is that varying amounts of
`
`
`
`
`
`
`ammonia are liberated from the alka-
`
`
`
`
`
`
`
`
`
`
`line hydrolysis of proteins, especially
`
`
`
`
`hemoglobin, and labile amides, espe-
`
`cially glutamine.!*29
`
`
`
`
`
`
`Ton-Exchange Chromatography
`
`
`
`
`
`
`
`
`In this approach ammonia gasis not
`
`
`
`
`
`liberated from the sample. Instead, a
`strongly acidic cation-exchange resin
`
`
`
`
`
`
`
`
`
`
`is used in batch mode to capture am-
`
`
`
`
`
`
`monium ions, NH,'. The resin is
`added to and subsequently separated
`
`
`
`
`
`
`
`
`
`
`
`from the sample by centrifugation, and
`
`
`
`
`
`
`the captured ammoniumionsare then
`eluted from the resin bysalt solutions
`
`
`
`
`
`
`
`
`
`
`
`
`
`or released as ammonia bythe addi-
`
`tion ofdilute alkali,?! a technique de-
`
`
`
`
`
`
`
`
`
`scribed in detail more recently by
`
`Brusilow.
`
`
`
`
`Deproteinization
`
`
`
`
`
`
`Whole blood or plasma proteins are
`precipitated by the addition oftri-
`
`
`
`
`
`
`
`
`
`
`chloroacetic or perchloric acids, and the
`
`
`
`
`
`ammoniais determined directly in the
`supernatant fluid after alkalization Pad
`
`
`
`
`
`
`
`
`QUANTITATION OF
`
`AMMONIA
`
`
`
`
`
`
`
`
`
`
`After the release or capture of ammo-
`nia or ammonium ions, several methods
`
`
`
`
`
`
`have been described to determine the
`
`
`
`
`
`
`
`
`
`
`
`amount of ammonia present. The gen-
`
`
`
`
`
`eral categories for these methods in-
`clude titration, colorimetric/fluorimet-
`
`
`
`
`
`
`ric, electrode-based, and enzymatic.
`
`
`
`Titration Method
`The ammonia liberated from the
`
`
`
`
`
`
`
`
`
`
`
`sample is trapped in an aliquotofdi-
`
`
`
`$13
`
`
`
`Par Pharmaceutical, Inc. Ex. 1022
`Par v. Horizon, IPR of Patent Nos. 9,254,278, 9,095,559, and 9,326,966
`Page 3 of 10
`
`

`

`BARSOTTI
`
`
`THE JOURNAL OF PEDIATRICS
`
`
`
`
`
`JANUARY 2001
`
`
`
`
`Gas Sensing Electrode
`
`
`
`
`
`
`
`
`
`
`
`PHelectrode Reference Electrode
`
`
`
`
`
`Reference Buffer
`
`
`Gas Permeable Membrane
`
`
`(water impervious)
`
`
`
`
`
`Temperature Controlled
`
`
`
`
`
`Both
`
`(_)<— Sample
`
`
`
`
`(-) 4— Alkaline Butfer
`
`Sealed Mixing Chamber
`
`
`
`
`Fig 4. Schematic of ion-selective electrode(left pane!) and suggested arrangement for continuous-flow analysis of ammonia
`
`
`
`
`
`
`
`
`
`
`
`
`
`in plasma samples (right panel).
`
`
`
`
`
`
`
`lute acid, and the amount is measured
`
`
`
`
`
`
`
`by back-titration of the acid solution
`
`
`
`
`
`
`with a base while the pH is monitored
`
`
`
`
`
`
`
`
`with an indicator or electrode. The
`
`
`
`
`
`
`
`
`
`
`
`
`principal advantage ofthis approachis
`thatit is inexpensive, requiring no spe-
`
`
`
`
`
`
`
`
`
`cialized equipment. The disadvan-
`
`
`
`
`
`tages, however, are significant. They
`
`
`
`
`include insensitivity, the requirement
`
`
`
`
`
`for large blood samples, and contami-
`
`
`
`
`
`
`nation byother volatile bases that may
`
`
`
`
`
`
`alfect the final value. Overall, this pro-
`cedureis laborious and slow andthere-
`
`
`
`
`
`
`fore is not used routinely.
`
`
`
`
`
`
`
`
`
`
`Colorimetric/Fluorimetric
`Reactions
`
`This method is based on the reaction
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`of ammonia with a reagent to form a
`colored complex that is measured by
`
`
`
`
`
`
`spectrometry or fluorometry. Among
`
`
`
`
`the first reactions used was the in-
`
`
`
`
`
`
`
`
`
`dophenolreaction, described by Berth-
`the formation of a
`in 18597":
`elot
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`bluish color by the reaction of ammo-
`
`
`
`
`
`
`nia with phenol and hypochlorite. This
`method is commonlyreferred to as the
`
`
`
`
`
`
`
`
`
`
`
`phenol reaction. Another colorimetric
`
`
`
`
`
`
`reaction is the Nessler reaction,
`in
`
`
`
`
`
`
`which a brown-orange color is formed
`
`
`
`
`
`
`by the reaction of ammonia with mer-
`
`
`
`
`
`
`cury or potassiumiodide in an alkaline
`solution. Some othercolorimetric reac-
`
`
`
`
`
`
`
`
`
`tions include the use ofisocyanurate,
`
`
`
`
`cyanate, ninhydrin, or thymol hypo-
`bromite. Detection of ammonia and
`
`
`
`
`
`
`
`
`
`
`
`primary amines down to the nanogram
`
`
`
`
`
`range is routinely performed with fluo-
`
`
`
`rescence derivatization reagents such
`
`
`
`
`
`
`
`
`
`
`
`
`
`as fluorescamine and o-phthalalde-
`hyde.?4-?? The principle advantages to
`
`
`
`
`this approach are speed, simplicity,
`
`
`
`
`
`
`
`
`
`specificity when used carefully, and ex-
`
`
`
`
`
`cellent sensitivity. The disadvantage is
`that other substances in the bloodaf-
`
`
`
`
`
`
`
`
`
`
`
`fect some reactions, for example, the
`Berthelot reaction is inhibited by ex-
`
`
`
`
`
`
`
`
`
`cess amino acids, creatine, glutamine,
`
`and sometherapeutic agents.*”
`
`
`
`
`
`
`
`
`
`
`
`Gas-sensing Electrode
`
`
`
`
`
`
`
`With the introduction of the gas-
`sensitive electrode (eg, Orion, Model
`
`
`
`
`
`
`
`951201), a number ofreports have ap-
`
`
`
`
`
`
`
`
`peared describing the methods re-
`
`
`
`
`
`
`
`
`quired and the use ofthe electrede in
`
`
`
`
`
`measuring ammonia levels in samples
`
`
`
`
`
`of blood, urine, cerebrospinal fluid,
`
`
`
`
`
`
`
`and saliva over a broad range from 10
`
`
`
`
`
`Lmol/Lto nearly 1 mmol/L.
`
`
`
`
`
`When immersed in the sample or
`
`
`
`
`
`held closely above it,
`the dissolved
`
`
`
`
`
`
`gas ofinterest diffuses across a gas-
`
`
`
`
`permeable membrane into a small vol-
`
`
`
`
`
`
`ume olfbuffer. The reaction changes
`
`
`
`
`
`
`
`
`the pH ofthe buffer, which is sensed
`
`
`
`
`
`
`
`by an internal pH electrode or sensing
`
`
`
`
`
`
`
`electrode. The change in pH results in
`
`
`
`
`
`
`
`a change in the potential between the
`
`
`
`
`
`sensing electrode and a reference elec-
`
`
`
`
`
`trode immersed in a separate reference
`buffer, all housed within the same elec-
`
`
`
`
`
`
`
`
`
`
`
`trode body. The arrangementis illus-
`
`
`
`
`
`
`
`
`
`trated in Fig 4,
`in which the scale of
`
`
`
`
`
`
`some of the components is exaggerated
`
`
`for clarity.
`The main advantages ofelectrode-
`
`
`
`
`
`
`
`
`
`based ammonia assay are its cheap-
`
`
`
`
`
`
`
`
`ness, ease of use, and because it never
`
`
`
`
`
`
`
`
`
`
`
`
`
`comes in contact with the sample, its
`imperviousness to sample color, tur-
`
`
`
`
`
`
`
`
`
`bidity, viscosity, or the presence of
`
`
`
`
`
`
`drugs or other metabolites in the sam-
`
`
`
`
`
`
`ple. The electrode is best arranged
`
`
`
`
`
`
`
`
`above the surface of the sample in a
`
`
`
`
`
`sealed environment
`(Fig 4). This
`avoids the accumulation of proteins
`
`
`
`
`
`
`
`
`
`
`
`and cell fragments on the membrane
`
`
`
`
`
`surface that would normally occur
`
`
`
`
`
`
`during immersion into plasma or blood
`
`
`
`
`
`
`sample and minimizes the loss of am-
`monia from the sample away from the
`
`
`
`
`
`
`
`measure-
`the
`electrode
`during
`
`
`
`men
`52 The disadvantages of the
`{2251
`
`
`
`
`
`
`
`
`electrode-based system include the re-
`
`
`
`
`
`
`quirementoflarge sample volumes and
`
`
`
`
`
`
`slow sample reads, especially at low
`
`ammonia levels, requiring 10 to 15
`
`
`
`
`
`minutes. In addition, major differences
`
`
`
`
`
`
`
`
`
`
`
`in either the osmolarity or temperature
`ofthe sample and the sensing electrode
`
`
`
`
`
`
`
`buffer must be avoided.
`
`
`
`
`
`
`
`Enzymatic Method
`
`
`The specificity of most methods for
`
`
`
`
`
`
`
`
`
`
`ammonia determination in biologic flu-
`ids relies on the physical separation of
`
`
`
`
`
`
`
`
`
`
`
`ammoniafrom interfering substances
`
`
`
`
`
`byvolatilization after alkalization. In
`contrast,
`the most common method
`
`
`
`
`
`usedin clinical laboratories is an enzy-
`
`
`
`
`
`
`matic method that measures ammonia
`
`
`
`
`
`
`
`
`
`
`directly. Thus sample preparation is
`
`
`
`
`
`relatively simple, because the previous
`
`
`
`
`
`
`liberation of ammonia from the sample
`
`
`
`
`
`
`
`
`is not required. This assay is based on
`
`
`
`
`the reductive amination of 2-oxoglu-
`
`Sl4
`
`
`
`Par Pharmaceutical, Inc. Ex. 1022
`Par v. Horizon, IPR of Patent Nos. 9,254,278, 9,095,559, and 9,326,966
`Page 4 of 10
`
`

`

`THE JOURNAL OF PEDIATRICS
`Votume 138, Numer |
`
`BARSOTTI
`
`injection quadru
`detection quadruple
`mass filter
`mass spectrometer
`Roots pump
`air or breath sample
`
`He carrier gas
`
`
`
`
`
`=> H30* or 0+
`
`
`ion source
`ion injection
`
`orifice
`orifice
`gas
`
`(e.g. AifH20)
`
`injection diffusion pump
`
`——
`10cm
`
`detection diffusion pump
`
`Fig 5. Simplified schematic of selected-ion flow tube (SIFT).
`
`tarate with glutamate dehydrogenase
`and reduced nicotinamideadenine di-
`nucleotide phosphate:
`
`2-Oxoglutarate + NH, + NADPH
`NGLDH
`Glutamate + NADP
`
`The decrease in absorbance at 340
`nm caused by the oxidation of
`NADPH is proportional to plasma am-
`monia. GLDH is specific for ammonia
`and does not react with methylated
`amines. Early studies describing the
`enzymatic determination of ammonia
`used NADH as a coenzyme. Because
`other NADH-consuming systems are
`present in blood, manyofthese reports
`overestimate the level of free ammonia,
`for example, Muting.** The effect of
`these systems can be minimized, usual-
`ly by a 30-minute preincubation peri-
`od. There are considerably fewer
`NADPH-consuming sources in plas-
`ma, so the preincubation time can be
`reduced to a few minutes.*
`The assaycan be used to measure am-
`monia levels over a broad range, from
`as low as 12 [mol/L to as high as 1
`mmol/L. The disadvantage ofthis ap-
`proachis the length and complexity of
`the procedure, thereby enhancing the
`potential for variation in reported blood
`ammonialevels. If not handled proper-
`ly, ammonia concentrations rise in
`standing blood or plasma. Indeed, be-
`cause standard procedures for pre-
`analysis sample processing do not in-
`corporate any attempts to inhibit the
`
`continued liberation of ammonia from
`the samples, except for lowering the
`temperature, ammonia levels in the
`sample will continue to increase during
`the assay and up to the time ofthe ini-
`tial readings. Thus overestimates of am-
`monia levels are most likely in poorly
`handled samples containing elevated
`levels oftransaminases and aminoacids.
`
`NormalValuesfor Blood
`Ammonia Levels
`Table | lists selected blood ammonia
`levels for various blood sample types
`and assay methods from a numberof
`studies. The average values for arterial
`blood, plasma, venous blood, and plas-
`ma are 18, 23, 28, and 32 |tmol/L, re-
`spectively. The average value for ve-
`nous blood and plasmais 30 [tmol/L.
`
`FUTURE
`DEVELOPMENTS
`
`The major limitations of convention-
`in vitro blood ammonia measure-
`al
`ments are the complexity involved in
`the proper drawing and handling of
`the sample, the time allowed between
`drawing and assaying, andfinally, the
`assayitself. A consequence ofthese
`limitations is that blood ammonia mea-
`surements are performed only a few
`times each day. Alternative methods
`are still sought that are noninvasive or
`require a small catheter in a peripheral
`vein but provide a continuous monitor
`of blood ammonia levels. Such meth-
`
`ods could alert medical staff to an im-
`pending hyperammonemic condition
`and would more easily permit earlier
`selection and regulation of therapeutic
`interventions. Two promising methods
`that are under development and may
`eventually be used clinically are the se-
`lected-ion flowtube technique, which
`analyzes trace gases in breath, and a
`fiber-optic catheter tipped with an am-
`monia-sensitive indicator.
`
`Selected-ion Flow Tube
`Selected-ion Flow Tube is a quanti-
`tative method for the rapid, real-time
`analysis ofthe trace gas content ofat-
`mospheric air. It was originally devel-
`opedto studyionic reactions in the gas
`phase and is particularly valuable
`for providing kinetics data on ion-
`molecule reactions, contributing to the
`current understanding of the chem-
`istry of some low-temperature gaseous
`plasmas, especially interstellar clouds.
`The same technology is currently
`being developed to analyze trace gases
`in breath. Previous methods for mea-
`surement of ammonia in breath have
`required large sampling flow rates”? or
`long sampling times® and are there-
`fore unsuitable for assessing the am-
`monia concentration from a single
`breath. A schematic of the SIFT appa-
`ratus is shown in Fig 5 taken from
`Smith and Spanel.*® This technologyis
`being further developed and may soon
`be a sensitive, quantitative method for
`the continuousreal-time analysis ofthe
`trace-gas content of human breath and
`
`SI5
`
`Par Pharmaceutical, Inc. Ex. 1022
`Par v. Horizon, IPR of Patent Nos. 9,254,278, 9,095,559, and 9,326,966
`Page 5 of 10
`
`

`

`BARSOTTI
`
`
`
`THE JOURNAL OF PEDIATRICS
`
`
`
`
`
`JANUARY 2001
`
`
`
`
`
`
`
`
`
`
`
`
`
`Table I, Reported assay techniques and blood ammonia levels in healthy subjects
`
`
`
`Reference
`
`Reference
`
`RSETEN nnennnnnennntnmnnnnenmneUMEeenONOeneanmnmnneevalue(Amol/L)
`
`
`
`
`
`
`Arterial blood
`
`
`
`
`
`
`Hutchinson and Labby(1962)
`
`
`
`Gips and Wibbens-Alberts (1968)
`
`
`
`
`Huizenga and Gips (1983)
`
`
`
`
`Huizengaet al (1992)
`
`
`
`
`
`Huizengaet al (1992)
`
`
`Arterial plasma
`
`
`
`Gips and Wibbens-Alberts (1968)
`
`
`
`
`Huizenga and Gips (1983)
`Venous blood
`
`
`Dienst (1961)
`
`
`
`
`Forman (1964)
`
`
`
`Hutchinson and Labby (1962)
`
`
`
`
`Proelss and Wright (1973)
`
`
`
`Gips and Wibbens-Alberts (1968)
`
`
`McCullough (1967)
`
`
`
`Gangolli and Nicholson (1966)
`
`
`
`
`Sinniahet al (1970)
`
`
`
`Huizengaet al (1992)
`
`
`
`Huizengaet al (1992)
`
`
`Venous plasma
`
`
`
`
`
`Gerronet al (1976)
`
`
`
`Oberholzer et al (1976)
`
`
`
`
`Butteryet al (1982)
`
`
`Brusilow (1991)
`
`
`
`
`Cooke and Jensen (1983)
`
`
`
`
`Willems and Steenssens (1988)
`
`
`
`
`Spooneret al (1975)
`
`
`
`Seligson and Hirahara (1957)
`
`
`
`
`Mondzacet al (1965)
`
`
`
`
`Muting etal (1968)
`
`
`
`
`van Anken and Schiphorst (1974)
`
`
`
`
`Howanitz et al (1984)
`
`
`
`da Fonseca-Wollheim (1990)
`
`
`21
`
`59
`
`42
`
`23
`
`41
`43
`40
`45
`16
`
`
`
`
`
`
`
`16
`
`
`
`
`lon-exchange/Nessler
`
`
`lon-exchange/phenol
`
`
`lon-exchange/Nessler
`
`Supernatant/electrode
`
`Supernatant/phenol
`
`Supernatant/phenol
`
`Supernatant/phenol
`
`Supernatant/phenol
`Microdiffusion/BAC I]
`
`
`
`
`Enzymatic
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`42
`
`4]
`15
`
`16
`
`
`
`16
`
`
`lon-exchange/Nessler
`
`Supernatant/phenol
`Microdiftusion/BAC I
`
`Microdiftusion/BAC II
`
`
`
`Enzymatic
`
`
`41
`
`15
`
`Supernatant/phenol
`
`Enzymatic
`
`
`
`
`9
`
`44
`38
`
`
`4
`52
`
`46
`
`
`27
`20
`
`
`2
`
`33
`
`S
`
`11
`
`8
`
`
`
`lon-exchange/phenol
`
`lon-exchange/phenol
`
`
`lon-exchange/phenol
`
`
`lon-exchange/phenol
`Electrode
`
`Electrode
`
`Supernatant/fluorometry
`Microdittusion/Nessler
`
`
`Enzymatic
`
`Enzymatic
`
`Enzymatic
`
`Enzymatic
`
`Enzymatic
`
`
`
`
`
`
`19
`
`22
`8
`
`2]
`
`
`
`21
`
`23
`mas
`
`
`
`
`
`15
`
`31
`
`21
`
`17
`
`22
`
`57
`
`30
`
`44
`21
`
`
`
`21
`
`
`19
`
`16
`
`21
`
`18
`44
`
`44
`
`
`52
`56
`
`
`30
`
`57
`
`22
`
`30
`
`29
`
`
`
`
`
`
`thus permit a continuous, noninvasive
`
`
`
`
`
`measure of systemic ammonialevels.
`This technique involves the genera-
`
`
`
`
`
`
`
`
`
`
`
`
`tion ofpositive ions that are created in
`
`
`
`
`
`a microwave discharge ion source,
`
`
`
`
`
`containing an appropriate gas mixture.
`
`
`
`
`In conventional mass spectrometry,
`
`
`
`
`
`
`
`ionization ofthe trace gas moleculesis
`
`
`
`
`achieved by electron bombardment,
`
`
`
`
`
`resulting in molecular “cracking” and
`
`
`
`
`
`the production of complicated spectra.
`
`
`
`
`
`
`
`The SIFT technique uses a current of
`
`
`
`
`
`precursor ions of a given masgs-to-
`
`charge ratio. In the case of ammonia
`
`
`
`
`
`
`
`
`
`
`
`
`detection, H,O* is extracted from this
`
`
`
`
`
`
`mixture of ions with a quadrupole
`mass filter. This current of selected
`
`
`
`
`
`
`
`
`
`
`
`
`
`ionsis then injected into a fast-flowing
`
`
`
`
`
`inert carrier gas stream, usually heli-
`
`
`
`
`
`
`
`um. The ions are carried along a 1-
`
`
`
`
`
`
`meter length flow tube and are sam-
`pled by a pinhole downstream ofthe
`
`
`
`
`
`
`
`
`
`
`
`injector. The sample ofbreathis intro-
`
`
`
`
`
`
`
`ducedinto the device by a sample port
`
`
`
`
`
`
`near the injector (Fig 5). Accurate
`trace gas analysis or quantificationis
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`possible because the reaction of the
`
`
`
`
`
`
`
`primaryions, in this case HO", with
`ammonia is to form ammoniumions
`
`
`
`
`
`
`
`
`
`
`
`
`
`and water is precisely defined in the
`
`
`
`
`
`
`
`SIFT. In general, the primary ions cho-
`
`
`
`
`
`
`
`sen must not react at significant rates
`
`
`
`
`
`
`with the major components of the
`
`
`
`
`breath sample, oxygen,
`nitrogen,
`water, or carbon dioxide, because such
`
`
`
`
`
`
`
`
`
`
`
`reactions would saturate the primary
`
`
`
`
`
`
`ions. In turn, the primary ions must
`
`
`
`
`
`
`react efficiently with the trace gases to
`
`
`
`
`
`be detected to form identifiable prod-
`
`
`
`
`$16
`
`
`Par Pharmaceutical, Inc. Ex. 1022
`Par v. Horizon, IPR of Patent Nos. 9,254,278, 9,095,559, and 9,326,966
`Page 6 of 10
`
`

`

`BARSOTTI
`
`
`
`peeeames
`
`C3HygN
`(200)
`
` —
`
`trimethylamine
`(209)
`
`own 8 ee “pe %
`oakJs<"
`Molecular Weight
`
`{“co
`
`104
`
`oO
`oO
`A 108
`io)~
`
`c3O
`
`o 1
`O
`
`401
`
`THE JOURNAL OF PEDIATRICS
`Votume 138, Numer |
`
`uct ions. Thus the primaryions and the
`specificity of their interactions with the
`target trace gases in the breath sample
`is a determinant ofthe selectivityofthe
`method. The identification and quanti-
`tation of the products formed are then
`detected by a mass spectrometer.
`Currently, the technique can measure
`trace gases downtoa partial pressure
`of approximately 10 ppb. The mean
`value of 960 ppb for breath ammonia in
`normaladults was reported in Davies
`et al®” and thus is well above the mini-
`mum sensitivity ol the apparatus. Fig 6
`shows a spectrum obtained from a
`breath sample taken from a uremic pa-
`tient (end-stage renal failure) before
`hemodialysis. The sample was taken
`and stored in a Teflar bag and then
`transferred to the laboratoryfor analy-
`sis by SIFT. The molecular weight (x-
`axis) for each molecule identilied is
`shown at the top of each bar. Open
`bars represent precursor ions, H;O+
`(molecular weight =
`19), and their
`water clusters with molecular weights
`of 37, 55, and 73. The solid bars show
`the counts for each trace gas, ammoni-
`um ions (molecular weight =
`18), and
`their water clusters, with molecular
`weights of 3

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket