throbber
Conference Proceedings
`
`Design Principles of Liquid Nebulization Devices Currently in Use
`
`Joseph L Ran PhD RRT FAARC
`
`Introduction
`.Iet Nebulizers
`
`Principle of Operation
`’ariahles ol’ Nebnlizer Performance
`
`Advantages anti Disadvantages of .let Nebuliaers
`Design Variations in 'l‘raditioaal .lel Nelnilizers
`Ultrasonic Nebuliaers
`
`Description and Basic Function
`Physics of Ultrasonic Nelnllization
`Theory of Aerosol Production
`Comparison of Ultrasonic and Jet Nebulizers
`Advantages and Disadvantages of Ultrasonic Nebulizers
`Snunnary
`
`Liquid nebuliaation is a comtnon method of medical aerosol generation. 1. 'ebulizers are ot’2 types: jet (or
`pneumatic) small-volume nehulizer, aml ultrasonic ncbulizer. .lct nebulizers are based on the venturi
`principle. whereas ultrasonic nebnliaers use the converse piezoelectric eITect to convert alternating
`current to high-frequency acoustic energy. Important variables for both types of nebulizer are treatment
`time required, particle size produced, and aerosol drug output. There are several advantages to jet
`nebulimtion. including that effective use requires only simple, tidal breathin ’, and that dose modifica-
`tion :qu dose compounding are possible. Disadvantages include the length of treatment titne and
`equipment size. Design modifications to the constant-output nebulizer have resulted in breath-enhanced,
`open-vent nehulizers such as the l’ari LC Plus and the dosimetrie AeroEclipse. Ultrasonic nebnlizers
`generally have a higher output rate than jet ucliulizers, but a larger average particle size. Ultrasonic
`nebnlizers can also substantially increase reservoir solution temperature. the opposite of‘jet nebulizer
`cooling. Drug concentration in the reservoir does not increase with ultrasonic uelniliration, as it does
`with jet nehulization. Ultrasonic nebulizers have the same advantages as jet nelnllizers. Ultrasonic
`nebulizers are more expensive and fragile than jet nellolizers, may cause drug degradation, anti do not
`nelnllize suspensions \vell. Neither type of nebulirer meets tlle criteria for an ideal inhaler: efficient and
`quick dose delivery with reproducibility, cost-el'l'cctivmless, and no ambient contamination by lost aero-
`sol. Krjr n'oi‘rls: m'hnli:orinn. jet ricbrilizer. aerosol. ultrasonic.
`IRespir Care 2002:1170 1):]257—l27fi]
`
`Introduction
`
`The term "nebulizer" derives from the Latin “nebula."
`meaning ”mist." and reportedly was first used in lii'l'l
`followed by an 1874 definition as “an instrument for con-
`verting a liquid into a line spray. especially l'or medical
`
`pul'poses."' The appealing logic ol‘ creating :1 vapor or
`aerosol for the inhalation treatment of lung disease is at
`least as old as written records of medicine. The Aynrvedic
`tradition of medicine in lndia. which dates back perhaps
`
`fitlthlit-.srlna‘t(JttYCakiiiuurnalConlercnce.LiquidNehuliz:ttion:F_merg-
`in; 'i'cchaologies. held June 25—30. Elltll. in Montreal. Quéhce. Canada.
`
`Joseph I. Ran i’hD ItR'I‘ FARRC isat'l'ilialcdwilhCardiopulnionary Care
`Sciences. Georgia State University. Atlanta. Georgia.
`Joseph l. Rim l'hl) RR'I‘ FAARE‘. presented :1 terxion ol‘ this report at the
`
`Cni'l'eswlitleltcc: Joseph 1, Run l’iil) RRT FAARC. Cardiopulntonar}
`Care Sciences. MSCSRUJI‘). Georgia Slate University. 33 Gihncr St SE.
`Unit ti. Atlanta GA 30303—3035. E-niaii: jr:m@gsu.cdu.
`
`RESPIRATORY CARE . Novesnnzn 2002 Vol. LitriNtr‘t'tlmmflF'fld
`attha NLM ard maybe
`EL eject US Copyright Laws
`
`[957
`
`WATSON LABORATORIES v. UNITED THERAPEUTICS, |PR2017-01621
`
`UNITED THERAPEUTICS, EX. 2044
`
`Page 1 of 24
`
`

`

`DESIGN PRINCIPLES or: LIQUID NRIIULIZA‘HUN DEVICES CURRiiN'I'l.Y 1N Usn
`
`
`
`6 " length tubing
`
`To patient —)
`
`4.000 years or more, used inhaled substances for manag—
`ing asthma.2 Although inhalation devices were described
`in the l91h century, the modern precursors of small-vol—
`ume nebulizer devices appeared with the glass and hand-
`bulb “atomizers” introduced for asthma treatment in the
`
`1930s. such as the DeVilbiss No. 40 glass nebulirer.I The
`Collison nebulizer became available in the late 1940s; it
`used a baffle to filter out large particles, thus distinguish—
`ing a "nebulizer" from an “atomizer."3 The Wright nebu-
`lirer, which appeared in the 19505;. was engineered l'rom
`cbonile and perspex; it was much more compact than the
`Collison and more closely resembled today's pneumati—
`cally powered nebulizers.4 A different method of creating
`liquid aerosols. the ultrasonic nebulizer, was introduced in
`the [9605.
`It relies on high-frequency sound waves to
`aerosolize the solution. Today the term ”nebulizcr." as
`used clinically in respiratory care. encompasses both gas—
`powercd jet nebulizcrs and ultrasonic nebulizers. This re-
`view considers the physical principles. designs, advan-
`tages. disadvantages. and factors affecting performance of
`jet and ultrasonic nebulizers that are in current clinical use.
`
`Jet Nchnlizers
`
`Principle of Operation
`
`Small—volume jet (pneumatic) nebulizcrs arc 2-lluid at-
`oinizers. The basic principle of operation involves com-
`pressed gas directed through a narrow orifice, with em
`trainmcnt of liquid through one or more capillary feeder
`tubes. The narrow—orifice gas nozzle has traditionally been
`termed a “vertiuri.” with liquid entrainment based on the
`Bernoulli principle. In laminar flow. and keeping height
`constant, Bernoulli’s equation‘ for an incompressible l'luid
`states:
`
`'/2 my1 + P = constant
`
`in which to —' mass. v = velocity, P = pressure. and
`1K2 niv2 represents kinetic energy. Fora gas in laminar flow
`through a tubing system, an increase in velocity. such as
`can occur at a restriction in the tithing. leads to a decrease
`in the lateral pressure. If kinetic energy increases. pressure
`energy must decrease, based on conservation of mass and
`energy. A “venturi‘l is based on Bernoulli's equation and
`uses the pressure drop to entrain a second fluid. The lower
`pressure causes drug solution to be drawn up to the gas jet
`through capillary or feeder tubes that
`lead down to the
`liquid reservoir. The entrainment is also attributed to ma-
`mentum transfer between the power-gas molecules and air
`or liquid in the feeder tubes!’ The high-velocity gas meet-
`ing the liquid causes primary generation of large droplets.
`typically 40—60 am. but with a possible range of 15 to >
`
`Secondary dmpleb
`11-10 pm)
`
`F311;?“firOleF
`
`Prll‘Z ow: = K
`
`tube
`rug solution
`
`
`baffle
`
`aplltary feeder
`
`Powor gas
`[flow motor, comprossori
`
`Fig. 1. The major elements of a jet {pneumatically powered) small-
`volume nebulizer. m = mass. v -- velocity. P = pressure. K =
`constant.
`
`500 am.” Droplets produced by primary generation are
`impacted onto bal'lles within the tichuliaer. producing a
`secondary generation of aerosol with particles of 1—H)
`pm, with typical sires of 4.0 and 5.0 uni.7 Du ring that second
`stage of aerosol generation the larger particles are baffled.
`coalesce. and drain back into the reservoir (Figure l).
`Aerosol droplet size depends on the geometry of the
`nozzle and bal‘t‘lcs. nozzle diameter. gas velocity, mass
`flow of gas, and the physical constants of the power-gas
`and drug solution. Jet nebulixcrs produce a range of par-
`ticle sizes. which is a distribution termed "polydisporse" or
`“heterodispersc.” in contrast with “monodispcrsc” acro-
`sols." The mass median aerodynamic diameter (MM/UH
`is the median droplet size of the cloud of aerosol particles.
`Particles of 1—5 pm diameter have a higher probability of
`reaching the lower respiratory tract than do larger parti—
`cles." However. there is a trade-off between particle size
`and drug amount. With spherical particles.
`the mass of
`drug varies directly as the third power of the particle ra—
`dius. An increase in particle sire gives an exponential
`increase in drug mass. but the converse is also trite. With
`aqueous solutions the extremely large surface area created
`by aerosol particles and a dry power—gas cause consider-
`able ovaporation and substantial cooling of the nebuliacr.
`because ol' latent heat of vaporization.
`to approximately
`If)” C below ambient temperature. within minutes. '” Higher
`power-gas llows can increase the amount of temperature
`drop in some nebulizcrs." The amount of temperature
`drop differs among nebulizer brands," and the amount of
`temperature drop is less with a partially liumidilied power-
`
`l253
`
`arthe Mud and .
`filjlinroav CARE 0 NOVEMBER 2002 VOL 47 No l
`mi‘mmria'“
`Subieet US, Copyright Laws
`
`l
`
`WATSON LABORATORIES v. UNITED THERAPEUTICS, |PR2017—01621
`
`UNITED THERAPEUTICS, EX. 2044
`
`Page 2 of 24
`
`

`

`Disstos PRINCIPLES or: LIouto NEBUUZA'I‘IUN Dimer-1: Coronas-rm lN Use
`
`0.5
`
`
`output ralo
`.-—-G—
`tempernlu r:
`cone. nlrallon—CI-
`
`.-.
`e
`_
`9:. 20
`E
`'57 93 «g-
`.-
`:I.
`e
`g
`'5
`-—
`a.
`g
`'5
`0 10
`o
`E
`g
`0:
`
`0.4
`
`0.2
`
`0.1
`
`D
`
`dead volume
`point
`
`
`
`
`
`t]
`
`5
`
`10
`
`.
`15
`
`1.5
`
`3
`E
`r
`g
`1.0 E
`c
`3
`8
`u
`.e
`o
`E
`hI:
`
`0.5
`
`0.0
`
`10
`
`
`
`
`
`NebulizatianTimetrninj
`
`
`
`0 ."amt. 4mL 5mL
`:15 Ll'rnin
`
`
`
`nt. mL 5 mL
`3 mL.
`al‘tfl Umin
`
`amt 4mL. Smt.
`at B Limin
`
`Fig. 3. Nebutization time is a function of both the power-gas flow
`rate and the fill volume. Lower flow rates and higher fill volumes
`increase the nebulization time, 'I he graph shows pooled data tron't
`1? nebulizers. {From Reference 1?. with permission]
`
`’ariahles of Nehulizer Performance
`
`Several variables describe the uebulizer performance.
`including the time required for Itebttlizalion. droplet size
`produced. and drug output.
`
`Nebuliaaliou Time. The time required is directly propor-
`tional to the volume oi solution placed in the reservoir and
`inversely proportional to the power—gas ttow (Fig. 3).”-17
`l)etcrmination ot' nebuliration titne in studies differs
`be ‘ause it
`is dil'l'icttlt to decide when aerosolization has
`ceased. The ucbttlization time depends on how the end
`point ol‘ nehulizalion is defined. Kradjan and Lakshmi—
`uarayan defined 3 possible end points for Itebulizalion:
`sputtering time. total time. and clinical time.” Sputtering
`is the point when aerosolization becomes en'atic. as noted
`by seeing and hearing. Total titne is when production ol‘
`aerosol ceases. Clinical
`time is between sputtering and
`total time. and approximates the point when a patient or
`therapist typically stops a treattnettt. With a 3 mL voltnne.
`averaged over 5 nebuliecr brands. the sputtering. clinical.
`and total times were measured as 13. 9.9. and |2.8 tuitt.
`
`respectively. by Kradjau and Laksluninarayau.‘3 Nebuli—
`zation time may also be defined as the point 30 seconds
`utter sputter. with or without tapping. although this can
`introduce greater subjectivity in the measurement. Regard—
`less ot' the end point chosen.
`it
`is important that studies
`define the end point clearly. in ordcrto allow comparisons.
`There is evidence to suggest that Itcbulizatiott titnc var-
`ies directly with the viscosity of solutions. Antibiotic so—
`lutions cart be more viscous than aqueous bronehodilator
`solutioits and can require more [Kiwcrful compressors tie,
`higher power-gas flows) to keep Itehulization times from
`being prolonged. Hess et al found that the average nebu-
`li‘ealion time pooled tor
`I? Itcbulizcrs, using albttterol,
`
`Nebulization time (mm
`
`Fig. 2. Functioning of a small-volume jot nebulizer with respect to
`output rate, solution temperature. and concentration of solute in
`the solution during nebulization. The data are lwpotheticat and not
`based on actual measurements of a particular brand of nebulizer.
`
`gas. such as l'rom a compressor. tltan with a drier gas. such
`as from a cylinder or wall outlet.l2 ’l‘he evaporation also
`increases llte solution's drug concentration during tlte tittte
`ol‘operation.”W-H lJemtis ct al l‘ottntl an increase from 1%
`to 38'»?- using 5 ml. of a solution at sodittttt fluoride in a
`Wright nehuliacr—a 200% increase in conecntration.'-‘
`Concentration increases ranging from 5% to over 3754-
`have been l‘ound with other jet nebuliaers. using albuterol
`solttlion.“' Figure 2 shovvs a generic illustration of otltptlt
`rate. temperature. and solution concentration for a typi ‘ztl
`small-volume jet nebulixer. In the first minutes the output
`rate falls as the soltttiott temperature drops and viscosity
`and surl'ace tension increase.” A plateau is tltctt reached in
`hotlt output and temperature. until a point at which output
`decreases precipitalcly; this is usually the point at whiclt
`the level ol‘ liquid in tlte reservoir falls below the bottom
`of the capillary t‘eeder tube. At that point. gas is drawn up
`with the liquid. the nehuliitcr begins to sputter. and aerosol
`creation is intermittent.
`'I'appiug the nebulizer can cause
`bal‘l'led tlroplcts to coalesce. run down the walls of the
`reservoir. increase the solution level. and temporarily re—
`store aerosoliaation. Ultimately nebulimtion ceases.
`al—
`though the nehulizer contains some residual solution on
`the apparatus walls. battles. and at the bottom of tire res-
`ervoir chamber. The amount ol‘ that remaining liquid is
`termed the "dead voluttte.“ which in most small—volume
`Itebttliaers is approximately l ml.."' The l‘act
`that most
`nebulirers canttot aerosolize well below 1 mL is the reason
`that a bland diluent ot‘ saline or water is added to a 0.5 ml.
`
`drug dose; for example. albtilerol from a multi-use bottle
`(110.5% solution. Addition of the diluent is not to weaken
`the drug solution. btll is in tact needed it atty aerosol is to
`be produced. and iuet'eases the amount of drug entitled by
`llte nebulizer.”
`
`
`
`
`
`5.__A.___§.._A_...............-.__.-_‘.____4-e_-_.__-p.._4-.......—.....-,--_—._.___._....—..
`
`Resrnm‘rotw Carat - NOVEMBER 2002 Vol. clit‘i’d'etalililmsmiid
`atthENLll-lardrnaybe
`Subject 05 Copyright Laws
`
`I.
`
`1259
`
`WATSON LABORATORIES v. UNITED THERAPEUTICS, IPR2017—01621
`
`UNITED THERAPEUTICS, EX. 2044
`
`Page 3 of 24
`
`

`

`DESIGN PRINCIPLES or: Ltotnn Neeuuzarton DIiVICI-IS CURRIiN'I'l.Y {N U51;
`
`'9
`
`
`
`
`
`DeadVolumetmgalbuterelj
`
`.L.L 3:30|
`
`_|
`
`-|-is:r
`
`LiI
`
`"Th—l
`
`
`
`
`
`0'1sz 4111!. 5mL
`at 6 Umin
`
`:imL amt 5mL
`at I! Urnin
`
`:FnL 4mL Smt.
`at IIIILl‘mIn
`
`Fig. 5. Effect of power‘gas [low rate [6. 8. or it] Umin] and till
`volume [3. 4. or 5 mL} on amount of drug remaining in the dead
`volume. The graph shows pooled data from 17 nebulizers. [From
`Reference 1?. with permissmn.)
`
`ccntage of 1—5 pm particles ranged from slightly over
`30% to 60%. Clay et al
`found an inverse relationship
`between the M MA!) and the geometric standard deviation.
`indicating that the MMAD was stttallerhul the aerosol was
`more hctct‘odispersc at higher llow rates.2|
`
`The primary factors affecting
`Aerosol [)rng Output.
`the amount of aerosolircd drug released front a jet nehu—
`liver are the till volume and. secondarily. the type of nett—
`tlli‘rcr. Increasing the till volume results in lower concen—
`trations of drug remaining in the dead volume of the
`ttchulizcr. when nchttlization ceases. This is hosed on the
`following relationship:
`
`[)eatl volume drug (mg) 2 dead volume toil.)
`
`K concentration {ntglntL}
`
`Willi a theoretical dead volume of Lt) ml. and assuming
`no increase in concentration ofthc residual solution due to
`
`evaponttion. a 2 ml. fill volume would leave at least half
`of the drug in the dead volttnte. A 4 ml. fill volume would
`only leave a quarter of the drttg in the dead volume. In
`reality. evaporation increases the concentration of the so-
`lution remaining in the nchttlizcr.‘5 Increasing the fill vol-
`tIIttc therefore reduces or "dilutes" the concentration of
`
`drug it] the dead volume. httt also increases nehttliaation
`little (Fig 5).” In the case of a 2.5 mg tlose of Itehulized
`albtttet'ol.
`l'ill volumes of 3. 4. and 5 ml. decrease the
`amount of drug left
`it]
`the dead volume from approxi—
`mately [.4 mg to approximately l.lS mg and LI mg.
`respectively.” Determination of aerosol output. dead vol—
`ume. and remaining drug amount depends on the defini—
`tiott of “end of itehulization."
`
`The type of nehuIizcr can also affect tltc amount of drug
`available for inhalation. Devatlasott ct al compared 4 neb-
`
`u
`
`
`
`MMAD[11ml M
`
`A I
`
`:
`
`e
`
`_a
`nebultzer flow
`
`10
`
`J
`
`U3
`
`a3
`
`
`
`(pm!0-‘N(-93|-0|MMAD
`
`ABCDEFGHIJKLMNOPD
`
`nebulizer brand
`
`Fig. 4. A. Effect of power-gas new rate on mass median aerody-
`namic diameter tMMAD), with smaller particle sizes at higher pow-
`er—gas flow rates. The graph shows pooled data from 1? nebuliz-
`ers. B. MMAD for 1'." nebulizer brands. pooled for all power—gas
`flow rate settings. (From Reference 1?, with permission.)
`
`with 4 ml, voltttne and 6 Limit: power—gas flow, was ap—
`proximately 13 min." Newman et al
`investigated neltu-
`Iized gentamicin solution with 4 different nchnliacrs and
`found tltat 4 mL (I60 mg) at 6 Until: power-gas flow
`required between I? and 25 min.” The approximate av-
`erage time pooled for the 4 nebulizcrs was 20 min. That
`study recommended flows of Ill—l2 limit] or a high-flow
`compressor for nebulization of antihiotic solutions.
`
`Particle size is a major factor influencing
`Particle Size.
`the probability ofdrug particles reaching the various levels
`of airway generation?“ Particle size produced by a small-
`volume jet nehuliaer is inversely proportional to power-
`gas flow (or compressor pressure).
`l-Iess et al found that
`the MMAD varied inversely with power—gas flow rates
`between o and Ill IJmin (Fig. 4A).” In that study the
`percentage of l—S pm particles increased with higher pow-
`er-gas flow rates. Similar data were reported from the
`United Kingdom by Clay et al for 4 disposable jet nehn-
`liaers.21 Increasing the power-gas flow rate produced
`smaller particles and reduced nebulization time. Particle
`size is also affected by the physical characteristics of the
`drug solution, such as surface tension and viscosity”:2
`lless et al found that the MMAD produced by I? ttchtt-
`Iizers differed significantly (see Fig. 48).” and the per-
`
`1260
`
`Thismaterialwasfiestttttarottv Cato; 0 NOVEMBER 2002 V01, 47 No I
`atthe NLM 3rd may be
`Subject: US Copyright Laws
`
`l
`
`WATSON LABORATORIES v. UNITED TH ERAPEUTICS, |PR2017—0162‘l
`
`UNITED THERAPEUTICS, EX. 2044
`
`Page 4 of 24
`
`

`

`Drasrotsl PRINCIPLES or LIQUID Nerturam'rron Dravrcr-s CURREN'I'LY IN USE
`
`trliver systems (Acorn. Acorn plus Mizer. Ventstrcam. and
`am LC) and fotrnd significant differences in the delivery
`of albtttet'ol to inspirator‘y tillers. ranging from 9.35% to
`l‘)% of tltc nominal dose?“ Barry et al compared outputs
`of a oehulired corticosteroid thudesonidc) from the Vent-
`stream. the lari LC Plus. and tlte l’ari LC Star. rising tlte
`compressors specified for each. The nebuliacrs differed in
`hell] the rate of otttpul (13—145 ugt'min] and the total
`output of drug collected on irtspiralory litters {32.3—‘)l.‘)
`lug). using simulated conditions of breathing?"
`The physical properties of the power-gas and the solit-
`tion to he nebulixed also affect uebulirer output. Substi-
`tuting helios for air decreased particle size and inhaled
`drug mass significantly. using albuterol.” Nebulired al-
`buterol a 'ailability was greater from a solution containing,
`the preservative benmlkoniunt chloride. becatrse of the
`lower surface tension produced by the preservative (35.0 i
`[1.5 millinewtons per meter with preservative vs 'lt).5 : 0.5
`nrillinewtotts per rrteter without)? The foaming seen with
`the lower surface tension from the preservative resulted in
`tnore return of liquid to the reservoir and a snraller dead
`volume than with the preservative-free solution. with which
`large droplets adhered to the nebulizer walls and increasctl
`the dead volume. Coates et al also found that nebuliver
`
`output of aerosolived tohrarnyciu was greater with the ad—
`ditiouot'albutero] containinghenzalltouiurnchloride. which
`lowered the sttrface tension?"
`
`Advantages and Disadvantages ol'.let Nehulizers
`
`lists the advantages and disadvantages ofsmall-
`Table 1
`voltrrne jet nehtrlizers. Chief among the advantages is tlte
`
`Advantages and Disadvantages of Small—Volume Jet
`Table l.
`tl’nerrrnalie: Nebulixers
`
`Atlmtrtrrgca'
`'
`l‘atient lizard-breathing emirdination not required
`IEI'I'ective when used with normal tidal breathing patterns. with no
`hrealh-Itold
`Effective with low inspiralory flows or volumes
`Ability to acrosolire many drugI solutions
`Ability to acrosolize drug mixtures (3’
`| drug}. assuming suitable
`testing of drug activity
`' Drug crutccntrations cart he modified and high doses are port-tilde
`“twirlt'rrtttrtgr'x
`- Expense of equipment
`liqoipment is cumbersome attd lacks portabiiity
`Treatment tunes are lengthy and include set-up and cleaning time
`Variability in performance characteristics
`Possible patient auto-contaminalion with inadequate cleaning
`Wet. cold spray is produced. which is unpleasant with face-mask
`delivery
`Need for an external power source teleelricity Illlllllflr compressed
`gas}
`
`tidal breathing pattern used. Complex breathing
`simple.
`maneuvers or coordination of breathing with device func-
`tion is not needed. Drug delivery to the airt 'ay is over a
`period of rnirttttes. with [10—90 breaths, rather than depen-
`dent on | or 2 carefully tinted inspirations with particular
`flow rates, as with a tneteretl‘tlose inhaler (MDI) or dry
`powder inhaler tDPI). A second major advantage is the
`ability to nebulize various solutions arid to modify solution
`concentration artd thereby the amottnt of drug delivered.
`Many of the drugs currently used in the United States for
`inhalation treatment ofpulmonary disease are available as
`ttchtllizel‘ solutions. and some of these drugs, strclt as dor—
`nase alfa (Puln‘rozyme) and inhaled tobr‘atnycin (TOBI)
`are only available In solution form for nebulixation. In the
`United States the use of ncbulirers in home therapy is paid
`for by Medicare. whereas other inhaled drug formulations
`such as MDI and DP] are not. since drug therapy for
`nonhospitalized patients is not generally covered by Medi—
`care at
`this time. This provides an additional economic
`advantage to nebulizcd drug therapy for a certain sector of
`patients.
`The chiefdisadvantages of small-volume jet nebulizers
`are the size of the equipment (which requires a power
`source]. the inconvenicntly long treatment time. and the
`equipment set-up antl cleaning. In addition to the lack of
`portability and slow dosing. disposable jet nebulizers vary
`in performance among samples of the satire brand and
`among different brands. Ilollie et al showed both intmncbu—
`litter arid intcrnebulizer variability in output and respira-
`bte-range output
`(the “fine particle fraction") for the
`IJcVilbiss 6-16 jet nebtrliaer.27 Alvine et at examined 8
`disposable jet nebttlixer models. from o manufacturers,
`and fottnd that 4 of the 8 models showed visual signs of
`malfunction. which included spraying of large. visible drop-
`lets. leakirtg of solution. and air leaks that prevented ttebuli-
`ration completely. Variability of the nebtrlimtiou rate within
`a specific model ranged from 57% to |39%.3’* The studies by
`llollie el al and Alvinc et al attribttted the lack of reliability
`in performance to poor quality control by manufacturers.
`
`Design Variations in Traditional .let Nebuliaers
`
`Although the basic design ofjct nebulizcrs has remained
`the satire over the last 40—50 years. within the last It) years
`there have been modifications aimed at reducing device
`loss and exhaled toss and improving drug availability to
`the patient. These modifications are in current clinical use.
`although newer developments in nebttlization of liquids
`are also occurring. Investigational developments are not
`reviewed herein.
`
`Dennis conceptualized 3 categories ofjet nebulizer: con-
`stant-output. breatlr-enhanecd. anti dosimetrie.” Figure (1
`
`Rnsr‘rrtarortv Cara; - Novrsrsuuart 2002 Vor. 47rbtarsttriamascapied
`at: the NLM are! may be
`Subject US Copwight Laws
`
`126]
`
`WATSON LABORATORIES V- UNITED THERAPEUTICS, |PR2017—01621
`
`UNITED THERAPEUTICS, EX. 2044
`
`Page 5 of 24
`
`

`

`Dt-LsniN PRINCIPLE-IS or LIQUID Ntzntrtizn'rtoa Dcvtcns Ctrtzsr;.v‘t't.‘r IN Usr:
`
`
`Inspiration
`Expiration
`
`
`6 “ length tubing
`To patient
`
`Output
`
`4. Power gas
`CONSTANT OUTPUT
`
`Mouth Piece
`
`Output
`
`Off
`
`On
`
`Exhale
`
`\_
`—> Patient
`
`
`
`
`4-Power Gas
`
`Output
`
`
`
`
`
`
`
`BREATH ENHANCED (OPEN VFNT}
`
`
`Inspiration
`
`
`[Expiration
`
`Inspiration
`
`[Expiration
`
`DOSIMETRIC (BREATH ACTUATED)
`
`Fig. 6. Jet nebulizer designs. Aerosol generation and delivery are shown for inspiration and expiration. for each type of jet nebulizer:
`constant—output, breath—enhanced (operhvent). and dosimetric. [Partially based on an illustration lrom Reference 13.}
`
`types of nebulircr and
`shows the basic designs of the Cl
`theiraerosol production in relation to the respiratory cycle.
`In a constant-output device. aerosol
`is produced and
`released constantly. during inspiration and expiration. so
`aerosol can he lost to the environment whenever inhalation
`is not occurring, An example is the traditional JIL'I‘Illli'fCl'
`with a mouthpiece connected to a 'I‘—piece on the ncliulirer
`cttp. such as the Allegiance Air-lite Misty-Net).
`A breath-enhanced Iiebulizcr routes inhaled flow through
`the nehnlizer. using a one—way inspiratory valve. thereby
`increasing the number of aerosol particles in the inspired
`volume. Daring exhalation the inspiratory valve closes and
`aerosol is somewhat contained in the nehuliaer. although
`
`aerosol generation continues and there is exhaled loss.
`This design has been referred to as "open vent. hreatli
`assistcti."-‘ [Example IIICIlltIt: the Ventstrcam and the l’ari
`LC I’Ius.
`
`A dosimclric nehulixcr releases aerosol only during the
`inhalation cycle.” This can he achieved lay generating
`aerosol only during inspiration, using a manual interrupter
`for the power-gas [as in the l’ari LI.) or a spring-loaded
`valve (as in the Acroliclipse) or by containing all aerosol
`in the device. with no release during expiration (as with
`the Circtilaire). Classifying the Circulaire as a dosimetric
`device could he questioned because the device nebulires
`continuously, but by providing a system ol‘ aerosol storage
`
`1262
`
`rm:mntmalwssflmnitm'roav CARE 0 Novrisaasa 2002 Vol. 47 No 1]
`attire NLM and may he
`Subject USCopyi-igh'. Laws
`
`WATSON LABORATORIES v. UNITED THERAPEUTICS, IPR2017-O1621
`
`UNITED THERAPEUTICS, EX. 2044
`
`Page 6 of 24
`
`

`

`DESIGN PRINCIPLES m: LIQUID NEBULIZATION DEVICES CURRENTLY IN USP.
`
`Table '2. Aernxui Output and Dispnsiiiun I'rnm Icl Nehuiizem’I‘
`
`Rcl'crenec nr
`. nuru.
`”I‘ntul Inhaled
`Lung
`O-I‘I'htunmch
`Device
`Exhaled
`Methods
`q
`..
`
`Cnnstttnt Output
`
`Tnlnl Salute ($12]
`
`14 {5. I)
`-
`0.359.714}
`11855.45]
`
`12.4-01.5)
`3.60.6}
`—
`—
`
`1.5 (0.9]
`51'! [3.3)
`—
`i
`
`65.3 {II-(1)
`3'? [9]
`54.5 “8.21
`55.? [Ill] plus
`'11.? (5.5] in Mixer
`
`19.? [3.9)
`31.56.81
`—-
`——
`
`Ratliuiul‘tcl
`RIItIIuIIIIIc]
`Insp filler
`[nx‘p tiller
`
`2‘}
`3t]
`23
`3]
`
`15.3
`22.2
`”L3
`321! tilti)
`
`31
`CIICIII 1FdCtlr
`35-4
`59-3
`NA
`123T
`3|
`Chem tracer
`2.2.3
`55.5
`NA
`13.41
`[(1.61
`NA
`(12.3
`18.9
`Chem tracer
`3|
`
`l5.2 [mm
`NA
`NA
`—
`Spectmphut
`32
`
`[Iruttlll Enhanced
`
`
`in Vim
`
`Mitti-NL‘I‘I (II Utttill]
`Miny-Nch [7 Umin}
`Atom [(3 IJIniII]
`AcmnIMILer {t1 Uminl
`
`In Vilru
`SIlIL‘bII‘L‘JlIII. I‘urttuleh
`Hutlmn, I’IIIIIInAitle
`Cirrus. I’IIIIIIIIAitle
`Aern’I'ee tt’I IJIIIiIIl
`
`In Vivu
`
`I’ttri LC. I’uriiitly
`I’ttri1.C Slur. I’mnch Turbo
`Puri 1542 tuwu L‘mupressnrl
`It] Vilm
`
`—
`—-
`—
`
`12.1103}
`11.1 (1.?)
`17.1 [8.36)
`
`142 (13)
`13.11111)
`NA
`
`56.3 (13.”
`55.! (3.4}
`NA
`
`16.2 (2.?)
`20.0 {1.21
`NA
`
`Radiolnhc]
`Rudiohtbei
`Rildifiiilhtl
`
`33
`34
`35
`
`I’ari [.C Plus. I’urI'IItIy
`l’uri I52. I’nri Muster
`Venthtreum. CR 50
`I'uri [.L‘. Inhalt‘rlin}.I
`Venlstrettm [It ”mint
`Part LC I‘IIIN. Turhnliny
`Ventslrcnm. l‘nrlanch
`I’uri LC Slur. 'I'urlufliny
`
`31
`Chem {meet
`ItI.'I'
`6-1.]
`—
`Illl‘i
`25.2
`31
`Chem truecr
`15.9
`44.?
`—
`334+
`30.4
`3i
`Chem tracer
`4.3
`70.9
`—
`1101'
`24.3
`23
`Ittsp filter
`—
`519(14]
`—
`NA
`19112633)
`23.
`Inst: IIIIt'r
`—
`563(115)
`—
`NA
`”-2 (6.83]
`’14
`Imp litter
`—-
`——
`—
`15.51:
`22.2 (LII)
`'JII (2.4]
`8.35:
`-—
`—-
`—-
`Imp tilter
`2.1.
`
`EILIII {11.4)
`lifti
`—
`—
`—-
`Insp filler
`2-1
`
`[)usimetric
`
`
`In Vivn
`
`Cireuhtirc [7 [.IIIIIIIII
`I‘ilri I.I.. I’uri
`I‘lny
`III Vitru
`Circuiaire t (1 limit”
`AeInI-chipxe [ti Uminl
`At'rnIEL-Iirnc (II ”mini
`
`—
`—
`
`12.85.11
`II.I I411}
`
`1.1 (0.5}
`I156.”
`
`3301'
`«Imam
`
`13.99.81
`13-20.“)
`
`Rntiiuluhci
`Radiolubct
`
`3t]
`33
`
`12.8 ttl.-l)
`fitlfi
`38.? £1.34}
`
`3.210.411
`41.2
`34.” [L251
`
`—
`9.!1teuicuiuted]
`-—
`
`—
`4M tcnicuialctll
`5|.llt2.|]
`
`—-
`3-1
`(1.68.1)
`
`32
`SpecImphnt
`Filler. “I’LC Manufacturer
`Spurtruphm
`Ari el
`til.
`EIK12§
`
`In mmn' in win. Mimic: II» lung. lllN' n- u tilI 'IIIIIIcII :unuunt hnml nn the particle sin- drumming" Ur th‘ 3mm"
`'Ji't Ik'hulin'n i-. l NIL-gqu [UII‘ILI'I‘ uulpIIl. I‘IIL'ilIII enlmm‘ml .‘uIII IIuhiIIK‘IIiC.
`
`I‘INIHCL‘II a-nl tin:
`llll'|I inlmlul man. 1|..- |Ill‘lIIlilI\ rnlumn IlilIIl'jl‘C'I Ilu- pmmry “than 01' :lIug enllcn-Iinn :Iml Int-autumn
`.
`chemical Irm'cr. Sven-null“ — LpL-u-tnIIIEuunIlerir. ILInliuluhei —‘ rmlinl.tht.'|I.-Il.
`I! I' — IntgllI.IIyIIgI-:Il. Imp liltcr --
`|!1\|'|lFiI|U')t I'llrt'r L'I’II‘1'II!‘I'I.(.III'III"all?! -=
`”lid“ ’ 5 put
`tMms I'. 0.1 “III
`II!‘1.(' : tug-II p\'l[0ITII.IIIL\’ IIquid (Itnltmlllymley
`§UII|||IIII|\11ct'. il:.t;|. Ali l'l .II. Iii-tight .‘ilatt' Illnu'nity. 1m}.
`
`
`
`between breaths. aerosol is released nuly during iIIIIuIII-
`titan, which is the defining feature of :1 [Insimctric device.13
`Table 2 summarizes studies of aerosol output and dispo-
`sititm t‘or constant-output. open-vent breath-enhanced. :uItl
`dusimctric IchtIIimI-s.
`
`Cunstant-Oulput Designs.
`Most dispusuble Ismail-v01-
`umcjcl nebulizcrs typify this functional category. Figure 7
`sltuws data from :1 study by Lewis and Fleming on drug
`disposition with :1 ConsttInt-uutput
`IICbLIIIZCI‘, along with
`cmnpnrisun data from an NIDI.2"'-"‘ That study found that
`
`RESPIRATORY CARI-2 o NOVIEMIlE-ZR 2002 VOL 4Wuafiialmsmiad
`51th: NLM ard may I}!
`Subject US Copyright Laws
`
`1263
`
`WATSON LABORATORIES V- UNITED THERAPEUTICS, |PR2017—01621
`
`UNITED THERAPEUTICS, EX. 2044
`
`Page 7 of 24
`
`
`l"_..,_._.-._._i..__._.___._....
`
`,.._..__
`
`

`

`0 Exhaled
`Storage Bag
`
`B
`
`Patient
`i
`
`Arousal
`
`Exhalatlon port
`IL.__
`—I- Patient
`
`T-plece
`——-—%*
`Patient 4—-
`..
`1-wayI I“! I
`[Inspiramryi
`
`A
`
`Mlzer
`
`{expiratory}
`1-way
`
`Hemispherical
`Deflector
`
`
`
`
`Acorn
`"‘ Nebulizer
`
`4- Power Gas
`
`DESIGN PRINCIPLES or LIQUtn Ntsnurtza’rton Dmcas CURRENTLY IN Usts
`
`100
`
`90
`
`__— 80
`5 " length tuhing
`To patient +
`
`T0
`
`80
`
`50
`
`40
`
`30
`
`20'
`
`1 0
`
`Capillary
`louder
`
`tube
`
`Drug
`- solution
`
`Power gas
`[flow meter,
`
`comprnssor]
`
`Fig. T. Disposition of total drug dose with a constant-output jet nebulizer (MiniNeb) versus a metered-dose inhaler {MIDI}. O-P = oropha‘
`ryngeal. SUN = Small-VOIUI’TIB nebutizer. {Modified from Relerence 29.)
`
`SVN
`
`MD]
`
`Negative
`pressure —
`area
`
`
`
`
`... Power Gas
`
`4- Power Gas
`
`Fig. 8‘ Three examples of constant—output nebulizers. A. Acorn with Mizer reservoir chamber attached. B. Slde
`air entrainment by nebulizer venturi effect. C. AeroTee with aerosol storage bag.
`
`stream with open vent for
`
`12% ofthc nominal dost: reached the lungs. with 2% to the
`mouth and stomach. 66% left in the apparatus. and ’lU‘i’r-
`exhaled?" Combining tlte I296 delivered to the lungs with
`the 2% lost to the (‘irUplltIl‘yilx and stomach gives a total
`inhaled amount of 14% of the total dose. The dcad~vo|ttme
`
`loss of 66% agrees well with the data ol‘ Kratljan and
`Lakshtninarayatt. who found that -’l l 416% of the total dose
`stayed in the nebulizcl' untler simulated clinical condi—
`lions}K Constant-output ttcbttlizcrs have been criticized as
`
`more inefficient than other aerosol delivery tncthods such
`as the MDl. because of the amount of drug lull
`in the
`
`apparatus and exhaled. The traditional jcl ttchttlizcr is I10
`more incl‘l‘icicut itt delivery of aerosol drug to the airways
`tltatt other lratlitional aerosol devices. The patlcrn of loss.
`however. tlil'l’ers among jct nchuliaers. MDls. and Dl‘ls.
`The sizable traction lost to the ttcbulirer apparatus is bal-
`anced by the (10—80% ot‘opharyngual and stomach thPU'
`sition with an MDI or the apparatus and oropharyngeal
`deposition with a I)l’l.-“"--” A convincing study by Zainu-
`tlin et al showed that only about
`IU% of aerosolixcd albu-
`tcrol reaches the lung. regardless of whether a nebulizcr.
`MDI. or DH is ttst:tJ.-‘T though that finding might not apply
`
`[264
`
`to;materratwasl-‘ttaaatta'rortv Came c NOVEMBER 2002 V01- 47 No II
`atthe NLM arcl maybe
`Subject L'S Copyright Laws
`
`WATSON LABORATORIES v. UNITED THERAPEUTICS, |PR201?—01621
`
`UNITED THERAPEUTICS, EX. 2044
`
`Page 8 of 24
`
`

`

`DizsioN Prostitutes or: LIQUID Niiituumriou Devices CURRENTLY IN USE
`
`Inspiration
`
`Exhale
`Valve
`
`1-way
`valve
`
`Mouth Piece \
`
`
`
`Power Gas -D-
`
`
`
`1-way
`(expiratory)
`
`1-way
`(inhaled)
`
`Patientw- T-pteee
`
`
`
`
`Fig. 9. Two examples of breath-enhanced. open-vent nebulizers. Left: Par] LC Plus. Right: Ventstream.
`
`to newer aerosol devices. some of which are discussed
`elsewhere in the conference proceedings in this and the
`
`following issue of RF—ZSI’iItATORt‘ Cain-z.
`The loss ol‘ aerosol from the 'I‘-piece of a co

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket